(249 days)
Lumipulse G TP-N Immunoreaction Cartridges Set For in vitro diagnostic use. WARNING: Lumipulse G TP-N is not intended for blood and tissue donor screening. United States federal law restricts this device to sale by or on the order of a physician. Lumipulse G TP-N is a Chemiluminescent Enzyme Immunoassay (CLEIA) for the qualitative determination of antibodies (1gG and 1gM) to Treponema pallidum in human serum and plasma (sodium citrate, or dipotassium EDTA) on the LUMIPULSE & System. Lumipulse G TP-N can be used as an initial diagnostic test or in conjunction with a nontreponemal laboratory test and clinical findings to aid in the diagnosis of syphilis infection.
The Lumipulse G TP-N is an assay system, including a set of immunoassay reagents, for the qualitative detection of anti-TP antibodies (IgG and IgM) in specimens based on CLEIA technology by a two-step sandwich immunoassay method on the LUMIPULSE G System.
Here's an analysis of the acceptance criteria and study detailed in the provided document:
Device: Lumipulse G TP-N Immunoreaction Cartridges Set (for qualitative determination of antibodies to Treponema pallidum)
Indications for Use: For in vitro diagnostic use. Can be used as an initial diagnostic test or in conjunction with a nontreponemal laboratory test and clinical findings to aid in the diagnosis of syphilis infection.
1. Table of Acceptance Criteria and Reported Device Performance
The document doesn't explicitly state "acceptance criteria" in a numerical, pre-defined format for PPA/NPA. Instead, it presents the performance results obtained and implicitly compares them to the predicate device's expected performance and clinical utility. For qualitative assays like this, the relevant performance metrics are Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA).
| Criteria / Performance Metric | Reported Device Performance (Lumipulse G TP-N) - Prospective Samples | Reported Device Performance (Lumipulse G TP-N) - Retrospective Samples |
|---|---|---|
| Positive Percent Agreement (PPA) vs. Final Comparator Result | 92.7% (95% CI: 88.6%, 95.4%) | 94.3% (95% CI: 92.0%, 96.0%) |
| Negative Percent Agreement (NPA) vs. Final Comparator Result | 99.6% (95% CI: 99.0%, 99.9%) | 98.1% (95% CI: 97.0%, 98.8%) |
Note: The document states, "The results of these analytical (nonclinical) and clinical studies demonstrate that the performance of the Lumipulse G TP-N assay is substantially equivalent to the performance of the ADVIA Centaur Syphilis (SYPH) assay." This implies that the observed PPA and NPA values were considered acceptable for demonstrating substantial equivalence to the predicate device.
2. Sample Size Used for the Test Set and Data Provenance
- Total Test Set Sample Size: 2,791 subjects were submitted for testing. After exclusions, 1,290 evaluable prospective samples and 1,472 evaluable retrospective samples were used.
- Data Provenance:
- Prospective Samples: Collected sequentially from patients prescribed a laboratory test for syphilis during a defined period. Collected from 7 sites representing different geographical regions of the US, including both low-prevalence and high-prevalence sites.
- Retrospective Samples: Pre-selected (enriched population) from various sources:
- 379 pregnant women (250 without syphilis, 129 with syphilis)
- 520 HIV positive subjects (298 remnant samples from reference laboratories, 222 collected at a research facility)
- 130 known to be T. pallidum (TP)-reactive by previous laboratory testing
- 68 samples from a research facility from patients clinically diagnosed with syphilis
- 375 samples consisting of remnants of specimens for routine syphilis testing
- 289 samples from subjects with well-characterized medically diagnosed syphilis (from Argentina (52%) and Florida (48%))
- 474 samples from apparently healthy subjects (75 pediatric, 399 adult/not pregnant).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The ground truth for the test set was not established by individual experts reviewing cases. Instead, it was based on an algorithmic combination of results from other FDA-cleared laboratory tests. Therefore, information regarding "number of experts" and their "qualifications" is not applicable in the traditional sense of image or clinical interpretation.
4. Adjudication Method for the Test Set
The adjudication method was a "2 out of 3 rule" based on the results of three FDA-cleared tests:
- A treponemal test (EIA)
- A non-treponemal Rapid Plasma Reagin (RPR) test
- A second treponemal test, Treponema pallidum particle agglutination (TP-PA)
The "Final Comparator Result" was determined as Positive or Negative if at least two of the three comparator tests yielded a consistent result.
- Example: Positive EIA + Positive RPR + Positive TP-PA = Positive Final Comparator Result.
- Example: Negative EIA + Negative RPR + Negative TP-PA = Negative Final Comparator Result.
- Example: Negative EIA + Positive RPR + Positive TP-PA = Positive Final Comparator Result.
- Indeterminate Results: Cases were considered "Indeterminate" if the EIA was "equivocal" and the TP-PA was "inconclusive". These were excluded from the final analysis.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. This device is an in-vitro diagnostic assay for qualitative determination of antibodies. Its performance is evaluated against a pre-established "ground truth" derived from other laboratory tests, not by comparing human reader performance with and without AI assistance in interpreting diagnostic images or clinical data.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
Yes, a standalone performance study was done. The entire clinical study presented (comparison against the "Final Comparator Result") represents the standalone performance of the Lumipulse G TP-N assay, as it is an automated laboratory test. There is no human interpretation or "in-the-loop" component in the direct output of this specific device.
7. The Type of Ground Truth Used
The ground truth used was a "Reference Comparator Algorithm" based on the consensus results of three established, FDA-cleared laboratory tests (EIA, RPR, and TP-PA), using a "2 out of 3 rule."
8. The Sample Size for the Training Set
The document describes performance characteristics and clinical comparison studies for the Lumipulse G TP-N. It does not explicitly mention a "training set" in the context of machine learning. For in-vitro diagnostic assays like this, the development typically involves:
- Internal studies for analytical performance (precision, interference, cross-reactivity).
- Clinical validation studies for method comparison to established comparator methods or reference standards.
The closest analogue to initial "training" or development data might be mentioned in the section on "Assay cut-off": "Lumipulse G TP-N was developed in 1997. 312 negative specimens, 99 positive specimens, and 6 intermediate specimens previously assigned using TP-PA were measured with Lumipulse G TP-N in order to establish its cut-off index."
- If this is considered the "training set" for cutoff establishment: 417 samples (312 negative, 99 positive, 6 intermediate).
9. How the Ground Truth for the Training Set Was Established
For the establishment of the assay cut-off, the "ground truth" for the 417 samples (312 negative, 99 positive, 6 intermediate) was established using TP-PA (Treponema pallidum particle agglutination), which is described as a previous method for assigning positivity/negativity.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are white against a dark background. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 5, 2016
Diana Dickson Director, Clinical and Regulatory Science Fujirebio Diagnostics, Inc. 201 Great Valley Parkway Malvern, PA 19355
Re: K153145
Trade/Device Name: LUMIPULSE G TP-N Immunoreaction Cartridges Set Regulation Number: 21 CFR 866.3830 Regulation Name: Treponema pallidum treponemal test reagents Regulatory Class: Class II Product Code: LIP Dated: May 31, 2016 Received: June 1, 2016
Dear Ms. Dickson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Steven R. Gitterman -S
for Uwe Scherf. M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K153145
Device Name
Lumipulse G TP-N Immunoreaction Cartridges set
Indications for Use (Describe) Lumipulse G TP-N Immunoreaction Cartridges Set
For in vitro diagnostic use.
WARNING: Lumipulse G TP-N is not intended for blood and tissue donor screening. United States federal law restricts this device to sale by or on the order of a physician.
Lumipulse G TP-N is a Chemiluminescent Enzyme Immunoassay (CLEIA) for the qualitative determination of antibodies (1gG and 1gM) to Treponema pallidum in human serum and plasma (sodium citrate, or dipotassium EDTA) on the LUMIPULSE & System. Lumipulse G TP-N can be used as an initial diagnostic test or in conjunction with a nontreponemal laboratory test and clinical findings to aid in the diagnosis of syphilis infection.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is the logo for Fujirebio Diagnostics, Inc. The logo features a stylized blue and white graphic on the left, resembling a water droplet or a stylized flower. To the right of the graphic is the company name, "FUJIREBIO" in large, bold, blue letters, with the words "Diagnostics, Inc." in smaller, gray letters underneath.
Section 5 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
A. 510(k) Number:
B. Purpose for Submission:
New device
C. Measurand:
IgG and IgM antibodies to Treponema pallidum (T. pallidum)
D. Type of Test:
Qualitative assay, automated chemiluminescent enzyme immunoassay (CLEIA)
E. Applicant:
| Address: | Fujirebio Diagnostics, Inc.201 Great Valley ParkwayMalvern, PA 19355 |
|---|---|
| Contact person: | Stacey Dolan(610) 240-3843 |
Summary preparation date: June 28, 2016
F. Proprietary and Established Names:
Lumipulse G TP-N Immunoreaction Cartridges Set
G. Regulatory Information:
-
- Regulation section: 21 CFR § 866.3830, Treponema pallidum treponemal test reagents
dolans@fdi.com
- Regulation section: 21 CFR § 866.3830, Treponema pallidum treponemal test reagents
-
- Classification: Class II
-
- Product code: LIP, Enzyme Linked Immunoabsorption Assay, Treponema Pallidum
-
- Panel: 83, Microbiology
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized graphic on the left, featuring blue and green shapes. To the right of the graphic, the word "FUJIREBIO" is written in large, bold, blue letters. Below "FUJIREBIO", the words "Diagnostics, Inc." are written in smaller, black letters.
H. Intended Use:
-
- Intended use(s): See indications for use below.
-
- Indication(s) for use:
Lumipulse G TP-N Immunoreaction Cartridges Set
For in vitro diagnostic use.
WARNING: LUMIPULSE G TP-N is not intended for blood and tissue donor screening. United States federal law restricts this device to sale by or on the order of a physician.
Lumipulse G TP-N is a Chemiluminescent Enzyme Immunoassay (CLEIA) for the qualitative determination of antibodies (IgG and IgM) to Treponema pallidum in human serum and plasma (sodium citrate or dipotassium EDTA) on the LUMIPULSE G System. Lumipulse G TP-N can be used as an initial diagnostic test or in conjunction with a nontreponemal laboratory test and clinical findings to aid in the diagnosis of syphilis infection.
-
- Special conditions for use statement(s): Prescription use only
-
- Special instrument requirements: LUMIPULSE G1200 System
. Device Description:
The Lumipulse G TP-N is an assay system, including a set of immunoassay reagents, for the qualitative detection of anti-TP antibodies (IgG and IgM) in specimens based on CLEIA technology by a two-step sandwich immunoassay method on the LUMIPULSE G System.
Lumipulse G TP-N Immunoreaction Cartridges Set REF 235041 The Lumipulse G TP-N Immunoreaction Cartridges Set contains the following:
-
- Lumipulse G TP-N Immunoreaction Cartridges 3 x 14 tests
Antibody-Coated Particle Solution (Liquid when used, 250 µL/Immunoreaction Cartridge) Contains 75 ug/mL recombinant TP antigen (Tp15-17)-coated particles, 75 ug/mL recombinant TP antigen (TpN47)-coated particles, protein stabilizers (bovine) and chemical stabilizers in 0.15 M sodium chloride/Tris buffer. This solution contains gelatin and turns into gel at 15 °C or lower. Preservative: sodium azide.
- Lumipulse G TP-N Immunoreaction Cartridges 3 x 14 tests
Enzyme-Labeled Antigen Solution (Liquid, 350 µL/Immunoreaction Cartridge) Contains 0.075 µg/mL alkaline phosphatase (ALP: calf) labeled recombinant TP antigen (Tp15-17), 0.075 µg/mL ALP (calf)-labeled recombinant TP antigen
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for Fujirebio Diagnostics, Inc. On the left side of the logo is a graphic of three overlapping shapes in blue, white, and green. To the right of the graphic is the company name in blue, with "Diagnostics, Inc." in a smaller font size and in black.
(TpN47) and protein stabilizers (bovine and calf) and chemical stabilizers in 0.1 M sodium chloride/Tris buffer. Preservative: sodium azide.
- Lumipulse G TP-N Calibrators CAL Liquid, 1 x 2 Concentrations 2.
Image /page/5/Picture/4 description: The image shows two lines of text describing TP calibrators. The first line says "CAL N TP calibrator-N (1 x 2.0 mL)". The second line says "CAL P TP calibrator-P (1 x 2.0 mL)".
TP calibrator-P (1 x 2.0 mL)
Contains 0.15 M sodium chloride in Tris buffer with protein stabilizer (bovine). The material of TP calibrator-P is processed from anti-TP positive but inactivated human serum. Preservative: sodium azide.
J. Substantial Equivalence Information:
-
- Predicate device name(s): ADVIA Centaur Syphilis (SYPH) Assay
-
- Predicate 510(k) number(s): K112343
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a blue and green abstract symbol on the left, followed by the company name in blue, bold letters. Below the company name, the words "Diagnostics, Inc." are written in a smaller, black font.
3. Comparison with predicate:
| Similarities | ||
|---|---|---|
| Lumipulse G TP-N(K153145) | ADVIA Centaur Syphilis(SYPH) Assay(Predicate Device)K112343 | |
| Device TypeClassification | In vitro diagnosticClass II | In vitro diagnosticClass II |
| Regulation Number | 21CFR § 866.3830Treponema pallidumtreponemal test reagents | 21CFR § 866.3830Treponema pallidumtreponemal test reagents |
| Product Usage | Clinical and Hospitallaboratories | Clinical and Hospitallaboratories |
| Intended Use | Lumipulse G TP-N is aChemiluminescent EnzymeImmunoassay (CLEIA) for thequalitative determination ofantibodies (IgG and IgM) toTreponema pallidum inhuman serum and plasma(sodium citrate, ordipotassium EDTA) on theLUMIPULSE GSystem. Lumipulse G TP-Ncan be used as an initialdiagnostic test or inconjunction with anontreponemal laboratory testand clinical findings to aid inthe diagnosis of syphilisinfection. | The ADVIA Centaur Syphilis(SYPH) assay is an in-vitrodiagnostics immunoassay forthe qualitative determinationof antibodies to Treponemapallidum in human serum orplasma (EDTA, lithium orsodium heparinized, citrate)using the ADVIA Centaur®and ADVIA Centaur® XPsystems as an aid in thediagnosis of syphilis. |
| Black box warning(PRECAUTION) | WARNING: LUMIPULSE GTP-N is not intended for bloodand tissue donor screening.United States federal lawrestricts this device to sale byor on the order of a physician. | WARNING. The ADVIACentaur SYPH assay is notintended for blood and tissuedonor screening.United States federal lawrestricts this device to sale byor on the order of a physician. |
| Assay Type | Direct sandwichimmunoassay based onchemiluminescenttechnology | Direct sandwichimmunoassay based onchemiluminescenttechnology |
| Specimen CollectionMethod | Routine PhlebotomyTechniques | Routine PhlebotomyTechniques |
| Differences | ||
| Lumipulse G TP-N(K153145) | ADVIA Centaur Syphilis(SYPH) Assay(Predicate Device)K112343 | |
| Instrument System | LUMIPULSE G System | ADVIA Centaur |
| Principle of Operation | AutomatedChemiluminescentMicroparticle Immunoassay(CMIA) | Chemiluminescence EnzymeImmunoassay(CLEIA) |
| Sample Volume | 60 µL | 100 µL |
| Capture/DetectionAntigen/Antibody | Recombinant antigensTpN15-17 and TpN47 onmicroparticles andrecombinant antigens TpN15-17 and TpN47 alkalinephosphatase conjugates | Recombinant antigensTpN17 and TpN15 as biotinconjugates and recombinantantigens TpN17 and TpN15as acridinium esterconjugates |
| Calibrators | 2 calibrators - anti-TP positivehuman plasma is used in theLumipulse G TP-N positivecalibrator and delipidizednormal human sera are usedin the Lumipulse G TP-Nnegative calibrator. | 2 calibrators - liquid inhuman plasma |
| Cut-off | < 1.0 Non-reactive≥ 1.0 Reactive | < 0.9 Non-reactive≥ 0.9 to < 1.1 Equivocal≥ 1.1 Reactive |
| Type of Specimen | Human serum or plasma(EDTA or sodium citrate) | Human serum or plasma(EDTA, lithium or sodiumheparinized, citrate) |
| Analyte Detected | IgG and IgM antibodies toTreponema pallidum | IgG antibodies to Treponemapallidum |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image contains the logo for Fujirebio Diagnostics, Inc. The logo features a stylized graphic on the left, consisting of blue and white shapes with green vertical lines. To the right of the graphic is the company name, "FUJIREBIO" in large, bold, blue letters. Below the company name is the text "Diagnostics, Inc." in a smaller, dark font.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image contains the logo for Fujirebio Diagnostics, Inc. On the left is a blue and green graphic design. To the right of the graphic is the company name in blue, with "FUJIREBIO" on the top line and "Diagnostics, Inc." on the second line.
K. Standard/Guidance Document Referenced (if applicable):
- . ISO 17511:2003 Measurement of Quantities in Biological Samples - Metrological Traceability of Values Assigned to Calibrator and Control Materials
- CLSI EP5-A3 - Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - Third Edition
- CLSI EP7-A2 Interference Testing in Clinical Chemistry; Approved Guideline-Second ● Edition
- CLSI EP28-A3c Defining, Establishing, and Verifying Reference Intervals in the Clinical . Laboratory; Approved Guideline-Third Edition
- CLSI EP12-A2 User Protocol for Evaluation of Qualitative Test Performance Second ● Edition
- Guidance on Informed Consent for In Vitro Diagnostic Device Studies Using Leftover Human Specimens that are Not Individually Identifiable - Guidance for Sponsors, Institutional Review Boards, Clinical Investigators and FDA Staff
L. Test Principle:
The Lumipulse G TP-N is an assay system, including a set of immunoassay reagents, for the qualitative detection of anti-TP antibodies (IgG and IgM) in specimens based on CLEIA technology by a two-step sandwich immunoassay method on the LUMIPULSE G System.
Anti-TP antibodies in specimens (IgG and IgM) specifically binds to recombinant TP antigens (Tp15-17 and TpN47) on the particles, and antigen-antibody complexes are formed. The particles are washed and rinsed to remove unbound materials. Alkaline phosphatase (ALP: calf)-labeled recombinant TP antigens (Tp15-17 and TpN47) specifically bind to anti-TP antibodies of the immunocomplexes on the particles, and additional immunocomplexes are formed. The particles are washed and rinsed to remove unbound materials. Substrate Solution is added and mixed with the particles. AMPPD contained in the Substrate Solution is dephosphorylated by the catalysis of ALP indirectly conjugated to particles. Luminescence (at a maximum wavelength of 477 nm) is generated by the cleavage reaction of dephosphorylated AMPPD. The luminescent signal reflects the amount of anti-TP antibodies.
*AMPPD: 3-(2'-spiroadamantane)-4-methoxy-4-(3"-phosphoryloxy) phenyl-1, 2-dioxetane disodium salt
M. Performance Characteristics (if/when applicable):
Data were generated using the LUMIPULSE G1200 System.
1. Analytical performance:
- a. Precision/Reproducibility:
The precision of the Lumipulse G TP-N was evaluated in a study at one internal site conducted according to the Clinical and Laboratory Standards Institute (CLSI) Protocol EP05-A3. One human sodium citrate-based sample (specimen pool) and five human serum-
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image contains the logo for Fujirebio Diagnostics, Inc. On the left side of the logo is a blue and green abstract design. To the right of the design is the company name in blue, with "FUJIREBIO" on the top line and "Diagnostics, Inc." on the bottom line.
based samples (specimen pools), two reactive controls and one nonreactive control were assayed in replicates of two at two separate times of the day for 20 days (n=80 for each sample) using one LUMIPULSE G1200 system. Data from this study are presented below.
| Sample | Mean(C.O.I.) | Within-run(Repeatability) | Between Run | Between-Day | Within-Laboratory*(Total) | ||||
|---|---|---|---|---|---|---|---|---|---|
| SD | CV(%) | SD | CV(%) | SD | CV(%) | SD | CV(%) | ||
| Sodium CitratePlasma(moderate positive) | 2.2 | 0.042 | 1.9 | 0.011 | 0.5 | 0.030 | 1.4 | 0.053 | 2.4 |
| Serum 1(low negative) | 0.2 | 0.000 | 0.0 | 0.000 | 0.0 | 0.000 | 0.0 | 0.000 | 0.0 |
| Serum 2(high negative) | 0.8 | 0.043 | 5.3 | 0.000 | 0.0 | 0.000 | 0.0 | 0.043 | 5.3 |
| Serum 3(low positive) | 1.2 | 0.037 | 3.0 | 0.000 | 0.0 | 0.013 | 1.1 | 0.039 | 3.2 |
| Serum 4(high positive) | 7.9 | 0.117 | 1.5 | 0.060 | 0.8 | 0.026 | 0.3 | 0.134 | 1.7 |
| Serum 5(moderate positive) | 2.2 | 0.047 | 2.2 | 0.016 | 0.7 | 0.005 | 0.3 | 0.050 | 2.3 |
| ReactiveControl 1 | 2.9 | 0.049 | 1.7 | 0.022 | 0.8 | 0.013 | 0.4 | 0.055 | 1.9 |
| ReactiveControl 2 | 18.5 | 0.299 | 1.6 | 0.179 | 1.0 | 0.071 | 0.4 | 0.355 | 1.9 |
| NonreactiveControl | 0.1 | 0.000 | 0.0 | 0.000 | 0.0 | 0.000 | 0.0 | 0.000 | 0.0 |
*Within-laboratory precision includes within-run, between-run and between-day components.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a blue and green abstract symbol on the left, followed by the company name in blue text. The word "FUJIREBIO" is in a larger, bolder font than "Diagnostics, Inc."
Reproducibility: The reproducibility of the Lumipulse G TP-N assay was evaluated at 3 laboratory sites. One human sodium citrate-based sample (specimen pool) and five human serum-based samples (specimen pools), two reactive controls and one nonreactive control were assayed in replicates of two at two separate times of the day at each of the sites for 10 days (n=120 for each sample) using one LUMIPULSE G1200 System at each site. Data from this study are presented below.
| Sample | Mean(C.O.I.) | Within-Run(Repeatability) | Between-Run | Between-Day | Between-Site | Reproducibility*(Total) | |||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| SD | CV(%) | SD | CV(%) | SD | SD | SD | CV(%) | SD | CV(%) | ||
| Sodium CitratePlasma(moderate positive) | 2.1 | 0.046 | 2.1 | 0.030 | 1.4 | 0.037 | 1.7 | 0.054 | 2.5 | 0.085 | 4.1 |
| Serum 1(low negative) | 0.2 | 0.000 | 0.0 | 0.000 | 0.0 | 0.000 | 0.0 | 0.000 | 0.0 | 0.000 | 0.0 |
| Serum 2(high negative) | 0.8 | 0.035 | 4.3 | 0.016 | 1.9 | 0.020 | 2.4 | 0.011 | 1.4 | 0.045 | 5.6 |
| Serum 3(low positive) | 1.2 | 0.033 | 2.7 | 0.027 | 2.3 | 0.038 | 3.2 | 0.012 | 1.0 | 0.058 | 4.9 |
| Serum 4(high positive) | 7.7 | 0.125 | 1.6 | 0.172 | 2.2 | 0.106 | 1.4 | 0.252 | 3.3 | 0.346 | 4.5 |
| Serum 5(moderate positive) | 2.2 | 0.044 | 2.0 | 0.039 | 1.8 | 0.038 | 1.8 | 0.048 | 2.2 | 0.085 | 3.9 |
| ReactiveControl 1 | 2.8 | 0.061 | 2.2 | 0.038 | 1.3 | 0.038 | 1.4 | 0.039 | 1.4 | 0.090 | 3.2 |
| ReactiveControl 2 | 18.3 | 0.268 | 1.5 | 0.143 | 0.8 | 0.224 | 1.2 | 0.510 | 2.8 | 0.634 | 3.5 |
| NonreactiveControl | 0.1 | 0.000 | 0.0 | 0.000 | 0.0 | 0.000 | 0.0 | 0.000 | 0.0 | 0.000 | 0.0 |
*Reproducibility includes within-run, between-run, between-site components.
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image is a logo for Fujirebio Diagnostics, Inc. The logo consists of a blue and white abstract shape on the left, followed by the company name in blue text. The word "FUJIREBIO" is in a larger, bolder font than "Diagnostics, Inc."
The lot-to-lot precision of the Lumipulse G TP-N assay was evaluated using 3 lots of Lumipulse GTP-N immunoreaction cartridges and calibrators. The above human samples (specimen pools) and three controls were assayed in replicates of two at two separate times of the day for each of the lots for 10 days (n=120 for each sample) using one LUMIPULSE G1200 System. Data from this study are presented below. The %CV for between-lot imprecision was ≤5.7% for any of the 9 samples.
| Sample | Mean(C.O.I.) | Within-Run(Repeatability) | Between-Run | Between-Day | Between-Lot | Within-Laboratory*(Total) | |||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| SD | CV (%) | SD | CV(%) | SD | CV(%) | SD | CV(%) | SD | CV(%) | ||
| Sodium CitratePlasma(moderate positive) | 2.2 | 0.049 | 2.2 | 0.018 | 0.8 | 0.059 | 2.7 | 0.057 | 2.6 | 0.097 | 4.4 |
| Serum 1(low negative) | 0.2 | 0.000 | 0.0 | 0.000 | 0.0 | 0.000 | 0.0 | 0.000 | 0.0 | 0.000 | 0.0 |
| Serum 2(high negative) | 0.8 | 0.026 | 3.1 | 0.000 | 0.0 | 0.024 | 2.9 | 0.041 | 4.9 | 0.054 | 6.8 |
| Serum 3(low positive) | 1.2 | 0.034 | 2.8 | 0.018 | 1.5 | 0.029 | 2.4 | 0.062 | 5.0 | 0.079 | 6.5 |
| Serum 4(high positive) | 7.9 | 0.109 | 1.4 | 0.120 | 1.5 | 0.144 | 1.8 | 0.344 | 4.3 | 0.407 | 5.1 |
| Serum 5(moderate positive) | 2.2 | 0.051 | 2.3 | 0.020 | 0.9 | 0.044 | 2.0 | 0.124 | 5.7 | 0.143 | 6.5 |
| ReactiveControl 1 | 2.9 | 0.049 | 1.7 | 0.047 | 1.6 | 0.067 | 2.3 | 0.117 | 4.0 | 0.151 | 5.2 |
| ReactiveControl 2 | 19.3 | 0.252 | 1.3 | 0.138 | 0.7 | 0.302 | 1.6 | 0.824 | 4.3 | 0.923 | 4.8 |
| NonreactiveControl | 0.1 | 0.000 | 0.0 | 0.000 | 0.0 | 0.000 | 0.0 | 0.000 | 0.0 | 0.000 | 0.0 |
*Within-laboratory precision includes within-run, between-day and between-lot components.
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a blue and green abstract symbol on the left, followed by the company name in blue text. The word "FUJIREBIO" is in a larger, bolder font, and the words "Diagnostics, Inc." are in a smaller font below.
b. Linearity/assay reportable range:
Not applicable
- C. Traceability, Stability, Expected values (controls, calibrators, or methods):
Calibrators
Calibration of the Lumipulse G TP-N is traceable to in-house reference calibrators, whose values have been assigned to correlate with SERODIA®-TP·PA.
CAL N TP calibrator-N (1 x 2.0 mL)
CAL P TP calibrator-P (1 x 2.0 mL)
Contains 0.15 M sodium chloride in Tris buffer with protein stabilizer (bovine). The material of TP calibrator-P is processed from anti-TP positive but inactivated human serum. Preservative: sodium azide.
Stability
Shelf life: The shelf life for Lumipulse G TP-N is 9 months at 2-10° C.
On board: The Lumipulse G TP-N cartridges remain stable at the end of shelf life when stored under on-board conditions (12° C). Labeling recommends calibrator storage for 30 days.
Transport: Lumipulse G TP-N is shipped at 2-10°C.
Sample Stability: Specimens collected in all four matrices (Red top serum tube, SST, K2-EDTA and Sodium Citrate) were stable at 23-27°C for up to 3 days, 2-10 °C for up to 4 days, or -10°C or colder for longer periods.
Serum samples kept on the clot and stored under ambient temperatures (not to exceed 40°C) must be tested within 5 days of collection. However, it is recommended that separated serum be removed from the clot as soon as possible. In analytical studies, a loss of reactivity ranging from 7-14% was observed when on-the-clot serum samples were held at 40°C for 5 days: although no change in positivity was observed for the low positive samples in the study, a false negative result in clinical samples with low antibody levels is possible.
d. Detection limit:
Samples with a cutoff index (C.O.I.) of < 1.0 are considered nonreactive for syphilis T. pallidum antibodies.
Samples with a cutoff index (C.O.I.) ≥ 1.0 are considered reactive for syphilis T. pallidum antibodies.
- e. Analytical specificity:
{13}------------------------------------------------
Image /page/13/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a blue and green abstract symbol on the left, followed by the company name in blue text. The word "FUJIREBIO" is in a larger, bolder font than "Diagnostics, Inc."
Lumipulse G TP-N was evaluated for potential interference from endogenous substances and from therapeutic drugs in a study consistent with the guidelines in the CLSI Protocol EP7-A2. Human serum specimens were supplemented with potentially interfering substances. The study showed that the Lumipulse G TP-N assay is not susceptible to interference from the following substances at the concentrations shown below.
| Endogenous | Test |
|---|---|
| Interferences | Concentration |
| Free Bilirubin (unconjugated) | 40 mg/dL |
| Conjugated Bilirubin | 40 mg/dL |
| Triglycerides (Intralipid 20% Emulsion) | 3000 mg/dL |
| Hemoglobin | 500 mg/dL |
| Human Serum Albumin | 12 g/dL |
| Gamma Globulin | 30 mg/mL |
| Biotin | 500 ng/mL |
| Human Anti-Mouse Antibodies (HAMA) | 1000 ng/mL |
| Rheumatoid Factor (RF) | 1000 IU/mL |
| Cholesterol | 400 mg/mL |
| Ascorbic Acid | 3 mg/mL |
| Therapeutic DrugInterferences | TestConcentration |
| Abacavir Sulfate | 3.85 µg/mL |
| Acetylsalicylic Acid | 3.62 mmol/L |
| Carbamazepine | 50.8 µmol/L |
| Diphenhydramine | 19.6 µmol/L |
| Metformin | 310 µmol/L |
| Metoprolol Tartrate | 18.7 µmol/L |
| Penicillin G Benzathine | 500 IU/mL |
| Rosuvastatin calcium | 30 ng/mL |
| Warfarin | 32.5 µmol/L |
{14}------------------------------------------------
Image /page/14/Picture/1 description: The image is the logo for Fujirebio Diagnostics, Inc. The logo features a stylized blue and green graphic on the left, followed by the company name in blue text. The word "FUJIREBIO" is in a larger, bolder font, and "Diagnostics, Inc." is in a smaller font below it. A registered trademark symbol is located to the right of the word "FUJIREBIO".
Lumipulse GTP-N on the LUMIPULSE G1200 System was evaluated for potential crossreactivity in other diseases and biological conditions using 282 specimens obtained from patients known to have a variety of microbial and viral infections, biological abnormalities and from drug users. The presence of the potential cross-reactant was confirmed with a FDA cleared assay (where applicable). The results are shown below.
| Clinical Category | Number Tested | Number of Reactive Specimens | |
|---|---|---|---|
| Lumipulse G TP-N | FDA Cleared EIA | ||
| Lyme Disease | 10 | 0 | 0 |
| Anti-Nuclear Antibody (ANA) | 10 | 0 | 0 |
| Rheumatoid Factor (RF) | 10 | 1 | 0 |
| Human Anti-Mouse Antibody (HAMA) | 10 | 0 | 0 |
| Hepatitis A Infection (HAV) total | 20 | 2 | 3 |
| Hepatitis A Infection (HAV) IgM | 10 | 0 | 0 |
| Hepatitis B Infection (HBV) | 10 | 0 | 0 |
| Hepatitis C Infection (HCV) | 10 | 2 | 3 |
| Human Immunodeficiency Virus (HIV) | 11 | 6 | 8 |
| Cytomegalovirus (CMV) IgG | 10 | 0 | 0 |
| Cytomegalovirus (CMV) IgM | 10 | 0 | 0 |
| Epstein-Barr Virus (EBV) IgG | 10 | 1 | 1 |
| Herpes Simplex Virus (HSV) IgG | 10 | 1 | 1 |
| Rubella IgG | 10 | 1 | 0 |
| Rubella IgM | 10 | 0 | 0 |
| Toxoplasma IgG | 10 | 0 | 0 |
| Toxoplasma IgM | 10 | 1 | 1 |
| Varicella Zoster Virus (VZV) IgG | 10 | 0 | 0 |
| Lupus (SLE) | 10 | 2 | 2 |
| Drug users | 20 | 14 | 17 |
| Myeloma patients | 13 | 0 | 1 |
| Flu Vaccine recipients | 26 | 0 | 0 |
| Hyper IgG | 10 | 2 | 2 |
| Hyper IgM | 10 | 0 | 1 |
| Leptospirosis | 2 | 0 | 0 |
| Total Samples Tested | 282 | 33 | 40 |
All samples reactive with the Lumipulse G TP-N assay were also reactive with the TP-PA assay, with the exception of 1 Rubella IgG sample and 2 samples from drug users, indicating reactivity to Syphilis (T. Pallidum antibodies) rather than cross reactivity.
Results comparing the Lumipulse G TP-N, a FDA cleared EIA and TP-PA assay for the study above are shown below.
{15}------------------------------------------------
Image /page/15/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized graphic on the left and the company name on the right. The graphic is a combination of blue and white shapes, with some green vertical lines. The text "FUJIREBIO" is in blue, with "Diagnostics, Inc." in a smaller font below it.
| Lumipulse GTP-N | EIA | Total Subjects | TP-PA(Reference Standard) | |
|---|---|---|---|---|
| + | - | |||
| + | + | 31 | 29 | 2 |
| + | - | 2 | 1 | 1 |
| - | + | 9 | 1 | 8 |
| - | - | 240 | 0 | NA* |
| Total | 282 | 31 | 11 |
- Not applicable. Samples were not tested on the TP-PA if both the Lumipulse G TP-N and the FDA Cleared EIA tested non-reactive
f. Assay cut-off:
Lumipulse G TP-N was developed in 1997. 312 negative specimens, 99 positive specimens, and 6 intermediate specimens previously assigned using TP-PA were measured with Lumipulse G TP-N in order to establish its cut-off index. There is clear separation of negative and positive results when using a cut-off value of 1.0.
2. Comparison studies:
a. Method Comparison
In a multi-center clinical study, samples from a total of 2791 subjects were submitted for testing. Among those, there were 1316 specimens prospectively collected from the intended use population and 1475 specimens that were pre-selected from a retrospective collection. After exclusion of 26 specimens from the prospective collection (due to hemolyzed/lipemic samples, lack of test results and/or protocol deviations) and three samples from pre-selected retrospective collection (due to lack of test results), there were a total of 1290 (46%) evaluable prospective samples and a total of 1472 (54%) evaluable retrospective samples. The study samples were tested at four clinical sites.
The prospective collection consisted of specimens collected sequentially from all patients prescribed a laboratory test for syphilis between a defined period of time. The specimens were collected from 7 sites representing different geographical regions of the US including both low prevalence sites and high prevalence sites.
The retrospective samples included specimens from 379 pregnant women (250 without syphilis, 129 with syphilis), 520 HIV positive subjects (298 remnant samples from reference laboratories and 222 collected at a research facility), 130 known to be T. pallidum (TP)reactive by previous laboratory testing. 68 samples collected at a research facility from patients clinically diagnosed with syphilis as well as 375 samples consisting of remnants of specimens sent to a laboratory for routine syphilis testing.
{16}------------------------------------------------
Image /page/16/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a blue abstract symbol on the left, followed by the company name in blue, bold letters. Below the company name, the words "Diagnostics, Inc." are written in a smaller, non-bold font.
Additionally, 289 samples from subjects with well-characterized medically diagnosed syphilis and 474 samples from apparently healthy subjects, including 75 pediatric subjects and 399 adult/not pregnant subjects were tested with the Lumipulse G TP-N assay.
The overall success rate during clinical testing was 99.7% (3540/3552). All 12 specimens were retested and valid results were obtained after a single retest.
1. Comparison of Results
The clinical performance of the Lumipulse G TP-N was evaluated by comparing the assay results with the comparator result based on an algorithm of results from three Food and Druq Administration (FDA) cleared tests: a treponemal test (EIA), a non-treponemal Rapid Plasma Reagin (RPR) test, and a second treponemal test, Treponema pallidum particle agglutination (TP-PA). A determination of positive, negative or indeterminate result was made for each test according to the directions in the labeling.
The final comparator result was determined using a 2 out of 3 rule (EIA, RPR, and TP·PA). In cases where the EIA result was "equivocal" (as per the device labeling), and the TP-PA result was "inconclusive", the final comparator result could not be determined; those results would be excluded from the final analysis (there were no indeterminate final comparator results in this study).
The clinical performance of the Lumipulse GTP-N assay was determined by calculating percent agreement between the Lumipulse G TP-N result and the Final Comparator Result described below.
{17}------------------------------------------------
Image /page/17/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. On the left side of the logo is a blue and green abstract design. To the right of the design is the company name in blue, with "FUJIREBIO" on the top line and "Diagnostics, Inc." on the bottom line.
| Reference Comparator Algorithm | |||
|---|---|---|---|
| EIA(TreponemalTest) | RPR(Non-treponemalTest) | TP-PA(Second TreponemalTest) | FinalComparatorResult |
| Negative | Negative | Positive | Negative |
| Negative | Negative | Negative | Negative |
| Negative | Negative | Inconclusive | Negative |
| Negative | Positive | Positive | Positive |
| Negative | Positive | Negative | Negative |
| Negative | Positive | Inconclusive | Negative |
| Positive | Positive | Positive | Positive |
| Positive | Positive | Negative | Positive |
| Positive | Positive | Inconclusive | Positive |
| Positive | Negative | Positive | Positive |
| Positive | Negative | Negative | Negative |
| Positive | Negative | Inconclusive | Positive |
| Equivocal | Negative | Positive | Positive |
| Equivocal | Negative | Negative | Negative |
| Equivocal | Negative | Inconclusive | Indeterminate |
| Equivocal | Positive | Positive | Positive |
| Equivocal | Positive | Negative | Negative |
| Equivocal | Positive | Inconclusive | Indeterminate |
Performance of the Lumipulse G TP-N assay with Prospective Samples
The age range for the 1290 evaluable subjects was 18 to 92 with a median age of 42. The population contained more male subjects compared to female subjects (66% vs. 34%, respectively). Subjects were from different locations across the US.
Percent agreement between the Lumipulse G TP-N and the Final Comparator Result
The Positive Percent Agreement (PPA) was 92.7% (Confidence Interval (CI) 88.6%, 95.4%) and the Negative Percent Agreement (NPA) was 99.6% (CI 99.0%, 99.9%).
{18}------------------------------------------------
Image /page/18/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized graphic on the left, followed by the company name in blue text. The graphic is a combination of blue and green shapes, with a white space in the center. The text "FUJIREBIO" is in a bold, sans-serif font, and "Diagnostics, Inc." is in a smaller font below it.
| Lumipulse G TP-N | Final Comparator Result | PPA (%) | 95% CI | NPA (%) | 95% CI | |
|---|---|---|---|---|---|---|
| Reactive | Nonreactive | |||||
| Reactive | 215 | 4 | 92.7 | 88.6, 95.4 | 99.6 | 99.0, 99.9 |
| Nonreactive | 17 | 1054 |
The summary of the serological test profile for the prospective study population is shown below.
| EIA | RPR | TP-PA | FinalComparatorResult | Lumipulse GTP-N | Number ofSubjects |
|---|---|---|---|---|---|
| Negative | Negative | Negative | Negative | Negative | 1031 |
| Positive | Negative | Positive | Positive | Positive | 119 |
| Positive | Positive | Positive | Positive | Positive | 90 |
| Positive | Negative | Positive | Positive | Negative | 14 |
| Positive | Negative | Negative | Negative | Negative | 14 |
| Negative | Positive | Negative | Negative | Negative | 5 |
| Positive | Negative | Negative | Negative | Positive | 4 |
| Positive | Negative | Indeterminate | Positive | Negative | 3 |
| Positive | Positive | Negative | Positive | Positive | 3 |
| Equivocal | Negative | Negative | Negative | Negative | 2 |
| Negative | Negative | Positive | Negative | Negative | 2 |
| Negative | Positive | Positive | Positive | Positive | 2 |
| Negative | Negative | Positive | Negative | Positive | 1 |
| Total | 1290 |
The study population consisted of subjects sent for routine syphilis testing, those with previous history of syphilis, pregnant women (3%), and persons positive for HIV. The percent agreement of the Lumipulse G TP-N assay results when compared to the Final Comparator Result in each category of the prospective samples is shown below.
{19}------------------------------------------------
Image /page/19/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a blue abstract symbol on the left, followed by the company name in blue text. The word "FUJIREBIO" is in a larger, bolder font than "Diagnostics, Inc.", which is placed below it. The overall design is clean and professional.
| Category | PPA(%) | 95% CI | NPA (%) | 95% CI |
|---|---|---|---|---|
| Routine Syphilis | 90.4 (94/104) | 83.2, 94.7 | 99.7 (933/936) | 99.1, 99.9 |
| Previously Diagnosedwith Syphilis* | 96.9 (94/97) | 91.3, 98.9 | 100 (12/12) | 75.8, 100 |
| Pregnant(Unknown Trimester) | NA | NA | 100 (41/41) | 91.4, 100 |
| HIV Positive* | 93.7 (74/79) | 86.0, 97.3 | 98.6 (72/73) | 92.6, 99.8 |
NA=Not applicable, no positive results obtained/determined
*52 subjects had a previous history of syphilis and were HIV positive. These subjects are counted in each category separately.
Due to variance in geographic locations or demographics, assay results obtained in individual laboratories may vary from data presented.
Performance of the Lumipulse G TP-N assay with Retrospective Samples
The age range for the 1472 evaluable subjects was 14 to 89 with a median age of 33. The retrospective population had more female than male subjects (57% vs. 43%, respectively). Subjects were from different locations across the US. The pregnant subjects were evenly split across each trimester.
Percent agreement between the Lumipulse G TP-N and the Final Comparator Result
The Positive Percent Agreement (PPA) was 94.3% (CI 92.0%, 96.0%) and the Negative Percent Agreement (NPA) was 98.1% (CI 97.0%, 98.8%).
| Lumipulse G TP-N | Final Comparator Result | PPA (%) | 95% CI | NPA (%) | 95% CI | |
|---|---|---|---|---|---|---|
| Reactive | 513 | 18 | 94.3 | 92.0, 96.0 | 98.1 | 97.0, 98.8 |
| Nonreactive | 31 | 910 |
{20}------------------------------------------------
Image /page/20/Picture/1 description: The image shows the logo for FUJIREBIO Diagnostics, Inc. The logo consists of a blue and green abstract shape on the left, followed by the company name in blue, bold letters. Below the company name, the words "Diagnostics, Inc." are written in a smaller, black font.
A summary of the serological test profile for the retrospective samples is summarized in the following table.
| EIA | RPR | TP-PA | FinalComparatorResult | Lumipulse GTP-N | Number ofSubjects |
|---|---|---|---|---|---|
| Negative | Negative | Positive | Negative | Negative | 6 |
| Negative | Negative | Negative | Negative | Negative | 844 |
| Negative | Negative | Negative | Negative | Positive | 10 |
| Negative | Negative | Indeterminate | Negative | Negative | 2 |
| Negative | Positive | Negative | Negative | Positive | 20 |
| Positive | Positive | Positive | Positive | Positive | 277 |
| Positive | Positive | Positive | Positive | Negative | 4 |
| Positive | Positive | Negative | Positive | Negative | 2 |
| Positive | Positive | Indeterminate | Positive | Negative | 1 |
| Positive | Negative | Positive | Positive | Positive | 234 |
| Positive | Negative | Positive | Positive | Negative | 20 |
| Positive | Negative | Negative | Negative | Negative | 28 |
| Positive | Negative | Negative | Negative | Positive | 8 |
| Positive | Negative | Indeterminate | Positive | Positive | 2 |
| Positive | Negative | Indeterminate | Positive | Negative | 4 |
| Equivocal | Negative | Negative | Negative | Negative | 10 |
| Total | 1472 |
The percent agreement of the Lumipulse G TP-N assay results when compared to the Final Comparator Result in each category of the retrospective samples is shown below
| Category | PPA(%) | 95% CI | NPA(%) | 95% CI |
|---|---|---|---|---|
| Pregnant | 96.8 (92/95) | 91.1, 98.9 | 96.8 (275/284) | 94.1, 98.3 |
| HIV | 90.3 (214/237) | 85.9, 93.4 | 97.5 (276/283) | 95.0, 98.8 |
| Reactive by PreviousLaboratory Testing | 99.2 (121/122) | 95.5, 99.9 | 100.0 (8/8) | 67.6, 100.0 |
| Routine SyphilisTesting | 91.2 (31/34) | 77.0, 97.0 | 99.7 (340/341) | 98.4, 99.9 |
| Medically DiagnosedSyphilis (UnknownStage) | 98.2 (55/56) | 90.6, 99.7 | 91.7 (11/12) | 64.6, 98.5 |
{21}------------------------------------------------
Image /page/21/Picture/1 description: The image shows the Fujirebio Diagnostics, Inc. logo. The logo consists of a blue abstract shape on the left and the company name in blue text on the right. The company name is written in two lines, with "FUJIREBIO" on the top line and "Diagnostics, Inc." on the bottom line.
The percent agreement between the Lumipulse G TP-N and the Final Comparator Result with samples from pregnant women in the prospective and retrospective populations is shown in the table below.
| Category | PPA(%) | 95% CI | NPA(%) | 95% CI |
|---|---|---|---|---|
| Prospective | ||||
| Unknown Trimester | NA (0/0) | NA | 100 (41/41) | 91.4, 100 |
| Retrospective | ||||
| First Trimester | 100 (21/21) | 84.5, 100 | 100 (93/93) | 96.0, 100 |
| Second Trimester | 96.8 (30/31) | 83.8, 99.4 | 94.7 (90/95) | 88.3, 97.7 |
| Third Trimester | 95.4 (41/43) | 84.5, 98.7 | 95.8 (92/96) | 89.8, 98.4 |
The percent agreement between the Lumipulse G TP-N and the Final Comparator Result with samples from HIV positive subjects in the prospective and retrospective populations is shown in the table below.
| Category | PPA(%) | 95% CI | NPA(%) | 95% CI |
|---|---|---|---|---|
| Prospective | ||||
| HIV Positive | 93.7 (74/79) | 86.0, 97.3 | 98.6 (72/73) | 92.6, 99.8 |
| Retrospective | ||||
| HIV Positive | 90.3 (214/237) | 85.9, 93.4 | 97.5 (276/283) | 95.0, 98.8 |
The performance of the Lumipulse G TP-N assay was also evaluated with samples from medically diagnosed syphilis subjects, based on clinical information and laboratory test results.
The age range for the 289 subjects evaluated was 18 to 78 with a median age of 44. The population was predominantly male (67%). Subjects were from Argentina (52%) and Florida (48%). The reactivity of the Lumipulse G TP-N assay with samples from subjects medically diagnosed with syphilis is presented by syphilis stage and treatment status.
{22}------------------------------------------------
Image /page/22/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a blue and green abstract symbol on the left, followed by the company name in blue text. The word "FUJIREBIO" is in a larger, bolder font than "Diagnostics, Inc."
| Medically Diagnosed Subjects | Lumipulse G TP-N | |||
|---|---|---|---|---|
| Syphilis Stage | Treatment Status | N | Number of Reactive (%) | Number of Nonreactive (%) |
| Primary | Treated | 2 | 2 (100%) | 0 (0%) |
| Primary | Untreated | 27 | 27 (100%) | 0 (0%) |
| Secondary | Treated | 25 | 25 (100%) | 0 (0%) |
| Secondary | Untreated | 30 | 30 (100%) | 0 (0%) |
| Latent | Treated | 5 | 5 (100%) | 0 (0%) |
| Latent | Untreated | 200 | 183 (91.5%) | 17* (8.5%) |
*13/17 nonreactive samples tested nonreactive by TP-PA and 15/17 tested nonreactive by RPR.
Performance in Apparently Healthy Individuals
The performance of the Lumipulse G TP-N assay was evaluated with samples from apparently healthy individuals. The 474 samples tested included 75 pediatric subjects and 399 adult/not pregnant subjects ranging in age from 2 months to 68 years of age (63% male and 37% female).
| Lumipulse G TP-N Result | |||
|---|---|---|---|
| Category | Number ofReactive (%) | Number ofNonreactive (%) | Total |
| Adults | 1* (0.3) | 398 (99.7) | 399 |
| Pediatrics | 0 (0.0) | 75 (100.0) | 75 |
| Total | 1 (0.2) | 473 (99.7) | 474 |
*The one reactive sample also tested reactive by TP·PA and RPR
As with all in vitro diagnostic assays, each laboratory should determine its own reference range(s) for the diagnostic evaluation of patient results.
- b. Matrix Comparison
Lumipulse G TP-N on the LUMIPULSE G1200 was evaluated for matrix differences by performing a study using sixty (60) matched sets of serum (red top and serum separator tubes (SST)) and plasma (K2EDTA and sodium citrate) samples. The results demonstrated equivalency between the matrices and are presented in the following table:
{23}------------------------------------------------
Image /page/23/Picture/1 description: The image contains the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized graphic on the left, resembling overlapping water droplets in blue and green. To the right of the graphic is the company name, "FUJIREBIO" in bold, blue letters, with the registered trademark symbol. Below the company name is the text "Diagnostics, Inc." in a smaller, black font.
| MatrixComparison | SampleRange(COI) | CorrelationCoefficient | Intercept(95% CI) | Slope(95%CI) | Bias(95%CI) |
|---|---|---|---|---|---|
| SSTversus Red Top | 0.1 – 6.2 | 0.997 | -0.002(-0.003 – 0.000) | 1.013(0.996 – 1.03) | .0111(-0.004 – 0.026) |
| K₂EDTAversus Red Top | 0.1 – 5.6 | 0.992 | 0.004(0.002 – 0.007) | 0.952(0.924 – 0.980) | -0.043(-0.069 – -0.0179) |
| Sodium Citrateversus Red Top | 0.1 – 5.6 | 0.987 | 0.007(0.004 – 0.010) | 0.922(0.893 – 0.952) | -0.070(-0.097 – -0.044) |
Clinical studies: 3.
-
a. Clinical sensitivity:
Not applicable (See item 2a above) -
b. Clinical specificity:
Not applicable (See item 2a above) -
c. Other clinical supportive data (when a. and b. are not applicable):
-
- Clinical cut-off:
Not applicable
5. Expected values/Reference range:
The distribution of Lumipulse G TP-N results in the prospective population in this study, stratified by age and gender, is shown in the table below.
{24}------------------------------------------------
Image /page/24/Picture/1 description: The image is a logo for Fujirebio Diagnostics, Inc. The logo features a stylized graphic on the left, consisting of blue and green shapes. To the right of the graphic is the company name, "FUJIREBIO" in large, blue, sans-serif font. Below the company name is the text "Diagnostics, Inc." in a smaller, gray, sans-serif font.
| Age Range | Gender | Lumipulse G TP-N ResultNumber ofReactive (%) | Number ofNonreactive (%) | Total |
|---|---|---|---|---|
| 18 to 21 | Female | 0 (0%) | 22 (100%) | 22 |
| 18 to 21 | Male | 3 (10%) | 26 (90%) | 29 |
| 22 to 29 | Female | 0 (0%) | 90 (100%) | 90 |
| 22 to 29 | Male | 27 (18%) | 119 (82%) | 146 |
| 30 to 39 | Female | 3 (3%) | 115 (97%) | 118 |
| 30 to 39 | Male | 40 (21%) | 149 (79%) | 189 |
| 40 to 49 | Female | 7 (11%) | 56 (89%) | 63 |
| 40 to 49 | Male | 28 (18%) | 127 (82%) | 155 |
| 50 to 59 | Female | 32 (26%) | 89 (74%) | 121 |
| 50 to 59 | Male | 53 (22%) | 188 (78%) | 241 |
| 60 to 64 | Female | 5 (20%) | 20 (80%) | 25 |
| 60 to 64 | Male | 16 (28%) | 42 (72%) | 58 |
| ≥ 65 | Female | 0 (0%) | 6 (100%) | 6 |
| ≥ 65 | Male | 5 (19%) | 22 (81%) | 27 |
| Total | 219 (17%) | 1071 (83%) | 1290 |
6. Conclusion
The results of these analytical (nonclinical) and clinical studies demonstrate that the performance of the Lumipulse G TP-N assay is substantially equivalent to the performance of the ADVIA Centaur Syphilis (SYPH) assay.
N. Proposed Labeling:
The labeling satisfies the requirements of 21 CFR Part 809.10.
§ 866.3830
Treponema pallidum treponemal test reagents.(a)
Identification. Treponema pallidum treponemal test reagents are devices that consist of the antigens, antisera and all control reagents (standardized reagents with which test results are compared) which are derived from treponemal sources and that are used in the fluorescent treponemal antibody absorption test (FTA-ABS), theTreponema pallidum immobilization test (T.P.I.), and other treponemal tests used to identify antibodies toTreponema pallidum directly from infecting treponemal organisms in serum. The identification aids in the diagnosis of syphilis caused by bacteria belonging to the genusTreponema and provides epidemiological information on syphilis.(b)
Classification. Class II (performance standards).