(170 days)
The ARCHITECT Syphilis TP assay is a chemiluminescent microparticle immunoassay (CMIA) for the qualitative detection of antibodies (IgG and IgM) directed against Treponema pallidum (TP) in human serum and plasma. The ARCHITECT Syphilis TP assay is intended to be used as an initial diagnostic test or in conjunction with a nontreponemal laboratory test and clinical findings to aid in the diagnosis of syphilis infection.
Warning: The ARCHITECT Syphilis TP assay is not intended for use in screening blood, plasma, or tissue donors. The effectiveness of the ARCHITECT Syphilis TP assay for use in screening blood, plasma, or tissue donors has not been established.
The ARCHITECT Syphilis TP Calibrator is for the ARCHITECT iSystem when used for the qualitative detection of antibody to Treponema pallidum (TP) in human serum and plasma.
The ARCHITECT Syphilis TP Controls are for the estimation of test precision of systematic analytical deviations of the ARCHITECT iSystem when used for the qualitative detection of antibody to Treponema pallidum (TP) in human serum and plasma.
The ARCHITECT Syphilis TP assay is a two-step immunoassay for the qualitative detection of antibodies (IgG and IgM) directed against TP in human serum or plasma using CMIA technology with flexible assay protocols, referred to as Chemiflex.
- Sample, assay diluent, and recombinant TP antigen (TpN15, TpN17 and TpN47) coated microparticles are combined. Anti-TP antibodies present in the sample bind to the TP coated microparticles.
- After washing, anti-human IgG and IgM acridinium-labeled conjugate is added to create a reaction mixture.
- Following another wash cycle, Pre-Trigger and Trigger Solutions are added to the reaction mixture.
- The resulting chemiluminescent reaction is measured as relative light units (RLUs). There is a direct relationship between the amount of anti-TP antibodies in the sample and the RLUs detected by the ARCHITECT iSystem optics.
The presence or absence of anti-TP antibodies in the specimen is determined by comparing the chemiluminescent signal in the reaction to the cutoff signal determined from an active calibration. If the chemiluminescent signal in the reaction is greater than or equal to the cutoff signal, the specimen is considered reactive for anti-TP antibodies.
Here's a breakdown of the acceptance criteria and study information for the ARCHITECT Syphilis TP Reagent Kit:
1. Acceptance Criteria and Reported Device Performance
The document does not explicitly state pre-defined acceptance criteria for positive and negative percent agreement. Instead, it presents the calculated percent agreements observed in the clinical study. The FDA's substantial equivalence decision implies that these observed performance metrics were deemed acceptable when compared to the predicate device.
| Performance Metric | Acceptance Criteria (Implied by FDA Acceptance) | Reported Device Performance (ARCHITECT Syphilis TP Assay) |
|---|---|---|
| Overall Positive Percent Agreement (Prospectively-Collected Intended Use Population) | Expected high agreement with composite reference for positive cases. | 96.2% (153/159) with 95% CI: 92.0% to 98.3% |
| Overall Negative Percent Agreement (Prospectively-Collected Intended Use Population) | Expected high agreement with composite reference for negative cases. | 99.0% (976/986) with 95% CI: 98.1% to 99.4% |
| Positive Percent Agreement (Routine Syphilis Category) | Expected high agreement for this subpopulation. | 97.3% (36/37) with 95% CI: 86.2-99.5% |
| Negative Percent Agreement (Routine Syphilis Category) | Expected high agreement for this subpopulation. | 99.5% (403/405) with 95% CI: 98.2-99.9% |
| Positive Percent Agreement (HIV Positive Category) | Expected high agreement for this subpopulation. | 95.9% (117/122) with 95% CI: 90.8-98.2% |
| Negative Percent Agreement (HIV Positive Category) | Expected high agreement for this subpopulation. | 97.5% (270/277) with 95% CI: 94.9-98.8% |
| Positive Percent Agreement (Pre-Selected Positive Specimens) | Expected high agreement for this enriched population. | 98.9% (376/380) with 95% CI: 97.3% to 99.6% |
| Negative Percent Agreement (Pre-Selected Positive Specimens) | Expected high agreement for this enriched population. | 92.3% (24/26) with 95% CI: 75.9% to 97.9% |
| Positive Percent Agreement (Presumed Positive Category) | Expected high agreement for this subpopulation. | 98.9% (356/360) with 95% CI: 97.2-99.6% |
| Negative Percent Agreement (Presumed Positive Category) | Expected high agreement for this subpopulation. | 92.3% (24/26) with 95% CI: 75.9-97.9% |
| Positive Percent Agreement (Reactive Pregnant Category) | Expected 100% agreement for this subpopulation. | 100.0% (20/20) with 95% CI: 83.9-100.0% |
| Negative Percent Agreement (Prospectively-Collected Pregnant Females) | Expected high agreement for this subpopulation. | 99.7% (303/304) with 95% CI: 98.2-99.9% |
Note: NA (not applicable) for some categories indicates that there were no positive or negative cases, respectively, in that specific subpopulation to calculate the agreement.
The document also describes several non-clinical performance acceptance criteria through studies such as:
- Precision: Evaluated within-run, within-laboratory, within-day, between-site, and between-lot variability. Specific SD and %CV values are reported for controls and panels. The implication is that these values demonstrate acceptable precision.
- Interference: Demonstrated that the assay is not susceptible to interference from various substances at specified levels (e.g., conjugated bilirubin ≤ 20 mg/dL, hemoglobin ≤ 500 mg/dL).
- Analytical Specificity: Evaluation against various medical conditions and disease states. While some cross-reactivity was noted for a few specimens in specific categories (e.g., CMV IgG, HIV), the overall performance and context are considered acceptable, with a specific limitation noted for yaws, pinta, and bejel.
- Tube Type Matrix Comparison: Acceptable mean/median S/CO differences and distribution of S/CO differences for different tube types compared to serum.
- Sample On-Board Stability: Up to 3 hours storage on the ARCHITECT i2000SR System.
- Sample Stability: Up to 7 days at 2-8°C, up to 72 hours at room temperature, up to 6 freeze/thaw cycles, and up to 30 days frozen (-10°C or colder).
- High Dose Hook Effect: Determined not to be susceptible.
- Within-Assay Sample Carryover: Determined not to be susceptible.
2. Sample Sizes Used for the Test Set and Data Provenance
Test Set Sample Sizes:
- Clinical Study Total Specimens: 2220
- Prospectively-Collected Intended Use Population: 1145 specimens
- Routine Syphilis Testing: 442
- Pregnant Females: 304
- HIV Positive: 399
- Pre-Selected Positive for TP Antibodies: 406 specimens
- Apparently Healthy Individuals: 480 specimens
- Medically Diagnosed (Primary, Secondary, Latent Syphilis): 179 specimens
- Spiked Pregnant Female Specimens: 10 specimens
- Prospectively-Collected Intended Use Population: 1145 specimens
Data Provenance:
- Country of Origin: United States. Specimens for the prospectively-collected intended use population were sourced or collected from various locations: Baltimore, Maryland; Colton, California; Fort Lauderdale, Florida; Hyannis, Massachusetts; Los Angeles, California; Miami, Florida; San Antonio, Texas; and Temple, Texas.
- Retrospective/Prospective:
- Prospectively-collected: 1145 specimens from the intended use population were collected prospectively.
- Pre-selected: 406 specimens were "pre-selected positive" based on previous laboratory testing, suggesting a retrospective selection criterion based on existing positive results.
- Medically diagnosed and Apparently healthy: These populations represent specific cohorts selected for the study.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
This document describes a diagnostic device for syphilis, not an imaging diagnostic AI. Therefore, the "ground truth" is a laboratory-based composite reference, not expert-read interpretations of images.
The ground truth for the clinical studies was established using comparator assays, not human experts interpreting data for a "ground truth" consensus in the way a radiologist would for an imaging study. The comparator results were determined using a "2 out of 3 rule" from the following assays:
- Treponemal Chemiluminescent Immunoassay (TP-CLIA)
- Rapid Plasma Reagin (RPR) - a nontreponemal assay
- Treponema Pallidum Particle Agglutination (TP-PA) - a second treponemal assay
For the "medically diagnosed" category, the diagnosis was made by a licensed physician based on clinical information and serological testing results (e.g., VDRL, RPR, TP-PA). While a physician's input is a form of expert opinion, it's not described as a multi-expert consensus process specifically for establishing ground truth for the algorithm's direct input.
4. Adjudication Method for the Test Set
The adjudication method for the clinical test set (specifically for the final comparator result) was a "2 out of 3 rule" involving the three comparator assays: TP-CLIA, RPR, and TP-PA. If at least two of these assays were concordant (e.g., both positive or both negative), that determined the final comparator result.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This device is an in-vitro diagnostic assay, not an AI-powered imaging or interpretation system designed to assist human readers. Therefore, the concept of "human readers improving with AI vs without AI assistance" does not apply here.
6. Standalone Performance Study
Yes, a standalone performance study was done. The entire document describes the performance of the ARCHITECT Syphilis TP assay as a standalone diagnostic device. Its performance metrics (positive percent agreement, negative percent agreement, precision, specificity, etc.) are all measures of the algorithm's (assay's) performance without human-in-the-loop intervention for result interpretation. The assay outputs a qualitative "Reactive" or "Nonreactive" result based on its internal cutoff calculation (S/CO).
7. Type of Ground Truth Used
The primary type of ground truth used was a composite reference standard derived from the results of three established laboratory assays: a treponemal chemiluminescent immunoassay (TP-CLIA), a nontreponemal assay (Rapid Plasma Reagin [RPR]), and a second treponemal assay (Treponema Pallidum Particle Agglutination [TP-PA]), adjudicated by a "2 out of 3 rule."
Additionally, for some categories, clinical context and medical diagnosis by a licensed physician were used to characterize the patient population (e.g., "medically diagnosed individuals with primary, secondary, or latent syphilis").
8. Sample Size for the Training Set
The document does not explicitly state a separate "training set" sample size for the ARCHITECT Syphilis TP assay. For in-vitro diagnostic devices like this, the "development" or "training" of the assay (e.g., selection of antigens, optimization of reagents, determination of cutoff) is often an iterative process using internal studies and samples, rather than a distinct, large "training set" in the machine learning sense.
The "internal study" for selecting and validating the cutoff is mentioned, indicating some data was used for this purpose, but the size is not specified. The clinical studies described are primarily for validation and performance assessment (akin to a test set).
9. How the Ground Truth for the Training Set was Established
Since a distinct "training set" in the machine learning context is not explicitly described, the method for establishing its ground truth is also not detailed. However, based on the general development of such assays, the "ground truth" for optimizing the assay's components and cutoff during development would likely involve:
- Known positive and negative samples: Samples from individuals with confirmed syphilis infections (e.g., by culture, Western blot, or a panel of established treponemal and non-treponemal tests) and samples from healthy individuals.
- Reference standards: Internal or international reference standards for Treponema pallidum antibodies.
The "internal study" for cutoff selection and validation would have used such characterized samples to ensure the assay's sensitivity and specificity met developmental targets before moving to large-scale clinical validation.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a seal for the Department of Health & Human Services USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. In the center of the seal is a stylized image of three human profiles facing to the right, stacked on top of each other.
June 15, 2016
Food and Drug Administration
Public Health Service
10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Abbott Laboratories Diagnostic Division Laura Yelvigi Joglekar, PharmD Regulatory Affairs Project Manager Dept. 9AA, Bldg. AP8-1, 100 Abbott Park Road Abbott Park, IL 60064-6038
Re: K153730
Trade/Device Name: ARCHITECT Syphilis TP Reagent Kit, ARCHITECT Syphilis TP Calibrator, ARCHITECT Syphilis TP Controls Regulation Number: 21 CFR 866.3830 Regulation Name: Treponema pallidum treponemal test reagents Regulatory Class: Class II Product Code: LIP, JIT, JJX Dated: May 20, 2016 Received: May 23, 2016
Dear Dr. Joglekar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{1}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Steven R. Gitterman -S
Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K153730
Device Name ARCHITECT Syphilis TP Reagent Kit ARCHITECT Syphilis TP Calibrator ARCHITECT Syphilis TP Controls
Indications for Use (Describe)
The ARCHITECT Syphilis TP assay is a chemiluminescent microparticle immunoassay (CMA) for the qualitative detection of antibodies (IgG and IgM) directed against Treponema pallidum (TP) in human serum and plasma. The ARCHITECT Syphilis TP assay is intended to be used as an initial diagnostic test or in conjunction with a nontreponemal laboratory test and clinical findings to aid in the diagnosis of syphilis infection.
Warning: The ARCHITECT Syphilis TP assay is not intended for use in screening blood, plasma, or tissue donors. The effectiveness of the ARCHITECT Syphilis TP assay for use in screening blood, plasma, or tissue donors has not been established.
The ARCHITECT Syphilis TP Calibrator is for the ARCHITECT iSystem when used for the qualitative detection of antibody to Treponema pallidum (TP) in human serum and plasma.
The ARCHITECT Syphilis TP Controls are for the estimation of test precision of systematic analytical deviations of the ARCHITECT iSystem when used for the qualitative detection of antibody to Treponema pallidum (TP) in human serum and plasma.
| Type of Use (Select one or both, as applicable) |
|---|
| Prescription Use (Part 21 CFR 801 Subpart D) |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Section 5: 510(k) Summary
This summary of the 510(k) is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
I. Applicant Name
Laura Yelvigi Joglekar, PharmD, ADD, Regulatory Affairs Project Manager Abbott Laboratories Diagnostics Division Dept. 9AA, AP8-1 100 Abbott Park Road Abbott Park, IL 60064-6095 Telephone Number: (224) 667-3503 Fax Number: (224) 667-4836 E-Mail: laura.joglekar@abbott.com Date Summary prepared: June 9, 2016
Grace LeMieux, ADD, Regulatory Affairs Director Telephone Number: (224) 668-0409 Fax Number: (224) 667-4836 E-Mail: grace.lemieux@abbott.com
II. Device Name
ARCHITECT Syphilis TP (8D06)
Reagents
Trade Name: ARCHITECT Syphilis TP Reagent Kit Device Classification: Class II Classification Name: Treponema pallidum treponemal test reagents Governing Regulation: 21 CFR § 866.3830 Code: LIP
Calibrator
Trade Name: ARCHITECT Syphilis TP Calibrator Device Classification: Class II Classification Name: Calibrator Governing Regulation: 862.1150 Code: JIT
{4}------------------------------------------------
Controls
Trade Name: ARCHITECT Syphilis TP Controls Device Classification: Class I Classification Name: Control Governing Regulation: 862.1660 Code: JJX
III. Predicate Device
DiaSorin LIAISON Treponema assay (K061247)
IV. Description of Device
Reagents
The ARCHITECT Syphilis TP reagent kit contains:
- . Microparticles: (1 bottle x 6.6 mL per 100-test / 1 bottle x 27.0 mL per 500-test) TP (E.coli, recombinant) antigen coated microparticles in HEPES buffer with detergent. Minimum concentration: 0.08% solids. Preservatives: sodium azide and other antimicrobial agents.
- . Conjugate: (1 bottle x 5.9 mL per 100-test / 1 bottle x 26.3 mL per 500-test). Murine anti-IgG/anti-IgM acridinium-labeled conjugate in MES buffer with protein (bovine) stabilizer. Minimum concentration: (anti-IgG) 26.6 ng/mL / (anti-IgM) 1.34 ng/mL. Preservatives: sodium azide and other antimicrobial agents.
- . Assay Diluent: (1 bottle x 10.0 mL per 100-test / 1 bottle x 52.5 mL per 500-test). Syphilis TP Assay Diluent containing MES buffer with detergent. Preservatives: ProClin 950 and other antimicrobial agents.
Calibrator
The ARCHITECT Syphilis TP Calibrator contains:
- 1 Bottle (4.0 mL) of ARCHITECT Syphilis TP Calibrator. . ARCHITECT Syphilis TP Calibrator is prepared in recalcified human plasma (inactivated); reactive for anti-TP. Preservatives: sodium azide and other antimicrobial agents.
The calibrator is referenced to an Abbott internal reference standard. This internal reference standard is manufactured by dilution of high titer positive plasma with HEPES buffer with nonreactive human plasma.
{5}------------------------------------------------
Controls
The ARCHITECT Syphilis TP Controls contain:
- . 1 Bottle (8.0 mL) Negative Control prepared in recalcified human plasma. Preservatives: sodium azide and other antimicrobial agents.
- . 1 Bottle (8.0 mL) Positive Control prepared in recalcified human plasma. The Positive Control (inactivated) is reactive for anti-TP. Preservatives: sodium azide and other antimicrobial agents.
The controls are at the following proposed target syphilis TP concentrations and ranges:
| Control | Control Range S/CO |
|---|---|
| Negative Control (Control –) | $\leq$ 0.40 |
| Positive Control (Control +) | 1.25 - 3.75 |
Biological Principles of the Procedure
The ARCHITECT Syphilis TP assay is a two-step immunoassay for the qualitative detection of antibodies (IgG and IgM) directed against TP in human serum or plasma using CMIA technology with flexible assay protocols, referred to as Chemiflex.
-
- Sample, assay diluent, and recombinant TP antigen (TpN15, TpN17 and TpN47) coated microparticles are combined. Anti-TP antibodies present in the sample bind to the TP coated microparticles.
-
- After washing, anti-human IgG and IgM acridinium-labeled conjugate is added to create a reaction mixture.
-
- Following another wash cycle, Pre-Trigger and Trigger Solutions are added to the reaction mixture.
-
- The resulting chemiluminescent reaction is measured as relative light units (RLUs). There is a direct relationship between the amount of anti-TP antibodies in the sample and the RLUs detected by the ARCHITECT iSystem optics.
The presence or absence of anti-TP antibodies in the specimen is determined by comparing the chemiluminescent signal in the reaction to the cutoff signal determined from an active calibration. If the chemiluminescent signal in the reaction is greater than or equal to the cutoff signal, the specimen is considered reactive for anti-TP antibodies.
{6}------------------------------------------------
V. Intended Use of the Device
The ARCHITECT Syphilis TP assay is a chemiluminescent microparticle immunoassay (CMIA) for the qualitative detection of antibodies (IgG and IgM) directed against Treponema pallidum (TP) in human serum and plasma. The ARCHITECT Syphilis TP assay is intended to be used as an initial diagnostic test or in conjunction with a nontreponemal laboratory test and clinical findings to aid in the diagnosis of syphilis infection.
Warning: The ARCHITECT Syphilis TP assay is not intended for use in screening blood, plasma, or tissue donors. The effectiveness of the ARCHITECT Syphilis TP assay for use in screening blood, plasma, or tissue donors has not been established.
VI. Comparison of Technological Characteristics
The ARCHITECT Syphilis TP assay (candidate assay) utilizes a chemiluminescent microparticle immunoassay (CMIA) methodology for the quantitative in vitro determination of syphilis TP and is intended for use on the ARCHITECT iSystem.
The similarities and differences between the candidate assay and the predicate assay are presented in the tables starting on page 5-5.
{7}------------------------------------------------
| Similarities | ||
|---|---|---|
| Characteristics | Candidate DeviceARCHITECT Syphilis TP (List No. 8D06) | Predicate DeviceDiaSorin LIAISON† Treponema Assay (K061247) |
| Intended Use andIndications for Use | The ARCHITECT Syphilis TP assay is achemiluminescent microparticle immunoassay(CMIA) for the qualitative detection of antibodies(IgG and IgM) directed against Treponema pallidum(TP) in human serum and plasma. The ARCHITECTSyphilis TP assay is intended to be used as an initialdiagnostic test or in conjunction with anontreponemal laboratory test and clinical findings toaid in the diagnosis of syphilis infection. | The LIAISON Treponema Assay useschemiluminescent immunoassay (CLIA) technologyon the LIAISON Analyzer family for the qualitativedetermination of total antibodies directed againstTreponema pallidum in human serum. The presenceof antibodies to Treponema pallidum specificantigen, in conjunction with non-treponemallaboratory tests and clinical findings may aid in thediagnosis of syphilis infection. |
| Methodology | Chemiluminescent microparticle immunoassay(CMIA) | Sandwich chemiluminescence immunoassay (CLIA) |
| Cut-off Index | 1.00 S/CO | Index value of 1.0 |
| Standardization | The calibrator is referenced to an internal referencestandard. | The calibrator concentrations are referenced to anin-house antibody preparation. |
| Controls | 2 (Negative and Positive) | 2 (Negative and Positive) |
{8}------------------------------------------------
| Differences | ||||
|---|---|---|---|---|
| Characteristics | Candidate DeviceARCHITECT Syphilis TP (List No. 8D06) | Predicate DeviceDiaSorin LIAISON Treponema Assay (K061247) | ||
| Antigen Used | recombinant TP antigens: TpN15, TpN17 and TpN47(obtained in E.coli) | DNA-Tp17 Recombinant antigen(obtained in E. coli) | ||
| Platform | ARCHITECT i2000SR System | DiaSorin LIAISON | ||
| Assay Protocol | 2-step | 1-step | ||
| Specimen Type | Human serum or plasma | Human serum | ||
| Equivocal Zone | No gray zone | Index value of 0.9 - 1.1 | ||
| Components | Microparticles – TP (E.coli, recombinant) antigencoated microparticles in HEPES buffer withdetergent. Minimum concentration: 0.08% solids.Preservatives: sodium azide and other antimicrobialagents.Conjugate - Murine anti-IgG/anti-IgM acridinium-labeled conjugate in MES buffer with protein(bovine) stabilizer. Minimum concentration:(anti-IgG) 26.6 ng/mL / (anti-IgM) 1.34 ng/mL.Preservatives: sodium azide and other antimicrobialagents.Assay Diluent - Syphilis TP Assay Diluentcontaining MES buffer with detergent. Preservatives:ProClin 950 and other antimicrobial agents. | Magnetic Particles - Magnetic particles coated withTp17 DNA recombinant protein (obtained in E. coli),BSA, phosphate buffer. Preservative: 0.099% SodiumAzide.Conjugate – Tp17 DNA recombinant protein(obtained in E. coli) conjugated to an isoluminolderivative, BSA, phosphate buffer. Preservatives:ProClin 300, Gentamycin Sulfate.Specimen Diluent - EDTA, phosphate buffer,bacterial proteins, an inert blue dye. Preservative:ProClin 300. |
{9}------------------------------------------------
| Differences | ||
|---|---|---|
| Characteristics | Candidate DeviceARCHITECT Syphilis TP (List No. 8D06) | Predicate DeviceDiaSorin LIAISON Treponema Assay (K061247) |
| Tube Types | Human serum:• Serum• Serum separatorHuman plasma:• Dipotassium EDTA• Tripotassium EDTA• Lithium heparin plasma separator• Lithium heparin• Sodium heparin | Serum |
| Calibrators | 1 Positive Calibrator | 2 Positive Calibrators |
| Calibration Storage | Maximum of 30 days | Maximum of 2 weeks |
{10}------------------------------------------------
VII. Summary of Nonclinical Performance
Assay Cutoff
An internal study and an external clinical study were performed in order to select and validate the appropriate cutoff multiplier for the ARCHITECT Syphilis TP assay. The cutoff calculation is as follows:
Cutoff (CO) = Calibrator Mean RLU* x 0.20
S/CO = Sample RLU / Cutoff RLU
Results < 1.00 S/CO are reported as nonreactive. Results ≥ 1.00 are reported as reactive.
Precision
Within-Laboratory Precision
A study was performed based on guidance from Clinical and Laboratory Standards Institute (CLSI) document EP05-A2. Testing was conducted using 3 lots of ARCHITECT Syphilis TP Reagents, 2 lots of ARCHITECT Syphilis TP Calibrator, and 1 lot of ARCHITECT Syphilis TP Controls, and 2 instruments. Two controls and 4 serum panels were assayed in a minimum of 2 replicates (target of 3 replicates) at 2 separate times per day on 22 different days.
| Sample | Instrument | Lot | N | MeanS/CO | Within-Run | Within-Laboratorya(Total) | ||
|---|---|---|---|---|---|---|---|---|
| NegativeControl | 1 | 1 | 130 | 0.04 | 0.002 | 4.3 | 0.002 | 4.9 |
| 2 | 131 | 0.04 | 0.004 | 11.1 | 0.005 | 13.9 | ||
| 3 | 131 | 0.04 | 0.004 | 11.7 | 0.005 | 13.8 | ||
| 2 | 1 | 130 | 0.04 | 0.002 | 5.4 | 0.002 | 5.4 | |
| 2 | 130 | 0.04 | 0.005 | 14.3 | 0.005 | 14.7 | ||
| 3 | 132 | 0.03 | 0.004 | 13.1 | 0.004 | 13.6 | ||
| PositiveControl | 1 | 1 | 132 | 2.71 | 0.035 | 1.3 | 0.053 | 2.0 |
| 2 | 132 | 2.68 | 0.034 | 1.3 | 0.058 | 2.2 | ||
| 3 | 131 | 2.70 | 0.034 | 1.3 | 0.056 | 2.1 | ||
| 2 | 1 | 132 | 2.66 | 0.048 | 1.8 | 0.060 | 2.3 | |
| 2 | 131 | 2.65 | 0.066 | 2.5 | 0.094 | 3.6 | ||
| 3 | 131 | 2.66 | 0.048 | 1.8 | 0.069 | 2.6 |
* RLU = Relative Light Units
{11}------------------------------------------------
| Sample | Instrument | Lot | N | MeanS/CO | Within-Run | Within-Laboratorya(Total) | ||
|---|---|---|---|---|---|---|---|---|
| NonreactivePanel | 1 | 1 | 132 | 0.08 | 0.004 | 4.3 | 0.005 | 5.6 |
| 1 | 2 | 132 | 0.08 | 0.004 | 5.2 | 0.004 | 5.5 | |
| 1 | 3 | 132 | 0.08 | 0.004 | 5.3 | 0.005 | 6.9 | |
| 2 | 1 | 131 | 0.08 | 0.003 | 4.4 | 0.004 | 4.7 | |
| 2 | 2 | 132 | 0.08 | 0.005 | 6.3 | 0.005 | 7.0 | |
| 2 | 3 | 131 | 0.07 | 0.004 | 5.8 | 0.004 | 6.1 | |
| HighNonreactivePanel | 1 | 1 | 132 | 0.57 | 0.010 | 1.8 | 0.013 | 2.3 |
| 1 | 2 | 132 | 0.56 | 0.010 | 1.8 | 0.013 | 2.3 | |
| 1 | 3 | 132 | 0.56 | 0.010 | 1.9 | 0.014 | 2.4 | |
| 2 | 1 | 132 | 0.54 | 0.012 | 2.2 | 0.013 | 2.4 | |
| 2 | 2 | 131 | 0.54 | 0.013 | 2.4 | 0.017 | 3.2 | |
| 2 | 3 | 131 | 0.53 | 0.013 | 2.4 | 0.015 | 2.8 | |
| Low ReactivePanel | 1 | 1 | 132 | 1.28 | 0.017 | 1.3 | 0.023 | 1.8 |
| 1 | 2 | 132 | 1.27 | 0.018 | 1.5 | 0.026 | 2.1 | |
| 1 | 3 | 131 | 1.26 | 0.020 | 1.6 | 0.026 | 2.1 | |
| 2 | 1 | 130 | 1.24 | 0.025 | 2.0 | 0.030 | 2.4 | |
| 2 | 2 | 132 | 1.23 | 0.024 | 2.0 | 0.031 | 2.5 | |
| 2 | 3 | 132 | 1.22 | 0.024 | 2.0 | 0.031 | 2.5 | |
| High ReactivePanel | 1 | 1 | 132 | 3.62 | 0.056 | 1.5 | 0.063 | 1.7 |
| 1 | 2 | 132 | 3.60 | 0.052 | 1.5 | 0.063 | 1.8 | |
| 1 | 3 | 132 | 3.59 | 0.043 | 1.2 | 0.059 | 1.6 | |
| 2 | 1 | 132 | 3.55 | 0.065 | 1.8 | 0.079 | 2.2 | |
| 2 | 2 | 132 | 3.53 | 0.146 | 4.1 | 0.157 | 4.4 | |
| 2 | 3 | 132 | 3.54 | 0.064 | 1.8 | 0.082 | 2.3 |
4 If no result was obtained for one of the target three replicates for a panel or control within a run, the lost replicate was not retested.
Interference
A study was performed based on guidance from CLSI EP07-A2. Potentially interfering substances were evaluated to determine whether S/CO values were affected when using the ARCHITECT Syphilis TP assay. Samples containing the potential interferents were prepared at 2 levels of syphilis (approximately 0.80 S/CO and 1.20 S/CO). The samples were assayed, and the S/CO values of the spiked samples were compared to the reference samples. The ARCHITECT Syphilis TP assay is not susceptible to interference effects from the following interferents at the interferent levels listed in the table below.
| Interferent | Interferent Level |
|---|---|
| Conjugated Bilirubin | $\leq$ 20 mg/dL |
| Unconjugated Bilirubin | $\leq$ 20 mg/dL |
| Cholesterol | $\leq$ 500 mg/dL |
{12}------------------------------------------------
| Interferent | Interferent Level |
|---|---|
| Gamma Globulin | $\leq$ 6 g/dL |
| Hemoglobin | $\leq$ 500 mg/dL |
| Triglycerides | $\leq$ 3000 mg/dL |
| Total Protein | $\leq$ 12 g/dL |
Analytical Specificity
The ARCHITECT Syphilis TP assay was evaluated for potential cross-reactivity from specimens from individuals with medical conditions and other disease states unrelated to a syphilis infection.
Six specimens reactive in ARCHITECT Syphilis TP and a treponemal chemiluminescent immunoassay (TP-CLIA) were also either positive (marked by 4 in table below) or indeterminate (marked by " in table below) by confirmation testing. Confirmed reactive results were not unexpected because the specimens had been obtained from vendors based on the required disease condition and documentation of prior syphilis laboratory test results was not provided by the vendor.
One of the 10 CMV IgG specimens, 1 of the 10 HIV specimens, 1 of the 7 monoclonal hyper IgG, and 1 of the 10 anti-E. coli specimens (marked by " in table below) were reactive in ARCHITECT Syphilis TP and were not confirmed with other treponemal or nontreponemal tests applied in this study.
One of the 11 gonorrhea specimens and 2 of the 6 HTLV-II specimens (marked by vin table below) were reactive in ARCHITECT Syphilis TP but could not be tested by the other tests (TP-CLIA, RPR, TP-PA, Fluorescent Treponemal Antibody Absorption [FTA-ABS]) because specimens from these disease states could only be sourced as plasma specimens.
| Clinical Category | N | Number of ARCHITECT Syphilis TPReactive Results |
|---|---|---|
| Chlamydia | 15 | 1a |
| Cytomegalovirus (CMV) IgG | 10 | 1b |
| Cytomegalovirus (CMV) IgM | 10 | 1a |
| Anti-dsDNA Autoantibodies | 3 | 0 |
| Epstein-Barr Virus (EBV) IgG | 10 | 1a |
| Epstein-Barr Virus (EBV) IgM | 24 | 0 |
| Anti-Escherichia coli (E. coli) | 10 | 1b |
{13}------------------------------------------------
| Clinical Category | N | Number of ARCHITECT Syphilis TPReactive Results |
|---|---|---|
| Gonorrhea | 11 | 1c |
| HAVAB IgG | 10 | 0 |
| HBc IgM | 4 | 0 |
| Hemodialysis | 10 | 0 |
| Hepatitis A Virus (HAV) | 10 | 1a |
| Hepatitis B Virus (HBV) | 10 | 0 |
| Hepatitis C Virus (HCV) | 10 | 1a |
| Herpes Simplex Virus (HSV) | 10 | 1d |
| Human Anti-Mouse Antibodies (HAMA) | 10 | 0 |
| Human Immunodeficiency Virus (HIV) | 10 | 1b |
| Human T-Lymphotropic Virus-I (HTLV-I) | 10 | 0 |
| Human T-Lymphotropic Virus-II (HTLV-II) | 6 | 2c |
| Influenza Vaccine Recipient | 20 | 0 |
| Leptospirosis | 6 | 0 |
| Leptospirosis IgM | 5 | 0 |
| Lyme Disease | 10 | 0 |
| Monoclonal Hyper IgG | 7 | 1b |
| Anti-Nuclear Antibody (ANA) | 10 | 0 |
| Polyclonal Hyper IgG | 3 | 0 |
| Pregnant | 90 | 0 |
| Rheumatoid Factor | 10 | 0 |
| Rubella IgG | 10 | 0 |
| Systemic Lupus Erythematosus | 10 | 0 |
| Toxoplasma gondii IgG | 12 | 0 |
| Toxoplasma gondii IgM | 3 | 0 |
| Transplant Recipient | 10 | 0 |
| Varicella Zoster Virus | 10 | 0 |
| Total | 409 | 13 |
a Specimen was confirmed to be positive by other tests (TP-CLIA, RPR, TP-PA, FTA-ABS).
b Specimen was not confirmed positive by other tests (TP-CLIA, RPR, TP-PA, FTA-ABS).
& Specimen was reactive by ARCHITECT Syphilis TP but was not confirmed due to tube type limitations of the other tests (TP-CLIA, RPR, TP-PA, FTA-ABS).
d Specimen was positive by TP-CLIA and indeterminate by TP-PA and FTA-ABS.
Assay interference due to circulating antibodies against yaws, pinta, and bejel has not been evaluated. Cross-reactivity with these conditions is to be expected. Therefore, the following statement will be included in the Limitations section of the ARCHITECT Syphilis TP reagent package insert: "Assay interference due to circulating antibodies against yaws, pinta, and bejel has not been evaluated. Cross-reactivity with these treponemal disease conditions is to be expected."
{14}------------------------------------------------
Tube Type Matrix Comparison
The following tube types are acceptable for use with the ARCHITECT Syphilis TP
assay:
- . Serum, including serum separator
- Plasma: dipotassium EDTA, tripotassium EDTA, lithium heparin plasma separator, . lithium heparin, and sodium heparin
The ARCHITECT Syphilis TP assay showed the following mean/median S/CO difference and distribution of S/CO difference for nonreactive samples when compared to the control tube type (serum).
| Nonreactive Samples (Unspiked) | High Nonreactive Samples (Target 0.80 S/CO) | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Mean/MedianS/CO | Distribution of S/CO Differences | Mean/MedianS/CO | Distribution of S/CO Differences | |||||||
| Tube Type | N | Differenceª | < 0.10S/CO | 0.10-0.20S/CO | > 0.20S/CO | N | Differenceª | < 0.10S/CO | 0.10-0.20S/CO | > 0.20S/CO |
| Serum Separator,Plastic | 28 | 0.00 | 100.0%(28/28) | 0.0%(0/28) | 0.0%(0/28) | 27 | 0.02 | 81.5%(22/27) | 18.5%(5/27) | 0.0%(0/27) |
| DipotassiumEDTA | 28 | -0.00 | 100.0%(28/28) | 0.0%(0/28) | 0.0%(0/28) | 27 | 0.02 | 81.5%(22/27) | 18.5%(5/27) | 0.0%(0/27) |
| TripotassiumEDTA | 27 | -0.00 | 100.0%(27/27) | 0.0%(0/27) | 0.0%(0/27) | 27 | 0.02 | 85.2%(23/27) | 11.1%(3/27) | 3.7%(1/27) |
| Lithium HeparinPlasma Separator | 28 | -0.01 | 100.0%(28/28) | 0.0%(0/28) | 0.0%(0/28) | 27 | 0.01 | 88.9%(24/27) | 11.1%(3/27) | 0.0%(0/27) |
| Lithium Heparin | 28 | -0.01 | 100.0%(28/28) | 0.0%(0/28) | 0.0%(0/28) | 27 | 0.02 | 88.9%(24/27) | 7.4%(2/27) | 3.7%(1/27) |
| Sodium Heparin | 28 | -0.01 | 100.0%(28/28) | 0.0%(0/28) | 0.0%(0/28) | 26 | 0.00 | 88.5%(23/26) | 11.5%(3/26) | 0.0%(0/26) |
4 If the Shapiro-Wilk p-value is ≤ 0.0100, then the value displayed is the median.
{15}------------------------------------------------
The ARCHITECT Syphilis TP assay showed the following mean/median percent difference and distribution of percent difference for reactive samples when compared to the control tube type (serum).
| Low Reactive Samples (Target 1.20 S/CO) | High Reactive Samples (Target 6.00 S/CO) | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Distribution of % Differences | Distribution of % Differences | |||||||||
| Tube Type | N | Mean/Median% Differencea | < 10% | 10-20% | > 20% | N | Mean/Median% Differencea | < 10% | 10-20% | > 20% |
| Serum Separator,Plastic | 28 | 0.3 | 96.4%(27/28) | 3.6%(1/28) | 0.0%(0/28) | 28 | -0.1 | 96.4%(27/28) | 3.6%(1/28) | 0.0%(0/28) |
| DipotassiumEDTA | 28 | 0.9 | 100.0%(28/28) | 0.0%(0/28) | 0.0%(0/28) | 28 | -0.7 | 100.0%(28/28) | 0.0%(0/28) | 0.0%(0/28) |
| TripotassiumEDTA | 28 | -0.0 | 96.4%(27/28) | 3.6%(1/28) | 0.0%(0/28) | 27 | -0.3 | 100.0%(27/27) | 0.0%(0/27) | 0.0%(0/27) |
| Lithium HeparinPlasma Separator | 27 | 3.7 | 63.0%(17/27) | 33.3%(9/27) | 3.7%(1/27) | 28 | -0.1 | 100.0%(28/28) | 0.0%(0/28) | 0.0%(0/28) |
| Lithium Heparin | 28 | -0.3 | 96.4%(27/28) | 3.6%(1/28) | 0.0%(0/28) | 28 | -0.0 | 100.0%(28/28) | 0.0%(0/28) | 0.0%(0/28) |
| Sodium Heparin | 28 | 0.8 | 96.4%(27/28) | 3.6%(1/28) | 0.0%(0/28) | 28 | -0.3 | 100.0%(28/28) | 0.0%(0/28) | 0.0%(0/28) |
a If the Shapiro-Wilk p-value is ≤ 0.0100, then the value displayed is the median.
Sample On-Board Stability
A study was performed to evaluate the stability of samples when stored on the ARCHITECT i2000SR System (on-board storage) and tested with the ARCHITECT Syphilis TP assay.
Samples may be stored on-board the ARCHITECT instrument for up to 3 hours.
Sample Stability
A study was performed to evaluate serum and plasma specimens when subjected to various conditions (room temperature storage, 2 to 8°C storage, and freeze/thaw) and tested with the ARCHITECT Syphilis TP assay.
Specimens may be stored before testing
- · for up to 7 days after draw at 2 to 8°C (on the cells/clot)
- · for up to 72 hours after draw at room temperature (30°C) (on the cells/clot)
{16}------------------------------------------------
After storage at 2 to 8°C off the cells/clot, specimens can be subjected to up to 6 freeze/thaw cycles.
After storage at 2 to 8°C off the cells/clot, specimens can be stored frozen (-10°C or colder) for up to 30 days before testing.
High Dose Hook Effect
A study was performed that demonstrated that the ARCHITECT Syphilis TP assay is not susceptible to interference from specimens with high levels of anti-syphilis TP.
Within-Assay Sample Carryover
A study was performed to evaluate the susceptibility of the ARCHITECT Syphilis TP assay to within-assay sample carryover from a sample containing a very high level of syphilis anti-TP antibodies.
The ARCHITECT Syphilis TP assay is not susceptible to within-assay sample carryover.
{17}------------------------------------------------
VIII. Summary of Clinical Performance
EXPECTED VALUES
Due to geographic locations or demographics, assay results obtained in individual laboratories may vary from data presented.
A total of 1145 specimens prospectively collected from the intended use population were tested using the ARCHITECT Syphilis TP assay; 673 (58.8%) were female and 472 (41.2%) were male. The mean age was 35 years (age range: 6 to 91 years).
The ARCHITECT Syphilis TP assay was reactive in 163 (14.2 %) of the prospectively-collected specimens in the intended use population. Testing of the specimens was performed at three clinical testing sites located in in San Antonio, Texas; Baltimore, Maryland; and Temple, Texas.
The distribution of ARCHITECT Syphilis TP reactive and nonreactive results by age and gender is summarized in the following table.
| Age Range(Years) | Gender | ARCHITECT Syphilis TP Result | ||
|---|---|---|---|---|
| Number ofReactive (%) | Number ofNonreactive (%) | Total | ||
| 2 to 12 | Female | 0 (0.0%) | 1 (100.0%) | 1 |
| 13 to 21 | Female | 1 (0.8%) | 118 (99.2%) | 119 |
| 13 to 21 | Male | 1 (6.3%) | 15 (93.8%) | 16 |
| 22 to 29 | Female | 6 (2.6%) | 229 (97.4%) | 235 |
| 22 to 29 | Male | 12 (23.5%) | 39 (76.5%) | 51 |
| 30 to 39 | Female | 6 (2.9%) | 199 (97.1%) | 205 |
| 30 to 39 | Male | 32 (23.7%) | 103 (76.3%) | 135 |
| 40 to 49 | Female | 9 (12.9%) | 61 (87.1%) | 70 |
| 40 to 49 | Male | 50 (29.6%) | 119 (70.4%) | 169 |
| 50 to 59 | Female | 5 (21.7%) | 18 (78.3%) | 23 |
| 50 to 59 | Male | 29 (40.8%) | 42 (59.2%) | 71 |
| 60 to 64 | Female | 3 (50.0%) | 3 (50.0%) | 6 |
| 60 to 64 | Male | 2 (16.7%) | 10 (83.3%) | 12 |
{18}------------------------------------------------
| Age Range(Years) | Gender | ARCHITECT Syphilis TP Result | ||
|---|---|---|---|---|
| Number ofReactive (%) | Number ofNonreactive (%) | Total | ||
| 65 to 100 | Female | 1 (7.1%) | 13 (92.9%) | 14 |
| Male | 6 (33.3%) | 12 (66.7%) | 18 | |
| Total | 163 (14.2%) | 982 (85.8%) | 1145 |
The 1145 prospectively-collected specimens from the intended use population included 442 specimens sent for routine syphilis testing (325 female and 117 male, 6-91 years old), 304 pregnant females (16-43 years old), and 399 human immunodeficiency virus (HIV) positive individuals (44 female and 355 male, 18-72 years old).
The ARCHITECT Syphilis TP results for each category in the intended use population are summarized in the following table.
| ARCHITECT Syphilis TP Result | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Category | Number ofReactive (%) | Number ofNonreactive (%) | Total | |||||||
| Routine Syphilis | 38 (8.6%) | 404 (91.4%) | 442 | |||||||
| Pregnant | 1 (0.3%) | 303 (99.7%) | 304 | |||||||
| HIV Positive | 124 (31.1%) | 275 (68.9%) | 399 | |||||||
| Total | 163 (14.2%) | 982 (85.8%) | 1145 |
Precision
System Reproducibilitv
A 5-day precision study was performed for the ARCHITECT Syphilis TP assay based on guidance from CLSI documents EP05-A2 and EP15-A2. Testing was conducted at 3 clinical sites using 3 lots each of ARCHITECT Syphilis TP Reagents, 2 lots of ARCHITECT Syphilis TP Calibrator, and 1 lot of ARCHITECT Syphilis TP Controls, on 1 instrument per site. Two levels of controls and 4 serum panels were assayed in replicates of 4 at 2 separate times of day for 5 days.
{19}------------------------------------------------
| Sample | N | GrandMeanS/CO | Within-Run | Within-Day | Within-Laboratory (Total) | Precision withAdditionalComponent ofBetween-Site | Precision withAdditionalComponent ofBetween-Lot | Precision withAdditionalComponents ofSite and Lot(Overall) | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| SD | %CV | SD | %CV | SD | %CV | 95% CI | SD | %CV | SD | %CV | SD | %CV | |||
| Negative Control | 360 | 0.03 | 0.002 | 6.4 | 0.002 | 6.6 | 0.002 | 6.6 | (6.24,7.08) | 0.003 | 9.3 | 0.003 | 9.5 | 0.003 | 9.6 |
| Positive Controla | 360 | 2.72 | 0.108 | 4.0 | 0.130 | 4.8 | 0.134 | 4.9 | (4.59,5.35) | 0.134 | 4.9 | 0.134 | 4.9 | 0.134 | 4.9 |
| Nonreactive Panel | 360 | 0.08 | 0.004 | 5.1 | 0.004 | 5.1 | 0.004 | 5.4 | (5.07,5.78) | 0.006 | 7.5 | 0.006 | 7.4 | 0.007 | 8.8 |
| High Nonreactive Panel | 360 | 0.55 | 0.012 | 2.3 | 0.015 | 2.7 | 0.015 | 2.8 | (2.61,3.04) | 0.017 | 3.2 | 0.017 | 3.2 | 0.017 | 3.2 |
| Low Reactive Panel | 360 | 1.24 | 0.023 | 1.9 | 0.027 | 2.2 | 0.029 | 2.4 | (2.18,2.56) | 0.034 | 2.7 | 0.035 | 2.8 | 0.035 | 2.8 |
| High Reactive Panel | 360 | 3.57 | 0.073 | 2.0 | 0.079 | 2.2 | 0.094 | 2.6 | (2.43,2.91) | 0.107 | 3.0 | 0.100 | 2.8 | 0.107 | 3.0 |
4 An outlying run was observed at one site for one reagent lot. The total %CV for the Positive Control using the replacement run was 2.6%.
Clinical Performance
A multicenter study was conducted on the ARCHITECT iSystem to evaluate the ability of the ARCHITECT Syphilis TP assay to detect antibodies (IgG and IgM) directed against Treponema pallidum (TP).
A total of 2222 specimens were tested in the ARCHITECT Syphilis TP clinical study. Two specimens were excluded due to specimen issues. The remaining 2220 specimens included 1145 prospectively collected from the intended use population; 406 pre-selected positive for antibodies directed against TP based on previous laboratory testing (including 20 pregnant women known to be reactive for syphilis antibodies); 480 from apparently healthy individuals; 179 from medically diagnosed individuals with primary, secondary, or latent syphilis; and 10 specimens from pregnant females spiked with high antibody-positive syphilis TP specimens.
Each specimen prospectively collected from the intended use population or pre-selected positive for antibodies directed against TP based on previous laboratory testing was tested using the ARCHITECT Syphilis TP assay and the following comparator assays: a treponemal chemiluminescent immunoassay (TP-CLIA), a nontreponemal assay (Rapid Plasma Reagin [RPR]), and a second treponemal assay (Treponema Pallidum Particle Agglutination [TP-PA]). The final comparator result was determined using a 2 out of 3 rule (TP-CLIA, RPR, and TP-PA).
{20}------------------------------------------------
The clinical performance of the ARCHITECT Syphilis TP assay was evaluated by calculating positive percent agreement and negative percent agreement of the assay with the final comparator result.
During the clinical study (total number of tests = 2220), 2 exception codes prevented initial results and were retested to generate a final result.
Clinical Performance in Prospectively-Collected Specimens in Intended Use Population
Of the total 2220 specimens analyzed in the ARCHITECT Syphilis TP clinical study, 1145 were prospectively-collected intended use population specimens. These included 442 specimens sent for routine syphilis testing (325 female and 117 male, 6-91 years old), 304 pregnant females (16–43 years old), and 399 HIV positive (44 female and 355 male, 18-72 years old). Of the 1145 prospectively-collected specimens in the intended use population, 136 were pediatric specimens (6-21 years old).
The 1145 specimens included in the intended use population were prospectively sourced or collected at the following locations:
- · Baltimore, Maryland: 12.4%
- · Colton, California: 5.9%
- · Fort Lauderdale, Florida: 7.8%
- · Hyannis, Massachusetts: 0.1%
- · Los Angeles. California: 19.7%
- · Miami, Florida: 27.2%
- · San Antonio, Texas: 11.5%
- · Temple, Texas: 14.8%
- Location Unknown: 0.4%
A summary of the serological test profile for all prospectively-collected specimens in the intended use population is summarized in the following table.
| TP-CLIA | RPR | TP-PA | FinalComparatorResult | ARCHITECT | Number ofSubjects |
|---|---|---|---|---|---|
| + | + | + | + | + | 113 |
| + | + | I | + | + | 1 |
{21}------------------------------------------------
| TP-CLIA | RPR | TP-PA | FinalComparatorResult | ARCHITECT | Number ofSubjects |
|---|---|---|---|---|---|
| + | - | + | + | + | 37 |
| + | - | - | - | - | 4 |
| + | - | - | - | + | 2 |
| - | + | + | + | - | 6 |
| - | + | + | + | + | 1 |
| - | + | - | - | - | 167 |
| - | + | - | - | + | 3 |
| - | + | I | - | - | 1a |
| - | + | I | - | + | 1a |
| - | - | + | - | - | 8 |
| - | - | + | - | + | 2 |
| - | - | - | - | - | 796 |
| - | - | - | - | + | 2 |
| E | - | + | + | + | 1b |
| Total | 1145 |
- = Positive/Reactive
- = Negative/Nonreactive
E = Equivocal
I = Inconclusive
a Two specimens that were TP-CLIA nonreactive, RPR reactive, and TP-PA inconclusive were assigned a final comparator result of negative based on the nonreactive treponemal test.
b One specimen that was TP-CLIA equivocal, RPR nonreactive, and TP-PA reactive was assigned a final comparator result of positive based on the reactive treponemal test.
Percent Agreement
The comparison between the ARCHITECT Syphilis TP result and the final comparator result for the prospectively-collected specimens in the intended use population is summarized in the following table.
| ARCHITECT Syphilis TPResult Interpretation | Final Comparator Result | |
|---|---|---|
| Reactive | Nonreactive | |
| Reactive | 153 | 10 |
| Nonreactive | 6a | 976 |
a Six specimens were nonreactive by TP-CLIA and reactive by TP-PA and RPR.
{22}------------------------------------------------
Positive percent agreement was 96.2% (153/159) with a 95% confidence interval of 92.0% to 98.3%. Negative percent agreement was 99.0% (976/986) with a 95% confidence interval of 98.1% to 99.4%.
Percent Agreement by Category
The percent agreement between the ARCHITECT Syphilis TP result and the final comparator result for each category of the prospectively-collected specimens in the intended use population is summarized in the following table.
| Category | Positive PercentAgreement% (x/n) | 95%ConfidenceInterval (%) | Negative PercentAgreement% (x/n) | 95%ConfidenceInterval (%) |
|---|---|---|---|---|
| Routine Syphilis | 97.3 (36/37) | 86.2-99.5 | 99.5 (403/405) | 98.2-99.9 |
| Pregnant | NA | NA | 99.7 (303/304) | 98.2-99.9 |
| HIV Positive | 95.9 (117/122) | 90.8-98.2 | 97.5 (270/277) | 94.9-98.8 |
NA = not applicable
Clinical Performance in Pre-Selected Positive Specimens
Of the total 2220 specimens analyzed in the ARCHITECT Syphilis TP clinical study, 406 specimens were pre-selected positive for antibodies directed against TP based on previous laboratory testing (RPR and/or TP-PA). These included 386 presumed positive (106 female and 278 male, 16-78 years old). Gender was not reported for 2 specimens. Pre-selected positive specimens also included 20 reactive pregnant female specimens (17-36 years old). Of the 406 pre-selected positive specimens, 25 were pediatric specimens (16-21 years old).
A summary of the serological test profile for all pre-selected positive specimens is summarized in the following table.
| TP-CLIA | RPR | TP-PA | FinalComparatorResult | ARCHITECT | Number ofSubjects |
|---|---|---|---|---|---|
| + | + | I | + | + | 1 |
| + | - | + | + | - | 2 |
{23}------------------------------------------------
| TP-CLIA | RPR | TP-PA | FinalComparatorResult | ARCHITECT | Number ofSubjects |
|---|---|---|---|---|---|
| + | - | + | + | + | 116 |
| + | - | - | - | - | 1 |
| + | - | - | - | + | 2 |
| + | - | I | + | - | 1a |
| - | + | - | - | - | 3 |
| - | - | - | - | - | 20 |
| E | - | + | + | - | 1b |
| Total | 406 |
- = Positive/Reactive
- = Negative/Nonreactive
E = Equivocal
I = Inconclusive
a One specimen that was TP-CLIA reactive, RPR nonreactive, and TP-PA inconclusive was assigned a final comparator result of positive based on the reactive treponemal test.
b One specimen that was TP-CLIA equivocal, RPR nonreactive, and TP-PA reactive was assigned a final comparator result of positive based on the reactive treponemal test.
Percent Agreement
The comparison between the ARCHITECT Syphilis TP result and the final comparator
result for all pre-selected positive specimens is summarized in the following table.
| ARCHITECT Syphilis TPResult Interpretation | Final Comparator Result | |
|---|---|---|
| Reactive | Nonreactive | |
| Reactive | 376 | 2 |
| Nonreactive | 4 | 24 |
Positive percent agreement was 98.9% (376/380) with a 95% confidence interval of 97.3% to 99.6%. Negative percent agreement was 92.3% (24/26) with a 95% confidence interval of 75.9% to 97.9%.
{24}------------------------------------------------
Percent Agreement by Category
The percent agreement between the ARCHITECT Syphilis TP result and the final comparator result for each category in the pre-selected positive population is summarized in the following table.
| Category | Positive PercentAgreement% (x/n) | 95%ConfidenceInterval (%) | Negative PercentAgreement% (x/n) | 95%ConfidenceInterval (%) |
|---|---|---|---|---|
| Presumed Positive | 98.9 (356/360) | 97.2-99.6 | 92.3 (24/26) | 75.9-97.9 |
| Reactive Pregnant | 100.0 (20/20) | 83.9-100.0 | NA | NA |
NA = not applicable
Clinical Performance in Apparently Healthy Individuals
Of the total 2220 specimens analyzed in the ARCHITECT Syphilis TP clinical study, 480 specimens were from apparently healthy individuals. These included 244 female and 236 male, adults (22-89 years old), and pediatrics (0-21 years old).
| Category | Number of Reactive (%) | Number of Nonreactive (%) | Total |
|---|---|---|---|
| Adults | 15 (4.1%)a | 352 (95.9%) | 367 |
| Pediatrics | 0 (0.0%) | 113 (100.0%) | 113 |
| Total | 15 (3.1%) | 465 (96.9%) | 480 |
a Fourteen specimens were reactive and 1 specimen was equivocal by TP-CLIA.
{25}------------------------------------------------
Clinical Performance in Medically Diagnosed Individuals
Of the total 2220 specimens analyzed in the ARCHITECT Syphilis TP clinical study, 179 were from individuals medically diagnosed with primary, secondary, or latent syphilis. Medical diagnosis was made by a licensed physician based on the patient's clinical information and the results of serological testing, such as a positive test for syphilis (Venereal Disease Research Laboratory [VDRL] test, RPR, and/or TP-PA) at the time the specimen was collected. These included 9 female and 170 male, adults (22-66 years old), and pediatrics (18-21 years old).
| Medically Diagnosed Individuals | ARCHITECT Syphilis TP Result | |||
|---|---|---|---|---|
| Syphilis Stage | Treatment Status | N | Reactive | Nonreactive |
| Primary | Treated | 44 | 33 | 11a |
| Untreated | 25 | 25 | 0 | |
| Secondary | Treated | 29 | 29 | 0 |
| Untreated | 27 | 27 | 0 | |
| Latent | Treated | 25 | 25 | 0 |
| Untreated | 29 | 29 | 0 |
4 Nine specimens were nonreactive by TP-CLIA.
{26}------------------------------------------------
CLINICAL PERFORMANCE IN PREGNANT FEMALES
A total of 334 pregnant female specimens were analyzed in the ARCHITECT Syphilis TP clinical study. These included adults (22-43 years old) and pediatrics (16-21 years old). Age was not reported for 10 specimens.
The percent agreement for the prospectively-collected and the pre-selected positive specimens between the ARCHITECT Syphilis TP results and the final comparator results for the pregnant female population by trimester is summarized in the following table.
| Category | Positive PercentAgreement% (x/n) | 95%ConfidenceInterval (%) | Negative PercentAgreement% (x/n) | 95%ConfidenceInterval (%) |
|---|---|---|---|---|
| Prospectively-Collected | NA | NA | 99.7 (303/304) | 98.2-99.9 |
| First Trimester | NA | NA | 100.0 (13/13) | 77.2-100.0 |
| Second Trimestera | NA | NA | 99.2 (126/127) | 95.7-99.9 |
| Third Trimester | NA | NA | 100.0 (161/161) | 97.7-100.0 |
| Trimester Unknown | NA | NA | 100.0 (3/3) | 43.8-100.0 |
| Pre-Selected Positive | 100.0 (20/20) | 83.9-100.0 | NA | NA |
| First Trimester | 100.0 (1/1) | 20.7-100.0 | NA | NA |
| Second Trimester | NA | NA | NA | NA |
| Third Trimester | 100.0 (5/5) | 56.6-100.0 | NA | NA |
| Trimester Unknown | 100.0 (14/14) | 78.5-100.0 | NA | NA |
NA = not applicable
a One specimen reactive on ARCHITECT was reactive by TP-CLIA and nonreactive by RPR and TP-PA.
In addition, 10 nonreactive pregnant female specimens that were spiked with high antibody-positive syphilis TP specimens were reactive by ARCHITECT Syphilis TP and TP-CLIA. Positive percent agreement was 100.0% with a confidence interval of 72.2% to 100.0%.
{27}------------------------------------------------
Conclusion Drawn from Nonclinical and Clinical Laboratory Studies
The results presented in this 510(k) premarket notification demonstrate that the candidate assay (ARCHITECT Syphilis TP [List No. 8D06]) performance is substantially equivalent to the predicate assay (DiaSorin LIAISON Treponema assay, K061247).
The similarities and differences between the candidate assay and the predicate assay are presented in the tables starting on page 5-5.
There is no known potential adverse effect to the operator when using this in vitro device according to the ARCHITECT Syphilis TP reagent package insert instructions.
§ 866.3830
Treponema pallidum treponemal test reagents.(a)
Identification. Treponema pallidum treponemal test reagents are devices that consist of the antigens, antisera and all control reagents (standardized reagents with which test results are compared) which are derived from treponemal sources and that are used in the fluorescent treponemal antibody absorption test (FTA-ABS), theTreponema pallidum immobilization test (T.P.I.), and other treponemal tests used to identify antibodies toTreponema pallidum directly from infecting treponemal organisms in serum. The identification aids in the diagnosis of syphilis caused by bacteria belonging to the genusTreponema and provides epidemiological information on syphilis.(b)
Classification. Class II (performance standards).