(125 days)
Not Found
No
The document describes standard MR imaging functionalities and workflow enhancements, with no mention of AI or ML algorithms for image analysis, diagnosis, or other functions.
No
The "Intended Use / Indications for Use" section explicitly states that "This system is a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device." It also mentions that the images, spectra, and measurements "may assist the diagnosis and therapy planning," but its primary indicated use is for diagnosis, not therapy itself.
Yes
The "Intended Use / Indications for Use" section explicitly states, "This system is a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device." It further elaborates that images, spectra, and measurements "provide information that may assist the diagnosis and therapy planning."
No
The device description explicitly mentions both hardware and software changes, indicating it is not a software-only medical device.
Based on the provided text, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly states it is a "Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device." It produces images and spectra of the internal structure of the body.
- Mechanism of Action: The device uses magnetic resonance to create images based on the spatial distribution of protons or other nuclei. This is an imaging modality, not a test performed on biological samples in vitro.
- Interpretation: The images, spectra, and measurements are interpreted by a trained physician to assist in diagnosis and therapy planning. This is typical for imaging devices.
- Lack of IVD Characteristics: There is no mention of analyzing biological samples (blood, urine, tissue, etc.) or performing tests on these samples outside of the body.
IVD devices are specifically designed to perform tests on biological specimens in vitro to provide information for diagnosis, monitoring, or treatment. This device is an in vivo diagnostic imaging system.
N/A
Intended Use / Indications for Use
This system is a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device.
The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities.
Magnetic Resonance images represent the spatial distribution of protons or other nuclei with spin. Image appearance is determined by many different physical properties of the tissue and the MR scan technique applied. The image acquisition process can be synchronized with the patient's breathing or cardiac cycle. The systems can use combinations of images to produce physical parameters, and related derived images.
Images, spectra, and measurements of physical parameters, when interpreted by a trained physician, provide information that may assist the diagnosis and therapy planning. The accuracy of determined physical parameters depends on system and scan parameters, and must be controlled and validated by the clinical user. For some studies the use of contrast agents can be essential. Their application is subject to local medico-legal regulations and to their appropriateness to assist the diagnosis and therapy planning as judged by a trained physician.
In addition the Philips MR systems provide imaging capabilities, such as MR fluoroscopy, to guide and evaluate interventional and minimally invasive procedures in the head, body or extremities.
MR Interventional procedures, performed inside or adjacent to the Philips MR system, must be performed with MR Conditional or MR Safe instrumentation as selected and evaluated by the clinical user for use with the specific MR system configuration in the hospital. The appropriateness and use of information from a Philips MR system for a specific interventional procedure and specific MR system configuration must be validated by the clinical user.
Product codes (comma separated list FDA assigned to the subject device)
LNH, LNI
Device Description
The proposed Ingenia 1.5T and Ingenia 1.5T S R5.2 with ScanWise Implant feature is provided with a 70 cm magnet. ScanWise Implant functionality enables MR technologists to implement an improved and controlled workflow for MR Conditional implants. The feature consists of an extension to the Patient Registration User Interface where the information relevant to MR Conditional device labeling can be assessed, controlled and reviewed. The ScanWise Implant feature allows the user at the examination level to define restrictions on the 'active fields' generated by the MR system.
The proposed Ingenia 1.5T and Ingenia 1.5T S R5.2 with ScanWise Implant feature also consolidates separately-cleared novel functionalities, and minor hardware and software changes since the clearance of the currently marketed and predicate device, Ingenia R4 (K110151, 03/22/2011).
Following minor hardware and software changes are covered in this submission:
- (Hardware) Enhanced Patient Communication User Interface Module, IEC/ISO compliant symbols.
- (Hardware) New computing platform and peripherals for MR Spectrometer (DDAS).
- (Software) User Interface layout modifications for scan preparation, sequence planning (geometries and parameters), and data processing and viewing.
- (Software) Planning on cine images.
- (Software) SAR related parameters (SED). Pregnancy status related to Normal Mode.
- (Software) Parameter optimization for the reconstruction algorithms.
- (Software) Partial NSA algorithm in reconstruction.
- (Software) AutoVoice, using pre-recorded spoken instructions.
- (Software) VCG, optimized electrode placement and enhanced algorithm.
- (Software) ComforTone: mechanical resonance frequency dependent timing adjustments of sequences for lower acoustic noise.
- Enhanced sequences:
a. AutoSpair.
b. TSE flow compensation enhancement.
c. Optimized 3D TSE flip angle sweeps per anatomy.
d. ENCASE: 3D encoding.
e. CardiacQuant: triggered T1 mapping sequence.
f. pCASL.
g. DTI enhancements.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Magnetic Resonance (MR)
Anatomical Site
head, body or extremities
Indicated Patient Age Range
Not Found
Intended User / Care Setting
trained physician, clinical user, MR technologists / hospital
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Non-Clinical verification and or validation tests have been performed with regards to the intended use, the technical claims, the requirement specifications and the risk management results. Specifically for ScanWise Implant, Human Factors Engineering testing was performed in line with FDA's guidance document entitled "Applying Human Factors and Usability Engineering to Optimize Medical Device Design - June 22, 2011". The verification, validation test results and sample clinical images demonstrate that the proposed Ingenia 1.5T and Ingenia 1.5T S R5.2 with ScanWise Implant feature: Complies with the aforementioned international and FDA-recognized consensus standards and Device specific guidance document, entitled "Guidance for the Submission Of Premarket Notifications for Magnetic Resonance Diagnostic Devices – November 14, 1998" Meets the acceptance criteria and is adequate for its intended use.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 892.1000 Magnetic resonance diagnostic device.
(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
0
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 22, 2016
Philips Medical Systems, Nederlands B.V. % Mr. Henrie Daniels Regulatory Affairs Engineer Veenpluis 4-6 Best, 5684 PC THE NETHERLANDS
Re: K153324
Trade/Device Name: Ingenia 1.5T and Ingenia 1.5T S R5.2 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH, LNI Dated: February 16, 2016 Received: February 18, 2016
Dear Mr. Daniels:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
1
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D.'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K153324
Device Name
Ingenia 1.5T and Ingenia 1.5T S R5.2
Indications for Use (Describe)
This system is a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device.
The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities.
Magnetic Resonance images represent the spatial distribution of protons or other nuclei with spin. Image appearance is determined by many different physical properties of the tissue and the MR scan technique applied. The image acquisition process can be synchronized with the patient's breathing or cardiac cycle. The systems can use combinations of images to produce physical parameters, and related derived images.
Images, spectra, and measurements of physical parameters, when interpreted by a trained physician, provide information that may assist the diagnosis and therapy planning. The accuracy of determined physical parameters depends on system and scan parameters, and must be controlled and validated by the clinical user. For some studies the use of contrast agents can be essential. Their application is subject to local medico-legal regulations and to their appropriateness to assist the diagnosis and therapy planning as judged by a trained physician.
In addition the Philips MR systems provide imaging capabilities, such as MR fluoroscopy, to guide and evaluate interventional and minimally invasive procedures in the head, body and extremities.
MR Interventional procedures, performed inside or adjacent to the Philips MR system, must be performed with MR Conditional or MR Safe instrumentation as selected and evaluated by the clinical user for use with the specific MR system configuration in the hospital. The appropriateness and use of information from a Philips MR system for a specific interventional procedure and specific MR system configuration must be validated by the clinical user.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
Image /page/3/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, which makes the blue letters stand out.
510(k) Summary
510(k) Summary of Safety and Effectiveness
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92.
Date Prepared: | March 16, 2016 | |
---|---|---|
Manufacturer: | Philips Medical Systems Nederland B.V. | |
Veenpluis 4-6, 5684 PC, Best, The Netherlands | ||
Establishment Registration Number: 3003768277 | ||
Primary Contact | ||
Person: | Henrie Daniels | |
Regulatory Affairs Engineer | ||
Phone: +31 40 2762192 | ||
E-mail: henrie.daniels@philips.com | ||
Secondary Contact | ||
Person | Ruojuan Zhang | |
Regulatory Affairs engineer | ||
Phone: +31 631685825 | ||
E-mail: ruojuan.zhang@philips.com | ||
Device Name: | Ingenia 1.5T and Ingenia 1.5T S R5.2 | |
Classification: | Classification Name: | Magnetic Resonance Diagnostic |
Device (MRDD) | ||
Classification Regulation: | 21CFR §892.1000 | |
Classification Panel: | Radiology | |
Device Class: | Class II | |
Primary Product code: | 90LNH | |
90LNI | ||
Primary Predicate | ||
Device: | Trade Name: | ACHIEVA R4 1.5T and ACHIEVA |
R4 3.0T (aka Ingenia) | ||
Manufacturer: | Philips Medical Systems Nederland | |
B.V. | ||
510(k) Clearance: | K110151 (March 22, 2011) | |
Classification Regulation: | 21 CFR, Part 892.1000 | |
Classification Name: | Magnetic Resonance Diagnostic | |
Device (MRDD) | ||
Classification Panel: | Radiology | |
Device Class: | Class II | |
Product Code: | 90LNH | |
90LNI | ||
Device description: | The proposed Ingenia 1.5T and Ingenia 1.5T S R5.2 with | |
ScanWise Implant feature is provided with a 70 cm magnet. | ||
ScanWise Implant functionality enables MR technologists to | ||
implement an improved and controlled workflow for MR | ||
Conditional implants. The feature consists of an extension to the | ||
Patient Registration User Interface where the information relevant | ||
to MR Conditional device labeling can be assessed, controlled and | ||
reviewed. The ScanWise Implant feature allows the user at the | ||
examination level to define restrictions on the 'active fields' | ||
generated by the MR system. |
The proposed Ingenia 1.5T and Ingenia 1.5T S R5.2 with
ScanWise Implant feature also consolidates separately-cleared novel
functionalities, and minor hardware and software changes since the
clearance of the currently marketed and predicate device, Ingenia
R4 (K110151, 03/22/2011).
Following minor hardware and software changes are covered in this
submission:
- (Hardware) Enhanced Patient Communication User Interface
Module, IEC/ISO compliant symbols. - (Hardware) New computing platform and peripherals for MR
Spectrometer (DDAS). - (Software) User Interface layout modifications for scan
preparation, sequence planning (geometries and parameters), and
data processing and viewing. - (Software) Planning on cine images.
- (Software) SAR related parameters (SED). Pregnancy status
related to Normal Mode. - (Software) Parameter optimization for the reconstruction
algorithms. - (Software) Partial NSA algorithm in reconstruction.
- (Software) AutoVoice, using pre-recorded spoken instructions.
- (Software) VCG, optimized electrode placement and enhanced
algorithm. - (Software) ComforTone: mechanical resonance frequency
dependent timing adjustments of sequences for lower acoustic
noise. - Enhanced sequences:
a. AutoSpair.
b. TSE flow compensation enhancement.
c. Optimized 3D TSE flip angle sweeps per anatomy. | |
| | | d. ENCASE: 3D encoding. |
| | | e. CardiacQuant: triggered T1 mapping sequence. |
| | | f. pCASL. |
| | | g. DTI enhancements. |
| Indications for Use: | This system is a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. | |
| | The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities. | |
| | Magnetic Resonance images represent the spatial distribution of protons or other nuclei with spin. Image appearance is determined by many different physical properties of the tissue and the anatomy, and the MR scan technique applied. The image acquisition process can be synchronized with the patient's breathing or cardiac cycle. The systems can use combinations of images to produce physical parameters, and related derived images. | |
| | Images, spectra, and measurements of physical parameters, when interpreted by a trained physician, provide information that may assist the diagnosis and therapy planning. The accuracy of determined physical parameters depends on system and scan parameters, and must be controlled and validated by the clinical user. For some studies the use of contrast agents can be essential. Their application is subject to local medico-legal regulations and to their appropriateness to assist the diagnosis and therapy planning as judged by a trained physician. | |
| | In addition the Philips MR systems provide imaging capabilities, such as MR fluoroscopy, to guide and evaluate interventional and minimally invasive procedures in the head, body and extremities. MR Interventional procedures, performed inside or adjacent to the Philips MR system, must be performed with MR Conditional or MR Safe instrumentation as selected and evaluated by the clinical user for use with the specific MR system configuration in the hospital. The appropriateness and use of information from a Philips MR system for a specific interventional procedure and specific MR system configuration must be validated by the clinical user. | |
| Fundamental
Scientific Technology: | The proposed Ingenia 1.5T and Ingenia 1.5T S R5.2 with ScanWise Implant feature is based on the principle that certain atomic nuclei present in the human body will emit a weak relaxation signal when placed in a strong magnetic field and excited by a radio signal at the precession frequency | |
| Summary of Non-Clinical Performance
Data: | The emitted relaxation signals are analyzed by the system and a computed image reconstruction is displayed on a video screen.
The principal technological components (magnet, transmit body coil, gradient coil, receive coils and patient support) of the proposed Ingenia 1.5T and Ingenia 1.5T S R5.2 with ScanWise Implant feature is identical to those used in the currently marketed and predicate device, Ingenia R4( K110151, 03/22/2011).
ScanWise Implant uses existing safety mechanisms to protect the patient against excessive RF exposures. This includes Whole Body and Head SAR, local SAR controls, and display of B1+rms. No modifications relative to the implementation of safety mechanisms relative to the predicate device was required.
ScanWise Implant extends existing software safety provisions to prevent peripheral nerve stimulation. In previous products, dB/dt was evaluated (at the compliance volume defined in IEC60601-2-33) and displayed for informational purposes. In this software, dB/dt is controlled not to exceed a user-specified value.
Based on the information provided above, the proposed Ingenia 1.5T and Ingenia 1.5T S R5.2 with ScanWise Implant feature does not raise different questions of safety and effectiveness compared to the marketed and predicate device, Ingenia R4(K110151,03/22/2011).
The proposed Ingenia 1.5T and Ingenia 1.5T S R5.2 with ScanWise Implant feature comply with the following international and FDA-recognized consensus standards:
IEC60601-1 Edition 3 Amendment 1 IEC60601-1-2 Edition 3 IEC60601-1-6 Edition 3 / IEC62366 IEC60601-1-8 Edition 2 IEC60601-2-33 Edition 3 Amendment 1 IEC 62304 NEMA MS-1 2008 NEMA MS-4 2008 NEMA MS-8 2008 ISO 14971 Application of risk management to medical devices (2007) Device specific guidance document, entitled "Guidance for the Submission Of Premarket Notifications for Magnetic Resonance Diagnostic Devices – November 14 1998" | |
| | Non-Clinical verification and or validation tests have been performed
with regards to the intended use, the technical claims, the requirement
specifications and the risk management results. | |
| | Specifically for ScanWise Implant, Human Factors Engineering
testing was performed in line with FDA's guidance document entitled
"Applying Human Factors and Usability Engineering to Optimize
Medical Device Design - June 22, 2011". | |
| | The verification, validation test results and sample clinical images
demonstrate that the proposed Ingenia 1.5T and Ingenia 1.5T S
R5.2 with ScanWise Implant feature: | |
| | Complies with the aforementioned international and FDA-
recognized consensus standards and Device specific
guidance document, entitled "Guidance for the Submission
Of Premarket Notifications for Magnetic Resonance
Diagnostic Devices – November 14, 1998" Meets the acceptance criteria and is adequate for its intended
use. | |
| | Therefore, the proposed Ingenia 1.5T and Ingenia 1.5T S R5.2
with ScanWise Implant feature is substantially equivalent to the
currently marketed and predicate device, Ingenia R4( K110151,
03/22/2011) in terms of safety and effectiveness. | |
| Summary of Clinical
Data: | The proposed Ingenia 1.5T and Ingenia 1.5T S R5.2 with
ScanWise Implant feature did not require clinical study since
substantial equivalence to the primary currently marketed and
predicate device was demonstrated with the following attributes: Design features; Indication for use; Fundamental scientific technology; Non-clinical performance testing; and Safety and effectiveness. | |
| Substantial
Equivalence
Conclusion: | The proposed Ingenia 1.5T and Ingenia 1.5T S R5.2 with
ScanWise Implant feature and the currently marketed and predicate
device, Ingenia R4( K110151, 03/22/2011) have the same primary
indications for use with respect to the following: Providing cross-sectional images based on the magnetic
resonance phenomenon Interpretation of the images is the responsibility of trained
physicians Images can be used for interventional and treatment
planning purposes | |
4
Image /page/4/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is plain white, which makes the blue letters stand out.
5
Image /page/5/Picture/0 description: The image shows the word "PHILIPS" in a bold, sans-serif font. The word is written in a bright blue color. The background is plain white, which makes the word stand out.
6
Image /page/6/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is plain white, which makes the blue letters stand out. The logo is simple and recognizable.
7
Image /page/7/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame.
8
Image /page/8/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame.
The proposed Ingenia 1.5T and Ingenia 1.5T S R5.2 with ScanWise Implant feature and control software are substantially equivalent to the currently marketed and predicate device, Ingenia R4 (K110151, 03/22/2011) in terms of improving the user control of safety related parameters in MR Conditional workflows. |
---|
The proposed Ingenia 1.5T and Ingenia 1.5T S R5.2 with ScanWise Implant feature are substantially equivalent to the currently marketed and predicate device, Ingenia R4,(K110151, 03/22/2011), in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness. Additionally, substantial equivalence was demonstrated with non-clinical performance (verification and validation) tests, which complied with the requirements specified in the international and FDA-recognized consensus standards and device-specific guidance. The results of these tests demonstrate that the proposed Ingenia 1.5T and Ingenia 1.5T S R5.2 with ScanWise Implant feature met the acceptance criteria and is adequate for its intended use. |