K Number
K152047
Date Cleared
2015-12-23

(153 days)

Product Code
Regulation Number
888.3660
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The INTEGRA® TITAN™ Modular Total Shoulder System Fin-Lock™ Glenoid is intended for use in Total Shoulder Arthroplasty which is indicated for:

  1. Severely painful and/or disabled joint resulting from osteoarthritis, traumatic arthritis or rheumatoid arthritis.
  2. Fracture dislocations of the proximal humerus where the articular surface is severely comminuted, separated from its blood supply or where the surgeon's experience indicated that alternative methods of treatment are unsatisfactory.
  3. Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g. – revision of a failed primary component)
    The humeral component is intended for cemented or uncemented use.
    The glenoid component is intended for cemented use only.
Device Description

The Integra TITAN™ Modular Total Shoulder System Fin-Lock™ Glenoid is a fully cemented all highly cross-linked polyethylene glenoid used in total shoulder arthroplasty for resurfacing the glenoid fossa. The glenoid component has one central peg with barbs and three peripheral pegs. The articulating surface of the glenoid device will be the same dimensions and sizes as existing Titan Modular Total Shoulder System three peg inline glenoids for consistent mismatching. The Fin-Lock™ Glenoid is intended to be a part of the TITAN™ Modular Shoulder System, 2.5 (K142413) and be used with the humeral component.

AI/ML Overview

Based on the provided text, the acceptance criteria and the study results for the INTEGRA® TITAN™ Modular Total Shoulder System Fin-Lock™ Glenoid are described in the "807.92(b)(1-2) – Nonclinical Tests Submitted" section.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The document states that "The results of these performance tests met their respective acceptance criteria." However, it does not explicitly list the quantitative acceptance criteria for each test. It only mentions that the device met them.

Acceptance CriteriaReported Device Performance
Not explicitly stated in the documentMet the respective acceptance criteria for each test.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the sample sizes used for the test sets in the performance tests. It also does not mention the data provenance (e.g., country of origin, retrospective or prospective nature). The tests are described as "nonclinical," implying they were conducted in a laboratory setting rather than using patient data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This information is not applicable and not provided. The tests described are "nonclinical" performance tests (e.g., mechanical testing, material characterization), which typically do not involve human experts establishing ground truth in the way clinical studies or diagnostic AI algorithms do.

4. Adjudication Method for the Test Set

This information is not applicable and not provided. As mentioned above, these are nonclinical performance tests, not clinical studies requiring adjudication of outcomes by experts.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not done. The document describes nonclinical performance tests, not studies involving human readers or AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

This is not applicable. The device is a physical medical implant (shoulder prosthesis), not an algorithm or AI. The tests performed are for the physical properties and performance of the implant itself.

7. The Type of Ground Truth Used

The ground truth for these nonclinical tests would be based on engineering specifications, material science standards, and mechanical testing protocols. For example, for a "Glenoid Pull-out Fixation Test," the ground truth would be a defined force threshold or displacement limit that the device must withstand according to established engineering principles for such implants. The document implies these standards were used by stating "met their respective acceptance criteria."

8. The Sample Size for the Training Set

This is not applicable. The device is a physical medical device. There is no "training set" in the context of an AI/algorithm. The development and testing of a physical implant involve design iterations, material selection, and extensive pre-clinical testing, but not an AI-style training set.

9. How the Ground Truth for the Training Set was Established

This is not applicable as there is no training set for a physical medical device. The "ground truth" for the device's design and manufacturing would be based on biomaterial science, biomechanical engineering principles, clinical needs, existing standards for orthopedic implants, and the performance characteristics of predicate devices.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol featuring three stylized human profiles facing to the right, overlapping to create a sense of unity and connection. The symbol is meant to represent the department's focus on health and human well-being.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 23, 2015

Integra LifeSciences Corporation Mr. William Garzon Senior Regulatory Affairs Specialist 311 Enterprise Drive Plainsboro, New Jersey 08536

Re: K152047

Trade/Device Name: INTEGRA® TITAN™ Modular Total Shoulder System Fin-Lock™ Glenoid Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: KWS Dated: November 24, 2015 Received: November 27, 2015

Dear Mr. Garzon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Casey Hanley -S

  • Mark Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
    Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K152047

Device Name

INTEGRA® TITAN™M Modular Total Shoulder System Fin-Lock™ Glenoid

Indications for Use (Describe)

The INTEGRA® TITAN™ Modular System Fin-Lock™ Glenoid is intended for use in Total Shoulder Arthroplasty which is indicated for:

  1. Severely painful and/or disabled joint resulting from osteoarthritis or rheumatoid arthritis.

  2. Fracture-dislocations of the proximal humerus where the articular surface is severely comminuted, separated from its blood supply or where the surgeon's experience indicated that alternative methods of treatment are unsatisfactory.

  3. Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g. – revision of a failed primary component)

The humeral component is intended for cemented or uncemented use.

The glenoid component is intended for cemented use only.

Type of Use (Select one or both, as applicable)

☒ Transportation Use in CFR 229.301(a)☐ For Other Use in CFR 229.301(a)
---------------------------------------------------------------------------------------------------------------------------

|X | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

807.92(a)(1) – Submitter Information
NameIntegra LifeSciences Corporation
Address311 Enterprise DrivePlainsboro NJ 08536
Name of Contact PersonWilliam Garzon
Phone Number(512) 596-8908
Fax Number(512) 836-6933
EstablishmentRegistration Number1651501
Date PreparedDecember 22, 2015
807.92(a)(2) – Name of device
Trade or Propriety NameINTEGRA® TITAN™ Modular Total Shoulder System Fin-Lock™ Glenoid
Common or Usual NameShoulder Joint Metal/Polymer Semi-Constrained CementedProsthesis
Classification NameProstheses, Shoulder, Semi-Constrained, Metal/PolymerCemented
Classification PanelOrthopedic
RegulationClass II (under 21CFR 888.3660)
Product CodeKWS
807.92(a)(3) - Legally marketed device(s) to which equivalence is claimed
Ascension® TITAN™ Modular Total Shoulder System Glenoid (K100448) –Primary Predicate
DePuy Anchor Peg Glenoid (K052472) – Reference Device
Tornier Affiniti Glenoid (K103007) – Reference Device
807.92(a)(4) - Device description
The Integra TITAN™ Modular Total Shoulder System Fin-Lock™ Glenoid is a fullycemented all highly cross-linked polyethylene glenoid used in total shoulder arthroplastyfor resurfacing the glenoid fossa. The glenoid component has one central peg with barbsand three peripheral pegs. The articulating surface of the glenoid device will be the samedimensions and sizes as existing Titan Modular Total Shoulder System three peg inlineglenoids for consistent mismatching. The Fin-Lock™ Glenoid is intended to be a part ofthe TITAN™ Modular Shoulder System, 2.5 (K142413) and be used with the humeralcomponent.
807.92(a)(5) – Intended Use of the device
Indications for UseThe INTEGRA® TITAN™ Modular Total Shoulder System Fin-Lock™ Glenoid is intended for use in Total Shoulder Arthroplastywhich is indicated for:1. Severely painful and/or disabled joint resulting fromosteoarthritis, traumatic arthritis or rheumatoid arthritis.2. Fracture dislocations of the proximal humerus where the
articular surface is severely comminuted, separated from itsblood supply or where the surgeon's experience indicatedthat alternative methods of treatment are unsatisfactory.3. Other difficult clinical problems where shoulderarthrodesis or resection arthroplasty are not acceptable (e.g.– revision of a failed primary component)The humeral component is intended for cemented or uncementeduse.The glenoid component is intended for cemented use only.

510(k) Summary

{4}------------------------------------------------

807.92(a)(6) Summary of the technological characteristics of the device compared to the predicate

The INTEGRA® TITAN™ Modular Total Shoulder System Fin-Lock™ Glenoid is similar in design and materials to the predicate device, Ascension® TITAN™Modular Total Shoulder System (K100448). The INTEGRA® TITAN™ Modular Total Shoulder System Fin-Lock™ Glenoid has similar indications for use, intended use and fundamental scientific technology as its predicate.

The INTEGRA® TITAN™ Modular Total Shoulder System Fin-Lock™ Glenoid is similar in central peg length to the predicate devices, DePuy Anchor Peg Glenoid (K052472) and Tornier Affiniti Glenoid (K103007).

The differences between the predicates and proposed device do not raise any new issues regarding safety and effectiveness; therefore, the INTEGRA® TITAN™M Modular Total Shoulder System Fin-Lock™ Glenoid is considered substantially equivalent to the predicate device.

807.92(b)(1-2) – Nonclinical Tests Submitted

Testing to verify the performance of the INTEGRA® TITAN™ Modular Total Shoulder System Fin-lock™ Glenoid included the following:

  • Rocking and Subluxation Characterization Test ●
  • Glenoid Pull-out Fixation Test ●
  • Extensively Cross-Linked UHMWPE Material Characterization Test ●
  • Extensively Cross-Linked UHMWPE Wear Test

The results of these performance tests met their respective acceptance criteria and demonstrate that the INTEGRA® TITAN™ Modular Total Shoulder System Fin-Lock™ Glenoid is substantially equivalent to the predicate devices identified.

807.92(b)(3) - Conclusions drawn from non-clinical data

The design features, materials, intended use, and overall fundamental scientific technology of the INTEGRA® TITAN™ Modular Total Shoulder System are substantially equivalent to the predicate device. The safety and effectiveness of the INTEGRA® TITAN™ Modular Total Shoulder System is adequately supported by the substantial equivalence information, materials information, and performance data provided within this Premarket Notification submission.

§ 888.3660 Shoulder joint metal/polymer semi-constrained cemented prosthesis.

(a)
Identification. A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).(b)
Classification. Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”