(143 days)
HYPERION X5 is an extraoral X-ray system for digital panoramic X-rays suitable for production of orthopanoramic images of the maxillofacial region, diagnostic examination of the dentition (teeth), arches and other structures of the oral cavity.
Hyperion X5 allows the following projections:
- Standard or pediatric panoramic views (PAN);
- i Full or partial views of the dentition selected by the user (DENT);
- Frontal and lateral view of the maxillary sinuses (SIN); .
- Lateral and posteroanterior view of the temporomandibular joints (TMJ) from several angles. -
The HYPERION X5 shall not be used on patients (child) less than approximately 1045mm in height and less than 19Kg in weight. These height and weight measurements approximately correspond to that of an average 4 year old. Furthermore the device shall not be used with patients not vigilant and cooperative, since the patient must be able to understand and follow the operator's instructions for a correct positioning
The Hyperion X5 apparatus consists of one rotating arm attached to the support column. The powered rotating arm is capable of rotating for about 210° and translating thereby allowing the x-ray emission system and image detector to be moved around the patient according to complex orbits that follow the morphological profile. The rotating arm is attached to a support column capable of moving vertically through a motorized movement in order to allow it to perform the panoramic exposition.
The system provides patient positioning means such as handles and bite point with chin rest. The correct patient positioning is guided by 3 LASER positioning lights. The x-ray generator provides up to 85 kV, 15 mA to the x-ray tube, and the scanning process is performed by an independent axes cinematic unit. The system is equipped with CMOS solid state x-ray image detector, with associated electronics. The solid state CMOS detector contain a Cs(Tl) scintillator able to capture the radiation emitted by the x-ray generator, converting it in fluorescence and then in electric charge, accumulated by each pixel according to the light intensity. The accumulated charge in each row is sequentially selected by vertical shift registers for row scan, transferred to the amplifier and converted in voltage signal and than converted in a 14bit digital signal that transfered to the PC, trough the Ethernet connection, allow to create a set of by dimensional images acquired during the movement of the rotating arm along the designed orbits. These set of images suitably interconnected, allow at the software to create the panoramic standard image in order to show the exam result, store and evaluate it.
Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the Hyperion X5 device:
Important Note: The provided document is a 510(k) Summary, which focuses on demonstrating substantial equivalence to a predicate device, not necessarily providing a full, detailed clinical study with pre-defined acceptance criteria and statistical power calculations typically found in a clinical trial report. Therefore, some information requested (like specific numerical acceptance criteria for clinical outcomes, sample sizes for test/training sets with data provenance, and details about expert ground truth establishment for AI performance) is not explicitly present because the document describes a benchmarking/comparison study against a predicate device, rather than a standalone clinical efficacy study with strict performance metrics.
1. Table of Acceptance Criteria and Reported Device Performance
Given the nature of the 510(k) Summary (demonstrating substantial equivalence), the "acceptance criteria" are implicitly defined by the predicate device's performance and compliance with relevant standards. The "reported device performance" refers to how the Hyperion X5 compares to the predicate.
| Feature / Performance Metric | Acceptance Criteria (Implicitly, equivalent to Predicate) | Reported Device Performance (Hyperion X5) |
|---|---|---|
| Indications for Use | Subset of predicate's 2D indications | Meets subset of predicate's 2D indications |
| Panoramic Image Resolution | Comparable to predicate (Hyperion X9) using QUART Phantom | Comparable to predicate (Hyperion X9) |
| Geometrical Performance (Magnification, Orthogonality) | Comparable to predicate (Hyperion X9) using CEFLA Phantom | Comparable to predicate (Hyperion X9) |
| Image Information Level (Under/Over-Exposition) | Similar to predicate (Hyperion X9) in extreme conditions | Similar to predicate (Hyperion X9) |
| MTF (Modulation Transfer Function) | 60% @ 1lp/mm, 10% @ 4lp/mm (Predicate) | 58% @ 1lp/mm, 8% @ 4 lp/mm (Hyperion X5) |
| DQE (Detective Quantum Efficiency) | 80% @ 0lp/mm, 20% @ 3lp/mm (Predicate) | 70% @ 0lp/mm, 18% @ 3lp/mm (Hyperion X5) |
| Patient Population | Adult & pediatric | Adult & pediatric |
| Selectable Expositions | PAN, DENT, SIN, TMJ (2D mode) | PAN, DENT, SIN, TMJ |
| Compliance to Standards | IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 62366, 60601-2-63, 62304; FDA Guidance on Solid State X-ray Imaging Devices; FDA Guidance on MD SW | Compliant with all mentioned standards and guidances |
2. Sample Size for Test Set and Data Provenance
The document describes non-clinical bench tests using phantoms rather than a clinical test set with patient data.
- Sample Size for Test Set: Not applicable in the context of human patient data. The "test set" consists of phantom images. The number of images or runs with the phantoms is not specified.
- Data Provenance: Not applicable, as no human patient data was used. The tests were performed internally by the manufacturer (CEFLA S.C.).
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Number of Experts: Not applicable. As non-clinical bench tests with phantoms were used, the "ground truth" is based on the known physical properties and measurements of the phantoms themselves, not expert interpretation of clinical images.
- Qualifications of Experts: Not applicable.
4. Adjudication Method for the Test Set
- Adjudication Method: Not applicable. Bench tests with phantoms rely on objective measurements against known phantom characteristics, not subjective human adjudication of clinical findings.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Was an MRMC study done? No. The submission focuses on direct technical comparison using phantoms and compliance with standards. There is no mention of human readers evaluating images or a comparative effectiveness study with or without AI assistance.
- Effect Size: Not applicable, as no MRMC study was performed.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
- Was a standalone study done? Yes, in a sense. The described "non clinical performance tests" (comparison with QUART Technical Phantom, Geometrical Comparison with CEFLA Phantom, and comparison during under/over-exposition) assess the device's image acquisition and reconstruction capabilities independent of human interpretation or intervention for diagnostic tasks. These are technical performance evaluations of the imaging system.
7. Type of Ground Truth Used
- Type of Ground Truth: Phantom-based objective measurements. The "ground truth" for these non-clinical tests refers to the known characteristics and values designed into the QUART Technical Phantom (DIN 6868-5) and the Standard Technical CEFLA Phantom for evaluating resolution, magnification, and orthogonality.
8. Sample Size for the Training Set
- Sample Size for Training Set: Not applicable. The Hyperion X5 is an X-ray imaging system, not an AI or machine learning algorithm in the context of this submission. There is no mention of a "training set" for an AI component. The software mentioned (firmware, VKB, NNT/IRYS) controls the device and processes images, but the document does not describe it as an AI system requiring a training set in the modern sense.
9. How the Ground Truth for the Training Set Was Established
- How Ground Truth for Training Set Was Established: Not applicable, as no training set for an AI component is mentioned.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 24, 2015
Cefla S.c. % Maurizio Pantaleoni CEO Isemed Srl Via A. Bonetti 3/a Imola, 40026 IT
Re: K152162
Trade/Device Name: Hyperion X5 Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral Source X-Ray System Regulatory Class: Class II Product Code: MUH Dated: July 21, 2015 Received: August 3, 2015
Dear Maurizio Pantaleoni:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
{1}------------------------------------------------
Page 2 - Maurizio Pantaleoni
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael D'Hara
For
Robert Ochs Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known)
Device Name HYPERION X5
Indications for Use (Describe)
HYPERION X5 is an extraoral X-ray system for digital panoramic X-rays suitable for production of orthopanoramic images of the maxillofacial region, diagnostic examination of the dentition (teeth), arches and other structures of the oral cavity.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
510(k) Summary of the HYPERION X5
This 510(k) Summary is being submitted as required by 21 CFR 807.92.
| 1. | General Information | |
|---|---|---|
| Submitter : | CEFLA S.C. | |
| Via Selice Prov.le 23/a | ||
| Imola, BO 40026 ITALY | ||
| Tel. +39 0542 653441 | ||
| Fax +39 0542 653607 | ||
| Consultant/ Contact: | Maurizio Pantaleoni | |
| ISEMED srl | ||
| Via Altobelli Bonetti 3/a | ||
| Imola, BO 40026 ITALY | ||
| Mob. +39 348 4435155 | ||
| Tel. +39 0542 683803 | ||
| Fax +39 0542 698456 | ||
| Email: regulatory@isemed.eu | ||
| Summary Prepared Date: | July 21, 2015 |
2. Names
Device Name: HYPERION X5 Common Name: Panoramic x-ray system Product Code: MUH Classification: ll Regulation Number: 872.1800 (Extraoral source x-ray system)
3. Predicate Devices
The HYPERION X5 is substantially equivalent to the following device:
| Applicant | Device name | 510(k) Number |
|---|---|---|
| CEFLA S.C. | HYPERION X9 | K123381 |
{4}------------------------------------------------
4. Device Description
The HYPERION X5 is manufactured by Cefla S.C. but it can be sold under two different brands and commercial name Hyperion X5 and NewTom GO. The device can be set in three different installation configurations: Suspended version, Floor version with column and wall bracket and Floor version with column and standard base.
The Hyperion X5 apparatus consists of one rotating arm attached to the support column. The powered rotating arm is capable of rotating for about 210° and translating thereby allowing the x-ray emission system and image detector to be moved around the patient according to complex orbits that follow the morphological profile. The rotating arm is attached to a support column capable of moving vertically through a motorized movement in order to allow it to perform the panoramic exposition.
The system provides patient positioning means such as handles and bite point with chin rest. The correct patient positioning is guided by 3 LASER positioning lights. The x-ray generator provides up to 85 kV, 15 mA to the x-ray tube, and the scanning process is performed by an independent axes cinematic unit. The system is equipped with CMOS solid state x-ray image detector, with associated electronics. The solid state CMOS detector contain a Cs(Tl) scintillator able to capture the radiation emitted by the x-ray generator, converting it in fluorescence and then in electric charge, accumulated by each pixel according to the light intensity. The accumulated charge in each row is sequentially selected by vertical shift registers for row scan, transferred to the amplifier and converted in voltage signal and than converted in a 14bit digital signal that transfered to the PC, trough the Ethernet connection, allow to create a set of by dimensional images acquired during the movement of the rotating arm along the designed orbits. These set of images suitably interconnected, allow at the software to create the panoramic standard image in order to show the exam result, store and evaluate it.
The HYPERION X5 system has been developed according to the FDA Guidance "Solid State X-ray Imaging Devices - Premarket Notification 510(k) submissions" of August 6, 1999.
5. Indications for Use
HYPERION X5 is an extraoral X-ray system for digital panoramic X-rays suitable for production of orthopanoramic images of the maxillofacial region, diagnostic examination of the dentition (teeth), arches and other structures of the oral cavity.
Hyperion X5 allows the following projections:
- Standard or pediatric panoramic views (PAN);
- i Full or partial views of the dentition selected by the user (DENT);
- Frontal and lateral view of the maxillary sinuses (SIN); .
- Lateral and posteroanterior view of the temporomandibular joints (TMJ) from several angles. -
The HYPERION X5 shall not be used on patients (child) less than approximately 1045mm in height and less than 19Kg in weight. These height and weight measurements approximately correspond to that of an average 4 year old. Furthermore the device shall not be used with patients not vigilant and cooperative, since the patient must be able to understand and follow the operator's instructions for a correct positioning
The Hyperion X5 can perform panoramic views, both for adult and child, frontal and lateral view of the maxillary sinuses and Lateral and posteroanterior view of the temporomandibular joints exactly as the predicate device, there are no differences in indications for use of the two devices
The Indication for use for the HYPERION X5 are a subset of the indications for use of the predicate device only related to the 2D mode applications, since HYPERION X5 is a simplified version of the HYPERION X9.
{5}------------------------------------------------
6. Summary of comparison discussion
| Proposed Device | Predicate Devices | |
|---|---|---|
| Device Name | HYPERION X5CEFLA S.C. | HYPERION X9CEFLA S.C.(K123381) |
| Figure | Image: HYPERION X5 | Image: HYPERION X9 |
| Classification & Intended use | ||
| Regulation Number | 21CFR 872.1800 | 21CFR 892.1750 |
| Regulation Name | Extraoral source x-ray system | Computed tomography x-ray system |
| Classification Product Code | MUH | OAS (Classification Product code )MUH (Subsequent Product code) |
| Regulatory Class | Class II | Class II |
| Indication for use statement | HYPERION X5 is an extraoral X-ray systemfor digital panoramic X-rays suitable forproduction of orthopanoramic images ofthe maxillofacial region, diagnosticexamination of the dentition (teeth),arches and other structures of the oralcavity. | HYPERION X9 is a digital panoramic, cephalometric andtomographic extra-oral X-ray system, indicated for usein:(i) producing panoramic X-ray images of themaxillofacial area, for diagnostic examination ofdentition (teeth), jaws and oral structures; and(ii) producing radiographs of jaws, parts of the skull andcarpus for the purpose of cephalometric examination,when equipped with tele-radiographic arm (CEPH);(iii) producing tomographic images of the oral andmaxillofacial structure, for diagnostic examination ofdentition (teeth), jaws, oral structures and some cranialbones if equipped with CBCT option. |
| Performances features comparison | ||
| Performance Specification | Panoramic | Panoramic. Cephalometric and Computed tomography |
| Patient population | Adult & paediatric | Adult & paediatric |
| Exposition selectable | PAN, DENT, SIN, TMJ | PAN, DENT, SIN, TMJ (2D mode) |
| Patient type selectable | Male, female, child | Male, female, child |
| Patient size selectable | Large, medium, small for male an female patient | Large, medium, small for male an female patient |
| Dose reduction forchild/adult (medium)exposition | 0,7 | 0,6 |
| Technical & Functional features comparison | ||
| X-ray emission | ||
| X-RAY GENERATORVOLTAGE | 60 - 85 kV, manually or automaticallyselectable, in steps of 1 kV | 2D mode: 60 - 85 kV, automatic and manuallyselectable image acquisition in steps of 1 in the kVrange.CBCT MODE: 60 - 90 kV, manually or automaticallyselectable, in steps of 1 kV |
| ANODIC CURRENT | 4 - 15 mA selectable, in steps of 1 mArange. | 1 – 10 mA, selectable, in steps of 1 mA range. |
| EXPOSURE TIME MAX(STANDARD PAN) | 1s - 15s continuous radiation(depending on the 2D examination typeselected) | 1- 14s, continued exposure |
| SHAPE OF X-RAY BEAM | Fan-shaped x-ray beam | Fan-shaped x-ray beam and cone beam for tomographyx-ray |
| TYPE OF X-RAY EMISSION | Continuous for PAN exams | Continuous for PAN and CEPH exams and pulsed for CBCTexams |
| EXPOSURE TIME(STANDARD CEPH) | N/A | 4.6 s, continued exposure |
| EXPOSURE TIME(CBCT) | N/A | 3.6 - 9 s, effective radiation time |
| STANDARDS | IEC60601-2-63 | IEC60601-2-63 |
| SSD Detector & IMAGE acquisition | ||
| • PANORAMICIMAGE DETECTOR | Digital CMOS linear detector | Digital CCD linear detector |
| • SCINTILLATOR MATERIAL | CSI | CSI |
| • MTF | 58% @ 1lp/mm8% @ 4 lp/mm | 60% @ 1lp/mm (1x1)10% @ 4lp/mm (1x1) |
| • DQE | 70% @0lp/mm18% @3lp/mm | 80% @ 0lp/mm (1x1)20% @3lp/mm (1x1) |
| • CEPHALOMETRIC• IMAGE DETECTOR | N/A | Digital CCD linear detector, Repluggable for panoramicexposure technique |
| • CBCT EXAMS• FLAT PANEL IMAGEDETECTOR | N/A | Amorphous Silicon flat panel X-ray detector(VARIAN manufacturer) |
| • IMAGE DETECTORSDIMENSION | PAN: 6 x 150 mmCEPH: N/ACBCT: N/A | PAN: 6 x 146 mmCEPH: 6 x 220 mmCBCT: 80x130, 130x130mm, 150x150 mm |
| • DETECTOR PIXEL SIZE: | PAN: 100 μmCEPH: N/ACBCT: N/A | PAN: 48 μmCEPH: 48 μmCBCT: 127 μm |
| • SOURCE TO IMAGEDETECTOR DISTANCE (SID) | PAN: 500 mmCEPH: N/ACBCT: N/A | PAN: 550mmCEPH:1500mmCBCT:660mm |
| • PANORAMIC IMAGEACQUISITION PATH | 2 axis - 210° | 3 axis - 360° |
| • IMAGE TRANSFER TO PC | Direct - Giga Ethernet | Direct - Giga Ethernet for CBCT |
| Laser & positioning | ||
| • NUMBER OF LASER POINTE | 3 lasers pointers | 3 lasers pointers |
| • LASER OPTICAL CLASS | Class 1 for IEC 60825-1: 2003 | Class 1 for IEC 60825-1: 2003 |
| Control & Viewing Software | ||
| • CONTROL SW | Firmware + VKB (on PC) | Firmware |
| • VIEWING SOFTWARE | NNT / IRYS(optional) | NNT / IRYS(optional) |
| • SOFTWARE VALDATION | IEC62304+ Guidance FDA on MD SW | IEC62304+ Guidance FDA on MD SW |
{6}------------------------------------------------
{7}------------------------------------------------
The HYPERION X5 is regulated under the 21 CFR 872.1800 as "Extraoral source x-ray system" with product code MUH, these regulation number and product code are different from the regulation number and classification product code of HYPERION X9 (K123381) that is regulated under 21 CFR892.1750 "Computed tomography x-ray system" with classification product code OAS , but the product code of the Hyperion X5 is equal to the subsequent product code of the predicate device (MUH), since the Hyperion X5 is a simplified version of the HYPERION X9 (K123381) able to perform only panoramic function instead of panoramic, cephalometric and computed tomography functions, performed by Hyperion X9.
Since HYPERION X5 is a simplified version of the HYPERION X9, it performs only panoramic X-ray in 2D mode and it did not produce other radiographic exposition as Cephalometric and Computed tomography. For this reason, the indications for use of the HYPERION X5 are a subset of the indications for use of the predicate device only related to the 2D mode applications.
Both the predicate device, are indicated to be used on adult and paediatric population, in both the machines, is possible select PAN, DENT, SIN, TMJ expositions selecting the kind of patients between Male, female and child, and in case of male and female (adults) is possible also select the size between large, medium and small. In both machines the child exposition is unique and no further subdivision of size are allowed.
About X-ray emission the X-ray tube head voltage for the Hyperion X5 is selectable in steps of 1 kV from 60 to 85 kV as in the predicate device when it works in 2D mode, while 5 the anodic current is selectable from 4 to 15mA, instead of the range of the anodic current selectable for the HYPERION X9 (K123381) that is from 1 to 10mA. The other features related to the X-ray emission, as the X-ray beam and the type of emission are respectively fan-shaped and continuous emission, both for Hyperion X5 that for predicate device. Since the X-ray emission has been validated in compliance with the same standard (IEC60601-2-63) both for Hyperion X5 that for Hyperion X9 (K123381) and that the effectiveness of the x-ray emission together with the SSD Detector and the image acquisition process has been compared in the specific comparative performance bench tests described in the section 7 "Performance data, these differences are not considered relevant for the substantial equivalence of the two devices
About SSD Detector & Image acquisition, the HYPERION X5 uses a linear CMOS detector made of Cs| (Cesiume lodide) material with a pixel dimension of 100 um instead of the digital CCD sensor made of Csl used by the predicate device, but with a pixel dimension of 48 µm. The MTF value is very similar between the two sensor as the DQE. The path used to acquire the image, is different between the subject device and the predicate because the Hyperion X5 moves the rotating arm with a trajectory designed on 2 axis with a 210-degree rotation of the rotational arm, instead of the three axis with 360-degree rotation of the rotational arm, used by the Hyperion X9.
All these differences are not considered relevant for the substantial equivalence of the two devices because the images produced by the two sensors are considered comparable in term of technical features of the images (comparable resolution, distortion performances and with similar information in term of gray level, also in extreme exposition conditions) and also for the more significant clinical aspects required by a dentist
{8}------------------------------------------------
to this kind of device. These considerations are supported by the specific comparative performance bench tests described in the section 7 "Performance data".
As the predicate devices HYPERION X9 (K123381), the HYPERION X5 uses luminous laser traces to guide patient positioning. These traces are projected by the identical class I laser beamers.
The viewing and control software used on the subject device is substantially equivalent to the one used for the predicate device, "Hyperion X9 (K123381)", unless for the user interface used by the dentist to select the process parameters, that for the subject device has been developed with a specific Virtual Keyboard module that runs on a PC, and not directly installed into the firmware of the device.
The firmware and the software (VKB on PC) of the Hyperion X5, for the PAN, DENT, SIN, TMJ expositions, performs the same functions and algorithms of the predicate device, omitting all functions not related to the intended use as ie. CBCT exposition. Furthermore both the software has been validate in compliance to the IEC62304 and FDA Guidance on MD SW.
7. Performance Data
Since the "Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices" require to provide samples of test pattern images (SSXI and predicate) that show the device to provide images equivalent to those of a predicate device(s), CEFLA has performed some non clinical bench tests in order to compare the performances of the subject device with the Hyperion X9 (K123381). Furthermore
The non clinical performance tests executed are:
-
- Comparison with QUART Technical Phantom (DIN 6868-5) on X9 and X5 Standard Panoramic X-rays. This test allow to demonstrate that both Hyperion X5 that the predicate device, together with their sensors and background software, are able to produce images with comparable level of resolution
-
- Geometrical Comparison with Standard Technical CEFLA Phantom on 2D Dental Panoramic Projections. This test allow to demonstrate that both Hyperion X5 that the predicate device are able to produce images with comparable magnification of vertical and horizontal lines and orthogonality errors)
-
- Comparison between Hyperion X5 and X9 Panoramic Under- and Over-Exposed X-ray. This test allow to evaluate the behaviour of the two devices, and in particular the behaviour of the two 2D sensors and of the two reconstruction software used in the subject device and in the predicate device, when a PANORAMIC exposition is performed in extreme exposition (under and over exposition).
Furthermore the Hyperion X5 is designed according to the FDA "Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices" and also designed, tested and found to be compliant with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6 and 62366, 60601-2-63 and 62304, exactly as its predicate devices
The non clinical performance tests allow to demonstrate that the images produced by the Hyperion X5 have similar resolution, similar geometrical performances (magnification of vertical lines and orthogonality errors), similar ability to maintain information level, also in very extreme exposition conditions, and are also comparable in term of clinical outcome.
For these reasons the results of the non clinical performance together with the compliance to the same standards allow to demonstrate the substantial equivalence of the Hyperion X5 and the predicate device.
Conclusions
Do to all the above described reasons, we affirm that Hyperion X5 is substantially equivalent to the predicate device since it can be consider as safety and effectiveness as the predicate device.
§ 872.1800 Extraoral source x-ray system.
(a)
Identification. An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.(b)
Classification. Class II.