K Number
K151617
Date Cleared
2016-01-22

(221 days)

Product Code
Regulation Number
870.1220
Panel
CV
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Medline ReNewal Reprocessed St. Jude Medical Livewire Electrophysiology Catheter can be used in the evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites.

Device Description

Medline ReNewal Reprocessed St. Jude Livewire Electrophysiology Catheters are commonly placed at the high right atrium, right ventricular apex and His bundle, and in the coronary sinus, and are used for electrogram recording and cardiac stimulation during diagnostic electrophysiology studies.

AI/ML Overview

The Medline ReNewal Reprocessed St. Jude Medical Livewire Electrophysiology Catheter (K151617) underwent performance testing to demonstrate substantial equivalence to its predicate device, the St. Jude Medical Livewire Electrophysiology Catheter (K022380).

1. Table of Acceptance Criteria and Reported Device Performance:

The document provides a list of functional performance studies, cleaning validation, biocompatibility testing, packaging/sterilization validation, and product stability tests performed. However, it does not explicitly define specific numerical acceptance criteria for each test nor provides detailed quantitative results (e.g., "bond strength was X N, meeting the criteria of >Y N"). Instead, it states that the functional characteristics were "evaluated and found to be substantially equivalent to the predicate device."

Here’s a summary of the tests performed, which implicitly form the basis for acceptance:

Acceptance Criteria (Implicitly: Performance comparable to predicate)Reported Device Performance (Implicitly: Met comparable performance)
Functional Performance Studies:
Bond strength (tensile testing)Tested and found comparable
Torsional strengthTested and found comparable
Leakage currentTested and found comparable
Steerable deflection and continuityTested and found comparable
Catheter/handle joint flexibility and continuity (flex fatigue)Tested and found comparable
Electrode adherenceTested and found comparable
Applied force deflection (tip flexibility)Tested and found comparable
Shaft flexibilityTested and found comparable
Tip bucklingTested and found comparable
Direct current resistanceTested and found comparable
Impedance at 5 kHzTested and found comparable
Corrosion resistanceTested and found comparable
Cleaning:
Protein, total organic carbon, and endotoxinsTested and cleaning performance qualified
Visual inspection under magnificationTested and cleaning performance qualified
Cleaning performance qualificationPerformed and qualified
Biocompatibility:
CytotoxicityTested and found acceptable
SensitizationTested and found acceptable
IrritationTested and found acceptable
Acute systemic toxicityTested and found acceptable
PyrogenicityTested and found acceptable
Hemocompatibility (hemolysis, thrombogenicity, complement activation)Tested and found acceptable
Packaging and sterilization validation:
Bioburden testingTested
Ethylene oxide and ethylene chlorohydrin residuals testingTested
Product stabilityTested and found acceptable

2. Sample Size Used for the Test Set and Data Provenance:

The document does not specify the exact sample sizes used for each of the performance tests. It also does not mention the country of origin of the data or whether the studies were retrospective or prospective.

3. Number of Experts and Qualifications for Ground Truth:

This information is not applicable and not provided in the context of this 510(k) summary, as it describes a reprocessed medical device's physical and functional performance, not an AI/algorithmic medical device requiring expert ground truth for classification or diagnosis.

4. Adjudication Method for the Test Set:

This information is not applicable and not provided for the reasons stated in point 3. The testing involves laboratory and benchtop evaluations of device properties.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

An MRMC comparative effectiveness study was not performed, as it is not relevant for the type of device (reprocessed catheter) described. This type of study is typically used for diagnostic imaging or AI devices that interact with human readers/clinicians.

6. Standalone (Algorithm Only) Performance:

Standalone performance (algorithm only) was not performed, as it is not relevant for this reprocessed physical medical device. This concept applies to AI/software as a medical device.

7. Type of Ground Truth Used:

The "ground truth" for this device's performance evaluation is based on established engineering and biological standards relevant to medical devices, as well as comparison to the performance characteristics of the original, new device (predicate device). For instance, for bond strength, the "ground truth" would be the required mechanical integrity, and for cleaning, it would be the absence of specified contaminants below acceptable thresholds.

8. Sample Size for the Training Set:

This information is not applicable and not provided. The listed tests are for verifying the performance of the reprocessed device itself, not for training an algorithm.

9. How the Ground Truth for the Training Set Was Established:

This information is not applicable and not provided for the reasons stated in point 8.

§ 870.1220 Electrode recording catheter or electrode recording probe.

(a)
Identification. An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.(b)
Classification. Class II (performance standards).