(316 days)
Measuring temporal characteristics of saccadic refixation responses when viewing lateral visual stimulus and identifying the individual time delays of moving the eyes toward the stimuli.
The Saccadometer system measures eye movements in horizontal axis with temporal 1kHz and spatial resolution < 5 arc min. The system incorporates a built in visual target presentation using miniature laser projectors mounted on the sensor forehead plate. Saccadometer Plus is equipped with three red laser projectors with 10° separation. Saccadometer Advanced has additional extra green laser and it allows to perform more complex saccadic task. Information about eye movement is acquired by analyzing the infrared light reflected from eye orbit. The left and right eve orbit are illuminated by two infrared LEDs, providing eve surface irradiance below 1mW/cm-(operating condition). Due to the conjugacy and synchronicity of the saccadic eye movements, the rotation of left and right eyes, can be added and averaged. The inner canthi of the left and right eyes are illuminated with the low intensity IR. The difference between the amounts of IR reflected back from the eye surfaces toward the photodetecting circuitry, carries the information about the eye position changes. The eye landmarks are taking part in generation of the eye movement signal. The corneal bulge and the limbus (the border between darker iris and white sclera). Main contributor to the eye movement signal is the corneal bulge. Being the relative eye position measurement system, it requires to define the initial eye position. It means that at the beginning of every measurement, a steady eye fixation point needs to be provided. In Saccadometer the central fixation target serves this purpose. The Saccadometer Plus and Saccadometer Advanced systems are composed of: - proprietary hardware and firmware, enclosed in - examination control unit with 2 x 1.5V AA batteries, and the - eye movement sensor with integrated miniature laser spot projectors - proprietary software application LatencyMeter. - optical fiber to USB data transmitter
The provided FDA 510(k) summary for the Saccadometer Plus and Saccadometer Advanced devices describes performance data, but does not explicitly state "acceptance criteria" in a quantitative, pass/fail manner typical for clinical trials or formal validation studies. Instead, it presents results of various tests demonstrating the device's characteristics and performance.
Therefore, the table below will list the reported device performance characteristics based on the provided text, rather than explicitly stated acceptance criteria. For questions where the information is not provided in the document, it will be stated as "Not provided."
Acceptance Criteria and Reported Device Performance
| Characteristic / Acceptance Criteria Type | Reported Device Performance |
|---|---|
| Saccade Detection Algorithm Accuracy | Mean absolute error of saccadic latency measurement is below 2 ms. The influence of parameters of pseudosaccade course (amplitude, rising edge, trigger to slope delay-simulated latency) changes on the latency of measurement error is also below the value 1ms. |
| Measurement Linearity | Mean averaged maximal linearity error for range ±15 degrees was observed. Linearity error increased with the increase of measurement ranges, but it is noted that human fixation has an error of ±0.3 degrees. The device is primarily used for ±10 degree measurement range. |
| Spatial Resolution | All gathered results were within the declared measurement resolution (5 arcmin or better). The device's eye movement sensor has a spatial resolution < 5 arc min. |
| Temporal Resolution and Bandwidth | Eye movement measurement bandwidth (defined as -3dB amplitude loss) is 0 - 200 Hz. Measurement temporal resolution - digitized signal output rate- is 1000 Hz. The device has a temporal resolution of 1kHz. |
| Reliability (Test-Retest) | Analysis did not reveal any significant difference between the saccadic parameters measured in the test experiment and retest. Results showed high test-retest reliability and did not differ significantly within the session. |
| Electrical Safety and EMC | Complies with standard IEC 60601-1 and 60601-1-2:2014 (emission and immunity). |
| Biocompatibility | Primary component (polyamide/synthetic rubber elastic tape) has OEKO-TEX 100 certificate. Other materials (silicone nose pads, PVC cable insulation) were used in the predicate device (Ober2). No new issues of safety or effectiveness are introduced. |
| IR Photoemission Safety | Both corneal and retinal exposure are at a safe level, based on IEC 62471 standard. |
| Laser Safety | Power of the lasers doesn't exceed Class II. |
| Dimension Stability | Saccadometer housing exposed to 70°C for 7 hours did not significantly change dimensions. |
| Predicated Equivalence | The Saccadometer Plus and Saccadometer Advanced are substantially equivalent to the predicate device (Permobil Meditech Ober2) given the similar infrared technology, high 1kHz sampling rate, and demonstrated safety and effectiveness profiles. Differences relate to integrated stimuli presentation and improved electrical safety (low voltage batteries, optical fiber isolation), not core functionality. Conducted studies revealed that Saccadometer allows the safe and effective measurement of temporal characteristics of saccadic refixation responses. |
Study Details
-
Sample size used for the test set and the data provenance:
- Saccade Detection Algorithm Accuracy: Test environment allowing simulation of saccade response with given parameters on a photoelectric converter. This implies a simulated, rather than human, test set. No human subject sample size is provided for this specific test.
- Measurement Linearity Testing: "12 subjects aged 22-69."
- Spatial Resolution of eye movement measurement: "12 subjects (aged 22-69)."
- Reliability study: "17 subjects aged 22-69."
- Data Provenance: For human subject studies (linearity, spatial resolution, reliability), it's explicitly stated: "Studies were performed only outside United States." This indicates the data is retrospective or prospective, but no further detail is given. For the algorithm accuracy, it was a simulated environment.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Saccade Detection Algorithm Accuracy: The ground truth was established by the simulated "pseudosaccade" parameters in a controlled test environment. No human experts were involved in establishing ground truth for this test.
- Measurement Linearity, Spatial Resolution, Reliability: The document does not describe the establishment of ground truth by human experts for these performance tests. The tests appear to measure the device's agreement with expected physical properties or its own consistency, rather than comparing it against a human interpretation of data.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- No adjudication method is described for any of the performance tests. The tests focus on device-centric performance metrics rather than interpretative output requiring human adjudication.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC comparative effectiveness study was done. The device (Saccadometer) is an "Eye Movement Monitor" that objectively measures characteristics; it is not described as an AI-assisted diagnostic tool that human readers would use to improve their performance in interpreting images or data.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Yes, the "Experimental verification of the saccade detection algorithm" was a standalone test of the device's algorithm performance using a simulated environment. Other tests like measurement linearity, spatial resolution, and temporal resolution also assess the device's standalone performance, though they involved human subjects whose eye movements were measured by the algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Saccade Detection Algorithm Accuracy: Simulated "pseudosaccade" parameters provided as input to the photoelectric converter.
- Measurement Linearity, Spatial Resolution, Reliability: Ground truth for these characteristics appears to be implicitly defined by the physical or physiological processes being measured (e.g., actual target displacement for linearity, intrinsic eye movement characteristics for resolution and reliability), which the device then attempts to quantify. It's not based on expert consensus, pathology, or outcomes data.
-
The sample size for the training set:
- The document does not mention machine learning or AI models with distinct "training sets." The device's "saccade detection algorithm" seems to be based on established algorithms for eye movement analysis rather than requiring a specifically labeled training dataset in the modern sense of deep learning.
-
How the ground truth for the training set was established:
- As no training set is described for a machine learning model, this information is not applicable.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an abstract symbol that resembles three human profiles facing right, stacked on top of each other, with a stylized wing-like shape extending from the top profile.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 11, 2016
Ober Consulting Sp. Z.o.o. Mr. Jan Krzysztof Ober Brzechwy 6, 60-195 Poznan Poland
Re: K152890 Trade/Device Name: Saccadometer Plus and Saccadometer Advanced Regulation Number: 21 CFR 886.1510 Regulation Name: Eye Movement Monitor Regulatory Class: Class II Product Code: HLL Dated: July 8, 2016 Received: July 11, 2016
Dear Mr. Ober:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Alexander
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K152890
Device Name Trade / family name: Saccadometer model name: Plus, Advanced
Indications for Use (Describe)
Measuring temporal characteristics of saccadic refixation responses when viewing lateral visual stimulus and identifying the individual time delays of moving the eyes toward the stimuli.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
EF
{3}------------------------------------------------
510(k) Summary
06 July 2016
I. SUBMITTER
Submitter's contact: Mr. Jan Krzysztof Ober Submitter's name: Ober Consulting sp. z o.o. Submitter address and contact data: Brzechwy 6, 60-195 Poznan Poland +48 505940787 info@ober-consulting.com
II. DEVICE
There is no prior submission for the device. The assigned 510(k) number is K152890/S001.
Proprietary name: Saccadometer Plus, Saccadometer Advanced
Common or usual name: Saccadometer
Classification name: EYE MOVEMENT MONITOR (21 CFR 886.1510)
Classification code: HLL
III. PREDICATE DEVICE
Permobil Meditech, Ober2, K902954
VI. DEVICE DESCRIPTION
The Saccadometer system measures eye movements in horizontal axis with temporal 1kHz and spatial resolution < 5 arc min. The system incorporates a built in visual target presentation using miniature laser projectors mounted on the sensor forehead plate. Saccadometer Plus is equipped with three red laser projectors with 10° separation. Saccadometer Advanced has additional extra green laser and it allows to perform more complex saccadic task.
Information about eye movement is acquired by analyzing the infrared light reflected from eye orbit. The left and right eve orbit are illuminated by two infrared LEDs, providing eve surface irradiance below 1mW/cm-(operating condition). Due to the conjugacy and synchronicity of the saccadic eye movements, the rotation of left and right eyes, can be added and averaged. The
{4}------------------------------------------------
inner canthi of the left and right eyes are illuminated with the low intensity IR. The difference between the amounts of IR reflected back from the eye surfaces toward the photodetecting circuitry, carries the information about the eye position changes. The eye landmarks are taking part in generation of the eye movement signal. The corneal bulge and the limbus (the border between darker iris and white sclera). Main contributor to the eye movement signal is the corneal bulge. Being the relative eye position measurement system, it requires to define the initial eye position. It means that at the beginning of every measurement, a steady eye fixation point needs to be provided. In Saccadometer the central fixation target serves this purpose.
The Saccadometer Plus and Saccadometer Advanced systems are composed of:
- proprietary hardware and firmware, enclosed in
- examination control unit with 2 x 1.5V AA batteries, and the
- eye movement sensor with integrated miniature laser spot projectors
- proprietary software application LatencyMeter.
- optical fiber to USB data transmitter
Saccadometer's control unit is made of ABS material and it is not used by the tested subject. Saccadometer uses only three materials that have direct contact with head skin:
- a) Elastic tape with OEKO-TEX 100 certificate
- b) Silicone nose pads
- c) Ring made from M90 acetal (POM) homopolymer plastic
- d) Flex flat cable insulation material: polyvinyl chloride (PVC)
Saccadometers are supplied in a plastic non sterile boxes filled with a sponge for mechanical safety.
Saccadometer is non sterile device and cannot be sterilized with high temperature processes.
The device is worn on head so it have direct contact with the skin, hairs and sweat. The proper method for cleaning Saccadometer surfaces (having contact with skin) is the disinfection with the use of cotton wipe moistured with ethyl alcohol.
Saccadometer Plus and Saccadometer Advanced is Rx only device. Manual includes the information that the device is for prescription only. The label with the statement (Rx Only) is placed on control unit of Saccadometer (small label) as well as the on plastic box in which Saccadometer is supplied (large label). Device is also labeled by Laser warning label (placed on sensor), Laser class label (sensor), Serial number label (control unit, box), Laser details label on control unit and box, Battery label (control unit, box).
V. INDICATION FOR USE
Indication for use:
Measuring temporal characteristics of saccadic refixation responses when viewing lateral visual stimulus and identifying the individual time delays of moving the eyes toward the stimuli.
{5}------------------------------------------------
VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Ober Consulting has determined that the Saccadometer Plus and Saccadometer Advanced is substantially equivalent to the predicate device Permobil Meditech Ober2.The main similarity between Saccadometer Plus/Advanced and Permobil Meditech Ober2 is the same infrared technology used to capture horizontal axis eye movements data with the same high 1kHz sampling rate. The eye movement sensor uses the same material - the PCB board, for carrying the IR illuminators, photodetectors and electronics for signal amplification and IR illumination control.
There are following differences between Saccadometer and Permobil Meditech Ober2:
-
Ober2 can be used with any external stimuli source,
-
Saccadometer Plus/Advanced have its own miniature laser spot projectors providing fixed separation stimuli angle and is specially designed for measuring latency of eyes movements in response to visually stimuli.
Saccadometer Plus/Advanced design provides higher electrical safety standards from its predicate by using low voltage AA batteries (no connection to AC supply) and by using optical fiber isolation from computer's USB port when uploading eye movements data.
VII. PERFORMANCE DATA
Following performance data were provided in support the substantial equivalence determination.
Biocompatibility testing
Primary component that have direct contact with forehead skin (a) is made from polyamide 86,5% and synthetic rubber 13,5% and have OEKO-TEX 100, that means it is free of harmful substances and meet human-ecological requirements for product with direct skin contact. The other materials include silicone nose pads (b) which have contact with nose bridge skin and flex flat cable made form PVC which can touch ear skin, head skin or hairs.
Saccadometer Plus/Advanced utilizes the materials that were used in Ober2. No new issues of safety or effectiveness are introduced by using Saccadometer device.
Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on Saccadometer Plus and Saccadometer Advanced. The Saccadometer Plus and Saccadometer Advanced, complies with standard IEC 60601-1.IEC standards for safety and the 60601-1-2:2014 (emission and immunity).
Dimension Stability of Saccadometer Enclosure in High Temperature
Saccadometer housing was exposed to a temperature of 70 deg of 7 hours duration. High temperatures did not significantly change the dimensions of the enclosure
Infrared (IR) photoemission analysis
Saccadometer is using the eye movement measuring technology, based on infrared illumination. For this reason we evaluated the effects of the used radiation (940 nm) on the safety of the person whose eye movement is measured. Our calculations and measurements are based on IEC
{6}------------------------------------------------
62471 standard: "Photobiological safety of lamps and lamp systems". Both corneal and retinal exposure are at safe level.
Lasers safety
Power of the lasers doesn't exceed the Class II and is verified at the stage of designing and manufacturing..
Experimental verification of the saccade detection algorithm
To verify the correctness of the Saccadometer algorithm for automatic saccade detection and measurement of saccadic latency we developed test environment, allowing to simulate the saccade response with given parameters on the photoelectric converter. Beside the evaluation of the latency measurement error we decided also to assess the impact of changes of pseudosaccade parameters on this error. The results showed that the mean absolute error of saccadic latency measurement is below 2 ms. The influence of parameters of pseudosaccade course (amplitude, rising edge, trigger to slope delay-simulated latency) changes on the latency of measurement error is also below the value 1ms.
Measurement linearity testing
Evaluation of linearity was performed based on the results of 12 subjects aged 22-69. Studies were performed only outside United States. The experiment was performed in the dark room using the head immobilization, visual stimuli was displayed on the LCD screen. Target was displayed randomly in one of 15 possible (0, ±5, ±20, ±25, ±30, ±35) horizontal location, subjects were asked to move their gaze following the target displacements.
Results showed that mean averaged maximal linearity error for range ±15 degree. Linearity error increased with the increase of measurement ranges, however it needs to notice that human fixation is not ideally accurate and it is characterized by the error ±0,3 degree.
Moreover it requires to point that experiments implemented on Saccadometer Plus/Advanced, require only ±10 degree of measurement range.
Spatial resolution of eve movement measurement
Resolution of measurement was calculated for eye position data of 12 subjects (aged 22-69), which took part in measurement linearity testing. All gathered results were within the declared measurement resolution (5 arcmin or better).
Temporal resolution and bandwith
The Saccadometer's eye movement measurement bandwith was measured using optical simulator described in "Experimental verification of the saccade detection algorithm". Simulator's IR emitters were driven by sinusoidal wave of controlled frequency. Amplitude change in digitized signal was observed. According to achieved results Saccadometer's eye movement measurement bandwidth (defined as -3dB amplitude loss) is 0 - 200 Hz. Measurement temporal resolution - digitized signal output rate- is 1000 Hz.
Reliability study
Step task experiment (100 trials) was performed on 17 subjects aged 22-69. Studied performed only outside United States. The same experiment was repeated after 5 minutes break (test and
{7}------------------------------------------------
after 5 minutest retest). For every person we calculated (using LatencyMeter software) mean saccadic latency and coefficient of variation (CV) as well as mean saccadic duration for test and retest results. We evaluated test-retest reliability, between-session effect, as well as withinsession effects. Evaluation of within-session effect was performed only on the results of first measurement (test). Analysis did not revealed any significant difference between the saccadic parameters measured in the test experiment. Results showed high test-retest reliability and did not differ within the session.
Summary
Results of conducted studies demonstrate that the safety and effectiveness profile of Saccadometer is similar to the predicate device.
VIII. CONCLUSIONS
Based upon the 510(k) summaries and statements (21 CFR 807) and information provided herein, Ober Consulting conclude that the Saccadometer Plus and Saccadometer Advanced are substantially equivalent to the predicate device under the Federal Food, Drug and Cosmetic Act. Conducted studies reveled that Saccadometer allows the safe and effective measurement of temporal characteristics of saccadic refixation responses.
§ 886.1510 Eye movement monitor.
(a)
Identification. An eye movement monitor is an AC-powered device with an electrode intended to measure and record ocular movements.(b)
Classification. Class II.