(160 days)
The Propeller System includes the Propeller Sensor is an accessory device intended for singlepatient use to assist physicians and patients in recording and monitoring the actuations of prescribed DPI usage for the Diskus devices.
The Propeller Mobile Application records, stores, and transmits usage events from Propeller Sensors, or via manual user entry, to a remote storage system. With the Propeller Mobile Application the user can review information collected from the DPI sensor, and review symptoms and other information about their disease management and its impact. The user may also share their information with their caregivers, physician, and healthcare providers.
The Propeller Web Application is software that, like the Propeller Mobile Application, is intended to allow users to review the collected information and characteristics of their DPI and its use, to capture other patient-reported information and outcomes, and to allow that information to be shared with their caregivers, physicians, and health care providers.
When used with a prescribed DPI, the system can report on information captured during the normal course of use, such as the time between actuations that can be helpful in assessing DPI technique.
The Propeller System is intended to be used in populations from Child (>2 years) to Adult.
The Propeller System can be used both indoors home, work, and clinical settings, as well as on aircraft.
The Propeller System may also be used in clinical trials where researchers need to know information about the use of DPI medication(s) by a participant.
The output of the Propeller System is not intended to diagnosis provided by a licensed physician. The Propeller System is not intended for use as an DPI dose counter, nor is it intended to indicate the quantity of medication remaining in an DPI.
Dry Powder Inhaler (DPI) / Diskus Accessory that monitors inhaler usage. The portable sensor mounts on top of a DPI and records inhalation events when the medication is used.
This document (K152882) describes the 510(k) premarket notification for the Propeller Sensor Model 2014-D. It explicitly states that no new testing was performed for the OTC indication, and no clinical testing was required. The device's substantial equivalence is based on its identity to a previously cleared predicate device (K143671) and another (K142516) with only a change from MDI to DPI.
Therefore, the detailed information requested regarding acceptance criteria, study sample sizes, ground truth establishment, expert involvement, and MRMC studies are not available in this specific regulatory submission as no new performance studies were conducted for this 510(k) clearance.
However, based solely on the provided document, here's what can be inferred and what remains unknown regarding device performance validation:
1. Table of Acceptance Criteria and Reported Device Performance:
Since no new performance testing was conducted, there are no specific numeric acceptance criteria or reported device performance metrics presented in this document for the Propeller Sensor Model 2014-D in its OTC indication. The primary "performance" assessed was functional equivalence to the predicate.
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Device continues to meet specified requirements for functionality (e.g., recording and monitoring DPI actuations). | "Validation testing including EMC, electrical, safety, mechanical durability, and functional testing has been completed and confirms that the device continues to meet the specified requirements for the change from 'Prescription Use' to 'OTC' status." |
| No new safety concerns with OTC indication. | "Hazard Analysis ... was identical to the predicate (K143671). ... No new concerns of safety with the proposed OTC indication were found." |
| Software verification and validation. | "Software verification and validation testing confirms this result." (referring to compliance with predetermined specifications) |
| Tech is identical to predicate. | "The Propeller Sensor Model 2014-D is identical to the predicate (K143671), and therefore there are no technology differences." |
| Effective and safe as predicate. | "The overall testing confirms that the Propeller Sensor Model 2014-D is as safe and as effective as the predicate device." |
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size: Not applicable/not provided for new performance testing, as none was conducted specific to this 510(k). The functional testing performed was "bench testing."
- Data Provenance: Not applicable/not provided for new performance testing. The document refers to prior testing for predicate devices.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
- Not applicable, as no new performance studies with ground truth establishment were conducted for this 510(k). Hazard analysis involved review by (unspecified) experts during the predicate device clearance process.
4. Adjudication Method for the Test Set:
- Not applicable, as no new performance studies were conducted.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:
- No, an MRMC study was not done. The device does not involve human interpretation of medical images or data that would necessitate such a study. It is an accessory to monitor inhaler usage.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- The "functional testing" performed would be considered analogous to standalone performance, as it tests the device's ability to record actuations independently. However, no specific metrics like accuracy, sensitivity, or specificity are provided for this functional test for the current submission, as it relied on the established performance of the predicate. The device's primary function is data acquisition and transmission, not algorithmic interpretation requiring performance metrics typical of AI/ML devices.
7. The Type of Ground Truth Used:
- For the predicate device's original clearance, functional ground truth would have been established by comparing the device's recorded actuations against directly observed or mechanically triggered actuations of the DPI/MDI. This document relies on the ground truth established for the predicate device.
8. The Sample Size for the Training Set:
- Not applicable. This device is not described as utilizing a machine learning algorithm that requires a training set in the context of this 510(k). Its function is based on sensor detection, not pattern recognition learned from data.
9. How the Ground Truth for the Training Set Was Established:
- Not applicable, as there is no mention of a training set for a machine learning algorithm in this submission.
In summary, this 510(k) clearance (K152882) for the Propeller Sensor Model 2014-D focused on demonstrating substantial equivalence to previously cleared predicate devices (K143671 and K142516) and a change in indication (Prescription Use to OTC). It explicitly states that no new clinical or performance testing was required or performed for this specific submission. The acceptance criteria were implicitly met by demonstrating that the device is identical in technology and functionality to the predicate and that the change to OTC status did not introduce new safety concerns based on hazard analysis and bench testing.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of three human profiles facing right, layered on top of each other to create a sense of depth and unity. The profiles are rendered in a dark color, contrasting with the white background.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 8, 2016
Reciprocal Labs Mr. David Hubanks VP Operations 634 W. Main Street, Suite 102 Madison, Wisconsin 53703
Re: K152882
Trade/Device Name: Propeller Sensor Model 2014-D Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: Class II Product Code: CAF Dated: February 4, 2016 Received: February 5. 2016
Dear Mr. David Hubanks:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Erin I. Keith -S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K152882
Device Name Propeller Sensor Model 2014-D
Indications for Use (Describe)
The Propeller System includes the Propeller Sensor is an accessory device intended for singlepatient use to assist physicians and patients in recording and monitoring the actuations of prescribed DPI usage for the Diskus devices.
The Propeller Mobile Application records, stores, and transmits usage events from Propeller Sensors, or via manual user entry, to a remote storage system. With the Propeller Mobile Application the user can review information collected from the DPI sensor, and review symptoms and other information about their disease management and its impact. The user may also share their information with their caregivers, physician, and healthcare providers.
The Propeller Web Application is software that, like the Propeller Mobile Application, is intended to allow users to review the collected information and characteristics of their DPI and its use, to capture other patient-reported information and outcomes, and to allow that information to be shared with their caregivers, physicians, and health care providers.
When used with a prescribed DPI, the system can report on information captured during the normal course of use, such as the time between actuations that can be helpful in assessing DPI technique.
The Propeller System is intended to be used in populations from Child (>2 years) to Adult.
The Propeller System can be used both indoors home, work, and clinical settings, as well as on aircraft.
The Propeller System may also be used in clinical trials where researchers need to know information about the use of DPI medication(s) by a participant.
The output of the Propeller System is not intended to diagnosis provided by a licensed physician. The Propeller System is not intended for use as an DPI dose counter, nor is it intended to indicate the quantity of medication remaining in an DPI.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
510(k) Summary
| Submission Date: | February 4, 2016 |
|---|---|
| Submitter: | Reciprocal Labs Corporation634 W. Main Street, Ste. 102Madison, WI 53703 |
| Submitter andOfficial Contact: | David HubanksVP OperationsReciprocal Labs634 W. Main Street, Ste. 102Madison, WI 53703+1 (608) 251-0470+1 (608) 338-0883 (fax)david.hubanks@propellerhealth.com |
| ManufacturingSite: | Reciprocal Labs Corporation634 W. Main Street, Ste. 102Madison, WI 53703 |
| Trade Name: | Propeller Sensor Model 2014-D |
| Common Name: | Nebulizer |
| ClassificationName: | NEBULIZER (DIRECT PATIENT INTERFACE) |
| ClassificationRegulation: | 21 CFR §868.5630 |
| Product Code: | CAF |
| DeviceDescription: | Dry Powder Inhaler (DPI) / Diskus Accessory that monitorsinhaler usage. The portable sensor mounts on top of a DPI andrecords inhalation events when the medication is used. |
| SubstantiallyEquivalentDevices: | K143671, Propeller Sensor Model 2014-D (Rx)K142516, Propeller Sensor Model 2 (OTC) |
| Intended Use: | The Propeller System includes the Propeller Sensor Model2014-D. The sensor is an accessory device intended for single-patient use to assist physicians and patients in recording |
{5}------------------------------------------------
and monitoring the actuations of prescribed DPI usage for the Diskus devices.
The Propeller Mobile Application records, stores, and transmits usage events from Propeller Sensors, or via manual user entry, to a remote storage system. With the Propeller Mobile Application the user can review information collected from the DPI sensor, and report and review symptoms and other information about their disease management and its impact. The user may also share their information with their caregivers, physician, and healthcare providers.
The Propeller Web Application is software that, like the Propeller Mobile Application, is intended to allow users to review the collected information and characteristics of their DPI and its use, to capture other patient-reported information and outcomes, and to allow that information to be shared with their caregivers, physicians, and health care providers.
When used with a prescribed DPI, the system can report on information captured during the normal course of use, such as the time between actuations that can be helpful in assessing DPI technique.
The Propeller System is intended to be used in populations from Child (>2 years) to Adult.
The Propeller System can be used both indoors and outdoors; home, work, and clinical settings, as well as on aircraft.
The Propeller System may also be used in clinical trials where researchers need to know information about the use of DPI medication(s) by a participant.
The output of the Propeller System is not intended to diagnose or replace a diagnosis provided by a licensed physician. The Propeller System is not intended for use as an DPI dose counter, nor is it intended to indicate the quantity of medication remaining in an DPI.
{6}------------------------------------------------
Technology Comparison and Device Description:
The Propeller Sensor Model 2014-D keeps track of medication use, with a record when the dry-powder inhaler is used. The sensor is a small device that attaches to the existing inhaler. Both the subject device and the predicate devices use technology that includes bluetooth wireless connectivity which connects to the previously cleared Propeller Health software system together with a mobile phone or wireless gateway .
The Propeller Sensor Model 2014-D is identical to the predicate (K143671), and therefore there are no technology differences.
| TechnologyComparison | Predicate Device:Propeller System,Propeller SensorModel 2014-D510k Number:K143671 | Candidate Device:Propeller System,Propeller SensorModel 2014-D (OTC) |
|---|---|---|
| Design - Attachmentto MedicationDispenser | Physically attaches toDPI without inhibitingpatient use | Same |
| Principle ofOperation | The Propeller HealthSensor attaches tothe medicationcanister and performswireless uploading ofusage history of theDPI. | Same |
| Output port andComputer Interface | Wireless uploading todatabase; viewed byPC or otherInternet-capabledevice. | Same |
| Data CollectionTechnology | Records date andtime of DPI usage bymonitoring actuationof the DPI via sensors | Same |
| Mobile Platforms | · iOS versions 7 orhigher• Androidoperating system | Same |
| Required Off theShelf Hardware | • Applesmartphones ordevices withBluetooth, iOS 7or higher• Androidsmartphones ordevices withBluetooth andoperating systemversion of 4.3and up for app• Internet capabledevice; noprocessor ormemoryrequirements(see RequiredBrowser) | Same |
| Required Browser | Firefox, Chrome,Safari , InternetExplorer | Same |
| Mobile Application | The Propeller HealthMobile Applicationrecords, stores, andtransmits usageevents from thePropeller HealthSensor via a featureor smart phone. Inaddition, the mobileapplication can beused to review theinformation capturedwhen using a smartphone. | Same |
| Software | The Propeller HealthWeb Application issoftware intended toallow users to reviewthe collectedinformation andcharacteristics of DPIuse, to add detailassociated with arecorded usageevent, and to sharethat information withtheir physician inorder to provideadditionalinformationassociated with thecondition for whichtheir DPImedication(s) areprescribed. | Same |
| Dose Counter | No | Same |
| Records Usage | Yes | Same |
| Records Location ofUsage (GPSCoordinates) | Geographiccoordinates can becaptured by thewireless device ifpaired with a sensor. | Same |
| Keyboard/InputInterface | Single buttoninterface | Same |
| Digital Display | No | Same |
| Power Source | 2 internal 3V DCLi-ion Batteries | Same |
| Battery Life | 1 Years | Same |
| Low battery indicator | Yes, lightcombination;software display ofbattery life. | Same |
| Patient Reminder | Yes | Same |
| Support | Yes | Same |
| Patient Data Storagewith Software | Yes | Same |
| Patient Data ReportGeneration withSoftware | Yes | Same |
| Patient Data GraphsGeneration | Yes | Same |
| Data Retrieval fromDevice w/ Software | Yes | Same |
| Case Material -Patient Contact byIntact Skin (hands) | Lexan Polycarbonate | Same |
| Case Material -Patient Contact bybreached skin (lips) | Lexan Polycarbonate | Same |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
The Propeller Sensor Model 2014-D and the predicate (K142516) have the same technological characteristics, both devices are bluetooth sensors fixed to inhalers designed to detect medication use. The changes below only reflect a change from MDI to DPI.
| TechnologyComparison | Predicate Device:Propeller System,Propeller SensorModel 2 OTC510k Number:K142516 | Candidate Device:Propeller System,Propeller SensorModel 2014-D (OTC) |
|---|---|---|
| Design - Attachmentto MedicationDispenser | Physically attachesto MDI withoutinhibiting patientuse | Physically attachesto DPI withoutinhibiting patientuse |
| Principle ofOperation | The Propeller HealthSensor attaches tothe medicationcanister and | The Propeller HealthSensor attaches tothe medicationcanister and |
| performs wirelessuploading of usagehistory of the MDI. | performs wirelessuploading of usagehistory of the DPI. | |
| Output port andComputer Interface | Wireless uploadingto database; viewedby PC or otherInternet-capabledevice. | Same |
| Data CollectionTechnology | Records date andtime of MDI usage bymonitoring actuationof the MDI viasensors | Records date andtime of DPI usage bymonitoring actuationof the DPI viasensors |
| Mobile Platforms | • iOS versions 7 orhigher• Androidoperatingsystem | Same |
| Required Off theShelf Hardware | • Applesmartphones ordevices withBluetooth, iOS 7 | Same |
| Required Browser | Firefox, Chrome,Safari , InternetExplorer | Same |
| Mobile Application | The Propeller HealthMobile Applicationrecords, stores, andtransmits usageevents from thePropeller HealthSensor via a featureor smart phone. Inaddition, the mobileapplication can beused to review theinformationcaptured when usinga smart phone. | Same |
| Software | The Propeller HealthWeb Application issoftware intended toallow users toreview the collectedinformation andcharacteristics ofMDI use, to adddetail associatedwith a recordedusage event, and toshare thatinformation withtheir physician inorder to provideadditionalinformationassociated with thecondition for whichtheir MDImedication(s) areprescribed. | The Propeller HealthWeb Application issoftware intended toallow users toreview the collectedinformation andcharacteristics ofDPI use, to adddetail associatedwith a recordedusage event, and toshare thatinformation withtheir physician inorder to provideadditionalinformationassociated with thecondition for whichtheir DPImedication(s) areprescribed. |
| Dose Counter | No | Same |
| Records Usage | Yes | Same |
| Records Location ofUsage (GPSCoordinates) | Geographiccoordinates can becaptured by thewireless device ifpaired with asensor. | Same |
| Keyboard/InputInterface | Single buttoninterface | Same |
| Digital Display | No | Same |
| Power Source | Internal 3V DC Li-ionBattery | 2 internal 3V DCLi-ion Batteries |
| Battery Life | 1.5 Years | 1 Years |
| Low batteryindicator | Yes, lightcombination;software display ofbattery life. | Same |
| Patient Reminder | Yes | Same |
| Support | Yes | Same |
| Patient Data Storagewith Software | Yes | Same |
| Patient Data ReportGeneration withSoftware | Yes | Same |
| Patient Data GraphsGeneration | Yes | Same |
| Data Retrieval fromDevice w/ Software | Yes | Same |
| Case Material -Patient Contact byIntact Skin (hands) | Lexan Polycarbonate | Same |
| Case Material -Patient Contact bybreached skin (lips) | N/A | Lexan Polycarbonate |
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
Validation testing including EMC, electrical, safety, Validation Testing mechanical durability, and functional testing has been for OTC completed and confirms that the device continues to meet the specified requirements for the change from "Prescription Use" to "OTC" status. Hazard Analysis for
Hazard Analysis for OTC was identical to the predicate ОТС (K143671). It included a review of existing hazards as well as how the patient obtains and learns about the system, registers for the system, installs the sensor, uses the Propeller System to track Diskus medication use, shares data with their physician/care team and obtains help & support with Diskus
{13}------------------------------------------------
| labeling. No new concerns of safety with the proposed OTCindication were found. | |
|---|---|
| Test Summary: | Test results indicate that the Propeller Sensor Model 2014-Dand its predicate Propeller System Model 2 complies withpredetermined specifications. Software verification andvalidation testing confirms this result. |
| Bench testing performed was identical to the predicate device.No new testing was performed for the OTC indication. | |
| Clinical Testing | No clinical testing was required |
| Conclusion: | The technology is identical to the predicate device. The overalltesting confirms that the Propeller Sensor Model 2014-D is assafe and as effective as the predicate device. |
§ 868.5630 Nebulizer.
(a)
Identification. A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.(b)
Classification. Class II (performance standards).