K Number
K152001

Validate with FDA (Live)

Date Cleared
2015-12-10

(143 days)

Product Code
Regulation Number
876.4400
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sterile Repositionable Hemostasis Clipping Device is indicated for endoscopic clip placement within the gastrointestinal tract for the purpose of:
(1) endoscopic marking,
(2) hemostasis for
(a) mucosal / sub-mucosal defects < 3cm,
(b) bleeding ulcers,
(c) polyps < 1.5cm in diameter,
(d) diverticula in the colon,
(3) as a supplementary method, closure of Gl tract luminal perforations <20mm that can be treated conservatively.

Device Description

The proposed device Sterile Repositionable Hemostasis Clipping Device is a sterile, single-use endoscopic clipping device, intended to be used for endoscopic marking, hemostasis for mucosal/submucosal defects in digestive tract. It consists of two main components, delivery system and clip assembly. And it is offered in different dimensions.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device, the "Sterile Repositionable Hemostasis Clipping Device". This document is submitted to the FDA to demonstrate substantial equivalence to a predicate device, not typically to prove a device meets specific clinical performance acceptance criteria through a study with ground truth. Therefore, many of the requested categories for AI/algorithm performance studies are not directly applicable.

Here's an analysis based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state quantitative acceptance criteria or a direct performance table in the typical sense of algorithm-based device performance (e.g., sensitivity, specificity). Instead, it lists bench tests performed to demonstrate equivalence to a predicate device. The general acceptance criterion is that the device "meets the requirements" of relevant ISO standards and that "the testing performed demonstrated that the proposed and predicate delivery systems are equivalent."

Test CategorySpecific Tests PerformedReported Device Performance
Biological EvaluationISO 10993-1 (Evaluation and Testing)Meets requirements
ISO 10993-7 (Ethylene oxide sterilization residuals)Meets requirements
SterilizationISO 11135-1 (Ethylene Oxide Sterilization)Meets requirements
Bench Tests - Mechanical IntegrityDimensional verificationDemonstrated equivalence to predicate delivery systems
Mechanical Integrity of Clip AssemblyDemonstrated equivalence to predicate delivery systems
Clamping Strength TestingDemonstrated equivalence to predicate delivery systems
Tensile Strength TestingDemonstrated equivalence to predicate delivery systems
Release Force TestingDemonstrated equivalence to predicate delivery systems
Rotation TestingDemonstrated equivalence to predicate delivery systems

2. Sample size used for the test set and the data provenance

Not applicable. This is a medical device for physical intervention, not an AI/algorithm-based diagnostic or prognostic device that uses a test set of data. The "tests" mentioned are bench tests on the physical device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This device does not rely on expert-established ground truth for its performance evaluation in the context of this 510(k) submission.

4. Adjudication method for the test set

Not applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No. The document explicitly states: "No clinical study is included in this submission." This is not an AI-assisted device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No. This is a physical hemostasis clipping device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

Not applicable in the context of AI/algorithm performance. For the bench tests, the "ground truth" would be the engineering specifications and performance characteristics of the predicate device and the relevant ISO standards.

8. The sample size for the training set

Not applicable. This is not an AI/algorithm device.

9. How the ground truth for the training set was established

Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of three human profiles facing right, arranged in a cascading manner. The profiles are connected and appear to be emerging from a single form. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the graphic.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 10, 2015

Micro-Tech (Nanjing) CO., Ltd. Becky Li Manager of Quality Department No. 10 Gaoke Third Road Nanjing, 210032 China

Re: K152001

Trade/Device Name: Sterile Repositionable Hemostasis Clipping Device Regulation Number: 21 CFR §876.4400 Regulation Name: Hemorrhoidal ligator Regulatory Class: II Product Code: PKL Dated: November 10, 2015 Received: November 12, 2015

Dear Becky Li,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Image /page/1/Picture/8 description: The image shows the name "Herbert P. Lerner -S" in a large, bold font. The text is black and is set against a white background. The letters are evenly spaced and easy to read. The image is simple and straightforward, with the focus on the name.

Benjamin R. Fisher, Ph.D. for Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K152001

Device Name

Sterile Repositionable Hemostasis Clipping Device

Indications for Use (Describe)

The Sterile Repositionable Hemostasis Clipping Device is indicated for endoscopic clip placement within the gastrointestinal tract for the purpose of:

(1) endoscopic marking,

(2) hemostasis for

(a) mucosal / sub-mucosal defects < 3cm,

(b) bleeding ulcers,

(c) polyps < 1.5cm in diameter,

(d) diverticula in the colon,

(3) as a supplementary method, closure of Gl tract luminal perforations <20mm that can be treated conservatively.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Micro-Tech. The logo consists of the letters "MT" in a stylized, three-dimensional font on the left side. To the right of the letters is the text "MICRO-TECH" in a sans-serif font, with a line underneath. Below the line are two Chinese characters.

510K Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92. The assigned 510(k) Number: K152001

1. Date of Preparation: 12/08/2015

2. Sponsor Identification

Micro-Tech (Nanjing) Co., Ltd.

No.10 Gaoke Third Road, Nanjing National Hi-Tech, Industrial Development Zone,

Nanjing, Jiangsu Province, PRC

Establishment Registration Number: 3004837686

Contact Person: Becky Li

Position: Manager of Quality Department

Tel: +86-25-58646378

Fax: +86-25-58744269

Email: In@micro-tech.com.cn

3. Identification of Proposed Device

Trade Name: Sterile Repositionable Hemostasis Clipping Device

Common Name: Hemostasis Clipping Device

Regulatory Information

Classification Name: Hemostatic Metal Clip for the GI Tract

Classification: 2

Product Code: PKL

Regulation Number: 876.4400

Review Panel: Gastroenterology/Urology

Intended Use Statement:

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image contains the logo for MICRO-TECH. The logo consists of the letters "MT" in a stylized, three-dimensional blue font. To the right of the letters, the words "MICRO-TECH" are written in a simple, sans-serif font. Below the words "MICRO-TECH", there are two Chinese characters.

This device is intended to be used for endoscopic marking, hemostasis for mucosal/submucosal defects in digestive tract.

4. Identification of Predicate Device

510(k) Number: K122660

Product Name: Resolution™ Hemomstasis Clipping Device

5. Indications for Use

The Sterile Repositionable Hemostasis Clipping Device is indicated for endoscopic clip placement within the gastrointestinal tract for the purpose of:

  • (1) endoscopic marking,
  • (2) hemostasis for
    • (a) mucosal / sub-mucosal defects < 3cm,
    • (b) bleeding ulcers,
    • (c) polyps < 1.5cm in diameter,
    • (d) diverticula in the colon,
  • (3) as a supplementary method, closure of GI tract luminal perforations <20mm that can be treated conservatively

6. Device Description

The proposed device Sterile Repositionable Hemostasis Clipping Device is a sterile, single-use endoscopic clipping device, intended to be used for endoscopic marking, hemostasis for mucosal/submucosal defects in digestive tract. It consists of two main components, delivery system and clip assembly. And it is offered in different dimensions.

7. Comparison of Technological Characteristics

The Sterile Repositionable Hemostasis Clipping Device incorporates

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the text "510K Summary". The text is in a simple, sans-serif font and is left-aligned. The background is plain white, which makes the text stand out.

Image /page/5/Picture/1 description: The image shows the logo for Micro-Tech. The logo consists of the letters "MT" in a stylized, three-dimensional font, with the words "MICRO-TECH" in a smaller, sans-serif font to the right. Below the words "MICRO-TECH" are two Chinese characters.

substantially equivalent device materials, design, configuration, packaging fundamental technology, manufacturing processes, sterilization process and

intended use as those featured in the Boston Scientific predicate devices.

Comparison to predicate Devices:

ItemProposed DeviceSterile Repositionable HemostasisClipping DeviceComparison toPredicateDevices
Product CodePKLSimilar
Regulation No.876.4400Same
Class2Same
Supplied SterileYesSame
ConfigurationDelivery system and clip assemblySame
Open width11mmSame
Minimal workingchannel2.8mmSame
Working Length1650mm, 1950mm, 2350mm, 2700mmSimilar
Indications for UseThe Sterile Repositionable Hemostasis ClippingDevice is indicated for endoscopic clipplacement within the gastrointestinal tract forthe purpose of:(1) endoscopic marking,(2) hemostasis for(a) mucosal / sub-mucosal defects < 3cm,(b) bleeding ulcers,(c) polyps < 1.5cm in diameter,(d) diverticula in the colon,(3) as a supplementary method, closure of GItract luminal perforations <20mm that can betreated conservativelySimilar
Single UseYesSame
PackagingSingle-use EO sterilized pouch with one deviceper pouchSimilar
Shelf LifeThree yearsSame

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for MICRO-TECH. The logo consists of a blue 3D block letter "MT" on the left, followed by the text "MICRO-TECH" in blue on the right. Below the text "MICRO-TECH" are two Chinese characters, also in blue.

8. Performance Data

The proposed device the Sterile Repositionable Hemostasis Clipping Device meets the requirements of ISO 10993 "Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing", ISO 11135-1 "Sterilization of Health Care products - Ethylene Oxide - Part 1: Requirements for Development, Validation, and Routine Control of Sterilization processes for Medical Devices", and ISO 10993-7 "Biological evaluation of medical devices - Part 7: ethylene oxide sterilization residuals",

The following bench tests were performed on the Sterile Repositionable

Hemostasis Clipping Device:

Dimensional verificationMechanical Integrity of Clip Assembly
Clamping Strength TestingTensile Strength Testing
Release Force TestingRotation Testing

The testing performed demonstrated that the proposed and predicate delivery systems are equivalent.

9. Clinical Test Conclusion

No clinical study is included in this submission.

10. Substantially Equivalent (SE) Conclusion

Based on the indications for use, technological characteristics, and safety and performance testing, the Sterile Repositionable Hemostasis Clipping Device has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the currently cleared the Resolution™ Hemostasis Clipping Device (K122660) .

§ 876.4400 Hemorrhoidal ligator.

(a)
Identification. A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.(b)
Classification. Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.