Intended Use

The Thora-3Di is intended for a one-time measurement of respiratory rate as part of a vital signs assessment. The device is indicated for hospital or clinical use in adult patients. The device is intended to be operated by clinicians and medically qualified personnel. It is available for sale only upon the order of a physician or licensed health care provider.

The Thora-3Di is not intended to monitor vital signs. This device is not an apnea monitor.

Device Description

The Thora-3Di uses Structured Light Plethysmography (SLP) to derive respiratory rate. With the Thora-3Di, a structured grid pattern of light is projected onto the thoracoabdominal wall and two digital video cameras record changes in the grid pattern due to movement of the wall. A dedicated parallel processor executes real-time pattern recognition algorithms to identify the position of each part of the virtual grid in each camera. The system computes the 3-dimensional coordinates of the different markers to assess respiratory rate.

The Thora-3Di does not include any wireless transmission features, nor does it include a cloud or other remote functionality (no cybersecurity-related features).

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Mean Squared Difference (MSD) between Thora-3Di and Capnography (Clinician Overscored CO2)

§ 868.2375 Breathing frequency monitor.

(a)
Identification. A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.(b)
Classification. Class II (performance standards).