K Number
K151379
Date Cleared
2015-08-13

(83 days)

Product Code
Regulation Number
888.3030
Panel
OR
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Conventus PRSTM System is indicated for treatment of proximal radial fractures when internal fixation is desired, and fracture fragments are not too numerous and/or too small to be stabilized with the use of the device.

Device Description

The Conventus PRSTM is an intramedullary device intended for proximal radius fractures. The PRSTM System is a self-expanding implant which is deployed into the medullary canal and provides a scaffold to which bone fragments are attached using fragment screws. The implant is made from titanium alloy (Ti-6Al-4V ELI) and Nitinol.

AI/ML Overview

This document is a 510(k) premarket notification decision letter from the FDA to Conventus Orthopaedics, Inc. It concerns the Conventus PRSTM System, an intramedullary device for proximal radial fractures. The letter confirms substantial equivalence to predicate devices based on pre-clinical testing mentioned in the 510(k) Summary.

Here's an analysis of the provided information, focusing on acceptance criteria and study details:

1. A table of acceptance criteria and the reported device performance

The provided document does not explicitly state specific acceptance criteria (e.g., pass/fail thresholds) for the pre-clinical tests. It lists the types of tests performed and concludes that "The results demonstrate that the PRSTM System is substantially equivalent to the legally marketed predicate devices." This implies that the device performance met the criteria necessary to demonstrate equivalency to the predicate devices, whatever those inherent criteria were for each test.

Acceptance Criteria Type (Implicit)Reported Device Performance (Summary)
Static and cyclic axial/bend testingMet performance equivalent to predicate devices.
Static and cyclic torsional testingMet performance equivalent to predicate devices.
Screw pullout testingMet performance equivalent to predicate devices.
Corrosion testingMet performance equivalent to predicate devices.
Wear testingMet performance equivalent to predicate devices.
Nickel ion release testingMet performance equivalent to predicate devices.
Surface analysis testingMet performance equivalent to predicate devices.
Nitinol phase compositionMet performance equivalent to predicate devices.
Nitinol transition temperature (Af)Met performance equivalent to predicate devices.
Biocompatibility testingMet performance equivalent to predicate devices.
Animal TestingMet performance equivalent to predicate devices.
MRI TestingMet performance equivalent to predicate devices.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

The document does not provide specific sample sizes for any of the listed pre-clinical tests. It also does not specify the data provenance in terms of country of origin or whether the studies were retrospective or prospective. Given these are pre-clinical (laboratory and animal) tests, the concept of retrospective/prospective human data provenance does not apply.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

This section is not applicable to the type of pre-clinical studies described. These studies involve material science, mechanical engineering, and animal models, not human diagnostic interpretation requiring expert consensus on ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This section is not applicable as the studies are physical/mechanical/biological tests, not diagnostic interpretations requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

An MRMC comparative effectiveness study was not conducted. This is a pre-clinical evaluation of a surgical implant, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This refers to AI algorithm performance and is not applicable to the pre-clinical evaluation of a medical device implant.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

For these pre-clinical studies, "ground truth" refers to the established scientific and engineering principles, material properties, and biological responses measured by validated test methods. For example:

  • Mechanical tests: Ground truth is derived from engineering standards, material specifications, and the established performance characteristics of the predicate devices.
  • Corrosion, wear, nickel ion release, surface analysis, Nitinol properties: Ground truth is based on established material science standards and acceptable ranges for medical implant materials.
  • Biocompatibility: Ground truth is established by ISO standards and recognized biological response criteria (e.g., cytotoxicity, sensitization, irritation).
  • Animal Testing: Ground truth involves histological analysis, imaging, and functional assessment to determine biological response and healing in comparison to expected outcomes or predicate device performance.

8. The sample size for the training set

This refers to AI model training and is not applicable to the pre-clinical evaluation of a medical device implant.

9. How the ground truth for the training set was established

This refers to AI model training and is not applicable to the pre-clinical evaluation of a medical device implant.

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.