(498 days)
The 3D Systems, Inc. VSP Cranial System is intended for use as a collection of software to provide image segmentation and transfer of imaging information from a CT based medical scanner. The is processed by the VSP Cranial System and the result is an output data file that may then be provided as digital models or used as input in the production of physical outputs including anatomical models, templates, and surgical guides for use in the marking of cranial bone in cranial surgery.
The 3D Systems VSP® Cranial System is a collection of Commercial Off-The-Shelf (COTS) software, third party medical device software, and custom software intended to provide a variety of outputs to support cranial reconstructive surgery. The system uses CT based imaging data of the patient's anatomy with input from the physician, to manipulate original patient images for planning and executing surgery. The system produces a variety of patient specific outputs including, anatomical models (physical and digital), surgical templates / guides, and patient specific case reports.
Here's a breakdown of the acceptance criteria and the study information for the VSP Cranial System, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
| Test | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Leveraged from Predicate Device | ||
| Biocompatibility testing | (Implied: Meets established biocompatibility standards for medical devices) | Not explicitly stated for this device, but leveraged from predicate. |
| Cleaning and sterilization testing | (Implied: Demonstrates effective cleaning and sterilization methods) | Not explicitly stated for this device, but leveraged from predicate. |
| Software verification and validation | (Implied: Software functions as intended, meets specified requirements, and is safe and effective) | Not explicitly stated for this device, but leveraged from predicate. |
| Performance testing (Process validation, Simulated Use, Mechanical testing) | (Implied: All aspects of the manufacturing process, simulated use in planning, and mechanical strength/durability meet pre-established specifications) | Not explicitly stated for this device, but leveraged from predicate. |
| Specific to VSP Cranial System | ||
| Packaging Validation | Packaging and labels tested according to ASTM D4577, ASTM D642 (Method A), ASTM D4728, ASTM D3580, ASTM D5276, ASTM D6179, ASTM D880, ASTM D6179, ASTM D6653, and National Motor Freight Classification Rule 180. (Implied: Successfully withstands shipping and handling without compromise to product integrity or labeling) | All packaging and labeling met the required acceptance criteria. |
| Sterilization Compatibility | VSP® Cranial System outputs subjected to a single sterilization cycle and visually/dimensionally inspected to ensure compatibility with the validated sterilization method. (Implied: Maintains visual and dimensional integrity after sterilization.) | All acceptance criteria was met. |
| Dimensional Analysis | Sizes and shapes of VSP® Cranial System templates and guides (selected to challenge the system) were dimensionally inspected to verify conformance to the product requirements. (Implied: Dimensions are within specified tolerances.) | All acceptance criteria was met. |
| Bioburden | Bioburden testing conducted on VSP® Cranial System templates, guides, anatomical models, and metal accessories per ISO 11737-1, USP <61> and USP <1227>. (Implied: Bioburden levels are within acceptable limits for sterilization.) | All acceptance criteria was met. |
| Pyrogenicity testing | Pyrogenicity testing conducted on VSP® Cranial System templates, guides, anatomical models, and metal accessories per AAMI ST72, USP <85>, and USP <161>. (Implied: Endotoxin levels are below a specified threshold.) | All samples met the acceptance criteria of ≤ 2.15 EU/device. |
2. Sample Size Used for the Test Set and Data Provenance:
The document details performance testing that was largely leveraged from the predicate device (VSP® System, K133907) because the planning/design process, materials, manufacturing process, cleaning methods, and sterilization methods are identical.
For the tests specific to the VSP Cranial System:
- Sample Size: Not explicitly stated but "sizes and shapes of VSP® Cranial System templates and guides were selected to challenge the system" for Dimensional Analysis. For bioburden and pyrogenicity, "VSP® Cranial System templates, guides, anatomical models, and metal accessories" were tested.
- Data Provenance: This appears to be prospective testing conducted by 3D Systems for the specific VSP Cranial System device. The country of origin of the data is not specified beyond being generated by the applicant, 3D Systems, Inc., which is based in Littleton, Colorado, USA.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:
This information is not provided in the document. The testing described focuses on engineering, material, and sterilization validation rather than clinical performance or diagnostic accuracy that would require expert-established ground truth for a test set.
4. Adjudication Method for the Test Set:
This information is not provided in the document. As mentioned above, the tests are primarily engineering and material validations, which don't typically involve an adjudication method in the way a clinical study for diagnostic accuracy would.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
No, an MRMC comparative effectiveness study was not done. The study described is not a clinical study involving human readers or comparative effectiveness of AI assistance. It's a technical performance and validation study to demonstrate substantial equivalence to a predicate device.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
The device is described as a "collection of software" that provides image segmentation and processing, resulting in an "output data file that may then be provided as digital models or used as input in the production of physical outputs." The system is operated by "trained 3D Systems employees," and the "physician does not directly input information" but provides "clinical input and review."
While there is "software verification and validation testing" (leveraged from the predicate), the document does not explicitly describe a standalone performance study for the algorithm's output in terms of accuracy against a ground truth independent of the human-in-the-loop (3D Systems employees and physician review) process. The nature of this device (planning tools for customized physical outputs) means the "human-in-the-loop" is integral to its intended use and output generation.
7. The Type of Ground Truth Used:
For the specific tests performed:
- Packaging Validation: Ground truth is defined by the technical specifications of the ASTM and National Motor Freight Classification standards.
- Sterilization Compatibility: Ground truth is visual and dimensional integrity after a validated sterilization cycle.
- Dimensional Analysis: Ground truth is the "product requirements" and specified dimensional tolerances.
- Bioburden: Ground truth is defined by the acceptable bioburden limits specified in ISO 11737-1, USP <61>, and USP <1227>.
- Pyrogenicity: Ground truth is the acceptance criteria of ≤ 2.15 EU/device as per AAMI ST72, USP <85>, and USP <161>.
For the overall system, the "ground truth" for its functionality is implicitly linked to the predicate device's established performance and safety/effectiveness, as the core technologies and processes are leveraged and deemed "substantially equivalent."
8. The Sample Size for the Training Set:
This information is not applicable/not provided. The VSP Cranial System is not described as a machine learning or AI algorithm that requires a "training set" in the conventional sense for developing predictive models. It's a software system for image segmentation and design of physical outputs based on CT data.
9. How the Ground Truth for the Training Set was Established:
This information is not applicable/not provided for the same reasons as point 8.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three heads, representing the department's focus on health, human services, and the well-being of the population. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 23, 2016
3D Systems Corporation Ms. Kim Torluemke Vice President, Quality & Regulatory, Healthcare 5381 South Alkire Circle Littleton, Colorado 80127
Re: K151285
Trade/Device Name: VSP Cranial System Regulation Number: 21 CFR 882.4310 Regulation Name: Powered Simple Cranial Drills, Burrs, Trephines, and Their Accessories Regulatory Class: Class II Product Code: PPT Dated: September 21, 2016 Received: September 22, 2016
Dear Ms. Torluemke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena -S
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K151285
Device Name VSP Cranial System
Indications for Use (Describe)
The 3D Systems, Inc. VSP Cranial System is intended for use as a collection of software to provide image segmentation and transfer of imaging information from a CT based medical scanner. The is processed by the VSP Cranial System and the result is an output data file that may then be provided as digital models or used as input in the production of physical outputs including anatomical models, templates, and surgical guides for use in the marking of cranial bone in cranial surgery.
| Type of Use (Select one or both, as applicable) |
|---|
| Prescription Use (Part 21 CFR 801 Subpart D) |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for 3D Systems. The logo consists of a geometric shape on the left and the text "3D SYSTEMS" on the right. The geometric shape is a stylized cube with the letters "3D" integrated into its design. The text "3D SYSTEMS" is written in a bold, sans-serif font.
510(K) SUMMARY
1. INTRODUCTION
This document contains the 510(k) summary for the VSP® Cranial System. The content of this summary is based on the requirements of 21 CFR 807.92.
2. APPLICANT NAME AND ADDRESS
| Name: | 3D Systems, Inc. |
|---|---|
| Address: | 5381 South Alkire CircleLittleton, CO 80127, USAPhone: (720) 643-1001Fax: (720) 643-1009 |
| Official Contact: | Kim Torluemke |
Vice President, Quality and Regulatory, Healthcare
Date Prepared: September 23, 2016
3. DEVICE NAME AND CLASSIFICATION
| Trade Name: | VSP® Cranial System |
|---|---|
| Common Name: | System for the creation of patient specific anatomical models, templates,guides, and surgical plans. |
| Classification Name: | Powered Simple Cranial Drills, Burrs, Trephines, and Their Accessories |
| Classification: | Class II, 21 CFR 882.4310 |
| Product Code: | PPT |
4. PREDICATE DEVICE
- · VSP® System, 3D Systems (K133907)
5. DESCRIPTION OF THE DEVICE
The 3D Systems VSP® Cranial System is a collection of Commercial Off-The-Shelf (COTS) software, third party medical device software, and custom software intended to provide a variety of outputs to support cranial reconstructive surgery. The system uses CT based imaging data of the patient's anatomy with input from the physician, to manipulate original patient images for planning and executing surgery. The system produces a variety of patient specific outputs including, anatomical models (physical and digital), surgical templates / guides, and patient specific case reports.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for 3D Systems. The logo consists of a geometric shape on the left and the text "3D SYSTEMS" on the right. The geometric shape is a cube-like structure with the letters "3D" integrated into its design. The text "3D SYSTEMS" is written in a bold, sans-serif font.
3D Systems employees utilize a combination of Commercial Off-The-Shelf (COTS) software, third party medical device software, and custom software to manipulate CT based imaging data to create patient-specific physical and digital outputs. The process requires clinical input and review from the physician during planning and prior to delivery of the final outputs. While the process and data-flow can vary somewhat based on the particular requirements of a given patient and physician, the following description outlines the functions of key sub-components of the system, and how they interact to produce the defined system outputs. It should be noted that the system is operated only by trained 3D Systems employees, and the physician does not directly input information. The physician only provides input for model manipulation and interactive feedback through viewing of digital models of system outputs that are modified by the engineer during the planning session.
The VSP® Cranial System is made up of 10 individual pieces of software and additional manufacturing equipment integrated to provide a range of anatomical models (physical and digital), surgical templates / guides and patient-specific case reports for reconstructive surgery in the cranial region.
The VSP® Cranial System requires an input 3D image file from CT based imaging systems. This input is then used, with support from the prescribing physician, to provide the following categories of outputs to support reconstructive surgery. Each system output is designed with physician input, and reviewed by the physician prior to finalization and distribution. All outputs are used only with direct physician involvement to reduce the criticality of the outputs.
System Outputs:
- Anatomical Models
- Surgical Positioning Templates / Guides
- Osteotomy Templates / Guides
- Patient Specific Case Reports
The VSP® Cranial System also contains Stainless Steel Cutting and Drill Inserts (VSP® Cranial System Accessories) which are intended to be used by the physician to guide cutting and drilling activities during the surgical procedure. The inserts fit into a standard slot / hole in the cutting / drill guides, and can be used across all VSP® Cranial System guides and templates.
6. INDICATIONS FOR USE
The 3D Systems, Inc. VSP® Cranial System is intended for use as a collection of software to provide image segmentation and transfer of imaging information from a CT based medical scanner. The input data file is processed by the VSP® Cranial System and the result is an output data file that may then be provided as digital models or used as input in the production of physical outputs including anatomical models, templates, and surgical guides for use in the marking or cutting of cranial bone in cranial surgery.
The Indications for Use statement for the VSP® Cranial System is nearly identical to the predicate device, differing only in anatomical region. The anatomical region for which the predicate device is
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for 3D Systems. The logo consists of a geometric shape on the left and the text "3D SYSTEMS" on the right. The geometric shape is a cube-like structure with the letters "3D" embedded within it. The text "3D SYSTEMS" is written in a bold, sans-serif font.
intended is listed as maxillofacial, indicating that the device may be used only for maxillofacial applications. The VSP® Cranial System is specific to only the cranial region, and excludes maxillofacial applications. The change in anatomical region presented in the Indications for Use statement of the VSP® Cranial System as compared to the predicate device do not change the therapeutic effects of the device.
Both the subject and predicate devices have the same intended use for providing image segmentation for the transfer of imaging information from a medical scanner, pre-operative software tools for simulating / evaluating surgical treatment options, and patient specific outputs for use in reconstructive surgery.
7. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The intended use and technological characteristics of the subject device (VSP® Cranial System) are either identical or substantially equivalent to the predicate device (VSP® System), differing only in the anatomical region for which they are intended. The potential impact on substantial equivalence of each technological difference was addressed by extensive risk analysis and verification and validation testing.
The VSP® Cranial System employs identical fundamental technologies as the predicate device:
- The VSP Cranial System uses a subset of the software previously cleared in the predicate device.
- Hardware for rapid manufacturing of patient-specific anatomical models, guides and templates is identical to the previously cleared predicate device.
The principles of operation and technological characteristics are all either identical or substantially equivalent between the VSP® Cranial System and the predicate device.
The VSP® Cranial System has the following identical technological characteristics:
- System Inputs: Images from CT based medical scanners
- System Outputs: Physical and digital outputs such as patient-specific anatomical models, cutting and drill guides, and templates
- Materials: Biocompatible polymers and surgical instrument grade stainless steel
- Sterility Assurance Level: 1 x 10-6
The intended use of the VSP® Cranial System is similar to the predicate device, differing only in the anatomical region in which the device is intended for:
- Both devices are intended to provide tools and accessories (software for image manipulation, anatomical models, guides and templates) for use in reconstructive surgery.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for 3D Systems. The logo consists of a geometric shape on the left and the text "3D SYSTEMS" on the right. The geometric shape is a stylized cube with the letters "3D" and "S" incorporated into its design. The text "3D SYSTEMS" is written in a bold, sans-serif font.
- The VSP® Cranial System and its predicate device are both intended to be used by trained personnel, in a non-medical manufacturing or office environment, with active support from the surgeon.
- The VSP® Cranial System and its predicate device are both intended for use on surgical candidates undergoing complex reconstructive surgery.
The VSP® Cranial System and its predicate's accessories are manufactured to ASTM F899-11 and are sterilized by the healthcare facility with the same cycle as the other system outputs.
8. SUMMARY OF PERFORMANCE TESTING
The planning/design process, materials, manufacturing process, cleaning methods and sterilization methods for VSP® Cranial System are identical to those used for the predicate device. Therefore, the following testing was leveraged from the predicate device:
- Biocompatibility testing
- Cleaning and sterilization testing
- Software verification and validation testing
- · Performance testing
- o Process validation (IQ/OQ/PQ)
- Simulated Use (Planning session) o
- o Mechanical testing
VSP® Cranial System output guides and templates differ from the predicate device only in size and shape to accommodate the anatomical region for which they are intended. To address any concerns with the similar performance of the VSP® Cranial System, the following performance data were provided in addition to the leveraged testing to support the substantial equivalence determination:
| Test | Test Method Summary | Results |
|---|---|---|
| Packaging Validation | Product packaging and labels were testedaccording to ASTM D4577, ASTM D642, MethodA, ASTM D4728, ASTM D3580, ASTM D5276,ASTM D6179, ASTM D880, ASTM D6179, ASTMD6653, and National Motor Freight ClassificationRule 180 | All packaging and labelingmet the requiredacceptance criteria |
| SterilizationCompatibility | To ensure VSP® Cranial System outputs arecompatible with the validated sterilizationmethod, outputs were subjected to a singlesterilization cycle and visually/dimensionallyinspected | All acceptance criteria wasmet |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for 3D Systems. The logo consists of a geometric shape on the left and the text "3D SYSTEMS" on the right. The geometric shape is a stylized cube with the letters "3D" incorporated into its design. The text "3D SYSTEMS" is written in a bold, sans-serif font.
| Test | Test Method Summary | Results |
|---|---|---|
| Dimensional Analysis | Sizes and shapes of VSP® Cranial Systemtemplates and guides were selected to challengethe system and were dimensionally inspected toverify conformance to the product requirements | All acceptance criteria wasmet |
| Bioburden | Bioburden testing was conducted on VSP®Cranial System templates, guides, anatomicalmodels, and metal accessories per ISO 11737-1,USP <61> and USP <1227> | All acceptance criteria wasmet |
| Pyrogenicity testing | Pyrogenicity testing was conducted on VSP®Cranial System templates, guides, anatomicalmodels, and metal accessories per AAMI ST72,USP <85>, and USP <161> | All samples met theacceptance criteria of ≤ 2.15EU/device |
Summary
All Design, Process, and other Verification and Validation testing, which were conducted as a result of risk analyses and design impact assessments, showed conformity to pre-established specifications and acceptance criteria. The acceptance criteria was established in support of device performance, and testing demonstrated substantial equivalence of the system to the predicate device.
9. CONCLUSION
Based on a comparison of the intended use and technological characteristics, the VSP® Cranial System is substantially equivalent to the identified predicate device. Minor differences in the indications for use do not raise new or different questions of safety and effectiveness. Validation data supports that the system performs in accordance with its intended use and is substantially equivalent to the predicate device.
§ 882.4310 Powered simple cranial drills, burrs, trephines, and their accessories.
(a)
Identification. Powered simple cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments are used with a power source but do not have a clutch mechanism to disengage the tip after penetrating the skull.(b)
Classification. Class II (performance standards).