(128 days)
The AZUR Peripheral Coil System is intended to reduce or blood flow in vessels of the peripheral vasculature. It is intended for use in the interventional radiologic management of arteriovenous malformations, arteriovenous fistulae, aneurysms, and other lesions of the peripheral vasculature.
The AZUR CX Coils consist of implant coil made of platinum alloy with inner hydrogel core. The coils are designed in 3D spherical structure in various loop sizes and lengths. The AZUR CX Coil System (Detachable) consists of an implantable coil attached to a delivery pusher. The coil system is delivered to the treatment site through the microcatheter. The proximal end of the delivery pusher is inserted to the AZUR detachment controller. The detachment controller is activated by the user and this detaches the coil. The AZUR coils are designed for use with the AZUR Detachment Controller (Also known as AZUR Detachment Controller), specifically designed for coil detachment and is sold separately.
This document is a 510(k) premarket notification for a medical device called the AZUR CX Peripheral Coil System - Detachable 35. It describes the device, its intended use, and compares it to predicate devices to establish substantial equivalence.
Here's an analysis of the provided information regarding acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance:
The document provides a "Verification of Test Summary" and "Biocompatibility Summary" which can be interpreted as the acceptance criteria and the results of tests performed.
| Acceptance Criteria Category | Specific Test / Standard | Reported Device Performance |
|---|---|---|
| Bench Testing | 1. Simulated use | Passed |
| 2. Advance/Retract | Passed | |
| 3. Gel Expansion | Passed | |
| 4. Appendix Strength | Passed | |
| 5. Spring Constant | Passed | |
| 6. Pusher Sleeve Retention | Passed | |
| Biocompatibility - Coil Implant Segment | 1. Cytotoxicity (MEM Elution Test, ISO Cell Culture Agar Overlay) | Passed (implied by "Summary of Substantial Equivalence") |
| 2. Sensitization (Guinea Pig Maximization Test) | Passed (implied) | |
| 3. Irritation (ISO Intracutaneous Reactivity Evaluation Test) | Passed (implied) | |
| 4. Hemocompatibility (Hemolysis, Prothrombin Time Assay) | Passed (implied) | |
| 5. Systemic Toxicity (IV injection, Rabbit Pyrogen Test) | Passed (implied) | |
| 6. Genetic Toxicology (Bacteria Reverse Mutation Assay) | Passed (implied) | |
| 7. Intramuscular Implantation (7-day, 13-week, 26-week) | Passed (implied) | |
| Biocompatibility - Delivery Pusher Segment | 1. Cytotoxicity (MEM Elution Test, ISO Cell Culture Agar Overlay) | Passed (implied) |
| 2. Sensitization (Guinea Pig Maximization Test) | Passed (implied) | |
| 3. Irritation (ISO Intracutaneous Reactivity Evaluation Test) | Passed (implied) | |
| 4. Hemocompatibility (Hemolysis, Prothrombin Time Assay) | Passed (implied) | |
| 5. Systemic Toxicity (IV injection, Rabbit Pyrogen Test) | Passed (implied) |
Note: The document explicitly states "Passed" for bench tests. For biocompatibility, it lists test methods and standards, and the "Summary of Substantial Equivalence" implies that these tests were passed and support the equivalency claim.
2. Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):
The provided document does not specify the sample sizes used for the bench tests or biocompatibility tests. It also does not provide information on the data provenance (e.g., country of origin, retrospective or prospective nature) for these tests. This information would typically be found in the detailed test reports, which are not included in this summary.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):
This document describes a medical device, the AZUR CX Peripheral Coil System, which is an implantable coil for embolization. The tests conducted are primarily mechanical/physical bench tests and biocompatibility tests, not studies involving expert interpretation of medical images or clinical outcomes that would require a ground truth established by medical experts for a test set. Therefore, this question is not applicable in the context of the information provided.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
As the tests are primarily bench and biocompatibility tests, there is no adjudication method described or relevant for establishing a clinical "ground truth" for a test set by human experts. The results are typically objectively measured or observed (e.g., a "pass" or "fail" for a mechanical test, or quantitative results against a standard for biocompatibility).
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No MrMC comparative effectiveness study was done or is mentioned in this document. This filing is for a physical medical device (embolization coil), not an AI-powered diagnostic or assistive tool.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
Not applicable. This device is a physical implant, not an algorithm, so standalone performance in the context of AI is irrelevant.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
For the bench tests, the "ground truth" is defined by engineering specifications and performance requirements for the device's physical and functional characteristics. For biocompatibility tests, the "ground truth" is established by international standards (ISO 10993 series), which define acceptable biological responses to medical devices. There isn't a "ground truth" in the sense of clinical diagnoses or outcomes used in AI or clinical trials with human subjects.
8. The sample size for the training set:
Not applicable. The device is a physical medical implant, not an AI model that requires a training set.
9. How the ground truth for the training set was established:
Not applicable. As there is no AI model or training set, there is no ground truth to establish for a training set.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of the human form.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 25, 2015
MicroVention, Inc. Cynthia Valenzuela Project Manager, Regulatory Affairs 1311 Valencia Avenue Tustin, California 92780
Re: K151358
Trade/Device Name: AZUR CX Peripheral Coil System - Detachable 35 Regulation Number: 21 CFR 870.3300 Regulation Name: Vascular Embolization Device Regulatory Class: Class II Product Code: KRD Dated: August 18, 2015 Received: August 20, 2015
Dear Cynthia Valenzuela:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K151358
Device Name
AZUR CX Peripheral Coil System - Detachable 35
Indications for Use (Describe)
The AZUR Peripheral Coil System is intended to reduce or blood flow in vessels of the peripheral vasculature. It is intended for use in the interventional radiologic management of arteriovenous malformations, arteriovenous fistulae, aneurysms, and other lesions of the peripheral vasculature.
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(K) Summary of Safety and Effectiveness
Date Prepared: 11SEP2015
510(K) Summary
| Trade Name: | AZUR CX Peripheral Coil System - Detachable 3- |
|---|---|
| Generic Name: | Vascular Embolization Device |
| Classification: | Class II, 21 CFR 870.3300 |
| Submitted By: | MicroVention, Inc.1311 Valencia AvenueTustin, California U.S.A. |
| Contact: | Cynthia ValenzuelaProject Manager, Regulatory AffairsMicroVention, Inc.Direct: 714-247-8053Cell: 949-413-0071 |
Predicate Device:
| Number | Description | Clearance Date |
|---|---|---|
| K123384 | AZUR Peripheral HydroCoilEndovascular Embolization System -Detachable 18 | 28NOV2015 |
| K130577 | AZUR Peripheral HydroCoilEndovascular Embolization System -Pushable 18 & 35 | 17JUN2013 |
Device Description
The AZUR CX Coils consist of implant coil made of platinum alloy with inner hydrogel core. The coils are designed in 3D spherical structure in various loop sizes and lengths. The AZUR CX Coil System (Detachable) consists of an implantable coil attached to a delivery pusher. The coil system is delivered to the treatment site through the microcatheter. The proximal end of the delivery pusher is inserted to the AZUR detachment controller. The detachment controller is activated by the user and this detaches the coil. The AZUR coils are designed for use with the AZUR Detachment Controller (Also known as AZUR Detachment Controller), specifically designed for coil detachment and is sold separately.
{4}------------------------------------------------
Indication For Use
The intended use as stated in the product labeling is as follows:
The AZUR Peripheral Coil System is intended to reduce or block the rate of blood flow in vessels of the peripheral vasculature. It is intended for use in the interventional radiologic management of arteriovenous malformations, arteriovenous fistulae, aneurysms, and other lesions of the peripheral vasculature.
Technological Comparison Chart
| Features | Cleared AZUR Pushable 18& 35 (K130577) | Cleared AZUR CX (D18)(K123384) | AZUR CX (D35)Proposed device |
|---|---|---|---|
| Design Attributes | |||
| Coil Shape | Helical | 3D - Spherical | 3D - Spherical |
| Coil Filar OD (inch) | 0.00275 - 0.0050 | 0.00250 - 0.00400 | 0.0035 - 0.0050 |
| Primary Coil Wind OD(inch) | 18: 0.012035: 0.0220 | 0.0140 - 0.0150 | 0.0290 |
| 3D Coil OD (mm) | 18: 2 – 10mm35: 4 – 16mm | 4 - 20* | 4 - 20* |
| Restrained Coil Length(cm) | 18: 2 – 20mm35: 4 – 20mm | 13 - 40* | 7 - 39 |
| Pusher Length | 18: 2 - 10mm35: 4 - 16mm | 175cm | 162cm |
| Materials | |||
| Main Coil Wire | 18: Platinum (92%)/ Tungsten(8%) | Platinum/Tungsten alloy(92/8) | 4mm - 8mm:Platinum/Tungsten alloy (92/8) |
| 35: Platinum alloy-clad tantalum[Platinum (90%)/Iridium (10%)] | 10mm -20mm:Platinum alloy-clad tantalum[Platinum (90%)/Iridium (10%)] | ||
| 35 (4 mm): Platinum (92%)/Tungsten (8%) | |||
| Coil-to-Pusher Coupler | - | Platinum / Iridium (90/10) | No coupler |
| Adhesive | UV cure adhesive (DYMAX 1128-AM-VT) | UV cure adhesive (DYMAX1128-AM-VT) | UV cure adhesive (DYMAX1128-AM-VT) |
| Implant to the pusher | - | Polyolefin Elastomer | Polyolefin Elastomer |
| Stretch Resistant Filament | - | Polyolefin Elastomer | Polyolefin Elastomer |
| Hydrogel | Hydrophilic acrylic copolymer(cross-linked copolymer ofpolyethylene glycol diacrylamideand acrylic acid) | Hydrophilic acryliccopolymer (cross-linkedcopolymer of polyethyleneglycol diacrylamide andacrylic acid) | Hydrophilic acrylic copolymer(cross-linked copolymer ofpolyethylene glycoldiacrylamide and acrylic acid) |
| General | |||
| Catheter Compatibility | 18: 0.021" - 0.022" ID35: 0.041" - 0.047" ID | Compatible withmicrocatheters having anID of > .019" | Compatible with cathetershaving an ID of 0.041"-0.047" |
| Method of Supply | Sterile, single use | Sterile, single use | Sterile, single use |
| Package Configuration | Sealed in a poly (ethyleneterephthalate) pouch coated withsilica or aluminum oxide andplaced in a shipping carton | Dispenser coil, pouch &shipping carton | Dispenser coil, pouch &shipping carton |
{5}------------------------------------------------
Verification of Test Summary
| Bench Testing | Result |
|---|---|
| Simulated use | Passed |
| Advance/Retract | Passed |
| Gel Expansion | Passed |
| Appendix Strength | Passed |
| Spring Constant | Passed |
| Pusher Sleeve Retention | Passed |
Biocompatibility Coil Implant Segment - Biocompatibility Summary
| Test Method | Standard |
|---|---|
| Cytotoxicity | |
| MEM Elution Test | ISO 10993-5 |
| ISO Cell Culture Agar Overlay | ISO 10993-5 |
| Sensitization | |
| Sensitization-Guinea Pig Maximization Test | ISO 10993-10 |
| Irritation | |
| ISO Intracutaneous Reactivity Evaluation Test | ISO 10993-10 |
{6}------------------------------------------------
Coil Implant Segment - Biocompatibility Summary
| Test Method | Standard |
|---|---|
| Hemocompatibility | |
| Hemolysis | ISO 10993-4 |
| Prothrombin Time Assay - ISO | ISO 10993-4 |
| Systemic Toxicity | |
| Systemic toxicity (IV injection) | ISO 10993-11 |
| Rabbit Pyrogen Test (material mediated) | ISO 10993-11 |
| Test Method | Standard |
|---|---|
| Genetic Toxicology | |
| Bacteria Reverse Mutation Assay (Ames Test) | ISO 10993-3 |
| Intramuscular Implantation | |
| 7-day Muscle Implantation | ISO 10993-6 |
| 13-week Intramuscular Implantation Test | ISO 10993-6 |
| 26-week Intramuscular Implantation Test | ISO 10993-6 |
{7}------------------------------------------------
Biocompatibility Summary for Delivery Pusher Segment
| Test Method | Standard |
|---|---|
| Cytotoxicity | |
| MEM Elution Test | ISO 10993-5 |
| ISO Cell Culture Agar Overlay | ISO 10993-5 |
| Sensitization | |
| Sensitization-Guinea Pig Maximization Test | ISO 10993-10 |
| Irritation | |
| ISO Intracutaneous Reactivity Evaluation Test | ISO 10993-10 |
| Hemocompatibility | |
| Blood compatibility Evaluation (Hemolysis) | ISO 10993-4 |
| Prothrombin Time Assay - ISO | ISO 10993-4 |
| Systemic Toxicity | |
| Systemic toxicity(IV injection) | ISO 10993-11 |
| Rabbit Pyrogen Test (material mediated) | ISO 10993-11 |
{8}------------------------------------------------
Summary of Substantial Equivalence:
The data presented in this submission demonstrates the technological similarity and equivalency of the AZUR Peripheral Coils System - CX Detachable 35 Coils when compared with the predicate devices, MicroVention, Inc. AZUR CX Detachable 18 (K1223384) and AZUR Pushable 18 & 35 (K130577).
- The device, ●
- Has the same intended use, ●
- Uses the same operating principle, ●
- Incorporates the same basic design, ●
- Uses similar construction and material, ●
- Are packaged and sterilized using the same material and processes. .
In summary, the AZUR Peripheral Coil System – CX Detachable 35 Coils described in this submission is, in our opinion equivalent to the predicate devices.
§ 870.3300 Vascular embolization device.
(a)
Identification. A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.(b)
Classification. Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).