(47 days)
Not Found
No
The summary describes a standard oscillometric blood pressure monitor and does not mention any AI or ML capabilities.
No
The device is intended for measurement and monitoring of blood pressure, not for therapy or treatment of a condition.
Yes
The device is intended to measure diastolic and systolic blood pressures and pulse rate, which are used to evaluate a person's cardiovascular health, indicating its role in diagnosis.
No
The device description explicitly states it is a "battery driven automatic non-invasive blood pressure meter" and mentions an "inflatable cuff," indicating it includes hardware components beyond just software. The performance studies also reference standards related to electrical equipment and biological evaluation, further supporting the presence of hardware.
No, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostics are medical devices used to perform tests on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections. The tests are performed outside the body (in vitro).
- Device Function: The AGE Automatic Wrist Blood Pressure Monitor measures blood pressure and pulse rate by applying a cuff to the wrist and using the oscillometry technique. This is a non-invasive measurement performed on the body, not on a sample taken from the body.
- Intended Use: The intended use clearly states it's a "non-invasive blood pressure measurement system."
Therefore, based on the provided information, the AGE Automatic Wrist Blood Pressure Monitor is a non-invasive medical device, not an In Vitro Diagnostic.
N/A
Intended Use / Indications for Use
AGE Automatic Wrist Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 13.5cm~19.5 cm.
Product codes
DXN
Device Description
AGE Automatic Wrist Blood Pressure Monitor is a battery driven automatic non-invasive blood pressure meter. It can automatically conduct the inflation and measurement, which can measure systolic and diastolic blood pressure as well as the pulse rate of adult at wrist within its claimed range and accuracy via the Oscillometry technique. The device also has low voltage indication, which will be triggered when the battery is low.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Wrist
Indicated Patient Age Range
Adult
Intended User / Care Setting
Medical professionals or at home
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
AGE Automatic Wrist Blood Pressure Monitor has been evaluated the safety and performance by lab bench testing according to the following standards:
- IEC 60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety, 2005 + A1:2012
- IEC 60601-1-2, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic Disturbances -Requirements and tests, 2014
- ISO 10993-5, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity, 2009
- ISO 10993-10, Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization, 2010
- AAMI / ANSI / ISO 81060-2 Second Edition, Non-Invasive Sphygmomanometers - Part 2: Clinical Validation Of Automated Measurement Type. (Cardiovascular)
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Pressure: +/-3 mmHg
Pulse: +/-5%
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.1130 Noninvasive blood pressure measurement system.
(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of three human profiles facing right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 30, 2015
Dongguan Ageless Health Industrial Co., Ltd c/o Ms. Cecilia Ceng Regulatory Manager Suite 306, Kecheng Mansion, No.121 Science Road, Guahgzhou Science Park Guangzhou, 510663 CN
Re: K151281
Trade/Device Name: AGE Automatic Wrist Blood Pressure Monitor, Models BW-601, BW-602. BW-603. BW-605. BW-606. BW-611. BW-612. BW-613 Regulation Number: 21 CFR 870.1130 Regulation Name: Non-Invasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: April 20, 2015 Received: May 14, 2015
Dear Ms. Cecilia Ceng,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
1
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mitchell Stein
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
2
Page __ 1_ of 1
510(k) Number (if known): K151281
De vice Name : 605, BW -606, BW -611, BW -612, and BW -613
Indications For Use: AGE Automatic Wrist Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 13.5cm.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
OR
Prescription Use (Per 21 CFR 801.109) Over-The-Counter Use
3
510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR 870.1130.
1. Submitter Information
Sponsor Name: Dongguan Ageless Health Industrial Co.,Ltd. Address: 3/F, A1 Bldg, Dongshen Sima Industrial Area, No.33 Shenbei Road, Sima Village, Changping Town, Dongguan City, Guangdong Province, China Contact Person: Victor Wan (Vice-president) Phone: +86-769-81158038 Fax: +86-769-82289331 E-mail: victor@agelh.com
Application Correspondent:
Contact Person: Ms. Cecilia Ceng / Mr. Tim Wong Guangzhou GLOMED Biological Technology Co., Ltd. Tel: +86-20-61099984 Email: regulatory@glomed-info.com
Subject Device Information 2.
Type of 510(k): | Traditional |
---|---|
----------------- | ------------- |
Common Name: Noninvasive blood pressure measurement systems
Trade Name: AGE Automatic Wrist Blood Pressure Monitor, Models BW-601,
BW-602, BW-603, BW-605, BW-606, BW-611, BW-612, BW-613
Classification Name: Noninvasive brood pressure measurement system
4
K151281 Page 2 of 6
Review Panel: | Cardiovascular |
---|---|
Product Code: | DXN |
Regulation Number: | 21 CFR 870.1130 |
Regulation Class: | 2 |
3. Predicate Device Information
Sponsor: | Health & Life Co.,Ltd. |
---|---|
Common Name: | Noninvasive blood pressure measurement systems |
Trade Name: | Full Automatic (NIBP) Blood Pressure Monitor, Model HL158LA |
510(k) number: | K130564 |
Review Panel: | Cardiovascular |
Product Code: | DXN |
Regulation Number: | 21 CFR 870.1130 |
Regulation Class: | 2 |
4. Device Description
AGE Automatic Wrist Blood Pressure Monitor is a battery driven automatic non-invasive blood pressure meter. It can automatically conduct the inflation and measurement, which can measure systolic and diastolic blood pressure as well as the pulse rate of adult at wrist within its claimed range and accuracy via the Oscillometry technique. The device also has low voltage indication, which will be triggered when the battery is low.
5. Intended Use
AGE Automatic Wrist Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 13.5 cm~19.5 cm.
6. Test Summary
5
K151281 Page 3 of 6
AGE Automatic Wrist Blood Pressure Monitor has been evaluated the safety and performance by lab bench testing according to the following standards:
- � IEC 60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety, 2005 + A1:2012
- � IEC 60601-1-2, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic Disturbances -Requirements and tests, 2014
- � ISO 10993-5, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity, 2009
- � ISO 10993-10, Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization, 2010
- ◆ AAMI / ANSI / ISO 81060-2 Second Edition, Non-Invasive Sphygmomanometers - Part 2: Clinical Validation Of Automated Measurement Type. (Cardiovascular)
7. Comparison to Predicate Device
Compare with predicate device, the subject device is very similar in design principle, intended use, indications for use, functions, material and the applicable standards. The differences between subject device and predicate device do not raise and new questions of safety or effectiveness.
| Elements of
Comparison | Subject Device | Predicate Device | Verdict |
---|---|---|---|
Product Name | AGE Automatic Wrist Blood | ||
Pressure Monitor | Full Automatic (NIBP) Blood | ||
Pressure Monitor, Model HL158LA | -- | ||
Intended Use and Indications for Use | |||
Intended Use | AGE Automatic Wrist Blood | ||
Pressure Monitor is intended for use | |||
by medical professionals or at home | |||
to monitor and display diastolic, | |||
systolic blood pressure and pulse | |||
rate on adult each time, with an air | |||
cuff buckled around one's wrist | HL158LA automatically measures | ||
human's Systolic, Diastolic blood | |||
pressure and heart rate by using the | |||
oscillometric method. All values can | |||
be read out in one LCD panel. | |||
Measurement position is at human | |||
being's wrist. The intended use of | SE | ||
Note 1 | |||
Elements of | |||
Comparison | Subject Device | Predicate Device | Verdict |
Indications for | |||
Use | according to the instruction in the | ||
user's guide manual. | this over-the-counter device is for | ||
use by people over the age of 18 | |||
with wrist circumference ranging | |||
from approx. 5.3 to 7.7 inches(13.5 | |||
cm to 19.5 cm) and for home use. | |||
AGE Automatic Wrist Blood | |||
Pressure Monitor is for use by | |||
medical professionals or at home | |||
and is a non-invasive blood | |||
pressure measurement system | |||
intended to measure the diastolic | |||
and systolic blood pressures and | |||
pulse rate of an adult individual by | |||
using a non-invasive technique in | |||
which an inflatable cuff is wrapped | |||
around the wrist. The cuff | |||
circumference is limited to | |||
13.5cm~19.5cm. | HL158LA automatically measures | ||
human's Systolic, Diastolic blood | |||
pressure and heart rate by using the | |||
oscillometric method. All values can | |||
be read out in one LCD panel. | |||
Measurement position is at human | |||
being's wrist. The intended use of | |||
this over-the-counter device is for | |||
use by people over the age of 18 | |||
with wrist circumference ranging | |||
from approx. 5.3 to 7.7 inches(13.5 | |||
cm to 19.5 cm) and for home use. | SE | ||
Note 1 | |||
ELECTRICAL REQUIREMENT | |||
Power Supply | 3Vdc (2 "AAA" batteries) | 3Vdc (2 "AAA" batteries) | SE |
PERFORMANCE SPECIFICATION | |||
Measuring | |||
Method | Oscillometry | Oscillometry | SE |
Measuring | |||
Range | Pressure: 0~294 mmHg | ||
Pulse: 40~199 beats/minute | Pressure: 0~300 mmHg | ||
Pulse: 40~199 beats/minute | SE | ||
Note 2 | |||
Accuracy | Pressure: ±3 mmHg | ||
Pulse: ±5% | Pressure: ±3mmHg | ||
Pulse: ±5% | SE | ||
Patient | |||
Population | Adult | Adult | SE |
Measurement | |||
Site of Body | Wrist | Wrist | SE |
Cuff | |||
Circumference | 13.5 ~ 19.5 cm | 13.5 ~ 19.5 cm | SE |
Inflation and | |||
Deflation | Automatic | Automatic | SE |
Memory Size | 2 x 90 sets record | 3×40 sets record | SE |
Note 3 | |||
Storage | |||
Environment | OPERATING & STORAGE CONDITIONS | ||
Temperature: -20°C ~ +65°C | |||
Humidity: 15~95%RH | |||
Atmospheric Pressure:86 kPa~106 | Temperature: -25°C ~ +70 °C | ||
Humidity: ≤ 93%RH | SE | ||
Note 3 | |||
Elements of | |||
Comparison | Subject Device | Predicate Device | Verdict |
kPa | |||
Working | |||
Environment | Temperature: 5°C ~ 40°C | ||
Humidity: 15~90%RH | |||
Atmospheric Pressure:86 kPa~106 kPa | Temperature: 5°C ~ 40°C | ||
Humidity: 15~93%RH | SE | ||
Note 3 | |||
COMPLIANCE STANDARDS | |||
Electrical, | |||
Mechanical and | |||
Thermal | |||
Evaluation | IEC 60601-1 | ||
IEC 60601-1-2 | IEC 60601-1 | ||
IEC 60601-1-2 | |||
IEC 60601-11 | SE | ||
Note 4 | |||
Biocompatibility | |||
Evaluation | All the patient contracting materials | ||
are evaluated by the | |||
biocompatibility standard ISO | |||
10993 -5, -10. | All the patient contracting materials | ||
are evaluated by the | |||
biocompatibility standard ISO | |||
10993 -1, 5, -10. | SE | ||
Note 5 | |||
Performance | ISO 81060-2 | AAMI SP10 | SE |
Note 6 |
6
7
Note 1
Although there is little difference for measurement cuff circumference of subject device and predicate device, both of them are complied with AAMI SP10 or ISO 81060-2. This difference does not affect the safety and effectiveness.
Note 2
Although the measuring range of pressure and pulse of subject device and predicate device are different, both of them are complied with AAMI SP10 or ISO 81060-2. The difference of their measuring range does not affect the safety and effectiveness.
Note 3
Although some specifications of operating & storage conditions, memory size are different for subject device and predicate device, they are both complied with IEC 60601-1. The differences do not affect the safety and effectiveness.
Note 4
8
K151281 Page 6 of 6
Although the electrical, mechanical and thermal evaluation of subject device and predicate device are different, they are both complied with IEC 60601-1 and IEC 60601-1-2. The differences do not affect the safety and effectiveness.
Note 5
Although the biocompatibility evaluation of subject device and predicate device are different, they are both complied with ISO 10993-5 and ISO 10993-10. The differences do not affect the safety and effectiveness.
Note 6
Although the standards of performance have updated and substituted from AAMI SP10 to ISO 81060-2, all of the requirements in these three standards are intended for blood pressure monitor. Therefore, both of them met the requirements. The differences do not affect the safety and effectiveness.
8. Conclusion
The subject device AGE Automatic Wrist Blood Pressure Monitor has all features of the predicate device. The few differences do not affect the safety and effectiveness of the subject device. Thus, the subject device is substantially equivalent to the predicate device.
9. Summary Prepared Date
29 June 2015