K Number
K153487
Date Cleared
2016-02-04

(62 days)

Product Code
Regulation Number
880.5970
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Pressure Injectable PICC with Chlorag+ard Antithrombogenic Technology is indicated for shortterm or long-term peripheral access to the central venous system for intravenous therapy, blood sampling, infusion, pressure injection of contrast media and allows for central venous pressure monitoring. The maximum pressure of pressure injection equipment used with the pressure injectable PICC may not exceed 300 psi.

Chlorag+ard Technology treatment on the external surface of the catheter body as well as the entire fluid pathway of the catheter has been shown to be effective in reducing microbial colonization and thrombus accumulation on catheter surfaces. Antimicrobial and antithrombogenic effectiveness were evaluated using in vitro and in vivo test nethods and no correlation between these test methods and clinical outcome has currently been ascertained. It is not intended to be used for the treatment of existing infections or vein thrombosis.

The VPS Stylet and Console are indicated for guidance and tip positioning for central venous catheters. The Stylet provides stiffess for use in placement of the catheter, intravascular capability for ECG detection and recording and intravascular ultrasound for catheter guiding and positioning. The VPS Stylet, when used with the VPS Console, provides real-time catheter tip location information by using the patient's physiological (cardiac electrical activity and blood flow) information. When the VPS System guidance indicator shows a Blue Bullseye, the catherer tip is in the desired location.

The VPS System is indicated for use as alternative method to fluoroscopy or chest x-ray for central venous catheter tip placement confirmation in adult patients when a steady Blue Bullseye is obtained. NOTE: If a steady Blue Bullseve is not obtained, standard hospital practice should be followed to confirm catheter tip location.

Limiting but not contraindicated situations for this technique are in patients where alterations of cardiac rhythm change the presentation of the P-wave as in atrial flutter, severe tachycardia and pacemaker-driven rhythm, and in central venous catheterization procedures performed through femoral vein access which change the presentation of the P-wave. In such patients, who are easily identifiable prior to central venous catheter insertion, the use of an additional method is required to confirm catheter tip location.

Device Description

The purpose of this premarket notification is to propose design and material modifications to the subject device: the Arrow Vascular Positioning System (VPS) stylet PLUS and CG+ Arrow PICC powered by Arrow VPS Stylet PLUS . The Arrow Vascular Positioning System (VPS) Stylet PLUS (hereafter referred to as Arrow VPS Stylet PLUS) is designed for use with a VPS Console to guide market available central catheters to the desired location, which is the lower third of the superior vena cava (SVC) or at the cavo-atrial junction. The CG+ Arrow PICC powered by Arrow VPS Stylet PLUS provides the user with an Arrow VPS Stylet PLUS already loaded into a central catheter. (The Arrow VPS Stylet PLUS and the Arrow VPS Stylet PLUS included in the CG+ Arrow PICC powered by Arrow VPS Stylet PLUS are exactly the same stylet.)

The subject device, the Arrow VPS Stylet PLUS has the following characteristics:

  • 6 ft overall length
  • < 0.021" outer diameter over working length of polyimide with polytetrafluoroethylene ● (PTFE)
  • Intravascular electrocardiogram (ivECG) signal sensing consists of a platinumiridium (Pt/Ir) conductor band assembly
  • Doppler transducer connected to coaxial cable at the distal end
  • Coaxial cable and ivECG wire attached to connector at the proximal end to be plugged in ● to VPS Console or extension cable (that in turn connects to the VPS Console)
  • Tuohy-Borst adapter
  • . Marking accessory

The Arrow VPS Stylet PLUS is for use in a hospital setting by trained clinicians. The Arrow VPS Stylet PLUS is the stylet portion of a vascular positioning system designed to be used with the VPS Console and a market-available catheter. The Arrow VPS Stylet PLUS body is a polyimide tube with a fluoropolymer (PTFE) coating. The tubing contains a Doppler sensor on a coaxial cable and an intravascular electrocardiogram (ivECG) signal sensing platinum-iridium (Pt/Ir) conductor band with an intermediate stainless steel cannula that is welded to the stainless steel ECG wire. The stainless steel portions of the ECG conductor are encased within the platinum-iridium band. The Doppler sensor and the only exposed portion of the ivECG, i.e. the platinum-iridium band, are located at the distal end of the stylet and are used to detect and transmit physiological information to the VPS Console for analysis. The proximal end contains a connector to be plugged into the VPS Console or to an extension cable (that in turn connects to the VPS Console).

The CG+ Arrow PICC Powered by Arrow VPS Stylet PLUS is the Arrow VPS Stylet PLUS preloaded into a 4.5 Fr 1-Lumen, 5.5 Fr 2-Lumen, 40-55 cm pressure injectable Chlorag+ard Peripherally Inserted Central Catheter (PICC).

The CG+ Arrow PICC is a short-term or long-term, single use catheter designed to provide access to the central venous system. It consists of a non-tapered, radiopaque polyurethane extruded catheter body with a softer, contoured Blue Flex Tip. The catheter can be used for the iniection of contrast media. The external catheter body and the internal fluid path of the device are treated with Chlorag+ard, a Chlorhexidine-based coating technology.

AI/ML Overview

This document describes the premarket notification (K153487) for the CG+ Arrow PICC powered by Arrow VPS Stylet PLUS and Arrow VPS Stylet PLUS. It details modifications made to the device and the nonclinical testing performed to demonstrate substantial equivalence to a predicate device (K141618).

Here's an analysis of the acceptance criteria and study information provided:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" with numerical thresholds for performance. Instead, it lists the nonclinical tests performed and their results, which were all "PASS". The implicit acceptance criterion for each test is that the device must pass it.

Test DeviceTestReported Device Performance ("Results")
StyletTensile per ISO 10555-1PASS
StyletTorque strength Per the Coronary and Cerebrovascular Guidewire guidancePASS
StyletTip Flexibility Per the Coronary and Cerebrovascular Guidewire guidancePASS
StyletFlexing BS EN ISO 11070PASS
StyletCorrosion BS EN ISO 10555-1PASS
StyletComponent Compatibility: Marking accessory grip strengthPASS
StyletComponent Compatibility: Tuohy-Borst Adapter grip strengthPASS
StyletTuohy-Borst Leak per 594-1PASS
StyletPhysical characteristicsPASS
StyletX-ray Detectability ASTM F640-07PASS
StyletElectrical Performance: Hi-Pot, Continuity, Doppler Bandwidth, Capacitance, Sensitivity stabilityPASS
StyletBiocompatibility: Cytotoxicity, Irritation, Sensitization, Acute Systemic Toxicity and Hemocompatibility per ISO 10993-1PASS
Combined deviceElectrical Safety and Electromagnetic Compatibility Testing (IEC 60601-1-2, IEC 60601-2-37)PASS
Combined deviceCatheter Compatibility Simulated Use Insertion/ RemovalPASS
Combined deviceCatheter Compatibility Force to Remove stylet from catheterPASS

2. Sample Size Used for the Test Set and Data Provenance

The document describes nonclinical testing (bench testing and biocompatibility) rather than a clinical study with a "test set" from patients. Therefore, information about patient sample size and data provenance (country of origin, retrospective/prospective) is not applicable or provided. The tests were performed on the device itself.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. This was nonclinical testing, not a study involving human interpretation of data where expert consensus would establish ground truth.

4. Adjudication Method for the Test Set

Not applicable. This was nonclinical testing.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

No. This document describes nonclinical, bench, and biocompatibility testing to demonstrate substantial equivalence of a modified medical device to a predicate device. It is not an MRMC comparative effectiveness study, nor does it involve human readers or AI assistance.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done

No. This document pertains to a physical medical device (catheter and stylet), not an algorithm or AI system.

7. The Type of Ground Truth Used

The "ground truth" for the nonclinical tests was established by adherence to specified international standards and guidances (e.g., ISO 10555-1, BS EN ISO 11070, ASTM F640-07, ISO 10993-1, IEC 60601 series). These standards define the test methods and criteria for acceptable functional and safety performance (e.g., tensile strength, corrosion resistance, biocompatibility).

8. The Sample Size for the Training Set

Not applicable. This is not an AI/algorithm study requiring a training set. The modifications are to a physical medical device.

9. How the Ground Truth for the Training Set Was Established

Not applicable. This is not an AI/algorithm study requiring a training set.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of a human figure, with three overlapping profiles facing to the right. The profiles are rendered in black and create a sense of depth and unity. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 4, 2016

Arrow International, Inc. (subsidiary of Teleflex Inc.) Elizabeth Duncan Sr. Regulatory Affairs Specialist 2400 Bernville Road Reading, PA 19605

Re: K153487

Trade/Device Name: CG+ Arrow PICC powered by Arrow VPS Stylet PLUS and Arrow VPS Stylet PLUS Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: II Product Code: LJS, OBJ Dated: January 6, 2016 Received: January 7, 2016

Dear Ms. Duncan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours, Tina Kiang -S

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K153487

Device Name

CG+ Arrow PICC powered by Arrow VPS Stylet PLUS

Indications for Use (Describe)

The Pressure Injectable PICC with Chlorag+ard Antithrombogenic Technology is indicated for shortterm or long-term peripheral access to the central venous system for intravenous therapy, blood sampling, infusion, pressure injection of contrast media and allows for central venous pressure monitoring. The maximum pressure of pressure injection equipment used with the pressure injectable PICC may not exceed 300 psi.

Chlorag+ard Technology treatment on the external surface of the catheter body as well as the entire fluid pathway of the catheter has been shown to be effective in reducing microbial colonization and thrombus accumulation on catheter surfaces. Antimicrobial and antithrombogenic effectiveness were evaluated using in vitro and in vivo test nethods and no correlation between these test methods and clinical outcome has currently been ascertained. It is not intended to be used for the treatment of existing infections or vein thrombosis.

The VPS Stylet and Console are indicated for guidance and tip positioning for central venous catheters. The Stylet provides stiffess for use in placement of the catheter, intravascular capability for ECG detection and recording and intravascular ultrasound for catheter guiding and positioning. The VPS Stylet, when used with the VPS Console, provides real-time catheter tip location information by using the patient's physiological (cardiac electrical activity and blood flow) information. When the VPS System guidance indicator shows a Blue Bullseye, the catherer tip is in the desired location.

The VPS System is indicated for use as alternative method to fluoroscopy or chest x-ray for central venous catheter tip placement confirmation in adult patients when a steady Blue Bullseye is obtained. NOTE: If a steady Blue Bullseve is not obtained, standard hospital practice should be followed to confirm catheter tip location.

Limiting but not contraindicated situations for this technique are in patients where alterations of cardiac rhythm change the presentation of the P-wave as in atrial flutter, severe tachycardia and pacemaker-driven rhythm, and in central venous catheterization procedures performed through femoral vein access which change the presentation of the P-wave. In such patients, who are easily identifiable prior to central venous catheter insertion, the use of an additional method is required to confirm catheter tip location.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

See PRA Statement below.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K153487

Device Name Arrow VPS Stylet PLUS

Indications for Use (Describe)

The VPS Stylet and Console are indicated for guidance and tip positioning for central venous catheters. The Stylet provides stiffness for use in placement of the catheter, intravascular capability for ECG detection and recording and intravascular ultrasound for catheter guiding and positioning. The VPS Stylet, when used with the VPS Console, provides real-time catheter tip location information by using the patient's physiological (cardiac electrical activity and blood flow) information. When the VPS System guidance indicator shows a Blue Bullseye, the catheter tip is in the desired location.

The VPS System is indicated for use as alternative method to fluoroscopy or chest x-ray for central venous catheter tip placement confirmation in adult patients when a steady Blue Bullseye is obtained. NOTE: If a steady Blue Bullseve is not obtained, standard hospital practice should be followed to confirm catheter tip location.

Limiting but not contraindicated situations for this technique are in patients where alterations of cardiac rhythm change the presentation of the P-wave as in atrial flutter, severe tachycardia and pacemaker-driven rhythm, and in central venous catheterization procedures performed through femoral vein access which change the presentation of the P-wave. In such patients, who are easily identifiable prior to central venous catheter insertion, the use of an additional method is required to confirm catheter tip location.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{5}------------------------------------------------

K153487

510(K) SUMMARY PER 21 CFR 807.92

Submitter Information

Name:Arrow International, Inc (subsidiary of Teleflex Inc.)
Address:2400 Bernville RoadReading, PA 19605-9607 USA
Contact Person:Elizabeth DuncanSr. Regulatory Affairs Specialist
Telephone Number:(610) 378-0131 Extension 603220
Fax Number:(610) 478-3179
Email:elizabeth.duncan@teleflex.com
Date Prepared:December 3, 2015

Device Name

Device Trade Name: CG+ Arrow PICC powered by Arrow VPS Stylet PLUS and Arrow VPS Stylet PLUS Device Common Name: Catheter: Peripherally Inserted Central Catheter Stylet: Catheter, Ultrasound, Intravascular Device Classification Name: Catheter: Percutaneous Implanted Long-Term Intravascular Catheter Stylet: Diagnostic Intravascular Catheter FDA Classification Regulation: Catheter: 21 CFR: 880.5970 FDA Classification (Catheter and Stylet): Class II FDA Product Code: Catheter: LJS Stylet: OBJ

Predicate Devices

  • K141618: Arrow Vascular Positioning System (VPS) stylet PLUS and CG+ Arrow PICC . powered by Arrow VPS Stylet PLUS

Modifications

This Special 510(k) is being submitted for design and material modifications to the predicate Arrow VPS Stylet PLUS and the Arrow VPS Stylet PLUS included in the predicate CG+ Arrow PICC powered by Arrow VPS Stylet PLUS (K141618). The following is a list of modifications being made to the subject device followed by the reasons for the changes:

{6}------------------------------------------------

Material modification

  • Addition of a platinum-iridium conductor band assembly and associated adhesive

Design Modification

  • Increase in stylet body inner diameter (ID); decrease in outer diameter (OD) ●
  • Increase in the ECG wire body OD
  • Decrease in the co-ax wire OD
  • Connector design changes
  • Process of PTFE coating

See the "Technological Characteristics and Substantial Equivalence" table below for more detailed information.

The predicate premarket notification for the Arrow VPS Stylet PLUS and the Arrow VPS Stylet PLUS included in the predicate CG+ Arrow PICC powered by Arrow VPS Stylet PLUS (K141618) did not contain information about the VPS Console information was contained within separate cleared 510(k)s (K103260 and K123813). There are no changes to the VPS Consoles associated with this submission.

There are no changes to the catheter or Chlorag+ard technology portion of the predicate Arrow CG+ Arrow PICC powered by the Arrow VPS Stylet (K141618).

Device Description

The purpose of this premarket notification is to propose design and material modifications to the subject device: the Arrow Vascular Positioning System (VPS) stylet PLUS and CG+ Arrow PICC powered by Arrow VPS Stylet PLUS . The Arrow Vascular Positioning System (VPS) Stylet PLUS (hereafter referred to as Arrow VPS Stylet PLUS) is designed for use with a VPS Console to guide market available central catheters to the desired location, which is the lower third of the superior vena cava (SVC) or at the cavo-atrial junction. The CG+ Arrow PICC powered by Arrow VPS Stylet PLUS provides the user with an Arrow VPS Stylet PLUS already loaded into a central catheter. (The Arrow VPS Stylet PLUS and the Arrow VPS Stylet PLUS included in the CG+ Arrow PICC powered by Arrow VPS Stylet PLUS are exactly the same stylet.)

The subject device, the Arrow VPS Stylet PLUS has the following characteristics:

  • 6 ft overall length
  • < 0.021" outer diameter over working length of polyimide with polytetrafluoroethylene ● (PTFE)
  • Intravascular electrocardiogram (ivECG) signal sensing consists of a platinumiridium (Pt/Ir) conductor band assembly
  • Doppler transducer connected to coaxial cable at the distal end
  • Coaxial cable and ivECG wire attached to connector at the proximal end to be plugged in ● to VPS Console or extension cable (that in turn connects to the VPS Console)

{7}------------------------------------------------

  • Tuohy-Borst adapter
  • . Marking accessory

The Arrow VPS Stylet PLUS is for use in a hospital setting by trained clinicians. The Arrow VPS Stylet PLUS is the stylet portion of a vascular positioning system designed to be used with the VPS Console and a market-available catheter. The Arrow VPS Stylet PLUS body is a polyimide tube with a fluoropolymer (PTFE) coating. The tubing contains a Doppler sensor on a coaxial cable and an intravascular electrocardiogram (ivECG) signal sensing platinum-iridium (Pt/Ir) conductor band with an intermediate stainless steel cannula that is welded to the stainless steel ECG wire. The stainless steel portions of the ECG conductor are encased within the platinum-iridium band. The Doppler sensor and the only exposed portion of the ivECG, i.e. the platinum-iridium band, are located at the distal end of the stylet and are used to detect and transmit physiological information to the VPS Console for analysis. The proximal end contains a connector to be plugged into the VPS Console or to an extension cable (that in turn connects to the VPS Console).

The CG+ Arrow PICC Powered by Arrow VPS Stylet PLUS is the Arrow VPS Stylet PLUS preloaded into a 4.5 Fr 1-Lumen, 5.5 Fr 2-Lumen, 40-55 cm pressure injectable Chlorag+ard Peripherally Inserted Central Catheter (PICC).

The CG+ Arrow PICC is a short-term or long-term, single use catheter designed to provide access to the central venous system. It consists of a non-tapered, radiopaque polyurethane extruded catheter body with a softer, contoured Blue Flex Tip. The catheter can be used for the iniection of contrast media. The external catheter body and the internal fluid path of the device are treated with Chlorag+ard, a Chlorhexidine-based coating technology.

Intended Use

The intended use of the subject and predicate (K141618) devices are identical and are listed below.

The intended use of the Arrow VPS Stylet PLUS and VPS Console (together with the VPS System) is to quickly and accurately guide market available central catheters to the desired location. which is the lower third of the superior vena cava (SVC) or at the cavo-atrial junction.

The CG+ Arrow PIC Catheters powered by Arrow VPS Stylet PLUS permit venous access to the central circulation through a peripheral vein with the guidance of the VPS system.

Indications for Use

The indications for use of the subject and predicate (K141618) devices are identical and are listed below.

Arrow VPS Stylet PLUS:

{8}------------------------------------------------

The VPS Stylet and Console are indicated for guidance and tip positioning for central venous catheters. The Stylet provides stiffness for use in placement of the catheter, intravascular capability for ECG detection and recording and intravascular ultrasound for catheter guiding and positioning. The VPS Stylet, when used with the VPS Console, provides real-time catheter tip location information by using the patient's physiological (cardiac electrical activity and blood flow) information. When the VPS System guidance indicator shows a blue bullseye, the catheter tip is in the desired location.

The VPS System is indicated for use as an alternative method to fluoroscopy or chest x-ray for central venous catheter tip placement confirmation in adult patients when a steady blue bullseye is obtained. NOTE: If a steady blue bullseye is not obtained, standard hospital practice should be followed to confirm catheter tip location.

Limiting but not contraindicated situations for this technique are in patients where alterations of cardiac rhythm change the presentation of the P-wave as in atrial fibrillation, atrial flutter, severe tachycardia and pacemaker-driven rhythm, and in central venous catheterization procedures performed through femoral or saphenous vein access which change the presentation of the Pwave. In such patients, who are easily identifiable prior to central venous catheter insertion, the use of an additional method is required to confirm catheter tip location.

CG+ Arrow PIC Catheters powered by Arrow VPS Stylet PLUS:

The Pressure Injectable PICC with Chlorag+ard Antimicrobial and Antithrombogenic Technology is indicated for short-term or long-term peripheral access to the central venous system for intravenous therapy, blood sampling, infusion, pressure injection of contrast media and allows for central venous pressure monitoring. The maximum pressure of pressure injection equipment used with the pressure injectable PICC may not exceed 300 psi.

Chlorag+ard Technology treatment on the external surface of the catheter body as well as the entire fluid pathway of the catheter has been shown to be effective in reducing microbial colonization and thrombus accumulation on catheter surfaces. Antimicrobial and antithrombogenic effectiveness were evaluated using in vitro and in vivo test methods and no correlation between these test methods and clinical outcome has currently been ascertained. It is not intended to be used for the treatment of existing infections or vein thrombosis.

The VPS Stylet and Console are indicated for guidance and tip positioning for central venous catheters. The Stylet provides stiffness for use in placement of the catheter, intravascular capability for ECG detection and recording and intravascular ultrasound for catheter guiding and positioning. The VPS Stylet, when used with the VPS Console, provides real-time catheter tip location information by using the patient's physiological (cardiac electrical activity and blood flow) information. When the VPS System guidance indicator shows a blue bullseye, the catheter tip is in the desired location.

{9}------------------------------------------------

The VPS System is indicated for use as an alternative method to fluoroscopy or chest x-ray for central venous catheter tip placement confirmation in adult patients when a steady blue bullseye is obtained. NOTE: If a steady blue bullseye is not obtained, standard hospital practice should be followed to confirm catheter tip location.

Limiting but not contraindicated situations for this technique are in patients where alterations of cardiac rhythm change the presentation of the P-wave as in atrial fibrillation, atrial flutter, severe tachycardia and pacemaker-driven rhythm, and in central venous catheterization procedures performed through femoral or saphenous vein access which change the presentation of the Pwave. In such patients, who are easily identifiable prior to central venous catheter insertion, the use of an additional method is required to confirm catheter tip location.

CharacteristicPredicate DeviceSubject Device
Design6 foot long stylet which contains a Dopplersensor and an intravascular electrocardiogram(ivECG) signal sensing wire at the distal tipSAME: 6 foot long stylet which contains aDoppler sensor and an intravascularelectrocardiogram (ivECG) signal sensing wire atthe distal tip
Signal ConductorTwo conductor stylet wires (ECG wire andcoaxial cable)SAME: Two conductor stylet wires (ECG wireand coaxial cable)
ECG Conductor pathwayExposed conductive surface at distal end ofECG wireSAME: Exposed conductive surface at distal endof ECG wire
ECG wire OD (nominal).006".007"
ConnectorSingle, 7 prong custom, 2 barbsSingle, 7 prong custom, with the addition of abarb, two clips and grooves
Stylet body OD and ID (nominal)ID: 0.0134"OD: 0.0187"ID: 0.0144"OD: 0.0174"
Coaxial cable OD (nominal)OD: .0063"OD: .0062"
Stylet Patient Contacting Materials(Circulating Blood contact for limitedduration)Stylet body: Polyimide with Fluoropolymer(PTFE) heat shrinkECG Conductor: 304V SSTStylet body: SAME material, different process,Polyimide coated with Fluoropolymer (PTFE)ECG Conductor: Platinum-Iridium band weldedto 304V SST
Tuohy-Borst: Polycarbonate, Polypropylene,silicone, PTFE, TPETuohy-Borst: SAME, Polycarbonate,Polypropylene, silicone, PTFE, TPE
Adhesive encased over transducer: Loctite3311 UV Cured AcrylicSAME: Loctite 3311 UV Cured Acrylic
Adhesive used to bond ECG assembly tostylet body: Loctite 3311 UV cured acrylicAdhesive used to bond ECG assembly to styletbody: Loctite 4014 Cyanoacrylate acrylic adhesive
Operating Frequency of Stylet DopplerTransducer11.667 MHzSAME: 11.667 MHz
Size of Stylet Doppler Transducer0.4 mmSAME: 0.4 mm
Stylet Doppler Transducer ShapeHexagonalSAME: Hexagonal
Marking AccessoryTwist-locking colletSAME: Twist-locking collet
PIC CathetersAll characteristics: materials, diameter (4.5,5.5, 6 Fr.), lengths, manufacturing processes,conditions and aidsSAME: All characteristics: materials, diameter(excluding the 6 Fr, since this sized catheter is notincluded in the subject device), lengths,manufacturing processes, conditions and aids
Chlorhexidine TechnologyAll characteristics: Identity, formulation,concentration (content per surface area),manufacturing processes, conditions and aids,method of application to the device andmechanism by which the agent is releasedfrom the deviceSAME: All characteristics

Technological Characteristics and Substantial Equivalence

{10}------------------------------------------------

CharacteristicPredicate DeviceSubject Device
Instruction s for UseContentStylet: SAMECatheter: SAME
Stylet Shelf life1 yearSAME: 1 year
Sterilization: All characteristics,including: method, assurance levelAll characteristics, including:Method: Ethylene OxideAssurance Level: 10-6SAME: All characteristics, including:Method: Ethylene OxideAssurance Level: 10-6
PackagingStandalone kits: PET tray sealed in a Tyvekheader pouchPreloaded kits: PETG tray with a TyveklidstockStandalone kits: Polystyrene (HIPS) tray withTyvek sterile barrierPreloaded kits: PETG tray with a Tyvek lidstock

The Arrow YPS Stylet PLUS and CG+ Arrow PICC powered by Arrow VPS Stylet PLUS is substantially equivalent to the predicate Arrow VPS Stylet PLUS and the CG+ Arrow PICC powered by Arrow VPS Stylet PLUS (K141618) in terms of: intended use, indications for use, patient population, contraindications, operating principle, functional performance, sterilization, fundamental scientific technology, features, material and design.

There are no changes to the catheter or Chlorag+ard Technology portion of the predicate CG+ Arrow PICC catheters powered by Arrow VPS Stylet PLUS (K141618). There are no changes being made to the VPS Consoles (K103260 and K123813), as part of this submission.

Nonclinical Testing

The modifications proposed to the subject device were evaluated using risk management. The risk analysis method used to assess the impact of the modifications was per ISO 14971 and included a Risk Management Plan. Hazards Analysis, and Risk Management Report. As a result of the risk assessment, the testing listed in the table below was determined to be required. The resultant testing listed in the table below was performed on the subject device: the Arrow VPS Stylet PLUS and the CG+ Arrow PICC powered by Arrow VPS Stylet PLUS after being preconditioned with EtO sterilization, ISTA simulated environmental conditions and one year accelerated aging:

Test DeviceTestResults
StyletTensile per ISO 10555-1PASS
StyletTorque strength Per the Coronary and Cerebrovascular Guidewire guidancePASS
StyletTip Flexibility Per the Coronary and Cerebrovascular Guidewire guidancePASS
StyletFlexing BS EN ISO 11070PASS
StyletCorrosion BS EN ISO 10555-1PASS
StyletComponent Compatibility: Marking accessory grip strengthPASS
StyletComponent Compatibility: Tuohy-Borst Adapter grip strengthPASS
StyletTuohy-Borst Leak per 594-1PASS
StyletPhysical characteristicsPASS
StyletX-ray Detectability ASTM F640-07PASS
StyletElectrical Performance: Hi-Pot, Continuity, Doppler Bandwidth, Capacitance,Sensitivity stabilityPASS
StyletBiocompatibility: Cytotoxicity, Irritation, Sensitization, Acute SystemicPASS

{11}------------------------------------------------

Test DeviceTestResults
Toxicity and Hemocompatibility per ISO 10993-1
Combined deviceElectrical Safety and Electromagnetic Compatibility Testing (IEC 60601-1-2,IEC 60601-2-37)PASS
Catheter Compatibility Simulated Use Insertion/ RemovalPASS
Catheter Compatibility Force to Remove stylet from catheterPASS

Conclusions

The subject device is substantially equivalent to the predicate device in terms of intended use, indications for use, patient population, contraindications, operating principle, functional performance, fundamental scientific technology, features, design and material. The results of the risk assessment and resultant testing performed have demonstrated that the design and material modifications is as safe, as effective and performs as well as the predicate device and therefore the subject device is considered substantially equivalent to the cited predicate device.

§ 880.5970 Percutaneous, implanted, long-term intravascular catheter.

(a)
Identification. A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.(b)
Classification. Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”