K Number
K152273
Date Cleared
2015-10-30

(79 days)

Product Code
Regulation Number
878.4635
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

INTENDED USE: This ultraviolet lamp is intended for use in sunlamp products for tanning of the human skin.

Device Description

This 510(k) applies to more than one hundred and sixty individual devices that fall into a generic class of ultraviolet lamps that have the same basic technological features and exactly the same intended use. The ultraviolet sunlamps of this 510(k) are classified scientifically as low pressure, mercury-rare gas discharge devices. The general lamp construction and technological principle of operation of all of the devices of this 510(k) is the same as that of a common fluorescent lamp. The main differentiators for the various sunlamps of this 510(k) include: lamp length, lamp diameter, lamp wattage, spectral characteristics and private labeling thereof.

The ultraviolet lamps named herein comprise a tubular glass envelope, the internal surface of which is coated with a fluorescent phosphor. To each end of the tube, a glass mount is sealed. The main purpose of the mount is to provide 1) for a means of hermetically sealing the internal atmosphere of the tube, 2) a means of supporting an electrode and a means of 3) electrically accessing the internal atmosphere of the device via electrically conducting lead-wires. Once assembled as described, the envelope is evacuated of air, the air is replaced by an inert gas at a pressure significantly lower than that of the normal atmosphere, and a drop of mercury is injected into the enclosed envelope. Finally, the envelope is sealed.

In operation, the device, now called a "lamp", is connected to an electrical source typically in series with a ballast that provides for 1) ignition of a discharge through the inert-gas-mercury vapor mixture and 2) its subsequent stabilization and sustainment of the electrical behavior of the lamp. Once connected to the applied electric field, free electrons are accelerated through the mercury vapor wherein collisions take place with the mercury atoms. These collisions cause excitation of the electrons in the mercury (Hq) atom and the production of photons at various wavelengths. Certain of these photons travel to the bulb wall and excite the phosphor powder. The phosphor, in turn, emits radiation that is transmitted through the transparent glass envelope. The radiation that is emitted from a standard fluorescent general lighting lamp is primarily in the visible region of the spectrum (380nm-760nm). The radiation that is emitted from an ultraviolet sunlamp is primarily in the UV range of the electromagnetic spectrum (300nm-400nm). Modern day, low pressure ultraviolet sunlamps typically emit most of their energy in the wavelength range that is closest to the visible: that is, in the UVA range (320nm-400nm) range. The emission spectrum of sunlamps usually contains a relatively small proportion of UVB radiation. (260nm-320nm)

AI/ML Overview

The provided document is a 510(k) premarket notification for "Cosmedico Ultraviolet Lamps of Various Models." It does not describe an AI/ML powered device, but rather conventional sunlamp products. Therefore, the specific information requested about acceptance criteria and studies related to an AI/ML device's performance, such as sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, and ground truth establishment, is not applicable to this document.

The document indicates that the device is a conventional sunlamp and relies on equivalence to legally marketed predicate devices rather than novel AI/ML technology requiring specific performance studies as outlined in the request.

However, I can extract the acceptance criteria and a summary of performance testing for the conventional device as described in the document:

1. Table of Acceptance Criteria and Reported Device Performance

Test (Performance Criteria)Lamps meet criteria of predicate devices (Yes/No)
Functionality / Light upYes
Electrical CharacteristicsYes
Burn-in BehaviorYes
Dimensional CharacteristicsYes
Mechanical SafetyYes
Electrical SafetyYes
Physical Design (e.g., reflector angle)Yes
UV PerformanceYes
UVA IrradianceYes
Erythemal Effective IrradianceYes
UVC/UVB Ratio (per 21 CFR 1040.20)Yes

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document states: "According to a Quality Sampling Plan, lamps are tested in-process and / or post-production". It does not specify the exact sample size for the test set or the data provenance beyond "Quality Sampling Plan". Given the nature of the device (sunlamps), the tests are likely conducted on manufactured samples, which implies prospective testing during or after production. There is no information regarding the country of origin of the data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. The device is a conventional sunlamp, and its performance is measured against technical specifications and predicate device characteristics, not through expert-labeled ground truth for diagnostic purposes. Performance testing involves physical and electrical measurements.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. Adjudication methods are typically relevant for subjective assessments by multiple human readers, often in image interpretation or clinical scenarios where ground truth is expert-derived. This device undergoes objective performance testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is not an AI-powered system designed to assist human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is not an algorithm or AI system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for these sunlamps is their conformance to established technical specifications and performance characteristics, which are compared to legally marketed predicate devices. This includes objective measurements of electrical characteristics, dimensions, mechanical safety, UV radiation characteristics (UVA Irradiance, Erythemal Effective Irradiance, UVC/UVB ratio), and functionality. It is not an expert consensus, pathology, or outcomes data ground truth as would be used for a diagnostic AI device.

8. The sample size for the training set

Not applicable. This device is not an AI/ML model that undergoes a training phase.

9. How the ground truth for the training set was established

Not applicable. This device is not an AI/ML model.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines suggesting hair or clothing. The profiles are arranged in a way that creates a sense of depth and unity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 30, 2015

Cosmedico Light Incorporated Mr. Steven C. Schlitt Director Engineering & Quality 233 Libbey Industrial Parkway Weymouth, Massachusetts 02189

Re: K152273 Trade/Device Name: Cosmedico Ultraviolet Lamps of Various Models Regulation Number: 21 CFR 878.4635 Regulation Name: Sunlamp products and ultraviolet lamps intended for use in sunlamp products Regulatory Class: Class II Product Code: LEJ Dated: August 15, 2015 Received: August 19, 2015

Dear Mr. Schlitt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR [SELECT ONE: Part 801 [or, for IVDs only] Parts 801 and 809]); medical device reporting (reporting of medical device-related adverse events) (21 CFR

{1}------------------------------------------------

803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Joshua C. Nipper -S

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K152273 Not used.

Device Name

Cosmedico Ultraviolet Lamps of Various Models (intended for use in Sunlamp Products and commonly referred to as "sunlamps")

Indications for Use (Describe)

INTENDED USE: This ultraviolet lamp is intended for use in sunlamp products for tanning of the human skin.

Type of Use (Select one or both, as applicable)
Research Use Only (Does not require IRB review) Gene Therapy (Requires IRB review) Research Use Only (Does not require IRB review) Gene Therapy (Requires IRB review)
Research Use Only (Does not require IRB review) Gene Therapy (Requires IRB review)

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Per Sec. 807.92(c)

807.92(a)(1)

Date Prepared:August 10, 2015/ Revised 10-30-15
510(k) Company / Holder Name:Cosmedico Light, Inc.
Contact Name:Steven C. Schlitt
Title:Director Engineering & Quality
Address:233 Libbey Industrial ParkwayWeymouth, MA 02189
Phone Number:781-331-0949 (Ext. 107)
Fax Number:781-331-4766

807.92(a)(2)

Device Proprietary Name:Cosmedico Ultraviolet Lamps of VariousModelsSee attachment F-1 for list of models and part numbers.
Device Common or Usual Name:Ultraviolet Sunlamps
Indication for Use / Intended UseINTENDED USE: This ultraviolet lamp isintended for use in sunlamp products fortanning of the human skin.
Classification Name:Sunlamp products and ultraviolet lampsintended for use in sunlamp products
Classification Code:LEJ
Regulation Number:21 CFR Part 878.4635
Device ClassificationII

{4}------------------------------------------------

807.92(a)(3)

Predicate Devices:

Prior to September 2, 2014. the basic ultraviolet sunlamps named in this 510(k) existed and were offered for sale as legally marketed, Class I, 510(k) exempt, medical devices. These devices are historically Over-the-Counter devices.

For the basis of this 510(k), Cosmedico Light claims substantial equivalence to these legally marketed OTC devices. Since none of the ultraviolet lamps named in this 510(k) have changed in 1) intended use or 2) technological design characteristics after September 2, 2014, essentially the contemporary ultraviolet lamps are identical to the predicate devices (i.e., those basic sunlamps offered for sale prior to the FDA's cutoff date of Sept. 2, 2014).

Such an approach of using the legally marketed Class I medical devices as predicate devices is validated in the Final Reclassification Order published on June 2, 2014 in the Federal Register Vo. 79, No. 105, Page 31212 whereby it is stated that:

" FDA cleared several 510(k)s for sunlamp products prior to exempting the devices from premarket notification submission. At least one 510(k) for a sunlamp product has been cleared since then under product code LEJ. These cleared sunlamp products, as well as any 510(k)-exempt sunlamp product or UV lamp intended for use in a sunlamp product legally offered for sale on or before September 2, 2014, can serve as predicates for substantial equivalence purposes."

lt will be demonstrated in this 510(k) that the named ultraviolet lamps:

    1. have the same intended use as the predicate OTC devices
    1. have the same technological characteristics as the predicate devices and
    1. conform to the special controls required by the reclassification order

In short, the contemporary ultraviolet sunlamps of this 510(k) are as safe and as effective as the predicate devices.

Description of the Devices:

This 510(k) applies to more than one hundred and sixty individual devices that fall into a generic class of ultraviolet lamps that have the same basic technological features and exactly the same intended use. The ultraviolet sunlamps of this 510(k) are classified scientifically as low pressure, mercury-rare gas discharge devices. The general lamp construction and technological principle of operation of all of the devices of this 510(k) is the same as that of a common fluorescent lamp. The main differentiators for the various sunlamps of this 510(k) include: lamp length, lamp diameter, lamp wattage, spectral

{5}------------------------------------------------

characteristics and private labeling thereof.

The ultraviolet lamps named herein comprise a tubular glass envelope, the internal surface of which is coated with a fluorescent phosphor. To each end of the tube, a glass mount is sealed. The main purpose of the mount is to provide 1) for a means of hermetically sealing the internal atmosphere of the tube, 2) a means of supporting an electrode and a means of 3) electrically accessing the internal atmosphere of the device via electrically conducting lead-wires. Once assembled as described, the envelope is evacuated of air, the air is replaced by an inert gas at a pressure significantly lower than that of the normal atmosphere, and a drop of mercury is injected into the enclosed envelope. Finally, the envelope is sealed.

In operation, the device, now called a "lamp", is connected to an electrical source typically in series with a ballast that provides for 1) ignition of a discharge through the inert-gas-mercury vapor mixture and 2) its subsequent stabilization and sustainment of the electrical behavior of the lamp. Once connected to the applied electric field, free electrons are accelerated through the mercury vapor wherein collisions take place with the mercury atoms. These collisions cause excitation of the electrons in the mercury (Hq) atom and the production of photons at various wavelengths. Certain of these photons travel to the bulb wall and excite the phosphor powder. The phosphor, in turn, emits radiation that is transmitted through the transparent glass envelope. The radiation that is emitted from a standard fluorescent general lighting lamp is primarily in the visible region of the spectrum (380nm-760nm). The radiation that is emitted from an ultraviolet sunlamp is primarily in the UV range of the electromagnetic spectrum (300nm-400nm). Modern day, low pressure ultraviolet sunlamps typically emit most of their energy in the wavelength range that is closest to the visible: that is, in the UVA range (320nm-400nm) range. The emission spectrum of sunlamps usually contains a relatively small proportion of UVB radiation. (260nm-320nm)

Some critical design components and aspects of the sunlamps include:

Glass Bulb (tubular envelope): Must be properly dimensioned and provides for transmission of ultraviolet energy in the range required for tanning.

Phosphor Powder: In addition to proper physical properties such as particle size and aqueous suspension compatibility, the phosphor must emit in the wavelength range required for a particular tanning application. It is the phosphor that primarily determines the spectral irradiance distribution of the radiation from the lamp.

Fill Gas and Pressure: The fill gas (in conjunction with the electrical control gear) determines the electrical characteristics of the lamp and has a large influence on the intensity of the radiation and the temporal depreciation of light output over the usage range.

{6}------------------------------------------------

Coated Electrodes (Cathodes / Coils): Provide a surface area for cathode coating and to a great extent influence starting of the lamp and its physical life.

Mercury: Mercury provides the basis for the operation of the lamp. It is through the excitation of mercury atoms and the subsequent quenching of the excited atoms in the vapor state that "light" is produced. It is the shortwave UV photons that resulting from the excitations that excite phosphor to produce light in a specific spectral range.

Bases: Allow the lamp to be installed in appropriate sockets for electrical connectivity to the electrodes.

Image /page/6/Figure/4 description: This image shows a diagram of a fluorescent lamp and its components. The diagram labels the different parts of the lamp, including the bulb, phosphor, cathode, exhaust tube, base, mercury, gas, stem press, and lead-in wires. The text provides a brief description of each component and its function within the lamp. For example, the phosphor coating inside the bulb transforms ultraviolet radiation into visible light.

For illustrative purposes:

Figure 2. Basic parts of a typical hot cathode fluorescent lamp.

Intended Use/Indications for Use

INTENDED USE: This ultraviolet lamp is intended for use in sunlamp products for tanning of the human skin.

TYPE OF USE: The predicate devices named in this 510(k) are historically used as Over the Counter (OTC) devices.

Technological Characteristics:

The technological characteristics of the ultraviolet sunlamps which are the subject of this 510(k) are identically the same as the technological characteristics of the predicate devices. The tables and data that are reported in Section 10 and the information summarized in Section 12, "Substantial Equivalence Discussion" of this 510(k), document that there are no significant differences between the subject and predicate devices. In fact, the subject and the predicate devices are "one in the same".

{7}------------------------------------------------

Performance Testing Summary:

  • . All lamps are 100% tested for functionality at the end of the manufacturing process. Lamps that do not meet critical characteristics for functionality and safety are rejected.
  • . According to a Quality Sampling Plan, lamps are tested in-process and / or post-production for:
    • Electrical Characteristics and Safety A
    • Mechanical Safety A
    • Dimensional Integrity

    • Physical Design Attributes

    • UV radiation Characteristics, including:

      • UVA Irradiance
      • -Erythemal Effective Irradiance
      • . UVC/UVB ratio per 21 CFR 1040.20

As noted, the subject lamps of this 510(k) are identical to the predicate devices. Therefore, the subject lamps and the predicate devices have the same performance requirements.

Test (Performance CriteriaLamps meet criteria of predicatedevices (Yes/No)
Functionality / Light upYes
Electrical CharacteristicsYes
Burn-in BehaviorYes
Dimensional CharacteristicsYes
Mechanical SafetyYes
Electrical SafetyYes
Physical Design (e.g., reflector angle)Yes
UV PerformanceYes
UVA IrradianceYes
Erythemal Effective IrradianceYes
UVC/UVB Ratio (per 21 CFR1040.20)Yes

Lamps that do not meet Acceptance Criteria are either re-worked or scrapped.

{8}------------------------------------------------

COSMEDICOPART NUMBERLAMP DESCRIPTION
10018FR18T8 / COSMOLUX HRF
10036F25WT5 / COSMOLUX LS
10044F15WT5 / COSMOLUX X
10180F59T12 / HI TAN / 80W
10186F71T12 / Bermuda Gold Premium Plus / 105W
10187F73T12 / Bermuda Gold / RDC / 105W
10198F71T12 / HI TAN / 100W
10199F73T12 / HI TAN / 100W / RDC
10201F73T12 / EXPRESS TAN HO / 100W / RDC
10202F73T12 / Genesis / 120W / RDC
10204F71T12 / Genesis Hyperbronze / 105W
10205F73T12 / Genesis Hyperbronze / 105W
10206F72T12 / Genesis Plus / 105W
10212F71T12 / Genesis / 120W
10217F71T12 / Eternal Sun XPFT / 100W
10218F73T12 / Eternal Sun XPFT / 100W / RDC
10219F71T12 / Eternal Sun XP Gold / 100W
10220F73T12 / Eternal Sun XP Gold / 100W / RDC
10222F59T12 / Eternal Sun XP Gold / 80W
10224F71T12 / Eternal Sun XP Gold / 120W
10318FR18T8 / Sundash HRF
10330F25T5 / UVA Plus
10331F25T5 / COSMOLUX LS
10332FR25WT5 / LSR
10990F71T12 / COSMOLUX GSP 9K90 / 100W
10994F71T12 / Eternal Sun XP Gold 9K90 / 100W
11282F59T12 / GOLDARIUM SP / 80W
11305F71T12 / GOLDARIUM SP / 100W
11306F73T12 / GOLDARIUM SP / 100W
12482F59T12 / COSMOSTAR LONGLIFE / 80W
12503F71T12 / COSMOSTAR LONGLIFE / 100W
12600F71T12 / COSMOLUX ESP / 100W
12601F73T12 / COSMOLUX ESP / 100W / RDC
12607F73T12 / Genesis GS 120W / RDC
15163FR59T12 / Eternal Sun XP*R / 80W
15165FR71T12 / Palm Beach Mega Tan S1-S2 / 100W
15211FR59T12 / COSMOLUX VLR-TT / 80W / 220°
15212FR71T12 / COSMOLUX VLR TT / 100 W
15214FR71T12 / Bermuda Gold PLUS / 100W
15215FR71T12 / COSMOSUN R / 100W
15216FR71T12 / Eternal Sun XP*R / 100W
15220FR79T12 / COSMOLUX VLR / 120W
15602FR71T12 / COSMOLUX ESP R / 100 W
COSMEDICOPART NUMBERLAMP DESCRIPTION
15603FR73T12 / COSMOLUX ESP / 100W / RDC
15640FR71T12 / Genesis Super Power VHP15 Plus / 120W
15642FR71T12 / Genesis Super Power VHP20 Plus / 120W
15649FR71T12 / Genesis Hyperbronze / 105W
15656FR71T12 / Bermuda Gold / Supreme 20 / 120W
15691FR71T12 / /COSMOLUX VLR 9K-90 / 100W
15700FR71T12 / COSMOLUX RA PLUS / 100W
15704FR24T12 / COSMOLUX RA PLUS / 40W
15710FR71T12 / COSMOLUX RCS / 100W
16077FR73T12 / HI TAN / VHO / RDC
16107FR73T12 / COSMOLUX VHR-TT / 160W / RDC
16112FR59T12 / COSMOLUX VHR-TT / 140W
16121FR71T12 / Genesis 170WRS-E / 160W
16161FR71T12 / COSMOLUX / VHO
16169FR71T12 / COSMOLUX VHR-TT / 160W
16172F72T12 / HI TAN / VHO / RDC
16180FR71T12 / COSMOLUX VHR / 160W
16202FR75T12 / COSMOLUX VHR 200 / 1.9 M
16211FR75T12 / COSMOLUX VHR / HP / 180W
16218FR71T12 / COSMOLUX VHR-200
16220FR79T12 / COSMOLUX VHR 200 / 2.0 M
16223FR79T12 / COSMOLUX 9K90 / 180W-200W
16234FR79T12 / COSMOLUX 9K90 / 225W
16236FR71T12 / COSMOSUN R / 160W
16238FR79T12 / Ergoline Turbo Power / 200W
16239FR74T12 / COSMOLUX VHR-TT / 160-200W
16250FR79T12 / Suncapsule 250 MAX / 180-250W
16258FR79T12 / Eternal Sun XP*r / 2.0M / 180-200W
16259FR79T12 / Eternal Sun XP*R / 9K90 / 2.0M / 180-200W
16286FR79T12 / COSMOLUX VHR HP 250 MAX / 180-250W
16307FR72T12 / Genesis 170WRS / RDC
16317FR73T12 / Express Tan VR 160W / RDC
16332FR71T12 / COSMOLUX VHR / 200 / HP
16341FR71T12 / Genesis Turbo Power VHP20 PLUS / 160W
16344FR71T12 / Genesis Turbo Power VHP15 PLUS / 160W
16367FR71T12 / Bermuda Gold Premium Plus / 160W
16385FR71T12 / Ergoline Turbo Power VHP12 / 160W
16386FR72T12 / Genesis GS 170W / RDC
16388FR71T12 / HI-TAN / VHR / 160W
16391FR71T12 / COSMOLUX VHR / 9K90 / 160W
16392FR71T12 / Genesis 9K90 / 160WR
16393FR71T12 / COSMOLUX VHR FLEX / 160-200W
16394FR71T12 / COSMOLUX VHR-XT 160W
COSMEDICOPART NUMBERLAMP DESCRIPTION
16395FR71T12 / COSMOLUX VHR / HP / 160W
16398FR71T12 / Genesis Hyperbronze / 170-200W
16399FR71T12 / Suncapsule SCR 160 / 160W
16445FR71T12 / Eternal Sun XP*R / 160W
16447FR71T12 / Eternal Sun XP*R 9K90 /160W
16459FR59T12 / Eternal Sun XP*R / 140W
16481FR71T12 / Eternal Sun XP / 160W
16542FR71T12 Eternal Sun XP HR 160W
18810FR71T12 / COSMOLUX VLR.T / 100 W
21069F25T5 / Radiance / 25W (07040)
21248F72T12 / Radiance 8000 7.5 160W RDC (03017)
21249F74T12 / Radiance 8.5 / 100-120W (07034)
21250F72T12 / Radiance 7000 6.5 / 100-120W (03016)
21626FR18T8 / Radiance HRF (07041)
22010FR79T12 / Genesis VHP 20 / 120W
22015FR79T12 / Genesis VHP 15 / 120W
22300F73T12 / Radiance 120 / 120W / RDC (07044)
22302F73T12 / Eternal Sun XP Gold / 120W
26111F73T12 / FABUTAN /100W / RDC
26113FR71T12 / Radiance 71i / 120W
26114FR75T12 / Radiance 75i / 120W
26180F71T12 / Radiance JAGUAR 10.0 / 100W (61747)
26189FR71T12 / Silver Sun / 100W R
26190F71T12 / Radiance 9000 8.5 / 100W (03011)
26191F71T12 / Radiance 5K 5.0 / 100W (03001)
26192F73T12 / Radiance 7000 RDC 6.5 / 100W (03006)
26193F71T12 / Radiance 8000 7.5 / 100W (03009)
26194F71T12 / Radiance PLUS 7000 6.5 / 100-120W (07011)
26195F73T12 / Radiance 9000 8.5 / 100-120W / RDC (03012)
26196FR73T12 / Radiance 7000 6.5 / 100W / RDC (03008)
26197F73T12 / Millennium 7000 6.5 / 100W (03027)
26198F73T12 / Radiance Plus 7000 6.5 / 100-120W (07013)
26199F59T12 / Radiance 5000 / 5.0 / 80W (03000)
26202F73T12 / Radiance +5000 5.0 / 100W (07003)
26203FR73T12 / Radiance+5000 5.0 / 100-120W / RDC (07006)
26267F73T12 / Radiance 8000 7.5 / 100-120W
26303F71T12 / California Sun Ultra / 105W
26304FR71T12 / California Sun Mega / 120W
26305FR71T12 / California Sun Premier / 170W
26308F71T12 / Radiance Black / 100W (03076)
26309F73T12 / Radiance Black / 100-120W / RDC (03076)
26673F73T12 / Radiance 8000 7.5 / 160W RDC (03018)
26682FR71T12 / Radiance R2 XLP 1.7 / 100-120W (03100)
COSMEDICOPART NUMBERLAMP DESCRIPTION
26774FR71T12 / SilverSun / 160WR
27355FR79T12 / Maxlight e-power XL / 200W-R
27450FR71T12 / Radiance R2 / 100-120W / HOR 3.7 (07043)
27451FR71T12 / Radiance R2 VHO R / 160-200W (07051)
27483FR72T12 / Radiance R2 VHOR / 160-200W (07052)
27524F73T12 / Radiance 5000 5.0 / RDC / 100W (03002)
27525FR71T12 / Radiance 5000 5.0 / 100-120W (03003)
27526F71T12 / Radiance 7000 6.5 / 100-120W (03005)
27527FR71T12 / Radiance 7000 6.5 / 100-120W (03007)
27528FR71T12 / Radiance Plus+ 7000 6.5 / 100-120W (07015)
27529FR59T12 / Radiance R2 / 140W (07050)
27530FR73T12 / Radiance R2 / 160-200W / RDC (07053)
27531F59T12 / Radiance 7000 6.5 / 80W (03014)
27532FR59T12 / Radiance Plus+ 6.5 / 80W (07014)
27533FR59T12 / Radiance R2 3.7 / 80W (07042)
27534FR71T12 / Palm Beach Mega Tan / 160W
27535FR59T12 / Premium Plus Bermuda Gold Plus / 140W
27538FR71T12 / Sundazzler Premium / 160W
27540FR2.0M / Genesis GSXR 200 / 180W
27541FR71T12 / Radiance R2 XLP 7.0 / 160W R (03098)
27548FR79T12 / Radiance R2 / 2M / 120W (03081)
27574FR74T12 / Eternal Sun XP*R / 200W / RDC
27712FR73T12 / Radiance R2 3.7 / 100-120W (07045)
27728F71T12 / Millennium 5000 5.0 /100W (03021)
27733FR72T12 / RadianceR2 VIP / FR71-RDC / 160W (07066)
27736FR71T12 / GOLDARIUM SR / 100W
27748FR79T12 / Eternal Sun BronzeR / 225W
27749FR71T12 / Eternal Sun XP20R / 120W
27751FR71T12 / Radiance Black / 100-120W (07075)
27756FR71T12 / Radiance Black / 160W (07076)
27757FR71T12 / Bermuda Gold / 100 W
28018FR79T12 / Radiance R2 2M / 200W (03080)
28019FR74T12 Radiance R3 VHOR / RDC / 200W (07048)
28021F71T12/ Millennium 7000 6.5 / 100W (03026)
28022FR71T12 / Radiance R2 Plus / 200W (07059)
28996FR71T12 Millennium M2 3.15 / 160-180W (03087)
901935140FR71T12 / Ergoline Super Power SR / 100W
926295040FR71T12 / CLEO ADVANTAGE XPT / 160W-R
971036940FR71T12 / CLEO ADVANTAGE / 100W-R

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

§ 878.4635 Sunlamp products and ultraviolet lamps intended for use in sunlamp products.

(a)
Identification. A sunlamp product is any device designed to incorporate one or more ultraviolet (UV) lamps intended for irradiation of any part of the living human body, by UV radiation with wavelengths in air between 200 and 400 nanometers, to induce skin tanning. This definition includes tanning beds and tanning booths. A UV lamp intended for use in sunlamp products is any lamp that produces UV radiation in the wavelength interval of 200 to 400 nanometers in air.(b)
Classification. Class II (special controls). The special controls for sunlamp products and UV lamps intended for use in sunlamp products are:(1) Conduct performance testing that demonstrates the following:
(i) Device meets appropriate output performance specifications such as wavelengths, energy density, and lamp life; and
(ii) Device's safety features, such as timers to limit UV exposure and alarms, function properly.
(2) Demonstrate that device is mechanically safe to prevent user injury.
(3) Demonstrate software verification, validation, and hazard analysis.
(4) Demonstrate that device is biocompatible.
(5) Demonstrate that device is electrically safe and electromagnetically compatible in its intended use environment.
(6)
Labeling —(i)Sunlamp products. (A) The warning statement below must appear on all sunlamp products and must be placed in a black box. This statement must be permanently affixed or inscribed on the product when fully assembled for use so as to be legible and readily accessible to view by the person who will be exposed to UV radiation immediately before the use of the product. It shall be of sufficient durability to remain legible throughout the expected lifetime of the product. It shall appear on a part or panel displayed prominently under normal conditions of use so that it is readily accessible to view whether the tanning bed canopy (or tanning booth door) is open or closed when the person who will be exposed approaches the equipment and the text shall be at least 10 millimeters (height). Labeling on the device must include the following statement: Attention: This sunlamp product should not be used on persons under the age of 18 years. (B) Manufacturers shall provide validated instructions on cleaning and disinfection of sunlamp products between uses in the user instructions.
(ii)
Sunlamp products and UV lamps intended for use in sunlamp products. Manufacturers of sunlamp products and UV lamps intended for use in sunlamp products shall provide or cause to be provided in the user instructions, as well as all consumer-directed catalogs, specification sheets, descriptive brochures, and Web pages in which sunlamp products or UV lamps intended for use in sunlamp products are offered for sale, the following contraindication and warning statements:(A) “Contraindication: This product is contraindicated for use on persons under the age of 18 years.”
(B) “Contraindication: This product must not be used if skin lesions or open wounds are present.”
(C) “Warning: This product should not be used on individuals who have had skin cancer or have a family history of skin cancer.”
(D) “Warning: Persons repeatedly exposed to UV radiation should be regularly evaluated for skin cancer.”
(c)
Performance standard. Sunlamp products and UV lamps intended for use in sunlamp products are subject to the electronic product performance standard at § 1040.20 of this chapter.