K Number
K151530
Date Cleared
2015-07-07

(29 days)

Product Code
Regulation Number
862.1345
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CardioChek Plus Test System is a small portable analyzer and test strip system intended for multiple patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices. This system is for in vitro diagnostic use only. The tests strips are for the quantitative determination of glucose, total cholesterol, HDL (high density lipoprotein) cholesterol and triglycerides in venous whole blood from the fingertip. A Chol/HDL ratio and estimated values for LDL (low density lipoprotein) cholesterol and non-HDL cholesterol are calculated by the CardioChek Plus analyzer.

• Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.

• HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.

• Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellius, nephrosis, liver obstruction, other diseases involving lipid metabolism or various endocrine disorders.

• Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Device Description

The CardioChek Plus Test System includes the CardioChek Plus professional analyzer and analyte-specific test strips (PTS Panels eGLU test strips, PTS Panels Glucose test strips and PTS Panels Lipid Panel test strips). The CardioChek Plus Test System is modified in this submission by the addition of the CardioChek ChekMate strips. CardioChek ChekMate strips are dry strips that mimic the use of the PTS PANELS test strips to check the CardioChek analyzer system optics, calibration and result handling algorithms. ChekMate strips should not be used as a substitute for liquid quality control materials. There is no change to any of the test strips in any of the above named systems.

AI/ML Overview

The provided text describes the CardioChek Plus Test System and its modification with the addition of CardioChek ChekMate strips. The 510(k) summary focuses on the performance characteristics of these new ChekMate strips, particularly their precision and stability.

Here's an analysis of the acceptance criteria and study information provided:

1. Table of Acceptance Criteria and Reported Device Performance:

The document does not explicitly state numerical acceptance criteria in a dedicated table format. However, it reports on the precision of the CardioChek ChekMate strips.

Performance CharacteristicAcceptance Criteria (Implicit from Results)Reported Device Performance (CardioChek ChekMate Strips)
Intermediate Precision (% CV)< 3.0% (as demonstrated by results)< 3.0% for within-run, between-run, between-day, and total % CV
Real-time StabilityStudy supports claim18 months
Re-use StabilityStudy supports claim500 uses

2. Sample size used for the test set and the data provenance:

  • Sample Size for Test Set:
    • For the Intermediate Precision evaluation, three lots of ChekMate strips were evaluated. Two analyzers were used for each lot. The evaluation was performed over a 20-day period. (The exact number of individual samples/measurements is not specified beyond "three lots" and "two analyzers per lot").
    • For Value Assignment and Stability, the document states "Previously reported in K142302", implying these studies were conducted and reported in a prior submission, and details are not provided in this document.
  • Data Provenance: Not explicitly stated (e.g., country of origin). The studies appear to be internal validation studies conducted by the manufacturer, Polymer Technology Systems, Inc., based in Indianapolis, IN, USA. The document does not specify if the data is retrospective or prospective, but clinical performance studies for IVD devices are typically prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not provided in the document. The studies described are analytical performance studies (precision, stability) of a diagnostic device component rather than studies requiring expert interpretation of clinical data to establish ground truth.

4. Adjudication method for the test set:

This information is not applicable and not provided. The studies described are analytical performance studies focused on the device's technical specifications and do not involve clinical adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This is not applicable and not provided. The CardioChek Plus Test System is an in vitro diagnostic device for quantitative determination of analytes (glucose, cholesterol, etc.) from blood samples. It is not an AI-assisted diagnostic imaging or interpretation system that would involve human readers or MRMC studies.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

This is not applicable and not provided in the context of an "algorithm only" performance for interpretive tasks. The "standalone" performance for this device refers to its analytical performance (e.g., precision, accuracy, linearity) in quantifying specific biomarkers, which is what the precision and stability studies address. It is an automated system providing quantitative results.

7. The type of ground truth used:

For the precision studies, the "ground truth" is typically established by the inherent statistical properties of repeated measurements, comparing individual readings to the mean of those readings to assess variability. For value assignment, the ground truth would typically be established against a reference method or certified reference material, although this detail is not provided in this document as it was "previously reported in K142302".

8. The sample size for the training set:

This information is not applicable and not provided. The CardioChek Plus Test System and ChekMate strips are not described as employing machine learning or AI algorithms that would require a distinct "training set" in the conventional sense. The "training" for such devices typically involves calibration curves and algorithms derived from analytical experiments, not a machine learning training dataset.

9. How the ground truth for the training set was established:

This information is not applicable and not provided for the same reasons as point 8.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is placed to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 30, 2015

POLYMER TECHNOLOGY SYSTEMS, INC. MEGAN BURNS REGULATORY AFFAIRS ASSOCIATE 7736 ZIONSVILLE ROAD INDIANAPOLIS IN 46268

Re: K151530

Trade/Device Name: CardioChek Plus Test System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: CGA, CHH, JGY, LBR Dated: June 5, 2015 Received: June 8, 2015

Dear Ms. Megan Burns:

This letter corrects our substantially equivalent letter of July 7, 2015.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Courtney H. Lias -S

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K151530

Device Name

CardioChek Plus Test System

Indications for Use (Describe)

The CardioChek Plus Test System is a small portable analyzer and test strip system intended for multiple patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices. This system is for in vitro diagnostic use only. The tests strips are for the quantitative determination of glucose, total cholesterol, HDL (high density lipoprotein) cholesterol and triglycerides in venous whole blood from the fingertip. A Chol/HDL ratio and estimated values for LDL (low density lipoprotein) cholesterol and non-HDL cholesterol are calculated by the CardioChek Plus analyzer.

· Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.

· HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.

· Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellius, nephrosis, liver obstruction, other diseases involving lipid metabolism or various endocrine disorders.

· Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5: 510(k) Summary

1. Applicant Information:

Polymer Technology Systems, Inc. 7736 Zionsville Road Indianapolis, IN 46268

Contact Person: Megan A. Burns, RAC Phone: 317-870-5610 Fax: 317-870-5608

2. Date of Preparation:

July 2, 2015

3. Trade Name:

CardioChek Plus Test System

4. Device Description:

The CardioChek Plus Test System includes the CardioChek Plus professional analyzer and analyte-specific test strips (PTS Panels eGLU test strips, PTS Panels Glucose test strips and PTS Panels Lipid Panel test strips). The CardioChek Plus Test System is modified in this submission by the addition of the CardioChek ChekMate strips. CardioChek ChekMate strips are dry strips that mimic the use of the PTS PANELS test strips to check the CardioChek analyzer system optics, calibration and result handling algorithms. ChekMate strips should not be used as a substitute for liquid quality control materials. There is no change to any of the test strips in any of the above named systems.

5. Classification Names:

Glucose Test System
Panel:Clinical Chemistry 75
Product Codes:CGA
Device Classification:Class 2
Regulation:21 CFR 862.1345
Cholesterol Test System
Panel:Clinical Chemistry 75
Product Codes:CHH
Device Classification:Class I [meets limitation of exemption as per 21 CFR 862.9(c)(4)]
Regulation:21 CFR 862.1175
Triglyceride Test System
Panel:Clinical Chemistry 75
Product Codes:JGY
Device Classification:Class I [meets limitation of exemption as per 21 CFR 862.9(c)(4)]
Regulation:21 CFR 862.1705

{4}------------------------------------------------

Lipoprotein Test System
Panel:Clinical Chemistry 75
Product Codes:LBR, MRR
Device Classification:Class I [meets limitation of exemption as per 21 CFR 862.9(c)(4)]
Regulation:21 CFR 862.1475

6. Intended Use

The CardioChek Plus Test System is a small portable analyzer and test strip system intended for multiple patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices. This system is for in vitro diagnostic use only. The tests strips are for the quantitative determination of glucose, total cholesterol, HDL (high density lipoprotein) cholesterol and triglycerides in venous whole blood and capillary whole blood from the fingertip. A Chol/HDL ratio and estimated values for LDL (low density lipoprotein) cholesterol and non-HDL cholesterol are calculated by the CardioChek Plus analyzer.

  • Cholesterol measurements are used in the diagnosis and treatment of ● disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.
  • . HDL (lipoprotein) measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
  • . Triglycerides measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism or various endocrine disorders.
  • Glucose measurements are used in the diagnosis and treatment of . carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

7. Reason for 510(k):

Device Modification

8. Predicate Device Information

Device Trade Name: CardioChek Plus Test System

Device Company: Polymer Technology Systems, Inc. (PTS)

K140068 510(k) number:

Comparison to Predicate:

The modified system is identical to the unmodified system with the exception of the addition of the CardioChek ChekMate strips.

{5}------------------------------------------------

9. Performance Characteristics

The strips for each system are identical to the previously cleared strips. The CardioChek Plus Test System is unmodified compared to the predicate device with the exception of the addition of the CardioChek ChekMate strips.

CardioChek ChekMate Strips Precision:

An Intermediate Precision evaluation was performed over a 20 day period as described in CLSI EP5-A2 Evaluation of Precision Performance of Quantitative Measurement Methods. Three lots of ChekMate strips were evaluated on the CardioChek Plus using two (2) analyzers for each lot. Within run, between run, between day and total % CV for all results were less than 3.0%.

CardioChek ChekMate Strips Value Assignment:

Previously reported in K142302

CardioChek ChekMate Strips Stability:

Previously reported in K142302

Real-time Stability:

The study supports an 18 month stability claim.

Re-use Stability:

The study supports the claimed re-use stability of 500 uses for the ChekMate strips.

10. Conclusion:

The performance characteristics above demonstrate that the modification of the above named test system by the addition of the CardioChek ChekMate strips is substantially equivalent to the unmodified system.

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.