K Number
K151693
Manufacturer
Date Cleared
2015-10-09

(108 days)

Product Code
Regulation Number
882.1320
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cutaneous Electrodes, "Dr-Ho's Foot Pad Electrodes", are intended to be used with legally marketed electrical stimulating devices such as transcutaneous electrical nerve stimulators or powered muscle stimulators. The cutaneous electrodes will deliver the stimulation signals generated by the stimulator to the feet which they are in contact with.

Device Description

The Dr-Ho's Foot Pad Electrode is a foot-shaped, flexible piece of conductive silicone rubber with a receptacle for a wire that connects the device with electrical stimulation devices such as transcutaneous electrical nerve stimulators (TENS) or powered muscle stimulators (PMS).

To stimulate the bottom of the foot the patient simply connects the Foot Pad to a TENS unit, places the Foot Pad onto the floor and then places the foot onto the Foot Pad for stimulation with the TENS device.

The Foot Pad Electrode has a female pin receptacle which accepts the lead wire from the stimulator. The entire Foot Pad Electrode is made up of conductive material to provide uniform current distribution when connected to a stimulator.

The devices can be used dry or wet when in contact with the skin. The entire surface (approx. 35 in') of the Foot Pad Electrode is very conductive, having a resistance of less than 7 ohms per inch. This low resistance provides low current density with uniform current distribution.

AI/ML Overview

This document is a 510(k) summary for the Dr-Ho's Foot Pad Electrode. It outlines the device's characteristics and its substantial equivalence to predicate devices, but it does not contain information about specific acceptance criteria, device performance metrics, or a study proving that the device meets such criteria.

The document focuses on demonstrating that the Dr-Ho's Foot Pad Electrode is substantially equivalent to legally marketed predicate devices, primarily through a comparison of technological characteristics and the results of biocompatibility testing. It explicitly states that no clinical tests were conducted or relied upon to determine substantial equivalence.

Therefore, I cannot fulfill your request for the following information based on the provided document:

  1. A table of acceptance criteria and the reported device performance: This information is not present. The document focuses on demonstrating equivalence rather than meeting specific performance criteria.
  2. Sample size used for the test set and the data provenance: No test set for performance comparison is described. The tests mentioned are biocompatibility tests (irritation and cytotoxicity studies), but these are not performance evaluations in the context of clinical efficacy or accuracy.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as no clinical test set with ground truth is described.
  4. Adjudication method for the test set: Not applicable.
  5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a passive electrode, not an AI-powered diagnostic tool.
  6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable.
  7. The type of ground truth used: Not applicable.
  8. The sample size for the training set: Not applicable.
  9. How the ground truth for the training set was established: Not applicable.

The document primarily states:

  • Non-Clinical Tests Performed: Biocompatibility tests (ISO 10993-10 for irritation and skin sensitization in rabbits and guinea pigs, and ISO 10993-5 for in vitro cytotoxitity).
  • Conclusion: The device is substantially equivalent to predicate devices based on intended use, technological characteristics, and safety information (biocompatibility and bench testing).

In summary, this document is a regulatory submission focused on substantial equivalence rather than a detailed performance study against specific acceptance criteria.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image contains the logo for the Department of Health & Human Services - USA. The logo features the department's seal on the left, which includes the department's name encircling a stylized image of a human face in profile. To the right of the seal, the text 'DEPARTMENT' is displayed in a bold, sans-serif font.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 9, 2015

DRTV Asia Ltd. c/o Media Trade Corporation Attn: Guenter Ginsberg, President 11820 Red Hibiscus Dr. Bonita Springs, FL 34135

Re: K151693 Trade Name: Dr-Ho's Foot Pad Electrode Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous electrode Regulatory Class: Class II Product Code: GXY Dated: September 8, 2015 Received: September 10, 2015

Dear Mr. Ginsberg:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code 27f Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{1}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Michael J. Hoffmann -S

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K151693

Device Name Dr-Ho's Foot Pad Electrode

Indications for Use (Describe)

The Cutaneous Electrodes, "Dr-Ho's Foot Pad Electrodes", are intended to be used with legally marketed electrical stimulating devices such as transcutaneous electrical nerve stimulators or powered muscle stimulators. The cutaneous electrodes will deliver the stimulation signals generated by the stimulator to the feet which they are in contact with.

Type of Use (Select one or both, as applicable)

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

Date of Summary Prepared: 06/02/2015 (revised 10.7.15)

  1. Submitter's Name: DRTV Asia, Ltd.

Address. 2nd Blg., Juntuo Industry Park, Xinye Dadao, Nancun Panyu, Guangzhou 511442, China

Contact person: Michael Ho

Tel: 416.303.7728

Fax: 905.471.2608

E-mail drtvasia@hotmail.com

2. Proposed New Device:

Trade Name: Dr-Ho's Foot Pad Electrode

Classification Name: Cutaneous Electrode

Regulation Number: 882.1320

Product Code: GXY

Classification Panel: Neurology

Device Class: II

3. Predicate (cleared) Devices:

Predicate 1:TENS BACK PAIN RELIEF SYSTEM, Model EM38(TENS Belt with sewn-in Electrode Patches)
510(k) Number: K113321Manufacturer: Savia Limited, China
Predicate 2:OTC TENS FOR LOW BACK PAIN RELIEF, MODEL WL-2407(TENS Belt with sewn-in Electrode Patches)
510(k) Number: K063660Manufacturer: Well-Life Healthcare Ltd., Taiwan

{4}------------------------------------------------

4. Description of Proposed Device:

The Dr-Ho's Foot Pad Electrode is a foot-shaped, flexible piece of conductive silicone rubber with a receptacle for a wire that connects the device with electrical stimulation devices such as transcutaneous electrical nerve stimulators (TENS) or powered muscle stimulators (PMS).

To stimulate the bottom of the foot the patient simply connects the Foot Pad to a TENS unit, places the Foot Pad onto the floor and then places the foot onto the Foot Pad for stimulation with the TENS device.

The Foot Pad Electrode has a female pin receptacle which accepts the lead wire from the stimulator. The entire Foot Pad Electrode is made up of conductive material to provide uniform current distribution when connected to a stimulator.

The devices can be used dry or wet when in contact with the skin. The entire surface (approx. 35 in') of the Foot Pad Electrode is very conductive, having a resistance of less than 7 ohms per inch. This low resistance provides low current density with uniform current distribution.

The following tests were performed:

ISO 10993-10 Biological evaluation of medical devices. Part 10: Tests for irritation and skin sensitization (Irritation Study in Rabbits)

ISO 10993-5 Biological evaluation of medical devices- Part 5: Tests for in vitro cytotoxitity. ISO

10993-10 Biological evaluation of medical devices, Part 10: Test for irritation and skin sensitization. (Irritation Study in Guinea Pigs)

5. Indication for Use:

The Cutaneous Electrodes, "Dr-Ho's Foot Pad Electrodes", are intended to be used with legally marketed electrical stimulating devices such as transcutaneous electrical nerve stimulators or powered muscle stimulators. The cutaneous electrodes will deliver the stimulation signals generated by the stimulator to the bottom of the feet with which they are in contact. The Foot Pad Electrodes are made up of conductive silicone rubber.

6. Environment of Use:

Clinics, Hospitals and home environments

7. Contraindications:

{5}------------------------------------------------

Do not use this device if you have a cardiac pacemaker, implanted defibrillator, or other implanted metallic or electronic device. Such use could cause electrical shock, bums, electrical interference, or death.

Do not use this device together with a life-supporting medical electronic device such as an artificial heart or lung.

Do not use this device together with a body-worn medical electronic device such as an ECG.

8. Technological Characteristics Compared to the Predicate Device

Both, the Dr-Ho's Foot Pad Electrode and the sewn-in Electrode Patches of the named predicate devices have the same intended use and fundamental technology. All units provide electrically generated pulses from a transcutaneous electrical nerve stimulator to the skin. Dr-Ho's Foot Pad Electrode can be viewed as substantially equivalent to the predicate devices' Electrode Patches.

The electrical stimulation provided by Dr-Ho's Foot Pad Electrodes is substantially equivalent to that of the predicates and is commonly employed by TENS devices that have been cleared for marketing without prescription labeling; i.e. for OTC sale. Technological characteristics, features, specifications, materials and intended uses of the subject device are substantially equivalent to the quoted predicate device. The differences that exist between the devices, mainly in shapes and sizes, are insignificant in terms of safety or effectiveness.

9. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence is as follows:

DRTV Asia, Ltd did not conduct, nor rely upon, clinical tests to determine substantial equivalence of the Dr-Ho's Foot Pad Electrodes vs. the predicate.

ISO 10993-10 Biological evaluation of medical devices, Part 10: Tests for irritation and skin sensitization (Irritation Study in Rabbits)

ISO 10993-5 Biological evaluation of medical devices- Part 5: Tests for in vitro cytotoxitity. ISO

10993-10 Biological evaluation of medical devices, Part 10: Test for irritation and skin sensitization. (Irritation Study in Guinea Pigs)

b. Biocompatibility

The material of the subject electrodes has been tested for biocompatibility as per

ISO Standards below:

ISO 10993-10 Biological evaluation of medical devices, Part 10: Tests for irritation and skin sensitization (Irritation Study in Rabbits)

{6}------------------------------------------------

ISO 10993-5 Biological evaluation of medical devices- Part 5: Tests for in vitro cytotoxitity.

ISO 10993-10 Biological evaluation of medical devices, Part 10: Test for irritation and skin sensitization. (Irritation Study in Guinea Pigs)

The skin contact duration is class A = ≤ 24 hrs.

c. Software

There is no software in this product.

d. Cleaning

The cleaning instructions as described the Instruction Manual have been tested to be sufficient. Testing involved validation of the manual cleaning method as per the instructions. All testing concluded that that the Dr-Ho's Foot Pad Electrodes can be cleaned by the use of the methods described in the Instruction Manual. Sterilization is not required.

8. Conclusions:

The Dr-Ho's Foot Pad Electrodes have the same intended use and technological characteristics as the predicate device. Moreover, bench testing and safety report documentation demonstrate that the submitted device could maintain the same safety and effectiveness as that of predicate device. In the other words, the differences do not affect the intended use and do not raise any new questions of safety or effectiveness or alter the fundamental scientific technology of the device. Thus, the Dr-Ho's Foot Pad Electrodes are substantially equivalent to the predicate device.

{7}------------------------------------------------

Typical Application of the Foot Pat Electrodes

.

Image /page/7/Picture/1 description: The image shows a pair of feet in what appears to be a foot scanner. The feet are positioned side by side, with the toes pointing upwards. A small metal bar is positioned between the heels. The number 5 is written in the bottom right corner of the image.

§ 882.1320 Cutaneous electrode.

(a)
Identification. A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.(b)
Classification. Class II (performance standards).