(28 days)
REXEED-S Series Dialyzer is intended for use in hemodialysis for the treatment of patients suffering from acute or chronic renal failure.
The line of Asahi REXEED-S Series Dialyzer (hereafter referred to as REXEED-S) is a family of high permeability hollow fiber dialyzers intended for the treatment of patients with acute or chronic renal failure.
REXEED-S is designed for single use.
REXEED-S is constructed of hollow fiber membrane housed within a plastic housing of Styrene-Butadiene block copolymer and is subject to gamma-ray irradiation prior to shipment.
This document is a 510(k) summary for a medical device called the Asahi REXEED-S Series Dialyzer, which is a high-permeability hemodialysis system. The submission describes modifications to the device's materials, specifically in the housing and stoppers, and aims to demonstrate substantial equivalence to previously cleared versions of the same device. Therefore, the "acceptance criteria" discussed are largely related to design verification rather than diagnostic performance metrics (like sensitivity, specificity, etc.) that would typically apply to an AI/ML medical device.
Based on the provided text, here's a breakdown of the requested information, focusing on the context of this specific device and submission:
1. A table of acceptance criteria and the reported device performance
Acceptance Criteria (General) | Reported Device Performance (General) |
---|---|
Material properties: | The modifications to the housing and stopper materials were evaluated. |
Functional performance: | The device's performance characteristics (implied to be filtration, clearance, etc.) were maintained. |
Biocompatibility: | The new materials are safe for patient contact. |
Sterilization effectiveness: | The gamma-ray irradiation process remained effective. |
Risk mitigation: | All identified risks associated with the modifications were adequately addressed. |
Specific numerical acceptance criteria and performance values are not detailed in this summary document. The document states that "All test results met the acceptance criteria, and proved those modifications to be appropriate."
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not specify the sample size used for the test set or the data provenance. It refers generally to "Design verification tests."
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not provided in the document. Given that this is a material and design modification for a physical device, the "ground truth" would likely be established through engineering specifications, material science testing, and possibly clinical performance data (though not explicitly detailed here). The experts would be engineers, material scientists, and potentially medical professionals involved in hemodialysis.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Adjudication methods like 2+1 or 3+1 are typically used for establishing ground truth in diagnostic studies involving expert review. This is not applicable to this type of device modification submission, which focuses on engineering and material performance.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
MRMC studies are relevant for diagnostic AI/ML devices. This is a submission for a physical medical device (dialyzer) with material modifications, so an MRMC study is not applicable and was not performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
"Standalone" performance is relevant for AI/ML algorithms. This is a submission for a physical medical device, so this concept is not applicable.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For this device modification, the "ground truth" was established through:
- Engineering specifications: Standards for material properties, mechanical integrity, and functional performance (e.g., filtration rates, solute clearance).
- Material testing: Chemical and physical analysis of the new materials.
- Biocompatibility testing: To ensure safety when in contact with blood.
- Risk analysis (FMEA): To identify and mitigate potential failure modes.
8. The sample size for the training set
This concept is relevant for AI/ML models. This is a physical device, so there is no "training set" in the AI/ML sense. Design verification and validation are based on testing prototypes and production samples.
9. How the ground truth for the training set was established
As there is no AI/ML "training set" for this device, this question is not applicable. The "ground truth" for verifying the device's performance (as described in point 7) was established through established engineering and scientific methodologies.
§ 876.5860 High permeability hemodialysis system.
(a)
Identification. A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K
uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.).
(3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
(b)
Classification. Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”