(141 days)
The Symphony breast pump is a powered breast pump to be used by lactating women to express and collect milk from their breast.
The Symphony breast pump is intended to be used by lactating women to express and collect milk from their breasts. The pumping can be performed on one breast (single pumping) or on both breasts at the same time (double pumping). The Symphony breast pump is a multi-user breast pump and therefore, designed for a safe and hygienic use by different mothers. The Symphony breast pump employs a control knob for the user to adjust the applied vacuum. The suction patterns are programmed with either constant or variable cycles (pump speed). The Symphony breast pump is capable of providing vacuum levels from -50 to -250 mmHg with cycling rates up to 120 cycles per minute.
Here's an analysis of the provided text regarding the acceptance criteria and study for the Medela Symphony Breast Pump. Please note that the document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting a de novo study with strict acceptance criteria against a clinical outcome. Therefore, some requested information, particularly around clinical acceptance criteria and related study details, may not be explicitly present or not applicable in the same way it would be for a novel device.
Acceptance Criteria and Device Performance for Medela Symphony Breast Pump (K151632)
The provided document (510(k) summary) focuses on demonstrating substantial equivalence of the Medela Symphony Breast Pump (current version) to a legally marketed predicate device (K020518). The "acceptance criteria" are primarily based on meeting the same performance specifications as the predicate device or demonstrating that modifications do not adversely affect safety and effectiveness.
1. Table of Acceptance Criteria and Reported Device Performance
| Parameter / Acceptance Criteria | Reported Device Performance (Symphony Current Version) |
|---|---|
| Indications for Use: To be used by lactating women to express and collect milk from their breast. (Identical to predicate) | The device has identical indications for use compared to the predicate (K020518). |
| Environment of Use: Hospital, Home (Identical to predicate) | Hospital, Home. |
| Power Source: 100-240 Vac, 50/60Hz, 0.3A. (Identical to predicate) | 100-240 Vac, 50/60Hz, 0.3A. |
| Batteries: 2 x 6V/1.2Ah Rechargeable Pb Batteries. (Identical to predicate) | 2 x 6V/1.2Ah Rechargeable Pb Batteries. |
| Battery charge/discharge: Operation time: approx. 1 hour; Charging time: approx. 4 hours (80%), approx. 8 hours (100%). (Identical to predicate) | Operation time: approx. 1 hour; Charging time: approx. 4 hours (80%), approx. 8 hours (100%). Performance testing demonstrated that the specification for battery operation time is met. |
| Maximum vacuum mmHg/kPa: Approx. -290 mmHg / -40 kPa. (Identical to predicate) | Approx. -290 mmHg / -40 kPa. |
| Vacuum range - double pumping (Standard Program and 24 mm / size M PersonalFit breastshield): -50 to -250 mmHg ± 20 mmHg / -7 to -33 kPa ± 3 kPa. (Identical to predicate) | Performance testing at minimum and maximum vacuum settings demonstrated that the specification for vacuum level is met. |
| Vacuum range - double pumping (Preemie+ Program and 24 mm / size M PersonalFit breastshield): -70 to -250 mmHg ± 20 mmHg / -7 to -33 kPa ± 3 kPa. (New program, compared against specified range). | Performance testing at minimum and maximum vacuum settings demonstrated that the specification for vacuum level is met. |
| Vacuum range - single pumping (Standard Program and 24 mm / size M PersonalFit breastshield): -50 to -250 mmHg ± 20 mmHg / -7 to -33 kPa ± 3 kPa. (Identical to predicate) | Performance testing demonstrated that the specification for vacuum level is met in single pumping mode. |
| Vacuum range - single pumping (Preemie+ Program and 24 mm / size M PersonalFit breastshield): -70 to -250 mmHg ± 20 mmHg / -7 to -33 kPa ± 3 kPa. (New program, compared against specified range). | Performance testing demonstrated that the specification for vacuum level is met in single pumping mode. |
| Adjustable Suction Levels: 16 levels. (Identical to predicate) | 16 levels. |
| Cycle Speed (Standard Program): 54 – 120 Cycles/Minute. (Identical to predicate) | Performance testing demonstrated that the specification for cycle rate is met. |
| Cycle Speed (Preemie+ Program): 35 – 120 Cycles/Minute. (New program, compared against specified range). | Performance testing demonstrated that the specification for cycle rate is met. |
| Overflow protection: A Protection Membrane on the pump side acts as a media separation and prevents milk from going into the pump in case of a milk overflow into the vacuum tubes. (Identical to predicate) | Present and functions as described. Demonstrated not to impact safety/effectiveness negatively. |
| Safety and Effectiveness: No new issues of safety and effectiveness introduced by modifications. | "Verification and Validation testing demonstrated that no adverse effects have been introduced by these differences and that the device performs as intended." Complies with ISO 14971, FDA guidance on sterility and biocompatibility, IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 60601-1-11 (home use), and EN ISO 10079-1 (suction equipment). |
| Software: Software changes (including Preemie+ program) are verified and validated according to FDA guidance for "moderate" level of concern. No failures leading to minor injury (pain or engorgement) should occur. | Software verification and validation testing were conducted, and documentation provided as recommended by FDA guidance. The software was deemed "moderate" level of concern. |
| Sterilization (for sterile items): Compliance with FDA guidance "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile." Accelerated aging per ASTM F1980 (07) to support shelf life. | Sterilization information provided in accordance with FDA guidance. Accelerated aging tests conducted per ASTM F1980 (07) to support labeled shelf life. |
| Cleaning Validation: Validation of manual cleaning, dishwasher cleaning, and sanitation (boiling water & micro-steam™ bags) to reduce organisms from human breast milk solution containing organisms. Protocol based on AAMI TIR30-2003. | Validation of cleaning methods (manual, dishwasher, and sanitation) conducted using a selection of reusable breast pump kit components challenged with human breast milk solution containing organisms. Evaluated for presence of surviving organisms post-cleaning/sanitizing. Protocol based on AAMI TIR30-2003. |
| Biocompatibility: Compliant with FDA guidance "Use of International Standard ISO-10993..." No new testing required as patient contacting materials are identical to predicate. | Biocompatibility evaluation completed; concluded no new testing required as all patient contacting materials are identical to those used in the predicate. |
| Drive Unit Lifespan: Verification of drive unit life time in support of breast pump service life specification. | Testing conducted to verify drive unit life time in support of the breast pump service life specification. |
2. Sample Size Used for the Test Set and Data Provenance
The document describes non-clinical performance testing primarily to demonstrate physical and functional specifications.
- Sample Size for Performance Testing:
- For testing with the Standard 2.0 program card: Six pumps were used (three with each of the two approved motor types).
- For testing with the Preemie+ program card: One pump of each motor type was used (total of 2 pumps).
- Additional tests (battery operation, varying input voltages, different breast shield sizes, single pumping) were conducted, but specific sample sizes for these sub-tests are not explicitly stated beyond implying they were performed on the tested pumps.
- Data Provenance: The tests are non-clinical, laboratory-based performance tests, likely conducted at the manufacturer's facilities in Switzerland (Medela AG is based in Baar, Switzerland). The data is prospective in the sense that it was generated specifically for this 510(k) submission to demonstrate compliance with specifications.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not applicable in the context of this 510(k) summary for a powered breast pump. The "ground truth" for the performance tests (e.g., vacuum levels, cycle rates, battery life) is established by direct measurement against engineering specifications and industry standards, not by expert consensus or interpretation of complex medical imagery/data.
4. Adjudication Method for the Test Set
This information is not applicable. Adjudication methods (like 2+1 or 3+1) are typically used in clinical studies for diagnostic devices where human reader interpretations are being assessed against a consensus ground truth. For performance testing of a physical device like a breast pump, measurements are objective and directly compared to specifications.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done
No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Clinical testing was not required to demonstrate the substantial equivalence of the Symphony breast pump to its predicate device." Therefore, there is no mention of human readers, AI assistance, or effect size improvements related to such a study.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
This concept is not applicable to this device. The Medela Symphony Breast Pump is a mechanical device with integrated software for controlling its operation (e.g., 2-Phase Expression, Preemie+ program). It does not involve "algorithms only" or "AI" in the sense of a standalone diagnostic or predictive algorithm being evaluated independently. The software is integral to the device's function, and its performance is assessed as part of the overall device performance (e.g., ensuring correct vacuum and cycle rates as programmed).
7. The Type of Ground Truth Used
The "ground truth" for the device's technical performance is based on:
- Engineering Specifications: Defined vacuum ranges, cycle speeds, and battery life.
- Voluntary Standards: Compliance with electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), home use safety (IEC 60601-1-11), and suction equipment performance (EN ISO 10079-1).
- Regulatory Guidance: Adherence to FDA guidance for sterility, biocompatibility, and software validation.
8. The Sample Size for the Training Set
This information is not applicable. The Medela Symphony Breast Pump is a traditional electromechanical medical device. It does not use "training sets" in the context of machine learning or artificial intelligence. Its software is programmed based on predefined algorithms and fixed parameters and validated through traditional software verification and validation, not through learning from data like an AI model.
9. How the Ground Truth for the Training Set Was Established
This information is not applicable for the same reasons as point 8.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal is circular and contains the department's name around the perimeter. In the center of the seal is an abstract symbol that resembles three human profiles facing to the right. The symbol is composed of three curved lines that overlap each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 5, 2015
Medela AG % Adrienne Lenz Member Pathway Regulatory Consulting, LLC W324 S3649 County Road E Dousman, WI 53118
Re: K151632
Trade/Device Name: Symphony Breastpump, Symphony Reconditioned, Symphony Preemie+ Breastpump, Symphony Plus Breastpump (battery Version), Symphony Consignment, Preemie Consignment Pump, Symphony Rental, Symphony Plus Rental, Symphony Preemie+ Rental Regulation Number: 21 CFR 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: October 2, 2015 Received: October 5 17, 2015
Dear Adrienne Lenz,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K151632
Device Name
Symphony breast pump, Models: Symphony Preeme+ Breastpump, Symphony Plus Breastpump (battery version), Symphony Consignment, Symphony Preemie Consignment Pump, Symphony Plus Rental, Symphony Preemie+ Rental, Symphony Reconditioned
Indications for Use (Describe)
The Symphony breast pump is a powered breast pump to be used by lactating women to express and collect milk from their breast.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(K) SUMMARY
MEDELA AG SYMPHONY BREAST PUMP
{4}------------------------------------------------
510(k) Summary
K151632
In accordance with 21 CFR 807.92 the following summary of information is provided:
DATE: October 30, 2015
SUBMITTER:
Medela AG Lättichstrasse 4b 6341 Baar / Switzerland Phone +41 (0)41 769 51 51 Fax + 41 (0)41 769 51 00
PRIMARY CONTACT PERSON:
Adrienne Lenz, RAC Member Pathway Regulatory Consulting, LLC T 262-290-0023
SECONDARY CONTACT PERSON:
Orlando Antunes Vice President Regulatory Affairs Medela AG
DEVICE:
TRADE NAME: Symphony breast pump, Models: Symphony Breastpump, Symphony Preemie+ Breastpump, Symphony Plus Breastpump (battery version), Symphony Consignment, Symphony Preemie Consignment Pump, Symphony Rental, Symphony Plus Rental, Symphony Preemie+ Rental, Symphony Reconditioned
COMMON/USUAL NAME: Powered Breast Pump
DEVICE CLASS: Class II
CLASSIFICATION NAMES: 21 CFR 884.516022 Pump, Breast, Powered
PRODUCT CODE: HGX
{5}------------------------------------------------
PREDICATE DEVICE(S):
K020518 Medela Symphony Breast Pump Model 024
DEVICE DESCRIPTION:
The Symphony breast pump is intended to be used by lactating women to express and collect milk from their breasts. The pumping can be performed on one breast (single pumping) or on both breasts at the same time (double pumping). The Symphony breast pump is a multi-user breast pump and therefore, designed for a safe and hygienic use by different mothers.
The Symphony breast pump employs a control knob for the user to adjust the applied vacuum. The suction patterns are programmed with either constant or variable cycles (pump speed). The Symphony breast pump is capable of providing vacuum levels from -50 to -250 mmHg with cycling rates up to 120 cycles per minute.
The Symphony® breast pump provides the following user features:
- Belt-driven vacuum pump design, for quiet operation.
- . Portable or trolley-mounted transport/operation options.
- LCD display, for user assistance/device status.
- Control knob, for user adjustment of vacuum level/pump speed.
- . 2-Phase Expression® Technology designed to mimic a baby's natural nursing rhythm:
- . Stimulation Phase (phase 1): Suction pattern with fast cycles and low vacuum to start milk flowing
- . Expression Phase (phase 2): Suction pattern with slower cycles and higher vacuum to express more milk gently and efficiently.
- Symphony program card with the initiation program (Preemie+) software. The Preemie+ program contains stimulation and expression phases with breast pump suction patterns which assist pump users in the initiation of milk production. The Preemie+ program mimics the irregular sucking pattern of healthy term infants right after birth.
- Symphony program card with the "Standard" pumping program software. The Standard program contains stimulation and expression phases with breast pump suction patterns which assist users in coming to volume, and/or maintenance of lactation. The Standard program imitates a breastfeeding baby during established lactation.
- "Let-down" control button to change between stimulation phase and expression phase.
- Option of either single or double breast pumping.
{6}------------------------------------------------
- Disposable protective membrane, designed to isolate the pump mechanism from the breast milk collection apparatus (prevent milk overflow into the pump). No vacuum can be created if the protective membrane is missing or damaged.
- User-friendly pump exterior, designed for ease of cleaning.
- A variety of accessories and breast pump kits.
INDICATIONS FOR USE:
The Symphony breast pump is a powered breast pump to be used by lactating women to express and collect milk from their breast.
DETERMINATION OF SUBSTANTIAL EQUIVALENCE:
SUMMARY OF TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE
The Symphony breast pump has identical indications for use compared to the previously cleared version (K020518). A number of modifications have been made to the Symphony Breast Pump since the last 510(k) clearance (K020518). These are types of changes expected during years of production, including changes to meet new or modified standards and non-US regulations, manufacturing improvements to increase yield, improve stability and precision of components and prevent damage during use. Changes to the software have also been made, including the introduction of the Preemie+ initiation program.
The table below identifies key similarities and differences of the proposed Symphony breast pump to the legally marketed predicate version of this device (K020518).
| Symphony (Current Version) | Symphony (Predicate Version) | |
|---|---|---|
| K020518 | ||
| Indications forUse | The Symphony breast pump is apowered breast pump to be used bylactating women to express and collectmilk from their breast. | The Symphony breast pump is apowered breast pump to be used bylactating women to express and collectmilk from their breast. |
| Environment ofUse | Hospital, Home | Hospital, Home |
| TABLE 5.1 COMPARISON OF SYMPHONY TO PREDICATE DEVICE | ||
|---|---|---|
| Symphony (Current Version) | Symphony (Predicate Version) | |
| K020518 | ||
| User Interface | Hardware interfaces | Hardware interfaces |
| User Control | State-dependent controls:On-Off membrane switchLet-Down Button (membrane switch),also used to select the Preemie+initiation programValue-adjustment control:Vacuum/Cycle-adjustment control | State-dependent controls:On-Off membrane switchLet-Down Button (membrane switch)Value-adjustment control:Vacuum/Cycle-adjustment control |
| Visual Indicator | Liquid-crystal display | Liquid-crystal display |
| Single/DoublePumping | Single and double pumping | Single and double pumping |
| Accessories | Breast shield Breast shield inserts Valve Membrane Tubing Bottles, with disks lids and caps Trolly Vehicle adapter Power cord Container Stand Cooler Colostrum and Collection Containers Pump and Save Bags | Breast shield Breast shield inserts Valve Membrane Tubing Bottles, with disks lids and caps Trolly Vehicle adapter Power cord |
| Cleaning | Breastpump/case - wipe with clean, damp cloth Tubing - wash or sanitize only if milk or condensation in tubing Breastpump kit and bottles - wash and sanitize | Tubing - wash or sanitize weekly or if milk or condensation in tubing Breastpump kit and bottles - wash and sanitize (home) or autoclave sterilization (hospital) |
| Specifications | ||
| Power Source | 100-240 Vac | 100-240 Vac |
| 50/60Hz | 50/60Hz | |
| 0.3A | 0.3A | |
| Symphony (Current Version) | Symphony (Predicate Version) | |
| K020518 | ||
| Batteries | 2 x 6V/1.2AhRechargeable Pb Batteries | 2 x 6V/1.2AhRechargeable Pb Batteries |
| Battery charge/discharge | Operation time: approx. 1 hourCharging time: approx.. 4 hours (80%)approx.. 8 hours (100%) | Operation time: approx. 1 hourCharging time: approx.. 4 hours (80%)approx.. 8 hours (100%) |
| Housing Material | Polyamide | PS coated |
| Vacuum aggregate type | Diaphragm pumpTwo diaphragms - for left and right breastLinear movement by driving bars | Diaphragm pumpTwo diaphragms - for left and right breastLinear movement by driving bars |
| Vacuum Regulation type | Electronic measuring of the displacement of the membranes which corresponds to the generated vacuum. | Electronic measuring of the displacement of the membranes which corresponds to the generated vacuum. |
| Maximum vacuum mmHg/kPa | Approx. -290 mmHg / -40 kPa | Approx. -290 mmHg / -40 kPa |
| Vacuum range - double mmHg/kPa | -50 to -250mmHg ± 20 mmHg /-7 to -33 kPa ± 3 kPa(with Standard Program and 24 mm / size M PersonalFit breastshield)-70 to -250mmHg ± 20 mmHg /-7 to -33 kPa ± 3 kPa(with Preemie+ Program and 24 mm / size M PersonalFit breastshield) | -50 to -250mmHg ± 20 mmHg /-7 to -33 kPa ± 3 kPa(with Standard Program and 24 mm / size M PersonalFit breastshield) |
| Vacuum range - single mmHg/kPa | -50 to -250mmHg ± 20 mmHg /-7 to -33 kPa ± 3 kPa(with Standard Program and 24 mm / size M PersonalFit breastshield)-70 to -250mmHg ± 20 mmHg /-7 to -33 kPa ± 3 kPa(with Preemie+ Program and 24 mm / size M PersonalFit breastshield) | -50 to -250mmHg ± 20 mmHg /-7 to -33 kPa ± 3 kPa(with Standard Program and 24 mm / size M PersonalFit breastshield) |
| Adjustable Suction Levels | 16 levels | 16 levels |
| Cycling Control Mechanism | Electronic measuring of the Motor speed and position. | Electronic measuring of the Motor speed and position. |
| Symphony (Current Version) | Symphony (Predicate Version) | |
| K020518 | ||
| Cycle Speed | 54 – 120 Cycles/Minute(with Standard Program)35 – 120 Cycles/Minute(with Preemie+ Program) | 54 – 120 Cycles/Minute(with Standard Program) |
| Overflowprotection | A Protection Membrane on the pumpside acts as a media separation andprevents milk from going into thepump in case of a milk overflow intothe vacuum tubes. | A Protection Membrane on the pumpside acts as a media separation andprevents milk from going into thepump in case of a milk overflow intothe vacuum tubes. |
| Software Cards | Standard 2.0 or Preemie+ programcards, each featuring stimulation andexpression phases to assist for theinitiation, coming to volume, and/ormaintenance of lactation.Vacuum range and cycle speed aredependent on the breast pump suctionpatterns.The Standard 2.0 program consists oftwo different curves.The Preemie+ initiation programconsists of three different curves:Standard 2.0 program card: Standardprogram only.Preemie+ program card: Standard 2.0and Preemie+ programs. | Standard program card featuringstimulation and expression phases toassist for the coming to volume,and/or maintenance of lactation.Vacuum range and cycle speed aredependent on the breast pump suctionpatterns.The Standard program consists of twodifferent curves.The Standard program card includesthe Standard program only |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
SUMMARY OF NON-CLINICAL TESTS:
The Symphony breast pump complies with voluntary standards for electrical safety, electromagnetic compatibility, use in the home healthcare environment and powered suction pumps. The following data were provided in support of the substantial equivalence determination:
- . Risk Analysis developed in accordance with ISO 14971: 2007.
- . Sterilization information in accordance with FDA's guidance document "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile".
{10}------------------------------------------------
- . Cleaning Validation. Validation of cleaning methods, both manual cleaning with dish soap and cleaning in the dishwasher, and validation of the sanitation process with boiling water and the quick clean™ micro-steam™ bags were conducted. A selection of reusable breast pump kit components that contact breast milk representative of all Medela breast pump components, including those specified for use with Symphony, were challenged with a human breast milk solution containing organisms, cleaned, or sanitized according to the Instructions for Use and evaluated for the presence of surviving organisms. The validation protocol was based upon AAMI TIR30-2003, A Compendium of Processes, Materials, Test Methods, and Acceptance Criteria for Cleaning Reusable Medical Devices, 8 October 2003.
- . For sterile items, accelerated aging tests to support the labeled shelf life were conducted according to ASTM F1980 (07) Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices. Testing was also conducted to verify drive unit life time in support of the breast pump service life specification.
- . Biocompatibility Evaluation was completed according to the FDA guidance "Use of International Standard ISO- 10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" and concluded that no new testing was required as all patient contacting materials are identical to those used in the predicate Symphony Breast Pump.
- . Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, since prior to mitigations of hazards, failure of the software could lead to minor injury, such as pain or engorgement.
- . Electrical safety and electromagnetic compatibility testing per IEC 60601-1:2005 (3rd Edition) with US deviations per AAMI/ANSI ES60601-1:2005/(R)2012 and A1:2012 standard and IEC 60601-1-2: 2007 standards, respectively
- . Safety testing for use in the home per IEC 60601-1-11: 2010 standard
- . Performance testing demonstrating compliance with EN ISO 10079-1: 2009 Particular requirements for the safety of electrically powered suction equipment
- . Performance testing at minimum and maximum vacuum settings to determine the minimum and maximum vacuum levels of the pump as compared to its specifications, which are identical to those of the predicate version of the Symphony pump. Tests were conducted over 20 minutes with the Standard 2.0 program card or 15 minutes with the Preemie+ program card to simulate a typical pumping session using size medium (M) breast shields. Tests began in stimulation mode and continued with expression mode after the first two minutes. There are two approved suppliers of the motor for the Symphony® breast pump. Testing of six pumps was completed with the Standard 2.0 program card, using three pumps with each motor type. One of each pump type was used for testing with the Preemie+ program card. Vacuum performance during normal use was first evaluated and demonstrated that, independent of the pump
{11}------------------------------------------------
motor type, the specification for vacuum level and cycle rate is met. Additional test cases evaluated performance of the pump when running on battery, performance with varying input voltages and with the other size breast shields. Finally, testing of performance for single pumping was completed. Results demonstrated that independent of the pump motor type, the specification for battery operation time is met, that the specification for vacuum level and cycle rate is met in single pumping mode as well as with all power source and with varying input voltage. With different breast shield sizes, the specification for the maximum vacuum level is met.
SUMMARY OF CLINICAL TESTS:
Clinical testing was not required to demonstrate the substantial equivalence of the Symphony breast pump to its predicate device. However, published research studies are referenced to support marketing claims.
CONCLUSION:
The differences between the Symphony breast pump and its predicate device do not introduce a new intended use and do not raise new issues of safety and effectiveness. Verification and Validation testing demonstrated that no adverse effects have been introduced by these differences and that the device performs as intended.
From the results of nonclinical testing described, Medela AG concludes that the Symphony breast pump is substantially equivalent to the legally marketed predicate device.
§ 884.5160 Powered breast pump.
(a)
Identification. A powered breast pump in an electrically powered suction device used to express milk from the breast.(b)
Classification. Class II (performance standards).