(62 days)
No
The document describes a vascular positioning system that uses physiological signals (ECG and ultrasound) for catheter tip placement guidance. While it analyzes these signals, there is no mention of AI or ML algorithms being used for this analysis or for any other function of the device. The system relies on real-time signal processing and a predefined indicator (Blue Bullseye) based on these signals.
No
The device is a system for guiding and confirming the placement of central venous catheters, not for treating a disease or condition. The PICC itself is for accessing the central venous system for various therapies, but the overall system described focuses on guidance and positioning.
Yes
The device, specifically the VPS Stylet and Console, provides real-time catheter tip location information by using the patient's physiological (cardiac electrical activity and blood flow) information, and is indicated for use as an alternative method to fluoroscopy or chest x-ray for central venous catheter tip placement confirmation, which is a diagnostic function.
No
The device description clearly details hardware components like a stylet with a Doppler transducer, ivECG signal sensing band, coaxial cable, and a connector. While it interacts with a console (which likely contains software), the device itself is a physical medical device with integrated hardware.
Based on the provided text, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections. They are used to provide information for diagnosis, monitoring, or screening.
- Device Function: The described device (Pressure Injectable PICC with Chlorag+ard Antithrombogenic Technology and the Arrow Vascular Positioning System (VPS) Stylet PLUS) is a medical device used for accessing the central venous system and guiding and positioning central venous catheters. It utilizes physiological signals (ECG and blood flow) and ultrasound for real-time tip location.
- Lack of Sample Analysis: The device does not analyze samples taken from the patient's body in a laboratory setting. It interacts directly with the patient's circulatory system and uses real-time physiological data for guidance and positioning.
The device is a therapeutic and diagnostic (in the sense of providing real-time positioning information) medical device, but it does not fit the definition of an in vitro diagnostic.
N/A
Intended Use / Indications for Use
The Pressure Injectable PICC with Chlorag+ard Antithrombogenic Technology is indicated for shortterm or long-term peripheral access to the central venous system for intravenous therapy, blood sampling, infusion, pressure injection of contrast media and allows for central venous pressure monitoring. The maximum pressure of pressure injection equipment used with the pressure injectable PICC may not exceed 300 psi. Chlorag+ard Technology treatment on the external surface of the catheter body as well as the entire fluid pathway of the catheter has been shown to be effective in reducing microbial colonization and thrombus accumulation on catheter surfaces. Antimicrobial and antithrombogenic effectiveness were evaluated using in vitro and in vivo test nethods and no correlation between these test methods and clinical outcome has currently been ascertained. It is not intended to be used for the treatment of existing infections or vein thrombosis. The VPS Stylet and Console are indicated for guidance and tip positioning for central venous catheters. The Stylet provides stiffess for use in placement of the catheter, intravascular capability for ECG detection and recording and intravascular ultrasound for catheter guiding and positioning. The VPS Stylet, when used with the VPS Console, provides real-time catheter tip location information by using the patient's physiological (cardiac electrical activity and blood flow) information. When the VPS System guidance indicator shows a Blue Bullseye, the catherer tip is in the desired location. The VPS System is indicated for use as alternative method to fluoroscopy or chest x-ray for central venous catheter tip placement confirmation in adult patients when a steady Blue Bullseye is obtained. NOTE: If a steady Blue Bullseve is not obtained, standard hospital practice should be followed to confirm catheter tip location. Limiting but not contraindicated situations for this technique are in patients where alterations of cardiac rhythm change the presentation of the P-wave as in atrial flutter, severe tachycardia and pacemaker-driven rhythm, and in central venous catheterization procedures performed through femoral vein access which change the presentation of the P-wave. In such patients, who are easily identifiable prior to central venous catheter insertion, the use of an additional method is required to confirm catheter tip location.
Product codes (comma separated list FDA assigned to the subject device)
LJS, OBJ
Device Description
The purpose of this premarket notification is to propose design and material modifications to the subject device: the Arrow Vascular Positioning System (VPS) stylet PLUS and CG+ Arrow PICC powered by Arrow VPS Stylet PLUS . The Arrow Vascular Positioning System (VPS) Stylet PLUS (hereafter referred to as Arrow VPS Stylet PLUS) is designed for use with a VPS Console to guide market available central catheters to the desired location, which is the lower third of the superior vena cava (SVC) or at the cavo-atrial junction. The CG+ Arrow PICC powered by Arrow VPS Stylet PLUS provides the user with an Arrow VPS Stylet PLUS already loaded into a central catheter. (The Arrow VPS Stylet PLUS and the Arrow VPS Stylet PLUS included in the CG+ Arrow PICC powered by Arrow VPS Stylet PLUS are exactly the same stylet.)
The subject device, the Arrow VPS Stylet PLUS has the following characteristics:
- 6 ft overall length
§ 880.5970 Percutaneous, implanted, long-term intravascular catheter.
(a)
Identification. A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.(b)
Classification. Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
0
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of a human figure, with three overlapping profiles facing to the right. The profiles are rendered in black and create a sense of depth and unity. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 4, 2016
Arrow International, Inc. (subsidiary of Teleflex Inc.) Elizabeth Duncan Sr. Regulatory Affairs Specialist 2400 Bernville Road Reading, PA 19605
Re: K153487
Trade/Device Name: CG+ Arrow PICC powered by Arrow VPS Stylet PLUS and Arrow VPS Stylet PLUS Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: II Product Code: LJS, OBJ Dated: January 6, 2016 Received: January 7, 2016
Dear Ms. Duncan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
1
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, Tina Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
2
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K153487
Device Name
CG+ Arrow PICC powered by Arrow VPS Stylet PLUS
Indications for Use (Describe)
The Pressure Injectable PICC with Chlorag+ard Antithrombogenic Technology is indicated for shortterm or long-term peripheral access to the central venous system for intravenous therapy, blood sampling, infusion, pressure injection of contrast media and allows for central venous pressure monitoring. The maximum pressure of pressure injection equipment used with the pressure injectable PICC may not exceed 300 psi.
Chlorag+ard Technology treatment on the external surface of the catheter body as well as the entire fluid pathway of the catheter has been shown to be effective in reducing microbial colonization and thrombus accumulation on catheter surfaces. Antimicrobial and antithrombogenic effectiveness were evaluated using in vitro and in vivo test nethods and no correlation between these test methods and clinical outcome has currently been ascertained. It is not intended to be used for the treatment of existing infections or vein thrombosis.
The VPS Stylet and Console are indicated for guidance and tip positioning for central venous catheters. The Stylet provides stiffess for use in placement of the catheter, intravascular capability for ECG detection and recording and intravascular ultrasound for catheter guiding and positioning. The VPS Stylet, when used with the VPS Console, provides real-time catheter tip location information by using the patient's physiological (cardiac electrical activity and blood flow) information. When the VPS System guidance indicator shows a Blue Bullseye, the catherer tip is in the desired location.
The VPS System is indicated for use as alternative method to fluoroscopy or chest x-ray for central venous catheter tip placement confirmation in adult patients when a steady Blue Bullseye is obtained. NOTE: If a steady Blue Bullseve is not obtained, standard hospital practice should be followed to confirm catheter tip location.
Limiting but not contraindicated situations for this technique are in patients where alterations of cardiac rhythm change the presentation of the P-wave as in atrial flutter, severe tachycardia and pacemaker-driven rhythm, and in central venous catheterization procedures performed through femoral vein access which change the presentation of the P-wave. In such patients, who are easily identifiable prior to central venous catheter insertion, the use of an additional method is required to confirm catheter tip location.
Type of Use (Select one or both, as applicable) | |
---|---|
X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
3
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
4
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K153487
Device Name Arrow VPS Stylet PLUS
Indications for Use (Describe)
The VPS Stylet and Console are indicated for guidance and tip positioning for central venous catheters. The Stylet provides stiffness for use in placement of the catheter, intravascular capability for ECG detection and recording and intravascular ultrasound for catheter guiding and positioning. The VPS Stylet, when used with the VPS Console, provides real-time catheter tip location information by using the patient's physiological (cardiac electrical activity and blood flow) information. When the VPS System guidance indicator shows a Blue Bullseye, the catheter tip is in the desired location.
The VPS System is indicated for use as alternative method to fluoroscopy or chest x-ray for central venous catheter tip placement confirmation in adult patients when a steady Blue Bullseye is obtained. NOTE: If a steady Blue Bullseve is not obtained, standard hospital practice should be followed to confirm catheter tip location.
Limiting but not contraindicated situations for this technique are in patients where alterations of cardiac rhythm change the presentation of the P-wave as in atrial flutter, severe tachycardia and pacemaker-driven rhythm, and in central venous catheterization procedures performed through femoral vein access which change the presentation of the P-wave. In such patients, who are easily identifiable prior to central venous catheter insertion, the use of an additional method is required to confirm catheter tip location.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
5
K153487
510(K) SUMMARY PER 21 CFR 807.92
Submitter Information
Name: | Arrow International, Inc (subsidiary of Teleflex Inc.) |
---|---|
Address: | 2400 Bernville Road |
Reading, PA 19605-9607 USA | |
Contact Person: | Elizabeth Duncan |
Sr. Regulatory Affairs Specialist | |
Telephone Number: | (610) 378-0131 Extension 603220 |
Fax Number: | (610) 478-3179 |
Email: | elizabeth.duncan@teleflex.com |
Date Prepared: | December 3, 2015 |
Device Name
Device Trade Name: CG+ Arrow PICC powered by Arrow VPS Stylet PLUS and Arrow VPS Stylet PLUS Device Common Name: Catheter: Peripherally Inserted Central Catheter Stylet: Catheter, Ultrasound, Intravascular Device Classification Name: Catheter: Percutaneous Implanted Long-Term Intravascular Catheter Stylet: Diagnostic Intravascular Catheter FDA Classification Regulation: Catheter: 21 CFR: 880.5970 FDA Classification (Catheter and Stylet): Class II FDA Product Code: Catheter: LJS Stylet: OBJ
Predicate Devices
- K141618: Arrow Vascular Positioning System (VPS) stylet PLUS and CG+ Arrow PICC . powered by Arrow VPS Stylet PLUS
Modifications
This Special 510(k) is being submitted for design and material modifications to the predicate Arrow VPS Stylet PLUS and the Arrow VPS Stylet PLUS included in the predicate CG+ Arrow PICC powered by Arrow VPS Stylet PLUS (K141618). The following is a list of modifications being made to the subject device followed by the reasons for the changes:
6
Material modification
- Addition of a platinum-iridium conductor band assembly and associated adhesive
Design Modification
- Increase in stylet body inner diameter (ID); decrease in outer diameter (OD) ●
- Increase in the ECG wire body OD
- Decrease in the co-ax wire OD
- Connector design changes
- Process of PTFE coating
See the "Technological Characteristics and Substantial Equivalence" table below for more detailed information.
The predicate premarket notification for the Arrow VPS Stylet PLUS and the Arrow VPS Stylet PLUS included in the predicate CG+ Arrow PICC powered by Arrow VPS Stylet PLUS (K141618) did not contain information about the VPS Console information was contained within separate cleared 510(k)s (K103260 and K123813). There are no changes to the VPS Consoles associated with this submission.
There are no changes to the catheter or Chlorag+ard technology portion of the predicate Arrow CG+ Arrow PICC powered by the Arrow VPS Stylet (K141618).
Device Description
The purpose of this premarket notification is to propose design and material modifications to the subject device: the Arrow Vascular Positioning System (VPS) stylet PLUS and CG+ Arrow PICC powered by Arrow VPS Stylet PLUS . The Arrow Vascular Positioning System (VPS) Stylet PLUS (hereafter referred to as Arrow VPS Stylet PLUS) is designed for use with a VPS Console to guide market available central catheters to the desired location, which is the lower third of the superior vena cava (SVC) or at the cavo-atrial junction. The CG+ Arrow PICC powered by Arrow VPS Stylet PLUS provides the user with an Arrow VPS Stylet PLUS already loaded into a central catheter. (The Arrow VPS Stylet PLUS and the Arrow VPS Stylet PLUS included in the CG+ Arrow PICC powered by Arrow VPS Stylet PLUS are exactly the same stylet.)
The subject device, the Arrow VPS Stylet PLUS has the following characteristics:
- 6 ft overall length