K Number
K153303
Manufacturer
Date Cleared
2016-08-10

(268 days)

Product Code
Regulation Number
866.3940
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EUROIMMUN Anti-West Nile Virus ELISA (IgG) is intended for the qualitative detection of IgG antibodies to West Nile virus in human serum and plasma (K+EDTA, Li+-heparin). This test is intended as an aid in the presumptive laboratory diagnosis of West Nile virus infection in patients with clinical symptoms consistent with meningitis encephalitis, in conjunction with other laboratory and clinical findings. Positive results must be confirmed by the plaque reduction test (PRNT) or by using the current CDC guidelines for diagnosis of this disease.

The assay characteristics have not been established for testing cord blood, neonates, prenatal screening, and general population screening of patients without symptoms of meningoencephalitis. This assay is not FDA cleared or approved for testing blood or plasma donors.

Warning: Cross-reactivity with IgG to Dengue, Chikungunya, Zika and Tick-borne Encephalitis (TBE) viruses has been observed with the EUROIMMUN Anti-West Nile Virus ELISA (IgG). Reactive results must be reported with a caution statement regarding possible IgG cross-reactivity with other flaviviruses.

Device Description

Patient samples are diluted 1:101 in sample buffer, 100 ul of each diluted patient sample and pre-diluted controls and the calibrator are added to the antigen coated microtiter wells and incubated for 60 minutes at +37°C. After incubation the microtiter well strips are washed 3 times with wash buffer to remove unbound antibodies and 100 ul of the anti-human IgG enzyme conjugate reagent is added to each microtiter well. After an additional 30-minutes incubation at room temperature, the microtiter wells are again washed 3 times with wash buffer to remove any unbound enzyme conjugate and 100 ul of the chromogen substrate is added. The strips are incubated for 15 minutes at room temperature and 100 ul stop solution is added. The microtiter plates are placed in an ELISA reader and read at a wavelength of 450 nm and a reference wavelength of between 620 nm and 650 nm within 30 minutes.

AI/ML Overview

The provided text describes the EUROIMMUN Anti-West Nile Virus ELISA (IgG) device, which is an in vitro diagnostic immunoassay. Therefore, the questions related to AI/MRMC studies, human-in-the-loop performance, and expert ground truth are not applicable in the context of this device. The acceptance criteria and study data are presented for the assay's analytical and clinical performance.

Here's the breakdown of the information based on the provided document:

1. Acceptance Criteria and Reported Device Performance

The acceptance criteria for the EUROIMMUN Anti-West Nile Virus ELISA (IgG) are primarily based on its analytical performance (repeatability, reproducibility, specificity) and clinical performance (sensitivity and specificity compared to a predicate device and PRNT).

Acceptance Criteria CategorySpecific Metric/TargetReported Device Performance
Analytical PerformanceRepeatability (Intra-assay precision): Demonstrated low variability across multiple runs and days. No specific %CV target is stated as an acceptance criterion, but low %CV values indicate acceptable repeatability.Repeatability: For 7 seropositive samples tested over 14 runs on 7 days (3 replicates/run), the total %CV ranged from 6.1% to 13.6%. For negative samples (0.1 ratio) it was 13.3%.
Reproducibility (Inter-site precision): Demonstrated low variability across multiple sites. No specific %CV target is stated as an acceptance criterion, but low %CV values indicate acceptable reproducibility.Reproducibility: For 7 seropositive samples tested at 3 different sites over 5 days (2 runs/day, 2 replicates/run), the total %CV ranged from 10.1% to 16.3%. For negative samples (0.1 ratio) it was 16.3%.
Analytical Specificity/Cross-Reactivity: Evaluate cross-reactivity with other common infections and autoimmune markers. While no specific rejection threshold is given, the warning in the Indications for Use (cross-reactivity with other flaviviruses) suggests that some level of cross-reactivity is acknowledged and managed through labeling. The goal is to minimize non-specific reactivity.Cross-Reactivity: Tested against 910 serologically characterized specimens from various diseases. Overall, 164 (18.0%) of these showed positive results with the WNV ELISA. Notably high cross-reactivity was observed with:
- Anti-Chikungunya virus: 23/72 (31.9% positive, 68.1% negative)
- Anti-Dengue virus: 50/58 (86.2% positive, 13.8% negative)
- Anti-TBE virus: 33/118 (28.0% positive, 72.0% negative)
- Anti-Zika virus: 47/47 (100% positive, 0% negative)
Interferences: Demonstrate no significant interference from common biological substances (hemoglobin, triglycerides, bilirubin).Interferences: No influence on results up to 1000 mg/dL hemoglobin, 2000 mg/dL triglycerides, and 40 mg/dL bilirubin. (Note: Interferences from high protein, cholesterol, and intralipids were not investigated).
Assay Cut-off: Establish a cut-off with optimal sensitivity and specificity (e.g., 100% sensitivity and 100% specificity in the initial ROC analysis). The borderline range should cover a high percentage of negative samples (e.g., 98%).Assay Cut-off: ROC analysis on 18 positive and 150 negative samples achieved 100.0% sensitivity and 100.0% specificity at OD 0.475. The Borderline range (ratio 0.8 to 1.1) covered 98% of negative samples. Using this cut-off/borderline with the same groups: Sensitivity 100.0% (95% C.I.: 81.5 - 100.0%), and specificity 98.7% (95% C.I.: 95.3 - 99.8% if borderline samples counted as positive).
Clinical PerformancePositive Agreement (PPA) with Predicate Device: High agreement is expected for positive samples. No explicit acceptance percentage is stated, but typically values above 90% are sought for substantial equivalence.Clinical Study I (Prospective, US): Positive agreement: 93.3% (42/45), 95% C.I. 81.7-98.6% vs. predicate.
Negative Agreement (NPA) with Predicate Device: High agreement is expected for negative samples. No explicit acceptance percentage is stated, but typically values above 90% are sought for substantial equivalence.Clinical Study I (Prospective, US): Negative agreement: 94.4% (101/107), 95% C.I. 88.2-97.9% vs. predicate.
Sensitivity vs. PRNT (Gold Standard): High sensitivity is critical for identifying infected individuals. For presumptive positives.Clinical Study I (Prospective, US): For 27 presumptive positives by the predicate, tested against PRNT (ground truth): Sensitivity 92.6% (25/27), 95% C.I. 75.7-99.1%.
Positive Agreement (PPA) with Predicate Device (Study II): (Replication/validation of agreement).Clinical Study II (Prospective, US): Positive agreement: 95.6% (43/45), 95% C.I. 84.9-99.5% vs. predicate.
Negative Agreement (NPA) with Predicate Device (Study II): (Replication/validation of agreement).Clinical Study II (Prospective, US): Negative agreement: 98.9% (352/356), 95% C.I. 97.1-99.7% vs. predicate.
Positive Agreement vs. PRNT (Retrospective, Non-US): High agreement with an established reference method, especially for positive samples. For outbreak samples.Clinical Study III (Retrospective, Germany/South Africa): Positive agreement: 99.5% (194/195), 95% C.I. 97.2-100.0% vs. RKI Anti-West Nile Virus PRNT.
Negative Agreement vs. PRNT (Retrospective, Non-US): High agreement with an established reference method, especially for negative samples.Clinical Study III (Retrospective, Germany/South Africa): Negative agreement: 97.0% (97/100), 95% C.I. 91.5-99.4% vs. RKI Anti-West Nile Virus PRNT.
Matrix ComparisonEquivalence of Serum vs. Plasma: The device should perform equivalently across tested sample matrices (serum, K+EDTA plasma, Li-heparin plasma). Regression equation should be near ideal (y=x), and % recovery should be within an acceptable range (e.g., 90-110%).Matrix Comparison: Based on 20 spiked paired samples each for EDTA and Li-heparin plasma vs. serum.
- EDTA plasma: Regression equation y = -0.01 + 1.02x (near ideal). R2 = 0.9983. Mean % recovery = 101% (Range 98-105%).
- Li-heparin plasma: Regression equation y = 0.04 + 0.96x (near ideal). R2 = 0.9759. Mean % recovery = 100% (Range 92-109%).

2. Sample Size Used for the Test Set and Data Provenance

The "test set" here refers to the clinical samples used for performance validation against a predicate device and ground truth (PRNT).

  • Clinical Study I (Predicate Comparison and PRNT Confirmation):
    • Sample Size: 152 samples for predicate comparison. Of these, 45 were presumptive positives by the predicate, and 27 of these were further tested by PRNT.
    • Data Provenance: Prospective collection from hospitals and clinics across the US in 2015.
  • Clinical Study II (Predicate Comparison):
    • Sample Size: 401 serum samples.
    • Data Provenance: Prospective collection from a clinical laboratory in the midwest (US).
  • Clinical Study III (PRNT Comparison):
    • Sample Size: 295 serum samples.
    • Data Provenance: Retrospective collection in cooperation with the Robert Koch Institute (RKI), Berlin, Germany, including 200 samples from major outbreaks of West Nile fever in South Africa in 1974 and 1984.
  • Analytical Specificity/Cross-Reactivity Study:
    • Sample Size: 910 serologically characterized seropositive specimens.
    • Data Provenance: Not explicitly stated, but implies diverse sources given the range of conditions.
  • Assay Cut-off Determination:
    • Sample Size: 18 sera from clinically characterized positive WNV patients and 150 sera from normal healthy blood donors.
    • Data Provenance: Implied clinical and healthy populations, source not specified.
  • Expected Values (US Studies):
    • Sample Size: 553 samples.
    • Data Provenance: Prospectively collected from US population.
  • Matrix Comparison:
    • Sample Size: 20 paired samples (serum, EDTA plasma, Li-heparin plasma), created by spiking 5 different sets of normal blood donor samples with 5 different positive serum samples.
    • Data Provenance: Normal blood donors (source not specified), spiked with serum samples (source not specified if human/animal).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

  • This is an in vitro diagnostic (IVD) device, not an AI/imaging device requiring expert reads for ground truth in the same way.
  • The ground truth for the clinical studies was established by Plaque Reduction Neutralization Test (PRNT), which is considered the gold standard for serological confirmation of West Nile Virus infection.
  • For the comparison studies, a legally marketed predicate device (Focus Diagnostics West Nile Virus IgG DxSelect™ (K031953)) also served as a comparative "ground truth" or reference, as is common in 510(k) submissions for substantial equivalence.
  • The document does not mention the use of human experts or their qualifications for establishing the ground truth for data used in the clinical studies. The ground truth relies on established laboratory tests (PRNT) or comparison to an already FDA-cleared device.

4. Adjudication Method for the Test Set

  • No explicit "adjudication method" in the sense of multiple human readers resolving discrepancies is mentioned for this IVD device.
  • For Clinical Study I, it states: "Average of three results for each clinical specimen tested at three sites was considered to calculate the positive percent and negative percent agreement between the EUROIMMUN Anti-West Nile Virus ELISA (IgG) vs the predicate assay." This indicates internal averaging/calculation, not expert adjudication.
  • The ground truth for discordant results was resolved by PRNT where applicable (e.g., of the 45 presumptive positives by predicate in Study I, 27 were sent for PRNT).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

  • Not applicable. This is an in vitro diagnostic (ELISA) device, not an AI or imaging device that uses human readers.

6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

  • Not applicable. This is an automated ELISA assay. Its "standalone performance" is what is described in the analytical and clinical studies. There is no human-in-the-loop component in the interpretation of the assay results, other than a lab technician running the test and reporting the quantitative ratio/index, which then falls into predefined qualitative categories (negative, borderline, positive).

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

The types of ground truth used were primarily:

  • Reference Laboratory Assay: Plaque Reduction Neutralization Test (PRNT) is considered the gold standard for confirming West Nile Virus infection. (Used in Clinical Study I and III).
  • Predicate Device Comparison: The Focus Diagnostics West Nile Virus IgG DxSelect™ (K031953), an already legally marketed device, served as a reference standard for comparison in Clinical Study I and II to demonstrate substantial equivalence.
  • Serological Characterization: For the cross-reactivity study, specimens were "serologically characterized," implying prior established diagnostic results for the various pathogens.
  • Clinically Characterized Patients: For the initial assay cut-off determination, samples were from "clinically characterized positive West Nile virus patients" and "normal healthy blood donors," suggesting a clinical diagnosis or healthy status served as the ground truth for defining the cut-off.

8. The Sample Size for the Training Set

  • This is an IVD kit, not a machine learning model, so there isn't a "training set" in the typical AI sense.
  • However, the assay cut-off determination can be seen as analogous to a model's calibration or training phase. For this, 18 sera from clinically characterized WNV positive patients and 150 sera from normal healthy blood donors were used to determine the optimal OD cut-off and borderline range.

9. How the Ground Truth for the Training Set Was Established

  • As noted above, for the assay cut-off determination (analogous to training/calibration):
    • The "positive" ground truth was established by using "clinically characterized positive West Nile virus patients." Further details about how these patients were characterized (e.g., by PRNT, by symptoms, by other diagnostics) are not provided in this document but are assumed to be based on an established clinical diagnosis of WNV infection.
    • The "negative" ground truth was established by using "normal healthy blood donors from a non-endemic region." These individuals are assumed to be free of WNV infection based on their health status and origin from a non-endemic area.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an abstract symbol that resembles a stylized caduceus or a series of interconnected human profiles. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 12, 2016

EUROIMMUN US, Inc. Michael Locke Director of Regulatory Affairs 1 Bloomfield Ave. Mountain Lakes, NJ 07046

Re: K153303

Trade/Device Name: EUROIMUN Anti-West Nile Virus ELISA (IgG) Regulation Number: 21 CFR 866.3940 Regulation Name: West Nile Virus Serological Reagent Regulatory Class: Class II Product Code: NPO Dated: July 8, 2016 Received: July 12, 2016

Dear Mr. Locke:

This letter corrects our substantially equivalent letter of August 10, 2016.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{1}------------------------------------------------

CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Stephen J. Lovell -S for

Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K153303

Device Name

EUROIMMUN Anti-West Nile Virus ELISA (IgG)

Indications for Use (Describe)

The EUROIMMUN Anti-West Nile Virus ELISA (IgG) is intended for the qualitative detection of IgG antibodies to West Nile virus in human serum and plasma (K+EDTA, Li+-heparin). This test is intended as an aid in the presumptive laboratory diagnosis of West Nile virus infection in patients with clinical symptoms consistent with meningitis encephalitis, in conjunction with other laboratory and clinical findings. Positive results must be confirmed by the plaque reduction test (PRNT) or by using the current CDC guidelines for diagnosis of this disease.

The assay characteristics have not been established for testing cord blood, neonates, prenatal screening, and general population screening of patients without symptoms of meningoencephalitis. This assay is not FDA cleared or approved for testing blood or plasma donors.

Warning: Cross-reactivity with IgG to Dengue, Chikungunya, Zika and Tick-borne Encephalitis (TBE) viruses has been observed with the EUROIMMUN Anti-West Nile Virus ELISA (IgG). Reactive results must be reported with a caution statement regarding possible IgG cross-reactivity with other flaviviruses.

Type of Use (Select one or both, as applicable)

∑ Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUBSTANTIAL EQUIVALENCE

510(k) Number

K153303

Applicant

EUROIMMUN US INC. 1 Bloomfield Ave., Mountain Lakes, New Jersey 07046 Phone: 800-913-2022 Fax: 973-656-1098

Contact: Michael Locke Director of Regulatory Affairs & Quality Management 1 Bloomfield Ave., Mountain Lakes, New Jersey 07046 Phone: 800-913-2022 Fax: 973-656-1098

Product Name

EUROIMMUN Anti-West Nile Virus ELISA (IgG)

Device Identification

Regulation: 21 CFR 866.3940 (West Nile virus serological reagents) Classification: Class II Product Code: NOP Panel: Microbiology

Substantial Equivalence Information:

Comparison to Predicate Device

Similarities

Antigen

New DevicePredicate Device
Similarities
ItemEUROIMMUN Anti-West Nile Virus ELISA(IgG) (K153303)Focus Diagnostics West Nile VirusIgG DxSelect™ (K031953)
Intended useDetection of IgG class antibodies against WestNile virusSame
Assay formatQualitativeSame
TechnologyELISASame
Assay platform96-well microtiter platesSame
AntigenCoated on microtiter plateSame
Calibrators and Controls1 calibrator (cut-off)2 controls: 1 positive; 1 negativeSame
ConjugateAnti-human IgG (rabbit) labelled withhorseradish peroxidaseSame
SubstrateTMBSame
Wash buffer10x concentrateSame
Serum sample dilution1:101Same
ProcedureSample incubation with micro-well antigencoated plate, followed by a wash step,incubation with an anti-human IgG enzymeconjugate; wash step, incubation withsubstrate; stopping of the reaction with stopsolution, photometric reading.Same
Differences
ItemNew DevicePredicate Device

| Recombinant, detergent-extracted

| Recombinant West Nile virus

{5}------------------------------------------------

glycoprotein E of West Nile virus from themembrane fraction of human cells,inactivated using high temperatures andgamma radiation; effectiveness ofinactivation tested by cultureantigen
Stop solution0.5 M sulphuric acid1 M sulfuric acid
Reagent preparationAll reagents, calibrator and controls are readyto use, except for the wash buffer.Calibrator and controls requiredilution before use.
Sample matrixSerum or plasma (EDTA, Li-heparin), CSFSerum
Reported resultsRatioIndex
Cut-off levelsRatioResultIndexResult
<0.8negative< 1.30negative
≥0.8 to <1.1borderline≥ 1.30 to < 1.50equivocal
>1.1positive> 1.50positive

Intended Use

The EUROIMMUN Anti-West Nile Virus ELISA (IgM) is intended for the qualitative detection of IgM antibodies to West Nile virus in human serum and plasma (K -EDTA, Li -heparin). This test is intended as an aid in the presumptive laboratory diagnosis of West Nile virus infection in patients with clinical symptoms consistent with meningitis/encephalitis, in conjunction with other laboratory and clinical findings. Positive results must be confirmed by the plaque reduction neutralization test (PRNT) or by using the current CDC guidelines for diagnosis of this disease.

The assay characteristics have not been established for testing cord blood, neonates, prenatal screening, and general population screening of patients without symptoms of meningoencephalitis. This assay is not FDA cleared or approved for testing blood or plasma donors.

Warning: Cross-reactivity with IgG from Dengue, Chikungunya, Zika and Tick-borne Encephalitis (TBE) viruses has been observed with the EUROIMMUN Anti-West Nile Virus ELISA (IgG). Reactive results must be reported with a caution statement regarding possible IgG cross-reactivity with other flaviviruses.

Device Description

Patient samples are diluted 1:101 in sample buffer, 100 ul of each diluted patient sample and pre-diluted controls and the calibrator are added to the antigen coated microtiter wells and incubated for 60 minutes at +37°C. After incubation the microtiter well strips are washed 3 times with wash buffer to remove unbound antibodies and 100 ul of the anti-human IgG enzyme conjugate reagent is added to each microtiter well. After an additional 30-minutes incubation at room temperature, the microtiter wells are again washed 3 times with wash buffer to remove any unbound enzyme conjugate and 100 ul of the chromogen substrate is added. The strips are incubated for 15 minutes at room temperature and 100 ul stop solution is added. The microtiter plates are placed in an ELISA reader and read at a wavelength of 450 nm and a reference wavelength of between 620 nm and 650 nm within 30 minutes.

Test Principle

The test kit contains 12 microtiter strips each with 8 break-off reagent wells coated with West Nile virus antigens. In the first reaction step, diluted patient samples, callbrator and controls are incubated in the wells. Anti-West Nile virus antibodies will bind to the antigens coated in the microtiter wells are washed to remove any unbound proteins and non-specific antibodies. In a second reaction step, rabbit anti-human IgG HRP enzyme conjugate is added to each well. The enzyme conjugate will bind to any wells that have human IgG binding to the West Nile virus antigen. The wells are washed to remove any unbound HRP enzyme conjugate. 3,3,5,5 tetramethylbenzidine (TMB) enzyme substrate is added. If the HRP enzyme is present in the well (positive reaction), the HRP enzyme will react with the TMB substrate and produce a blue color. After an additional incubation time to allow the color development, a stop solution is added which turns the blue color yellow and inhibits further color development to allow for a stable spectrophotometric reading. The test strips are placed in a microplate reader and the optical density of the color is measured. The amount of antigen specific bound antibody is proportional to the color intensity.

{6}------------------------------------------------

Performance Characteristics

Analytical Performance

Repeatability/Reproducibility

Repeatability: The repeatability of the EUROIMMUN Anti-West Nile Virus ELISA (IgG) was investigated by testing of a panel of 7 members prepared using natural patient samples seropositive at different concentrations. The inter-assay repeatability is based on 42 determinations per sample performed in 14 different runs on 7 different days (with 2 runs per day and 3 replicates per run). The data from the repeatability study is presented in the table below.

Repeatability

No.MeanRatioWithin-RunWithin DayBetween DaysTotal
SD%CVSD%CVSD%CVSD%CV
10.10.017.4%0.0113.2%0.001.3%0.0113.3%
20.40.025.3%0.0411.7%0.036.9%0.0513.6%
30.80.034.2%0.079.7%0.000.0%0.079.7%
40.90.033.7%0.0910.4%0.000.0%0.0910.4%
51.20.054.2%0.086.7%0.000.0%0.086.7%
62.20.083.8%0.146.1%0.000.0%0.146.1%
73.70.061.7%0.205.5%0.133.6%0.246.6%
84.10.143.5%0.235.7%0.092.3%0.256.1%

Reproducibility: The reproducibility of the EUROIMMUN Anti-West Nile Virus ELISA (IgG) was investigated by testing of a panel of 7 members prepared using natural patient samples seropositive at different concentrations. The reproducibility is based on 60 determinations per sample performed at 3 different sites (in-house; and 2 laboratories in the north-east) for 5 days with 2 runs per day and 2 replicates per run. The data from the reproducibility study is presented in the table below.

Reproducibility

No.MeanRatioWithin-RunWithin DayBetween DaysBetween SitesTotal
SD%CVSD%CVSD%CVSD%CVSD%CV
10.10.01014.1%0.01115.0%0.0056.4%0.0045.3%0.01216.3%
20.70.07211.0%0.09614.7%0.0000.0%0.0639.7%0.09614.7%
30.80.0628.0%0.0759.8%0.0172.2%0.0435.7%0.07710.1%
40.80.0708.3%0.09811.6%0.0000.0%0.0698.1%0.09811.6%
51.10.0897.8%0.11910.5%0.0000.0%0.0797.0%0.11910.5%
62.20.2119.7%0.2119.7%0.0602.7%0.0100.5%0.21910.1%
74.30.4059.4%0.43610.1%0.0000.0%0.1613.7%0.43610.1%

Analytical Specificity/Cross Reactivity

Cross Reactivity: Cross Reactivity was investigated using 910 serologically characterized seropositive specimens from patients with diseases other than WNV. Each of the specimens included in the study was characterized with respect to disease state prior to analysis of the specimens with the EUROIMMUN Anti-West Nile Virus ELISA (IgG). Cross-reactivity across the flavivirus group is common (i.e., St. Louis encephalitis, Dengue 1, 2, 3 & 4; Murray Valley encephalitis, Japanese encephalitis, Yellow fever viruses and Zika virus); as well as persons vaccinated for flaviviruses.

Cross Reactivity

No.PanelnAnti-West Nile Virus ELISA (IgG)
PositiveNegative% Negative
1Anti-Adenovirus12012100.0%
2Anti-Barmah Forest virus20020100.0%
3Anti-Borrelia burgdorferi5435194.4%
4Anti-Chikungunya virus72234968.1%
5Anti-Chlamydia pneum.12012100.0%
6Anti-CMV12012100.0%

{7}------------------------------------------------

No.PanelnAnti-West Nile Virus ELISA (IgG)
PositiveNegative% Negative
7Anti-Dengue virus5850813.8%
8Anti-EBV6025896.7%
9Anti-Hanta virus11011100.0%
10Anti-Hepatitis virus32032100.0%
11Anti-Helicobacter pylori12012100.0%
12Anti-HSV-13923794.9%
13Anti-Influenza A12012100.0%
14Anti-Influenza B12012100.0%
15Anti-Leptospira11011100.0%
16Malaria/anti-Plasmodium falciparum808100.0%
17Anti-Measles virus12012100.0%
18Anti-Mumps virus12012100.0%
19Anti-Mycoplasma pneumoniae12012100.0%
20Anti-Parainfluenza types 1-412012100.0%
21Anti-Polio virus3713697.3%
22Anti-Ross River virus2011995.0%
23Anti-RSV12012100.0%
24Anti-Rubella virus12012100.0%
25Anti-TBE virus118338572.0%
26Anti-Toxoplasma gondii909100.0%
27Anti-VZV32032100.0%
28Anti-West Nile Virus IgM10010100.0%
29Yellowfever virus immunization12012100.0%
30Anti-Zika virus474700.0%
31Rheumatoid arthritis/polyarthritis/anti-CCP16016100.0%
32Anti-Rheumatoid factor3713697.3%
33Anti-nuclear autoantibodies3313297.0%
34ANCA-associated small vessel vasculitides/ANCA606100.0%
35Celiac disease/anti-endomysium10010100.0%
36Plasma cell myeloma14014100.0%
Total91016474682.0%

Interferences: Hemolytic, lipemic and icteric samples showed no influence on the result up to a concentration of 1000 mg/dl for hemoglobin, 2000 mg/dl for triglycerides and 40 mg/dl for bilirubin in testing with the EUROIMMUN Anti-West Nile Virus ELISA (IgG). Interferences to high protein (albumin), cholesterol, and intralipids were not investigated.

Assay Cut-off:

The recommended assay cut-off is based on OD results of 18 sera from clinically characterized positive West Nile virus patients, and of 150 sera from normal healthy blood donors from a non-endemic region (100 men, 50 women; mean age 39.9 years, age range 18 to 68 years). The samples were investigated using the EUROIMMUN Anti-West Nile Virus ELISA (IgG) and a ROC analysis was performed using the OD's obtained. The ROC analysis demonstrated optimal sensitivity (100.0%) and specificity (100.0%) at the OD value of 0.475. The calibrator was established at this cut-off OD.

The borderline range of ratio 0.8 to ratio 1.1 was established to cover at least 98% of the negative samples (148 of 150 samples) in the negative range.

Using the cut-off of ratio 1.0 and borderline range of ratio 0.8 to 1.1 with the positive and negative groups mentioned above, the EUROIMMUN Anti-West Nile Virus ELISA (IgG) showed a sensitivity of 100.0% (95% C.I.: 81.5 - 100.0%) with a specificity of 98.7% (95% C.I.: 95.3 - 99.8%; if borderline samples counted as positive).

{8}------------------------------------------------

Clinical Study I:

A prospective clinical study was performed with 152 samples from patients suspected of West Nile Virus infection collected at hospitals and clinics across the US in 2015. The panel consisted of 81 men and 71 women, age ranged from 6 to 85 years with a mean age of 49 years. Each specimen was tested at one internal and two external sites with the EUROIMMUN Anti-West Nile Virus ELISA (IgG) in parallel with the predicate assay. Average of three results for each clinical specimen tested at three sites was considered to calculate the positive percent and negative percent agreement between the EUROIMMUN Anti-West Nile Virus ELISA (IgG) vs the predicate assay. The following results were obtained.

Serumn = 152Predicate
PositiveBorderlineNegative
EUROIMMUN Anti- West NileVirus ELISA (IgG)Positive4211
Borderline204
Negative10101
Positive agreement93.3% (42/45)95% C.I.81.7-98.6%
Negative agreement94.4% (101/107)95% C.I.88.2-97.9%

Of the 45 presumptive positives by the predicate device, 27 samples were further tested by PRNT (Plaque Reduction Neutralization Test) and the EUROIMMUN Anti-West Nile Virus ELISA (IgG). The results are shown below.

Serumn = 27PRNT Results
PositiveBorderlineNegative
EUROIMMUN Anti- West NileVirus ELISA (IgG)2500
100
100

Sensitivity 92.6% (25/27) 95% C.I. 75.7-99.1%

Clinical Study II:

A study was performed at a clinical laboratory in the midwest, with 401 serum samples collected prospectively from patients suspected of west nile virus infection. The serum panel consisted of 195 men and 206 women, age ranged from 3 to 102 years with a mean age of 47 years. The samples were tested with the EUROIMMUN Anti-West Nile Virus ELISA (IgG) in parallel with the Predicate ELISA.

Serumn = 152Predicate
PositiveBorderlineNegative
EUROIMMUN Anti- West NileVirus ELISA (IgG)Positive4303
Borderline101
Negative10352
Positive agreement95.6% (43/45)95% C.I.84.9-99.5%
Negative agreement98.9% (352/356)95% C.I.97.1-99.7%

Clinical Study III:

A retrospective clinical study was performed in cooperation with the Robert Koch Institute (RKI), Berlin, Germany with 295 serum samples that included 200 samples from major outbreaks of West Nile fever in South Africa in 1974 and 1984. The samples were tested by the EUROIMMUN Anti-West Nile Virus ELISA (IgG). The results are shown below.

Serumn = 295RKI Anti-West Nile Virus PRNT
PositiveNegative
EUROIMMUN Anti- West NileVirus ELISA (IgG)Positive1943
Borderline00
Negative197

{9}------------------------------------------------

Positive agreement99.5% (194/195)95% C.I.97.2-100.0%
Negative agreement97.0% (97/100)95% C.I.91.5-99.4%

Expected Values

Euroimmun assessed reactivity with 553 samples prospectively collected from US population. The samples consisted of 50% females, and 50% males The range of positivity of different populations from the US prospective studies with the EUROIMMUN Anti-West Nile Virus ELISA (IgG) test kit are presented below.

US Studies

AgenNegativeBorderlinePositive% Positive95% C.I.
0-91616000.0 – 20.6%
10-193230026.3% (2/32)0.8 – 20.8%
20-296659257.6% (5/66)2.5 – 16.8%
30-39867411112.8% (11/86)6.6 – 21.7%
40-49897611213.5% (12/89)7.2 – 22.4%
50-591007921919.0% (19/100)11.8 – 28.1%
60+16412124125.0% (41/164)18.6 – 32.3%
total55345589016.3% (90/455)13.3 – 19.6%

Note: It is recommended that each laboratory determine its own normal range based on the population and equipment used.

Matrix Comparison (Serum vs Plasma)

The usability of plasma was investigated using sample pairs each of serum and corresponding plasma (EDTA. Li-heparin). As no real plasma sample pairs from patients containing anti-West Nile virus antibodies were available, the samples were created from 5 different sets of normal blood donor sample pairs (serum, EDTA plasma, Li-heparin plasma) that were (after drawing) spiked with 5 different positive serum samples. After spiking, the sample sets were further diluted and processed according to the package insert.

Passing-Bablok regression was calculated for the comparison of serum to plasma. The regression equation is near the ideal correlation (intercept 0; slope 1.0) indicating equivalence of concentrations between serum and the corresponding plasma matrices. Coefficients of determination were found to be above 0.975 and % recovery compared to serum was in the range of 92 to 109% (serum = 100%).

EDTA plasmaLi-heparin plasma
Determinations (n)2020
Concentration range (serum)Ratio 0.6 - 4.1Ratio 0.6 - 4.1
Concentration range (plasma)Ratio 0.6 - 4.1Ratio 0.6 - 4.1
Regression equation(y = plasma, x = serum)y = -0.01 + 1.02xy = 0.04 + 0.96x
95% C.I. of intercept-0.05 - 0.02-0.11 - 0.15
95% C.I. of slope0.99 - 1.050.89 - 1.10
Coefficient of determination R20.99830.9759
Mean %recovery101 %100 %
Range of %recovery98 - 105 %92 - 109 %

Matrix Comparison (serum vs plasma)

Conclusion

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

§ 866.3940 West Nile virus serological reagents.

(a)
Identification. West Nile virus serological reagents are devices that consist of antigens and antisera for the detection of anti-West Nile virus IgM antibodies, in human serum, from individuals who have signs and symptoms consistent with viral meningitis/encephalitis. The detection aids in the clinical laboratory diagnosis of viral meningitis/encephalitis caused by West Nile virus.(b)
Classification. Class II (special controls). The special control is FDA's guidance entitled “Class II Special Controls Guidance Document: Serological Reagents for the Laboratory Diagnosis of West Nile Virus.” See § 866.1(e) for the availability of this guidance document.