(240 days)
Not Found
No
The device is a chemical indicator that reacts to steam sterilization parameters, with a visual readout. There is no mention of computational processing, algorithms, or learning from data.
No
The device is an integrating indicator used to monitor steam sterilization cycles, not to treat a medical condition or disease.
No
This device is a chemical indicator used to monitor steam sterilization cycles in healthcare settings, ensuring that sterilization criteria have been met. It indicates whether items are sterile, but it does not diagnose a medical condition in a patient.
No
The device description clearly states it is a "single use device" that is "included in a pack or load in a steam sterilizer" and has a "dark bar on the device" that enters an "ACCEPT (OK) window". This describes a physical, chemical indicator, not a software-only device.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
- Device Function: The described device is a chemical indicator used to monitor the effectiveness of steam sterilization cycles for medical instruments and supplies. It reacts to the physical parameters of the sterilization process (temperature, time, steam) to indicate whether the cycle was successful in achieving sterilization conditions.
- Intended Use: The intended use is to monitor the sterilization process of medical devices, not to diagnose or provide information about a patient's health.
- Device Description: The description clearly states it's a "single use device used by healthcare providers to monitor steam sterilization cycles."
- Lack of Biological Samples: The device does not interact with or analyze any biological samples from a patient.
Therefore, this device falls under the category of a medical device used in the sterilization process, but it is not an in vitro diagnostic device.
N/A
Intended Use / Indications for Use
The integrating indicator is designed to chemically react over time with the critical parameters of steam sterilization cycle within a specified tolerance. The integrating indicator strip is intended to be placed in each pack, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles:
Steam Sterilization Cycles:
- · 250°F/121°C, 30 minutes Gravity
- · 270°F/132°C, 4 minutes Dynamic Air Removal
- · 270°F/132°C, 15 minutes Gravity
- · 275°F/135°C, 3 minutes Dynamic Air Removal
- · 275°F/135°C, 10 minutes Gravity
Steam Sterilization Cycles (IUSS):
- · 270°F/132°C, 4 minutes Dynamic Air Removal
- · 270°F/132°C, 3 minutes Gravity
- · 270°F/132°C, 10 minutes Gravity
- · 275°F/135°C, 3 minutes Dynamic Air Removal
- · 275°F/135°C, 3 minutes Gravity
- · 275°F/135°C, 10 minutes Gravity
Product codes (comma separated list FDA assigned to the subject device)
JOJ
Device Description
The VERIFY® STEAM Integrating Indicator is a single use device used by healthcare providers to monitor steam sterilization cycles. The VERIFY® STEAM Integrating Indicator is included in a pack or load in a steam sterilizer and the load is processed in accordance with the sterilizer manufacturer's directions. Prior to the use of the processed items, the Integrator is observed. If the dark bar on the device enters the ACCEPT (OK) window, the integrator is read as a PASS to indicate that the steam sterilization criteria for the cycle have been met. If the dark bar on the device does not enter the ACCEPT (OK), the integrator is read as a FAIL, indicating that sufficient steam sterilization criteria has not been met and processed materials should be subjected to another steam sterilization cycle prior to use.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
healthcare providers
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Performance testing was conducted to verify that the proposed VERIFY® STEAM Integrating Indicator meets the requirements for integrating indicators in accordance with the Guidance for Industry and FDA Staff: Premarket Notification [510(k)] Submissions for Chemical Indicators for integrating indicators as well as ANSI/AAMI/ISO 11140-1:2014.
Study Type: BIER Vessel Testing
Sample Size: 3 Lots
Key Results: Pass result at SV for each temperature claimed (23.0 min at 121°C, 2.0 min at 132 °C, 1.2 min at 135 °C). Fail result at 15% below SV for each temperature claimed. Fail result at 1°C below SV for each temperature claimed. (Result: PASS)
Study Type: Simulated Use Testing in Claimed Sterilization Cycles
Key Results: 100% Pass result under pass conditions. 100% Fail result under fail conditions. (Result: PASS)
Study Type: Dry Heat Testing
Key Results: Fail result when exposed to 140°C dry heat for 30 min. (Result: PASS)
Study Type: Verification of integrator stated values at biological indicator growth-negative cycle conditions
Key Results: Integrator does not reach endpoint before the biological indicator is inactivated. (Result: PASS)
Study Type: Offset/Bleeding
Key Results: Integrator shall not bleed or offset. (Result: PASS)
Conclusion: The results of the VERIFY® STEAM Integrating Indicator performance testing demonstrate that both perform as intended in the claimed steam sterilization cycles and the proposed device is substantially equivalent to the predicate device.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 880.2800 Sterilization process indicator.
(a)
Biological sterilization process indicator —(1)Identification. A biological sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization.(2)
Classification. Class II (performance standards).(b)
Physical/chemical sterilization process indicator —(1)Identification. A physical/chemical sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.(2)
Classification. Class II (performance standards).
0
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features the department's name in a circular arrangement around an emblem. The emblem is a stylized representation of a human figure, composed of three overlapping profiles, symbolizing health and well-being.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 12, 2016
Steris Corporation Bill Brodbeck, Ph.D. Director, Regulatory Affairs 5960 Heisley Road Mentor, Ohio 44060
Re: K152630
Trade/Device Name: VERIFY® STEAM Integrating Indicator Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization process indicator Regulatory Class: Class II Product Code: JOJ Dated: April 11, 2016 Received: April 12, 2016
Dear Dr. Brodbeck:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
1
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K152630
Device Name VERIFY® STEAM Integrating Indicator
Indications for Use (Describe)
The integrating indicator is designed to chemically react over time with the critical parameters of steam sterilization cycle within a specified tolerance. The integrating indicator strip is intended to be placed in each pack, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles:
Steam Sterilization Cycles:
- · 250°F/121°C, 30 minutes Gravity
- · 270°F/132°C, 4 minutes Dynamic Air Removal
- · 270°F/132°C, 15 minutes Gravity
- · 275°F/135°C, 3 minutes Dynamic Air Removal
- · 275°F/135°C, 10 minutes Gravity
Steam Sterilization Cycles (IUSS):
- · 270°F/132°C, 4 minutes Dynamic Air Removal
- · 270°F/132°C, 3 minutes Gravity
- · 270°F/132°C, 10 minutes Gravity
- · 275°F/135°C, 3 minutes Dynamic Air Removal
- · 275°F/135°C, 3 minutes Gravity
- · 275°F/135°C, 10 minutes Gravity
Type of Use (Select one or both, as applicable)
_ Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
3
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
4
Image /page/4/Picture/1 description: The image shows the logo for STERIS. The logo consists of the word "STERIS" in a bold, sans-serif font, with a registered trademark symbol to the right. Below the word is an image of several horizontal, wavy lines in a light blue color, resembling water waves.
510(k) Summary for VERIFY® STEAM Integrating Indicator
Sponsor Facility
STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (216) 354-2600 Fax No: (216) 639-4459
Manufacturing Facility
Dana Products, Inc. 11457 Melrose Street Franklin Park, IL 60131
Contact:
Bill Brodbeck, Ph.D. Director, Regulatory Affairs Telephone: (440) 392-7690 (440) 357-9198 Fax No: e-mail: William brodbeck(@steris.com
May 12, 2016 Submission Date:
STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600
5
1. Device Name
Trade Name: | VERIFY® STEAM Integrating Indicator |
---|---|
Device Classification: | Class II |
Common/usual Name: | Indicator, physical/chemical sterilization process |
Classification Name: | Indicator, physical/chemical sterilization process |
(21 CFR 880.2800, JOJ) |
2. Predicate Device
Dana Steam Sterilization Integrator (K012195), Dana Products Inc.
3. Description of Device
The VERIFY® STEAM Integrating Indicator is a single use device used by healthcare providers to monitor steam sterilization cycles. The VERIFY® STEAM Integrating Indicator is included in a pack or load in a steam sterilizer and the load is processed in accordance with the sterilizer manufacturer's directions. Prior to the use of the processed items, the Integrator is observed. If the dark bar on the device enters the ACCEPT (OK) window, the integrator is read as a PASS to indicate that the steam sterilization criteria for the cycle have been met. If the dark bar on the device does not enter the ACCEPT (OK), the integrator is read as a FAIL, indicating that sufficient steam sterilization criteria has not been met and processed materials should be subjected to another steam sterilization cycle prior to use.
4. Intended Use
The integrating indicator is designed to chemically react over time with the critical parameters of steam sterilization cycle within a specified tolerance. The integrating indicator strip is intended to be placed in each pack, pouch, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles:
Steam Sterilization Cycles:
- · 250°F/121°C, 30 minutes Gravity
- · 270°F/132°C, 4 minutes Dynamic Air Removal
- · 270°F/132°C, 15 minutes Gravity
- · 275°F/135°C, 3 minutes dynamic air removal
- · 275°F/135°C, 10 minutes Gravity
6
Steam Sterilization Cvcles (IUSS):
- · 270°F/132°C, 4 minutes Dynamic Air Removal
- · 270°F/132°C, 3 minutes Gravity
- · 270°F/132°C, 10 minutes Gravity
- · 275°F/135°C, 3 minutes dynamic air removal
- · 275°F/135°C, 3 minutes Gravity
- · 275°F/135°C, 10 minutes Gravity
ડ. Description of Safety and Substantial Equivalence
The proposed and predicate devices are single use integrating indicators for use in monitoring steam sterilization cycles. The products are identical in design with the exception of the labeling, which is being revised as a result of rebranding the device under the VERIFY product family (STERIS purchased Dana Products Inc. in March, 2015). The differences between the proposed VERIFY® |STEAM Integrating Indicator and the predicate Dana Steam Sterilization Integrator are limited to expanding the indications for use statement to include use of the device to monitor various steam sterilization cycles. These differences do not raise any new issues of safety and efficacy. Performance testing has been conducted for monitoring all claimed steam sterilization cycles in accordance with the Guidance for Industry and FDA Staff: Premarket Notification [510(k)] Submissions for Chemical Indicators.
6. Technological Characteristics
The proposed and predicate devices are single use integrating indicators for use in monitoring particular steam sterilization cycles. The device components, mechanism of action, and endpoint are the same as the predicate's and when exposed to the defined processing conditions, the proposed and predicate devices exhibit a visible migration of the included chemistry to indicate that stated values have been achieved.
Table 5-1 contains a comparison of technological characteristics and specifications of the proposed VERIFY® STEAM Integrating Indicator to the predicate Dana Steam Sterilization Integrator.
7
Proposed | K012195 | Comparison | |
---|---|---|---|
Feature | VERIFY® STEAM | ||
Integrating Indicator | Dana Steam Sterilization | ||
Integrator | |||
Intended use | The integrating indicator is | ||
designed to chemically react | |||
over time with the critical | |||
parameters of steam | |||
sterilization cycle within a | |||
specified tolerance. The | |||
integrating indicator strip is | |||
intended to be placed in each | |||
pack, pouch, container, tray or | |||
other containment device to | |||
function as an independent | |||
monitor of critical parameters | |||
for the following sterilization | |||
cycles: | |||
Steam Sterilization Cycles: | |||
•250°F/121°C, 30 minutes | |||
Gravity | |||
•270°F/132°C, 4 minutes | |||
Dynamic Air Removal | |||
•270°F/132°C, 15 minutes | |||
Gravity | |||
•275°F/135°C, 3 minutes | |||
dynamic air removal | |||
•275°F/135°C, 10 minutes | |||
Gravity | |||
Steam Sterilization Cycles | |||
(IUSS): | |||
•270°F/132°C, 4 minutes | |||
Dynamic Air Removal | |||
•270°F/132°C, 3 minutes | |||
Gravity | |||
•270°F/132°C, 10 minutes | |||
Gravity | |||
•275°F/135°C, 3 minutes | |||
dynamic air removal | |||
•275°F/135°C, 3 minutes | |||
Gravity | |||
•275°F/135°C, 10 minutes | |||
Gravity | Dana Steam Sterilization | ||
Integrators are intended to be | |||
used to monitor critical | |||
parameters in steam | |||
sterilization cycles in | |||
healthcare facilities. The | |||
critical parameters for which | |||
the integrators will respond | |||
are time and temperature | |||
(when steam is present). The | |||
end point is determined by the | |||
migration of the steam | |||
sensitive dye to an area | |||
marked safe on the indicator. | |||
The integrators are intended to | |||
be used in 30 minute gravity | |||
cycles at 250 degrees F and 3 | |||
minute gravity and pre- | |||
vacuum cycles at 270 degrees | |||
F. | The intended use | ||
statement has been | |||
updated to reflect | |||
specific uses of the | |||
subject device as | |||
well expand the | |||
indicated cycles to | |||
include those | |||
commonly utilized | |||
in healthcare | |||
facilities. Testing in | |||
conformance with | |||
the Guidance for | |||
Industry and FDA | |||
Staff: Premarket | |||
Notification | |||
[510(k)] | |||
Submissions for | |||
Chemical Indicators | |||
for integrating | |||
indicators addresses | |||
any concerns | |||
regarding the safety | |||
and efficacy of the | |||
proposed device in | |||
regards to the | |||
claimed steam | |||
sterilization cycles. |
Table 5-1. Device Comparison Table
8
K152630 Response to 5/12/16 Interactive Review
VERIFY® STEAM Integrating Indicator
Proposed | K012195 | Comparison | |
---|---|---|---|
Feature | VERIFY® STEAM | Dana Steam Sterilization | |
Integrating Indicator | Integrator | ||
Device | |||
design - | |||
components | Backing material with | ||
embossed cavity containing | |||
temperature sensitive chemical | |||
and coloring dye, wicking strip, | |||
covered with laminated paper | |||
containing labeling and | |||
windows. | Backing material with | ||
embossed cavity containing | |||
temperature sensitive chemical | |||
and coloring dye, wicking strip, | |||
covered with laminated paper | |||
containing labeling and | |||
windows. | Device design and | ||
components are | |||
identical with | |||
exception of the | |||
printed labeling. | |||
Indicator | |||
agent | Proprietary formulation | Proprietary formulation | The indicator agent |
is identical to the | |||
predicate. | |||
Sterilization | |||
method and | |||
cycles | Per intended use above. | 30 minute gravity cycles at | |
250 degrees F and 3 minute | |||
gravity and pre-vacuum cycles | |||
at 270 degrees F. | The intended cycles | ||
vary between | |||
indicators. Intended | |||
cycles for the | |||
proposed device will | |||
be verified and | |||
validated in | |||
accordance with the | |||
Guidance for | |||
Industry and FDA | |||
Staff: Premarket | |||
Notification | |||
[510(k)] | |||
Submissions for | |||
Chemical Indicators. | |||
Mechanism | |||
of action | Proprietary | Proprietary | Mechanism of |
action is identical to | |||
predicate. | |||
Endpoint | |||
specifications | The end point is determined | ||
by the migration of the steam | |||
sensitive dye to an area | |||
marked ACCEPT (OK) on the | |||
indicator. Endpoint is reached | |||
at the stated value (SV) for | |||
each claimed temperature. | |||
Endpoint is not reached at the | |||
stated value - 15% time and/or | |||
-1°C. | The end point is determined | ||
by the migration of the steam | |||
sensitive dye to an area | |||
marked safe on the indicator. | |||
Endpoint is reached at the | |||
stated value (SV) for each | |||
claimed temperature. Endpoint | |||
is not reached at the stated | |||
value - 15% time and/or -1°C. | The end point is the | ||
same for proposed | |||
and predicate | |||
devices. | |||
Feature | Proposed | ||
VERIFY® STEAM | |||
Integrating Indicator | K012195 | ||
Dana Steam Sterilization | |||
Integrator | Comparison | ||
Comparison | |||
of integrator | |||
stated values | |||
at biological | |||
indicator | |||
growth- | |||
negative | |||
cycle | |||
conditions | Integrator does not reach | ||
endpoint before the biological | |||
indicator is inactivated. | Integrator does not reach | ||
endpoint before the biological | |||
indicator is inactivated | Same performance | ||
observed. | |||
Shelf-life | 5 years | 5 years | Proposed and |
predicate devices | |||
have the identical | |||
design and therefore | |||
carry the same shelf- | |||
life claims. | |||
Standard / | |||
Guidance | Conforms to the Guidance for | ||
Industry and FDA Staff: | |||
Premarket Notification | |||
[510(k)] Submissions for | |||
Chemical Indicators for | |||
integrating indicators and | |||
ANSI/AAMI/ISO 11140- | |||
1:2014: Sterilization of Health | |||
Care Products - Chemical | |||
Indicators - Part 1: General | |||
Requirements. | Conforms to ST60: | ||
Sterilization of Health Care | |||
Products - Chemical Indicators |
- Part 1: General
Requirements. | Both proposed and
predicate devices
meet the applicable
standards at the time
of their introduction
into interstate
commerce. |
9
K152630 Response to 5/12/16 Interactive Review VERIFY® STEAM Integrating Indicator
7. Performance Testing
Performance testing was conducted to verify that the proposed VERIFY® STEAM Integrating Indicator meets the requirements for integrating indicators in accordance with the Guidance for Industry and FDA Staff: Premarket Notification [510(k)] Submissions for Chemical Indicators for integrating indicators as well as ANSI/AAMI/ISO 11140-1:2014.
Table 5-2 summarizes the verification activities that were performed, with their respective acceptance criteria and results, to demonstrate that the proposed VERIFY® STEAM Integrating Indicator is as safe and is as effective. These studies confirm that the proposed device's performance meets the requirements of its predefined acceptance criteria and intended uses, and qualify the proposed device for use in the claimed steam sterilization cycles.
10
Stated Values
The subject device (integrator indicator) testing was performed in a bier vessel using the following stated values shown below at the following temperatures (see Table 5-2). The integrator indicator showed a passing result when the chemicals in the subject device reacted to all critical parameters in the bier vessel.
Test of 3 Lots | Acceptance Criteria | Study Result |
---|---|---|
BIER Vessel Testing | Pass result at SV for each temperature claimed | |
Stated values are: | ||
• 23.0 min at 121°C | ||
• 2.0 min at 132 °C | ||
• 1.2 min at 135 °C | ||
Fail result at 15% below SV for each | ||
temperature claimed | ||
Fail result at 1°C below SV for each | ||
temperature claimed | PASS | |
Simulated Use Testing in | ||
Claimed Sterilization | ||
Cycles | 100% Pass result under pass conditions | |
100% Fail result under fail conditions | PASS | |
Dry Heat Testing | Fail result when exposed to 140°C dry heat | |
for 30 min | PASS | |
Verification of integrator | ||
stated values at biological | ||
indicator growth-negative | ||
cycle conditions | Integrator does not reach endpoint before the | |
biological indicator is inactivated | PASS | |
Offset/ Bleeding | Integrator shall not bleed or offset | PASS |
Table 5-2. Verification Results Summary
The results of the VERIFY® STEAM Integrating Indicator performance testing demonstrate that both perform as intended in the claimed steam sterilization cycles and the proposed device is substantially equivalent to the predicate device.
8. Conclusion
The proposed device design is identical to that of predicate device, with the exception of the relabeling under the VERIFY brand. The intended use of the proposed device adds additional cycles and clarification of use conditions. The VERIFY® STEAM Integrating Indicator is as safe and is as effective when used to monitor the claimed steam sterilization cycles and is substantially equivalent to the predicate device.