K Number
K152859
Date Cleared
2015-12-03

(64 days)

Product Code
Regulation Number
872.1800
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is a diagnostic X-ray system, which is intended to be used by trained dentists and dental technicians as an extra-oral X-ray source for producing diagnostic X-ray images using intra-oral image receptors. Its use is intended for both adults and pediatric subjects.

Device Description

The portable x-ray system (Models: MINIX-V, MINIX-S) is an x-ray generating device which is mainly designed for dental examination (teeth and jaw). Operating principle is that the high voltage electricity is getting into x-ray tube and generates x-ray source. This x-ray source comes out from the emitting cone and penetrates teeth and jaw, and makes x-ray images through image receptors (chemical film or digital sensor). Once the power is supplied from the rechargeable lithium-ion polymer battery pack, the inverter generates high frequency and high voltage. Boosted and rectified DC high voltage (MINIX-V: 70kV, MINIX-S: 60kV) is loaded onto the x-ray tube, and the tube generates x-ray. Micro control unit (MCU) and other control parts function to display data on LCD and control the system by the input information from user interface. When operator controlled (fixed amount of) x-ray is exposed to patient's teeth and jaw, clinician can examine the parts with processed images from intra-oral image receptors (analog film, digital phosphor plate or digital sensor). The Portable X-ray System (Models: MINIX-V, MINIX-S) is composed of an x-ray generating part with an x-ray tube, a device controller (Control PCB), a power controller (Power PCB), an user interface (LCD window), a beam limiting part (x-ray emitting cone), a backscatter shield, and a remote control switch (hand-switch). The Portable X-ray System (Models: MINIX-V, MINIX-S) is a diagnostic x-ray system, which is intended to be used by trained dentists or dental technicians. Its use is intended for both adult and pediatric subjects. This device includes a high frequency inverter that changes direct current to alternating current, x-ray tube, electrical protective devices, and other elements. The system with high frequency x-ray provides sharp and clear images, and the total lead protection seal protects patients and dentists from leakage or scattered radiation from the device.

AI/ML Overview

The provided documentation does not contain information about acceptance criteria or a study proving the device meets them in the context of an AI/ML system. The device described, "Portable X-ray System (Models: MINIX-V, MINIX-S)," is an X-ray generating device for dental examinations. The submission focuses on demonstrating substantial equivalence to predicate devices based on safety, electrical, mechanical, and X-ray performance standards, rather than evaluating the performance of an AI/ML algorithm.

Therefore, I cannot provide the requested information regarding acceptance criteria, device performance, sample sizes, data provenance, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types, or training set details, as these are relevant to AI/ML device evaluations which are not described in this document.

The document states: "Clinical images were provided however they were not necessary in order to establish substantial equivalence with the predicate devices," further indicating that image analysis performance (which would typically involve AI/ML) was not the focus of this submission for substantial equivalence.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

December 3, 2015

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Digimed Co., Ltd. (Digimed Corporation until Dec. 31, 2012) % Mr. Yong Park President Innoden LLC. 212 Wells Avenue, S# 102 RENTON WA 98057

Re: K152859

Trade/Device Name: Portable X-ray System (Model: MINIX-V / MINIX-S ) Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: EHD Dated: September 30, 2015 Received: September 30, 2015

Dear Mr. Park:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert Oels

Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K152859

Device Name Portable X-ray System (Models: MINIX-V, MINIX-S)

Indications for Use (Describe) The device is a diagnostic X-ray system, which is intended to be used by trained dentists and dental technicians as an extra-oral X-ray source for producing diagnostic X-ray images using intra-oral image receptors. Its use is intended for both adults and pediatric subjects.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

See PRA Statement below.

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary of Portable X-ray System (MINIX-V&MINIX-S)

Date Prepared:Sep. 16, 2015
Company Name:DIGIMED CO., LTD.(DIGIMED CORPORATION. until Dec. 31, 2012)
Address:309~311, 318-ho,145, Gasan digital 1-ro, Geumcheon-gu,Seoul, 153-787 Korea
Telephone number:+82-2-2624-1551
Fax:+82-2-2624-1553
E-mail:kwon@digimed.co.kr
Contact:Youngbae Kwon, Managing Director

1. Company and Correspondent Making the Submission:

2. US Agent for FDA Contact:

NameMr. Yong Park
Company Name:INNODEN LLC.
Address:212 Wells Ave. S #102,Renton, WA 98057
Telephone number:425-572-0283
Fax:425-988-1256
E-mail:drpark@innoden.com

3. Device Information:

Proprietary-Trade Name:Portable X-Ray System (Models: MINIX-V, MINIX-S)
Classification Name:Extra-oral Source X-Ray System: EHD, Class II perregulation 21CFR 872.1800
Common/Usual Name:Portable X-Ray System

4. Predicate Devices:

Manufacturer:GENORAYDIGIMED
Device:Portable X-Ray System(Model: PORT-X II)Portable X-Ray System(Model: BIOX,PROX/DIOX-602)
Classification:Extra-oral Source X-Ray System: EHD, Class II per regulation 21CFR 872.1800Extra-oral Source X-Ray System: EHD, Class II per regulation 21CFR 872.1800
510(k) Number:K063121(Decision Date: Jan. 11, 2007)K103600 - BIOX(Decision Date: Mar. 10, 2011)K082167 - PROX / DIOX-602(Decision Date: July 29, 2008)

{4}------------------------------------------------

5. Indications for Use (Intended Use):

The Portable X-ray System (Models: MINIX-V, MINIX-S) is intended to be used by trained dentists and dental technicians as extra-oral x-ray source for producing diagnostic x-ray images using intra-oral image receptors. Its use is intended for both adult and pediatric subjects.

Compatible intra-oral image receptors are described as below.

    1. Analog dental x-ray films
    1. CCD and CMOS digital sensors
    1. Digital phosphor plates
    1. Most of the intra-oral image receptors which intended for dental x-ray examination

6. Description of Device

The portable x-ray system (Models: MINIX-V, MINIX-S) is an x-ray generating device which is mainly designed for dental examination (teeth and jaw).

Operating principle is that the high voltage electricity is getting into x-ray tube and generates x-ray source. This x-ray source comes out from the emitting cone and penetrates teeth and jaw, and makes x-ray images through image receptors (chemical film or digital sensor). Once the power is supplied from the rechargeable lithium-ion polymer battery pack, the inverter generates high frequency and high voltage. Boosted and rectified DC high voltage (MINIX-V: 70kV, MINIX-S: 60kV) is loaded onto the x-ray tube, and the tube generates x-ray.

Micro control unit (MCU) and other control parts function to display data on LCD and control the system by the input information from user interface. When operator controlled (fixed amount of) x-ray is exposed to patient's teeth and jaw, clinician can examine the parts with processed images from intra-oral image receptors (analog film, digital phosphor plate or digital sensor).

The Portable X-ray System (Models: MINIX-V, MINIX-S) is composed of an x-ray generating part with an x-ray tube, a device controller (Control PCB), a power controller (Power PCB), an user interface (LCD window), a beam limiting part (x-ray emitting cone), a backscatter shield, and a remote control switch (hand-switch).

The Portable X-ray System (Models: MINIX-V, MINIX-S) is a diagnostic x-ray system, which is intended to be used by trained dentists or dental technicians. Its use is intended for both adult and pediatric subjects. This device includes a high frequency inverter that changes direct current to alternating current, x-ray tube, electrical protective devices, and other elements. The system with high frequency x-ray provides sharp and clear images, and the total lead protection seal protects patients and dentists from leakage or scattered radiation from the device.

{5}------------------------------------------------

7. Safety and Effectiveness. Comparison to Predicate Device:

Safety and effectiveness of the subject device is considered with the latest version of test regulations. The subject device was shown to provide an equivalent level of safety and performance as compared to the predicate devices.

"Clinical images were provided however they were not necessary in order to establish substantial equivalence with the predicate devices"

8. Safety, EMC and Performance Data:

The subject device complies with the safety and performance standards listed in the chart below. Test reports were provided to demonstrate conformance. All required documents and reports are submitted to the appropriate oversight agency to establish compliance with the applicable requirements.

9. The differences between the subject device and the predicate devices

The subject device has little differences with its design, size, and user interface compare to the predicate devices. Detailed differences can be identified from "Substantial Equivalence Chart".

{6}------------------------------------------------

10. Substantial Equivalence Chart

GENORAYDIGIMEDDIGIMED
FEATUREPORT-X IIMINIX-VMINIX-S
(K063121)(NEW)(NEW)
INTENDED USEAll three systems are intended as extraoral x-ray sources to be used with intraoral image receptorsfor diagnostic imaging by dentists or dental technicians.
MECHANICAL
Size7.8" x 5.8" x 3.7"4.8" x 8.7" x 5.3"4.8" x 8.7" x 5.3"
Source toskin distance10 cm20 cm20 cm
Cone diameter7 cm5.5 cm5.5 cm
User interfaceUp-down buttons for exposuretime selection with timerdisplay. Additionally, severaluser-selectable preset timeswith patient size and toothselection icons on an LCDdisplayUp-down buttons for exposuretime selection with timerdisplay. Additionally, severaluser-selectable preset timeswith patient size, image-receptor type, and toothselection icons on an LCDdisplayUp-down buttons for exposuretime selection with timerdisplay. Additionally, severaluser-selectable preset timeswith patient size, image-receptor type, and toothselection icons on an LCDdisplay
Backscatterradiation protectionNone6.3" dia. Pb-filled acrylicplastic scatter shield6.3" dia. Pb-filled acrylicplastic scatter shield
Exposure switchExposure button at x-raycontrol panelExposure button at front coveron right hand side,or a remote control switchExposure button at front coveron right hand side,or a remote control switch
Tubehead mountingHandheld, or on a tripodHandheld, on optional arms oron a tripodHandheld, on optional arms oron a tripod
Half-value layer(not available)2.3 mm Al2.0 mm Al
ELECTRICAL
Energy sourceRechargeable 22.2 V DCLithium Polymer battery packRechargeable 22.2 V DCLithium Polymer battery packRechargeable 22.2 V DCLithium Polymer battery pack
Exposure time0.01~2.0 seconds in 46 steps0.01~2.0 seconds in 0.01increments0.01~2.0 seconds in 0.01increments
Time Accuracy± (10% +1 ms)± (10% +1 ms)± (10% +1 ms)
mA2 mA fixed2 mA fixed2 mA fixed
kVp60 kV fixed70 kV fixed60 kV fixed
WaveformConstant Potential (DC)Constant Potential (DC)Constant Potential (DC)
Duty Cycle(not available)1:601:60
EMI standardsEN60601-1-2EN60601-1-2, IEC60601-1-2CISPR 11, IEC61000-3-2IEC61000-3-3EN60601-1-2, IEC60601-1-2CISPR 11, IEC61000-3-2IEC61000-3-3
X-RAY PERFORMANCE
PerformancestandardsEN 60601-1EN 60601-1-3EN 60601-2-7EN 60601-2-28EN 60601-2-32IEC 60601-1IEC 60601-1-3IEC 60601-2-65IEC 60601-1IEC 60601-1-3IEC 60601-2-65
DifferenceSSD, Cone diameter, Backscatter radiation protection, Exposure switch, kV (MiniX-V only)
  1. Predicate device (Model: PORT-X II) from GENORAY

{7}------------------------------------------------

2) Predicate device (Model: PROX/DIOX-602) from DIGIMED
FEATUREDIGIMEDPROX/DIOX-602(K082167)DIGIMEDMINIX-V(NEW)DIGIMEDMINIX-S(NEW)
INTENDED USEAll three systems are intended as extraoral x-ray sources to be used with intraoral image receptorsfor diagnostic imaging by dentists or dental technicians.
MECHANICAL
SizePROX: 7" x 9.3" x 5.4"DIOX-602: 5.3" x 9.4" x 6.9"4.8" x 8.7" x 5.3"4.8" x 8.7" x 5.3"
Source toskin distance20 cm20 cm20 cm
Cone diameter6 cm5.5 cm5.5 cm
User interfaceUp-down buttons for exposuretime selection with timerdisplay. Anatomical toothbuttons, patient size changebutton, and image receptortype change button. Eachbutton shows icons on an LCDdisplay. Several user-selectablepreset times.Up-down buttons for exposuretime selection with timerdisplay. Additionally, severaluser-selectable preset timeswith patient size, image-receptor type, and toothselection icons on an LCDdisplayUp-down buttons for exposuretime selection with timerdisplay. Additionally, severaluser-selectable preset timeswith patient size, image-receptor type, and toothselection icons on an LCDdisplay
Backscatterradiation protection6.3" dia. Pb-filled acrylicplastic scatter shield6.3" dia. Pb-filled acrylicplastic scatter shield6.3" dia. Pb-filled acrylicplastic scatter shield
Exposure switchExposure button at front coveron right hand side,or a remote control switchExposure button at front coveron right hand side,or a remote control switchExposure button at front coveron right hand side,or a remote control switch
Tubehead mountingHandheld, on optional arms oron a tripodHandheld, on optional arms oron a tripodHandheld, on optional arms oron a tripod
Half-value layer2.0 mm Al2.3 mm Al2.0 mm Al
ELECTRICAL
Energy sourceRechargeable 24 V DCLithium Polymer battery packRechargeable 22.2 V DCLithium Polymer battery packRechargeable 22.2 V DCLithium Polymer battery pack
Exposure time0.01~1.6 seconds in 0.01increments0.01~2.0 seconds in 0.01increments0.01~2.0 seconds in 0.01increments
Time Accuracy$\pm$ (10% +1 ms)$\pm$ (10% +1 ms)$\pm$ (10% +1 ms)
mA2 mA fixed2 mA fixed2 mA fixed
kVp60 kV fixed70 kV fixed60 kV fixed
WaveformConstant Potential (DC)Constant Potential (DC)Constant Potential (DC)
Duty Cycle1:601:601:60
EMI standardsEN60601-1-2, IEC60601-1-2CISPR 11, IEC61000-3-2IEC61000-3-3EN60601-1-2, IEC60601-1-2CISPR 11, IEC61000-3-2IEC61000-3-3EN60601-1-2, IEC60601-1-2CISPR 11, IEC61000-3-2IEC61000-3-3
X-RAY PERFORMANCE
PerformancestandardsEN 60601-1EN 60601-1-3EN 60601-2-7EN 60601-2-28EN 60601-2-32IEC 60601-1IEC 60601-1-3IEC 60601-2-65IEC 60601-1IEC 60601-1-3IEC 60601-2-65
DifferenceCone diameter, Energy source, Exposure time, kV (MiniX-V only)
3) Predicate device (Model: BIOX) from DIGIMED
FEATUREDIGIMEDBIOX(K103600)DIGIMEDMINIX-V(NEW)DIGIMEDMINIX-S(NEW)
INTENDED USEAll three systems are intended as extraoral x-ray sources to be used with intraoral image receptorsfor diagnostic imaging by dentists or dental technicians.
MECHANICAL
Size7.6" x 9.3" x 5.4"4.8" x 8.7" x 5.3"4.8" x 8.7" x 5.3"
Source toskin distance21 cm20 cm20 cm
Cone diameter6 cm5.5 cm5.5 cm
User interfaceUp-down buttons for exposuretime selection with timerdisplay. Anatomical toothbuttons, patient size changebutton, and image receptortype change button. Eachbutton shows icons on an LCDdisplay. Several user-selectablepreset times.Up-down buttons for exposuretime selection with timerdisplay. Additionally, severaluser-selectable preset timeswith patient size, image-receptor type, and toothselection icons on an LCDdisplayUp-down buttons for exposuretime selection with timerdisplay. Additionally, severaluser-selectable preset timeswith patient size, image-receptor type, and toothselection icons on an LCDdisplay
Backscatterradiation protection6.3" dia. Pb-filled acrylicplastic scatter shield6.3" dia. Pb-filled acrylicplastic scatter shield6.3" dia. Pb-filled acrylicplastic scatter shield
Exposure switchExposure button at front coveron right hand side,or a remote control switchExposure button at front coveron right hand side,or a remote control switchExposure button at front coveron right hand side,or a remote control switch
Tubehead mountingHandheld, on optional arms oron a tripodHandheld, on optional arms oron a tripodHandheld, on optional arms oron a tripod
Half-value layer2.0 mm Al2.3 mm Al2.0 mm Al
ELECTRICAL
Energy sourceRechargeable 24 V DCLithium Polymer battery packRechargeable 22.2 V DCLithium Polymer battery packRechargeable 22.2 V DCLithium Polymer battery pack
Exposure time0.01~1.6 seconds in 0.01increments0.01~2.0 seconds in 0.01increments0.01~2.0 seconds in 0.01increments
Time Accuracy$\pm$ (10% +1 ms)$\pm$ (10% +1 ms)$\pm$ (10% +1 ms)
mA3 mA fixed2 mA fixed2 mA fixed
kVp60 kV fixed70 kV fixed60 kV fixed
WaveformConstant Potential (DC)Constant Potential (DC)Constant Potential (DC)
Duty Cycle1:601:601:60
EMI standardsEN60601-1-2, IEC60601-1-2CISPR 11, EN61000-3-2,EN61000-3-3EN60601-1-2, IEC60601-1-2CISPR 11, IEC61000-3-2IEC61000-3-3EN60601-1-2, IEC60601-1-2CISPR 11, IEC61000-3-2IEC61000-3-3
X-RAY PERFORMANCE
PerformancestandardsEN 60601-1EN 60601-1-3EN 60601-2-7EN 60601-2-28EN 60601-2-32IEC 60601-1IEC 60601-1-3IEC 60601-2-65IEC 60601-1IEC 60601-1-3IEC 60601-2-65
DifferenceSSD, Cone diameter, Energy source, Exposure time, mA (BIOX), kV (MiniX-V only)

{8}------------------------------------------------

3) Predicate device (Model: BIOX) from DIGIMED

{9}------------------------------------------------

11. Conclusion

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided above comparison table, the Portable X-ray System (Models: MINIX-V, MINIX-S) has little difference with its size and user interface as the information in the table. But the system is substantially equivalent to the predicate devices with its design, mechanical and electrical performance as described.

Performance evaluation (test) reports and device inspection report confirmed that the Portable X-ray System (Model: MINIX-V, MINIX-S) is suitable for its intended use and user instruction of the device.

§ 872.1800 Extraoral source x-ray system.

(a)
Identification. An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.(b)
Classification. Class II.