K Number
K151671

Validate with FDA (Live)

Date Cleared
2015-10-16

(116 days)

Product Code
Regulation Number
876.5010
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable Multistage Dilation Balloon Catheter is indicated for use in adult and adolescent populations to endoscopically dilate strictures of the gastrointestinal tract.

Device Description

The Disposable Multistage Dilation Balloon Catheter is capable of 3 distinct and progressively larger size diameters via controlled radial expansion. Specific balloon sizes are printed on each package and hub label. The Disposable Multistage Dilation Balloon Catheter is designed to pass through a 2.8mm or greater working channel of an endoscope. It will also accept a 0.035 in (0.89 mm) guidewire through its guidewire lumen. This catheter comes packaged with a 0.035 in (0.89mm) guidewire preloaded in the guidewire lumen. The guidewire is about 20cm longer than the balloon catheter with the excess length extending from single lumen tube. The guidewire locking device is attached to the guidewire hub of the catheter. The locking device will be packaged in locked.

AI/ML Overview

The Micro-Tech (Nanjing) CO., Ltd. Disposable Multistage Dilation Balloon Catheter did not undergo a clinical study. Instead, the device's substantial equivalence to predicate devices was established through bench testing.

Therefore, the following information cannot be extracted from the provided text:

  • Acceptance Criteria Table
  • Sample size used for the test set and data provenance
  • Number of experts and their qualifications for establishing ground truth
  • Adjudication method
  • Multi-reader multi-case (MRMC) comparative effectiveness study
  • Standalone performance study
  • Type of ground truth used
  • Sample size for the training set
  • How ground truth for the training set was established

1. Acceptance Criteria and Reported Device Performance

As no clinical study was performed, detailed acceptance criteria and reported device performance in a clinical context are not provided. The substantial equivalence was based on bench testing.

Bench Test Performed:

The following in-vitro performance tests were completed on the Disposable Multistage Dilation Balloon Catheter to demonstrate substantial equivalence to predicate devices:

  • Balloon Diameter
  • Balloon Length
  • Rated Pressure
  • Working Length
  • Retraction Testing
  • Tensile Strength
  • Tip Pull Testing
  • Endoscopic Compatibility Testing
  • Balloon Burst Pressure
  • Balloon Fatigue Testing
  • Radiopacity
  • Balloon Inflation/Deflation Testing
  • Package Integrity
  • Shelf Life Testing

The conclusion drawn from these tests was that "The Disposable Multistage Dilation Balloon Catheter performs as well as the predicate and is substantially equivalent to the predicate devices."

2. Sample size used for the test set and the data provenance

Not applicable, as no clinical study or test set with patient data was utilized. Substantial equivalence was demonstrated through bench testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable, as no clinical study or ground truth based on expert consensus was established. Substantial equivalence was demonstrated through bench testing.

4. Adjudication method for the test set

Not applicable, as no clinical study or test set requiring adjudication was performed.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable, as this device is a physical medical device (balloon catheter), not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable, as this device is a physical medical device, not an algorithm.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

Not applicable, as the evaluation was based on in-vitro bench testing, rather than a clinical ground truth.

8. The sample size for the training set

Not applicable, as this device is a physical medical device and not an AI/machine learning algorithm that requires a training set.

9. How the ground truth for the training set was established

Not applicable, as this device is a physical medical device and not an AI/machine learning algorithm that requires a training set.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 16, 2015

Micro-Tech (Nanjing) CO., Ltd. Becky Li Quality Manager No.10 Gaoke Third Rd., Nanjing National Hi-Tech Industrial Development Zone Nanjing, Jiangsu 210032 China

Re: K151671

Trade/Device Name: Disposable Multistage Dilation Balloon Catheter Regulation Number: 21 CFR §876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: II Product Code: FGE, KNQ Dated: September 6, 2015 Received: September 10, 2015

Dear Becky Li,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K151671

Device Name

Disposable Multistage Dilation Balloon Catheter

Indications for Use (Describe)

The Disposable Multistage Dilation Balloon Catheter is indicated for use in adult and adolescent populations to endoscopically dilate strictures of the gastrointestinal tract.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use (21 CFR 801 Subpart C)
----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete

and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for MICRO-TECH. The logo consists of the letters "MT" in a stylized, three-dimensional font, with the word "MICRO-TECH" to the right of the letters. Below the word "MICRO-TECH" are two Chinese characters. The logo is blue and white.

510K Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

The assigned 510(k) Number: K151671

1. Date of Preparation: 10/14/2015

2. Sponsor Identification

Micro-Tech (Nanjing) Co., Ltd.

No.10 Gaoke Third Road, Nanjing National Hi-Tech, Industrial Development Zone, Nanjing 210032, Jiangsu Province, PRC Establishment Registration Number: 3004837686 Contact Person: Becky Li Position: Manager of Quality Department Tel: +86-25-58646378 Fax: +86-25-58744269

Email: In@micro-tech.com.cn

3. Identification of Proposed Device

Product Name: Disposable Multistage Dilation Balloon Catheter Common Name: Disposable Multistage Dilation Balloon Catheter Requlatory Information Classification Name: 1) Catheter, Biliary, diagnostic, 2) Dilator, esophageal Classification: 2 Product Code: 1) FGE, 2) KNQ Regulation Number: 1) 876.5010, 2) 876.3565 Review Panel: Gastroenterology/Urology

4. Identification of Predicate Device

510(k) Number: K112994 Trade Name: CRE Dilatation Balloon Common Name: CRE Wireguided Dilatation Balloon

5. Indications for Use

The Disposable Multistage Dilation Balloon Catheter is indicated for use in

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows a logo for Micro-Tech. The logo consists of a blue, stylized "MT" symbol on the left, followed by the text "MICRO-TECH" in a sans-serif font. Below the text, there are two Chinese characters. The overall design is clean and modern.

adult and adolescent populations to endoscopically dilate strictures of the gastrointestinal tract.

6. Device Description

The Disposable Multistage Dilation Balloon Catheter is capable of 3 distinct and progressively larger size diameters via controlled radial expansion. Specific balloon sizes are printed on each package and hub label. The Disposable Multistage Dilation Balloon Catheter is designed to pass through a 2.8mm or greater working channel of an endoscope. It will also accept a 0.035 in (0.89 mm) guidewire through its guidewire lumen. This catheter comes packaged with a 0.035 in (0.89mm) guidewire preloaded in the guidewire lumen. The guidewire is about 20cm longer than the balloon catheter with the excess length extending from single lumen tube.

The guidewire locking device is attached to the guidewire hub of the catheter. The locking device will be packaged in locked.

7. Comparison of Technological Characteristics

The Disposable Multistage Dilation Balloon Catheter incorporates substantially equivalent device materials, design, catheter configuration, packaging fundamental technology, manufacturing processes, sterilization process and intended use as those featured in the Boston Scientific predicate devices.

ltemProposed DeviceComparison to
Disposable Multistage Dilation Balloon CatheterPredicate Devices
Product Code1) FGE, 2) KNQSimilar
Regulation No.1) 876.5010, 2) 876.3565Similar
Class2Same
Supplied SterileYesSame
Balloon Diameter6-7-8, 8-9-10, 10-11-12, 12-13.5-15, 15-16.5-18,Same
(mm)18-19-20
Balloon length (mm)30, 55, 80.Similar
Rated pressure(atm)3-6-10, 3-5.5-9, 3-5-8,3-4.5-8,3-4.5-7,3-4.5-6Same
Working Length1800, 2300Similar
(mm )

Comparison to predicate Devices:

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for MICRO-TECH. The logo consists of the letters "MT" in a stylized, blue font on the left side of the image. To the right of the letters is the word "MICRO-TECH" in a similar blue font, with a line underneath it. Below the line are two Chinese characters.

510K Summary

ItemProposed DeviceDisposable Multistage Dilation Balloon CatheterComparison toPredicate Devices
Indications for UseThe Disposable Multistage Dilation BalloonCatheter is indicated for use in adult andadolescent populations to endoscopically dilatestrictures of the gastrointestinal tract.Similar
ConfigurationTip, balloon, marker band, handle junction, andguidewire.Similar
Single UseYesSame
PackagingSingle-use EO sterilized pouch with one deviceper pouchSimilar
Shelf LifeThree years (36 months)Similar

8. Performance Data

Bench testing was performed to support a determination of substantial equivalence. The Disposable Multistage Dilation Balloon Catheter performs as well as the predicate and is substantially equivalent to the predicate devices. The following in-vitro performance tests were completed on the Disposable

Multistage Dilation Balloon Catheter:

Balloon DiameterEndoscopic Compatibility Testing
Balloon lengthBalloon Burst Pressure
Rated pressureBalloon Fatigue Testing
Working LengthRadiopacity
Retraction testingBalloon Inflation/Deflation Testing
Tensile StrengthPackage integrity
Tip Pull TestingShelf Life Testing

9. Clinical Test Conclusion

No clinical study is included in this submission.

10. Substantially Equivalent (SE) Conclusion

Based on the indications for use, technological characteristics, and safety and performance testing, the Disposable Multistage Dilation Balloon Catheter has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the Boston Scientific CRE Wireguided Dilatation Balloon 。

§ 876.5010 Biliary catheter and accessories.

(a)
Identification. A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.(b)
Classification. Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.