K Number
K151940
Manufacturer
Date Cleared
2016-03-10

(240 days)

Product Code
Regulation Number
868.2375
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Thora-3Di is intended for a one-time measurement of respiratory rate as part of a vital signs assessment. The device is indicated for hospital or clinical use in adult patients. The device is intended to be operated by clinicians and medically qualified personnel. It is available for sale only upon the order of a physician or licensed health care provider.

The Thora-3Di is not intended to monitor vital signs. This device is not an apnea monitor.

Device Description

The Thora-3Di uses Structured Light Plethysmography (SLP) to derive respiratory rate. With the Thora-3Di, a structured grid pattern of light is projected onto the thoracoabdominal wall and two digital video cameras record changes in the grid pattern due to movement of the wall. A dedicated parallel processor executes real-time pattern recognition algorithms to identify the position of each part of the virtual grid in each camera. The system computes the 3-dimensional coordinates of the different markers to assess respiratory rate.

The Thora-3Di does not include any wireless transmission features, nor does it include a cloud or other remote functionality (no cybersecurity-related features).

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Mean Squared Difference (MSD) between Thora-3Di and Capnography (Clinician Overscored CO2) < 0.81 brpm (corresponding to limits of agreement of +/- 2 brpm)MSD = 0.136 brpm
95% Limits of Agreement between Thora-3Di and Capnography (Clinician Overscored CO2)(-0.72, 0.75 brpm)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 35 adult subjects
  • Data Provenance: The study involved "healthy adults and adults with respiratory disease or impairment." The document does not explicitly state the country of origin, but given the manufacturer (PneumaCare Limited) is based in the United Kingdom, it's plausible the study was conducted there or in a related region. The study is described as a "clinical trial," which implies a prospective design.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

The ground truth for the clinical study was established using "COSC (Clinician Overscored CO2 derived from the reference capnograph device)." The document explicitly mentions "Clinician Overscored," indicating that clinicians were involved in evaluating the capnography data to establish the ground truth. However, the number of clinicians and their specific qualifications (e.g., years of experience, specialty) are not specified in the provided text.

4. Adjudication Method for the Test Set

The adjudication method is described as "Clinician Overscored CO2 derived from the reference capnograph device." This implies that clinicians reviewed and, presumably, validated or refined the CO2 measurements from the capnograph to establish the most accurate respiration rate. The exact process of adjudication (e.g., if multiple clinicians reviewed and resolved discrepancies, or if it was a single clinician oversight) is not detailed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. The study compared the device's performance against a reference standard (capnography with clinician overscoring), not against human readers with and without AI assistance.

6. Standalone Performance Study

Yes, a standalone performance study was done. The clinical trial directly evaluated the Thora-3Di's respiratory rate measurements against a gold standard (capnography). The primary endpoint analysis was to compare the output respiratory rate of Thora-3Di with that of the COSC.

7. Type of Ground Truth Used

The type of ground truth used for the clinical study was "expert consensus" (clinician overscored) applied to "physiological measurement data" (Capnography).

8. Sample Size for the Training Set

The document does not specify a training set sample size. The description of the device as executing "real-time pattern recognition algorithms" suggests machine learning may be involved, requiring training data, but this information is not provided. The clinical study described appears to be a validation or test set rather than a training set.

9. How the Ground Truth for the Training Set Was Established

As no training set sample size is specified, the method for establishing its ground truth is also not provided in the document.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of a human figure, with three overlapping profiles facing to the right. The profiles are rendered in black and create a sense of depth and unity. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 15, 2016

PneumaCare Limited c/o Ms. Nandini Murthy Regulatory Consultant to PneumaCare Alacrita, One Broadway, 14th Floor, Kendall Square Cambridge, Massachusetts 02142

Re: K151940

Trade/Device Name: Thora-3Di, Model T-01 Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: Class II Product Code: BZO Dated: February 9, 2016 Received: February 10, 2016

Dear Ms. Nandini Murthy:

This letter corrects our substantially equivalent letter of March 10, 2016.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{1}------------------------------------------------

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page

510(k) Number (if known) K151940

Device Name

Thora-3Di, Model T-01

Indications for Use (Describe)

The Thora-3Di is intended for a one-time measurement of respiratory rate as part of a vital signs assessment. The device is indicated for hospital or clinical use in adult patients. The device is intended to be operated by clinicians and medically qualified personnel. It is available for sale only upon the order of a physician or licensed health care provider.

The Thora-3Di is not intended to monitor vital signs. This device is not an apnea monitor.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D) Subpart C)

□ Over-The-Counter Use (21 CFR 801

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995. DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW. The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number

{3}------------------------------------------------

510(k) SUMMARY

Submitter Name:PneumaCare Limited
Submitter Address:Prospect House, 3 St Thomas' PlaceCambridgeshire Business Park, ElyCB7 4EX, United Kingdom
Contact Person:Mark Harwood
Phone Number:00 44 (0) 1223 967414
Fax Number:00 44 (0) 1353 666 692
Date Prepared:10 March 2016
Device Trade Name:Thora-3Di, Model T-01
Device Common Name:Breathing Frequency Monitor
Device Classification:21 CFR 868.2375, Breathing Frequency MonitorProduct Code BZQ
Predicate Device:K090273, Kai Spot KMS200, Kai Medical
Predicate DeviceClassification:21 CFR 868.2375, Breathing Frequency MonitorProduct Code BZQ

Device Description:

The Thora-3Di uses Structured Light Plethysmography (SLP) to derive respiratory rate. With the Thora-3Di, a structured grid pattern of light is projected onto the thoracoabdominal wall and two digital video cameras record changes in the grid pattern due to movement of the wall. A dedicated parallel processor executes real-time pattern recognition algorithms to identify the position of each part of the virtual grid in each camera. The system computes the 3-dimensional coordinates of the different markers to assess respiratory rate.

The Thora-3Di does not include any wireless transmission features, nor does it include a cloud or other remote functionality (no cybersecurity-related features).

Indications for Use:

The Thora-3Di is intended for a one-time measurement of respiratory rate as part of a vital signs assessment. The device is indicated for hospital or clinical use in adult

{4}------------------------------------------------

patients. The device is intended to be operated by clinicians and medically qualified personnel. It is available for sale only upon the order of a physician or licensed health care provider.

The Thora-3Di is not intended to monitor vital signs. This device is not an apnea monitor.

CharacteristicsPneumaCare's Thora-3DiKai Non-Contact RR Spot CheckK090273
Product CodeBZQBZQ
Primary outputDisplays respiratory rate (RR)Displays respiratory rate (RR)
Intended useThe Thora-3Di is intended for a one-time measurement of respiratory rateas part of a vital signs assessment.The device is indicated for hospital orclinical use in adult patients. Thedevice is intended to be operated byclinicians and medically qualifiedpersonnel. It is available for sale onlyupon the order of a physician orlicensed health care provider.The Thora-3Di is not intended tomonitor vital signs. This device is notan apnea monitor.Intended for a one-time measurementof respiratory rate as part of a vitalsigns assessment. The device isindicated for hospital or clinical use inadult patients. The device is intendedto be operated by clinicians andmedically qualified personnel. It isavailable for sale only uponthe order of a physician or licensedhealth care provider.The Kai Sensors Non-ContactRespiratory Rate Spot Check Model100 is not intended to monitor vitalsigns. This device is not an apneamonitor.
Intended useenvironmentIn hospital or clinical care settingsIn hospital or clinical care settings
Duration of monitoringUp to 5 minutes (acute measurement)Spot check measurement
TechnologyNon patient contacting, works fromdistance through clothingNon patient contacting, works fromdistance through clothing
Measurement positionsSubject is seatedSubject can be sitting or lying down
Clinical validationValidated against Capnography(Clinician Over-Scored Capnography)Validated against Embletta, WelchAllyn Propaq
Parameters displayedRespiratory rateRespiratory rate

Overview of Substantial Equivalence:

{5}------------------------------------------------

CharacteristicsPneumaCare's Thora-3DiKai Non-Contact RR Spot CheckK090273
Technology PlatformVisual technique using two camerasand a projector to look at chest andabdominal wall movementUses a low power radar to sense chestdisplacement.

Rationale for Substantial Equivalence:

The Thora-3Di indications for use is identical to the predicate Kai Spot. The Thora-3Di and the predicate Kai Spot devices are similar in design in that both devices assess respiratory rate as a spot measurement, i.e. neither device is a continuous long term monitor. Both devices are non-patient contacting and have a similar technological approach of looking at signals generated from chest wall displacement to assess respiratory rate. The Thora-3Di uses Structured Light Plethysmography (SLP), an optical technique while the Kai Spot uses Doppler Radar to assess respiratory rate.

Both devices' respiratory rate output were compared to reference or predicate devices in clinical studies, and met the prospectively defined acceptance criteria, demonstrating substantial equivalence to their respective comparator devices.

Relevant Standards:

The Thora-3Di has been tested and meets the requirements of the relevant sections of the following performance standards:

  • IEC 60601-1 Standard: Medical Electrical Equipment Part 1: General O Requirements for Basic Safety and Essential Performance (2005)
  • IEC 60601-1-2 Standard: Medical Electrical Equipment Part 1-2: General o Requirements for Basic Safety and Essential Performance, Collateral Standard: Electromagnetic Compatibility (2007)
  • IEC 62304: First edition 2006-05, medical device software software life cycle o processes.
  • ISO 14971: Second edition 2007-03-01, medical devices application of risk O management to medical devices.

Software documentation and processes also complied with the following FDA guidance documents:

  • Draft Guidance for Industry and FDA Staff: Content of Premarket Submissions o for Software Contained in Medical Devices( May 2005)
  • FDA General Principles of Software Validation: Final Guidance for Industry and o Staff (January 2002)

{6}------------------------------------------------

  • Guidance for Industry, FDA Reviewers and Compliance on Off-The-Shelf o Software Use in Medical Devices (September 1999)

Summary of Performance Data:

The Thora-3Di has been tested on the bench to demonstrate accuracy of respiratory rate between 2-60 brpm, where test conditions simulated different subject anatomies such as large abdomen and chest size, and weak breathing. The bench test also verified that the software displayed applicable user alerts as intended when measurements were taken outside of the labeled setup instructions. Software bench testing verified all software system requirements.

The Thora-3Di has also been evaluated in a clinical trial against the reference standard of capnography for the respiratory range of 8-25 brpm. 35 adult subjects between the ages of 19 and 77 meeting the study inclusion criteria were enrolled. The primary endpoint analysis was to compare the output respiratory rate of Thora-3Di with that of the COSC (Clinician Overscored CO2 derived from the reference capnograph device) measured over one randomly selected epoch of one minute for each subject.

The acceptance criterion was based on a mean squared difference (MSD) <0.81 brpm between the two devices which would indicate limits of agreement of +/- 2 brpm.

Results: The MSD was estimated to be 0.136 brpm which is significantly less than the null hypothesis value of 0.81 brpm (p-value=0.0002). The 95% limits of agreement were (-0.72, 0.75 brpm) which are much narrower than the pre-defined limits of agreement of (-2, 2 brpm).

Conclusion:

The Thora-3Di has similar intended use and indications statements as the predicate device, for the measurement of respiratory rate. In the bench study, the Thora-3Di showed an accuracy of </= +/-1 brpm for different simulated anatomies and simulated weak breathing. In a pivotal clinical study, Thora-3Di was compared to the reference capnograph device, using COSC for the measurement of respiratory rate (RR) in healthy adults and adults with respiratory disease or impairment. The 95% CI was in between -0.72 to 0.75 breaths per minute, which is equivalent to the predicate device's published t 95% CI of +1.8 and -4.5 BPM. Therefore, the Thora-3Di is substantially equivalent to the predicate device.

1 Droitcour et al, Non-Contact Respiratory Rate Measurement Validation for Hospitalized Patients, Conf Proc IEEE Eng Med Biol Soc. 2009; 2009: 4812-4815

§ 868.2375 Breathing frequency monitor.

(a)
Identification. A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.(b)
Classification. Class II (performance standards).