K Number
K152635
Date Cleared
2016-06-01

(260 days)

Product Code
Regulation Number
866.5100
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

QUANTA Flash® Scl-70 is a chemiluminescent immunoassay for the semi-quantitative determination of IgG anti-Scl-70 autoantibodies in human serum. The presence of anti-Scl-70 autoantibodies, in conjunction with clinical findings and other laboratory tests, aids in the diagnosis of systemic sclerosis.

QUANTA Flash® Scl-70 Calibrators are intended for use with the QUANTA Flash® Scl-70 chemiluminescent immunoassay for the determination of IgG anti-Scl-70 autoantibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values.

QUANTA Flash® Scl-70 Controls are intended for use with the OUANTA Flash® Scl-70 chemiluminescent immunoassay for quality control in the determination of IgG anti-Scl-70 autoantibodies in human serum.

Device Description

The QUANTA Flash Scl-70 assay is designed to run on the BIO-FLASH® instrument. This platform is a fully automated closed system with continuous load and random access capabilities that automatically processes the samples, runs the assay and reports the results. It includes liquid handling hardware, luminometer and computer with software-user interface. The QUANTA Flash Scl-70 assay utilizes a reagent cartridge format, which is compatible with the BIO-FLASH instrument.

Recombinant Scl-70 is coated onto paramagnetic beads. The bead suspension is lyophilized and stored in the bead tube. Prior to use in the BIO-FLASH system, the sealed reagent tubes are pierced with the reagent cartridge lid and the beads are rehydrated and resuspended using resuspension buffer by pipetting up and down with a transfer pipette. The reagent cartridge is then loaded onto the BIO-FLASH instrument. Samples are also loaded onto the instrument in sample racks. A patient serum sample is prediluted 1:23.5 by the BIO-FLASH with system rinse in a disposable plastic cuvette. Small amounts of the diluted patient serum, the beads, and assay buffer are all combined into a second cuvette, and mixed. This cuvette is then incubated at 37°C. The beads are magnetized and washed several times. Isoluminol conjugated anti-human IgG antibodies are then added to the cuvette, and again incubated at 37°C. The beads are magnetized and washed repeatedly. The isoluminol conjugate is oxidized when Trigger 1 (Fe(III) coproporphyrin in sodium hydroxide solution) and Trigger 2 (urea-hydrogen peroxide in sodium chloride solution) are added to the cuvette, and the flash of light produced from this reaction is measured as Relative Light Units (RLU) by the BIO-FLASH optical system. The RLU are proportional to the amount of isoluminol conjugate that is bound to the human IgG, which is in turn proportional to the amount of anti-Scl-70 antibodies bound to the corresponding Scl-70 on the beads.

For quantitation, the QUANTA Flash Scl-70 assay utilizes a predefined lot specific Master Curve that is uploaded onto the instrument through the reagent cartridge barcode. Every new lot number of reagent cartridge must be calibrated before first use, with the QUANTA Flash Scl-70 Calibrators. Based on the results obtained with the two Calibrators included in the Calibrator Set (sold separately), an instrument specific Working Curve is created, which is used to calculate chemiluminescent units (CU)mL from the instrument signal (RLU) obtained for each sample.

AI/ML Overview

Here's a summary of the acceptance criteria and study findings based on the provided text, structured as requested:

1. Table of Acceptance Criteria and Reported Device Performance

Performance CharacteristicAcceptance CriteriaReported Device Performance
PrecisionTotal %CV values within 10%.All total %CV values for 13 samples across various concentrations were within 10%. (Range: 3.4% - 5.9%)
Reproducibility (Between sites)All %CV values within 15%.All %CV values for 8 samples across three sites were within 15%. (Range: 1.3% - 8.3%)
Reproducibility (Between lots)All %CV values within 10%.All %CV values for 8 samples across three different lots were within 10%. (Range: 1.5% - 9.6%)
Limit of Quantitation (LoQ)Total error (TE) < 25% (accuracy goal).LoQ is 1.2 CU, meeting the accuracy goal.
Limit of Detection (LoD)Proportions of false positives (alpha) < 5% and false negatives (beta) < 5%.LoD is 0.2 CU, determined with alpha and beta less than 5%.
Limit of Blank (LoB)Not explicitly stated as a separate criterion, but derived from method.LoB is 0.1 CU.
Analytical Measuring Range (AMR)Defined by LoQ and highest Master Curve Standard.1.2 CU - 786.3 CU.
Auto-rerun function% recovery values for auto-rerun results compared to manual dilution within ± 20%.% recovery values for auto-rerun were 104%, 101%, and 91% (average 102%), all within ± 20%.
Linearity80%-120% recovery, 0.9-1.1 slope, and ≥0.95 R² for linear regression analysis.Percent recovery for all data points ranged from 80.2% to 118.4%. Slopes for individual samples ranged from 0.96 to 1.03 (all within 0.9-1.1). R² values for all samples were 0.99 or 1.00 (all ≥0.95). Combined results: Slope 1.01, R² 0.99.
Interference85% - 115% recovery for unit values.No interference detected. Recoveries for Bilirubin (88-101%), Hemoglobin (93-105%), Triglycerides (89-97%), Cholesterol (93-106%), Human IgG (90-109% or < 4 CU), RF IgM (97-111%), Prednisone (100.0-109.7%), and Naproxen (97.7-109.7%) were all within the 85-115% range.
Sample Stability90-110% average recovery compared to control (Day Zero, 2-8°C).All samples fulfilled acceptance criteria at each time point (up to 21 days at 2-8°C, up to 48 hours at room temperature, up to 3 freeze/thaw cycles).
Reagent Stability (Accelerated)Microparticles: 95% CI of regression line between 85-115% at 2 weeks, no individual data point outside 75-125% recovery at 2 weeks. Controls & Calibrators: Lower 95% CI of regression line between 90-110% at 2 weeks, no individual data point outside 80-120% recovery at 2 weeks.All components tested fulfilled the acceptance criteria, leading to a one-year expiration dating assignment.
Reagent Stability (Real-time)Controls: Results within established acceptable ranges. Calibrators: % recovery of average of triplicates between 85-115%, %CV of triplicates < 10%. Reagent Cartridge: %CV of replicates < 15%, % recovery of each sample between 80-120%.All results were within the acceptance limits for controls, calibrators, and reagent cartridges up to 6 months.
In-use Stability (Calibrators)5 successful calibrations in 8.5 hours, Calibrator average RLU recovery 90-110% compared to first use.5 successful calibrations performed over 8.5 hours. Calibrator RLU values remained within 90-110% range. Patient samples ran within expected range. Supports claim of up to 4 calibrations over 8 hours.
In-use Stability (Controls)Replicates run within established range, linear regression line of %recovery vs. runs between 85-115% at run 15.Low and High Controls ran within their acceptable range for all runs. Linear regression line of %recovery for both controls was within 85-115% at run 15. Supports claim of up to 15 uses.
In-use Stability (Reagent Cartridge)Stability claim precedes 95% CI of regression line reaching 85% or 115% recovery OR 2 data points or ≥2% of recovery data is ≤ 75% or ≥ 125% recovery.In-use (onboard) stability of 60 days was set. (Implies criteria were met up to that point).
Clinical Sensitivity (SSc)Not explicitly stated as a numerical acceptance criterion, but validation sought to demonstrate performance.42.3% (95% CI: 33.9% - 51.1%)
Clinical Specificity (Controls)Not explicitly stated as a numerical acceptance criterion, but validation sought to demonstrate performance.98.7% (95% CI: 96.9% - 99.4%)
Overall Agreement with PredicateNot explicitly stated as a numerical acceptance criterion (e.g., >X%). However, the comparison aimed to show substantial equivalence.Total Agreement: 97.6% (95% CI: 95.9% – 98.6%) for all 539 samples. Agreement within AMR (193 samples): Negative Agreement = 93.3% (95% CI: 88.1% – 96.3%), Positive Agreement = 95.5% (95% CI: 84.9% – 98.7%), Total Agreement = 93.8% (95% CI: 89.4% – 96.4%).

2. Sample Size Used for the Test Set and Data Provenance

  • Clinical Performance (Validation Set):

    • Sample Size: 498 samples.
    • Data Provenance: The text does not explicitly state the country of origin. It indicates the samples were a "separate set of samples, none of which were used in establishing the reference range." The patient groups include Systemic Sclerosis (SSc) and various control groups (Systemic Lupus Erythematosus, Rheumatoid Arthritis, Idiopathic Inflammatory Myopathy, Mixed Connective Tissue Disease, Celiac disease, Autoimmune thyroiditis, Sjögren's syndrome, Infectious disease, Crohn's disease, Osteoarthritis, COPD, Chronic Kidney Disease, Vasculitis, Raynaud's, Diabetes, Asthma, Skin Disease). The infectious disease samples specify Hepatitis C virus, Epstein-Barr virus, Toxoplasmosis, Cytomegalovirus, Mycoplasma infection, and Borrelia virus. The study is presented as evidence for the device's performance, suggesting prospective collection or a well-characterized retrospective cohort.
  • Comparison with Predicate Device:

    • Sample Size: 539 samples (the 498 samples from the Validation Set plus 41 additional contrived samples).
    • Data Provenance: Same as the clinical performance validation set, with additional contrived samples (diluted Scl-70 positive serum with negative serum).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not specify the number of experts or their qualifications for establishing the ground truth for the clinical diagnosis of systemic sclerosis (SSc) or other conditions used in the clinical validation set. It simply refers to "Diagnosis" in the tables relating to sensitivity and specificity. Given this is an in vitro diagnostic device, the ground truth for clinical conditions would typically be established by clinical diagnosis by treating physicians, potentially corroborated by other clinical and laboratory findings.

4. Adjudication Method for the Test Set

The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for establishing the ground truth diagnoses in the test set.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No multi-reader multi-case (MRMC) comparative effectiveness study is mentioned. This device is an automated chemiluminescent immunoassay; therefore, human reader assistance and improvement with AI assistance are not applicable. The comparison is between the new automated device and a predicate ELISA (enzyme-linked immunosorbent assay), not human readers.

6. Standalone Performance Study

Yes, a standalone performance study was done. The entire document describes the standalone performance of the QUANTA Flash® Scl-70 assay system (algorithm/device only without human-in-the-loop performance, as it's an automated immunoassay) across various analytical and clinical characteristics. The clinical sensitivity and specificity are direct measures of this standalone performance.

7. Type of Ground Truth Used

  • Clinical Performance (Sensitivity/Specificity): The ground truth was clinical diagnosis. For systemic sclerosis (SSc), this means patients diagnosed with SSc. For control groups, this refers to patients diagnosed with other autoimmune diseases or infectious diseases, or healthy individuals.
  • Analytical Performance: Ground truth was established by controlled experimental conditions, such as known concentrations for precision, linearity, and interference studies, or controlled conditions for stability studies.
  • Comparison with Predicate Device: The ground truth for this comparison was the result obtained from the legally marketed predicate device, QUANTA Lite® Scl-70 ELISA.

8. Sample Size for the Training Set

  • Reference Range Establishment (for Cut-off): 254 subjects were used. The sample groups included individuals with Rheumatoid Arthritis, Systemic Lupus Erythematosus, Hashimoto's Thyroiditis, Hepatitis B Virus, Hepatitis C Virus, Inflammatory Bowel Disease, Drug Induced Lupus, Autoimmune atrophic gastritis, Biliary anastomatic stricture, and Healthy Individuals.
  • Cut-off Adjustment: 19 systemic sclerosis samples that were positive on the predicate device were also used to aid in the final determination of the cutoff.
  • The document does not explicitly describe a separate "training set" in the context of machine learning, as this is an immunoassay, but rather samples used for establishing the reference range/cut-off.

9. How the Ground Truth for the Training Set was Established

  • Reference Range Establishment: The ground truth for the 254 subjects used to establish the reference range was based on their clinical diagnosis (e.g., Rheumatoid Arthritis, Systemic Lupus Erythematosus, or Healthy Individuals).
  • Cut-off Adjustment: The additional 19 systemic sclerosis samples had a ground truth based on their positive results on the predicate device (QUANTA Lite® Scl-70 ELISA) and their clinical diagnosis of systemic sclerosis.

The cut-off was established in accordance with CLSI C28-A3c ("Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition") and adjusted based on the results from the SSc samples positive on the predicate device to optimize differentiation.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three heads facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 1, 2016

INOVA Diagnostics, Inc. Dr. Gabriella Lakos Director, Assay Development 9900 Old Grove Road San Diego, CA 92131

Re: K152635

Trade/Device Name: QUANTA Flash® Scl-70 QUANTA Flash® Scl-70 Calibrators OUANTA Flash® Scl-70 Controls Regulation Number: 21 CFR 866.5100 Regulation Name: Antinuclear Antibodies Immunological Test System Regulatory Class: II Product Code: LLL, JIT, JJX Dated: April 27, 2016 Received: April 29, 2016

Dear Dr. Lakos:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements

{1}------------------------------------------------

as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kelly Oliner -S

FOR

Leonthena R. Carrington, MS, MBA, MT(ASCP) Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

Device Name

QUANTA Flash® Scl-70, QUANTA Flash® Scl-70 Calibrators, QUANTA Flash® Scl-70 Controls

Indications for Use (Describe)

QUANTA Flash Scl-70 is a chemiluminescent immunoassay for the semi-quantitative determination of IgG anti-Scl-70 autoantibodies in human serum. The presence of anti-Scl-70 autoantibodies, in conjunction with clinical findings and other laboratory tests, aids in the diagnosis of systemic sclerosis.

QUANTA Flash Scl-70 Calibrators are intended for use with the QUANTA Flash Scl-70 chemiluminescent immunoassay for the determination of IgG anti-Scl-70 autoantibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values.

QUANTA Flash Scl-70 Controls are intended for use with the OUANTA Flash Scl-70 chemiluminescent immunoassay for quality control in the determination of IgG anti-Scl-70 autoantibodies in human serum.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Inova Diagnostics. The logo consists of a red and blue abstract shape on the left, followed by the words "Inova" and "Diagnostics" stacked on top of each other on the right. The text is in a gray sans-serif font.

A Werfen Company

510(k) Summary

QUANTA Flash® Scl-70 QUANTA Flash® Scl-70 Calibrators QUANTA Flash® Scl-70 Controls

Page 1 of 19

510(k) Summary

Table of Contents

Administrative data
Device description
Intended use(s)
Substantial equivalence
Comparison to predicate device
Value assignment and traceability of Calibrators and Controls
Analytical performance characteristics
Precision
Reproducibility
Reproducibility between sites (instruments)
Reproducibility between lots
Limit of Blank (LoB) and Limit of Detection (LoD)
Analytical Measuring Range (AMR)
Auto-rerun function and reportable results
High concentration hook effect
Linearity
Interference
Cross-reactivity
Sample stability
Reagent stability
Cut-off ( reference range) establishment and verification
Clinical performance characteristics
Clinical sensitivity, specificity
Expected values
Comparison with predicate device

{4}------------------------------------------------

510(k) Summary QUANTA Flash® Scl-70

This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Administrative data

Submitter:Inova Diagnostics, Inc.9900 Old Grove Road,San Diego, CA, 92131
Purpose of submission:New device(s)
Devices in the submission:QUANTA Flash® Scl-70QUANTA Flash® Scl-70 CalibratorsQUANTA Flash® Scl-70 Controls
Scientific contact:Gabriella Lakos, Director of Assay DevelopmentInova Diagnostics, Inc.9900 Old Grove Road, San Diego, CA, 92131Phone: 858-586-9900/1393Fax: 858-863-0025email: glakos@inovadx.com
Quality Systems contact:Ronda Elliott, VP of Quality Systems and Regulatory AffairsInova Diagnostics, Inc.9900 Old Grove Road, San Diego, CA, 92131Phone: 858-586-9900Fax: 858-863-0025/1381email: relliott@inovadx.com
Preparation date:08/28/2015
Device name (assay kit):Proprietary name: QUANTA Flash® Scl-70
Common name:Anti-Scl-70 Chemiluminescent Immunoassay
Classification name:anti-Scl-70 antibody, antigen and control
Regulation DescriptionAntinuclear antibody immunological test system
Regulation Medical SpecialtyImmunology
Review PanelImmunology
Product CodeLLL
Regulation Number866.5100
Device Class2
Device name (Calibrators):Proprietary name:QUANTA Flash® Scl-70 Calibrators
Common name:Scl-70 Calibrators
Classification name:Calibrator, secondary
Regulation DescriptionCalibrator
Regulation Medical SpecialtyClinical Chemistry
Product CodeJIT
Regulation Number862.1150
Device Class2
Device name (Controls):Proprietary name:QUANTA Flash® Scl-70 Controls
Common name:Scl-70 Controls
Classification name:Single (specified) analyte controls (assayed and unassayed)
Regulation DescriptionQuality control material (assayed and unassayed)
Regulation Medical SpecialtyClinical Chemistry
Product CodeJJX
Regulation Number862.1660
Device Class1 (reserved)

{5}------------------------------------------------

{6}------------------------------------------------

Predicate device:

QUANTA Lite® Scl-70 ELISA, 510(k) number: K924898

Device description

The QUANTA Flash Scl-70 assay is designed to run on the BIO-FLASH® instrument. This platform is a fully automated closed system with continuous load and random access capabilities that automatically processes the samples, runs the assay and reports the results. It includes liquid handling hardware, luminometer and computer with software-user interface. The QUANTA Flash Scl-70 assay utilizes a reagent cartridge format, which is compatible with the BIO-FLASH instrument.

Recombinant Scl-70 is coated onto paramagnetic beads. The bead suspension is lyophilized and stored in the bead tube. Prior to use in the BIO-FLASH system, the sealed reagent tubes are pierced with the reagent cartridge lid and the beads are rehydrated and resuspended using resuspension buffer by pipetting up and down with a transfer pipette. The reagent cartridge is then loaded onto the BIO-FLASH instrument. Samples are also loaded onto the instrument in sample racks. A patient serum sample is prediluted 1:23.5 by the BIO-FLASH with system rinse in a disposable plastic cuvette. Small amounts of the diluted patient serum, the beads, and assay buffer are all combined into a second cuvette, and mixed. This cuvette is then incubated at 37°C. The beads are magnetized and washed several times. lsoluminol conjugated anti-human IgG antibodies are then added to the cuvette, and again incubated at 37°C. The beads are magnetized and washed repeatedly. The isoluminol conjugate is oxidized when Trigger 1 (Fe(III) coproporphyrin in sodium hydroxide solution) and Trigger 2 (urea-hydrogen peroxide in sodium chloride solution) are added to the cuvette, and the flash of light produced from this reaction is measured as Relative Light Units (RLU) by the BIO-FLASH optical system. The RLU are proportional to the amount of isoluminol conjugate that is bound to the human IgG, which is in turn proportional to the amount of anti-Scl-70 antibodies bound to the corresponding Scl-70 on the beads.

For quantitation, the QUANTA Flash Scl-70 assay utilizes a predefined lot specific Master Curve that is uploaded onto the instrument through the reagent cartridge barcode. Every new lot number of reagent cartridge must be calibrated before first use, with the QUANTA Flash Scl-70 Calibrators. Based on the results obtained with the two Calibrators included in the Calibrator Set (sold separately), an instrument specific Working Curve is created, which is used to calculate chemiluminescent units (CU)mL from the instrument signal (RLU) obtained for each sample.


The QUANTA Flash Scl-70 kit contains the following materials:

One (1) QUANTA Flash Scl-70 Reagent Cartridge

One (1) vial of Resuspension buffer

One (1) Transfer pipette

The QUANTA Flash Scl-70 Reagent Cartridge, containing the following reagents for 50 determinations:

a. Scl-70 antigen coated paramagnetic beads, lyophilized.

{7}------------------------------------------------

  • Assay Buffer 3 buffer containing protein stabilizers and preservatives. b.
  • C. Tracer IgG - Isoluminol labeled anti-human IgG antibodies in buffer, containing protein stabilizers and preservative.

The QUANTA Flash Scl-70 Calibrators kit contains two vials of Calibrator 1 and two vials of Calibrator 2.

  • -QUANTA Flash Scl-70 Calibrator 1: Two (2) barcode labeled tubes containing 0.3 mL prediluted, ready to use reagent. Calibrators contain human antibodies to Scl-70 in stabilizers and preservatives.
  • QUANTA Flash Scl-70 Calibrator 2: Two (2) barcode labeled tubes containing 0.3 mL prediluted, ready to use reagent. Calibrators contain human antibodies to Scl-70 in stabilizers and preservatives.

The QUANTA Flash Scl-70 Controls kit contains two vials of Negative Control and two vials of Positive Control.

  • QUANTA Flash Scl-70 Negative Control: Two (2) barcode labeled tubes containing 0.5 mL, ready to use reagent. Controls contain human antibodies to Scl-70 in stabilizers and preservatives.
  • । QUANTA Flash Scl-70 Positive Control: Two (2) barcode labeled tubes containing 0.5 mL, ready to use reagent. Controls contain human antibodies to Scl-70 in stabilizers and preservatives.

Intended use(s)

QUANTA Flash Scl-70 is a chemiluminescent immunoassay for the semi-quantitative determination of lgG anti-Scl-70 autoantibodies in human serum. The presence of anti-Scl-70 autoantibodies, in conjunction with clinical findings and other laboratory tests, aids in the diagnosis of systemic sclerosis.

QUANTA Flash Scl-70 Calibrators are intended for use with the QUANTA Flash Scl-70 chemiluminescent immunoassay for the determination of IgG anti-Scl-70 autoantibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values.

QUANTA Flash Scl-70 Controls are intended for use with the QUANTA Flash Scl-70 chemiluminescent immunoassay for quality control in the determination of IgG anti-Scl-70 autoantibodies in human serum.

{8}------------------------------------------------

Substantial equivalence

The QUANTA Flash Scl-70, the QUANTA Flash Scl-70 Calibrators and the QUANTA Flash Scl-70 Controls have the same intended use and assay principle as the predicate device.

Comparison to predicate device

QUANTA Flash Scl-70 reagent kit

Similarities
ItemQUANTA Flash Scl-70Predicate Device
Intended useQUANTA Flash Scl-70 is achemiluminescent immunoassay forthe semi-quantitative determinationof IgG anti-Scl-70 autoantibodies inhuman serum. The presence of anti-Scl-70 autoantibodies, in conjunctionwith clinical findings and otherlaboratory tests, aids in the diagnosisof systemic sclerosis.QUANTA Lite Scl-70 is an enzyme-linked immunosorbent assay (ELISA)for the semi-quantitative detection ofScl-70 antibodies in human serum. Thepresence of Scl-70 antibodies can beused in conjunction with clinicalfindings and other laboratory tests toaid in the diagnosis of scleroderma.
Assay methodologySolid phase (heterogenous)immunoassaySolid phase (heterogenous)immunoassay
Sample typeSerumSerum
Shelf lifeOne yearOne year
Differences
ItemQUANTA Flash Scl-70Predicate Device
Detection/Operating principleChemiluminescent immunoassayEnzyme-linked immunosorbent assay
Solid phaseParamagnetic microparticles (beads)96-well plate
AntigenRecombinantNative
ConjugateIsoluminol conjugated anti-human IgGHRP conjugated anti-human IgG
CalibrationLot specific Master Curve + twoCalibrators (Sold separately)Single standard(Included in the kit)

{9}------------------------------------------------

ItemQUANTA Flash Scl-70 CalibratorsPredicate Device
Intended useQUANTA Flash Scl-70 Calibrators areintended for use with the QUANTAFlash Scl-70 chemiluminescentimmunoassay for the determination ofIgG anti-Scl-70 autoantibodies inhuman serum. Each calibratorestablishes a point of reference for theworking curve that is used to calculateunit values.No separate intended use; calibratorsare part of the kit.
AnalyteAnti-Scl-70 antibodiesAnti-Scl-70 antibodies
MethodQUANTA Flash Scl-70chemiluminescent immunoassayQUANTA Lite Scl-70 ELISA
UnitCU (Chemiluminescent units)(arbitrary)units (arbitrary)
MatrixHuman serum, stabilizers, andpreservativeHuman serum, stabilizers, andpreservative
Physico-chemicalcharacteristicsLiquid, prediluted, ready to useLiquid, prediluted, ready to use
Storage2-8 °C2-8 °C
Shelf lifeOne yearOne year

QUANTA Flash Scl-70 Calibrators

QUANTA Flash Scl-70 Controls

ItemQUANTA Flash Scl-70 ControlsPredicate Device
Intended useQUANTA Flash Scl-70 Controls areintended for use with the QUANTAFlash Scl-70 chemiluminescentimmunoassay for quality control in thedetermination of IgG anti-Scl-70autoantibodies in human serum.No separate intended use; controls arepart of the kit.
AnalyteAnti-Scl-70 antibodiesAnti-Scl-70 antibodies
MethodQUANTA Flash Scl-70chemiluminescent immunoassayQUANTA Lite Scl-70 ELISA
UnitCU (Chemiluminescent units)(arbitrary)units (arbitrary)
MatrixHuman serum, stabilizers, andHuman serum, stabilizers, and

{10}------------------------------------------------

ItemQUANTA Flash Scl-70 ControlsPredicate Device
preservativepreservative
Physico-chemicalcharacteristicsLiquid, ready to useLiquid, prediluted, ready to use
Levels2 (negative and positive)2 (negative and positive)
Storage2-8 °C2-8 °C
Shelf lifeOne yearOne year

Value assignment and traceability of Calibrators and Controls

The QUANTA Flash Scl-70 Calibrators and Controls are manufactured by diluting human serum that contains high titer of antibodies with antibody stabilizer buffer, containing preservative. The human serum is obtained from commercial sources and it is tested for markers of infectious substances.

The target CU is achieved through trial dilutions on small scale. Once a dilution is selected, the Calibrators and Control are bulked, tested, and adjusted. Upon completion of the manufacturing process, the Calibrators and Controls are tested on at least two instruments, on at least two lots of reagent cartridge, in replicates of 10 to determine final value assignment.

Calibrator and Control values are directly traceable to the in-house Standards that are used to create the Master Curves for the QUANTA Flash Scl-70 assay.

MaterialAssigned Value (CU)
Scl-70 Master Curve Standard 10.0
Scl-70 Master Curve Standard 212.6
Scl-70 Master Curve Standard 342.4
Scl-70 Master Curve Standard 4106.4
Scl-70 Master Curve Standard 5290.8
Scl-70 Master Curve Standard 6786.3

List of Scl-70 Standards, Calibrators and Controls:

{11}------------------------------------------------

MaterialManufacturingTarget ValueManufacturingTarget Range
Scl-70 Calibrator 113 CU11 - 15 CU
Scl-70 Calibrator 2290 CU261 - 319 CU
Scl-70 Negative Control10 CU8-12 CU
Scl-70 Positive Control50 CU40-60 CU

Analytical performance characteristics

Precision

The precision of the QUANTA Flash Scl-70 assay was evaluated on 13 samples containing various concentrations of Scl-70 antibodies in accordance with CLSI EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Procedures - Approved Guideline: samples were run in duplicates, twice a day, for 20 days. Data were analyzed with the Analyse-it for Excel method evaluation software, and within run, between run, between day and total imprecisions are summarized in the Table below. Total %CV values were within the acceptance limit, 10%.

WithinRunBetween-RunBetween-DayTotal
SampleNMeanSD%CVSD%CVSD%CVSD%CV
Precision 18022.90.52.1%0.52.0%0.52.3%0.93.7%
Precision 28089.41.41.6%1.92.2%3.33.7%4.14.6%
Precision 38022.70.52.3%0.52.1%0.62.5%0.93.9%
Precision 48028.30.62.0%0.72.5%0.31.1%1.03.4%
Precision 580700.313.92.0%11.91.7%30.44.3%35.55.1%
Precision 68010.70.32.7%0.44.1%0.00.0%0.54.9%
Precision 78010.80.65.3%0.21.9%0.21.6%0.65.9%
Precision 88058.51.01.8%1.42.3%1.62.6%2.34.0%
Precision 98023.620.41.6%0.52.3%0.93.7%1.14.6%
Precision 108021.170.52.1%0.62.6%0.73.1%1.04.6%
Precision 118023.910.41.8%0.52.1%1.04.1%1.25.0%
Precision 1280368.548.02.2%5.21.4%13.23.6%16.34.4%
Precision 1380534.3310.21.9%12.02.2%19.13.6%24.74.6%

{12}------------------------------------------------

Reproducibility

Reproducibility between sites (instruments)

Eight samples were tested on three different instruments at three different sites. Samples were run in replicates of five, once a day for 5 days, to generate 25 data points per sample, per site.

Data were analyzed with the Analyse-it for Excel method evaluation software, between sites imprecision was calculated, and the results are summarized in the Table below. All %CV values were within the acceptance limit, 15%.

QUANTA Flash Scl-70Between Site Precision (Reproducibility)
Sample IDNumber of ReplicatesMean (CU)SD (CU)CV (%)
175682.411.71.7
275404.915.63.9
375109.65.04.5
47535.80.51.3
57510.10.54.5
67520.01.78.3
77518.50.95.0
87520.01.26.1

Reproducibility between lots

Lot to lot reproducibility study was performed according to CLSI EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures, by testing eight samples with three different lots of reagents in five replicates for 5 days, to generate 25 data points per lot, 75 data points total.

Data were analyzed with the Analyse-it for Excel method evaluation software, between lots imprecision was calculated, and the results are summarized in the Table below. All %CV values were within the acceptance limit, 10%.

{13}------------------------------------------------

Between LotImprecision
Mean(CU)SD (CU)CV
Sample IDN
17514.11.06.8%
27561.10.91.5%
375202.412.96.4%
475513.939.47.7%
575564.236.26.4%
67520.71.46.8%
77519.80.84.2%
87519.61.99.6%

Limit of Quantitation (LoQ), Limit of Blank (LoB) and Limit of Detection (LoD)

The Limit of Quantitation (LoQ) of the QUANTA Flash Scl-70 assay is 1.2 CU, which defines the lower limit of the AMR. The LoQ was determined consistent with CLSI EP17-A2 guideline the total error (TE) of each sample per reagent lot, using the Westgard model (TE= /Bias/ + 2s) based on 120 measurements of four low level samples. The total error (TE) for the LoQ is < 25% (accuracy goal). The sample with the lowest CU that met the accuracy goal specification is assigned LoQ.

The LoD of the QUANTA Flash Scl-70 assay is 0.2 CU, which is below the analytical measuring range of the assay. It was determined consistent with CLSI EP17-A guideline with proportions of false positives (alpha) less than 5% and false negatives (beta) less than 5%; based on 240 determinations, with 120 measurements on blank samples and 120 measurements of low level samples, per reagent lot. The LoB is 0.1 CU.

For the LoB study, four different aliquots of System Rinse were tested in five replicates over 3 days, on two reagent lots (60 data points per lot). The LoB was calculated separately on the two lots, and the higher value was used as the final LoB.

For the LoD study, four low level samples were tested in five replicates for 3 days on two reagent lots (60 data points per lot). The LoD was calculated separately on the two lots, and the higher value was used as the final LoD.

Analytical Measuring Range (AMR)

1.2 CU - 786.3 CU

The AMR is defined by the values of the Limit of Quantitation and highest Master Curve Standard.

{14}------------------------------------------------

510(k) Summary QUANTA Flash® Scl-70

Auto-rerun function and reportable results

The BIO-FLASH software has an Auto-rerun option available. If this option is selected, the instrument will automatically re-test any sample that has a result of result >786.3 CU by further diluting it by a factor specified in the assay definition file (20 fold), thereby bringing the measured value within the AMR. The final result will be calculated by the software by taking into account the additional dilution factor. As the highest value that can be measured is 786.3 CU, the highest value that can be reported is 15726 CU.

To validate the Auto-rerun function, three high positive specimens with results above the analytical measuring range were selected. The samples were run with the Auto-rerun function enabled on the BIO-FLASH. Then the specimens were manually diluted 20 fold and tested on the BIO-FLASH. The results were within the analytical measuring range after auto-rerun or manual dilution for all specimens. The % recovery values for results obtained with the auto-rerun results compared to the results obtained by manual dilution were 104%, 101% and 91% (average 102%) and are within the ± 20% acceptance limit.

High concentration hook effect

N/A

Linearity

The linearity of the AMR was evaluated by a study according to CLSI EP6-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline. Five serum samples with various Scl-70 antibody concentrations were diluted with negative serum in 10% increments (from 0% to 90% negative serum) to obtain values that cover the AMR. Diluted samples were assayed in duplicate. Percent recovery was calculated compared to expected results (based on dilution). Percent recovery for all data points ranged from 80.2% to 118.4%. Obtained values were plotted against expected values, and linear regression analysis was performed on each samples, and also on the combined results. Acceptance criteria were 80%-120% recovery, 0.9-1.1 slope and ≥0.95 R . Linear regression results are shown in the Table below.

Sample IDTest Range (CU)Slope (95% CI)
Sample 272.0 - 727.51.00 (0.96 to 1.04)0.99
Sample 37.9 - 101.31.03 (1.00 to 1.05)1.00
Sample 41.8 - 20.21.01 (0.98 to 1.04)1.00
Sample 51.7 - 9.71.01 (0.98 to 1.04)1.00
Sample 681.2 - 739.60.96 (0.92 to 1.00)1.00
All1.7 - 739.61.01(0.99 to 1.02)0.99

Interference

The interference study was performed according to CLSI EPO7-A2, Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition. Three specimens were tested (negative: 10.7 CU; low: 21.5 CU; positive: 49.4 CU). Interfering substances were spiked into every specimen at three different

{15}------------------------------------------------

510(k) Summary QUANTA Flash® Scl-70

concentrations in 10% of total specimen volume, and the resulting samples were assessed in triplicates with the Scl-70 assay. Recovery of the unit values was calculated compared to control samples spiked with the same volume of diluent (10% of total). Acceptance criteria for the interference studies were 85% - 115% recovery. The following interfering substances were tested:

Interferingsubstanceconcentration #1testedconcentration #2testedconcentration #3tested
Bilirubin, conjugated10 mg/dL5 mg/dL2.5 mg/dL
Hemoglobin200 mg/dL100 mg/dL50 mg/dL
Triglicerydes1000 mg/dL500 mg/dL250 mg/dL
Cholesterol332.5 mg/dL166.3 mg/mL83.1 mg/mL
Human IgG70 mg/mL35 mg/mL17.5 mg/mL
RF IgM500 IU/mL300 IU/mL100 IU/mL
Prednisone0.3 mg/mL0.15 mg/mL0.08 mg/mL
Naproxen25.6 mg/mL12.8 mg/mL6.4 mg/mL

No interference was detected with bilirubin up to 10 mg/dL (recovery: 88% to 101%), hemoglobin up to 200 mg/dL (recovery: 93% to 105%), triglycerides up to 1000 mg/dL (recovery: 89% to 97%), cholesterol up to 224.3 mg/dL (recovery: 93% to 106%), human IgG up to 70 mg/mL (recovery 90-109%m or < 4 CU), RF IgM up to 500 IU/mL (recovery: 97% to 11%), predinose up to 0.3 mg/mL (recovery: 100.0 to 109.7) and naproxen up to 25.6 mg/mL (recovery: 97.7 to 109.7%).

Cross-reactivity

To test potential cross-reactivity with autoantibodies and infection-induced antibodies, results obtained on 375 control samples that were included in the clinical validation study were assessed. These samples were from patients with autoimmune diseases that are characterized with disease specific autoantibodies, or from patients with infection. The composition of the anti-Scl-70 positivity rate is shown in the Table below:

Patient groupNumber of samples# Positive% Positive
Systemic Lupus Erythematosus3200.0%
Rheumatoid Arthritis3100.0%
Idiopathic Inflammatory Myopathy2500.0%
Mixed Connective Tissue Disease2500.0%
Celiac disease2500.0%
Autoimmune thyroiditis2500.0%
Sjögren's syndrome2000.0%
Infectious disease*3013.3%

{16}------------------------------------------------

Crohn's disease5423.7%
Osteoarthritis2813.6%
COPD1500.0%
Chronic Kidney Disease1000.0%
Vasculitis1500.0%
Raynaud's1000.0%
Diabetes500.0%
Asthma1500.0%
Skin Disease10110.0%
Total controls37541.4%

Sample stability

Six samples, encompassing negative, around the cutoff and low to high positive samples were tested in duplicates for up to 21 days at 2-8°C, up to 48 hours at room temperature, moreover, after repeated freeze/thaw cycles up to 3 cycles. Results were compared to those obtained on control samples (day zero, at 2-8°C)

Acceptance criteria: 90-110% average recovery.

All samples fulfilled the acceptance criteria at each time point for each condition.

Based on these result, we recommend that samples are stored up to 48 hours at RT, up to 21 days of at 2-8 °C, and can be subjected to up to 3 freeze/thaw cycles (when samples are stored at or below -20 °C).

Reagent stability

Shelf life

Accelerated Stability

To establish the initial claim for shelf life, accelerated stability studies were performed for 4 weeks at 37 °C.

Accelerated stability testing was performed on three lots of each of the following sealed components of the QUANTA Flash Scl-70 to establish initial stability claim:

  • Scl-70 beads
  • Calibrators 1 and 2
  • Low and High controls

Each week a new sealed component was placed in the incubator, and all components were tested at the end of the experiment together with the one that was stored at 5 ± 3°C. The recovery of the measured values was calculated for each time point (compared to those obtained with 5 ± 3°C stored reagent). All calculations were performed by comparing results of sealed components stored at 5 ± 3℃ (control) to those stored at 37 ± 3℃ (test) for 1, 2, 3, and 4 weeks, where one week is equal to six months at 5 ± 3℃. Linear regression analysis was performed between recovery values and the number of days.

{17}------------------------------------------------

Acceptance criteria for one year preliminary expiration dating were:

-Microparticles (beads):

With regression analysis, the 95% Cl interval of the regression line is between 85 and 115 % at 2 weeks, and no individual data point is outside the 75-125 %recovery range at 2 weeks.

  • Controls and Calibrators:

With regression analysis, the lower 95% Cl interval of the regression line is between 90 and 110 % at 2 weeks, and no individual data point is outside the 80-120 %recovery range at 2 weeks.

All components tested fulfilled the acceptance criteria above, so one year expiration dating was assigned to each component.

Real time stability

To confirm the shelf life that was assigned based on accelerated stability studies, real time stability studies have been initiated.

Real time stability testing will be performed on Calibrators, Controls and Reagents at regular time points to support the one year expiration. We are providing real time stability data for Calibrators, Controls, and Reagents up to 6 months

Controls were tested in triplicates at each time point.

  • Acceptance criteria: results should fall within their acceptable ranges as it was established at the release of the controls.

Calibrators were tested in triplicates at each time point as samples. Averages of the triplicates were compared to the value that was assigned to the Calibrators at release.

  • Acceptance criteria: % recovery of the average of the triplicates is between 85 and 115%, and %CV of the triplicates is < 10%

For reagent cartridge, 3 samples were tested at each time point, At every time point, all samples must be tested in three replicates.

Acceptance criteria for results obtained at consecutive time points:

  • %CV of the replicates is < 15% at each time point.

  • The percent recovery of each sample is between 80-120%.

All results were within the acceptance limits.

In-use (onboard) stability

Calibrators

Onboard stability claim: 4 calibrations, or 8 hours onboard

{18}------------------------------------------------

During assessing on-board stability, Calibrators were placed, uncapped, onboard the instrument, and calibration was performed altogether five times over 8.5 hours. Controls and a panel of characterized patient specimens were run on each calibration curve.

Acceptance criteria: Calibrators are considered stable if all five calibrations performed in the 8.5 hour period are successful, and Calibrator average RLU recovery values are between 90% and 110% compared to the first use.

A total of 5 successful calibrations were performed over an 8.5 hour period. Calibrator RLU values remained within the 90-110% range. Moreover, all characterized patient samples ran within their expected range. This supports the claim that calibrators can be used for up to 4 calibrations over an 8 hour period.

Controls

Onboard stability claim: up to 15 uses, at 10 minutes onboard per use

During assessing on-board stability, Low and High Controls were assayed for a total of 21 runs. The controls were left uncapped, onboard the instrument for 15 minutes per run. When not in use, the controls were capped, and stored at 5 ± 3℃ for at least 2 hours between runs.

Acceptance criteria: Controls are considered stable when all replicates run within their established range, moreover, the linear regression line obtained by plotting %recovery values against the number of runs stays is between 85 % and 115 % at run 15.

Low and High Controls ran within their respective acceptable range for all runs. The linear regression line obtained by plotting %recovery values against the number of runs was within 85 % and 115 % at run 15 for both Controls.

Reagent Cartridge

To determine the in-use stability of the QUANTA Flash Scl-70 reagent cartridge, two lots of reagent were tested with four to six serum specimens (with different reactivity levels) along with the Negative and Positive Controls were tested. The specimens were tested periodically for up to 60 days. Recoveries were calculated compared to the day zero average values, and linear regression analysis was performed. The claim is established using the following criteria (using the one that is fulfilled first):

  • The stability claim is established on the actual measurement day preceding the 95% confidence interval of the regression line reaches 85% or 115% recovery, or

  • On the actual measurement day preceding the day when 2 data points or ≥2% of the recovery data (whichever is greater) is ≤ 75% or ≥ 125% recovery.

The in-use (onboard) stability of Scl-70 reagent cartridge was set at 60 days.

Cut-off (reference range) establishment and verification

The reference population for establishing and verifying the reference interval for the Scl-70 assay consisted of 254 subjects:

{19}------------------------------------------------

Sample GroupN
Rheumatoid Arthritis49
Systemic Lupus Erythematosus50
Hashimoto's Thyroiditis30
Hepatitis B Virus26
Hepatitis C Virus10
Inflammatory Bowel Disease17
Drug Induced Lupus16
Autoimmune atrophic gastritis4
Biliary anastomatic stricture2
Healthy Individuals50
Total254

All specimens were the same matrix (serum) as specified in the Intended Use. All specimens were unaltered. The cut-off was established in accordance to CLSI C28-A3c: Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition. Using Analyse-it for Excel, the upper 99th percentile Reference Interval limit was calculated as 7387.3 RLUs.

Additionally, 19 systemic sclerosis samples that were positive on the predicate device were tested to aid in the determination of the cutoff. Taking into account the results of these samples, the cutoff was increased to 15,000 RLU to ensure optimal differentiation between negatives and positives, and a 20 CU value was assigned to this RLU value.

Clinical performance characteristics

Clinical sensitivity, specificity

A separate set of samples, none of which were used in establishing the reference range, was used to validate the clinical performance of the QUANTA Flash Scl-70. A total of 498 samples were included in the Validation Set for the QUANTA Flash Scl-70.

Distribution of the cohort used in the QUANTA Flash Scl-70 validation study:

Patient groupN# Positive% Positive
Systemic Lupus Erythematosus3200.0%
Rheumatoid Arthritis3100.0%
Idiopathic Inflammatory Myopathy2500.0%
Mixed Connective Tissue Disease2500.0%
Celiac disease2500.0%
Autoimmune thyroiditis2500.0%
Sjögren's syndrome2000.0%

{20}------------------------------------------------

Infectious disease*3013.3%
Crohn's disease5423.7%
Osteoarthritis2813.6%
COPD1500.0%
Chronic Kidney Disease1000.0%
Vasculitis1500.0%
Raynaud's1000.0%
Diabetes500.0%
Asthma1500.0%
Skin Disease10110.0%
Total controls37551.3%
Systemic Sclerosis (SSc)1235242.3%
Total498
  • The specific infectious disease samples tested are outlined in the table below.
Patient groupN# Positive% Positive
Hepatitis C virus10110.0%
Epstein-Barr virus1000.0%
Toxoplasmosis400.0%
Cytomegolovirus400.0%
Mycoplasma infection100.0%
Borrelia virus100.0%

Clinical sensitivity and specificity of the QUANTA Flash Scl-70 in systemic sclerosis

Clinical Study (N=498)QUANTA Flash®Scl-70Analysis(95% confidence)
PositiveNegativeTotal
DiagnosisSSC5271123Sensitivity = 42.3% (33.9 – 51.1%)
Controls5370375Specificity = 98.7% (96.9-99.4%)
Total57441498

Expected values

The expected value in the normal population is "negative". Anti-Scl-70 antibody levels were analyzed in a cohort of 100 apparently healthy blood donors (42 females and 58 males, ages 21 to 67 years, with an average and median age of 46 years) using the QUANTA Flash Scl-70. This patient population was

{21}------------------------------------------------

different from the one that was used to establish the cutoff. The mean concentration was <1.2 CU, and the values ranged from <1.2 to 2.2 CU.

Comparison with predicate device

All samples from the Validation Set study were tested on both the QUANTA Flash Scl-70 and on the predicate ELISA, along with 41 additional samples. These additional samples were contrived by diluting Scl-70 positive serum with negative serum. The comparison is shown in the tables below. Method comparison on all clinical samples:

All Samples (n=539)QUANTA Flash® Scl-70Percent Agreement (95% confidence)
NegativePositiveTotal
Scl-70 ELISANegative44111452Negative Agreement = 97.6% (95.78 – 98.6%)
Positive28587Positive Agreement = 97.7 % (92.0 – 99.4%)
Total44396539Total Agreement = 97.6% (95.9 – 98.6%)

Out of the 539 samples, results were within the AMR of both the QUANTA Flash assay and of the predicate ELISA for 193 samples. Agreement on samples within the AMR is shown below:

Samples within theAMR (n=193)QUANTA Flash®Scl-70Percent Agreement(95% confidence)
NegativePositiveTotal
Scl-70ELISANegative13910149Negative Agreement = 93.3% (88.1 – 96.3%)
Positive24244Positive Agreement = 95.5% (84.9 – 98.7%)
Total14152193Total Agreement = 93.8% (89.4 – 96.4%)

§ 866.5100 Antinuclear antibody immunological test system.

(a)
Identification. An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).(b)
Classification. Class II (performance standards).