K Number
K152188
Date Cleared
2016-01-28

(176 days)

Product Code
Regulation Number
872.3640
Panel
DE
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

DenTack QUAD Dental Implants and Prostheses are intended for surgical placement in the maxilla and/or in the mandible to provide a means for prosthetic attachment in single tooth restorations and in partially or fully edentulous jaws utilizing delayed loading or immediate loading when good primary stability is achieved and with appropriate occlusal loading.

Device Description

DenTack has developed a range of expandable dental implants made of Titanium to serve the need of patients that require partial or complete tooth restoration. The implant is placed so its apical end is in the trabecular (spongy) bone like any other implant. Once in place, the apical portion of the implant is expanded to achieve increased contact surface area with the surrounding bone. The system also includes various accessories that attach to the implant. All DenTack's implant bodies are made of a combination of Titanium alloy according to ASTM F136 and c.p. titanium according to ASTM F67. Abutments are made of same Titanium alloy while denture components are made of stainless-steel and polymers.

DenTack's QUAD Implants are available in outer diameter (OD) of: 3.75 and 4.1 mm. The implants' length dimensions are from 7 to 11 mm.

DenTack's Prostheses are identified as follows:
Metal housing
Plastic cup
Straight Multi-unit Sleeve
Straight Multi-unit Screw
Angled abutments 15° and 25°
Angled abutments 15° Large
Ball Attachments (1, 2 & 3 mm)
Direct screw platform
Straight abutments
Screw retained abutments
Straight Multi-unit abutments
Healing caps
Cover Screw

The QUAD expansion is performed using a reusable Expansion Tool and Ratchet Torque.

AI/ML Overview

Here's an analysis of the acceptance criteria and supporting studies for the DenTack QUAD Dental Implants and Prostheses based on the provided document:

Acceptance Criteria and Device Performance for DenTack QUAD Dental Implants and Prostheses

1. Table of Acceptance Criteria and Reported Device Performance

Test/CharacteristicAcceptance CriteriaReported Device Performance
Mechanical Performance
Fatigue Test (ISO 14801:2007)Device performs at least as good as or better than the predicate device."The QUAD met the acceptance criteria and performed at least as good as or better than the predicate device." Tested on worst-case configuration (25° abutment with narrowest implant OD).
Evaluation of Minimal Rotation Torque after Placement and ExpansionTest results to be similar to and thus equivalent to the predicate device."The tests results found to be very similar and thus equivalent. The QUAD met the acceptance criteria and performed at least as good as or better than the predicate device."
Partially expanded QUAD Implant Reciprocating Effect TestNo additional damage to surrounding bone or implant from counterclockwise/clockwise rotation; performs similarly to reference device."The study demonstrated that both the surrounding bone and the implant were not negatively affected by counterclockwise/clockwise rotation and that the QUAD performed similarly to the reference device."
Material & Surface Characteristics
Surface Analysis (SEM)-Implant surface tested using SEM. (No specific acceptance criteria or performance metric beyond "tested" is stated).
Biocompatibility (ISO 10993-1, -5, -12)No incompatibility potential or adverse biological effect."The chemical and biological tests have not revealed any incompatibility potential or any adverse effect."
Sterilization & Packaging
Gamma Sterilization Validation (ISO 11137-2, AAMI TIR 33 / ISO TS 13004)Successful sterilization according to standards."Sterilization was conducted with successful results, using Gamma Irradiation according to VDmax method..."
Steam Sterilization Validation (ISO 17665-1:2006, ANSI AAMI ST79:2010, ANSI AAMI ST77:2013)Validation results support SAL 10⁻⁶; IFU in-line with validation."The validation results supported SAL 10-6. IFU (instructions for use) are in-line with the validation results."
Shelf Life Validation (ISO 11607-1)Sterility maintained for the device shelf life."This validation confirmed that the sterility of DenTack's sterile products will be remained for the device shelf life."
Functional/Clinical Equivalence
QUAD Removal after ExpansionThe expandable design does not cause additional damage to the surrounding tissue."The evaluation post implant removal showed that both implants performed equally in terms of interaction with the bone."

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the numerical sample sizes for most of the individual tests. It generally refers to testing being conducted "for Dentistry - Implants - Dynamic fatigue test for endosseous dental implants" and "in comparison to predicate device" or "legally marketed device."

  • Fatigue Test (ISO 14801:2007): "Worst case configuration was selected." (Implies a specific combination of implant and abutment was tested, but no number of samples is given).
  • QUAD Removal after Expansion: "artificial bone was used and a reference legally marketed device for comparison." (No number of samples or explicit provenance beyond "artificial bone").
  • Evaluation of Minimal Rotation Torque: "artificial bone." (No number of samples).
  • Partially expanded QUAD Implant Reciprocating Effect Test: "a legally marketed device with similar instructions for use." (No number of samples).
  • Biocompatibility (ISO 10993-1, -5, -12): "final and sterilized products." (No number of samples).
  • Sterilization and Shelf Life Validations: Not specified, but generally follow specific standards.

The data provenance is largely implied to be laboratory-based testing using in vitro models (e.g., artificial bone) and comparisons to existing devices, rather than in vivo or human data. There is no information regarding country of origin for the data other than the company being based in Israel. The studies are by nature prospective as they are specifically conducted to support the 510(k) submission.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

No direct mention of experts or their qualifications establishing "ground truth" for the test set is provided. The studies described are primarily engineering and laboratory performance tests against industry standards (e.g., ISO, AAMI) or in comparison to predicate devices, rather than clinical studies requiring expert consensus on outcomes.

4. Adjudication Method for the Test Set

Not applicable. The tests are primarily physical, chemical, and mechanical evaluations against defined standards and comparative performance with predicate devices, not subjective assessments requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

No MRMC comparative effectiveness study was done. The document describes laboratory and engineering tests, not clinical studies involving human readers or cases.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study Was Done

Not applicable. This is a physical medical device (dental implants and prostheses), not a software algorithm.

7. The Type of Ground Truth Used

The "ground truth" for the various studies is typically defined by:

  • Industry Standards: Compliance with international standards (e.g., ISO 14801, ISO 10993 series, ISO 11137-2, ISO 17665-1, ISO 11607-1).
  • Comparative Performance: The performance of legally marketed predicate devices or reference devices, establishing a benchmark for substantial equivalence.
  • Defined Benchmarks/Thresholds: For tests like sterilization success (SAL 10⁻⁶) or fatigue limits implied by ISO standards.
  • Qualitative Assessment: Such as "no additional damage" or "no incompatibility potential" demonstrated in material science or biological tests.

8. The Sample Size for the Training Set

Not applicable. There is no training set mentioned, as this is for a physical medical device clearance, not an AI/ML algorithm that requires training data.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set.

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.