K Number
K152869
Manufacturer
Date Cleared
2016-01-14

(106 days)

Product Code
Regulation Number
886.1120
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LipiView® II Ocular Surface Interferometer is an ophthalmic imaging device intended for use by a physician in adult patients to capture, archive, manipulate and store digital images of:
• Specular (interferometric) observations of the tear film. Using these images, LipiView® II measures the absolute thickness of the tear film lipid layer.
• Meibomian glands under near-infrared (NIR) illumination
• The ocular surface and eyelids under white illumination
All of these image types can be photographically documented and visually monitored.

Device Description

The LipiView® II Ocular Surface Interferometer is a bench-top device used as an ophthalmic camera for imaging the lipid layer of the tear film, meibomian glands, ocular surface and eyelids. There are three imaging modes: Lipid, Gland and Ocular. LipiView II contains a computer system and electronics, two position chin and forehead rest, camera and lens, motion stage, illuminator and touch screen display. LipiView II includes pushbutton controls and handheld near-infrared lid everter for gland imaging.

AI/ML Overview

The provided document is a 510(k) summary for the LipiView II Ocular Surface Interferometer. It focuses on demonstrating substantial equivalence to predicate devices and does not contain detailed acceptance criteria and performance data from a specific clinical study for the device's diagnostic or predictive capabilities.

The document states that "Performance testing demonstrates that the design changes in LipiView II do not raise new questions of safety and effectiveness, and LipiView II is substantially equivalent to the predicate devices." It also mentions usability testing and software validation. However, it does not provide quantitative acceptance criteria or detailed results from a study that "proves the device meets the acceptance criteria" in terms of clinical accuracy (e.g., sensitivity, specificity, agreement with a gold standard).

Therefore, I cannot fulfill your request for:

  1. A table of acceptance criteria and the reported device performance in a clinical context.
  2. Sample size used for the test set and the data provenance, number of experts for ground truth, adjudication method, MRMC study details, standalone performance, type of ground truth, training set sample size, and ground truth establishment for the training set.

The document describes the device's technological characteristics, its intended use (imaging the tear film, meibomian glands, ocular surface, and eyelids), and how it compares to predicate devices. It emphasizes safety and effectiveness through compliance with standards (e.g., ISO, ANSI/IEC for light hazard, electrical safety, EMC, material flammability) and through the absence of new safety or effectiveness questions based on design changes.

Summary of what is present in the document regarding "performance testing":

  • Usability testing and software validation: These were done to demonstrate that LipiView II is as safe and effective as predicate devices for its intended use.
  • Compliance with standards: The device meets standards for light hazard protection (ISO15004-2), electrical safety (ANSI/IEC 60601), electromagnetic compatibility (EMC), and material flammability (UL 94V-1). The use of NIR illumination "does not raise new safety or effectiveness questions."
  • Hardware features: New hardware features (two-position chin rest, high-definition camera, NIR illumination, handheld NIR lid everter) "meet all design requirements for safety and performance, and therefore, do not adversely affect safety and effectiveness."

In conclusion, this document primarily asserts that the device's performance, particularly regarding safety and technological characteristics, is substantially equivalent to its predicate devices, rather than providing a detailed report of clinical study acceptance criteria and results for diagnostic performance.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles forming the staff and a wing-like shape above them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 14, 2016

Tearscience, Inc. % Dr. Christy Coleman Vice President, Clinical, Quality and Regulatory Affairs 5151 McCrimmon Pkwy., Suite 250 Morrisville, North Carolina 27650

Re: K152869

Trade/Device Name: Lipiview II Ocular Surface Interferometer Regulation Number: 21 CFR 886.1120 Regulation Name: Opthalmic Camera Regulatory Class: Class II Product Code: HKI, HJO Dated: December 8, 2015 Received: December 9, 2015

Dear Dr. Coleman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kesia Y. Alexander -A

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K152869

Device Name

LipiView® II Ocular Surface Interferometer

Indications for Use (Describe)

The LipiView II Ocular Surface Interferometer is an ophthalmic imaging device intended for use by a physician in adult patients to capture, archive, manipulate and store digital images of:

· Specular (interferometric) observations of the tear film. Using these images, LipiView® II measures the absolute thickness of the tear film lipid layer.

• Meibomian glands under near-infrared (NIR) illumination

· The ocular surface and eyelids under white illumination

All of these image types can be photographically documented and visually monitored.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for TearScience. The logo features a stylized blue eye on the left, followed by the word "TearScience" in a blue, slightly italicized font. A registered trademark symbol is present to the right of the word.

510(k) SUMMARY

PREPARATION DATE:September 29, 2015
APPLICANT:TearScience, Inc.5151 McCrimmon Parkway, Suite 250Morrisville, NC 27560
Tel: (919) 467-4007Fax: (919) 467-3300
CONTACT PERSON:Christy Coleman, OD, MPHVice President, Clinical, Quality and Regulatory Affairs
DEVICE TRADE NAME:LipiView® II Ocular Surface Interferometer
COMMON NAME:Ophthalmic Imaging Device
CLASSIFICATION NAME:Ophthalmic Camera
DEVICE CLASSIFICATION:Class II; 21 CFR 886.1120 and 21 CFR 886.1850
PRODUCT CODE:HKI, HJO
PREDICATE DEVICES:PRIMARY: LipiView® Ocular Surface InterferometerClass II under 21 CFR 886.1120 and 21 CFR 886.1850;Product Code HKI, HJO; Applicant: TearScience, Inc.;Originally cleared under K091935 on October 23, 2009;Expanded Indications for Use cleared under K122481 onDecember 31, 2012SECONDARY: Optomed Smartscope M5 Digital CameraClass II under 21 CFR 886.1120; Product Code HKI;Applicant: Optomed Oy.;Cleared under K132186 on December 18. 2013

DEVICE DESCRIPTION:

The LipiView® II Ocular Surface Interferometer is a bench-top device used as an ophthalmic camera for imaging the lipid layer of the tear film, meibomian glands, ocular surface and eyelids. There are three imaging modes: Lipid, Gland and Ocular. LipiView II contains a computer system and electronics, two position chin and forehead rest, camera and lens, motion stage, illuminator and touch screen display. LipiView II includes pushbutton controls and handheld near-infrared lid everter for gland imaging.

{4}------------------------------------------------

In the Lipid Imaging mode, LipiView II operates on the principle of white light interferometry and provides an interferometry color assessment of the tear film by specular reflection. The computer system captures a video image file that is recorded over time since the interference pattern changes as the tear film is distributed across the cornea during blinking. The video image of the ocular surface may be viewed on the computer screen display, in a printed report, or captured on video and exported to USB-attached storage or a mapped network drive.

In the Gland Imaging Mode, Lipi View II uses near infrared (NIR) illumination and an NIR-sensitive camera mounted on a motion control system to image the meibomian glands. The tissue between the meibomian glands and the surface of the eyelid are transparent to NIR light. The glands reflect NIR wavelengths, allowing them to be imaged.

In the Ocular Imaging mode, LipiView II captures high-resolution still images or video to record relevant findings on the ocular surface and/or eyelids. The images can be taken using room light only or the device's built-in white light.

The LipiView II Interferometer does not provide a diagnosis. The device has no treatment or no life-sustaining functions, and presents no potential for serious risk to the health, welfare or safety of patients.

INTENDED USE:

The LipiView® II Ocular Surface Interferometer is intended to image the eye and surrounding area. The LipiView II Interferometer is a prescription device for use by a physician during an in-office exam.

Indications for Use: The LipiView II Ocular Surface Interferometer is an ophthalmic imaging device intended for use by a physician in adult patients to capture, archive, manipulate and store digital images of:

  • . Specular (interferometric) observations of the tear film. Using these images, Lipi View II measures the absolute thickness of the tear film lipid layer.
  • Meibomian glands under near-infrared (NIR) illumination
  • . The ocular surface and eyelids under white illumination

All of these image types can be photographically documented and visually monitored.

As an ophthalmic camera, LipiView II has the same intended use to image the eye and surrounding area as the predicate devices, LipiView Interferometer (K091935, K122481) and Optomed Smartscope M5 Digital Camera (K132186). The Indications for Use for LipiView II are similar to the predicate LipiView device, except that the parts of the eye and surrounding area being imaged have been expanded to include the meibomian glands, ocular surface and eyelids. The expanded Indications for Use do not alter the intended use of the LipiView II Interferometer. LipiView II does not provide a diagnosis, and imaging other parts of the eye does not affect the safety and effectiveness of the device.

{5}------------------------------------------------

TECHNOLOGICAL CHARACTERISTICS:

The LipiView II Interferometer has the same fundamental scientific technology as the predicate devices. As summarized below, most of the technological characteristics of the LipiView II Interferometer remain unchanged from the predicate LipiView device. In addition, LipiView II and Optomed Smartscope M5 with optics module ES2 use similar visible and infrared LED illumination to image the eye. A comparison of the technology between LipiView II and each of the predicate devices is summarized below.

Comparison to LipiView Interferometer: The predicate LipiView device and LipiView II have the same principle of operation as an ophthalmic camera to image the eye and for specular observations of the tear film using white light interferometry. Both bench-top devices have similar overall dimensions and components, including computer system and electronics, chin and forehead rest, camera and lens, motion stage, illuminator and touch screen display. Both devices may be used with off-the-shelf USB accessories and operate wirelessly with networks.

Similar design features include: adjustable forehead and chin rest with the same patient contact material and disinfection method; digital video camera adjusted using a motion stage; white light Class I LED illuminator; and a visible LED fixation light. Both devices comply with ISO15004-2 standard for light hazard protection. Both devices have an AC power source in compliance with ANSI/IEC 60601 standards for electrical safety and electromagnetic compatibility (EMC). Both devices have flame retardant materials per UL 94V-1 standard. Both devices have ambient temperature of parts of device held by operator or accessible to patient in compliance with ISO 15004-1 standard.

The software on both devices runs on a Microsoft Windows-based operating system with a touchscreen display graphical user interface. Similar software design features include: password-protected user login: patient entry into a database: real-time video display to acquire and view images; touchscreen user controls for camera and video playback; image acquisition process with storage of lossless AVI format video images; printed or saved reports; and export of images to USB external media or electronic medical record system via wireless network. In addition, both devices use the same tear film lipid layer image processing for analysis and measurement of the lipid layer thickness.

Different design features include new hardware components in LipiView II: pushbutton controls, a two-position adjustable chin rest, a high definition camera, surface NIR illumination, and a handheld NIR lid everter. These design changes are to support imaging other parts of the eye and surrounding area, such as the meibomian glands. The primary software changes in LipiView II are an update to the Windows-based operating system, new graphical user interface appearance; and new meibomian gland imaging and ocular imaging functionality. Other minor hardware and software changes relative to the predicate LipiView device were made for user convenience or product enhancement.

Comparison to Optomed M5 Digital Camera: The Optomed M5 Digital Camera with ES2 optics module and LipiView II have the same principle of operation as an ophthalmic camera for digital imaging of the eye. Both devices have a camera, lens, visible and infrared LED illumination and visible LED fixation light. Both devices

{6}------------------------------------------------

comply with ISO15004-2:2007 standard for light hazard protection. Both devices have a power source in compliance with ANSI/IEC 60601-1-1 and 60601-1-2 standards for electrical safety and EMC. Also, both devices have display systems to view images and both provide image storage capability on external media.

LipiView II is a bench-top device with a computer system, software and a touchscreen display graphical user interface. Conversely, Optomed Smartscope M5 with optic module ES2 consists of a camera handset, attachable ophthalmic lens, cradle for charging and image transfer to a PC by USB connection. LipiView II has an integrated chin rest to support the patient's head, whereas. Optomed Smartscope utilizes an adapter to a slit lamp biomicroscope, which has support for the patient's head.

PERFORMANCE TESTING:

Performance testing demonstrates that the design changes in LipiView II do not raise new questions of safety and effectiveness, and LipiView II is substantially equivalent to the predicate devices.

  • Usability testing and software validation demonstrate that LipiView II is as safe O and as effective as the predicate devices in the intended use to image the eye and surrounding area, including meibomian glands, ocular surface and eyelids.
  • As for the predicate devices, LipiView II meets the standards for light hazard o protection, electrical safety and electromagnetic compatibility and material flammability. Therefore, the use of near infrared illumination does not raise new safety or effectiveness questions.
  • The new hardware features in LipiView II, including two-position chin rest, high o definition camera, NIR illumination and handheld NIR lid everter, meet all design requirements for safety and performance, and therefore, do not adversely affect safety and effectiveness as compared to the predicate device.

CONCLUSIONS:

The LipiView II Ocular Surface Interferometer has the same intended use and the same fundamental scientific technology as the predicate devices. The expanded Indications for Use for LipiView II relative to the predicate devices do not alter the intended use and do not affect safety and effectiveness of the device. Performance testing demonstrates the LipiView II Interferometer is substantially equivalent in intended use and technological characteristics to the predicate devices. The technology differences do not raise new questions of safety and effectiveness, and do not adversely affect safety and effectiveness of the device. Based on overall design verification and validation and bench performance testing, LipiView II is at least as safe and effective as the predicate devices.

{7}------------------------------------------------

TABLE 5-1. PREDICATE DEVICE COMPARISON TABLE
Comparison FeatureLipiView II InterferometerPredicate LipiView Interferometer(K091935 and K122481)Optomed Smartscope M5Digital Camera (K132186)
Intended UseIntended to take photographs of the eye and thesurrounding areaIntended to take photographs of the eye and thesurrounding areaIntended to take photographs ofthe eye and the surrounding area
Indications for UseThe LipiView II Ocular Surface Interferometer isan ophthalmic imaging device intended for use bya physician in adult patients to capture, archive,manipulate and store digital images of:• Specular (interferometric) observations of the tearfilm. Using these images, LipiView II measuresthe absolute thickness of the tear film lipid layer.• Meibomian glands under near-infrared (NIR)illumination.• The ocular surface and eyelids under whiteillumination.All of these image types can be photographicallydocumented and visually monitored.The LipiView Ocular Surface Interferometer is anophthalmic imaging device that is intended for useby a physician in adult patients to capture, archive,manipulate, and store digital images of specular(interferometric) observations of the tear film,which can be visually monitored andphotographically documented. Using theseimages, the LipiView® Interferometer measuresthe absolute thickness of the tear film lipid layer.Optomed Smartscope M5 withoptics module ES2 is intended tocapture images and video of thesurface of the human eye andsurrounding areas.
Rx only/OTCPrescription DevicePrescription DevicePrescription Device
Method of OperationOphthalmic ImagingOphthalmic camera with digital imagingOphthalmic camera with digital imagingOphthalmic camera with digitalimaging
Method of OperationSpecular Observationof Tear FilmTear film images and lipid layer thicknessmeasurement based on specular reflection usingwhite light interferometryTear film images and lipid layer thicknessmeasurement based on specular reflection usingwhite light interferometryNone
Illumination Source• Visible White Class I LED illumination• Visible LED fixation light• Near-Infrared LED illumination andtransillumination of the eyelid• Visible White Class I LED illumination• Visible LED fixation light• Visible and Infrared LEDillumination• Red LED Fixation Light• White and Cobalt Blue Light
Illumination ExposureComplies with ISO 15004-2:2007 Group 2Complies with ISO 15004-2:2007 Group 1Complies with ISO 15004-2:2007Group 2 instrument for safety
Parametersinstrument for safetyinstrument for safety
Brightness ControlNo brightness control adjustment by userNo brightness control adjustment by userUnknown
Material FlammabilityMaterials near light source comply with UL 94V-1Materials near light source comply with UL 94V-1Unknown
Maximum Temperature ofHeld or Accessible PartsAmbient temperature of parts of device held byoperator or accessible to patientAmbient temperature of parts of device held byoperator or accessible to patientUnknown
TABLE 5-1. PREDICATE DEVICE COMPARISON TABLE
Comparison FeatureLipiView II InterferometerPredicate LipiView Interferometer(K091935 and K122481)Optomed Smartscope M5Digital Camera (K132186)
Data Collection and/orDisplay Systems• Touchscreen display to view images• Stores images on computer and external media• Support for off-the-shelf USB printer and mediathat are compatible with Windows 7 and USB2.0• Touchscreen display to view images• Stores images on computer and external media• Support for off-the-shelf USB printer and mediacompatible with Windows XP and USB1.0• Active matrix color TFT LCD• Stores images on externalmedia (flash memory card)• USB1.1 terminal compatiblewith Windows XP/Vista/7
Electrical Safety andElectromagneticCompatibility• AC-powered• Complies with ANSI 60601-1:2005 (3rd edition)and IEC 60601-1-2: 2007 (3rd edition)• AC-powered• Complies with IEC 60601-1:1995 (2nd edition)and IEC 60601-1-2: 2001 (2nd edition)• Rechargeable NiMH Batterywith charging unit• Complies with IEC 60601-1:2005 (3rd edition) and IEC60601-1-2: 2007 (3rd edition)
Chin/Forehead RestSupportIntegrated chin rest and forehead support isdisinfected with alcohol prior to use and storageIntegrated chin rest and forehead support isdisinfected with alcohol prior to use and storageMay be used with slit lampadapter for chin rest support

{8}------------------------------------------------

§ 886.1120 Ophthalmic camera.

(a)
Identification. An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.(b)
Classification. Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.