K Number
K151572

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-07-10

(29 days)

Product Code
Regulation Number
882.5950
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Penumbra Smart Coil System is indicated for the embolization of:

  • · Intracranial aneurysms
  • · Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae
  • · Arterial and venous embolizations in the peripheral vasculature
Device Description

The Smart Coil functions to selectively embolize targeted segments of the vasculature by packing a sufficient quantity of bare platinum coils to achieve occlusion. The Smart Coil System consists of three components: a Coil Implant attached to a Detachment Pusher and a Detachment Handle.

AI/ML Overview

The provided text is a 510(k) premarket notification for the Penumbra Smart Coil. This document is a regulatory submission to the FDA aiming to demonstrate that a new medical device is substantially equivalent to a legally marketed predicate device.

Crucially, this document is NOT a study that proves the device meets acceptance criteria. Instead, it's a justification for market clearance based on a comparison to an already approved device. Therefore, it does not contain the information requested in your prompt regarding acceptance criteria, device performance metrics, sample sizes for test or training sets, expert qualifications, or adjudication methods for a new study.

The document explicitly states:

  • "A literature review was conducted to substantiate the safe and effective performance of the Smart Coil and Smart Coil Detachment Handle, as well as its substantial equivalence to the predicate device." This indicates that no new primary clinical study was performed for this specific 510(k) submission to establish performance against acceptance criteria.
  • "Review concluded that revised labeling is appropriate." The core of this 510(k)'s purpose is to update the Instructions for Use (IFU), specifically revising a precaution and updating the manufacturer's address. It's not about proving new performance.
  • "The Penumbra Smart Coil was found to have a safety and effectiveness profile that is similar to the predicate device." The "study" here is a comparison to a predicate device that was already cleared for market (K143218, cleared March 18, 2015).

Therefore, I cannot populate the requested table and answer the questions because the provided text describes a regulatory submission based on demonstrating substantial equivalence to a predicate device, not a new clinical or performance study with defined acceptance criteria and results.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with three curved lines above them. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 10, 2015

Penumbra, Inc. Mr. Charles DeNault Regulatory Affairs Specialist One Penumbra Place Alameda, California 94502

Re: K151572

Trade/Device Name: Penumbra Smart Coil™ Regulation Number: 21 CFR 882.5950 Regulation Name: Neurovascular Embolization Device Regulatory Class: Class II Product Code: HCG, KRD Dated: June 10, 2015 Received: June 11, 2015

Dear Mr. DeNault:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Carlos L. Pena -S 同公

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K151572

Device Name Penumbra Smart Coil

Indications for Use (Describe)

The Penumbra Smart Coil System is indicated for the embolization of:

  • · Intracranial aneurysms
  • · Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae
  • · Arterial and venous embolizations in the peripheral vasculature

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CER 801 Subpart C)
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

1 510(k) Summary

(as required by 21 CFR 807.92)

Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Penumbra Inc. is providing the summary of Substantial Equivalence for the Penumbra Smart Coil™

1.1 Sponsor/Applicant Name and Address

Penumbra, Inc. One Penumbra Place Alameda, CA 94502 USA

1.2 Sponsor Contact Information

Charles DeNault Regulatory Affairs Specialist Phone: (510) 748-3302 Fax: (510) 217-6414 Email: cdenault@penumbrainc.com

Date of Preparation of 510(k) Summary 1.3

July 6, 2015

Device Trade or Proprietary Name 1.4

Penumbra Smart Coil™

Primary Device Classification 1.5

Regulatory Class:II
Classification Panel:Neurology
Classification Name:Neurovascular embolization device
Regulation Number:21 CFR 882.5950
Product Code:HCG

1.6 Secondary Device Classification

Regulatory Class:II
Classification Panel:Cardiovascular
Classification Name:Vascular embolization device
Regulation Number:21 CFR 870.3300
Product Code:KRD

1.7 Predicate Devices

510(k) NumberClearance DateName of Predicate DeviceName of Manufacturer
K143218March 18, 2015Penumbra Smart Coil SystemPenumbra, Inc.

{4}------------------------------------------------

Predicate Comparison 1.8

Penumbra Smart CoilPenumbra Smart Coil
ClassificationClass II, HCG, KRDSame
Indications for UseIndicated for the embolization of:• Intracranial aneurysms• Other neurovascular abnormalitiessuch as arteriovenous malformationsand arteriovenous fistulae• Arterial and venous embolizations inthe peripheral vasculatureSame
Materials
Coil ImplantPlatinum/Tungsten, Nitinol, Polymer,AdhesivesSame
Inner CoilStainless Steel, Polymer,Platinum/Tungsten, NitinolSame
Detachment HandlePlasticSame
Dimensions/Shape
Coil Secondary Diameter1-18 mmSame
Coil Length1-60 cmSame
Coil Secondary ShapeComplex, Helical (Curve)Same
Pusher Length185 cmSame
Sterilization
Sterilization MethodEtOSame

Device Description 1.9

The Smart Coil functions to selectively embolize targeted segments of the vasculature by packing a sufficient quantity of bare platinum coils to achieve occlusion. The Smart Coil System consists of three components: a Coil Implant attached to a Detachment Pusher and a Detachment Handle.

The purpose of this 510(k) pre-market notification is to implement the following modifications to the Instructions for Use (IFU):

  • . Revise precaution
  • . Update legal address of manufacturer

1.10 Indications for Use

The Penumbra Smart Coil System is indicated for the embolization of:

  • . Intracranial aneurysms

{5}------------------------------------------------

  • . Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae
  • . Arterial and venous embolizations in the peripheral vasculature

1.11 Summary of Non-Clinical Data

A literature review was conducted to substantiate the safe and effective performance of the Smart Coil and Smart Coil Detachment Handle, as well as its substantial equivalence to the predicate device. Current literature was reviewed in relationship to device precautions. Review concluded that revised labeling is appropriate. Additionally, revised labeling was found to have a safety and effectiveness profile that is similar to the predicate device.

1.12 Summary of Substantial Equivalence

The Penumbra Smart Coil was found to have a safety and effectiveness profile that is similar to the predicate device.

1.13 Conclusion

The non-clinical data supports the safety of the device and the Penumbra Smart Coil should perform as intended in the specified use conditions.

§ 882.5950 Neurovascular embolization device.

(a)
Identification. A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.(b)
Classification. Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).