K Number
K153201
Manufacturer
Date Cleared
2016-06-24

(233 days)

Product Code
Regulation Number
862.1345
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SD GlucoNavii® Mentor BT Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood from fingertip, palm, upper arm, or forearm. SD GlucoNavii® Mentor BT Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. It is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control.

The SD GlucoNavii® Mentor BT Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes. The SD GlucoNavii® Mentor BT Blood Glucose Monitoring System is not for use in neonates. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

SD GlucoNavii® Mentor BT Blood Glucose Test Strips are for use with SD GlucoNavii® Mentor BT Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm.

SD GlucoNavii® Mentor BT Multi Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood from fingertip, palm, upper arm, or forearm. The SD GlucoNavii® Mentor BT Multi Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control program. This system should only be used with auto-disabling, single-use lancing devices.

The SD GlucoNavii® Mentor BT Multi Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes. The SD GlucoNavi® Mentor BT Multi Blood Glucose Monitoring System is not for use in neonates Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

SD GlucoNavii® Mentor BT Multi Blood Glucose Test Strips are for use with SD GlucoNavii® Mentor BT Multi Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm.

Device Description

SD GlucoNavii® Mentor BT and BT Multi Blood Glucose Monitoring Systems are OTC/ Rx blood glucose monitoring systems. The SD GlucoNavii® Mentor BT Blood Glucose Monitoring System is indicated for singlepatient use at home (over-the-counter; OTC) and should not be shared, while the SD GlucoNavii® Mentor BT Multi Blood Glucose Monitoring System is for multi-patient use in a professional healthcare setting(over-thecounter; OTC and prescription; POC), in order to help monitor the effectiveness of diabetes control.

A drop of blood sample from the finger prick works with glucose oxidase and the mediators in the test strip to make a small electric current proportional to the glucose concentration in the blood. The meter reads the current and displays the blood glucose result equivalent to the current.

The user can search the stored results with three presentations of 7, 14 and 30-day averages of test results stored in memory: normal, pre-meal and post-meal state averages. The system can set the beep, hypo warning, date, time, post-meal alarm and alarm. The system can also set the pre-meal mark. Test results are displayed with mg/dL unit. A check strip allows the meter to check a problem and the control solution allows the meter and test strip to be checked.

A Bluetooth unit for data transfer has been added to the subject devices in addition to NFC and USB cable communication features.

The following modifications were made on the subject device from the predicate device (K132929).

    1. The addition of Bluetooth for wireless transmission of data to mobile device or PC, and which comprises the following modifications:
    • i. Changed printed circuit board (PCB) layout.
    • ii. Added Bluetooth module to PCB.
    • iii. Added and deleted electronic components to support the Bluetooth Module.
    • iv. Deleted USB communication circuit.
    • Changed Firmware. V.
    • Added Bluetooth icon to display. vi.
    • Extended automatic meter shutoff times to allow enough time for data transfer via Bluetooth. vii.
    1. Memory capacity for stored glucose test results decreased from 300 to 100 records.
    1. The trade name of the systems has changed from SD GlucoNavii® Mentor NFC Blood Glucose Monitoring System and SD GlucoNavii® Mentor NFC Multi Blood Glucose Monitoring System to SD GlucoNavii® Mentor BT Blood Glucose Monitoring System and SD GlucoNavii® Mentor BT Multi Blood Glucose Monitoring System.
    1. The trade name of the test strips has changed from SD GlucoNavii® Mentor NFC Blood Glucose Test Strips and SD GlucoNavii® Mentor NFC Multi Blood Glucose Test Strips to SD GlucoNavii® Mentor BT Blood Glucose Test Strips and SD GlucoNavii® Mentor BT Multi Blood Glucose Test Strips.
    1. Labeling was modified to reflect the changes to the device.
AI/ML Overview

This document is a 510(k) Premarket Notification from the FDA regarding the SD GlucoNavii® Mentor BT Blood Glucose Monitoring System and SD GlucoNavii® Mentor BT Multi Blood Glucose Monitoring System. It primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting a standalone study of the device's performance against specific acceptance criteria.

Therefore, much of the requested information, particularly regarding detailed study design, sample sizes for test and training sets, expert qualifications, and adjudication methods, is not provided in this document as it pertains to clinical performance which the FDA states is "not applicable."

However, based on the provided text, we can extract the acceptance criteria (implicitly defined by the device's specifications and comparison to the predicate) and the reported device performance for certain aspects, primarily through the "Similarities and Differences" table.

Here's a breakdown of the available information:


1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state "acceptance criteria" in a numerical or statistical format, but rather details the device's specifications, which act as de facto criteria. The reported device performance is largely presented by stating that the modified device's performance characteristics are "identical" to the predicate or within acceptable changes.

Feature / Criteria (Implied Acceptance)Reported Device Performance (SD GlucoNavii® Mentor BT)
ManufacturerIdentical (SD Biosensor, Inc.)
Product CodeIdentical (CGA, NBW)
Indications for Use (Self-testing)Identical (Quantitative measurement of glucose in fresh capillary whole blood from fingertip, palm, upper arm, or forearm; for self-testing by people with diabetes at home as an aid to monitor effectiveness of diabetes control; not for diagnosis/screening or neonates; alternative site testing only during steady-state)
Indications for Use (Multi-patient use)Identical (Quantitative measurement of glucose in fresh capillary whole blood from fingertip, palm, upper arm, or forearm; for multiple-patient use in professional healthcare settings as an aid to monitor effectiveness of diabetes control program; to be used with auto-disabling, single-use lancing devices; not for diagnosis/screening or neonates; alternative site testing only during steady-state)
Test TimeIdentical (5 seconds)
Measuring RangeIdentical (20-600 mg/dL)
Operating TemperatureIdentical (10-45°C (50-113°F) for blood sample / 18-30°C (64-86°F) for control solution)
Operating HumidityIdentical (15-95% RH)
Operating AltitudeIdentical (up to 11,351ft.)
Hematocrit Range20-60%
Memory Capacity100 test results (Differs from predicate's 300, but deemed acceptable)
CodingIdentical (N/A)
Meter DimensionsIdentical (50mm x 93mm x 18mm)
Meter WeightIdentical (50g with battery)
Unit of MeasureIdentical (mg/dL)
Sample TypeIdentical (Fresh capillary whole blood)
Sample SitesIdentical (Fingertip, palm, forearm or upper arm)
Sample VolumeIdentical (0.3 µL)
Monitor TypeIdentical (LCD display)
BacklightIdentical (No)
Power SupplyIdentical (3V CR2032 Battery x1, Replaceable)
Power Saving (Automatic Shut-off)Identical (After 1 minute of inactivity without test strip / 3 minutes with test strip inserted)
Battery LifeIdentical (Approximately 1,000 Tests)
Test Strip TechnologyIdentical (Glucose Oxidase (GOD))
Test PrincipleIdentical (Electrochemical biosensor)
Sample ApplicationIdentical (Test strip capillary draw)
CalibrationIdentical (Plasma-calibrated)
Test Strip Storage TemperatureIdentical (2-32°C (36-90°F))
Test Strip Storage HumidityIdentical (10-95% RH)
PC Link FeatureYes, NFC, or Bluetooth (Differs from predicate's USB Cable/NFC, but deemed acceptable)
Smart Device Link FeatureYes, NFC or Bluetooth (Differs from predicate's NFC, but deemed acceptable)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document explicitly states: "Clinical sensitivity and clinical specificity testing are not applicable." This implies that no clinical performance study involving a patient test set was conducted for this 510(k) submission, as the focus was on demonstrating substantial equivalence based on non-clinical tests (design control, risk analysis, verification & validation activities) due to the nature of the modifications (primarily adding Bluetooth connectivity). Therefore, sample size for a clinical test set and its provenance are not provided.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable, as no clinical test set was used for this submission.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as no clinical test set was used for this submission.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a blood glucose monitoring system, not an AI-assisted diagnostic imaging device that would typically involve human readers and MRMC studies.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is a standalone blood glucose meter. Its performance is evaluated intrinsically, and the "study" demonstrating its performance relies on the device itself against a reference method (not detailed here, but expected for glucose meters). The document indicates "design control, risk analysis and verification & validation activities" were performed, which would include testing the device's accuracy and precision, but the specific details of these tests (e.g., sample size, reference method) are not included in this summary.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For a blood glucose monitoring system, the "ground truth" would typically be established by a laboratory-grade reference method for glucose measurement (e.g., YSI glucose analyzer), which is considered the gold standard. However, the document does not explicitly state the specific reference method used for the verification and validation activities.

8. The sample size for the training set

Not applicable in the context of an AI/algorithm training set. For a traditional medical device like a glucose meter, "training" might refer to calibration data, but the document does not provide details on the sample size for such data.

9. How the ground truth for the training set was established

Not applicable in the context of an AI/algorithm training set. For a glucose meter, the ground truth for calibration (if considered "training") would be established by measuring known glucose concentrations against a reference method.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of three human profiles facing to the right, arranged in a cascading manner. The profiles are depicted in a dark color, creating a silhouette effect. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 24, 2016

SD BIOSENSOR C/O PRISCILLA CHUNG LK CONSULTING GROUP, USA, INC. 2651 E CHAPMAN AVE STE 110 FULLERTON CA 92831

Re: K153201

Trade/Device Name: SD GlucoNavii® Mentor BT Blood Glucose Monitoring System, SD GlucoNavii® Mentor BT Multi Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: NBW, CGA Dated: May 23, 2016 Received: May 26, 2016

Dear Ms. Priscilla Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Katherine Serrano -S

For: Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K153201

Device Name

SD GlucoNavii® Mentor BT Blood Glucose Monitoring System

Indications for Use (Describe)

SD GlucoNavii® Mentor BT Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood from fingertip, palm, upper arm, or forearm. SD GlucoNavii® Mentor BT Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. It is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control.

The SD GlucoNavii® Mentor BT Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes. The SD GlucoNavii® Mentor BT Blood Glucose Monitoring System is not for use in neonates. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

SD GlucoNavii® Mentor BT Blood Glucose Test Strips are for use with SD GlucoNavii® Mentor BT Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K153201

Device Name

SD GlucoNavii® Mentor BT Multi Blood Glucose Monitoring System

Indications for Use (Describe)

SD GlucoNavii® Mentor BT Multi Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood from fingertip, palm, upper arm, or forearm. The SD GlucoNavii® Mentor BT Multi Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control program. This system should only be used with auto-disabling, single-use lancing devices.

The SD GlucoNavii® Mentor BT Multi Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes. The SD GlucoNavi® Mentor BT Multi Blood Glucose Monitoring System is not for use in neonates Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

SD GlucoNavii® Mentor BT Multi Blood Glucose Test Strips are for use with SD GlucoNavii® Mentor BT Multi Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for SD Biosensor, Inc. The logo features an orange abstract symbol resembling a stylized flame or plant on the left. To the right of the symbol, the text "SD BIOSENSOR, INC." is displayed in a simple, sans-serif font. The letters "SD" are larger and in gray, while the rest of the text is in black.

510(k) Summary

(K153201)

This summary of 510(k) information is being submitted in accordance with the requirements of 21 CFR 807.92.

Date of Summary: June 22, 2016

1. SUBMITTER'S IDENTIFICATION:

Manufacturer

SD Biosensor, Inc. C-4th&5th, 16, Deogyeong-Daero,1556beon-Gil, Yeongtong-Gu, Suwon-si, Gyeonggi-Do, KOREA, REPUBLIC OF 443-813 TEL: 82-31-300-0423 FAX: 82-31-300-0497

Contact Person

SunYoung Jeong Email: syj@sdbiosensor.com Official Correspondent (U.S. designated agent) Priscilla Chung (Regulatory Consultant) SD Biosensor, Inc. c/o LK Consulting Group USA, Inc. 2651 E Chapman Ave Ste 110, Fullerton CA 92831 TEL: (714) 202-5789 FAX: (714) 409-3357 E-MAIL: juhee.c@lkconsultinggroup.com

2. DEVICE NAME:

Proprietary Name:SD GlucoNavii® Mentor BT Blood Glucose Monitoring SystemSD GlucoNavii® Mentor BT Multi Blood Glucose Monitoring System
Common Name:Blood Glucose Monitoring System
Regulation Number:21 CFR §862.1345
Classification Name:Blood Glucose Test System
Product Code:NBW
SubsequentProduct Code:CGA
Regulatory Class:II

{5}------------------------------------------------

3. PREDICATE DEVICES:

SD GlucoNavii® Mentor NFC Blood Glucose Monitoring System and SD GlucoNavii® Mentor NFC Multi Blood Glucose Monitoring System (K132929) by SD Biosensor, Inc.

4. DEVICEDESCRIPTION:

SD GlucoNavii® Mentor BT and BT Multi Blood Glucose Monitoring Systems are OTC/ Rx blood glucose monitoring systems. The SD GlucoNavii® Mentor BT Blood Glucose Monitoring System is indicated for singlepatient use at home (over-the-counter; OTC) and should not be shared, while the SD GlucoNavii® Mentor BT Multi Blood Glucose Monitoring System is for multi-patient use in a professional healthcare setting(over-thecounter; OTC and prescription; POC), in order to help monitor the effectiveness of diabetes control.

A drop of blood sample from the finger prick works with glucose oxidase and the mediators in the test strip to make a small electric current proportional to the glucose concentration in the blood. The meter reads the current and displays the blood glucose result equivalent to the current.

The user can search the stored results with three presentations of 7, 14 and 30-day averages of test results stored in memory: normal, pre-meal and post-meal state averages. The system can set the beep, hypo warning, date, time, post-meal alarm and alarm. The system can also set the pre-meal mark. Test results are displayed with mg/dL unit. A check strip allows the meter to check a problem and the control solution allows the meter and test strip to be checked.

A Bluetooth unit for data transfer has been added to the subject devices in addition to NFC and USB cable communication features.

The following modifications were made on the subject device from the predicate device (K132929).

    1. The addition of Bluetooth for wireless transmission of data to mobile device or PC, and which comprises the following modifications:
    • i. Changed printed circuit board (PCB) layout.
    • ii. Added Bluetooth module to PCB.
    • iii. Added and deleted electronic components to support the Bluetooth Module.
    • iv. Deleted USB communication circuit.
    • Changed Firmware. V.
    • Added Bluetooth icon to display. vi.
    • Extended automatic meter shutoff times to allow enough time for data transfer via Bluetooth. vii.
    1. Memory capacity for stored glucose test results decreased from 300 to 100 records.
    1. The trade name of the systems has changed from SD GlucoNavii® Mentor NFC Blood Glucose Monitoring System and SD GlucoNavii® Mentor NFC Multi Blood Glucose Monitoring System to SD GlucoNavii® Mentor BT Blood Glucose Monitoring System and SD GlucoNavii® Mentor BT Multi Blood Glucose Monitoring System.

{6}------------------------------------------------

C-4th&5th, 16, Deogyeong-Daero, 1556beon-Gil, Yeongtong-Gu, Suwon-si, Gyeonggi-Do, KOREA, REPUBLIC OF 443-813

    1. The trade name of the test strips has changed from SD GlucoNavii® Mentor NFC Blood Glucose Test Strips and SD GlucoNavii® Mentor NFC Multi Blood Glucose Test Strips to SD GlucoNavii® Mentor BT Blood Glucose Test Strips and SD GlucoNavii® Mentor BT Multi Blood Glucose Test Strips.
    1. Labeling was modified to reflect the changes to the device.

{7}------------------------------------------------

5. INDICATION FOR USE:

SD GlucoNavii® Mentor BT Blood Glucose Monitoring System

SD GlucoNavii®Mentor BT Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood from fingertip, palm, upper arm, or forearm. SD Glucose Monitoring System is intended to be used by a single person and should not be shared. It is intended for self-testing outside the by people with diabetes at home as an aid to monitor the effectiveness of diabetes ontrol.

The SD Gluccilentor BT Blood Glucose Montoning System should not be used for the diabetes. The SD GlucoNavi™Mentor BT Blood Glucose Monitoring System is not for use in neonates . Alternative site done only during steady-state times (when glucose is not changing rapidly).

SD GlucoNavii®Mentor BT Blood Glucose Test Strips are with SD GlucoNavii®Mentor BT Blood Glucose Meter to quantitatively measure glucose (succee (succee (succee (succee (succ capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm.

SD GlucoNavii® Mentor BT Multi Blood Glucose Monitoring System

SD Glucollari®Mentor BT Multi Blood Glucose Monitative mesurement of glucose (sugaritiative messurement of glucose (sugar) in fresh capillary whole blood from fingertip, palm, upper arm, or foreari® Nentor BT Multi Blood Guccose Monitoring System is intended for testing outside the body in vitro diagnosic use) and is intended for multiple-patings as an aid to monitor the effectivenes of diabetes control program. This system should only be used with auto-disabling, single-use lancing devices.

The SD Gucollari®Mentor BT Multi Blood Goose Montoning System should not be used for the diabets. The SD Gucollari® Nentor BT Multi Blood Glucose Monitoring System is not for use in news in testing should be done only during steady-state times (when glucose is not thanging rapidly).

SD GlucoNavii®Mentor BT Multi Blood Glucose with SD GlucoNavii®Mentor BT Multi Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm.

{8}------------------------------------------------

6. COMPARISION TO PREDICATE DEVICE:

The subject devices are substantially equivali® Mentor NFC and Mentor NFC Multi Blood Glucose Monitoring System, cleared under K132929. Both the subject and predicate devices are the same in intended uses and basic fundamental scientific technology. Please refer to the following similarities and difference comparison chart:

Similarities and Differences
ItemUnmodified device(K132929)Modified(subject) device(K153201)Discussion ofDifferences
Trade NameSD GlucoNavii® Mentor NFC BloodGlucose Monitoring SystemSD GlucoNavii® Mentor NFC MultiBlood Glucose Monitoring SystemSD GlucoNavii® Mentor BT BloodGlucose Monitoring SystemSD GlucoNavii® Mentor BT MultiBlood Glucose MonitoringSystemdifferent
ManufacturerSD Biosensor, Inc.SD Biosensor Inc.identical
Product CodeCGA, NBWCGA, NBWidentical
Indications for UseSD GlucoNavii® Mentor NFC BloodGlucose Monitoring System isintended to be used for thequantitative measurement ofglucose (sugar) in fresh capillarywhole blood from fingertip, palm,upper arm, or forearm. SDGlucoNavii® Mentor NFC BloodGlucose Monitoring System isintended to be used by a singleperson and should not be shared. Itis intended for self-testing outsidethe body (in vitro diagnostic use)by people with diabetes at home asan aid to monitor the effectivenessof diabetes control.The SD GlucoNavii® Mentor NFCBlood Glucose Monitoring Systemshould not be used for thediagnosis of or screening fordiabetes. The SDSD GlucoNavii® Mentor NFC MultiBlood Glucose Monitoring System isintended to be used for thequantitative measurement ofglucose (sugar) in fresh capillarywhole blood from fingertip, palm,upper arm, or forearm. The SDGlucoNavii® Mentor NFC MultiBlood Glucose Monitoring System isintended for testing outside thebody (in vitro diagnostic use) and isintended for multiple-patient use inprofessional healthcare settings asan aid to monitor the effectivenessof diabetes control program. Thissystem should only be used withauto-disabling, single-use lancingdevices.The SD GlucoNavii® Mentor NFCMulti Blood Glucose MonitoringSystem should not be used for theidentical
Similarities and Differences
ItemUnmodified device(K132929)Modified(subject) device(K153201)Discussion ofDifferences
GlucoNavii®Mentor NFC BloodGlucose Monitoring System is notfor use in neonates. Alternative sitetesting should be done only duringsteady-state times (when glucose isnot changing rapidly). SDGlucoNavii® Mentor NFC BloodGlucose Monitoring System isintended to be used to transmitglucose values to compatiblemobile application or PC softwarethrough use of radio frequencycommunication.SD GlucoNavii® Mentor NFC BloodGlucose Test Strips are for use withSD GlucoNavii® Mentor NFC BloodGlucose Meter to quantitativelymeasure glucose (sugar) in freshcapillary whole blood samplesdrawn from the fingertips, forearm,upper arm, or palm.GlucoNavii®Mentor BT BloodGlucose Monitoring System is notfor use in neonates. Alternative sitetesting should be done only duringsteady-state times (when glucose isnot changing rapidly).SD GlucoNavii® Mentor BT BloodGlucose Test Strips are for use withSD GlucoNavii® Mentor BT BloodGlucose Meter to quantitativelymeasure glucose (sugar) in freshcapillary whole blood samplesdrawn from the fingertips, forearm,upper arm, or palm.diagnosis of or screening fordiabetes. The SD GlucoNavii®Mentor NFC Multi Blood GlucoseMonitoring System is not for use inneonates Alternative site testingshould be done only during steady-state times (when glucose is notchanging rapidly). SD GlucoNavii®Mentor NFC Multi Blood GlucoseMonitoring System is intended tobe used to transmit glucose valuesto compatible mobile application orPC software through use of radiofrequency communication.SD GlucoNavii® Mentor NFC MultiBlood Glucose Test Strips are foruse with SD GlucoNavii® MentorNFC Multi Blood Glucose Meter toquantitatively measure glucose(sugar) in fresh capillary wholeblood samples drawn from thefingertips, forearm, upper arm, orpalm.diagnosis of or screening fordiabetes. The SD GlucoNavii®Mentor BT Multi Blood GlucoseMonitoring System is not for use inneonates Alternative site testingshould be done only during steady-state times (when glucose is notchanging rapidly).SD GlucoNavii® Mentor BT MultiBlood Glucose Test Strips are foruse with SD GlucoNavii® Mentor BTMulti Blood Glucose Meter toquantitatively measure glucose(sugar) in fresh capillary wholeblood samples drawn from thefingertips, forearm, upper arm, orpalm.
Test Time5 secondsidentical
Measuring Range20-600 mg/dLidentical
Operating Temperature10-45°C (50-113°F) for blood sample / 18-30°C (64-86°F) for control solutionidentical
Operating Humidity15-95% RHidentical
Operating Altitudeup to 11,351ft.identical
Hematocrit20-60%
Memory Capacity300 test results100 test resultsdifferent
CodingN/Aidentical
Meter Dimensions50mm x 93mm x 18mmidentical
Meter Weight50g with batteryidentical
Similarities and Differences
ItemUnmodified device(K132929)Modified(subject) device(K153201)Discussion ofDifferences
Unit of measuremg/dLmg/dLidentical
Sample typeFresh capillary whole bloodFresh capillary whole bloodidentical
Sample sitesFingertip, palm, forearm or upper armFingertip, palm, forearm or upper armidentical
Sample volume0.3 µL0.3 µLidentical
MonitorLCD displayLCD displayidentical
BacklightNoNoidentical
Power Supply3V CR2032 Battery x1(Replaceable)3V CR2032 Battery x1(Replaceable)identical
Power SavingAutomatic shut off after 1 minute of inactivity WITHOUT inserting test strip / after 3 minutes of inactivity WITH test strip insertedAutomatic shut off after 1 minute of inactivity WITHOUT inserting test strip / after 3 minutes of inactivity WITH test strip insertedidentical
Battery LifeApproximately 1,000 TestsApproximately 1,000 Testsidentical
Test Strip TechnologyGlucose Oxidase (GOD)Glucose Oxidase (GOD)identical
Test PrincipleElectrochemical biosensorElectrochemical biosensoridentical
Sample ApplicationTest strip capillary drawTest strip capillary drawidentical
CalibrationPlasma-calibratedPlasma-calibratedidentical
Test Strip StorageTemperature2-32°C(36-90°F)2-32°C(36-90°F)identical
Test Strip StorageHumidity10-95%RH10-95%RHidentical
PC link FeatureYes, USB Cable or NFCYes, NFC, or Bluetoothdifferent
Smart device link FeatureYes, NFCYes, NFC or Bluetoothdifferent

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image contains the logo for SD Biosensor, Inc. The logo consists of an orange abstract symbol resembling a stylized "Y" on the left. To the right of the symbol are the letters "SD" in gray, followed by the company name "SD BIOSENSOR, INC." in black.

C-4th&5th, 16, Deogyeong-Daero, 1556beon-Gil, Yeongtong-Gu, Suwon-si, Gyeonggi-Do, KOREA, REPUBLIC OF 443-813

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for SD Biosensor, Inc. The logo consists of an orange abstract symbol resembling a stylized "Y" shape, followed by the letters "SD" in gray. To the right of "SD" is the text "SD BIOSENSOR, INC." also in gray. The overall design is clean and corporate.

C-4th&5th, 16, Deogyeong-Daero, 1556beon-Gil, Yeongtong-Gu, Suwon-si, Gyeonggi-Do, KOREA, REPUBLIC OF 443-813

{11}------------------------------------------------

DR, INC.

C-4th&5th, 16, Deogyeong-Daero, 1556beon-Gil, Yeongtong-Gu, Suwon-si, Gyeonggi-Do, KOREA, REPUBLIC OF 443-813

7. DISCUSSION OF NON-CLINICAL TESTS PERFORMED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE ARE AS FOLLOWS:

There are a number of differences between the unmodified devices and the modified devices as presented in the table above in section 6, however, the differences were evaluated through design control, risk analysis and verification & validation activities, and test results demonstrated that the differences do not raise a question of safety and effectiveness. Based on the information provided in this submission, we conclude that the modified device is substantially equivalent to the predicate device.

8. DISCUSSION OF CLINICAL TESTS PERFORMED:

Clinical sensitivity and clinical specificity testing are not applicable.

9. CONCLUSION:

Based on documentation supplied with this submission, conclusions drawn from design control, risk and verification & validation activities demonstrate that the subject devices are substantially equivalent to the predicate devices.

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.