(264 days)
Bed linens made with DermaTherapy® fabrics are intended for use by patients 18 years of age and older in hospital settings who are susceptible to pressure ulcers. The DermaTherapy Bed Linens help to reduce the likelihood of patients developing pressure ulcers by reducing moisture, friction and shear on the patient's skin.
DermaTherapy uses a silk-like fabric to minimize moisture, friction, and shear between the skin and the bedsupport surface. Continuous-filament yarns woven into the silk-like synthetic DermaTherapy fabric provide a smooth surface, free of broken or discontinuous fibers. This enhanced smoothness helps minimize the potential for irritation and abrasion of sensitive skin.
DermaTherapy Bed Linens consists of pillow cases, top flat sheets, and underpads. The subject devices are made from DermaTherapy fabrics which are plain-weave constructions of 100% continuous-filament yarns. The polyester yarns have a non-round fiber cross-section to create micro-channels to facilitate moisture wicking and rapid drying. The yarns used in DermaTherapy fabrics are commercially available products, typically used in apparel.
The only technological difference in the fabric construction between the subject device and the predicate device is the addition of a secondary static-dissipative polyester yarn at 0.8% of the total fabric weight. The static-dissipative yarn is incorporated to reduce the potential for static electricity that may occur during institutional laundering processes. The company has made no other modifications to the fabric or the manufacturing process of turning the fabric into the final product.
This document describes data for a medical device called "DermaTherapy Bed Linens," which are intended to reduce the likelihood of pressure ulcers in hospital patients. Here's a breakdown of the requested information:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state "acceptance criteria" with numerical targets for each performance characteristic. Instead, it lists the "Performance Characteristics" and the "Test Method / Standard" used to evaluate the device. The conclusion states that "bench testing demonstrated that the device met performance specifications" and "demonstrated substantially equivalent device performance." It also includes clinical study results showing a reduction in pressure ulcer incidence.
Given the information, the table below will combine the performance characteristics (which can be inferred as criteria for evaluation) and the clinical performance results.
| Performance Characteristic | Test Method / Standard (Acceptance Criteria are implied by meeting these standards) | Reported Device Performance |
|---|---|---|
| Weave Pattern | Visual | Not explicitly stated, implied to meet visual standards. |
| Weight | ASTM-D-3776 | Not explicitly stated, implied to meet standards. |
| Ends | ASTM-D-3775 | Not explicitly stated, implied to meet standards. |
| Picks | ASTM-D-3775 | Not explicitly stated, implied to meet standards. |
| Grab Tensile - Warp & Fill | ASTM-D-5034 | Not explicitly stated, implied to meet standards. |
| Tongue Tear - Warp & Fill | ASTM-D-2261 | Not explicitly stated, implied to meet standards. |
| Circular Bend | ASTM-D-4032 | Not explicitly stated, implied to meet standards. |
| Pore Size | ASTM-E-1294 | Not explicitly stated, implied to meet standards. |
| Moisture Regain | ASTM-D-2654 | Not explicitly stated, implied to meet standards. |
| Geometric Roughness | Kawabata KES-FB4 Surface Tester (measured in microns) | Not explicitly stated, implied to meet intended smoothness. |
| Fabric Wicking Rate | Distance water travels in five minutes (measured in millimeters) | Not explicitly stated, implied to meet intended wicking rate. |
| % Dry after 0-60 minutes | % moisture loss of fabric until dryness (measured in 15-minute intervals) | Not explicitly stated, implied to meet intended rapid drying. |
| Coefficient of Friction | Kawabata KES-FB4 Surface Tester (values of 0 to 1) | Not explicitly stated, implied to meet intended low friction. |
| Biocompatibility | ISO 10993 consensus standards (Cytotoxicity, sensitization, and irritation testing) | Test results demonstrated biocompatibility. |
| Antimicrobial Agent Leaching | PFG conducted leachability study | Test results demonstrated substantially equivalent amount of antimicrobial agent leaching, with no safety risk to the patient. |
| Reduction in Pressure Ulcers (Clinical) | Clinical studies comparing control vs. DermaTherapy | Study 1: Control: 12.3%, DermaTherapy: 4.6% (P = .01) Study 2: Control: 7.5%, DermaTherapy: 0.0% (P = .01) Study 3: Control: 11.5%, DermaTherapy: 3.1% (P < .001) (Consistent reduction in incidence of facility-acquired pressure ulcers) |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size (Clinical Test Set): A total of 2,009 study subjects across three studies.
- Study 1: 307 patients
- Study 2: 275 patients
- Study 3: 1,427 patients
- Data Provenance: The studies were conducted in an "acute care setting at multiple healthcare sites," implying they were conducted in clinical environments. The document does not specify the country of origin but is submitted to the FDA, suggesting it is either US-based or intended for US market approval. The nature of "clinical trials" and "comparison of data between study and control groups" suggests these were prospective studies.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
The document does not provide information on the number or qualifications of experts used to establish the ground truth for the clinical studies (e.g., diagnosis of pressure ulcers).
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
The document does not describe any adjudication method used for the clinical test set.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- This device is bed linens, not an AI-powered diagnostic tool. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and was not performed.
- The clinical studies performed were comparative effectiveness studies of the bed linens themselves, showing a reduction in pressure ulcer incidence. The "effect size" can be inferred from the percentage reduction in pressure ulcers in the DermaTherapy group compared to the control group (e.g., Study 1 showed an 8.3% absolute reduction: 12.3% - 4.6%).
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This point is not applicable as the device is not an algorithm or AI-powered system.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the clinical studies, the ground truth was the incidence of facility-acquired pressure ulcers. This is a form of outcomes data, likely determined through clinical assessment and documentation within the hospital settings.
8. The sample size for the training set
This is not applicable. The device is not an AI algorithm that requires a "training set" in the machine learning sense. The "training" that occurs is the manufacturing process of the linens to meet specific physical characteristics.
9. How the ground truth for the training set was established
This is not applicable for the same reason as point 8.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized caduceus or a family of three people.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 20, 2016
Precision Fabrics Group, Inc. % Mr. Jonathan Kahan Partner Hogan Lovells US LLP 555 13th Street NW Washington, DC 20004
Re: K152884
Trade/Device Name: DermaTherapy Bed Linens Regulation Number: 21 CFR 880.6190 Regulation Name: Mattress Cover for Medical Purposes Regulatory Class: I Product Code: FMW Dated: May 20, 2016 Received: May 20, 2016
Dear Mr. Kahan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -
for Erin I. Keith, M.S Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
Device Name
DermaTherapy Bed Linens
Indications for Use (Describe)
Bed linens made with DermaTherapy® fabrics are intended for use by patients 18 years of age and older in hospital settings who are susceptible to pressure ulcers. The DermaTherapy Bed Linens help to reduce the likelihood of patients developing pressure ulcers by reducing moisture, friction and shear on the patient's skin.
Type of Use (Select one or both, as applicable)
□ Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{3}------------------------------------------------
510(k) SUMMARY K152884
Precision Fabrics Group's DermaTherapy Bed Linens
Submitter Information:
| Applicant: | Precision Fabrics Group, Inc.301 North Elm Street, Suite 600Greensboro, NC 27401 |
|---|---|
| Phone: | (336) 510-8009 |
| Fax: | (336) 510-8003 |
| Contact Person: | Terry Montgomery, PhD, Vice President |
| Date Prepared: | June 15, 2016 |
Subject Device:
| Device Name: | DermaTherapy Bed Linens |
|---|---|
| Common Name: | Bed linens |
| Classification Name: | Mattress Cover for Medical Purposes |
| Regulation: | Class I, 21 C.F.R. § 880.6190 |
| Product Code: | FMW |
| Predicate Devices: |
Primary Predicate: Precision Fabrics Group's Institutional Bedding made with DermaTherapy Fabrics (K061242)
Indications for Use:
Bed linens made with DermaTherapy fabrics are intended for use by patients 18 years of age and older in hospital settings who are susceptible to pressure ulcers. The DermaTherapy Bed Linens help to reduce the likelihood of patients developing pressure ulcers by reducing moisture, friction and shear on the patient's skin.
{4}------------------------------------------------
Device Description
DermaTherapy uses a silk-like fabric to minimize moisture, friction, and shear between the skin and the bedsupport surface. Continuous-filament yarns woven into the silk-like synthetic DermaTherapy fabric provide a smooth surface, free of broken or discontinuous fibers. This enhanced smoothness helps minimize the potential for irritation and abrasion of sensitive skin.
DermaTherapy Bed Linens consists of pillow cases, top flat sheets, and underpads. The subject devices are made from DermaTherapy fabrics which are plain-weave constructions of 100% continuous-filament yarns. The polyester yarns have a non-round fiber cross-section to create micro-channels to facilitate moisture wicking and rapid drying. The yarns used in DermaTherapy fabrics are commercially available products, typically used in apparel.
The only technological difference in the fabric construction between the subject device and the predicate device is the addition of a secondary static-dissipative polyester yarn at 0.8% of the total fabric weight. The static-dissipative yarn is incorporated to reduce the potential for static electricity that may occur during institutional laundering processes. The company has made no other modifications to the fabric or the manufacturing process of turning the fabric into the final product.
Substantial Equivalence
The DermaTherapy Bed Linens are substantially equivalent to the primary predicate device, Institutional Bedding made with DermaTherapy Fabrics (K061242). Both devices have intended use as mattress covers that protect the skin, and similar technological characteristics. The new indications for use do not raise new questions of safety and effectiveness. The additional static-dissipative yarn does not affect the intended use or the device performance. Furthermore, bench testing demonstrated that the device met performance specifications, which support the new indications for use. Thus, the subject device is substantially equivalent to the primary predicate devices.
A summary of substantial equivalence between the subject, primary predicate, and reference devices are provided in Table 1 below.
| Subject Device:DermaTherapy Bed Linens | Primary Predicate Device:Institutional Bedding made with DermaTherapyfabrics (K061242) | |
|---|---|---|
| Indications for Use | Bed linens made with DermaTherapy® fabrics areintended for use by patients 18 years of age andolder in hospital settings who are susceptible topressure ulcers. The DermaTherapy Bed Linens helpto reduce the likelihood of patients developingpressure ulcers by reducing moisture, friction andshear on the patient's skin. | Institutional Bedding (bed sheets and pillow cases)made with DermaTherapy fabrics is intended for useby patients in a hospital, healthcare or home settingwho are susceptible to or may have mild atopicdermatitis. |
| User Population | Patients in hospital settings | Patients in hospital and long-term care settings |
Table 1: Substantial Equivalence Comparison
{5}------------------------------------------------
| Subject Device:DermaTherapy Bed Linens | Primary Predicate Device:Institutional Bedding made with DermaTherapyfabrics (K061242) | |
|---|---|---|
| Classification | Mattress Cover for Medical PurposesClass I21 CFR 880.6190Product Code FMW | Mattress Cover for Medical PurposesClass I21 CFR 880.6190Product Code FMW |
| TechnologicalCharacteristics | DermaTherapy fabrics made of nylon, polyester, andstatic-dissipating yarn with antimicrobial treatment | DermaTherapy fabrics made of nylon and polyesteryarn with antimicrobial treatment |
| Features | Bed linens designed to reduce moisture, friction andshear on the patient's skin.Fabric is undyed in natural white color. | Bed linens designed to reduce moisture, friction andshear on the patient's skin.Fabric is undyed in natural white color. |
| Components | Pillow cases, top flat sheets, bottom fitted sheets,and underpadsAntimicrobial agent: quaternary ammoniumcompound (3-Trimethoxy silyl propyl dimethyloctadecyl ammonium chloride), is applied to theDermaTherapy fabric | Pillow cases, top flat sheets, and bottom fitted sheetsAntimicrobial agent: quaternary ammoniumcompound (3-Trimethoxy silyl propyl dimethyloctadecyl ammonium chloride), is applied to theDermaTherapy fabric |
| Dimensions | Pillow Case at 21" width x 32" length (Standard).Top Flat Sheet at 70" width x 108" length (XL Twin).Bottom Fitted Sheet at 39" width x 81" length x 15"height (XL Twin).Underpad at 32″ width x 30″ length. | Pillow Case at 21" width x 32" length (Standard).Top Flat Sheet at 70" width x 108" length (XL Twin).Bottom Fitted Sheet at 39" width x 81" length x 15"height (XL Twin). |
| Biocompatibility | Cytotoxicity, sensitization, and irritation testing | Cytotoxicity, sensitization, and irritation testing |
| Sterilization | Not sterile | Not sterile |
Performance Data
Extensive bench testing has been performed to various standards, both industry- and company-specific, to ensure that the device meets the device performance specifications at the beginning and at the end of its use life. A summary of the testing conducted is provided in Table 2 below:
Table 2: Performance Characteristics and Test Methodology
| Performance Characteristics | Test Method / Standard |
|---|---|
| Weave Pattern | Visual |
| Weight | ASTM-D-3776 |
| Ends | ASTM-D-3775 |
| Picks | ASTM-D-3775 |
| Grab Tensile - Warp & Fill | ASTM-D-5034 |
| Tongue Tear - Warp & Fill | ASTM-D-2261 |
| Circular Bend | ASTM-D-4032 |
| Pore Size | ASTM-E-1294 |
| Moisture Regain | ASTM-D-2654 |
{6}------------------------------------------------
| Performance Characteristics | Test Method / Standard |
|---|---|
| Geometric Roughness | The surface contour (geometric roughness) is determined using the Kawabata KES-FB4Surface Tester. Measurements are made using a standard specimen size of 20 x 20 cm inthree replications. Using a calibrated surface probe, geometric roughness is measured inmicrons. Higher values correspond to a geometrically rougher surface. |
| Fabric Wicking Rate - Warp & Fill | A determination of the rate at which a fabric wicks water, measured as the distance watertravels in five minutes. A 1" x 6" sample is marked and lowered into deionized water. Thewicking rate is calculated by determining the distance, measured in millimeters, waterrises along the sample fabric in a 5-minute period. |
| % Dry after 0 - 60 minutes | A determination of % moisture loss of a fabric until dryness, measured in 15-minuteintervals up to 60 minutes, whichever comes first. An 8"x10" fabric sample is tested atintervals. The rate is calculated based upon the time it takes for the sample to return to itsoriginal dry weight. |
| Coefficient of Friction | The surface properties of friction (resistance/drag) are determined using the KawabataKES-FB4 Surface Tester. Measurements are made using a standard specimen size of 20 x20 cm in three replications. Using a calibrated friction probe, coefficient of friction (COF)values of 0 to 1 are determined, with the higher COF value corresponding to higherfriction. |
Cytotoxicity, intracutaneous reactivity, and skin sensitization testing were conducted according to ISO 10993 consensus standards to evaluate the biocompatibility of the device at various points of its use life.
PFG conducted a leachability study to quantify and to analyze the antimicrobial agent leaching from the subject device and the predicate device. Test results demonstrated that the subject device exhibited substantially equivalent amount of antimicrobial agent leaching from the device. In addition, a risk assessment of the antimicrobial agent completely leaching from the product demonstrated that there is no safety risk to the patient.
Clinical Data
To further demonstrate the substantial equivalence and performance of the device in reducing the likelihood of patients developing pressure ulcers, the company has conducted three studies since 2008 that included a total of 2,009 study subjects. A summary of the studies is provided in Table 3 below. Study subjects included patients 18 years of age and older in acute care setting at multiple healthcare sites including medical renal, urology, surgical intensive care, non-surgical intensive care, and telemetry units. Comparison of data between the study and control groups showed a consistent reduction in the incidence of facility-acquired pressure ulcers in a wide range of clinical settings and patient populations. No product-related adverse events or complications were reported over the entire study period.
| Study | Clinical Trials | Total Patients | Control | DermaTherapy | P |
|---|---|---|---|---|---|
| 1. | Acute Care:Medical Renal Unit | 307 | 12.3% | 4.6% | .01 |
| 2. | Acute Care:Surgical ICU | 275 | 7.5% | 0.0% | .01 |
| 3. | Acute Care:Telemetry/Urology/ICU | 1,427 | 11.5% | 3.1% | <.001 |
{7}------------------------------------------------
Conclusions
Precision Fabrics Group believes that DermaTherapy Bed Linens are substantially equivalent to the primary predicate device Institutional Bedding made with DermaTherapy fabrics (K061242). Both devices have the same intended use and similar indications for use. The only difference to the predicate device is the addition of static-dissipative polyester yarn. Bench testing has demonstrated substantially equivalent device performance and supports the expanded indications for use to reduce the formation of pressure ulcers. Therefore, the subject device is substantially equivalent to the primary predicate device.
§ 880.6190 Mattress cover for medical purposes.
(a)
Identification. A mattress cover for medical purposes is a device intended for medical purposes that is used to protect a mattress. It may be electrically conductive or contain a germicide.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. If the device is not labeled or otherwise represented as sterile, it is also exempt from the current good manufacturing practice requirements of the quality system regulation in part 820 of this chapter, with the exception of § 820.180, with respect to general requirements concerning records, and § 820.198, with respect to complaint files.