K Number
K151378
Date Cleared
2015-11-24

(186 days)

Product Code
Regulation Number
866.6010
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Lumipulse G HE4 Immunoreaction Cartridges:
For in vitro diagnostic use.
Lumipulse G HE4 is a Chemiluminescent Enzyme Immunoassay (CLEIA) for the quantitative determination of HE4 in human serum and plasma (lithium heparin or dipotassium EDTA) on the Lumipulse G System.
The assay is to be used as an aid in monitoring recurrence or progressive disease in patients with epithelial ovarian cancer. Serial testing for patient HE4 values should be used in conjunction with other clinical methods used for monitoring epithelial ovarian cancer.

Lumipulse G HE4 Calibrators:
Lumipulse G HE4 Calibrators are for use in the calibration of the Lumipulse G System for the quantitative measurement of HE4 in human serum and plasma (lithium heparin or dipotassium EDTA).

Device Description

The Lumipulse G HE4 is an assay system, including a set of immunoassay reagents, for the quantitative measurement of HE4 in specimens based on CLEIA technology by a two-step sandwich immunoassay method on the LUMIPULSE G System.

AI/ML Overview

The provided document describes the analytical and clinical performance of the Lumipulse G HE4 Immunoassay for monitoring recurrence or progressive disease in patients with epithelial ovarian cancer.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for this device are primarily demonstrated through its analytical performance, comparison to a predicate device, and clinical concordance in monitoring disease progression. While explicit "acceptance criteria" are not listed in a separate section, they are implicitly defined by the satisfactory results presented in the analytical and clinical studies.

CategoryAcceptance Criteria (Implicit)Reported Device Performance
PrecisionTotal (within-laboratory) CV ≤10% (as stated by the manufacturer for Lumipulse G HE4). Excellent inter-laboratory and lot-to-lot precision.Within-laboratory: Total %CV ≤3.5% (observed).Serum 1: 1.9%, Serum 2: 1.9%, Serum 4: 1.6%, Serum 5: 2.3%, Serum 6: 2.1%, Serum 7: 3.5%, Serum 8: 1.4%, Serum 9: 2.2%, Control 1: 2.4%, Control 2: 1.8%.External Laboratory (Inter-site) Precision: Total %CV ≤6.1% (observed).Serum 1: 3.8%, Serum 2: 3.7%, Serum 4: 4.2%, Serum 5: 5.1%, Serum 6: 5.7%, Serum 7: 6.1%, Serum 8: 5.4%, Serum 9: 5.2%, Control 1: 5.8%, Control 2: 4.0%.Lot-to-Lot Precision: Total %CV ≤3.2% (observed).Serum 1: 2.2%, Serum 2: 2.3%, Serum 4: 2.3%, Serum 5: 3.2%, Serum 6: 2.2%, Serum 7: 2.7%, Serum 8: 2.2%, Serum 9: 2.5%, Control 1: 2.3%, Control 2: 1.6%.
Linearity/RangeDemonstrated linearity across the assay range (20.0 to 1500.0 pM). Recovery 100 ± 10%.Linearity: Found in the range of 20.0 to 1500.0 pM for both serum and plasma. R-squared = 1.0000 for linear regression.Recovery: Ranged from 91% to 107% (within 100 ± 10%).
Detection LimitsLoB, LoD, and LoQ ≤ 20.0 pM (as stated by the manufacturer).LoB: 0.1 pM.LoD: 3.5 pM.LoQ: 3.5 pM (since percent total error estimate is ≤ 30%, LoQ equals LoD).
Analytical Specificity/InterferenceAverage interference ≤ 10% for potentially interfering compounds.Demonstrated average interference ≤ 10% for all tested endogenous (Bilirubin, Triglycerides, Hemoglobin, Human Serum Albumin, Immunoglobulin G, Biotin, HAMA, RF) and therapeutic drugs (Bevacizumab, Carboplatin, Cisplatin, Clotrimazole, Cyclophosphamide, Dexamethasone, Docetaxel, Doxil®, Doxorubicin, Erlotinib, Etoposide, 5-Fluorouracil, Gemcitabine, Leucovorin, Magestrol Acetate, Melphalan, Methotrexate, Olaparib, Paclitaxel, Rituximab, Tamoxifen, Topotecan, Trastuzumab). No high dose hook effect observed up to 300,000 pM.
Method ComparisonSubstantial equivalence to predicate device (Fujirebio Diagnostics, Inc. HE4 EIA) indicated by strong correlation and acceptable bias.Correlation Coefficient (r): 0.9891 (143 samples, range 35.2-969.5 pM) and 0.9917 (168 samples, range 35.2-4602.0 pM).Slope: 1.0349 (95% CI: 1.0074-1.0624) and 1.0289 (95% CI: 1.0045-1.0533).Average Bias: 8.8 pM and 13.8 pM.
Clinical ConcordanceReasonable sensitivity and specificity for monitoring disease progression/recurrence in epithelial ovarian cancer patients alongside clinical methods.Sensitivity: 49.2% (95% CI: 27.8-70.6%).Specificity: 79.6% (95% CI: 65.9-93.2%).Total Concordance: 73.9% (95% CI: 62.6-85.3%).Positive Predictive Value (PPV): 35.3% (95% CI: 19.9-50.7%).Negative Predictive Value (NPV): 87.4% (95% CI: 72.3-102.4%).

2. Sample Sizes Used for the Test Set and Data Provenance

  • Test Set (Clinical Study):

    • Sample Size: 72 patients with epithelial ovarian cancer, providing a total of 330 pairs of observations (serial samples). The average number of observations per patient was 5.6.
    • Data Provenance: Not explicitly stated regarding country of origin or specific institutions. The nature of the study (monitoring recurrence/progression) implies a retrospective collection of serial samples from patients already diagnosed, or a prospective collection over time. Given the context of a 510(k) submission, it is common for such clinical performance studies to be retrospective using existing clinical samples.
  • Test Set (Analytical Studies - Precision, Linearity, Detection Limits, Interference):

    • Precision (within-lab): 8 human serum-based samples (specimen pools) and 2 commercially available serum-based controls. Each assayed in replicates of two, twice a day for 20 days (n=80 for each sample).

    • Precision (external lab): Human serum-based samples (specimen pools) and 2 commercially available controls. Assayed in replicates of two, twice a day for 10 days at 3 external sites (n=40 for each sample per site, total n=120 per sample across sites).

    • Precision (lot-to-lot): Human serum-based samples (specimen pools) and 2 commercially available controls. Assayed in replicates of two, twice a day for 10 days with 3 lots of cartridges and calibrators (n=40 for each sample per lot, total n=120 per sample across lots).

    • Linearity/Recovery: Human serum and dipotassium EDTA plasma samples (pools). The document doesn't specify the exact number of unique "serum/plasma samples" but rather "specimen pools" and subsequent dilutions/additions. Tables show 3 serum and 3 plasma samples assessed for recovery, each with 6 target HE4 levels.

    • Detection Limits (LoB, LoD, LoQ): Seven low-level specimens were tested over 3 days using three LUMIPULSE G Systems and three Lumipulse G HE4 lots, yielding 180 determinations for each panel.

    • Analytical Specificity (Interference): Human serum specimens supplemented with potentially interfering compounds. The exact number of unique "specimens" is not given, but a long list of compounds tested.

    • Method Comparison: 143 samples for the common measuring range, and 168 samples including those requiring dilution.

    • Matrix Comparison: Not specified, but involved comparing SST, K2EDTA, and Lithium Heparin tube types against Red top serum.

    • Data Provenance (Analytical): Not explicitly stated, but typically these are laboratory-generated studies using human donor samples, commercial controls, or spiked samples. The CLSI guidelines followed (EP5-A3, EP6-A, EP7-A2, EP9-A3, EP17-A2) are standard for in vitro diagnostic device validation, implying controlled laboratory conditions.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • For the clinical study (monitoring disease status): The ground truth was defined by "changes in disease status," which is a clinical outcome. The document states: "A positive change in HE4 was defined as an increase in the value that was at least 18% greater than the previous value of the test. This level of change takes into account the variability of the assay. Forty-nine percent (49%) or 30/61 of the patient samples with a positive change correlated with the disease progression while eighty percent (80%) or 214/269 of the patient serial samples with no significant change in HE4 value correlated with no progression."
    • This implies that the "disease progression" ground truth was established by independent clinical assessments, likely by treating physicians or oncology teams using standard clinical methods (imaging, physician assessment, other biomarkers, symptoms).
    • The document does not specify the number of experts or their specific qualifications (e.g., "Radiologist with 10 years of experience"). This is typical for IVD submissions where the ground truth is derived from standard clinical practice rather than a specific expert consensus panel for the study.

4. Adjudication Method for the Test Set

  • For the clinical study: Not explicitly stated as a formal adjudication process involving multiple independent reviewers in the sense of medical imaging reader studies. The ground truth for "disease progression" appears to be derived from the patient's overall clinical status as assessed by their primary care or oncology team. The "correlation" mentions comparing HE4 changes with observed disease progression/no progression. There's no indication of multiple independent adjudicators for the disease status itself within the context of this study.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, an MRMC comparative effectiveness study was not done. This is an in vitro diagnostic (IVD) device, specifically an immunoassay for measuring a biomarker (HE4). MRMC studies are typically performed for imaging devices (e.g., AI for radiology) where human readers (e.g., radiologists) interpret images with and without AI assistance to measure improvement in diagnostic performance. This submission focuses on the analytical and clinical performance of a quantitative laboratory test.

6. Standalone Performance

  • Yes, standalone performance was done. The entire analytical performance section (Precision, Linearity, Detection Limits, Analytical Specificity) describes the performance of the Lumipulse G HE4 assay as a standalone system (Lumipulse G System + HE4 kit).
  • The clinical study also reports the performance characteristics (sensitivity, specificity, concordance) of the HE4 assay itself in correlating with disease progression, essentially its standalone diagnostic aid performance. The indication for use clearly states it is an "aid in monitoring recurrence or progressive disease," implying its utility as a standalone test in this context, albeit "in conjunction with other clinical methods."

7. Type of Ground Truth Used

  • Analytical Studies:
    • Known concentrations: For linearity, recovery, and detection limits, the ground truth is established by preparing samples with known concentrations of HE4.
    • Reference materials/specimen pools: For precision and interference, ground truth relies on the characteristics of the prepared specimen pools or commercial controls.
  • Clinical Study:
    • Clinical Outcomes/Disease Status: The ground truth for the clinical study was the "change in disease status" (progression or no progression). This is derived from an unstated combination of "other clinical methods used for monitoring epithelial ovarian cancer" as applied by the treating clinicians. This likely includes imaging, physical exams, symptoms, and potentially other biomarker trends, but the document does not specify.

8. Sample Size for the Training Set

  • The document describes a 510(k) submission for an in vitro diagnostic assay, not an AI/machine learning algorithm that requires a "training set" in the computational sense.
  • For an immunoassay, the equivalent of "training" would be the initial development and optimization phase, which uses various characterized samples (e.g., patient samples, spiked solutions, controls) to establish the assay's operational parameters, calibrate it, and define its intended performance characteristics before formal validation studies. The exact sample sizes for this development phase are typically not detailed in a 510(k) summary. The studies presented are validation studies (test sets) for regulatory approval.

9. How the Ground Truth for the Training Set Was Established

  • As above, the concept of a computational "training set" ground truth doesn't directly apply here. For the development and initial calibration of the immunoassay:
    • Ground truth for calibrators is established via a hierarchical process traceable to an in-house reference preparation, which correlates to the predicate device (HE4 EIA). This is a metrological traceability chain, not an expert panel.
    • Ground truth for developing the assay's measurement range, precision, and specificity would involve using characterized human samples (e.g., samples from healthy individuals, individuals with known HE4 levels, and individuals with confounding conditions) and spiked samples (where known amounts of HE4 are added to a matrix). The "ground truth" (i.e., the target value) for such samples is determined by precise laboratory methods, often against a reference standard or by the known amount of analyte added.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a human figure with outstretched arms.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 8, 2016

Fujirebio Diagnostics, Inc. c/o Stacey Dolan, RAC-US Manager, Regulatory Affairs 201 Great Valley Parkway Malvern, PA 19355

Re: K151378

Trade/Device Name: Lumipulse G HE4 Immunoreaction Cartridges, Lumipulse G HE4 Calibrators Regulation Number: 21 CFR 866.6010 Regulation Name: Tumor-associated antigen immunological test system Regulatory Class: II Product Code: OIU, JIT Dated: October 14, 2015 Received: October 15, 2015

Dear Ms. Dolan:

This letter corrects our substantially equivalent letter of November 23, 2015.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kelly Oliner -S

FOR

Leonthena R. Carrington, MS, MBA, MT(ASCP) Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K151378

Device Name

Lumipulse G HE4 Immunoreaction Cartridges Lumipulse G HE4 Calibrators

Indications for Use (Describe) Lumipulse G HE4 Immunoreaction Cartridges:

For in vitro diagnostic use.

Lumipulse G HE4 is a Chemiluminescent Enzyme Immunoassay (CLEIA) for the quantitative determination of HE4 in human serum and plasma (lithium heparin or dipotassium EDTA) on the Lumipulse G System.

The assay is to be used as an aid in monitoring recurrence or progressive disease in patients with epithelial ovarian cancer. Serial testing for patient HE4 values should be used in conjunction with other clinical methods used for monitoring epithelial ovarian cancer.

Lumipulse G HE4 Calibrators:

Lumipulse G HE4 Calibrators are for use in the calibration of the Lumipulse G System for the quantitative measurement of HE4 in human serum and plasma (lithium heparin or dipotassium EDTA).

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a blue and white abstract symbol on the left, followed by the company name in blue text. The word "FUJIREBIO" is in a larger, bolder font than "Diagnostics, Inc.", which is placed below it. The overall design is clean and professional.

Section 5 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

A. 510(k) Number:

K151378

B. Purpose for Submission:

New device

C. Measurand:

HE4

D. Type of Test:

Quantitative assay, automated chemiluminescent enzyme immunoassay (CLEIA)

E. Applicant:

Address:Fujirebio Diagnostics, Inc.201 Great Valley ParkwayMalvern, PA 19355
Contact person:Stacey Dolan, RAC (US)(610) 240-3917dolans@fdi.com

Summary preparation date: May 21, 2015

F. Proprietary and Established Names:

Lumipulse G HE4 Immunoreaction Cartridges Lumipulse G HE4 Calibrators

G. Regulatory Information:

    1. Regulation section: 21 CFR § 866.6010, Tumor-associated antigen immunological test system 21 CFR § 862.1150, Calibrator
    1. Classification: Class II
    1. Product code: OIU, Test, Epithelial Ovarian Tumor-Associated Antigen (HE4) JIT, Calibrator, Secondary

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized blue and green graphic on the left, followed by the company name in blue text. The word "FUJIREBIO" is in a larger, bolder font than "Diagnostics, Inc.", which is placed below it. The overall design is clean and professional.

    1. Panel:
      82, Immunology (Assay) 75, Chemistry (Calibrators)

H. Intended Use:

    1. Intended use(s): See indications for use below.
    1. Indication(s) for use:

Lumipulse G HE4 Immunoreaction Cartridges For in vitro diagnostic use.

The Lumipulse G HE4 is a Chemiluminescent Enzyme Immunoassay (CLEIA) for the quantitative determination of HE4 in human serum and plasma (lithium heparin or dipotassium EDTA) on the LUMIPULSE G System.

The assay is to be used as an aid in monitoring recurrence or progressive disease in patients with epithelial ovarian cancer. Serial testing for patient HE4 assay values should be used in conjunction with other clinical methods used for monitoring epithelial ovarian cancer.

Lumipulse G HE4 Calibrators

The Lumipulse G HE4 Calibrators are for use in the calibration of the LUMIPULSE G System for the quantitative measurement of HE4 in human serum and plasma (lithium heparin or dipotassium EDTA).

    1. Special conditions for use statement(s): Prescription use only
    1. Special instrument requirements: LUMIPULSE G1200 System

-Device Description:

The Lumipulse G HE4 is an assay system, including a set of immunoassay reagents, for the quantitative measurement of HE4 in specimens based on CLEIA technology by a two-step sandwich immunoassay method on the LUMIPULSE G System.

Lumipulse G HE4 Immunoreaction Cartridges |REF| 234174

The Lumipulse G HE4 Immunoreaction Cartridges consists of 3 × 14 tests. Each kit contains the following:

Antibody-Coated Particle Solution (Liquid when used, 250 µL/Immunoreaction Cartridge)

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image contains the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized graphic on the left, featuring blue and white shapes, next to the company name in blue, bold letters. Below the company name, the words "Diagnostics, Inc." are written in a smaller, thinner font.

Contains 150 ug/mL anti-HE4 monoclonal antibody (mouse)-coated particles, protein stabilizers (bovine and mouse) and chemical stabilizers in 0.15 M sodium chloride/Tris buffer. This solution contains gelatin and turns into gel at 15 °C or lower. Preservative: Proclin 300.

Enzyme-Labeled Antibody Solution

(Liquid, 350 µL/Immunoreaction Cartridge)

Contains 0.25 µg/mL alkaline phosphatase (ALP: calf) labeled anti-HE4 monoclonal antibody (mouse), protein stabilizers (bovine, calf and mouse) and chemical stabilizers in 0.1 M sodium chloride/MES buffer. Preservative: Proclin 300.

LUMIPULSE G HE4 Calibrators CAL 234181

Each calibrator kit contains one bottle each of Calibrators 1 and 2. The calibrator kit is packaged separately.

CAL Liquid, 1 × 2 Concentrations

CAL 1 0 pM HE4 calibrator (1 × 1.5 mL)

CAL 2 1500 pM HE4 calibrator (1 × 1.5 mL)

Contains sodium chloride and potassium chloride in phosphate buffer with protein stabilizer (bovine). Preservative: Proclin 950.

J. Substantial Equivalence Information:

    1. Predicate device name(s): Fujirebio Diagnostics, Inc. HE4 EIA
    1. Predicate 510(k) number(s): K072939

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image contains the logo for Fujirebio Diagnostics, Inc. The logo features a stylized graphic element on the left, consisting of blue and green shapes that resemble a water droplet or a stylized leaf. To the right of the graphic is the company name, "FUJIREBIO," in a bold, sans-serif blue font. Below the company name is the text "Diagnostics, Inc." in a smaller, sans-serif font, also in blue.

3. Comparison with predicate:

Similarities
Lumipulse G HE4 Assay(Proposed Device)Fujirebio Diagnostics, Inc. HE4EIA Assay(Predicate Device) K072939
Device TypeIn vitro diagnosticIn vitro diagnostic
ClassificationClass IIClass II
CFR section866.6010866.6010
Product CodeOIUOIU
Product UsageClinical and Hospital laboratoriesFor in vitro diagnostic use.Clinical and Hospital laboratoriesFor in vitro diagnostic use.
Intended UseThe Lumipulse G HE4 is aChemiluminescent EnzymeImmunoassay (CLEIA) for thequantitative determination of HE4 inhuman serum and plasma (lithiumheparin or dipotassium EDTA) on theLUMIPULSE G System.The assay is to be used as an aid inmonitoring recurrence or progressivedisease in patients with epithelialovarian cancer. Serial testing forpatient HE4 assay values should beused in conjunction with other clinicalmethods used for monitoring epithelialovarian cancer.The HE4 EIA is an enzymeimmunometric assay for thequantitative determination of HE4 inhuman serum.The assay is to be used as an aid inmonitoring recurrence or progressivedisease in patients with epithelialovarian cancer. Serial testing forpatient HE4 assay values should beused in conjunction with other clinicalmethods used for monitoring epithelialovarian cancer
Specimen CollectionMethodRoutine Phlebotomy TechniquesRoutine Phlebotomy Techniques
Calibrator AntigenIg-HE4Ig-HE4
AnalyteHuman HE4Human HE4

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo features a stylized graphic on the left, consisting of blue and green shapes. To the right of the graphic, the word "FUJIREBIO" is written in large, blue, sans-serif font. Below "FUJIREBIO", the words "Diagnostics, Inc." are written in a smaller, black, sans-serif font.

Differences
Lumipulse G HE4 Assay(Proposed Device)Fujirebio Diagnostics, Inc. HE4EIA Assay(Predicate Device) K072939
Instrument SystemLUMIPULSE G1200 SystemNone
Principle of OperationChemiluminscent EnzymeImmunoassay (CLEIA)Enzyme Immunometric Assay(EIA)
Type of SpecimenHuman Serum or Plasma (, lithiumheparin, or dipotassium EDTA)Human Serum
Assay Range20 - 1500 pM15 - 900 pM
Interpretation of ResultsCalibrator CurveStandard Curve
Calibrators2 Levels (0 and 1500 pM)Ready to Use-Liquid-Supplied as separate kit5 Levels (Lot specific)-Lyophilized
Antibodies12A2 and 3D8 (mousemonoclonal)2H5 and 3D8 (mouse monoclonal)

K. Standard/Guidance Document Referenced (if applicable):

  • . ISO 17511:2003 Measurement of Quantities in Biological Samples - Metrological Traceability of Values Assigned to Calibrator and Control Materials
  • . CLSI EP5-A3 - Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - Third Edition
  • CLSI EP7-A2 Interference Testing in Clinical Chemistry; Approved Guideline-Second . Edition
  • . CLSI C28-A3c - Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory: Approved Guideline-Third Edition
  • . CLSI EP17-A2 - Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline Second Edition
  • . CLSI EP6-A - Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline
  • . CLSI EP9-A3 - Measurement Procedure Comparison and Bias Estimation Using Patient Samples; approved Guideline - Third Edition
  • FDA Guidance Document for the Submission of Tumor Associated Antigen Premarket ● Notifications, [510(k)], to FDA
  • . Guidance on Informed Consent for In Vitro Diagnostic Device Studies Using Leftover Human Specimens that are Not Individually Identifiable - Guidance for Sponsors, Institutional Review Boards, Clinical Investigators and FDA Staff

L. Test Principle:

Lumipulse G HE4 is an assay system, including a set of immunoassay reagents, for the quantitative measurement of HE4 in specimens based on CLEA technology by a two-step sandwich immunoassay method on the LUMIPULSE G1200 System. HE4 in specimens

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image is a logo for Fujirebio Diagnostics, Inc. The logo features a stylized blue and white graphic on the left, resembling a water droplet or a stylized letter. To the right of the graphic, the word "FUJIREBIO" is written in a bold, blue sans-serif font. Below "FUJIREBIO", the words "Diagnostics, Inc." are written in a smaller, dark gray sans-serif font.

specifically binds to anti-HE4 monoclonal antibody (mouse) on the particles, and antigenantibody immunocomplexes are formed. The particles are washed and rinsed to remove unbound materials. Alkaline phosphatase (ALP: calf)-labeled anti-HE4 monoclonal antibody (mouse) specifically binds to HE4 of the immunocomplexes on the particles, and additional immunocomplexes are formed. The particles are washed and rinsed to remove unbound materials. Substrate Solution is added and mixed with the particles. AMPPD contained in the Substrate Solution is dephosphorylated by the catalysis of ALP indirectly conjugated to particles. Luminescence (at a maximum wavelength of 477 nm) is generated by the cleavage reaction of dephosphorylated AMPPD. The luminescent signal reflects the amount of HE4.

M. Performance Characteristics (if/when applicable):

Data were generated using the LUMIPULSE G1200 System.

    1. Analytical performance:
    • a. Precision/Reproducibility:

Lumipulse G HE4 is ≤10% total (within-laboratory) CV (coefficient of variation). Lumipulse G HE4 demonstrated precision ≤3.5% (total %CV) in a study run according to the Clinical and Laboratory Standards Institute (CLSI) guideline EP5-A3. Eight human serum-based samples (specimen pools), and two commercially available serum-based controls were assayed in replicates of two at two separate times of the day for 20 for each sample) using one LUMIPULSE G1200 System. Data from this study are presented below.

SampleMeanConc.(pM)n=80Within-runSD(pM)Within-run%CVTotalSD(pM)Total%CV
Serum 158.90.701.21.151.9
Serum 277.40.781.01.451.9
Serum 4119.01.581.31.871.6
Serum 5310.73.181.07.122.3
Serum 61125.114.031.224.192.1
Serum 728.60.351.20.993.5
Serum 8627.87.061.18.491.4
Serum 91329.921.891.629.882.2
Control 195.11.992.12.272.4
Control 2751.810.271.413.881.8

Lumipulse G HE4 demonstrated precision ≤6.1% total %CV when tested at 3 external laboratory sites. The human serum-based samples (specimen pools) and two commercially available serum-based controls were assayed in replicates of two separate times of the day at each of the sites for 10 days (n=40 for each sample) using one LUMIPUSE G1200 System. Data from this study are presented below.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized graphic on the left and the company name on the right. The graphic is composed of blue and green shapes, with a white teardrop shape in the center. The text "FUJIREBIO" is in blue, and "Diagnostics, Inc." is in black.

SampleMeanConc.(pM)n=120Within-runSD (pM)Within-run%CVTotalSD(pM)Total%CV
Serum 157.10.701.22.143.8
Serum 275.40.891.22.783.7
Serum 4115.91.471.34.854.2
Serum 5303.02.850.915.445.1
Serum 61110.214.991.462.985.7
Serum 727.50.351.31.686.1
Serum 8609.38.171.332.895.4
Serum 91323.225.922.068.745.2
Control 195.41.902.05.575.8
Control 2733.412.271.729.164.0

Lumipulse G HE4 demonstrated precision ≤3.2% total %CV when tested using 3 lots of Lumipulse G HE4 immunoreaction cartridges and calibrators. The human serum-based samples (specimen pools) and two commercially available serum-based controls were assayed in replicates of two at two separate times of the day for each of the lots for 10 days (n=40 for each sample) using one LUMIPULSE G1200 System. Data from this study are presented below.

SampleMeanConc.(pM)n=120Within-runSD (pM)Within-run%CVTotalSD (pM)Total%CV
Serum 157.50.561.01.242.2
Serum 275.30.761.01.752.3
Serum 4116.11.421.22.712.3
Serum 5302.73.811.39.763.2
Serum 61119.315.361.424.912.2
Serum 727.60.351.20.752.7
Serum 8617.59.661.613.832.2
Serum 91334.025.331.933.602.5
Control 194.91.661.82.202.3
Control 2746.58.351.111.821.6

b. Linearity/assay reportable range:

Lumipulse G HE4 demonstrated linearity in a study consistent with the quidelines in the CLSI Protocol EP6-A. One human serum specimen pool and one dipotassium EDTA plasma specimen pool with high HE4 levels were diluted with one human serum specimen pool and one dipotassium EDTA plasma specimen pool with low HE4 levels throughout the range of the assay. The linearity was found in the range of 20.0 to 1500.0 pM. Lumipulse G HE4 correlated with expected concentrations according to the linear regression formulas:

Serum:

Plasma: y= 1.0104 (x) + 12.4869; R-squared: 1.0000

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image is a logo for Fujirebio Diagnostics, Inc. The logo consists of a blue and white abstract shape on the left, followed by the company name in blue, and the words "Diagnostics, Inc." in black below the company name. The abstract shape is made up of a blue curved shape, a white teardrop shape, and green vertical lines.

Lumipulse G HE4 recovery is 100 ± 10%, ranging from 91% to 107%. A study was performed where known concentrations of HE4 antigen were added to human serum and dipotassium EDTA plasma samples with low endogenous HE4 levels. The concentration of HE4 was determined using the Lumipulse G HE4 and the resulting percent recovery was calculated.

SampleTargetHE4Levels(pM)MeasuredConc.(pM)(n=3)ExpectedConc.(pM)%Recovery
Serum180132.8125.2106
150188.2186.7101
450484.4474.2102
750768.0761.0101
10501056.51064.499
13501350.41390.597
Serum280148.3140.8105
150196.0191.5102
450488.7479.0102
750767.2765.8100
10501046.91069.298
13501316.91395.394
Serum380116.8109.3107
150202.4197.4103
450491.7484.9101
750763.8771.799
10501081.11075.1101
13501324.11401.294
Plasma180127.5122.2104
150242.4240.4101
450518.1527.998
750796.3814.798
10501069.31118.196
13501366.41444.295
Plasma280138.5134.5103
150206.5203.4102
450492.7490.9100
750780.6777.7100
10501035.81081.196
13501274.71407.291
Plasma380137.9130.9105
150204.8201.1102
450494.0488.6101
750784.3775.4101
10501057.11078.898
13501344.91404.996

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo features a stylized blue and white graphic on the left, resembling a water droplet or a stylized letter 'F'. To the right of the graphic is the company name, "FUJIREBIO" in bold blue letters, with "Diagnostics, Inc." written in a smaller, black font below it. The logo is clean and professional, conveying a sense of innovation and reliability.

High dose hook is a phenomenon whereby very high level specimens may read within the dynamic range of the assay. For Lumipulse G HE4, no high dose hook effect was observed for samples containing up to 300,000 pM of HE4 antigen. However values above 30,000 pM should be interpreted with caution because reading may be inaccurate above this concentration.

Traceability, Stability, Expected values (controls, calibrators, or methods): ﯼ

The Lumipulse G HE4 Calibrators are for use in the calibration of the LUMIPULSE G System when used for the quantitative measurement of HE4 in human serum and plasma (lithium heparin or dipotassium EDTA). 2 Bottles (1.5 mL each) are supplied for the LUMIPULSE G HE4 Calibrators. Preservatives: Proclin 950. The calibrators are at the following concentrations:

CalibratorConcentration (pM)
CAL 10
CAL 21500

There is currently no known internationally recognized consensus reference method or reference material for standardization. Lumipulse G HE4 values are expressed as pM (or pmol/L). This value is related to a Fujirebio Diaqnostics maintained reference preparation. Calibration of the Lumipulse G HE4 is traceable to in-house reference calibrators, whose values have been assigned to correlate to Fujirebio Diagnostics' HE4 EIA.

The real time stability (including open use) has been demonstrated for 10 months. The instrument cartridge and calibrators were found to be stable under simulated transport. conditions. Shelf life assignment for the Lumipulse G HE4 Immunoreaction Cartridges and Calibrators will be 8 months at 2-10°C. .

d. Detection limit:

The Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantitation (LoQ) of Lumipulse G HE4 is ≤ 20.0 pM.

1) LoB and LoD

The LoB for Lumipulse G HE4 was 0.1 pM.

The LoD for Lumipulse G HE4 was 3.5 pM, determined consistent with the quidelines in the CLSI Protocol EP17-A2. Seven low-level specimens were tested over 3 days using three LUMIPULSE G Systems and three Lumipulse G HE4 lots giving 180 determinations for each panel.

2) LoQ

The LOQ for Lumipulse G HE4 was 3.5 pM, determined consistent with the guidelines in the CLSI quideline EP17-A2. Since the percent total error estimate is ≤ 30%, LoQ equals LoD.

{12}------------------------------------------------

Image /page/12/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a blue and white abstract graphic on the left, followed by the company name in blue and black text on the right. The word "FUJIREBIO" is in a bold, sans-serif font, and the words "Diagnostics, Inc." are in a smaller, lighter font below.

e. Analytical specificity:

The Lumipulse G HE4 demonstrated an average interference of ≤ 10% (for each compound) in a study consistent with the guidelines in the CLSI guideline EP7-A2. Human serum specimens were supplemented with potentially interfering compounds. TThe following compounds were tested and found not to interfere with the test.

EndogenousInterferencesTestConcentration
Free Bilirubin (unconjugated)60 mg/dL
Conjugated Bilirubin60 mg/dL
Triglycerides (Intralipid 20% Emulsion)3000 mg/dL
Hemoglobin500 mg/dL
Human Serum Albumin12 g/dL
Immunoglobulin G (IgG)5 g/dL
Biotin19.8 mg/dL
Human Anti-Mouse Antibodies (HAMA)1,000 ng/mL
Rheumatoid Factor (RF)1,000 IU/mL
Therapeutic drugInterferencesTestConcentration
Bevacizumab750 µg/mL
Carboplatin600 µg/mL
Cisplatin180 µg/mL
Clotrimazole0.3 µg/mL
Cyclophosphamide800 µg/mL
Dexamethansone20 µg/mL
Docetaxel10.5 mg/dL
Doxil® (Pegylated Liposomal Doxorubicin)5.2 mg/dL
Doxorubicin120 µg/mL
Erlotinib150 ug/mL
Etoposide10 ug/mL
5-Fluorouracil900 ug/mL
Gemcitabine100 ug/mL
Leucovorin750 µg/mL
Magestrol Acetate10 ug/mL
Melphalan15 µg/mL
Methotrexate450 µg/mL
Olaparib28.1 mg/dL
Paclitaxel3.5 ng/mL
Rituximab750 µg/mL
Tamoxifen60 ug/mL
Topotecan0.155 mg/dL
Trastuzumab600 ug/ml

f. Assay cut-off:

{13}------------------------------------------------

Image /page/13/Picture/1 description: The image contains the logo for Fujirebio Diagnostics, Inc. The logo consists of a blue and white abstract symbol on the left, followed by the company name in blue, with "Diagnostics, Inc." in a smaller font size below "FUJIREBIO". The overall design is clean and corporate.

See Clinical Cutoff in M (4) below

2. Comparison studies:

a. Method Comparison

Lumipulse G HE4 method comparison study was performed using specimens consistent with CLSI guideline EP9-A3. The weighted Deming regression method was used to compare Lumipulse G HE4 to HE4 EIA. The HE4 EIA has a different measuring range than Lumipulse G HE4, therefore the results were limited to the measuring range of both devices (143 samples). The data are summarized in the following table.

Range of samples: 35.2 - 969.5 pM (Lumipulse G HE4) 33.4 - 899.0 pM (HE4 EIA)

Lumipulse G HE4 vs. HE4 EIA
nCorrelationCoefficient(r)Intercept(95% CI)Slope(95% CI)Average Bias(pM)
1430.9891-3.23501.03498.8
-5.3628 –1.0074 –
–1.10711.0624

The date summarized in the following table include results from a study with specimens above the measurement range of both devices requiring dilution (168 samples).

Range of samples:

35.2 - 4602.0 pM (Lumipulse G HE4) 33.4 - 4067.0 pM (HE4 EIA)

Lumipulse G HE4 vs. HE4 EIA

nCorrelationCoefficient(r)Intercept(95% CI)Slope(95% CI)Average Bias(pM)
1680.9917-2.85181.028913.8
-4.8632 --0.84041.0045 -1.0533

b. Matrix Comparison

{14}------------------------------------------------

Image /page/14/Picture/1 description: The image is a logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized graphic on the left and the company name on the right. The graphic is a combination of blue and green shapes, with a white teardrop shape in the center. The company name is written in blue, with "Diagnostics, Inc." written in a smaller, black font below the main name.

The Lumipulse G HE4 matrix comparison study was performed to evaluate the difference across tube types (SST, K2EDTA, and Lithium Heparin,) versus the means of the control samples (Red top serum) analyzed per CLSI guideline EP9-A3. The slope for each tube type when compared to the control had 95% confidence intervals that lay entirely within the range 0.9 to 1.1 and the correlation coefficients were ≥ 0.9.

    1. Clinical studies:
    • a. Clinical sensitivity:

See 3(c) below

  • b. Clinical specificity:

See 3(c) below

  • c. Other clinical supportive data (when a. and b. are not applicable):

Monitoring of Disease status in Patients Diagnosed with Epithelial Ovarian Cancer

The effectiveness of Lumipulse G HE4 as an aid in monitoring of disease status in EOC patients was determined by assessing changes in HE4 levels in serial serum samples from 72 patients compared to changes in disease status. A study involving a total of 330 pairs of observations was undertaken with an average number of 5.6 observations per patient. A positive change in HE4 was defined as an increase in the value that was at least 18% greater than the previous value of the test. This level of change takes into account the variability of the assay. Forty-nine percent (49%) or 30/61 of the patient samples with a positive change correlated with the disease progression while eighty percent (80%) or 214/269 of the patient serial samples with no significant change in HE4 value correlated with no progression. The total concordance was seventy-four percent (74% or 245/330).

Performance MeasurementSELower 95%CIUpper 95%CI
Sensitivity49.210.7427.870.6
Specificity79.66.8765.993.2
Total Concordance73.95.6962.685.3
PPV35.37.7119.950.7
NPV87.47.5472.3102.4

Lumipulse G HE4 Performance Measurements (%):

SE = Standard Error

Cl = Confidence Interval

The following table presents the data in a 2 x 2 format.

{15}------------------------------------------------

Image /page/15/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized graphic on the left, featuring blue and green elements that resemble a water droplet and vertical lines. To the right of the graphic, the company name "FUJIREBIO" is displayed in bold, blue letters, with "Diagnostics, Inc." appearing in a smaller font size below it.

Change in Disease State per Sequential Pair
Change inHE4ConcentrationNoprogressionProgressionTotal
≤ 18%21431245
> 18%553085
Total26961330

4. Clinical cut-off:

HE4 is significantly elevated if it is at least 18% higher than the previous assay.

5. Expected values/Reference range:

The distribution of HE4 levels determined in healthy premenopausal and postmenopausal women is shown in the table below:

AllHealthy SubjectsPremenopausalHealthy SubjectsPostmenopausalHealthy Subjects
N240120120
Mean59.558.061.0
(SD)(68.8)(94.8)(22.5)
Median51.347.055.9
Range (min, max)29.4, 1076.629.9, 1076.629.4, 176.7
Reference Interval(2.5th percentile,97.5th percentile)32.5, 108.431.9, 87.132.5, 112.2
Proportion below90.0 pM96%98%93%
Proportion below135.0 pM99%99%98%

All Lumipulse G HE4 concentrations are presented as pM

{16}------------------------------------------------

Image /page/16/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized graphic on the left, featuring blue and green elements, possibly representing water droplets or abstract shapes. To the right of the graphic, the company name "FUJIREBIO" is displayed in bold, blue letters, with "Diagnostics, Inc." written in a smaller, black font underneath.

The distribution of HE4 levels determined in benign conditions is shown in the table below:

BenignGynecologicalDiseaseOtherBenignDiseaseCongestiveHeartFailureHypertensionPregnant
N36640404040
Mean70.7232.6185.3397.045.4
(SD)(44.0)(548.8)(99.7)(999.9)(8.3)
Median57.881.4173.9108.844.5
Range (min, max)26.3,386.842.7,2762.030.6,417.739.0,5773.031.9,71.0
Reference Interval(2.5th percentile ,97.5th percentile)35.0,202.044.2,2374.943.5,407.540.9,2874.332.2,58.3
Proportion below90.0 pmol/L82%55%18%43%100%
Proportion below135.0 pmol/L94%70%38%65%100%

*All Lumipulse G HE4 concentrations are presented as pM

{17}------------------------------------------------

Image /page/17/Picture/1 description: The image is a logo for Fujirebio Diagnostics, Inc. The logo features a stylized blue and white graphic on the left, resembling a water droplet or a stylized leaf. To the right of the graphic is the company name, "FUJIREBIO" in bold, blue letters, with "Diagnostics, Inc." underneath in a smaller font. The logo is clean and professional, conveying a sense of innovation and reliability.

The distribution of HE4 levels determined in malignancy conditions is shown in the table below:

EpithelialOvarianCancerBladderCancerBreastCancerEndometrialCancerGICancerLungCancer
N1554040404040
Mean(SD)1188.7(2447.4)188.3(230.0)173.0(398.0)145.9(119.5)103.4(51.8)157.6(127.5)
Median341.0127.595.399.591.2115.8
Range (min, max)24.6,>15000.044.9,1381.932.4,2591.050.0,646.942.2,258.941.1,658.5
Reference Interval(2.5th percentile ,97.5th percentile)48.6,7602.350.3,569.938.3,419.860.5,500.044.3,247.742.3,562.9
Proportion below90.0 pmol/L14%30%48%35%48%33%
Proportion below135.0 pmol/L25%58%78%68%83%63%

All Lumipulse G HE4 concentrations are presented as pM

It is recommended that each laboratory establish its own reference value for the population of interest for Lumipulse G HE4 on the LUMIPULSE G1200.

6. Conclusion

The results of these analytical (nonclinical) and clinical studies demonstrate that the Lumipulse G HE4 assay is substantially equivalent to the performance of the Fujirebio Diagnostics, Inc. HE4 EIA assay.

N. Proposed Labeling:

The labeling satisfies the requirements of 21 CFR Part 809.10.

§ 866.6010 Tumor-associated antigen immunological test system.

(a)
Identification. A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.(b)
Classification. Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.