K Number
K152310
Manufacturer
Date Cleared
2016-02-24

(194 days)

Product Code
Regulation Number
870.1280
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CardioFocus Deflectable Sheath is indicated for percutaneous catheter introduction into the vasculature and into the chambers of the heart. The sheath deflection facilitates catheter positioning.

Device Description

The HeartLight® Deflectable Sheath is a sterile, disposable device which provides a conduit for catheter introduction into the vasculature and into the chambers of the heart. The Deflectable Sheath is able to move its distal section in one direction in a single plane and is activated by a rotating knob at the front of the device handle. The deflection of the sheath and the torquing of the device allow the distal end of the sheath to be positioned at the ostium of pulmonary veins.

AI/ML Overview

The provided document describes the HeartLight Deflectable Sheath, a medical device, and its comparison to a predicate device for the purpose of a 510(k) premarket notification. The information pertains to engineering testing and biocompatibility rather than the performance of an AI/ML powered device impacting diagnostic accuracy. Therefore, many of the requested elements for an AI/ML study, such as sample size for test sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, and training set information, are not applicable in this context.

However, I can extract the acceptance criteria and performance data as presented for this device.

1. Table of Acceptance Criteria and Reported Device Performance

The document details performance and functional requirements testing for the HeartLight Deflectable Sheath, primarily through a comparison to the predicate device, FlexCath® Advance™ Steerable Sheath (K123591), and a reference device, HeartSpan® Steerable Introducer (K132164). The acceptance criteria are generally implied by stating "Same" or "Similar" to the predicate, with specific values provided where there are differences.

Acceptance Criteria / Element of ComparisonPredicate Device Performance (FlexCath® Advance™ Steerable Sheath (K123591))HeartLight® Deflectable Sheath (K152310) Performance
Intended Use/Indications for UseFacilitates introducing various cardiovascular catheters into the heart. The sheath is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart. The sheath deflection facilitates catheter positioning.Identical: Facilitates introducing various cardiovascular catheters into the heart. The sheath is indicated for percutaneous catheter introduction into the vasculature and into the chambers of the heart. The sheath deflection facilitates catheter positioning.
UsageSingle use and disposableSame (Single use and disposable)
Sterilization MethodSterilized with ethylene oxide.Same (Sterilized with ethylene oxide)
Shelf Life2 yearsSame (2 years)
Storage ConditionsStore in a cool, dark, dry place.Same (Store in a cool, dark, dry place.)
Steerable sheath improves access to hard-to-reach anatomyYesSame (Yes)
Radiopaque for visualization under fluoroscopy - Radiopaque tip marker for enhanced fluoro visibilityYesSame (Yes)
Uni-directional curve deflection (Angle)Yes ≥ 135°Similar ≥ 130°. Comment: Minimal difference, does not affect safety/effectiveness.
Hemostasis valve to prevent blood leakage and/or air ingressYesSame (Yes)
Ergonomic handle to provide comfortYesSame (Yes)
Soft, atraumatic tip to minimize likelihood of perforation/damageYesSame (Yes)
Braided shaft provides exceptional torquability, pushability, and kink resistanceYesSame (Yes)
Sideport for infusion and contrast injectionYesSame (Yes)
Usable Length65 cm75 cm. Comment: Different from predicate, provides greater accessibility. Bench testing supports design.
Inner Diameter12F12F - Same
Dilator to Guidewire CompatibilityAccepts up to 0.035" guidewireAccepts 0.035" or 0.038" guidewire. Comment: Similar, bench testing supports design.
Biocompatibility - Contact with circulating blood for up to 24 hoursYesSame nature and duration of bodily contact.
Shaft MaterialPebax with Barium Sulfate, stainless steel braid and PTFE linerSame: Stainless Steel Braid Reinforced Pebax with Barium Sulfate with Stainless Steel Pull-Wire, PTFE Lined.
Hemostasis valve materialSiliconeSame (Silicone)
Hemostasis hub and cap materialPolycarbonateIsoplast. Comment: Material difference noted, but not explicitly stated as non-conforming.
Hemostasis valvePolycarbonateSame (Polycarbonate). (Note: This appears to be a typo in the original document, listed twice with different materials)
Adhesive (shaft to hub-sheath valve assembly)CyanoacrylateSame (Cyanoacrylate)
Dilator shaft and hub materialPolyethylenePolyethylene with BaSO4. Comment: Similar, biocompatibility testing supports materials.
Pouch materialTyvek/nylon polyethylene pouchSame (Tyvek/nylon polyethylene pouch)

Study Proving Device Meets Acceptance Criteria:

The study involved a series of Design Verification tests and Biocompatibility tests.

The document states: "Design Verification testing were performed for the HeartLight 12F Deflectable Sheath to verify that design outputs meet relevant performance and functional requirements and conforms to the intended use of the device." And "The device specification requirements for functionality, performance specifications, physical characteristics, safety characteristics, labeling, packaging and shipping, shelf life, and sterilization all passed the specific tests and engineering evaluations. The testing demonstrates that the device does not raise questions of safety or effectiveness when compared to the predicate."

2. Sample size used for the test set and the data provenance:

  • Test Set Sample Size: The document does not specify the exact sample size for each performance test (e.g., how many sheaths were tested for deflection, kink resistance, etc.). The testing appears to be bench testing of the device itself rather than human-subject data.
  • Data Provenance: The data is generated from bench testing and laboratory biocompatibility studies of the device components. There is no information regarding country of origin of the data as it's not human subject data. The studies are prospective in the sense that they were conducted specifically for this 510(k) submission.

3. Number of experts used to establish the ground truth for the test set and their qualifications:

  • Not Applicable. This is a medical device clearance based on engineering performance and biocompatibility bench testing, not an AI/ML diagnostic or prognostic device requiring expert-established ground truth from patient data.

4. Adjudication method for the test set:

  • Not Applicable. As above, no human expert adjudication on clinical cases was involved. The evaluation relied on predefined engineering specifications and ISO standards for biocompatibility.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not Applicable. This is not an AI/ML powered device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not Applicable. This is not an AI/ML powered device. The "standalone" performance here refers to the device's physical and functional attributes in bench tests.

7. The type of ground truth used:

  • Engineering Specifications and ISO Standards: For performance testing (e.g., deflection angle, usable length, hemostasis, kink resistance), the "ground truth" was the predefined engineering specifications and criteria that the device needed to meet, often referenced against the predicate device's characteristics.
  • ISO 10993-1 Standards: For biocompatibility, the "ground truth" was compliance with the requirements outlined in ISO 10993-1 for an external communicating device in contact with circulating blood for less than 24 hours. Specific tests included: C3a Complement Activation Assay, SC5b-9 Complement Activation Assay, ASTM Partial Thromboplastin Time, USP Pyrogen Study, ISO Systemic Toxicity Study in Mice, ISO Intracutaneous Study in Rabbits, ASTM Hemolysis Study, Cytotoxicity Study Using the ISO Elution Method, ISO Guinea Pig Maximization Sensitization Test, and Thromboresistance Study in the Sheep (Heparinized Model).

8. The sample size for the training set:

  • Not Applicable. This is not an AI/ML powered device. The device itself is manufactured; it does not have a "training set" in the AI/ML sense.

9. How the ground truth for the training set was established:

  • Not Applicable. See point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized image of three human profiles facing right, arranged in a way that suggests a sense of community and support. The profiles are black and are set against a white background. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the logo.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 24, 2016

Cardiofocus, Inc. Seema Paliwal Regulatory Affairs Specialist II 500 Nickerson Road Suite 500-200 Marlborough, Massachusetts 01752

Re: K152310

Trade/Device Name: HeartLight Deflectable Sheath Regulation Number: 21 CFR 870.1280 Regulation Name: Steerable Catheter Regulatory Class: Class II Product Code: DRA, DYB Dated: February 17, 2016 Received: February 18, 2016

Dear Seema Paliwal:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Mitchell Stein

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K152310

Device Name HeartLight® Deflectable Sheath

Indications for Use (Describe)

The CardioFocus Deflectable Sheath is indicated for percutaneous catheter introduction into the vasculature and into the chambers of the heart. The sheath deflection facilitates catheter positioning.

Type of Use (Select one or both , as applicable)
----------------------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for CardioFocus. The logo consists of the word "CardioFocus" in a stylized font, with a curved line extending from the top of the "C" in "Cardio" to the top of a red heart shape to the right of the word "Focus". The font is black, and the heart is red with a white highlight.

5.0510(k) Summary - K152310
5.1General Information:
Date Summary Prepared:February 17, 2016
Applicant:CardioFocus Inc.
500 Nickerson Road
Marlborough, MA 01752
Telephone: (508) 658-7200
Fax: (508) 480-0600
Establishment Registration No.:1225698
Contact Person:Seema Paliwal, PhD, RAC
Regulatory Affairs Specialist II
Telephone (direct dial): (508) 658-7208
Fax: (508) 480-0600
Email: spaliwal@cardiofocus.com
Trade Name:HeartLight® Deflectable Sheath
Common Name:Steerable Sheath and Dilator
Classification Name:Steerable Catheter
Classification & Panel:Class II, 21 CFR § 870.1280, Cardiovascular
Product Code:DRA, DYB
Predicate Device(s):FlexCath® Advance™ Steerable Sheath & Dilator, Model 4FC12 (K123591)
Reference Device(s):HeartSpan® Steerable Introducer, Models FST-085-00 FST-085-01 and FST-085-02 (K132164)

5.2 Device Description

The HeartLight® Deflectable Sheath is a sterile, disposable device which provides a conduit for catheter introduction into the vasculature and into the chambers of the heart. The Deflectable Sheath is able to move its distal section in one direction in a single plane and is activated by a rotating knob at the front of the device handle. The deflection of the sheath and the torquing of the device allow the distal end of the sheath to be positioned at the ostium of pulmonary veins.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for CardioFocus. The logo consists of the word "CardioFocus" in a black, sans-serif font. Above the text is a black curved line that starts above the "C" in "CardioFocus" and ends at a red heart shape to the right of the text. The heart has a white spot on it, giving it a three-dimensional appearance.

The Deflectable Sheath is comprised of the following basic elements and features:

  • A radio-opaque flexible shaft, 12F ID and 16F OD, having a distal section which is deflectable by means of a wire connected to a mechanism in the handle. The shaft is hollow, comprised of a central lumen through which a catheter or a dilator is able to pass.
  • A handle providing a means to torque, deflect and axially position the sheath. Deflection is accomplished by twisting a knob at the distal end of the handle.
  • A hemostasis valve located at the proximal end of the central lumen and the handle forms a seal with an inserted object such as a catheter or a guidewire.
  • . A side port and stopcock connected to the hemostasis valve to permit evacuation of air inside the sheath and to permit constant irrigation of the sheath during use.
  • Vent holes located within the soft distal tip allow purging of the sheath when the end of the sheath is pressed against tissue.
  • A dilator that is matched to the sheath to provide a smooth transition from a guide wire to the outside diameter of the distal tip of the sheath. The dilator's hub has a lock feature to allow the dilator to be axially locked in relation to the sheath's handle when desired.
  • A radio-opaque marker at the distal tip.

This device is intended to be used within a hospital or medical facility only by physicians trained in standard transseptal techniques and familiar with angiography and electrophysiology catheters and delivery systems.

5.3 Intended Use

Facilitates introducing various cardiovascular catheters into the heart.

5.4 Indications for Use

The HeartLight" Deflectable Sheath is indicated for percutaneous catheter introduction into the vasculature and into the chambers of the heart. The sheath deflection facilitates catheter positioning.

The new device has identical indications for use as the predicate device, FlexCath® Advance™ Steerable Sheath. Both, the new device and the predicate device are intended for introducing and positioning catheters to reach the superior and inferior pulmonary veins of the heart and thus have clinically equivalent therapeutic use.

The reference device, HeartSpan® Steerable Introducer from Merit Medical is indicated for introducing various cardiovascular catheters into the heart, including the left side of the interatrial septum. The CardioFocus sheath has similar indications for use as the reference device.

5.4.1 Comparison to the Predicate Device

The CardioFocus Deflectable Sheath is compared to the currently marketed predicate device FlexCath Advance™ Steerable Sheath (K123591) in the following table:

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "CardioFocus" in black font. A black curved line extends from above the "C" in CardioFocus to a red heart shape to the right of the word. The heart is slightly blurred, giving it a soft, glowing appearance.

Element of ComparisonFlexCath® Advance™ SteerableHeartLight® Deflectable Sheath
Sheath (K123591)(K152310)
(Predicate device)(New Device)
Intended Use/ Indications for Use
Intended UseFacilitates introducing variousFacilitates introducing various
cardiovascular catheters into thecardiovascular catheters into the
Indications for Useheart.The FlexCath Advance Steerableheart.The CardioFocus Deflectable
Sheath is intendedforSheath is indicatedfor
percutaneouscatheterpercutaneouscatheter
introductionintotheintointroductionthe
vasculature and intotheand intovasculaturethe
chambers of the heart. Thechambers of the heart. The
sheath deflection facilitatessheath deflection facilitates
catheter positioning.catheter positioning.
Sterility, Shelf Life
UsageSingle use and disposableSame
Sterilization MethodSterilized with ethylene oxide.Same
Shelf Life2 yearsSame
Storage ConditionsStore in a cool, dark, dry place.Same
Performance/ Technology/Design
Steerablesheathimprovestoaccesshard-to-reachanatomyYesSame
Radiopaque for visualizationunder fluoroscopy - Radiopaquetip marker for enhanced fluorovisibilityYesSame
Uni-directional curve deflectionHemostasis valve to preventblood leakage and or air ingressYes ≥ 135°YesSimilar ≥ 130°The deflection angle providesmaneuverability to the cathetersthat advance through the sheathfor access to the chambers ofthe heart. The difference indeflection angle is minimal anddoes not affect the safety andeffectiveness of the new devicerelative to the predicate device.Bench testing of the sheathsupports the device design.Same
Ergonomic handle to provideYesSame
comfort
Element of ComparisonFlexCath® Advance™ SteerableSheath (K123591)(Predicate device)HeartLight® Deflectable Sheath(K152310)(New Device)
Soft, atraumatic tip to minimizethe likelihood of perforation ordamage to atrial or venoustissue.YesSame
Braided shaft providesexceptional torquability,pushability, and kink resistanceYesSame
Sideport for infusion andcontrast injectionYesSame
Usable Length65 cm75 cmDifferent from predicate.Usable length of the sheathprovides greater accessibility.The shaft usable length shouldbe long enough to place 13 cm ofsheath into the chambers of theheart. Bench testing of thesheath supports the devicedesign.
Inner Diameter12F12F - Same
Dilator to Guide wireCompatibilityAccepts up to 0.035" guidewireAccepts 0.035" or 0.038"guidewireSimilar. Dilator to Guide wirecompatibility defines thespecification of guide wireaccepted by the device. Benchtesting of the sheath supportsthe device design.
Biocompatibility - The devicemay contact circulating blood orhave indirect blood contact forup to 24 hoursYesSame nature and duration ofbodily contact.
Materials
Shaft MaterialPebax with Barium Sulfate,stainless steel braid and PTFElinerSame. Stainless Steel BraidReinforced Pebax with BariumSulfate with Stainless Steel Pull-Wire, PTFE Lined
Hemostasis valve materialSiliconeSame
Hemostasis hub and capmaterialPolycarbonateIsoplast
Element of ComparisonFlexCath® Advance™ SteerableSheath (K123591)(Predicate device)HeartLight® Deflectable Sheath(K152310)(New Device)
Hemostasis valvePolycarbonateSame
Adhesive (shaft to hub-sheathvalve assembly)CyanoacrylateSame
Dilator shaft and hub materialPolyethylenePolyethylene with BaSO4Similar. Biocompatibility testingof the sheath supports thedevice materials.
Pouch materialTyvek/nylon polyethylene pouchSame

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "CardioFocus" in a stylized font. A black curved line extends from the top of the "C" in "Cardio" to a red heart shape that is positioned to the right of the word "Focus". The heart shape is slightly blurred, giving it a soft appearance.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for CardioFocus. The logo consists of the word "CardioFocus" in a sans-serif font, with a black curved line extending from the top of the "C" in "Cardio" to a red heart shape to the right of the word "Focus". The heart shape has a white highlight on the upper left side.

5.4.2 Technological Characteristics - Comparison of New Device to Predicate Device

The HeartLight Deflectable Sheath has the following similarities to the predicate device FlexCath Advance Steerable Sheath (K123591):

  • Identical Intended Use/ Indications for Use
  • . Equivalent technological characteristics and basic sheath design - steerable and deflectable sheath for accessibility, radiopaque for visibility, ergonomic handle and braided shaft for functionality, hemostasis valve and soft tip for safety.
  • Similar materials
  • Same sterilization process and Shelf Life

The new device has minor differences from the predicate in the deflection angle, usable length, dilator to guide wire compatibility and material used on the dilator shaft and hub and hemostasis hub and cap.

The design verification and biocompatibility testing of the HeartLight Deflectable Sheath support the design, performance and use of the materials for their indications for use. The safety and effectiveness of the new device is not compromised relative to the predicate device.

5.5 Summary of Performance Data

Design Verification testing were performed for the HeartLight 12F Deflectable Sheath to verify that design outputs meet relevant performance and functional requirements and conforms to the intended use of the device.

Performance and functional requirements testing included:

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for CardioFocus. The logo consists of the word "CardioFocus" in a stylized font, with a black curved line extending from the top of the "C" in "Cardio" to a red heart shape to the right of the word "Focus". The heart shape is slightly blurred, giving it a soft, glowing appearance.

12F Sheath and dilator size:

Sheath overall and free length, Sheath tube O.D., tip I.D., tip wall thickness, marker band location, Sheath side port diameter and location, Dilator free length.

12F Sheath functionality

Deflectable sheath testing for accessibility, radiopaque for visibility, ergonomic handle and braided shaft for functionality, hemostasis valve and soft tip for safety. The specific testing included:

  • Sheath hemostasis and air leak resistance
  • -Sheath distal articulation stability, tip articulation max and min, bend radius, kink resistance
  • -Sheath functionality after foreseeable worst case use: constrained geometry, life cycle test, multiple insertions and retractions
  • -Sheath deflection and deflection stability
  • -Sheath handle actuation: torque required to articulate tip, functionality after handle overtorque
  • -Sheath injection pressure
  • -Sheath delamination of PTFE liner
  • -Catheter rotation test
  • -Packaging integrity
  • -Sheath side port joint integrity, Stopcock tubing kink specification

Shelf life testing

The 2 year accelerated shelf life design verification testing was performed to support 2 year shelf life of the 12F Deflectable Sheath. The test results passed the design verification and established a two year shelf life for the HeartLight® 12F Deflectable Sheath.

Sterilization Validation (Ethylene Oxide)

Biocompatibility Testing (Verification)

Biocompatibility tests for the 12F Deflectable sheath and the 12 Dilator were performed according to ISO 10993-1 for an external communicating device in contact with circulating blood for less than 24 hours. The tests were leveraged from the reference device HeartSpan® Steerable Introducer Kit (K132164). The results conclude that the Sheath material passed the required ISO 10993 biocompatibility tests and thus support the use of the materials according to their indications for use.

The device were subjected to the following tests:

C3a Complement Activation Assay SC5b-9 Complement Activation Assay ASTM Partial Thromboplastin Time USP Pyrogen Study – Material Mediated ISO Systemic Toxicity Study in Mice ISO Intracutaneous Study in Rabbits ASTM Hemolysis Study

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the word "CardioFocus" in a stylized font. A curved line extends from above the "C" in "Cardio" to the right side of the word, ending at a red heart shape. The heart is positioned to the right of the word "Focus", visually connecting the text and the heart symbol.

Cytotoxicity Study Using the ISO Elution Method ISO Guinea Pig Maximization Sensitization Test Thromboresistance Study in the Sheep, Heparinized Model

The device specification requirements for functionality, performance specifications, physical characteristics, safety characteristics, labeling, packaging and shipping, shelf life, and sterilization all passed the specific tests and engineering evaluations. The testing demonstrates that the device does not raise questions of safety or effectiveness when compared to the predicate.

5.6 Conclusion

CardioFocus, Inc. considers its HeartLight Deflectable Sheath to be substantially equivalent to the currently marketed predicate device FlexCath (K123591). This assessment is based upon analysis of identical indications for use and similar technological characteristics, design, performance and materials used.

Bench testing was performed to support substantial equivalence between the proposed device, HeartLight® 12F Deflectable Sheath and the predicate FlexCath® Advance™ Steerable Sheath (K123591). The Design Verification testing demonstrated that design outputs met relevant performance and functional requirements and conforms to the intended use/indications of use for the device. The 12F Deflectable Sheath has identical intended use/ indications for use and performance specifications to that of the FlexCath® Advance™ Steerable Sheath cleared under K123591. Therefore the bench testing performed demonstrates substantial equivalence with the testing and intended use of primary predicate. The biocompatibility testing does not raise questions of safety or effectiveness since the material used for the 12F Deflectable sheath is the same as the reference device cleared under K132164 and has been in the market for a similar indications for use.

Design verification and biocompatibility testing results indicate that the properties and performance of the device are suitable for its intended use and no new issues of safety or effectiveness have been raised.

§ 870.1280 Steerable catheter.

(a)
Identification. A steerable catheter is a catheter used for diagnostic and monitoring purposes whose movements are directed by a steering control unit.(b)
Classification. Class II (performance standards).