AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Capiox® FX Advance Oxygenator with Integrated Arterial Filter and Hardshell Reservoir is intended to be used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during cardiopulmonary bypass surgery. The integrated arterial filter is intended to filtrate non-biologic particles and emboli and to facilitate air bubble removal from the blood flowing through the cardiopulmonary bypass circuit.

The integrated heat exchanger is used to warm or cool blood and/or perfusion fluid as it flows through the device.

The hardshell reservoir is used to store blood during extra-corporeal circulation from the venous line and the cardiotomy line. The reservoir contains a venous section that is comprised of a filter and defoamer to facilitate air bubble removal. The cardiotomy section of the reservoir contains a filter to remove particulate matter and a defoamer to facilitate air bubble removal. The 3-liter and 4-liter reservoirs may be used for Vacuum Assisted Drainage procedures and Post Operative Chest Drainage Procedures.

The Capiox® FX15 Advance Oxygenator with Integrated Arterial Filter and Hardshell Reservoir (both 3-liter and 4-liter) is for use with patients when the required blood flow rate will not exceed 5.0 L/min. The Capiox® FX25 is for use with patients when the required blood flow rate will not exceed 7.0 L/min.

The Capiox® FX Advance Oxygenator/Reservoir/Arterial Filter assemblies can be used in procedures lasting up to 6 hours.

Device Description

The modified Capiox® FX15 and FX25 Advance Oxygenators (with 4-liter Reservoir) utilize porous fiber technology to facilitate the transfer of gases between a blood-phase environment and a gas-phase environment for the intent of satisfying the gas exchange needs of a patient during cardiopulmonary bypass surgery. A fiber bundle offers the porous membrane surface to sufficiently permit the movement of gases through the walls of the hollow fibers via diffusion.

The modified Capiox® FX15 and FX25 Advance devices have an integrated heat exchanger that is comprised of stainless steel encased in a polycarbonate housing. The stainless steel acts as a heat transfer material that permits heat that is generated from a temperature controlled external water bath to transverse across the walls of the stainless steel to effect the necessary temperature change upon circulating blood.

With respect to the filtration of arterial blood, the modified Capiox® FX15 and FX25 Advance Oxygenators rely upon mechanical entrapment of particulates and emboli within the filter mesh as a means to remove those particulates from the blood.

The design of the subject and predicate devices is nearly identical. No modifications are being made to the oxygenator or arterial filter components of the reservoir component of the devices remain identical to the design of the original reservoir that was cleared by FDA with K140774, except for the following:

  • The geometry of the polycarbonate Venous Inlet Port is being modified from an angled design to a curved design. The curved design for the inlet port will facilitate enhanced air handling within the device.
  • The venous port drop tube within the reservoir is being modified from a straight design at the lower end of the tube to a flared design, which is intended to improve flow dynamics through the reservoir.
  • The polyvinyl chloride flexible tubing used in the Reservoir is being changed from PVC tubing with DEHP plasticizer to PVC tubing with TOTM plasticizer. This change will address growing market concerns with the use of DEHP and to meet pending regulatory requirements in the global market. The modified tubing affects the venous drop tubes, the sampling manifold tubing, the arterial quick disconnect and the arterial purge line tubing.
  • The minimum operating level of the Reservoir will be reduced from 200mL to 150mL.
  • The decal-type graduate scale label affixed to the outside surface of the Reservoir is being increased in length to offer convenience to the user-perfusionist.

The materials that are used in the construction of the Capiox® FX15 and FX25 Advance Oxygenators include nylon, polycarbonate, stainless steel, polyvinyl chloride, polyurethane, polyester, polypropylene, polyethylene terephthalate, polyethylene and X-Coating™.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the Terumo Capiox FX15 and FX25 Advance Oxygenators with Integrated Arterial Filter and Hardshell Reservoir. This is a medical device submission, and the focus is on demonstrating substantial equivalence to a predicate device rather than presenting a study proving performance against specific acceptance criteria for a novel AI device. Therefore, many of the requested elements for AI device studies (like sample size for test sets, data provenance, ground truth establishment, MRMC studies, training set size) are not applicable or not explicitly detailed in this type of document.

However, I can extract information related to the device's performance evaluation and acceptance criteria through the lens of device substantial equivalence.

Here's an analysis based on the provided text, addressing the applicable points:


Acceptance Criteria and Device Performance (Demonstrated through Substantial Equivalence)

The core "acceptance criteria" in this context are tied to demonstrating substantial equivalence to a predicate device (Terumo® Capiox® FX Hollow Fiber Oxygenator with Integrated Arterial Filter and Hardshell Reservoir - K140774). The performance of the modified device (Capiox® FX15 and FX25 Advance Oxygenators with 4-liter Reservoir) must be equivalent to the predicate device, especially considering the changes made to the reservoir.

Since this is a Special 510(k) focusing on changes to the reservoir, the performance criteria are applied to the reservoir's functions. The main changes are:

  • Geometry of the polycarbonate Venous Inlet Port (angled to curved)
  • Venous port drop tube (straight to flared)
  • Plasticizer in PVC tubing (DEHP to TOTM)
  • Minimum operating level of the Reservoir (200mL to 150mL)
  • Decal-type graduate scale label (increased length)

The document does not provide a table with specific numerical acceptance criteria (e.g., "hemolysis index < X") and corresponding reported performance values for each test. Instead, it broadly states that "Performance evaluations have demonstrated substantial equivalence" or "The subject Capiox® FX15 and FX25 Advance devices exhibit equivalent performance to the predicate devices."

Table of Performance Evaluations and Outcome Statements:

Performance Evaluation TypeReported Device Performance (vs. Predicate)
Verification Testing of Curved Inlet Reservoir ConnectionsDemonstrated substantial equivalence across varying elements of design.
Verification Testing of Pressure DropExhibited equivalent performance.
Verification Testing of HemolysisExhibited equivalent performance.
Verification Testing of Reservoir Venous Filter DefoamingExhibited equivalent performance.
Verification Testing of Clotting PerformanceExhibited equivalent performance.
Verification Testing of Air Handling PerformanceExhibited equivalent performance, facilitated by curved inlet port.
Flow Dynamics Evaluation of Curved Inlet PortExhibited equivalent performance, improved flow dynamics.
Aged Mechanical Integrity TestingExhibited equivalent performance.
Biocompatibility (for new plasticizer)Found to be biocompatible (with TOTM plasticizer).
Sterilization conditionsValidated to provide a Sterility Assurance Level (SAL) of 10⁻⁶.

Since this is a medical device submission for substantial equivalence and not an AI device study, most of the remaining points are not directly applicable or explicitly stated in the document.

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not explicitly stated for each performance test. The document refers to "simulated use testing of the reservoirs" and "performance evaluations."
  • Data Provenance: The studies were conducted by Terumo Cardiovascular Systems Corporation. Location information is "Elkton, Maryland, USA." The studies are prospective in the sense that they were conducted specifically for this submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable as this is not an expert-driven AI study. Performance tests generally rely on established engineering and physiological testing standards for medical devices.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Not applicable as this is not an expert-driven AI study requiring adjudication of interpretations. The performance evaluations are objective measurements.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This is a medical device (oxygenator, reservoir, filter) and not an AI or imaging diagnostic device.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This device does not have an "algorithm" in the context of AI. Its performance is mechanical and physiological.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • The "ground truth" for the device's performance tests would be established by industry standards, engineering specifications, and physiological requirements for cardiopulmonary bypass devices, as determined by validated test methods (e.g., for hemolysis, gas exchange efficiency, pressure drop, air handling).

8. The sample size for the training set:

  • Not applicable. This is not an AI device trained on a dataset.

9. How the ground truth for the training set was established:

  • Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the seal for the Department of Health & Human Services (USA). The seal features the department's name encircling a stylized caduceus symbol. The caduceus is depicted as a series of three interconnected, abstract human profiles. The seal is in black and white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 19, 2015

Terumo Cardiovascular Systems Corporation Jaime Lee Regulatory Affairs Associate 125 Blue Ball Road Elkton, Maryland 21921

Re: K151791

Trade/Device Name: Terumo Capiox FX15 Advance Oxygenator with Integrated Arterial Filter and Hardshell Reservoir, Terumo Capiox FX25 Advance Oxygenator with Integrated Arterial Filter and Hardshell Reservoir Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary Bypass Oxygenator Regulatory Class: Class II Product Code: DTZ. DTN. DTR. DTM Dated: October 21, 2015 Received: October 22, 2015

Dear Jaime Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

M.A. Hillebrand

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the Terumo Corporation logo. The logo consists of a green circle with a white "T" inside, followed by the word "TERUMO" in green, sans-serif font. A small registration mark is present to the right of the word.

SECTION 4 - Indications for Use Capiox® FX15 and FX25 Advance Oxygenators with Integrated Arterial Filter and 4-Liter Hardshell Reservoir

510(k) Number (if known): K151791

Device Name: CAPIOX® FX15 and FX25 Advance Oxygenators with Integrated Arterial Filter and Hardshell Reservoir

Indications for Use:

The Capiox® FX Advance Oxygenator with Integrated Arterial Filter and Hardshell Reservoir is intended to be used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during cardiopulmonary bypass surgery. The integrated arterial filter is intended to filtrate non-biologic particles and emboli and to facilitate air bubble removal from the blood flowing through the cardiopulmonary bypass circuit.

The integrated heat exchanger is used to warm or cool blood and/or perfusion fluid as it flows through the device.

The hardshell reservoir is used to store blood during extra-corporeal circulation from the venous line and the cardiotomy line. The reservoir contains a venous section that is comprised of a filter and defoamer to facilitate air bubble removal. The cardiotomy section of the reservoir contains a filter to remove particulate matter and a defoamer to facilitate air bubble removal. The 3-liter and 4-liter reservoirs may be used for Vacuum Assisted Drainage procedures and Post Operative Chest Drainage Procedures.

The Capiox® FX15 Advance Oxygenator with Integrated Arterial Filter and Hardshell Reservoir (both 3-liter and 4-liter) is for use with patients when the required blood flow rate will not exceed 5.0 L/min. The Capiox® FX25 is for use with patients when the required blood flow rate will not exceed 7.0 L/min.

The Capiox® FX Advance Oxygenator/Reservoir/Arterial Filter assemblies can be used in procedures lasting up to 6 hours.

NOTE: The indication change that increases the maximum flow rate of the FX15 with the 3-Liter reservoir from 4.0 liters per minute to 5.0 liters per minute has been submitted as a part of Traditional 510(k) application K151389. This Special 510(k) application is solely intended for the FX15 and FX25 oxygenator used with the 4-Liter reservoir. There is no indication change with respect to the 4-Liter reservoir.

XX ___________________________________________________________________________________________________________________________________________________________________________ Prescription Use (Part 21 CFR 801 Subpart D)

OROver-The-Counter Use (Part 21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{3}------------------------------------------------

Submitter Information

This submission was prepared in June 2015 by: Jaime Lee Regulatory Affairs Associate Terumo Cardiovascular Systems Corporation 125 Blue Ball Road Elkton, MD 21921 Telephone: 1-800-262-3304, Ext. 7650 Fax: 410-392-7171

This submission was prepared for: Terumo Cardiovascular Systems Corporation 125 Blue Ball Road Elkton, MD 21921 Facility Registration No. 1124841

Device Names/ClassificationsProprietary NameClassification NameCommon Name
Capiox® FX15 and FX25 AdvanceOxygenator with Integrated ArterialFilter and Hardshell ReservoirCardiopulmonary BypassOxygenator (Code: DTZ)Oxygenator
Cardiopulmonary BypassBlood Reservoir (Code: DTN)Blood Reservoir
Cardiopulmonary BypassHeat Exchanger (Code: DTR)Heat Exchanger
Cardiopulmonary BypassArterial Line Blood Filter (Code: DTM)Arterial Filter

Predicate Device

The device submitted in this 510(k) maintains characteristics that are substantially equivalent in intended use, design, technology/principles of operation, materials and specifications to the following devices:

Terumo® Capiox® FX Hollow Fiber Oxygenator with Integrated Arterial Filter and Hardshell Reservoir - K140774.

Intended Use

The Capiox® FX Advance Oxygenator with Integrated Arterial Filter and Hardshell Reservoir is intended to be used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during cardiopulmonary bypass surgery. The integrated arterial filter is intended to filtrate non-biologic particles and emboli and to facilitate air bubble removal from the blood flowing through the cardiopulmonary bypass circuit.

The integrated heat exchanger is used to warm or cool blood and/or perfusion fluid as it flows through the device.

The hardshell reservoir is used to store blood during extra-corporeal circulation from the venous line and the cardiotomy line. The reservoir contains a venous section that is comprised of a filter and defoamer to facilitate air bubble removal. The cardiotomy section of the reservoir contains a

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Terumo company logo. The logo consists of a green circle with a stylized "T" inside it, followed by the word "TERUMO" in green, sans-serif capital letters. A small registration mark is present to the right of the word.

filter to remove particulate matter and a defoamer to facilitate air bubble removal. The 3-liter and 4-liter reservoirs may be used for Vacuum Assisted Drainage procedures and Post Operative Chest Drainage Procedures.

The Capiox® FX15 Advance Oxygenator with Integrated Arterial Filter and Hardshell Reservoir (both 3-liter and 4-liter) is for use with patients when the required blood flow rate will not exceed 5.0 L/min. The Capiox® FX25 is for use with patients when the required blood flow rate will not exceed 7.0 L/min.

The Capiox® FX Advance Oxygenator/Reservoir/Arterial Filter assemblies can be used in procedures lasting up to 6 hours.

NOTE: The indication change that increases the maximum flow rate of the FX15 with the 3-Liter reservoir from 4.0 liters per minute to 5.0 liters per minute has been submitted as a part of Traditional 510(k) application K151389. This Special 510(k) application is solely intended for the FX15 and FX25 oxygenator used with the 4-Liter reservoir. There is no indication change with respect to the 4-Liter reservoir.

Principles of Operation and Technology

The modified Capiox® FX15 and FX25 Advance Oxygenators (with 4-liter Reservoir) utilize porous fiber technology to facilitate the transfer of gases between a blood-phase environment and a gas-phase environment for the intent of satisfying the gas exchange needs of a patient during cardiopulmonary bypass surgery. A fiber bundle offers the porous membrane surface to sufficiently permit the movement of gases through the walls of the hollow fibers via diffusion.

The modified Capiox® FX15 and FX25 Advance devices have an integrated heat exchanger that is comprised of stainless steel encased in a polycarbonate housing. The stainless steel acts as a heat transfer material that permits heat that is generated from a temperature controlled external water bath to transverse across the walls of the stainless steel to effect the necessary temperature change upon circulating blood.

With respect to the filtration of arterial blood, the modified Capiox® FX15 and FX25 Advance Oxygenators rely upon mechanical entrapment of particulates and emboli within the filter mesh as a means to remove those particulates from the blood.

Design and Materials

The design of the subject and predicate devices is nearly identical. No modifications are being made to the oxygenator or arterial filter components of the reservoir component of the devices remain identical to the design of the original reservoir that was cleared by FDA with K140774, except for the following:

  • The geometry of the polycarbonate Venous Inlet Port is being modified from an angled . design to a curved design. The curved design for the inlet port will facilitate enhanced air handling within the device.
  • . The venous port drop tube within the reservoir is being modified from a straight design at the lower end of the tube to a flared design, which is intended to improve flow dynamics through the reservoir.
  • The polyvinyl chloride flexible tubing used in the Reservoir is being changed from PVC . tubing with DEHP plasticizer to PVC tubing with TOTM plasticizer. This change will address growing market concerns with the use of DEHP and to meet pending regulatory

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Terumo logo. The logo consists of a green circle with a white "T" inside of it, followed by the word "TERUMO" in green, sans-serif font. The logo is simple and modern, and the green color gives it a sense of health and vitality.

requirements in the global market. The modified tubing affects the venous drop tubes, the sampling manifold tubing, the arterial quick disconnect and the arterial purge line tubing.

  • . The minimum operating level of the Reservoir will be reduced from 200mL to 150mL.
  • The decal-type graduate scale label affixed to the outside surface of the Reservoir is being o increased in length to offer convenience to the user-perfusionist.

The materials that are used in the construction of the Capiox® FX15 and FX25 Advance Oxygenators include nylon, polycarbonate, stainless steel, polyvinyl chloride, polyurethane, polyester, polypropylene, polyethylene terephthalate, polyethylene and X-Coating™.

Performance Evaluations

Clinical studies involving patients are not necessary to demonstrate substantial equivalence of the subject device to the predicate devices. For this submission, Terumo Cardiovascular has completed the following (simulated use) performance evaluations on the reservoir:

  • Verification Testing of Curved Inlet Reservoir Connections .
  • o Verification Testing of Pressure Drop
  • Verification Testing of Hemolysis .
  • Verification Testing of Reservoir Venous Filter Defoaming .
  • Verification Testing of Clotting Performance .
  • Verification Testing of Air Handling Performance o
  • Flow Dynamics Evaluation of Curved Inlet Port .
  • . Aged Mechanical Integrity Testing

For this submission, it is not necessary to conduct performance testing on the Note: oxygenator/arterial filter module since there have been no changes to the oxygenator/arterial filter. The performance of the oxygenator/arterial filter module is completely independent of the reservoir. Changes made to the reservoir have no impact whatsoever on the performance of the oxygenator.

Substantial Equivalence Comparison

The information presented in this section depicts a comparison between the subject Capiox® FX15 and FX25 Advance devices and the predicate Terumo Cardiovascular Systems Capiox® FX15 and FX25 Hollow Fiber Oxygenator with Integrated Arterial Filter and Hardshell Reservoir (K140774).

Comparison of Intended Use .

The subject Capiox® FX15 and FX25 Advance devices with the 4-liter Reservoir and the predicate Capiox® FX15 and FX25 devices with a 4-liter Reservoir are identical.

NOTE: The use of the FX15 with the 3-Liter reservoir has been submitted as a part of Traditional 510(k) application K151389. This Special 510(k) application is solely intended for the FX15 and FX25 oxygenator used with the 4-Liter reservoir.

Duration of Use .

The subject Capiox® FX15 and FX25 Advance devices and the predicate devices can be used in procedures lasting up to 6 hours.

Comparison of Labeling .

The subject Capiox® FX15 and FX25 Advance devices and the predicate Capiox® FX15 and FX25 devices are offered with adequate Instructions for Use and other product

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the Terumo company logo. The logo consists of a green circle with a stylized "T" inside it, followed by the word "TERUMO" in green, sans-serif capital letters. A small registration mark is present to the right of the word.

labeling as required by regulation. The product IFUs are equivalent and differ only indicating a higher maximum flow rate for the FX15 devices when used with a 3-Liter reservoir – and in some product illustrations.

NOTE: The indication change that increases the maximum flow rate of the FX15 with the 3-Liter reservoir from 4.0 liters per minute to 5.0 liters per minute has been submitted as a part of Traditional 510(k) application K151389. This Special 510(k) application is solely intended for the FX15 and FX25 oxygenator used with the 4-Liter reservoir. There is no indication change with respect to the 4-Liter reservoir.

Comparison of Principles of Operation & Technology .

Both the subject Capiox® FX15 and FX25 Advance Oxygenators and the predicate Capiox® FX15 and FX25 devices utilize the exact same technologies and principles of operation. There are absolutely no differences between the subject and predicate devices with respect to product technology or product operation.

. Comparison of Design

The design of the subject Capiox® FX15 and FX25 Advance Oxygenators are equivalent to the design of the predicate Capiox® FX15 and FX25 devices, except for the following:

  • . The geometry of the polycarbonate Venous Inlet Port is being modified from an angled design to a curved design. The curved design for the inlet port will facilitate enhanced air handling within the device.
  • . The venous port drop tube within the reservoir is being modified from a straight design at the lower end of the tube to a flared design, which is intended to improve flow dynamics through the reservoir.

Performance evaluations have demonstrated substantial equivalence across varying elements of design.

Comparison of Materials .

The materials of construction between the subject Capiox® FX15 and FX25 Advance Oxygenators and the predicate Capiox® FX15 and FX25 devices are identical with the exception of the plasticizer that is used in the polyvinyl chloride tubing (drop tubes, sampling manifold, purge lines). The subject devices include PVC tubing with TOTM plasticizer while the predicate devices include PVC tubing with DEHP as the plasticizer. Performance evaluations have demonstrated substantial equivalence between the two materials.

  • . Comparison of Performance
    The subject Capiox® FX15 and FX25 Advance devices exhibit equivalent performance to the predicate devices. A series of simulated use testing of the reservoirs demonstrates substantial equivalence between the two designs.

Substantial Equivalence Statement

The Terumo Cardiovascular Systems Corporation Capiox® FX15 and FX25 Advance devices (with 4-liter Reservoir) and the predicate Capiox® FX15 and FX25 devices (with 4-liter Reservoir) are substantially equivalent in intended use, principles of operation and technology, design, materials, and performance. The noted differences between the subject devices and the predicate devices do not raise new issues of safety and effectiveness.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the Terumo Corporation logo. The logo consists of a green circle with a stylized "T" inside, followed by the word "TERUMO" in green, sans-serif font. The logo is simple and modern, and the green color suggests health and vitality.

Additional Safety Information

  • Sterilization conditions have been validated to provide a Sterility Assurance Level (SAL) of . 10°. Terumo further asserts that the ethylene oxide residues will not exceed the maximum residue limits at the time of product distribution.
  • . Terumo maintains biocompatibility studies as recommended in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International ISO 10993, "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing." [External Communicating Devices, Circulating Blood, Limited Exposure (≤ 24 hours) Contact Duration]. The blood contacting materials were found to be biocompatible.

Conclusion

In summary, the Terumo Cardiovascular Systems Corporation Capiox® FX15 and FX25 Advance devices with 4-liter Reservoir are substantially equivalent to the predicate (K140774) Capiox® FX15 and FX25 devices with 4-liter Reservoir with respect to intended use, principles of operation and technology, design, materials, performance, and specifications. It is further noted that any recognized differences do not raise any new issues of patient/user safety or product effectiveness.

§ 870.4350 Cardiopulmonary bypass oxygenator.

(a)
Identification. A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”