(133 days)
In vitro test for the quantitative determination of carbamazepine in serum and plasma on Roche/Hitachi cobas c systems.
Measurements obtained are used in monitoring levels of carbamazepine to help ensure appropriate therapy.
The ONLINE TDM Carbamazepine Gen. 4 assay is for the quantitative determination of carbamazepine in human serum or plasma on automated clinical chemistry analyzers. It is a homogeneous microparticle agglutination immunoassay based on the kinetic interaction of microparticles in solution (KIMS). Biotinylated drug hapten serves as the binding partner to anticarbamazepine antibody and streptavidin coated latex beads. A competitive reaction to a limited amount of specific anti-carbamazepine antibody takes place between the hapten and free carbamazepine in the sample. A decrease in the apparent signal produced by the microparticle agglutination is proportional to the amount of drug present in the sample.
Here's an analysis of the provided text, focusing on acceptance criteria and the study proving the device meets them:
Device Name: ONLINE TDM Carbamazepine Gen.4
1. Table of Acceptance Criteria and Reported Device Performance:
| Acceptance Criteria Category | Specific Criteria/Metric | Reported Device Performance |
|---|---|---|
| Detection Limit | Limit of Blank (LoB) | 0.3 µg/mL (Claimed: 0.5 µg/mL) |
| Limit of Detection (LoD) | 0.5 µg/mL (Claimed: 1.0 µg/mL) | |
| Limit of Quantitation (LoQ) | 1.4 µg/mL (Claimed: 2.0 µg/mL) | |
| Precision | Repeatability (CV%) | TDM Control 1: 2.2% TDM Control 2: 1.4% TDM Control 3: 1.3% Human Serum 1: 2.7% Human Serum 2: 2.1% Human Serum 3: 1.8% Human Serum 4: 1.3% Human Serum 5: 1.4% |
| Intermediate Precision (CV%) | TDM Control 1: 2.8% TDM Control 2: 2.3% TDM Control 3: 1.8% Human Serum 1: 3.3% Human Serum 2: 2.9% Human Serum 3: 2.6% Human Serum 4: 2.4% Human Serum 5: 2.9% | |
| Linearity | Pearson correlation coefficient (R) | Serum: 0.997 K2-EDTA, Plasma: 0.999 |
| Measuring Range | Claimed Measuring Range | 2.0 to 20.0 µg/mL |
| Matrix Comparison | Correlation (r) | Serum vs. Serum Gel Separation: 0.989 Serum vs. Li-heparin: 0.991 Serum vs. Na-heparin: 0.990 Serum vs. K2-EDTA: 0.988 Serum vs. K3-EDTA: 0.994 |
| Interferences (Endogenous) | Hemolysis (H index) | Up to 1000 (approx. Hgb 1000 mg/dL) |
| Icterus (I index) | Up to 50 (approx. bilirubin 50 mg/dL) | |
| Lipemia (L index) | Up to 2000 (approx. triglyceride 1000 mg/dL) | |
| Cholesterol | Up to 600 mg/dL | |
| Rheumatoid Factor | Up to 1200 IU/mL | |
| Total Protein | Up to 13 g/dL | |
| Interferences (Drugs) | Non-interference at specific concentrations for 16 common drugs | (See Table 7 in document for specific concentrations) |
| Cross-reactivity | % Cross reactivity at 3 µg/mL and 12 µg/mL for various compounds | (See Table 8 in document for specific values for 27 compounds) |
| Method Comparison to Predicate | Deming Regression (r) | 0.993 |
2. Sample Size Used for the Test Set and Data Provenance:
- Detection Limit (LoB): 60 measured values (10-fold determinations per run on one instrument, 6 runs over 3 days).
- Detection Limit (LoD): 36 determinations (5 samples, 2-fold determination per run, 6 runs over 3 days).
- Detection Limit (LoQ): Not explicitly stated, but "Nine samples are prepared... tested in two aliquots over at least three days on one analyzer, 2 runs per day for 3 lots." This implies a significant number of data points.
- Precision: "Two runs per day for ≥ 21 days on the same analyzer." This amounts to at least 42 runs for each sample type. Numerous samples were tested (TDM Controls 1, 2, 3 and Human Serums 1-5).
- Linearity: "Eleven levels (including the high concentration pool and diluent)." Tested for both human serum and plasma.
- Matrix Comparison: "33 full tubes" of paired serum and plasma samples from single donors.
- Interferences (Endogenous): Two human sample pools, varying concentrations of interfering substances.
- Interferences (Drugs): Two human sample pools, spiked with 16 different drugs.
- Cross-reactivity: Two carbamazepine concentrations (3 µg/mL and 12 µg/mL) tested against 27 different compounds.
- Method Comparison to Predicate: "One hundred single native human samples of patients taking carbamazepine."
Data Provenance: The document does not explicitly state the country of origin for the samples. It mentions "human serum," "human serum sample pool," and "single native human samples of patients." It is implied to be a retrospective analysis of samples collected for the purpose of testing, but it's not explicitly stated as retrospective or prospective.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
Not applicable. This device is an in-vitro diagnostic (IVD) for quantitative determination of a drug concentration. The "ground truth" for these tests is established by reference methods, calibrators, and known concentrations, not by expert interpretation of images or clinical cases. For example, LoQ uses LC/MS as the expected/target value.
4. Adjudication Method for the Test Set:
Not applicable. As this is an IVD for quantitative measurement, there is no subjective interpretation requiring adjudication among experts. The "ground truth" is determined by analytical methods with known accuracy and precision (e.g., LC/MS for LoQ).
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:
No, an MRMC study was not done. This type of study is relevant for imaging devices or tests where human readers interpret results, often with and without AI assistance. This device is an automated quantitative assay, so human reader involvement in the primary measurement is not a factor.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:
Yes, the entire performance evaluation presented is a standalone (algorithm/device-only) performance assessment. The device quantifies carbamazepine levels directly from samples. The results are compared against established analytical standards and a predicate device.
7. The Type of Ground Truth Used:
The ground truth for the performance studies is established by:
- Reference Methods: For LoQ, LC/MS (Liquid Chromatography/Mass Spectrometry) is explicitly stated as the method for determining the expected or target value.
- Known Concentrations: Samples with known concentrations of carbamazepine are used for various tests like detection limits, precision, linearity, and interference studies.
- Predicate Device Comparison: The reference standard for the "Method Comparison to Predicate" is the results obtained from the predicate device (ONLINE TDM Carbamazepine, K031902) on the cobas c 501.
- USP Reference Standards: The traceability for the assay calibration is stated as being standardized against USP reference standards.
8. The Sample Size for the Training Set:
Not applicable. This device is an immunoassay (KIMS technology), not a machine learning or AI-based algorithm that requires a "training set" in the conventional sense. The "training" of such a system would involve optimizing assay reagents and parameters during development, not a data-driven training dataset for an AI model.
9. How the Ground Truth for the Training Set Was Established:
Not applicable for the same reasons as point 8. The device's "training" is in its chemical and biological design, optimization, and manufacturing, rather than data-driven ground truth establishment for a training set.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of three stylized human profiles facing to the right, with a design that resembles an abstract bird or wing shape above them. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 22, 2015
ROCHE DIAGNOSTICS OPERATIONS (RDO) DAVID TRIBBETT REGULATORY AFFAIRS PRINCIPAL 9115 HAGUE ROAD INDIANAPOLIS IN 46250
Re: K151578
Trade/Device Name: ONLINE TDM Carbamazepine Gen 4 Regulation Number: 21 CFR 862.3645 Regulation Name: Neuroleptic drugs radioreceptor assay test system Regulatory Class: II Product Code: KLT Dated: September 21, 2015 Received: September 22, 2015
Dear Mr. Tribbett:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Courtney Hias -S
Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K151578
Device Name ONLINE TDM Carbamezapine Gen.4
Indications for Use (Describe)
In vitro test for the quantitative determination of carbamazepine in serum and plasma on Roche/Hitachi cobas c systems.
Measurements obtained are used in monitoring levels of carbamazepine to help ensure appropriate therapy.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|---|---|
| ------------------------------------------------------------ | ----------------------------------------------------------- |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
ONLINE TDM Carbamazepine Gen. 4 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
| Submitter Name | Roche Diagnostics Operations (RDO) |
|---|---|
| Address | 9115 Hague RoadIndianapolis, IN, 46250, USA |
| Contact | David TribbettPhone: (317)-521-2964FAX: (317)-521-2324Email: david.tribbett@roche.com |
| Date Prepared | October 22, 2015 |
| Proprietary Name | ONLINE TDM Carbamazepine Gen. 4 |
| Common Name | Enzyme immunoassay, Carbamazepine |
| Classification Name | Neuroleptic drugs radioreceptor assay test system |
| Product Codes | KLT, 21 CFR § 862.3645 |
| Predicate Devices | ONLINE TDM Carbamazepine, K031902 |
| Establishment Registration | 1823260, Roche Diagnostics Corporation |
{4}------------------------------------------------
DEVICE DESCRIPTION 1.
The ONLINE TDM Carbamazepine Gen. 4 assay is for the quantitative determination of carbamazepine in human serum or plasma on automated clinical chemistry analyzers. It is a homogeneous microparticle agglutination immunoassay based on the kinetic interaction of microparticles in solution (KIMS). Biotinylated drug hapten serves as the binding partner to anticarbamazepine antibody and streptavidin coated latex beads. A competitive reaction to a limited amount of specific anti-carbamazepine antibody takes place between the hapten and free carbamazepine in the sample. A decrease in the apparent signal produced by the microparticle agglutination is proportional to the amount of drug present in the sample.
2. INDICATIONS FOR USE
In vitro test for the quantitative determination of carbamazepine in serum and plasma on Roche/Hitachi cobas c systems.
Measurements obtained are used in monitoring levels of carbamazepine to help ensure appropriate therapy.
TECHNOLOGICAL CHARACTERISTICS 3.
The ONLINE TDM Carbamazepine Gen. 4 assay is a homogeneous microparticle agglutination immunoassay.
It is a two-reagent system used for the detection of carbamazepine in serum. Kinetic interaction of microparticles in solution (KIMS) will be measured using automated analyzers. In this technology biotinylated drug hapten attached to streptavidin coated latex beads serves as the binding partner to anti-carbamazepine antibody. A competitive reaction to a limited amount of specific anti-carbamazepine antibody takes place between the latex bound hapten and free carbamazepine in the serum sample. A decrease in the apparent signal is proportional to the amount of drug present in the sample.
Reagents - working solutions which are ready for use, are packaged in a cassette labeled with their instrument positioning B (Reagent 1) and C (Reagent 2).
{5}------------------------------------------------
- R1, Anti-carbamazepine antibody (sheep monoclonal); MES® buffer, pH 6.4; . preservative
- R2, Carbamazepine biotinylated hapten; streptavidin coated latex microparticles: 0.1 . %; HEPES") buffer, pH 7.4; preservative
- 2-(N-Morpholino) ethanesulfonic acid a.
- b. N-(2-Hydroxyethyl)piperazine-N' -(2-ethanesulfonic acid)
The following table compares the similarities and differences of the candidate device to the predicate device.
| Assay Comparison Similarities | ||
|---|---|---|
| Feature | Predicate Device:ONLINE TDM Carbamazepine (K031902) | Candidate Device:ONLINE TDM Carbamazepine Gen. 4 |
| Intended Use | The ONLINE TDM Carbamazepine assay isfor the quantitative determination ofcarbamazepine in human serum or plasmaon automated clinical chemistry analyzers. | In vitro test for the quantitativedetermination of carbamazepine in serumand plasma on Roche/Hitachi cobas csystems. |
| Sample Types | Serum: Collect serum using standardsampling tubes.Plasma: Potassium EDTA, sodium orlithium heparin plasma. | Serum: Collect serum using standardsampling tubes.Plasma: K2- and K3-EDTA, sodium orlithium heparin plasma. |
| Test Principle | Homogeneous microparticleagglutination immunoassay. | Same |
| Reagent Shelf LifeStability | 2-8 °C until expiration date | 2-8 °C until expiration date |
| Reagent On-BoardStability | 30 days opened and refrigerated on theanalyzer. Do not freeze. | On-board in use and refrigerated on theanalyzer: 4 weeks |
| Measuring Range | 0.2-20 µg/mL | 2-20 µg/mL (8.5-85 µmol/L) |
| Traceability | This method has been standardized againstUSP reference standards. | Same |
| Calibrator | COBAS-FP Carbamazebine Calibrator,CAL A-F | Preciset TDM I, calibrators B-F(Previously cleared 510(k): K031856) |
| Calibrationfrequency | After reagent cassette change, after reagentlot change and as required following qualitycontrol procedures | After reagent lot change and as requiredfollowing quality control procedures |
| Controls | TDM Control Set, levels I, II and III | Same |
| Table 1: Substantial Equivalence – Assay Similarities and Differences |
|---|
| ----------------------------------------------------------------------- |
{6}------------------------------------------------
| Assay Comparison Similarities | ||
|---|---|---|
| Feature | Predicate Device:ONLINE TDM Carbamazepine (K031902) | Candidate Device:ONLINE TDM Carbamazepine Gen. 4 |
| Lower Limits ofMeasurement | Lower Detection Limit = 0.2 µg/mL(0.81 µmol/L) | LoB = 0.3 µg/mL (1.3 µmol/L)LoD = 0.5 µg/mL (2.1 µmol/L)LoQ = 1.4 µg/mL (5.9 µmol/L) |
| Test Principle | The ONLINE TDM Carbamazepine assay isa homogeneous microparticle agglutinationimmunoassay. It is a two-reagent systemused for the detection ofcarbamazepine in serum. Kinetic interactionof microparticles in solution (KIMS) will bemeasured using RD/Hitachi families ofautomated analyzers. In this technologybiotinylated drug hapten serves as thebinding partner to 1) anti-carbamazepineantibody and 2) streptavidin coated latexbeads. A competitive reaction to a limitedamount of specific anti-carbamazepineantibody takes place between the haptenand free carbamazepine in the serumsample. A decrease in the apparent signalis proportional to the amount of drugpresent in the sample. | The ONLINE TDM Carbamazepine Gen. 4assay is a homogeneous microparticleagglutination immunoassay. It is a two-reagent system used for the detection ofcarbamazepine in serum. Kineticinteraction of microparticles (KIMS) will bemeasured using automated analyzers. Inthis technology biotinylated drug haptenattached to streptavidin coated latexbeads serves as the binding partner toanti-carbamazepine antibody. Acompetitive reaction to a limited amount ofspecific anti-carbamazepine antibodytakes place between the latex boundhapten and free carbamazepine in theserum sample. A decrease in the apparentsignal is proportional to the amount ofdrug present in the sample. |
| Reagent Composition | R1 Conjugate Reagent/BufferCarbamazepine biotinylated hapten;2-(N-Morpholino) ethanesulfonic acid (MES)buffer, pH 6.4; preservative; surfactantR2 Latex-Antibody Reagent/BufferAnti-carbamazepine antibody (mousemonoclonal); streptavidin coated latexmicroparticles: 0.08%; N-(2-Hydroxyethyl)piperazine-N'-(2-ethanesulfonic acid)(HEPES) buffer, pH 7.5; preservative | R1 Anti-carbamazepine antibody (sheepmonoclonal); MES) buffer, pH 6.4;preservativeR2 Carbamazepine biotinylated hapten;streptavidin coated latex microparticles:0.1 %; HEPES) buffer, pH 7.4;preservative |
{7}------------------------------------------------
NON-CLINICAL PERFORMANCE EVALUATION 4.
The following performance data were provided in support of the substantial equivalence determination:
Detection Limit: LoB, LoD and LoQ according to CLSI EP17-A2 Precision according to CLSI EP5-A2 Linearity according to CLSI EP6-A Matrix Comparison - Anticoagulants Interferences Interference - Drugs Method Comparison to Predicate
Detection Limit 4.1.
LoB, LoD, and LoQ studies were performed based upon CLSI EP17-A2.
LoB:
The diluent is measured with 10-fold determinations per run on one instrument. Six runs distributed over 3 days are performed. Data analysis will be based on determination of the 95th percentile of the 60 measured values.
LoD:
The 5 samples with low analyte content spiked with Carbamazepine (with concentrations ranging from LoB to approx. 4 times specified LoB) are measured with 2-fold determination per run. Six runs distributed over 3 days are performed.
LoD is defined as the concentration, at which there is a 95% probability that a sample contains analyte.
{8}------------------------------------------------
LoQ is the lowest amount of analyte in a sample that can be detected quantitatively within specified precision and accuracy ranges.
Nine samples are prepared which cover the concentration range between LoB and 2x LoQ. Those samples are tested in two aliquots over at least three days on one analyzer, 2 runs per day for 3 lots.
Expected or target value is determined with LC/MS.
Table 2: LoB, LoD, and LoQ Experimental Determination
| Representative Result (µg/mL) | Claimed in labeling (µg/mL) | |
|---|---|---|
| Limit of Blank (LoB) | 0.3 | 0.5 |
| Limit of Detection (LoD) | 0.5 | 1.0 |
| Limit of Quantitation (LoQ) | 1.4 | 2.0 |
Precision according to CLSI EP5-A 4.1.
Two runs per day for ≥ 21 days on the same analyzer. Repeatability (within run precision) and intermediate precision (within lab precision) is calculated. The samples have to be randomized in each run separately.
The data set has to be complete for the 21 days.
| Table 3: Repeatability Summary | |||||
|---|---|---|---|---|---|
| -- | -- | -- | -- | -------------------------------- | -- |
| Specimen | Mean (ug/mL) | SD (ug/mL) | CV (%) |
|---|---|---|---|
| TDM Control 1 | 3.4 | 0.08 | 2.2 |
| TDM Control 2 | 9.7 | 0.14 | 1.4 |
| TDM Control 3 | 15.7 | 0.20 | 1.3 |
| Human Serum 1 | 2.9 | 0.08 | 2.7 |
| Human Serum 2 | 4.2 | 0.09 | 2.1 |
| Human Serum 3 | 9.4 | 0.17 | 1.8 |
| Human Serum 4 | 14.6 | 0.19 | 1.3 |
| Human Serum 5 | 19.5 | 0.27 | 1.4 |
{9}------------------------------------------------
| Specimen | Mean (ug/mL) | SD (ug/mL) | CV (%) |
|---|---|---|---|
| TDM Control 1 | 3.4 | 0.10 | 2.8 |
| TDM Control 2 | 9.7 | 0.22 | 2.3 |
| TDM Control 3 | 15.7 | 0.28 | 1.8 |
| Human Serum 1 | 2.9 | 0.10 | 3.3 |
| Human Serum 2 | 4.2 | 0.12 | 2.9 |
| Human Serum 3 | 9.4 | 0.25 | 2.6 |
| Human Serum 4 | 14.6 | 0.36 | 2.4 |
| Human Serum 5 | 19.5 | 0.56 | 2.9 |
Table 4: Intermediate Precision Summary
Linearity according to CLSI EP6-A 4.2.
A dilution series was prepared from a human serum sample pool and diluent (Analyte-free serum). The dilution series are prepared to obtain eleven levels (including the high concentration pool and diluent). The diluted samples shall span the measuring range including a sample at the lower end of the measuring range, a sample over the measuring range and samples at the medical decision points. The process was repeated for plasma samples.
The calculation is according to the CLSI guideline EP6-A. All measurement data of the dilution steps are calculated by regression.
| Sample Type | Linear Regression Equation | Claimed Measuring Range |
|---|---|---|
| Serum | $y = 1.000x - 0.0$Pearson correlation coefficient (R) = 0.997 | 2.0 to 20.0 µg/mL |
| K2-EDTA, Plasma | $y = 1.013x - 0.195$Pearson correlation coefficient (R) = 0.999 | 2.0 to 20.0 µg/mL |
Table 5: Linearity Results
{10}------------------------------------------------
4.3. Matrix Comparison - Anticoagulants
Each pair of serum and plasma of a single donor are spiked with carbamazepine. Included in the data are 33 full tubes as described below.
A matrix comparison is executed by taking the serum as reference. Only samples within the limit of icteric, lipemic and hemolytic interference are allowed to be used. Samples cover the measuring range.
The following matrix comparisons are provided:
- Gel separation tubes .
- K2-EDTA plasma vs serum .
- K3-EDTA plasma vs serum .
- Li-Heparin plasma vs serum .
- Na-Heparin plasma vs serum .
Table 6: Matrix Comparison
| Anticoagulant | Correlation |
|---|---|
| Serum vs. Serum Gel Separation | $y = 1.01x + 0.177, r = 0.989$ |
| Serum vs. Li-heparin | $y = 1.01x -0.290, r = 0.991$ |
| Serum vs. Na-heparin | $y = 1.02x -0.382, r = 0.990$ |
| Serum vs. K2-EDTA | $y = 1.02x -0.059, r = 0.988$ |
| Serum vs. K3-EDTA | $y = 0.993x + 0.147, r = 0.994$ |
Interferences - H, L and I Indices 4.4.
The effect on quantitation of analyte in the presence of endogenous interfering substances is determined at two carbamazepine concentrations and a dilution set of the added interfering substances. The following lists the potential interfering substances evaluated and the highest concentration which was shown to not interfere with the assay:
Hemolysis: up to an H index of 1000 (approximate hemoglobin concentration 1000 mg/dL)
{11}------------------------------------------------
Icterus/Bilirubin: up to an I index of 50 (approximate bilirubin concentration 50 mg/dL) Lipemia: up to an L index of 2000 (approximate triglyceride concentration up to 1000 mg/dL) Cholesterol: up to 600 mg/dL Rheumatoid Factor: up to 1200 IU/mL
Total Protein: up to 13 g/dL
Test procedure:
High concentrated stock solutions of the interference substances are prepared in a suitable solvent. Two human sample pools are spiked with the defined carbamazepine concentrations and each divided into two aliguots. The potential interfering substance is added to one aliquot of each pool, while the other aliquot is mixed with the same amount of solvent without the interfering substance. A dilution series is prepared with at least 10 dilution steps for each interferent by mixing the two aliquots.
The parts containing the interfering substance have the same carbamazepine concentrations as the aliquots containing no interfering substance. When diluting those two aliquots, the carbamazepine concentration remains constant while the concentration of interferent varies. Thus, the effect of increasing concentrations of interferent can be determined.
Median of the measured results is compared to the expected result (aliquot with no interfering substance) and the recovery is determined (paired difference testing).
This procedure is repeated for each of the interfering substances.
Interferences – Drugs 4.5.
Sixteen commonly used drugs were examined for potential interference on measurement with ONLINE TDM Carbamazepine Gen. 4.
Two human sample pools spiked with the defined carbamazepine concentrations are divided into two aliquots. One aliquot of each concentration is used as the reference sample for carbamazepine concentration and is not spiked with the drugs but the solvent for the drug.
{12}------------------------------------------------
The other aliquots, with either the high or low carbamazepine concentration, are spiked with the respective amount of drug. The carbamazepine concentration of the spiked aliquots is determined in triplicate and compared to the carbamazepine concentration determined for the reference aliquot.
The defined pharmaceutical compounds are spiked into samples with concentrations according to EP7-A2 or higher concentrations. The mean values are calculated as well as the % deviations from the interference samples compared to the reference sample (paired difference testing).
| Drug | Highest Concentration ShownNot to Interfere withCarbamazepine (drugconcentrations in (mg/L)) |
|---|---|
| Acetylcysteine | 1660 |
| Ampicillin-Na | 1000 |
| Ascorbic acid | 300 |
| Cefoxitin | 2500 |
| Heparin | 5000 U/L |
| Levodopa | 20 |
| Methyldopa | 20 |
| Metronidazol | 200 |
| Doxycyclin | 50 |
| Acetylsalicylic Acid | 1000 |
| Rifampicine | 60 |
| Cyclosporine | 5 |
| Acetaminophen | 200 |
| Ibuprofen | 500 |
| Phenylbutazone | 400 |
| Theophyllin | 100 |
Table 7: Common Drug Interferences
{13}------------------------------------------------
4.6. Cross-reactivity
The following compounds were tested for cross-reactivity at low (3 µg/mL) and high (12 µg/mL) carbamazepine concentration:
Table 8: Cross-reactivity Testing
| Compound | Concentrationtested(µg/mL) | % Crossreactivity at 3µg/mL | % Crossreactivity at 12µg/ml |
|---|---|---|---|
| Carbamazepine-10,11-epoxide | 29.6 µg/mL | 2.87 | 1.4 |
| Oxcarbazepine (Oxc) | 100 µg/mL | 0.88 | 0.2 |
| 10-Hydroxycarbamaze-pine (MHD) | 100 µg/mL | 0.63 | 0.2 |
| Nortriptyline | 50 µg/mL | 0 | 0.3 |
| Amitriptyline | 100 µg/mL | 0 | 0 |
| Imipramine | 200 µg/mL | 0 | 0 |
| Phenothiazine | 200 µg/mL | 0 | 0 |
| Phenylbutazone | 450 µg/mL | 0.05 | 0 |
| Promethazine | 1000 µg/mL | 0.02 | 0 |
| Phenytoin | 1000 µg/mL | 0 | 0 |
| Mephenytoin | 1000 µg/mL | 0.5 | 0.1 |
| 2-Phenyl-2-ethylmalonamide | 1000 µg/mL | 0.31 | 0.2 |
| Ethotoin | 1000 µg/mL | 0.12 | 0.10 |
| Valproic acid | 1000 µg/mL | 0.02 | 0 |
| Amobarbital | 1000 µg/mL | 0 | 0 |
| Chlordiazepoxide | 30 µg/mL | 0.33 | 0.4 |
| Clonazepam | 12 µg/mL | 0.44 | 0.3 |
| Ethosuximide | 1000 µg/mL | 0.01 | 0 |
| Diazepam | 25 µg/mL | 0.21 | 0.4 |
| Gluthethimide | 1000 µg/mL | 0 | 0 |
| Methosuximide | 100 µg/mL | 0.01 | 0 |
| p-Hydroxypheno-barbital | 100 µg/mL | 0.05 | 0.4 |
| 5-(p-Hydroxyphenyl)-phenylhydantoin | 1000 µg/mL | 0.01 | 0 |
| Phenobarbital | 1000 µg/mL | 0.01 | 0 |
| Primidone | 1000 µg/mL | 0.02 | 0 |
| Probenecid | 500 µg/mL | 0.03 | 0 |
| Secobarbital | 1000 µg/mL | 0.02 | 0 |
{14}------------------------------------------------
Method Comparison to Predicate 4.7.
One hundred single native human samples of patients taking carbamazepine covering the reportable range are tested.
The samples are tested in singlicate on the candidate and predicate device (cobas c 501).
The data was evaluated using Deming Regression analysis.
y = 0.994x - 0.054 µg/mL r = 0.993
CONCLUSIONS 5.
The submitted information in this premarket notification supports a substantial equivalence decision.
§ 862.3645 Neuroleptic drugs radioreceptor assay test system.
(a)
Identification. A neuroleptic drugs radioceptor assay test system is a device intended to measure in serum or plasma the dopamine receptor blocking activity of neuroleptic drugs and their active metabolites. A neuroleptic drug has anti-psychotic action affecting principally psychomotor activity, is generally without hypnotic effects, and is a tranquilizer. Measurements obtained by this device are used to aid in determining whether a patient is taking the prescribed dosage level of such drugs.(b)
Classification. Class II.