(222 days)
The Pulse Oximeter is intended for continuous monitoring, spot-checking of functional pulse oxygen saturation (SpO2) and pulse rate (PR) of single adult, adolescent, child and infant patients in hospitals and clinics.
The Pulse Oximeter MD300M/MD300K2 can display the SpO2%, pulse rate, Pulse Amplitude Index and other indication parameters, such as time, ID number, pulse amplitude bar and battery power status, alarm limits and the connections of sensors. It is used for adult, adolescent, child and infant patients. The device has physiological alarm and technology alarm function. There are two-level alarm priorities in oximeter. High priority: "Di-Di-Di-----------Di-Di" indicates the patient is in the very dangerous situation. Low priority: "Di" indicates the technical alarm caused by the device itself. The device also has visual alarm function to indicate users by lamp and information on the device. The power supply of the applicant device is 3 AA alkaline batteries, rechargeable batteries or adapter.
The proposed device consists of photo detector, display screen, signal amplify unit, CPU, display unit and power supply unit.
Principle of the oximeter is as follows: A mathematic formula is established making use of Lambert Beer Law according to Spectrum Absorption Characteristics of Reductive hemoglobin (RHb) and Oxyhemoglobin (HbO2) in red and near-infrared zones. Operation principle of the instrument: Photoeletric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacity Pulse Scanning and Recording Technology, so that two beams of different wavelength of lights (660nm red and 905nm near infrared light) can be focused on a human nail tip through a clamping finger-type sensor. A measured signal obtained by a photosensitive element, will be shown on the oximeter's display through process in electronic circuits and microprocessor.
The proposed device is not for life-supporting or life-sustaining, not for implant.
The device is not sterile and the transducers are reusable and do not need sterilization and re-sterilization.
The device is for prescription.
The device does not contain drug or biological products.
The device is software-driven and the software validation is provided in software.
Here's an analysis of the acceptance criteria and study details for the Pulse Oximeter (MD300M, MD300K2) based on the provided FDA 510(k) summary:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for SpO2 and PR accuracy are presented as part of the device's technical specifications and were evaluated through a clinical study.
| Parameter | Acceptance Criteria (Target/Claimed) | Reported Device Performance (MD300M with M-50E012CS09/M-50J033CS045 probes) | Reported Device Performance (MD300M with M-50B008CS09 probe) |
|---|---|---|---|
| SpO2 Accuracy (70-100%) | $\pm$ 2% | Arms of 1.75% | Arms of 1.48% |
| PR Accuracy | 30bpm | (Implicitly met as accuracy results provided for SpO2) | (Implicitly met as accuracy results provided for SpO2) |
Note: The document explicitly states the SpO2 accuracy (reported as Arms) as a result of the clinical study, which implicitly confirms it meets the claimed $\pm$ 2% accuracy. PR accuracy is listed as a specification but a specific Arms value from the clinical study is not provided in the summary, though it's implied to have been met. The relevant standard (ISO 80601-2-61) generally requires Arms to be less than or equal to the claimed accuracy.
2. Sample Size and Data Provenance for the Test Set
- Sample Size: 12 healthy adult volunteer subjects.
- Data Provenance: The study was conducted in Yue Bei People's Hospital (China). It was a prospective study.
- Subject Demographics: 6 females and 6 males, ages 21-43 years, 47-82 kg, 155-185 cm. The subject pool included diverse pigmentation: 6 with light pigmentation from Asian, 3 with light (white) pigmentation from Caucasian, and 3 with dark pigmentation from African.
3. Number of Experts and Qualifications for Ground Truth
The document does not explicitly state the number or qualifications of experts used to establish the ground truth. For pulse oximetry clinical studies, the ground truth for SpO2 is typically established using a co-oximeter on arterial blood samples, which is a gold standard measurement, not usually based on expert consensus of visual or imaging data.
4. Adjudication Method
Not applicable. This is not a study involving expert readers or image interpretation requiring adjudication. The ground truth for SpO2 accuracy is based on objective laboratory measurements (co-oximetry).
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for imaging devices or AI-assisted diagnostic tools where human interpretation is a primary component. The Pulse Oximeter is a physiological monitoring device.
6. Standalone Performance Study
Yes, a standalone study was performed. The clinical test described evaluates the accuracy of the device itself (algorithm only, as there’s no human-in-the-loop interpretation) by comparing its SpO2 readings to arterial blood gas measurements (co-oximetry), which serve as the ground truth.
7. Type of Ground Truth Used
The ground truth for the clinical accuracy study was established using arterial blood co-oximetry, which is considered the gold standard for measuring arterial oxygen saturation (SaO2) in pulse oximeter validation studies.
8. Sample Size for the Training Set
The document does not specify a separate training set. For pulse oximeters, the core algorithm for SpO2 calculation is based on established physiological principles and spectrophotometry (Lambert Beer Law) rather than a deep learning model requiring a large training dataset in the same way an AI diagnostic algorithm would. While manufacturers develop and refine their algorithms, this development process is not typically detailed as a "training set" in a 510(k) in the way AI/ML algorithms are. The "training" in this context would be more analogous to calibration and algorithm refinement based on engineering principles and potentially internal validation data.
9. How Ground Truth for the Training Set Was Established
As noted above, a distinct "training set" with ground truth in the AI/ML sense is not described. The fundamental principle (Lambert Beer Law) and device design are based on established science. Any internal calibration or algorithm development would use similar ground truth methodologies (e.g., co-oximetry data from desaturation studies) as the validation study to ensure accuracy across the measurement range, but these details are not provided as a "training set ground truth" in this summary.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three heads, representing the department's commitment to health, human services, and science. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 18, 2016
Beijing Choice Electronic Technology Co., Ltd. Lei Chen Quality Director North Building 3F, No.9 Shuangyuan Road Badachu Hi-tech Zone, Shijingshan District Beijing, China 100041
Re: K152563
Trade/Device Name: Pulse Oximeter (MD300M, MD300K2) Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: March 7, 2016 Received: March 11, 2016
Dear Mr. Lei Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K152563
Device Name Pulse Oximeter
Indications for Use (Describe)
The Pulse Oximeter is intended for continuous monitoring, spot-checking of functional pulse oxygen saturation (SpO2) and pulse rate (PR) of single adult, adolescent, child and infant patients in hospitals and clinics.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Section III 510(k) Summary
This summary of 510(k) is being submitted in accordance with the requirements of 21 CFR 807.92.
There is no prior submission for the device.
3.1Submitter Information
● Manufacturer Name:
Establishment Registration Number: 3005569927 Beijing Choice Electronic Technology Co., Ltd. Room 4104,No.A12 Yuquan Road Haidian District 100143 Beijing, P.R.China
. Contact Person:
Mr. Lei Chen Beijing Choice Electronic Technology Co., Ltd. North Building 3F, No. 9 Shuangyuan Road, Badachu Hi-tech Zone, Shijingshan District Beijing China 100041 Phone: +86-10-88798300 Ext 6020 Fax:215-4052545 Email: cc@choicemmed.com
- Date prepared: March 17, 2016
3.2Proposed Device Information
Device Common Name: Pulse Oximeter Device Trade/Proprietary Name: Pulse Oximeter Model: MD300M, MD300K2 Classification Name: Oximeter Regulation Number: 870.2700 Product Code: DQA
Class: II
Panel: Anesthesiology
3.3Predicate Device
510(k) Number: K090599
{4}------------------------------------------------
Premarket Notification 510(k) Submission-Section III 510(k) Summary Common Name: Pulse Oximeter Device Trade/Proprietary Name: Pulse Oximeter Model: MD300K1 Classification Name: Oximeter Product Code: DQA Regulation Number: 870.2700
Device Class:
Panel: Anesthesiology
Manufacturer: Beijing Choice Electronic Technology Co., Ltd.
Intended Use: MD300K1 Pulse Oximeter is intended for continuous monitoring, spot-checking of functional pulse oxygen saturation (SpO2) and pulse rate (PR) of single adult and pediatric patients in hospitals and home care.
3.4Device Description
The Pulse Oximeter MD300M/MD300K2 can display the SpO2%, pulse rate, Pulse Amplitude Index and other indication parameters, such as time, ID number, pulse amplitude bar and battery power status, alarm limits and the connections of sensors. It is used for adult, adolescent, child and infant patients. The device has physiological alarm and technology alarm function. There are two-level alarm priorities in oximeter. High priority: "Di-Di-Di-----------Di-Di" indicates the patient is in the very dangerous situation. Low priority: "Di" indicates the technical alarm caused by the device itself. The device also has visual alarm function to indicate users by lamp and information on the device. The power supply of the applicant device is 3 AA alkaline batteries, rechargeable batteries or adapter.
The proposed device consists of photo detector, display screen, signal amplify unit, CPU, display unit and power supply unit.
Principle of the oximeter is as follows: A mathematic formula is established making use of Lambert Beer Law according to Spectrum Absorption Characteristics of Reductive hemoglobin (RHb) and Oxyhemoglobin (HbO2) in red and near-infrared zones. Operation principle of the instrument: Photoeletric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacity Pulse Scanning and Recording Technology, so that two beams of different wavelength of lights (660nm red and 905nm near infrared light) can be focused on a human nail tip through a clamping finger-type sensor. A measured signal obtained by a photosensitive element, will be shown on the oximeter's display through process in electronic circuits and microprocessor.
The proposed device is not for life-supporting or life-sustaining, not for implant.
The device is not sterile and the transducers are reusable and do not need sterilization and re-sterilization.
The device is for prescription.
{5}------------------------------------------------
The device does not contain drug or biological products.
The device is software-driven and the software validation is provided in software.
{6}------------------------------------------------
3.5 Comparison list of the technological characteristics
| Comparison Elements | Proposed Device | Predicate Device | ||
|---|---|---|---|---|
| Product Name | Pulse Oximeter | Pulse Oximeter | ||
| Model | MD300M | MD300K1 | ||
| Regulation No. | 21 CFR 870.2700 | 21 CFR 870.2700 | ||
| Classification | II | II | ||
| Classification Name | Oximeter | Oximeter | ||
| Product Code | DQA | DQA | ||
| Indications for Use | The MD300M pulse oximeter is intended for continuous monitoring, spot-checking of functional pulse oxygen saturation (SpO2) and pulse rate (PR) of single adult, adolescent, child and infant patients in hospitals and clinics. | MD300K1 Pulse Oximeter is intended for continuous monitoring, spot-checking of functional pulse oxygen saturation (SpO2) and pulse rate (PR) of single adult and pediatric patients in hospitals and home care. | ||
| Comparison Statement | The proposed device and the predicated device have the similar intended use and classification. The only difference is that the proposed device cannot be used in the home care environment while the predicated device can be used in the home care environment. | |||
| Components | Photo detector, display screen, signal amplify unit, CPU, display unit and power unit. | Photo detector, display screen, signal amplify unit, CPU, display unit and power unit. | ||
| Design Principle | Principle of the oximeter is as follows: A mathematic formula is established making use | The MD300K1 Pulse Oximeter works by applying a sensor to a pulsating arteriolan | ||
| of Lambert Beer Law according to Spectrum | vascular bed. The sensor contains a dual light | |||
| CharacteristicsAbsorptionofReductive | source and photo detector. The one wavelength | |||
| (RHb) andOxyhemoglobinhemoglobin | of light source is 660nm, which is red light; the | |||
| red and near-infrared zones.(HbO2) in | other is 940nm, which is ultra red light. Skin, | |||
| principleOperationoftheinstrument: | bone, tissue, and venous normally absorb a | |||
| OxyhemoglobinPhotoeletricInspection | constant amount of light during systole and | |||
| Technology is adopted in accordance with | diastole, as blood volume increases and | |||
| Capacity Pulse Scanning and Recording | decreases. The ratio of light absorbed at systole | |||
| Technology, so that two beams of different | and diastole is translated into an oxygen | |||
| wavelength of lights (660nm red and 905nm | saturation measurement. This measurement is | |||
| near infrared light) can be focused on a human | referred to as SpO2. | |||
| nail tip through a clamping finger-type sensor. | ||||
| signalmeasuredobtainedAbyದ | ||||
| photosensitive element, will be shown on the | ||||
| display through processoximeter'sin | ||||
| electronic circuits and microprocessor. | ||||
| Measurement | Red | 660±3nm | 660 nanometers | |
| Wavelength | Infrared | 905 ± 8nm | 940 nanometers | |
| Comparison Statement | The proposed device and the predicate device have the same design principle and similarmeasurement wavelength. | |||
| Display Type | TFT (colorized) | OLED | ||
| Battery | 3*AA alkaline batteries, rechargeable | 2*AA alkaline batteries | ||
| batteries or adapter | ||||
| Power adapter | 5V 2A | Not available | ||
| SpO2 display range | 0%~100% | 0~100% | ||
| SpO2 measurementrange | 0~100% | 0~100% | ||
| SpO2 accuracy | 70% | 70% | ||
| PR display range | 0~250bpm | 0~235bpm | ||
| PR measurementrange | 30~250bpm | 30~235bpm | ||
| PR accuracy | 30bpm | 30bpm | ||
| PR Resolution | 1% | 1% | ||
| Pulse amplitudeindex measurementrange | 0.1%~20% | --- | ||
| Operatingtemperature | 0°C~40°C | 5°C~40°C | ||
| Relative humidity | $ \le $ 80% no condensation in operation;$ \le $ 93% no condensation in storage | $ \le $ 80% no condensation in operation;$ \le $ 93% no condensation in storage | ||
| Atmosphere pressure | 86kPa~106kpa | 86kPa~106kpa | ||
| Alarm type | Audio type, visual alarm and information | Audio type, visual alarm and information | ||
| Alarm limit range | SpO2 70% | SpO2 70% | ||
| DefaultLimit | SpO2 | High 100%, Low 90% | High 99%, Low 90% | |
| PR | High 100bpm, Low 60bpm | High 100bpm, Low 60bpm | ||
| Comparison Statement | The proposed device has similar product specification as predicate device. | |||
| ContactingMaterial | Battery Cover | ABS | ABS | |
| Plastic Case Cover | ||||
| Fingertip Cushion | Medical Silicone Gel (Finger probe M-50E012CS09, M-50B008CS09) | Medical Silicone Gel (Finger probe M-50 series) | ||
| Fingertip Cushion | Microfoam (Finger probe M-50J033CS045) | N/A | ||
| Comparison Statement | The contacting materials of the proposed device are similar to those of the predicate device. The only difference is that the material of MD300M supporting finger probe M-50J033CS045 is microfoam while which is not available to the predicate device. | |||
| Performance Testing | Laboratory Testing | The laboratory tests include SpO2 and PR accuracy Test, Weak Perfusion Test, High and Low Temperature and Humidity Test, Performance Test After Cleaning and ISO80601-2-61 | Meet the requirements of FDA Guidance | |
| Clinical Testing | Conformed to ISO9919&ISO 80601-2-61 | Conformed to ISO9919 | ||
| Clinical Test for the device accuracy is conducted in Yue Bei people's Hospital. Theclinical test report and protocol are providedin Performance Testing-Clinical Test Report | ||||
| Electrical | Electrical Safety | Conformed to IEC60601-1, IEC 60601-1-8 | Conformed to IEC60601-1 | |
| Electromagnetic Compatibility | Conformed to IEC60601-1-2 | Conformed to IEC60601-1-2 | ||
| EMCandSafety | Software | Moderate level of concern | ||
| Compliance with FDA Guidance for thecontent of Premarket Submissions forSoftware Contained in Medical Devices | ||||
| Risk Management in Compliance withISO14971:2007 | ||||
| Biocompati | Microfoam | In Vitro Cytotoxicity | No cytotoxic potential | N/A |
| Animal skin irritation test | No evidence of significant irritation from the test extract to rabbits | |||
| Skin Sensitization Test | No evidence of sensitization was observed. | |||
| Comparison Statement | Compliance with ISO10993 | |||
| Label and Labeling | Compliance with the Guidance of pulse oximeter-premarket notification issued on March 4,2013 | Compliance with FDA guidance |
Table 3-1 Performance Specification Comparison Table between the Proposed Device (MD300M) and Predicate Device
{7}------------------------------------------------
{8}------------------------------------------------
Performance Specification
{9}------------------------------------------------
Premarket Notification 510(k) Submission—Section III 510(k) Summary
{10}------------------------------------------------
{11}------------------------------------------------
Table 3-2 Performance Specification Comparison Table between the Proposed Device (MD300K2) and Predicate Device
| Comparison Elements | Proposed Device | Predicate Device |
|---|---|---|
| Product Name | Pulse Oximeter | Pulse Oximeter |
| Model | MD300K2 | MD300K1 |
| Regulation No. | 21 CFR 870.2700 | 21 CFR 870.2700 |
| Classification | II | II |
| Classification Name | Oximeter | Oximeter |
| Product Code | DQA | DQA |
| Indications for Use | The MD300K2 pulse oximeter is intended forcontinuous monitoring, spot-checking offunctional pulse oxygen saturation (SpO2) andpulse rate (PR) of single adult, adolescent,child and infant patients in hospitals andclinics. | MD300K1 Pulse Oximeter is intended forcontinuous monitoring, spot-checking offunctional pulse oxygen saturation (SpO2)and pulse rate (PR) of single adult andpediatric patients in hospitals and home care. |
{12}------------------------------------------------
| Comparison Statement | The proposed device and the predicated device have the similar intended use andclassification. The only difference is that the proposed device cannot be used in the homecare environment while the predicated device can be used in the home care environment. | |||
|---|---|---|---|---|
| Components | Photo detector, display screen, signal amplifyunit, CPU, display unit and power unit. | Photo detector, display screen, signal amplifyunit, CPU, display unit and power unit. | ||
| Design Principle | Principle of the oximeter is as follows: Amathematic formula is established making useof Lambert Beer Law according to SpectrumAbsorption Characteristics of Reductivehemoglobin (RHb) and Oxyhemoglobin(HbO2) in red and near-infrared zones.Operation principle of the instrument:Photoeletric Oxyhemoglobin InspectionTechnology is adopted in accordance withCapacity Pulse Scanning and RecordingTechnology, so that two beams of differentwavelength of lights (660nm red and 905nmnear infrared light) can be focused on a humannail tip through a clamping finger-type sensor.A measured signal obtained by a photosensitiveelement, will be shown on the oximeter'sdisplay through process in electronic circuitsand microprocessor. | The MD300K1 Pulse Oximeter works byapplying a sensor to a pulsating arteriolarvascular bed. The sensor contains a dual lightsource and photo detector. The onewavelength of light source is 660nm, whichis red light; the other is 940nm, which is ultrared light. Skin, bone, tissue, and venousnormally absorb a constant amount of lightduring systole and diastole, as blood volumeincreases and decreases. The ratio of lightabsorbed at systole and diastole is translatedinto an oxygen saturation measurement. Thismeasurement is referred to as SpO2. | ||
| MeasurementWavelength | Red | 660±3nm | 660 nm | |
| Infrared | 905 ± 10nm | 940 nm |
{13}------------------------------------------------
| Comparison Statement | The proposed device and the predicate device have the same design principle and similar measurement wavelength. | ||
|---|---|---|---|
| Performance Specification | Display Type | TFT (colorized) | OLED |
| Battery | 3*AA alkaline batteries, rechargeable batteries or adapter | 2*AA alkaline batteries | |
| Power adapter | 5V 2A | Not available | |
| SpO2 Display Range | 0~100% | 0~100% | |
| SpO2 Measurement Range | 0~100% | 0~100% | |
| SpO2 Accuracy | 70% | 70% | |
| SpO2 resolution | 1% | 1% | |
| PR Display Range | 0~250bpm | 0~235bpm | |
| PR Measurement Range | 30bpm~250bpm | 30bpm~235bpm | |
| PR Accuracy | 30bpm | 30bpm | |
| PR Resolution | 1% | 1% | |
| Pulse amplitude index display range | 0.1%~20% | --- |
{14}------------------------------------------------
| OperatingTemperature | 0°C~40°C | 5°C~40°C | |||
|---|---|---|---|---|---|
| Relative Humidity | ≤80% no condensation in operation;≤93% no condensation in storage | ≤80% no condensation in operation;≤93% no condensation in storage | |||
| Atmosphere Pressure | 86kPa~106kpa | 86kPa~106kpa | |||
| Alarm type | Audio type, visual alarm and information | Audio type, visual alarm and information | |||
| Alarm limit range | SpO2 70% | SpO2 70% | |||
| DefaultLimits | SpO2 | High 100%, Low 95% | High 99%, Low 90% | ||
| PR | High 100bpm, Low 60bpm | High 100bpm, Low 60bpm | |||
| Comparison Statement | The proposed device has similar product specification as predicate device. | ||||
| ContactingMaterial | Battery CoverPlastic Case Cover | ABS | ABS | ||
| Fingertip Cushion | Medical Silicone Gel(Finger probeM-50E012CS09, M-50B008CS09) | Medical Silicone Gel(M-50 series) | |||
| Fingertip Cushion | Microfoam(Finger probe M-50J033CS045) | N/A | |||
| Comparison Statement | The contacting materials of the proposed device are similar to those of the predicatedevice. The only difference is that the material of MD300K2 supporting finger probeM-50J033CS045 is microfoam while which is not available to the predicate device. |
{15}------------------------------------------------
| PerformanceTesting | Laboratory Testing | The laboratory tests include SpO2 and PRaccuracy Test, Weak Perfusion Test, High andLow Temperature and Humidity Test,Performance Test After Cleaning andISO80601-2-61 | Meet the requirements of FDA Guidance | |
|---|---|---|---|---|
| EMC andElectricalSafe | Electrical Safety | Conformed to IEC60601-1, IEC 60601-1-8 | Conformed to IEC60601-1 | |
| ElectromagneticCompatibility | Conformed to IEC60601-1-2 | Conformed to IEC60601-1-2 | ||
| Software | Moderate level of concernCompliance with FDA Guidance for the contentof Premarket Submissions for SoftwareContained in Medical DevicesRisk Management in Compliance withISO14971:2007 | Moderate level of concernCompliance with FDA Guidance for thecontent of Premarket Submissions forSoftware Contained in Medical DevicesRisk Management in Compliance withISO14971:2007 | ||
| Biocompatibility | Microfoam | In Vitro Cytotoxicity | No cytotoxicpotential | N/A |
| Animal skin irritationtest | No evidence ofsignificant irritationfrom the test extractto rabbits |
{16}------------------------------------------------
| Skin Sensitization Test | No evidence ofsensitization wasobserved. | |
|---|---|---|
| Comparison Statement | Compliance with ISO10993 | |
| Label and Labeling | Compliance with the Guidance of pulseoximeter-premarket notification issued onMarch 4,2013 | Compliance with FDA guidance |
{17}------------------------------------------------
3.6Intended use
The MD300M/MD300K2 pulse oximeter is intended for continuous monitoring, spot-checking of functional pulse oxygen saturation (SpO2) and pulse rate (PR) of single adult, adolescent, child and infant patients in hospitals and clinics.
3.7 Functional and Safety Testing:
Non-Clinical Test
The Pulse Oximeter MD300M/MD300K2 is designed and will be manufactured in accordance with both mandatory and voluntary standards, including:
IEC 60601-1:2005/AC:2012 Medical electrical equipment-Part 1: General requirements for basic safety and essential performance
IEC 60601-1-2:2007 Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
IEC 60601-1-8:2012 Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems.
ISO 80601-2-61:2011 Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment.
The Software Validation is in compliance with FDA Guidance to Compliance with FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
The compatibility of the skin-contact component material in the finished product meets the requirement of Biocompatibility. The Biological Evaluation Tests are in compliance with the standards of ISO10993. "Biological Evaluation of Medical Devices".
| No. | Test Name |
|---|---|
| 1 | System Performance Test |
| 2 | Shelf Life Test |
| 3 | Performance Test after Cleaning |
| 4 | Performance Test according to ISO 80601-2-61 |
| 5 | Electromagnetic Compatibility Test According to IEC 60601-1-2 |
| 6 | Electrical Safety Test According to IEC 60601-1 |
The list of non-clinical test performed on the proposed device is shown as following:
{18}------------------------------------------------
| Premarket Notification 510(k) Submission-Section III 510(k) Summary | |
|---|---|
| --------------------------------------------------------------------- | -- |
| 7 | Alarm System Tests According to IEC60601-1-8 |
|---|---|
| 8 | Irritation,Sensitization and Cytotoxicity Test according to ISO 10993 |
Clinical Test
The Clinical Test was conducted following the testing described in clause 201.12.1 of ISO 80601-2-61:2011 Medical electrical equipment- Part 2-61 Particular requirements for basic safety and essential performance of pulse oximeter equipment.
The supporting Oximeter probes of Pulse Oximeter MD300M are adult finger probe M-50E012CS09, pediatric finger probe M-50B008CS09, single use probe M-50J033CS045. The clinical study of MD300M was conducted with its supporting finger probe respectively.
The Clinical Test of MD300M was conducted in Yue Bei people's Hospital. The clinical s study report was presented in Performance Testing-Clinical Test Report.
Subjects information:
12 healthy adult volunteer subjects (6 females and 6 males ages 21-43yr, 47-82kg, 155-185cm, among which 6 with light pigmentation from Asian, 3 with light (white) pigmentation from Caucasian, 3 with dark pigmentation from African) were included in the study connected Sep. 20-22, 2014 to evaluate the SpO2 accuracy performance of the MD300M Pulse Oximeter and its supporting M-50E012CS09/M-50B008CS09/ M-50J033CS045 Oximeter Probe.
Results:
The SpO2 accuracy performance results showed the MD300M Pulse Oximeter and its supporting M-50E012CS09 Oximeter probe to have an Arms of 1.75 during steady state conditions over the range of 70-100%.
The SpO2 accuracy performance results showed the MD300M Pulse Oximeter and its supporting M-50B008CS09 Oximeter probe to have an Arms of 1.48 during steady state conditions over the range of 70-100%.
The SpO2 accuracy performance results showed the MD300M Pulse Oximeter and its supporting M-50J033CS045 Oximeter probe to have an Arms of 1.75 during steady state conditions over the range of 70-100%.
Conclusion:
The results of the study provide supporting evidence that the pulse oximeter MD300M with its supporting finger probes was compliance to the accuracy specification claimed by the manufacturer.
The proposed device MD300K2 and MD300M have similar hardware and software design. They have same contain identical materials, electro-optical component, SpO2 module,
{19}------------------------------------------------
same supporting finger probes and have equivalent sensor characteristics. The main difference between them is the appearance design (enclosure). So we think MD300M clinical study results can be to support clearance of the proposed device of MD300K2.
The clinical statement was presented in Performance Testing-Clinical Statement.
3.8Determination of substantial equivalence
The proposed device of Pulse Oximeter MD300M/MD300K2 has the same classification information, similar intended use, same design principle, similar product design and specifications as the predicated device. The differences exist in display screen, SpO2 accuracy and PR display range. The main difference is that the proposed device cannot be used in the homecare environment while the predicated device can be used in the home care environment. The differences are slight and do not influence the effectiveness and safety of the device. According to the laboratory and clinical test results, the proposed device is as safe and as effective as the predicate device. So the proposed device is Substantially Equivalent (SE) to the predicate device which is US legally market device.
§ 870.2700 Oximeter.
(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).