(112 days)
A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or fingers to prevent contamination between patient and examiner.
Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Powder-Free Vinyl Patient Examination Gloves, LYZ, and meets all requirements of ASTM D5250-06.
This document is a 510(k) Pre-market Notification for Zibo Huiying Medical Products, Co. Ltd.'s "Synmax Synthetic Patient Examination Vinyl Gloves, Powder Free, Blue". It aims to demonstrate substantial equivalence to a predicate device. The information provided heavily relies on non-clinical testing and comparison to established ASTM standards.
Here's an analysis of the acceptance criteria and the study that proves the device meets those criteria, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
| Acceptance Criteria (Standard / Predicate Device) | Reported Device Performance (Synmax Synthetic Patient Examination Vinyl Gloves) |
|---|---|
| Physical Properties (ASTM D5250, ASTM D6319) | |
| Before Aging: Tensile Strength: 16 Mpa min | Average Tensile Strength: 16.9 Mpa |
| Before Aging: Elongation: 500% min | Average Ultimate Elongations: 550% |
| After Aging: Tensile Strength: 14 Mpa min | Average Tensile Strength: 14.4 Mpa |
| After Aging: Elongation: 400% min | Average Ultimate Elongations: 500% |
| Overall Length on Medium Size: 230 mm min | Average over 234mm |
| Width of Palm on Medium Size: 95 ± 10 mm | Average 96mm |
| Palm Thickness: 0.05 mm min | Average 0.096 mm |
| Finger Thickness: 0.05 mm min | Average 0.98 mm |
| Residual Powder (ASTM D6124-06) | |
| < 2 mg powder per glove | < 2 mg per glove (no defect glove found) |
| Pinhole Results (FDA 1000 ml Water Fill Test / ASTM D5151-06) | |
| AQL 2.5 | Pinhole were found less than two pieces gloves out of 125 pieces gloves. AQL 2.5 is met. |
| Biocompatibility (ISO 10993-10) | |
| Primary Skin Irritation: Not an irritant | Under the conditions tested, the subject device was not an irritant |
| Dermal Sensitization: Not a sensitizer | Under the conditions tested, the subject device was not a sensitizer |
2. Sample Size Used for the Test Set and Data Provenance:
- Pinhole Results (ASTM D5151-06): 125 pieces of gloves were tested.
- Physical and Dimensions Testing (ASTM D-5250): Conducted at inspection level S-2, AOL 2.5. The exact sample size is not explicitly stated as number of gloves, but is implied by the AQL (Acceptance Quality Limit) which defines the maximum percentage of defective units in a batch that can be considered satisfactory.
- Residual Powder Test (ASTM D-6124): "start to finish inspection" and "no defect glove found according to ASTM D6124-06." The exact sample size as a number of gloves is not explicitly stated.
- Biocompatibility (ISO 10993-10): Sample sizes for "Primary Skin Irritation" and "Skin Sensitization" tests are not specified in the document, but these typically involve animal testing following established protocols.
- Data Provenance: The document does not explicitly state the country of origin of the test data (e.g., where the product was tested). The tests were performed by or for Zibo Huiying Medical Products, Co. Ltd., which is located in China. The data would be considered prospective, as it was generated specifically for this 510(k) submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
- Not applicable for this type of device. The assessment of patient examination gloves relies primarily on objective, quantifiable physical and chemical properties and performance tests against established standards (ASTM, ISO). There is no "ground truth" established by human experts in the way clinical images or diagnostic outcomes would be. The ground truth is defined by the scientific and engineering standards themselves.
4. Adjudication Method for the Test Set:
- Not applicable. As there are no human expert readers or subjective assessments requiring adjudication, no adjudication method was used. The tests are objective and quantitative, comparing results directly against specified numerical criteria in the standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
- No. An MRMC study is relevant for diagnostic devices where human readers interpret results, often to compare human performance with and without AI assistance. This device (patient examination gloves) is not a diagnostic device and does not involve human interpretation or AI assistance in its primary function.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- No. This device is a medical device, not an algorithm or AI system. Therefore, the concept of "standalone performance" for an algorithm is not applicable. The performance is the inherent physical and chemical properties of the glove itself.
7. The Type of Ground Truth Used:
- The ground truth for this device is based on established industry standards and regulatory requirements. Specifically:
- ASTM standards: ASTM D5250, ASTM D6319-00al (for physical properties), ASTM D6124-06 (for residual powder), ASTM D5151-06 (for pinhole detection).
- ISO standards: ISO 10993-10 (for biocompatibility).
- Regulatory requirements: 21 CFR 880.6250 for patient examination gloves.
- The predicate device's performance also serves as a benchmark for comparison.
8. The Sample Size for the Training Set:
- Not applicable. This device is a physical product, not an AI/ML algorithm that requires a "training set." The manufacturing process itself (which includes quality control and testing) ensures consistency, but there isn't a separate "training" of the product in the way an algorithm is trained on data.
9. How the Ground Truth for the Training Set was established:
- Not applicable. As there is no training set for a physical product like examination gloves, there's no ground truth to be established for it.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of three human profiles facing right, arranged in a cascading manner. The profiles are connected and appear to be emerging from a single form. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the graphic.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 5, 2016
Zibo Huiying Medical Products, Co. Ltd. % Sophie Hao Official Correspondent Basic Medical Industries, Inc. 12390 East End Ave Chino, California 91710
Re: K153028
Trade/Device Name: Synthetic Patient Examination Vinyl Gloves, Powder Free, Blue Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: December 18, 2015 Received: December 18, 2015
Dear Sophie Hao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K153028
Device Name
Synmax Synthetic Patient Examination Vinyl Gloves, Powder Free, Blue
Indications for Use (Describe)
A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or fingers to prevent contamination between patient and examiner.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(K) SUMMARY
This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR §807.92.
1. Submitter's Identification:
Zibo Huiying Medical Products, Co. Ltd. No. 3, Da Wu Road Linzi, Shandong China Tel: 86-18953386552
Contact Person
Sophie Hao Tel: 909-548-4828, Fax: 909-548-4808 Sophiehao@basicmedical.com
Date summary prepared: Jan 7, 2016
-
- Name of the Device: Synmax Synthetic Patient Examination Vinyl Gloves. Powder Free, Blue
-
- Common Name: Patient Examination Gloves
4. Predicate Device Information: Grand Work Plastic Products Co., Ltd. Vinyl Nitrile Co-Polymer Powder Free Exam Blue Gloves (K051662)
5. Device Description:
Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Powder-Free Vinyl Patient Examination Gloves, LYZ, and meets all requirements of ASTM D5250-06.
-
- Intended Use:
A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or fingers to prevent contamination between patient and examiner.(21 CFR 880.6250)
- Intended Use:
7. Comparison to Predicate Devices on Indication for Use:
Zibo Huiying Medical Products, Co. Ltd. Synmax Synthetic Patient Examination Vinyl Gloves, Powder Free, Blue is substantially equivalent in safety, and effectiveness to Grand Work Plastics Product Vinyl Nitrile Co-Polymer Powder Free Exam Blue Glove (K051662).
{4}------------------------------------------------
8. Discussion of Non-Clinical Test Performed for Determination of Substantial Equivalence are as Follows:
The standards used for Zibo Huiying Medical Products, Co. Ltd.'s gloves product are based on ASTM D-5250. All testing meets requirements for physical and dimensions testing conducted on gloves. Inspection level S-2, AOL 2.5.
The FDA 1000 ml Water Fill Test based on ASTM-D5151-06 was also conducted samplings of AQL 2.5 inspection level G-1, meeting these requirements. Primary Skin Irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted in accordance with ISO 10993-10 Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Sensitization with results showing no primary skin irritation or sensitization reactions under the conditions tested.
There are no special labeling claims and we do not claim our gloves as hypoallergenic on our labels.
A Residual Powder Test that based on ASTM D-6124 for start to finish inspection is conducted to insure that our gloves meet our "powder-free" claims (contains no more then 2 mg powder per glove).
9. Patient Contact
The glove is available for surface-contacting with less than 24 hours duration.
10. Sterilization
The subject device is non-sterile.
l 1. Discussion of Clinical Tests Performed:
Not Applicable - There is no hypoallergenic Claim.
12. Device Comparison Table
Please see below for detailed updated comparison between our device and predicate device.
13. Conclusions:
The Synmax Synthetic Patient Examination Vinyl Glove, Powder Free (Blue) is substantially equivalent to the Grand Works Plastics Vinyl Nitrile Co-Polymer Powder Free Exam Gloves (Blue). Based on the nonclinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device, Grand Works Plastics Vinyl Nitrile Co-Polymer Powder Free Exam Gloves (Blue) cleared under K051662.
{5}------------------------------------------------
Table. Side-by-Side Comparison of Intended Use, Design, Material, Physical, Biocompatibility, and Performance Testing
| Proposed Device | Predicate Device (K051662) | |
|---|---|---|
| Description | Zibo Huiying Medical Products, Co.Ltd. Synmax Synthetic ExaminationVinyl Gloves, Powder Free, Blue | Grand Work Plastics ProductVinyl Nitrile Co-PolymerPowder Free Exam BlueGloves |
| Indication for use | Cover the hand and wrist area. It is adisposable device which is worn uponthe examiner's hands or fingers toprevent contamination between patientand examiner. | Cover the hand and wrist area.It is a disposable device whichis worn upon the examiner'shands or fingers to preventcontamination between patientand examiner. |
| Labeling: Labels onthe carton | Labels include: Product name; color;"single use Only" size, piece count, lotnumber, distributor name, andmanufacturer address. | Labels include: Product name;color; "single use Only" size,piece count, lot number,distributor name, andmanufacturer address. |
| Device Materials | Poly Vinyl ChloridePolyurethaneDiisononyl Phthalate (DINP) | Poly Vinyl ChlorideNitrilePolyurethaneDiisononyl Phthalate (DINP) |
| Before Aging:TensileStrength(Mpa) andUltimateElongations | Average Tensile Strength (Mpa): 16.9Average Ultimate Elongations: 550% | Substantially EquivalentConforms to ASTM D5250 andASTM D6319 -00al.Tensile Strength: 16 mpa minElongation: 500 % min |
| After Aging: TensileStrength(Mpa) andUltimateElongations | Average Tensile Strength (Mpa): 14.4Average Ultimate Elongations: 500% | Substantially equivalentConforms to ASTM D5250 andASTM D6319 -00al.Tensile Strength: 14 mpa min,Elongation: 400 % min |
| Overall Length onMedium Size | Average over 234mm | Substantially equivalentConforms to ASTM D5250 andASTM D6319 -00al.230 min |
| Width of Palm onMedium Size | Average 96mm | Substantially equivalentConforms to ASTM D5250 andASTM D6319 -00al.95 ± 10 mm |
| Palm Thickness | Average 0.096 mm | Substantially equivalentConforms to ASTM D5250 andASTM D6319 -00al.0.05 mm min. |
| Finger Thickness | Average 0.98 mm | Substantially equivalentConforms to ASTM D5250 andASTM D6319 -00al.0.05 mm min |
| Residual Powder | According to ASTM D6124-06Standard Test Method for ResidualPowder on Medical gloves for thedetermination of residual powdercontent. Testing result indicates theweight of all types of residual orpowder on finished powder-freegloves as < 2 mg per glove and thereis no defect glove found according toASTM D6124-06. | Substantially equivalentConforms to ASTMD6124-06The weight of all types ofresidual or powder on finishedpowder-free gloves as < 2 mgper glove and there is no defectglove found |
| Pinhole Results | According to ASTM D5151-06,Testing result indicates pinhole werefound less than two pieces gloves outof 125 pieces gloves. AQL 2.5 is met. | Substantially equivalentConforms to ASTM D5151-06AQL 2.5 |
| BiocompatibilityResult: Primary SkinIrritation | Under the conditions tested, thesubject device was not an irritant | Substantially equivalentThe subject device was not anirritant |
| Dermal Sensitization | Under the conditions tested, thesubject device was not a sensitizer | Substantially equivalentThe subject device was not asensitizer |
| Summary ofcomparison | The Synmax Synthetic Patient Examination Vinyl Glove, Powder Free(Blue) is substantially equivalent to the Grand Works Plastics VinylNitrile Co-Polymer Powder Free Exam Gloves (Blue). Based on thenonclinical tests performed, the subject device is as safe, as effective,and performs as well as the legally marketed predicate device, GrandWorks Plastics Vinyl Nitrile Co-Polymer Powder Free Exam Gloves(Blue) cleared under K051662. |
{6}------------------------------------------------
:
.
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.