K Number
K152624
Manufacturer
Date Cleared
2015-12-07

(84 days)

Product Code
Regulation Number
878.4452
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Intended Use:

The XSTAT 30 is intended for the control of bleeding from wounds in the groin or axilla that are not amenable to tourniquet application in adults and adolescents.

Indications for Use:

XSTAT 30 is a hemostatic device for the control of severe, life-threatening bleeding from junctional wounds in the groin or axilla not amenable to tourniquet application in adults and adolescents.

XSTAT 30 is a temporary device for use up to four (4) hours until surgical care is acquired. It should only be used for patients at high risk for immediate life-threatening bleeding from, hemodynamically significant (Advanced Trauma Life Support class 3 or 4 hemorrhagic shock), non- compressible junctional wounds, and when definitive care at an emergency care facility cannot be achieved within minutes.

XSTAT 30 is NOT indicated for use in: the thorax; the pleural cavity; the mediastinum; the abdomen; the retroperitoneal space; the sacral space above the inguinal ligament; or tissues above the clavicle.

Device Description

The XSTAT 30 dressing is composed of a standard, regenerated cellulose medical sponge that is compressed and formed into a group of approximately 92 mini-sponges. Each minisponge has a height of 4-5 mm and a circular surface diameter of 9.8 mm. The compressed regenerated cellulose sponge is coated with chitosan. Upon contact with blood, the mini-sponges absorb blood and, if unencumbered, are capable of expanding to a pre-compressed height of 40-50 mm within approximately 20 seconds. A radiopaque marker is embedded into a circular surface of the mini-sponges to render each sponge detectable via X-ray. The XSTAT 30 dressing includes an applicator that facilitates delivery of the mini-sponges to external bleeding wounds. Three applicators filled with the XSTAT 30 dressing (i.e., mini-sponges) are packaged in a sealed foil pouch and terminally sterilized by gamma radiation to a sterility assurance level of 10°.

For the treatment of severe bleeding from pelvis or shoulder wounds not amenable to tourniquet application, the XSTAT 30 sponges are applied to the wound using the applicator. Once applied to the wound, the XSTAT 30 sponges absorb blood and expand, thereby packing the wound. All mini-sponges must be removed from wounds before surgical repair and closure of the wounds. Following removal of the mini-sponges and definitive surgical repair of the wound, a radiograph is required prior to wound closure to confirm that every mini-sponge has been removed.

AI/ML Overview

The provided document is a 510(k) summary for the XSTAT 30 device and does not contain detailed information about a study with acceptance criteria and reported device performance in the way typically found for AI/ML device evaluations. This document declares substantial equivalence to a predicate device (XSTAT, K130218) primarily based on identical technological characteristics and similar intended use, with a change to indications for use to include the civilian population.

Therefore, many of the requested points, particularly those related to a direct study proving device performance against acceptance criteria for an AI/ML device, ground truth establishment, sample sizes for training/test sets, expert adjudication, or MRMC studies, are not present in this regulatory submission.

However, I can extract the relevant information regarding performance data and substantial equivalence as described in the document.

1. A table of acceptance criteria and the reported device performance

The document states: "The technological characteristics for XSTAT 30 are identical to the predicate device. Thus, the performance testing of the XSTAT 30 has been demonstrated by the clearance of the XSTAT predicate device. The testing provided in the cleared XSTAT notification (K130218) is incorporated herein by reference."

This means that the device (XSTAT 30) is considered to meet acceptance criteria by demonstrating substantial equivalence to its predicate device (XSTAT). The performance criteria mentioned relate to the inherent properties and safety of the device rather than a measurable diagnostic accuracy outcome.

Acceptance Criteria CategoryReported Device Performance (XSTAT 30 vs. XSTAT predicate)
Intended UseSame
Technological CharacteristicsIdentical (Mini-sponges, Applicator, Casualty Card, Packaging)
DimensionsIdentical
WeightIdentical
Safety FeaturesIdentical (Radiopaque marker)
BiocompatibilityIdentical testing performed and passed (Cytotoxicity, Sensitization, Irritation, Acute systemic toxicity, Hemocompatibility per ISO 10993 standards)
SterilizationIdentical (Gamma radiation sterilization)
Indications for Use (Key difference)XSTAT 30 expands to include "patients at high risk for immediate life-threatening, hemodynamically significant (Advanced Trauma Life Support class 3 or 4 hemorrhagic shock), non-compressible junctional wounds, and when definitive care at an emergency care facility cannot be achieved within minutes." (Predicate was more restricted to battlefield only for user population).
User Population (Key difference)Civilian and battlefield patients (Predicate was battlefield patients only)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the document as it references performance testing from the predicate device (K130218) and does not detail a new clinical study. The performance testing mentioned refers to bench and biocompatibility tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable/provided. The performance evaluation is based on device characteristics and biocompatibility, not expert-adjudicated diagnostic results.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable/provided.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable/provided. The device is a hemostatic sponge, not an AI/ML diagnostic or assistive tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable/provided.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the biocompatibility and material properties, the "ground truth" would be established by the results of standardized tests (e.g., ISO 10993 series) showing lack of toxicity, irritation, etc. For the expansion capabilities, the ground truth would be direct measurement of expansion time and volume. This is not "ground truth" in the diagnostic sense.

8. The sample size for the training set

This information is not applicable/provided. There is no "training set" in the context of this device's evaluation.

9. How the ground truth for the training set was established

This information is not applicable/provided.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of three faces in profile, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 7, 2015

RevMedx Incorporated John J. Smith, M.D., J.D. Hogan Lovells US LLP 555 Thirteenth Street, Northwest Washington, DC 20004

Re: K152624

Trade/Device Name: XSTAT 30 Regulation Number: 21 CFR 878.4452 Regulation Name: Nonabsorbable expandable hemostatic sponge for temporary internal use Regulatory Class: Class II Product Code: PGZ Dated: September 14, 2015 Received: September 14, 2015

Dear Dr. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Binita S. Ashar -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known)

Device Name

XSTAT 30

Indications for Use (Describe)

XSTAT 30 is a hemostatic device for the control of severe, life-threatening bleeding from wounds in the groin or axilla that are not amenable to tourniquet application in adults and adolescents.

XSTAT 30 is a temporary device for use up to four (4) hours until surgical care is acquired. It should only be used for patients at high risk for immediate life-threatening, hemodynamically significant (Advanced Trauma Life Support class 3 or 4 hemorrhagic shock), non-compressible junctional wounds, and when definitive care at an emergency care facility cannot be achieved within minutes.

XSTAT 30 is NOT indicated for use in the thorax; the pleural cavity; the mediastinum; the abdomen; the retroperitoneal space, the sacral space above the inguinal ligament; or tissues above the clavicle.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY RevMedx, Inc. XSTAT 30

Manufacturer Information:

RevMedx, Inc. 25999 SW Canvon Creek Road. Suite C Wilsonville, OR 97070 Phone: 503-218-2172 Facsimile: 503-218-2274

Contact Person: Amy K. Pointer, RA/QA Manager

Date Prepared: September 14, 2015

Trade/Proprietary Name:

XSTAT 30

Classification Name:

Non-Absorbable, Expandable, Hemostatic Sponge for Temporary Internal Use

Product Classification & Code:

21 CFR 878.4452; PGZ

Predicate Devices:

XSTAT (DEN130016; K130218)

Intended Use / Indications for Use:

Intended Use:

The XSTAT 30 is intended for the control of bleeding from wounds in the groin or axilla that are not amenable to tourniquet application in adults and adolescents.

Indications for Use:

XSTAT 30 is a hemostatic device for the control of severe, life-threatening bleeding from junctional wounds in the groin or axilla not amenable to tourniquet application in adults and adolescents.

XSTAT 30 is a temporary device for use up to four (4) hours until surgical care is acquired. It should only be used for patients at high risk for immediate life-threatening bleeding from, hemodynamically significant (Advanced Trauma Life Support class 3 or 4 hemorrhagic shock), non- compressible junctional wounds, and when definitive care at an emergency care facility cannot be achieved within minutes.

XSTAT 30 is NOT indicated for use in: the thorax; the pleural cavity; the mediastinum; the abdomen; the retroperitoneal space; the sacral space above the inguinal ligament; or tissues above the clavicle.

{4}------------------------------------------------

Technological Characteristics

The XSTAT 30 dressing is composed of a standard, regenerated cellulose medical sponge that is compressed and formed into a group of approximately 92 mini-sponges. Each minisponge has a height of 4-5 mm and a circular surface diameter of 9.8 mm. The compressed regenerated cellulose sponge is coated with chitosan. Upon contact with blood, the mini-sponges absorb blood and, if unencumbered, are capable of expanding to a pre-compressed height of 40-50 mm within approximately 20 seconds. A radiopaque marker is embedded into a circular surface of the mini-sponges to render each sponge detectable via X-ray. The XSTAT 30 dressing includes an applicator that facilitates delivery of the mini-sponges to external bleeding wounds. Three applicators filled with the XSTAT 30 dressing (i.e., mini-sponges) are packaged in a sealed foil pouch and terminally sterilized by gamma radiation to a sterility assurance level of 10°.

For the treatment of severe bleeding from pelvis or shoulder wounds not amenable to tourniquet application, the XSTAT 30 sponges are applied to the wound using the applicator. Once applied to the wound, the XSTAT 30 sponges absorb blood and expand, thereby packing the wound. All mini-sponges must be removed from wounds before surgical repair and closure of the wounds. Following removal of the mini-sponges and definitive surgical repair of the wound, a radiograph is required prior to wound closure to confirm that every mini-sponge has been removed.

Performance Data

The technological characteristics for XSTAT 30 are identical to the predicate device. Thus, the performance testing of the XSTAT 30 has been demonstrated by the clearance of the XSTAT predicate device. The testing provided in the cleared XSTAT notification (K130218) is incorporated herein by reference.

Substantial Equivalence

XSTAT 30 and XSTAT have the same intended use, same technological characteristics, same principles of operation and similar indications. The difference between the devices is a change to the indications for use. The X-Stat devices have a specific indication for use to control hemorrhage from non-compressible junctional wounds. This type of wound is equally life threatening when sustained in either a combat environment or a civilian environment. where there may be a delay to expeditious surgical intervention. When available at the point of injury. X Stat 30 can help minimize blood loss and provide a bridge to surgery. Therefore, XSTAT 30 is substantially equivalent to XSTAT.

Conclusions

The XSTAT 30 is as safe and effective as the predicate XSTAT device. The intended use is the same for both XSTAT 30 and XSTAT, as are the technological characteristics and principles of operation. The Substantial Equivalence Summary table below details and compares the XSTAT 30 to the XSTAT.

{5}------------------------------------------------

of33
----------------
XSTAT 30XSTAT (K130218)
Intended UseSameSame
Indications for UseXSTAT 30 is a hemostatic devicefor the control of severe, life-threatening bleeding fromjunctional wounds in the groin oraxilla not amenable to tourniquetapplication in adults andadolescents.XSTAT is a hemostatic device forthe control of bleeding fromjunctional wounds in the groin oraxilla not amenable to tourniquetapplication in adults andadolescents.
XSTAT 30 is a temporary devicefor use up to four (4) hours untilsurgical care is acquired. It shouldonly be used for patients at highrisk for immediate life-threateningbleeding from, hemodynamicallysignificant (Advanced Trauma LifeSupport class 3 or 4 hemorrhagicshock), non-compressiblejunctional wounds, and whendefinitive care at an emergencycare facility cannot be achievedwithinminutes.XSTAT 30 is NOT indicated foruse in the thorax; the pleuralcavity; the mediastinum; theabdomen; the retroperitonealspace, the sacral space above theXSTAT is a temporary device foruse up to four (4) hours untilsurgical care is acquired. XSTATis intended for use in thebattlefield.XSTAT is NOT indicated for usein: the thorax; the pleural cavity;the mediastinum; the abdomen;the retroperitoneal space; thesacral space above the inguinalligament; or tissues above the
inguinal ligament; or tissues abovethe clavicle.clavicle.
User PopulationCivilian and battlefield patientsBattlefield patients
TechnologicalCharacteristics1. Mini-sponges2. Applicator3. Casualty Card4. Packaging1. Mini-sponges2. Applicator3. Casualty Card4. Packaging
Dimensions3-Pack: 254mm x 165mm x 38mm3-Pack: 254mm x 165mm x 38mm
(l x w x h)1-Pack: 254mm x 165mm x 38mm1-Pack: 254mm x 165mm x 38mm
Weight3- Pack: 0.25kg1-Pack: 0.1kg3- Pack: 0.25kg1-Pack: 0.1kg
Safety FeaturesRadiopaque markerRadiopaque marker
BiocompatibilityCytotoxicity (ISO 10993-5);Sensitization (ISO 10993-10);Irritation (ISO 10993-10);Acute systemic toxicity (ISO10993-11); andCytotoxicity (ISO 10993-5);Sensitization (ISO 10993-10);Irritation (ISO 10993-10);Acute systemic toxicity (ISO10993-11); and
Hemocompatibility (ISO 10993-4).Hemocompatibility (ISO 10993-4).
SterilizationGamma radiation sterilizationGamma radiation sterilization

Substantial Equivalence Summary Table

§ 878.4452 Nonabsorbable expandable hemostatic sponge for temporary internal use.

(a)
Identification. A nonabsorbable expandable hemostatic sponge for temporary internal use is a prescription device intended to be placed temporarily into junctional, non-compressible wounds, which are not amenable to tourniquet use, to control bleeding until surgical care is acquired. The sponges expand upon contact with blood to fill the wound cavity and provide a physical barrier and pressure that facilitates formation of a clot. The device consists of sterile, nonabsorbable radiopaque compressed sponges and may include an applicator to facilitate delivery into a wound.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Performance data must demonstrate the biocompatibility of patient-contacting components.
(2) Performance data must demonstrate the sterility of patient-contacting components including endotoxin and pyrogenicity assessments.
(3) Performance data must support device stability by demonstrating continued sterility of the patient-contacting components of the device, package integrity, and device functionality over the requested shelf life.
(4) Assessment of material characteristics must be sufficient to support safety under anticipated conditions of use. Assessments must include the following:
(i) Material specifications.
(ii) Immunogenicity.
(iii) Viral inactivation for animal-derived materials.
(5) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Absorption capacity.
(ii) Extent of swelling.
(iii) Mechanical properties.
(iv) Expansion force/pressure.
(v) Radiopacity.
(vi) Deployment/applicator functionality.
(6) In vivo performance data must demonstrate safe and effective use by verifying that the device performs as intended under anticipated conditions of use. Appropriate analysis/testing must demonstrate that the product: Controls bleeding, does not promote adverse local or systemic effects, and can be completely removed from the wound. The following performance characteristics must be tested:
(i) Deployment.
(ii) Control of bleeding.
(iii) Radiopacity.
(iv) Retrieval.
(v) Assessment of local and systemic effects.
(7) Human factors testing and analysis must validate that the device design and labeling are sufficient for appropriate use by emergency responders deploying the device as well as surgeons retrieving the device from wounds.
(8) Labeling must include:
(i) Specific instructions for deployment by emergency responders and retrieval by surgeons.
(ii) Warnings, cautions, and limitations needed for safe use of the device.
(iii) Information on how the device operates and the typical course of treatment.
(iv) A detailed summary of the in vivo and human factors testing pertinent to use of the device.
(v) Appropriate imaging information to ensure complete retrieval of device.
(vi) An expiration date/shelf life.