K Number
K153099
Date Cleared
2016-05-23

(210 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Inclusive® Tapered Implant System is indicated for use in maxillary and mandibular partially or fully edentulous cases, to support single, multiple-unit, and overdenture restorations. The implants are to be used for immediate loading only in the presence of primary stability and appropriate occlusal loading.

Device Description

The proposed device, Inclusive® Tapered Implant System, is manufactured from biocompatible Titanium alloy. The implant is designed with an internal hex connection with a diameter of implant with a 6-point flower (drawings are provided in the table below). Equipped with a tapered body and used to replace one or more missing teeth. The surface is blasted with Hydroxyl Apatite and acid etched. The dental implant (proposed device) will be provided sterile using gamma sterilization. The Inclusive Tapered Implant System, includes abutments in the form of healing abutments, engaging and non-engaging temporary abutments, straight and angled titanium abutments, titanium abutment fixation screws and straight multi-unit abutments.

AI/ML Overview

The provided text is a 510(k) summary for the Inclusive® Tapered Implant System. It describes the device, its indications for use, and a comparison to predicate devices to establish substantial equivalence.

Based on the document, here's an analysis of the acceptance criteria and study information:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly present a table of acceptance criteria with corresponding performance metrics like sensitivity, specificity, or AUC, as would be typical for an AI/algorithm-driven medical device. This is because the device in question is an endosseous dental implant, a physical medical device, not a diagnostic algorithm.

Instead, the "acceptance criteria" are implied by adherence to recognized standards and successful completion of non-clinical performance tests for mechanical integrity and biocompatibility. The "reported device performance" is the conclusion that the device meets these standards and functions equivalently to predicate devices.

Here's a table summarizing the performance evaluation strategy:

Evaluation AspectAcceptance Criteria (Implied)Reported Device Performance
BiocompatibilityCompliance with ISO 10993-1 or justification through equivalent materials, manufacturing, and patient contact to previously cleared predicate devices.The device is manufactured from biocompatible Titanium Grade 23 (Ti-6AL-4V ELI) and meets ASTM F-136 Standard. Biocompatibility is justified by equivalence to primary predicate (K121406) and reference predicate (K143353) which use identical materials, same manufacturing process, and same type/duration of patient contact.
Mechanical StrengthDynamic fatigue testing and static load failure testing of finished assembled implant/abutment systems, performed in accordance with FDA Guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments" and ISO 14801:2007.Performance testing (Fatigue Testing and Static Load Failure Testing) was performed and demonstrated implant-to-abutment compatibility. The testing was performed in accordance with the specified FDA Guidance and ISO 14801:2007. The conclusion is that the device has the "same performance specifications" as the predicate devices.
SterilityAdherence to relevant sterilization standards (ANSI-AAMI ST79, ISO 17665-1, ISO 11137-1, ISO 11137-2).The sterilization and shelf life of all device components are supported by prior validation conducted for the applicant's own primary and reference predicate devices (K121406 and K143353).
Design/MaterialsDevice materials, design, and indications for use must be substantially equivalent to legally marketed predicate devices.Comparison table (Table 1) demonstrates that the proposed device is "essentially the same as currently marketed devices for the same indications for use, same materials, same fundamental scientific technology and design." Any differences do not introduce new safety/efficacy concerns.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not applicable in the context of this 510(k) submission for a physical dental implant. The "test set" here refers to the physical samples of the device and its components used for non-clinical (laboratory) testing, not a dataset for an algorithm.

  • Sample Size: The document does not specify the exact number of implants or abutments tested in the fatigue and static load failure tests. It refers to "finished assembled implant/abutment systems" being tested.
  • Data Provenance: The data is generated from laboratory testing of the manufactured devices, not from patient data. Therefore, concepts like "country of origin of the data" or "retrospective/prospective" studies do not apply in this context.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is not applicable. "Ground truth" in the context of this device refers to the physical properties and performance characteristics of the implant, which are established through standardized engineering and material science tests, not through expert human interpretation or consensus. The tests themselves (e.g., fatigue testing, material composition analysis) provide objective data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This is not applicable. Adjudication methods like "2+1" or "3+1" are used in studies where human interpretations of data (e.g., medical images) need to be reconciled to establish a ground truth. For mechanical and material testing of a physical device, the results are derived from standardized test procedures and measurements, not human consensus on interpretation.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. The device is a physical dental implant, not an AI or algorithm. Therefore, an MRMC study related to AI assistance for human readers is irrelevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable, as the device is a physical dental implant, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For this physical device, the "ground truth" for demonstrating performance and safety is established through:

  • Adherence to recognized standards: E.g., ISO 10993-1 (biocompatibility), ASTM F136 (material properties), ISO 14801 (dynamic fatigue).
  • Non-clinical (laboratory) performance testing: Mechanical strength (fatigue and static load failure testing).
  • Material verification: Titanium Grade 23 (Ti-6AL-4V ELI) meeting ASTM F-136.
  • Sterilization validation: Conformance to ISO and AAMI standards.

These are objective, measurable criteria verified through specific test protocols, rather than subjective expert consensus or clinical outcomes data.

8. The sample size for the training set

This is not applicable. There is no "training set" as this is not an AI/machine learning device.

9. How the ground truth for the training set was established

This is not applicable, as there is no "training set."

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the symbol. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 23, 2016

Prismatik Dentalcraft, Inc. Ms. Maria E. Wagner Sr. Specialist, Regulatory Affairs 2212 Dupont Drive, Suite P Irvine, California 92612

Re: K153099

Trade/Device Name: Inclusive® Tapered Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: May 4, 2016 Received: May 5, 2016

Dear Ms. Wagner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tina Kiang
-s

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K153099

Device Name Inclusive® Tapered Implant System

Indications for Use (Describe)

Inclusive Tapered Implant System is indicated for use in maxillary and mandibular partially or fully edentulous cases, to support single, multiple-unit, and overdenture restorations. The implants are to be used for immediate loading only in the presence of primary stability and appropriate occlusal loading.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo consists of a colorful triangle on the left and the company name on the right. The triangle is divided into several sections, each with a different color. The company name is written in blue, with "PRISMATIK" on the top line and "DENTALCRAFT, INC." on the bottom line.

B.

007_510 (K) Summary-807.92

This 510 (k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.

A. SUBMITTER INFORMATION

Company Name:Prismatik Dentalcraft, Inc.
Company Address:2212 Dupont Dr., Suite P,Irvine, CA 92612
Company Phone:949-440-2636
Company FAX:949-553-0924
Facility Registration Number:2031503
Primary Contact Person:Maria Wagner, (949) 440-2636Senior Specialist, Regulatory AffairsMaria.wagner@glidewelldental.com
Secondary Contact PersonShelly Harris, (949) 440-2629Director of Regulatory Affairs & QualityAssuranceShelly.harris@glidewelldental.com
Date Summary Prepared:May 18, 2016
DEVICE IDENTIFICATION
Trade/Proprietary Name:Inclusive® Tapered Implant System
21 CFR Reference:21 CFR 872.3640
21 CFR Common Name:Endosseous dental implant
Classification:Class II, DZE, NHA
Panel:Dental

C. IDENTIFICATION OF PREDICATE DEVICE

Primary Predicate:Inclusive Tapered Implant System (K121406, Cleared on 2/22/2013)
Reference Predicate:Hahn Tapered Implant System (K143353, Cleared on 4/2/2015)

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. On the left is a triangle with different colors inside. To the right of the triangle is the company name, "PRISMATIK DENTALCRAFT, INC.", in blue text.

PROPOSED DEVICE DESCRIPTION D.

The proposed device, Inclusive® Tapered Implant System, is manufactured from biocompatible Titanium alloy. The implant is designed with an internal hex connection with a diameter of implant with a 6-point flower (drawings are provided in the table below). Equipped with a tapered body and used to replace one or more missing teeth. The surface is blasted with Hydroxyl Apatite and acid etched. The dental implant (proposed device) will be provided sterile using gamma sterilization. The Inclusive Tapered Implant System, includes abutments in the form of healing abutments, engaging and non-engaging temporary abutments, straight and angled titanium abutments, titanium abutment fixation screws and straight multi-unit abutments.

E. INDICATIONS FOR USE

Inclusive® Tapered Implant System is indicated for use in maxillary and mandibular partially or fully edentulous cases, to support single, multiple-unit, and overdenture restorations. The implants are to be used for immediate loading only in the presence of primary stability and appropriate occlusal loading.

F. DETERMINATION OF SUBSTANTIAL EQUIVALENCE

The comparison table below outlines and provides the similarities and the substantial equivalency of the proposed device, the Inclusive® Tapered Implant System, to the predicate device, Inclusive® Tapered Implant System, K121406. Prismatik believes that the comparative data presented in the preceding paragraphs demonstrate that the proposed device is essentially the same as currently marketed devices for the same indications for use, same materials same fundamental scientific technology and design and supports our claim of substantial equivalence to predicate Class II devices under the classification of endosseous dental implant (21 CFR 872.3640) that have previously been found to be substantially equivalent. Any differences between the proposed device and the predicate device does not introduce new concerns in safety and efficacy.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo consists of a colorful triangle on the left and the company name on the right. The triangle is divided into several sections, each with a different color. The company name is written in blue, with "PRISMATIK" on the top line and "DENTALCRAFT, INC." on the bottom line.

AttributesPredicate DeviceReferenced DeviceProposed DeviceSimilarities andDifferences
Inclusive Tapered ImplantSystem (K121406)Hahn Tapered ImplantSystem (K143353)Inclusive Tapered Implant -3.2 implant
Picture ofImplantImage: Inclusive Tapered Implant System (K121406)Image: Hahn Tapered Implant System (K143353)Image: Inclusive Tapered Implant - 3.2 implantSame asPredicate
Indications forUseInclusive Tapered Implants areindicated for use in maxillaryand mandibular partially orfully edentulous cases, tosupport single, multiple-unit,and overdenture restorations.The implants are to be used forimmediate loading only in thepresence of primary stabilityand appropriate occlusalloading.Hahn Implants are indicatedfor use in maxillary andmandibular partially or fullyedentulous cases, to supportsingle, multiple-unit, andoverdenture restorations. Theimplants are to be used forimmediate loading only in thepresence of primary stabilityand appropriate occlusalloading.Inclusive Tapered ImplantSystem is indicated for use inmaxillary and mandibularpartially or fully edentulouscases, to support single,multiple-unit, and overdenturerestorations. The implants are tobe used for immediate loadingonly in the presence of primarystability and appropriate occlusalloading.Same
SurfaceBlasted with HydroxylApatite and acid etchedBlasted with HydroxylApatite and acid etchedBlasted with Hydroxyl Apatiteand acid etchedSame
ConnectionImage: Connection 3.7 - 5.2: Internal HexImage: Connection 3.0: 6-point flower3.5-7.0: Internal HexImage: Connection 3.2: 6-point flowerSame asreference
DesignThreaded root-form implantThreaded root-form implantThreaded root-form implantSame
Implant BodyGeometryScrew TypeScrew TypeScrew TypeSame
Diameters(mm)3.7, 4.7, 5.23.0, 3.5, 4.3, 5.0, 7.03.2New: AdditionalSize
Lengths (mm)3.7 - 5.2: 8, 10, 11.5, 13 & 163.0: 11.5, 13 & 163.5 - 5.0: 8, 10, 11.5, 13 & 167.0: 8, 10 & 11.53.2: 11.5, 13, and 16Same
AttributesPredicate DeviceReferenced DeviceProposed DeviceSimilarities andDifferences
Inclusive Tapered ImplantSystem (K121406)Hahn Tapered ImplantSystem (K143353)Inclusive Tapered Implant -3.2 implant
SterilityPackaged SterilePackaged SterilePackaged SterileSame
MaterialTitanium Alloy, Grade 23Titanium Alloy, Grade 23Titanium Alloy, Grade 23Same
Description ofAbutmentTitanium abutments are madefrom titanium alloy. TitaniumAbutments are prefabricatedprosthetic componentsdirectly connected toendosseous dental implantsand are intended to use as anaid to prosthetic rehabilitation.Titanium abutments are madefrom, titanium alloy. TitaniumAbutments are prefabricatedprosthetic componentsdirectly connected toendosseous dental implantsand are intended to use as anaid to prostheticrehabilitation.Titanium abutments are madefrom the same material, titaniumalloy, Titanium Abutments areprefabricated prostheticcomponents directly connectedto endosseous dental implantsand are intended to use as an aidto prosthetic rehabilitation.Same
Picture ofScrewImage: ScrewImage: ScrewImage: ScrewSame asreference
AbutmentAngle0°-20°0°-30°0°-15°within range ofpreviousclearances
Description ofChangeNoneNoneAddition of the 3.2 - Externalthread pattern of the InclusiveImplant combined with internalconnection of the Hahn 3.0(reference).Additional size3.2 mm
AbutmentSeatImage: Abutment SeatImage: Abutment SeatImage: Abutment SeatSame asreference
AttributesPredicate DeviceReferenced DeviceProposed DeviceSimilarities andDifferences
Inclusive Tapered ImplantSystem (K121406)Hahn Tapered ImplantSystem (K143353)Inclusive Tapered Implant -3.2 implant
Screw SeatSits on a flatSits on a taperSits on a TaperSame asreference
ConstructionMachinedMachinedMachinedSame

Table 1 – Comparison between Predicate and Proposed Device

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo consists of a colorful triangular prism on the left and the company name on the right. The company name is written in a sans-serif font, with "PRISMATIK" on the top line and "DENTALCRAFT, INC." on the bottom line.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the logo for Prismatik Dentalcraft, Inc. The logo consists of two parts: a colorful triangular prism on the left and the company name on the right. The prism is divided into multiple sections, each filled with a different color, creating a rainbow effect. The text "PRISMATIK DENTALCRAFT, INC." is written in a bold, sans-serif font, with "PRISMATIK" on the top line and "DENTALCRAFT, INC." on the bottom line.

G. SUMMARY OF NON-CLINICAL TESTING (PERFORMANCE DATA)

Non-clinical test data was used to support the substantial equivalency. The functionality of the proposed Inclusive® Tapered Implant System, 3.2 as well as their conformance to design input was further determined by performance testing (Fatigue Testing and Static Load Failure Testing of finished assembled implant/abutment systems). The testing was performed in accordance with FDA Guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments. The testing performed demonstrated implant to abutment compatibility. The applicable standards that are used in this submission are listed below:

Applicable Standards
ISO 10993-1: 2009Biological Evaluation of Medical Devices-Evaluation and testing
ISO 14971: 2012Medical Devices —Application of riskmanagement to medical devices
ASTM F136-13: 2013Standard for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy forSurgical Implant Applications
ISO 14801:2007Dentistry - Implants - Dynamic fatigue test forendosseous dental implants
BS EN 1641:2009Dentistry — Medical devices for dentistry —Materials
ANSI-AAMI ST79: 2010 & A1: 2010& A2: 2011 & A3: 2012Comprehensive guide to steam sterilization andsterility assurance in health care facilities
ISO 17665-1:2006Sterilization of health care products — Moistheat — Part 1: Requirements for thedevelopment, validation and routine control of asterilization process for medical devices

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains the logo for Prismatik Dentalcraft, Inc. The logo consists of a colorful triangle on the left and the company name on the right. The triangle is divided into sections of different colors, including pink, blue, and yellow. The company name is written in a sans-serif font, with "PRISMATIK" on the top line and "DENTALCRAFT, INC." on the bottom line.

ISO 11137-1:2006+A: 2013Sterilization of health care products - Radiation -Part 1: Requirements for development,validation and routine control of a sterilizationprocess for medical devices
ISO 11137-2:20132013 Sterilization of health care products —Radiation Part 2: Establishing the sterilization

The proposed Inclusive® Tapered Implant System, 3.2 is manufactured from biocompatible titanium grade 23 (Ti-6AL-4V ELI) and meets ASTM F-136 Standard. In accordance with FDA Guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments", in lieu of performing biocompatibility testing per ISO 10993-1 for the subject device, we have identified a Prismatik predicate device and a Prismatik reference device with identical materials, same manufacturing process, and same type/duration of patient contact:

  • . -P r i m a r y Predicate: K121406, Inclusive Tapered Implant System
  • . -Reference Predicate: K143353, Hahn Tapered Implant System

The proposed device, the Inclusive® Tapered Implant System, 3.2 consists of dental implants, abutments, and screws. The implant, healing abutments, and multi-unit abutments will be provided sterile. Abutments and fixation screws are provided non-sterile and sterilized prior to use. The sterilization and shelf life of all components of the device are supported by prior validation conducted for the applicant's own primary and reference predicate devices K121406 and K143353

CONCLUSION FROM THE NON-CLINICAL TESTING (PERFORMANCE DATA) H.

The proposed device, the Inclusive® Tapered Implant System has the same performance specifications, fundamental scientific technology and intended use as that of the predicate device, Inclusive® Tapered Implant System, K121406 (Clearance letter dated February 22, 2013) and the reference device Hahn Tapered Implant System, K143353 (Clearance letter dated April 2, 2015). The performance data and a declaration of conformity with design controls supports a determination of continuing substantial equivalence of the proposed device to the predicate and referenced devices.

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.