K Number
K152720
Date Cleared
2016-03-10

(171 days)

Product Code
Regulation Number
880.6250
Panel
HO
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Nitrile Powder Free Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Nitrile films form a barrier to body fluids and bloodborne Pathogens. The nitrile rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure. Nitrile glove is known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D6319 and ASTM D5151 requirements.

AI/ML Overview

This document describes the acceptance criteria and performance of the "Xingda Nitrile Powder Free Patient Examination Gloves, Blue Color" (K152720) in comparison to a predicate device (K120970) to establish substantial equivalence for FDA 510(k) clearance.

1. Table of Acceptance Criteria and Reported Device Performance:

CharacteristicsStandard / Acceptance CriteriaReported Device PerformanceResult
DimensionASTM standard D 6319-10.MeetsMeets
Length≥ 230mm min (ASTM D6319-10)230mm min for all sizesSubstantially Equivalent
Width (Small)70-90 mm (ASTM D6319-10)84-85 mmSubstantially Equivalent
Width (Medium)85-105 mm (ASTM D6319-10)94-96 mmSubstantially Equivalent
Width (Large)100-120 mm (ASTM D6319-10)105-107 mmSubstantially Equivalent
Width (X-Large)110-130 mm (ASTM D6319-10)115-118 mmSubstantially Equivalent
Thickness (Finger)0.05mm min (ASTM D6319-10)0.10 mm min.Substantially Equivalent
Thickness (Palm)0.05mm min (ASTM D6319-10)0.06 mm min.Substantially Equivalent
Physical PropertiesASTM standard D 6319-10.MeetsMeets
Elongation (Before/After aging)≥ 500% (ASTM D6319-10)520-580%Substantially Equivalent
Tensile Strength (Before/After aging)≥ 14MPa (ASTM D6319-10)22-34 MPaSubstantially Equivalent
Freedom from Pinholes21 CFR 800.20, ASTM D6319-10, ASTM D5151-06 (Reapproved 2011)Meets (Holes at Inspection Level I AQL2.5)Meets / Substantially Equivalent
Powder ResidualASTM D6319-10 and D6124-06 (Reapproved 2011)Meets,

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.