(29 days)
Not Found
No
The device description and performance studies focus on immunochromatographic assay technology and standard laboratory testing methods (GC/MS). There is no mention of AI or ML in the document.
No
This device is an in vitro diagnostic (IVD) test used to detect the presence of Secobarbital and Oxycodone in urine. It provides preliminary test results for drug screening, not for treatment or therapy.
Yes
The "Intended Use / Indications for Use" section explicitly states "For in vitro diagnostic use only," which is a key characteristic of diagnostic devices.
No
The device description clearly states that the devices are "immunochromatographic assays that use a lateral flow system," which are physical test strips and cups, not software. The performance studies also describe testing physical samples and devices.
Yes, this device is an IVD (In Vitro Diagnostic).
The intended use statement explicitly states: "For in vitro diagnostic use only." This phrase is a key indicator that the device is intended to be used outside of the body to examine specimens (in this case, urine) for diagnostic purposes.
Furthermore, the device description and performance studies confirm its use in analyzing human urine samples to detect the presence of specific substances (Secobarbital and Oxycodone), which is a typical function of an in vitro diagnostic device.
N/A
Intended Use / Indications for Use
The AssureTech Secobarbital Strip test is an immunochromatographic assay for the qualitative determination of Secobarbital in human urine at a Cut-Off concentration of 300ng/mL.
The test may yield preliminary positive results when prescription drug Secobarbital is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.
The AssureTech Oxycodone Strip test is an immunochromatographic assay for the qualitative determination of Oxycodone in human urine at a Cut-Off concentration of 100ng/mL.
The test may yield preliminary positive results when prescription drug Oxycodone is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Oxycodone in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.
The AssureTech Secobarbital/Oxycodone Panel Dip test is an immunochromatographic assay for the qualitative determination of Secobarbital and Oxycodone in human urine at a Cut-Off concentration of 300ng/mL and 100 ng/mL, respectively.
The test may yield preliminary positive results when prescription drug Secobarbital or Oxycodone is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital or Oxycodone in urine. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.
The AssureTech Secobarbital/Oxycodone Quick Cup test is an immunochromatographic assay for the qualitative determination of Secobarbital and Oxycodone in human urine at a Cut-Off concentration of 300ng/mL and 100 ng/mL, respectively.
The test may yield preliminary positive results when prescription drug Secobarbital or Oxycodone is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital or Oxycodone in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.
The AssureTech Secobarbital/Oxycodone Turn Key-Split Cup test is an immunochromatographic assay for the qualitative determination of Secobarbital and Oxycodone in human urine at a Cut-Off concentration of 300ng/mL and 100 ng/mL. respectively.
The test may yield preliminary positive results when prescription drug Secobarbital or Oxycodone is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital or Oxycodone in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.
Product codes (comma separated list FDA assigned to the subject device)
DIS, DJG
Device Description
Secobarbital Strip. The AssureTech AssureTech Oxycodone Strip. AssureTech Secobarbital/Oxycodone Panel Dip, AssureTech Secobarbital/Oxycodone Quick Cup and AssureTech Secobarbial/Oxycodone Turn Key-Split Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of Secobarbital and/or Oxycodone (target analytes) in human urine. The Quick Cup does not contain a turn-key for device activation. The tests are the first step in a two-step process. The second step is to send the sample for laboratory testing if preliminary positive results are obtained.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Precision Test Set: Samples with concentrations of -100% cut off, -75% cut off, -50% cut off, -25% cut off, +25% cut off, +50% cut off, +75% cut off and +100% cut off. These samples were prepared by spiking drug in negative samples. Each drug concentration was confirmed by GC/MS. All sample aliquots were blindly labeled by the person who prepared the samples and didn't take part in the sample testing. For each concentration, tests were performed two runs per day for 25 days per device in a randomized order.
Cut-off Test Set: A total of 150 samples equally distributed at concentrations of -50% Cut-Off: -25% Cut-Off: Cut-Off: +25% Cut-Off: +50% Cut-Off were tested using three different lots of each device by three different operators.
Interference Test Set: Potential interfering substances found in human urine of physiological or pathological conditions were added to drug-free urine and target drugs urine with concentrations at 25% below and 25% above Cut-Off levels. These urine samples were tested using three batches of each device.
Specificity Test Set: Drug metabolites and other components that are likely to interfere in urine samples were tested using three batches of each device.
Urine Specific Gravity and pH Effect Test Set: Urine samples, with 1.000 to 1.035 specific gravity or urine samples with pH 4 to 9 were spiked with target drugs at 25% below and 25% above Cut-Off levels. These samples were tested using three batches of each device.
Comparison Studies Test Set: 80 (40 negative and 40 positive) unaltered clinical samples. The samples were blind labeled and compared to GC/MS results.
Lay-user Study Test Set: Urine samples were prepared at the following concentrations; negative, +/-75%, +/-50%, +/-25% of the cutoff by spiking drug(s) into drug free-pooled urine specimens. The concentrations of the samples were confirmed by GC/MS. Each sample was aliquoted into individual containers and blind-labeled.
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
1. Analytical Performance
- Precision: Precision studies were carried out for samples at various concentrations (-100% cut off, -75% cut off, -50% cut off, -25% cut off, +25% cut off, +50% cut off, +75% cut off and +100% cut off). For each concentration, tests were performed two runs per day for 25 days per device in a randomized order (total 50 tests per concentration per lot).
- Secobarbital: For all four devices (Secobarbital Strip, Secobarbital/Oxycodone Panel Dip, Secobarbital/Oxycodone Turn-Key Split Cup, Secobarbital/Oxycodone Quick Cup), across three lots, results were consistently negative for -100%, -75%, -50%, -25% cut off concentrations (50-/0+ for all). Results were consistently positive for +25%, +50%, +75%, +100% cut off concentrations (50+/0- for all). At the cut-off concentration, the results varied slightly across lots and devices (e.g., 2-/48+ or 3-/47+).
- Oxycodone: For all four devices (Oxycodone Strip, Secobarbital/Oxycodone Panel Dip, Secobarbital/Oxycodone Turn-Key Split Cup, Secobarbital/Oxycodone Quick Cup), across three lots, results were consistently negative for -100%, -75%, -50%, -25% cut off concentrations (50-/0+ for all). Results were consistently positive for +25%, +50%, +75%, +100% cut off concentrations (50+/0- for all). At the cut-off concentration, the results varied slightly across lots and devices (e.g., 1-/49+ or 2-/48+ or 3-/47+).
- Linearity: Not applicable.
- Stability: The devices are stable at 4-30 ℃ for 24 months based on accelerated stability study at 45°C and real time stability determination at both 4°C and 30°C.
- Cut-off: A total of 150 samples equally distributed at concentrations of -50% Cut-Off: -25% Cut-Off: Cut-Off: +25% Cut-Off: +50% Cut-Off were tested using three different lots of each device by three different operators. Results were all positive at and above +25% Cut-off and all negative at and below -25% Cut-off for both Secobarbital and Oxycodone.
- Secobarbital: Cut-off 300 ng/mL
- Oxycodone: Cut-off 100 ng/mL
- Interference: Compounds that showed no interference at a concentration of 100µg/mL were summarized. No differences observed for different devices.
- Specificity: Drug metabolites and other components were tested. Lowest concentrations causing positive results were listed with cross-reactivity percentages. No differences observed for different devices.
- Secobarbital (Cut-off=300 ng/mL): Secobarbital (100%), Amobarbital (30%), Alphenol (480%), Aprobarbital (120%), Butabarbital (300%), Butathal (60%), Butalbital (6%), Cyclopentobarbital (60%), Pentobarbital (150%), Phenobarbital (100%).
- Oxycodone (Cut-off=100 ng/mL): Oxycodone (100%), Codeine (0.1%), Acetylmorphine (0.1%), Oxymorphone (40%), Dihydrocodeine (0.1%), Hydromorphone (0.1%), Hydrocodone (3.2%), Morphine (0.1%), Buprenorphine (0.1%), Ethylmorphine (0.1%).
- Effect of Urine Specific Gravity and Urine pH: Urine samples with 1.000 to 1.035 specific gravity or pH 4 to 9, spiked with target drugs at 25% below and 25% above Cut-Off levels, were tested. Results were all positive for samples at and above +25% Cut-Off and all negative for samples at and below -25% Cut-Off. No differences observed for different devices.
2. Comparison Studies
- Method comparison studies were performed in-house with three laboratory assistants for each device.
- Operators ran 80 (40 negative and 40 positive) unaltered clinical samples.
- Samples were blind labeled and compared to GC/MS results.
- Secobarbital Strip: 1 discordant result (Sample 817130, GC/MS 286 ng/mL, Strip Viewer Results Positive by all three viewers).
- Secobarbital Panel Dip: 2 discordant results (Sample 817130, GC/MS 286 ng/mL, Panel Dip Viewer Results Positive by all three viewers; Sample 933321, GC/MS 306 ng/mL, Negative by Viewer A).
- Secobarbital Turn-Key Split Cup: 5 discordant results.
- Secobarbital Quick Cup: 4 discordant results.
- OXY Strip: 4 discordant results.
- OXY Panel Dip: 3 discordant results.
- OXY Turn-Key Split Cup: 3 discordant results.
- OXY Quick Cup: 5 discordant results.
Lay-user study
- Sample Size: 1060 lay persons.
- Sites: Three intended user sites.
- Participants: Diverse educational and professional backgrounds, aged 18 to >50 years.
- Study Design: Participants received package insert, 1 blind labeled sample, and a device.
- Results:
- Secobarbital Strip:
- -100% to -50% Cutoff: 100% correct negative results.
- -25% Cutoff: 90% correct negative results (2 positive, 18 negative out of 20 samples with GC/MS 228 ng/mL).
- +25% to +75% Cutoff: 100% correct positive results, except for +25% (20 positive, 0 negative for +50%, +75%; 20 positive, 0 negative for +25%).
- For +25% Cutoff (GC/MS 369 ng/mL), 20 positive, 0 negative for a 100% correct result.
- Oxycodone Strip:
- -100% to -50% Cutoff: 100% correct negative results.
- -25% Cutoff: 95% correct negative results (1 positive, 19 negative out of 20 samples with GC/MS 80 ng/mL).
- +25% Cutoff: 95% correct positive results (19 positive, 1 negative out of 20 samples with GC/MS 127 ng/mL).
- +50% to +75% Cutoff: 100% correct positive results.
- BAR/OXY Panel Dip Card:
- Secobarbital: For -100% to -50% Cutoff, 100% correct negative. For -25% Cutoff, 95% correct (1 positive, 19 negative). For +25% Cutoff, 95% correct (19 positive, 1 negative). For +50% to +75% Cutoff, 100% correct positive.
- Oxycodone: For -100% to -50% Cutoff, 100% correct negative. For -25% Cutoff, 95% correct (1 positive, 19 negative). For +25% Cutoff, 90% correct (18 positive, 2 negative). For +50% to +75% Cutoff, 100% correct positive.
- BAR/OXY Turn-Key Split Cup:
- Secobarbital: For -100% to -50% Cutoff, 100% correct negative. For -25% Cutoff, 90% correct (2 positive, 18 negative). For +25% to +75% Cutoff, 100% correct positive.
- Oxycodone: For -100% to -50% Cutoff, 100% correct negative. For -25% Cutoff, 95% correct (1 positive, 19 negative). For +25% to +75% Cutoff, 95% to 100% correct positive (+25% was 95%, others 100%).
- BAR/OXY Quick Cup:
- Secobarbital: For -100% to -50% Cutoff, 100% correct negative. For -25% Cutoff, 95% correct (1 positive, 19 negative). For +25% to +75% Cutoff, 100% correct positive.
- Oxycodone: For -100% to -50% Cutoff, 100% correct negative. For -25% Cutoff, 95% correct (1 positive, 19 negative). For +25% to +75% Cutoff, 100% correct positive.
- Secobarbital Strip:
- Survey Results: All lay users indicated that the device instructions can be easily followed.
- Flesch-Kincaid Reading Analysis: Grade Level of 7.
3. Clinical Studies
- Not applicable.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found (Raw data for calculations provided, but not the calculated metrics themselves.)
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 862.3150 Barbiturate test system.
(a)
Identification. A barbiturate test system is a device intended to measure barbiturates, a class of hypnotic and sedative drugs, in serum, urine, and gastric contents. Measurements obtained by this device are used in the diagnosis and treatment of barbiturate use or overdose and in monitoring levels of barbiturate to ensure appropriate therapy.(b)
Classification. Class II (special controls). A barbiturate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
0
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The text is in all caps and appears to be in a sans-serif font. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 4, 2015
ASSURE TECH. CO., LTD C/O JOE SHIA REGULATORY CONSULTANT 504 E DIAMOND AVE., SUITE F GAITHERSBURG MD 20877
Re: K151211
Trade/Device Name: AssureTech Secobarbital Strip AssureTech Oxycodone Strip AssureTech Secobarbital/Oxycodone Panel Dip AssureTech Secobarbital/Oxycodone Quick Cup AssureTech Secobarbital/Oxycodone Turn Kev-Split Cup Regulation Number: 21 CFR 862.3150 Regulation Name: Barbiturate test system Regulatory Class: II
Product Code: DIS. DJG Dated: April 3, 2015 Received: May 6, 2015
Dear Mr. Joe Shia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements
1
as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Katherine Serrano -S
For: Courtney H. Lias, Ph.D. Director
Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K151211
Device Name
AssureTech Secobarbital Strip, AssureTech Oxycodone Strip, AssureTech Secobarbital/Oxycodone Panel Dip AssureTech Secobarbital/Oxycodone Quick Cup AssureTech Secobarbial/Oxycodone Turn Key-Split Cup
Indications for Use (Describe)
The AssureTech Secobarbital Strip test is an immunochromatographic assay for the qualitative determination of Secobarbital in human urine at a Cut-Off concentration of 300ng/mL.
The test may yield preliminary positive results when prescription drug Secobarbital is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.
The AssureTech Oxycodone Strip test is an immunochromatographic assay for the qualitative determination of Oxycodone in human urine at a Cut-Off concentration of 100ng/mL.
The test may yield preliminary positive results when prescription drug Oxycodone is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Oxycodone in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.
The AssureTech Secobarbital/Oxycodone Panel Dip test is an immunochromatographic assay for the qualitative determination of Secobarbital and Oxycodone in human urine at a Cut-Off concentration of 300ng/mL and 100 ng/mL, respectively.
The test may yield preliminary positive results when prescription drug Secobarbital or Oxycodone is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital or Oxyoodone in urine. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.
The AssureTech Secobarbital/Oxycodone Quick Cup test is an immunochromatographic assay for the qualitative determination of Secobarbital and Oxycodone in human urine at a Cut-Off concentration of 300ng/mL and 100 ng/mL, respectively.
The test may yield preliminary positive results when prescription drug Secobarbital or Oxycodone is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital or Oxycodone in urine. The test
3
provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.
The AssureTech Secobarbital/Oxycodone Turn Key-Split Cup test is an immunochromatographic assay for the qualitative determination of Secobarbital and Oxycodone in human urine at a Cut-Off concentration of 300ng/mL and 100 ng/mL. respectively.
The test may yield preliminary positive results when prescription drug Secobarbital or Oxycodone is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital or Oxycodone in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.
Type of Use (Select one or both, as applicable) | |
---|---|
Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
4
510(k) SUMMARY
1. Date: | June 3, 2015 |
---|---|
2. Submitter: | Assure Tech. Co., Ltd. |
Building 1, No.10, Xiyuansan Road, Westlake | |
Economic Zone | |
Hangzhou, China, 310030 | |
3. Contact person: | Eric Lin |
Assure Tech. Co., Ltd. | |
Building 1, No.10, Xiyuansan Road, Westlake | |
Economic Zone | |
Hangzhou, China, 310030 | |
Telephone: 510-860-4680 | |
Email: customerservice@assurelabs.com. | |
4. Device Name: | AssureTech Secobarbital Strip |
AssureTech Oxycodone Strip | |
AssureTech Secobarbital/Oxycodone Panel Dip | |
AssureTech Secobarbital/Oxycodone Quick Cup | |
AssureTech Secobarbital/Oxycodone Turn Key-Split Cup |
Classification:
Product Code | CFR | Panel |
---|---|---|
DIS | 21 CFR, 862.3150 Barbiturate Test System | Toxicology |
DJG | 21 CFR, 862.3650 Opiate Test System | Toxicology |
5. Predicate Devices: K052115
FIRST CHECK DIAGNOSTICS LLC FIRST CHECK MULTI DRUG CUP 12
-
- Intended Use
The AssureTech Secobarbital Strip test is an immunochromatographic assay for the qualitative determination of Secobarbital in human urine at a Cut-Off concentration of 300ng/mL.
- Intended Use
The test may vield preliminary positive results when prescription drug Secobarbital is ingested. even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.
The AssureTech Oxycodone Strip test is an immunochromatographic assay for the qualitative determination of Oxycodone in human urine at a Cut-Off concentration of 100ng/mL.
5
The test may yield preliminary positive results when prescription drug Oxycodone is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Oxycodone in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.
The AssureTech Secobarbital/Oxycodone Panel Dip test is an immunochromatographic assay for the qualitative determination of Secobarbital and Oxycodone in human urine at a Cut-Off concentration of 300ng/mL and 100 ng/mL, respectively.
The test may vield preliminary positive results when prescription drug Secobarbital or Oxycodone is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital or Oxycodone in urine. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.
The AssureTech Secobarbital/Oxycodone Quick Cup test is an immunochromatographic assay for the qualitative determination of Secobarbital and Oxycodone in human urine at a Cut-Off concentration of 300ng/mL and 100 ng/mL, respectively.
The test may yield preliminary positive results when prescription drug Secobarbital or Oxycodone is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital or Oxycodone in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.
The AssureTech Secobarbital/Oxycodone Turn Key-Split Cup test is an immunochromatographic assay for the qualitative determination of Secobarbital and Oxycodone in human urine at a Cut-Off concentration of 300ng/mL and 100 ng/mL. respectively.
The test may yield preliminary positive results when prescription drug Secobarbital or Oxycodone is ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Secobarbital or Oxycodone in urine. The test provides only preliminary test
6
results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
For in vitro diagnostic use only. The test is intended for over-the-counter and for prescription use.
-
- Device Description
Secobarbital Strip. The AssureTech AssureTech Oxycodone Strip. AssureTech Secobarbital/Oxycodone Panel Dip, AssureTech Secobarbital/Oxycodone Quick Cup and AssureTech Secobarbial/Oxycodone Turn Key-Split Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of Secobarbital and/or Oxycodone (target analytes) in human urine. The Quick Cup does not contain a turn-key for device activation. The tests are the first step in a two-step process. The second step is to send the sample for laboratory testing if preliminary positive results are obtained.
- Device Description
-
- Substantial Equivalence Information
A summary comparison of features of the candidate devices and the predicate devices is provided in Tables 1 to 4 below.
- Substantial Equivalence Information
Item | Device | Predicate - K052115 |
---|---|---|
Indication(s) | ||
for Use | For the qualitative determination of drugs of | |
abuse in human urine. | Same (but the number of | |
drugs detected is different) | ||
Calibrator | Secobarbital (member of Barbiturates drug class) | Barbiturates |
Methodology | Competitive binding, lateral flow | |
immunochromatographic assays based on the | ||
principle of antigen antibody | ||
immunochemistry. | Same | |
Type of Test | Qualitative | Same |
Specimen Type | Human Urine | Same |
Cut-Off Values | 300 ng/mL | Same |
Intended Use | For over-the-counter and prescription uses. | For over-the-counter use. |
Configurations | Strip | Cup |
Table 1: Features Comparison of AssureTech Secobarbital Strip and the Predicate Devices
Table 2: Features Comparison of AssureTech Oxycodone Strip and the Predicate Devices
Item | Device | Predicate - K052115 |
---|---|---|
Indication(s) | ||
for Use | For the qualitative determination of | |
drugs of abuse in human urine. | Same (but the number of | |
drugs detected is different) | ||
Calibrator | Oxycodone | Same |
7
| Methodology | Competitive binding, lateral flow
immunochromatographic assays based on the
principle of antigen antibody
immunochemistry. | Same |
|----------------|-------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Type of Test | Qualitative | Same |
| Specimen Type | Human Urine | Same |
| Cut-Off Values | 100 ng/mL | Same |
| Intended Use | For over-the-counter and prescription uses. | For over-the-counter use. |
| Configurations | Strip | Cup |
Table 3: Features Comparison of AssureTech Secobarbital/Oxycodone Panel Dip and the Predicate Devices
Item | Device | Predicate - K052115 |
---|---|---|
Indication(s) | ||
for Use | For the qualitative determination of | |
drugs of abuse in human urine. | Same (but the number of | |
drugs detected is different) | ||
Calibrator | Secobarbital and Oxycodone | Same |
Methodology | Competitive binding, lateral flow | |
immunochromatographic assays based on the | ||
principle of antigen antibody | ||
immunochemistry. | Same | |
Type of Test | Qualitative | Same |
Specimen Type | Human Urine | Same |
Cut-Off Values | 300ng/mL for Secobarbital and 100 ng/mL for | |
Oxycodone | Same | |
Intended Use | For over-the-counter and prescription uses. | For over-the-counter use. |
Configurations | Panel Dip | Cup |
Table 4: Features Comparison of AssureTech Secobarbital/Oxycodone Cup and the Predicate Devices
Item | Device | Predicate - K052115 |
---|---|---|
Indication(s) | ||
for Use | For the qualitative determination of | |
drugs of abuse in human urine. | Same (but the number of | |
drugs detected is different) | ||
Calibrator | Secobarbital and Oxycodone | Same |
8
| Methodology | Competitive binding, lateral flow
immunochromatographic assays based on the
principle of antigen antibody
immunochemistry. | Same |
|----------------|-------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Type of Test | Qualitative | Same |
| Specimen Type | Human Urine | Same |
| Cut-Off Values | 300ng/mL for Secobarbital and 100 ng/mL for
Oxycodone | Same |
| Intended Use | For over-the-counter and prescription uses. | For over-the-counter use. |
| Configurations | Cup with or without turn-key | Cup |
9. Test Principle
AssureTech Secobarbital Strip, AssureTech Oxycodone Strip, AssureTech Secobarbital/Oxycodone Panel Dip, AssureTech Secobarbital/Oxycodone Quick Cup and AssureTech Secobarbial/Oxycodone Turn Key-Split Cup are rapid tests for the qualitative detection of Secobarbital and/or Oxycodone in urine samples. The tests are lateral flow chromatographic immunoassays. During testing, a urine specimen migrates upward by capillary action. If target drugs present in the urine specimen are below the cut-off concentration, it will not saturate the binding sites of its specific monoclonal mouse antibody coated on the particles. The antibodycoated particles will then be captured by immobilized drug-conjugate and a visible colored line will show up in the test line region. The colored line will not form in the test line region if the target drug level exceeds its cutoff-concentration because it will saturate all the binding sites of the antibody coated on the particles. A band should form in the control region of the devices regardless of the presence of drug or metabolite in the sample to indicate that the tests have been performed properly.
10. Performance Characteristics
1. Analytical Performance
- a. Precision
Precision studies were carried out for samples with concentrations of -100% cut off, -75% cut off, -50% cut off, -25% cut off, +25% cut off, +50% cut off , +75% cut off and +100% cut off. These samples were prepared by spiking drug in negative samples. Each drug concentration was confirmed by GC/MS. All sample aliquots were blindly labeled by the person who prepared the samples and didn't take part in the sample testing. For each concentration, tests were performed two runs per day for 25 days per device in a randomized order. The results obtained are summarized in the following tables.
9
Secobarbital
AssureTech Secobarbital Strip | ||
---|---|---|
------------------------------- | -- | -- |
Lot | -100% | -75% | -50% | -25% | cut off | +25% | +50% | +75% | +100% |
---|---|---|---|---|---|---|---|---|---|
Number | cut off | cut off | cut off | cutoff | cut off | cut off | cut off | cut off | |
Lot 1 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 2-/48+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
Lot 2 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 2-/48+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
Lot 3 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 2-/48+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
AssureTech Secobarbital/Oxycodone Panel Dip | |||||||||
Lot | -100% | -75% | -50% | -25% | cut off | +25% | +50% | +75% | +100% |
Number | cut off | cut off | cut off | cutoff | cut off | cut off | cut off | cut off | |
Lot 1 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 2-/48+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
Lot 2 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 2-/48+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
Lot 3 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 3-/47+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
AssureTech Secobarbital/Oxycodone Turn-Key Split Cup | |||||||||
Lot | -100% | -75% | -50% | -25% | cut off | +25% | +50% | +75% | +100% |
Number | cut off | cut off | cut off | cutoff | cut off | cut off | cut off | cut off | |
Lot 1 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 2-/48+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
Lot 2 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 3-/47+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
Lot 3 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 2-/48+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
AssureTech Secobarbital/Oxycodone Quick Cup | |||||||||
Lot | -100% | -75% | -50% | -25% | cut off | +25% | +50% | +75% | +100% |
Number | cut off | cut off | cut off | cutoff | cut off | cut off | cut off | cut off | |
Lot 1 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 2-/48+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
Lot 2 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 2-/48+ | 50+/0- | 50+/0- |
---|---|---|---|---|---|---|---|
Lot 3 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 1-/49+ | 50+/0- | 50+/0- |
Oxycodone
AssureTech Oxycodone Strip
| Lot
Number | -100%
cut off | -75%
cut off | -50%
cut off | -25%
cutoff | cut off | +25%
cut off | +50%
cut off | +75%
cut off | +100%
cut off |
|---------------------------------------------|------------------|-----------------|-----------------|----------------|---------|-----------------|-----------------|-----------------|------------------|
| Lot 1 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 2-/48+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 2 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 2-/48+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 3 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 1-/49+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| AssureTech Secobarbital/Oxycodone Panel Dip | | | | | | | | | |
| Lot
Number | -100%
cut off | -75%
cut off | -50%
cut off | -25%
cutoff | cut off | +25%
cut off | +50%
cut off | +75%
cut off | +100%
cut off |
|---------------|------------------|-----------------|-----------------|----------------|---------|-----------------|-----------------|-----------------|------------------|
| Lot 1 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 1-/49+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 2 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 2-/48+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 3 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 2-/48+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
AssureTech Secobarbital/Oxycodone Turn-Key Split Cup
Lot Number | -100% cut off | -75% cut off | -50% cut off | -25% cutoff | +25% cut off | +50% cut off | +75% cut off | +100% cut off | |
---|---|---|---|---|---|---|---|---|---|
Lot 1 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 3-/47+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
Lot 2 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 2-/48+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
Lot 3 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 1-/49+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
50+/0-
50+/0-
50+/0-
50+/0-
10
AssureTech Secobarbital/Oxycodone Quick Cup
| Lot
Number | -100%
cut off | -75%
cut off | -50%
cut off | -25%
cutoff | cut off | +25%
cut off | +50%
cut off | +75%
cut off | +100%
cut off |
|---------------|------------------|-----------------|-----------------|----------------|---------|-----------------|-----------------|-----------------|------------------|
| Lot 1 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 3-/47+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 2 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 2-/48+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 3 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 2-/48+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
b. Linearity
Not applicable.
c. Stability
The devices are stable at 4-30 ℃ for 24 months based on the accelerated stability study at 45 °C and real time stability determination at both 4 °C and 30 °C.
d. Cut-off
A total of 150 samples equally distributed at concentrations of -50% Cut-Off: -25% Cut-Off: Cut-Off: +25% Cut-Off: +50% Cut-Off were tested using three different lots of each device by three different operators. Results were all positive at and above +25% Cut-off and all negative at and below -25% Cut-off for both Secobarbital and Oxycodone.
The following cut-off values for the candidate devices have been verified.
Calibrator | Cut-off (ng/mL) |
---|---|
Secobarbital | 300 |
Oxycodone | 100 |
e. Interference
Potential interfering substances found in human urine of physiological or pathological conditions were added to drug-free urine and target drugs urine with concentrations at 25% below and 25% above Cut-Off levels. These urine samples were tested using three batches of each device. Compounds that showed no interference at a concentration of 100µg/mL are summarized in the following tables. There were no differences observed for different devices.
Secobarbital:
Acetaminophen | Erythromycin | O-Hydroxyhippuric acid |
---|---|---|
Acetophenetidin | β-Estradiol | D,L-Octopamine |
Acetylsalicylic acid | Estrone-3-sulfate | Oxalic acid |
Aminopyrine | Ethyl-p-aminobenzoate | Oxazepam |
Amitryptyline | Fenoprofen | Oxolinic acid |
Amoxicillin | Furosemide | Oxycodone |
DL-Amphetamine sulfate | Gentisic acid | Oxymetazoline |
Ampicillin | Hemoglobin | Papaverine |
Apomorphine | Hydralazine | Penicillin-G |
Ascorbic acid | Hydrochlorothiazide | Pentazocaine |
Aspartame | Hydrocodone | Perphenazine |
Atropine | Hydrocortisone | Phencyclidine |
Benzilic acid | p-Hydroxyamphetamine | Phenelzine |
7
11
Benzoic acid | p-Hydroxymethamphetamine | β-Phenylethlamine |
---|---|---|
Benzoylecgonine | 3-Hydroxytyramine | Phenylpropanolamine |
Bilirubin | Ibuprofen | Prednisolone |
Brompheniramine | Imipramine | Prednisone |
Caffeine | (-) Isoproterenol | Procaine |
Cannabidiol | Isoxsuprine | Promazine |
Cannabinol | Ketamine | Promethazine |
Chloralhydrate | Ketoprofen | D,L-Propanolol |
Chloramphenicol | Labetalol | D-Propoxyphene |
Chlorothiazide | Levorphanol | Quinidine |
(±) Chlorpheniramine | Loperamide | Quinine |
Chlorpromazine | L-Phenylephrine | Ranitidine |
Chlorquine | Maprotiline | Salicylic acid |
Cholesterol | Meperidine | Serotonin |
Clomipramine | Meprobamate | Sulfamethazine |
Clonidine | Morphine-3-β-D glucuronide | Sulindac |
Cocaine hydrochloride | Methadone | Temazepam |
Codeine | Methamphetamine | Tetracycline |
Cortisone | (±) - 3,4-Methylenedioxy- | |
amphetamine | Tetrahydrozoline | |
(-) Cotinine | (±)-3,4-Methylenedioxy- | |
methamphetamine | ||
hydrochloride | Thebaine | |
Creatinine | Morphine Sulfate | Thiamine |
Deoxycorticosterone | N-Acetylprocainamide | Thioridazine |
Dextromethorphan | Nalidixic acid | Triamterene |
Diazepam | Naloxone | Trifluoperazine |
Diclofenac | Naltrexone | Trimethoprim |
Diflunisal | Naproxen | Trimipramine |
Digoxin | Niacinamide | Tryptamine |
Diphenhydramine | Nifedipine | D, L-Tyrosine |
Doxylamine | Norcodein | Uric acid |
Ecgonine hydrochloride | Norethindrone | Verapamil |
Ecgonine methylester | D-Norpropoxyphene | Zomepirac |
(IR 2S)-Ephedrine | Noscapine |
Oxycodone
Acetophenetidin | L-Ephedrine | Oxymetazoline |
---|---|---|
Nalidixic acid | Ecgonine methylester | Papaverine |
Acetylsalicylic acid | Ethyl-p-aminobenzoate | Penicillin-G |
Aminopyrine | B-Estradiol | Perphenazine |
Amoxicillin | Estrone-3-sulfate | Phenelzine |
Ampicillin | Erythromycin | L-Phenylephrine |
L-Phenylephrine | Fenoprofen | B-Phenylethylamine |
Apomorphine | Furosemide | Phenylpropanolamine |
Aspartame | Gentisic acid | Prednisone |
Atropine | Hemoglobin | Loperamide |
12
Benzilic acid | Hydralazine | Quinine |
---|---|---|
Benzoic acid | Hydrochlorothiazide | Quinidine |
Benzphetamine | Hydrocortisone | Ranitidine |
Bilirubin | O-Hydroxyhippuric acid | Salicylic acid |
Deoxycorticosterone | 3-Hydroxytyramine | Serotonin |
Caffeine | Labetalol | Sulfamethazine |
Chloralhydrate | D, L-Isoproterenol | Sulindac |
Chloramphenicol | Meprobamate | Tetracycline |
Chlorothiazide | Methoxyphenamine | Tetrahydrocortisone |
D,L-Chlolrpheniramine | Nalidixic acid | Morphine-3-β-D-gluc |
Chlorpromazine | Naloxone | uronide |
Chlorquine | Naltrexone | Tetrahydrozoline |
Cholesterol | Naproxen | Thiamine |
Clonidine | Niacinamide | Thioridazine |
L-Cotinine | Nifedipine | D,L-Tyrosine |
Cortisone | Isoxsuprine | Tolbutamide |
Creatinine | D,L-Propanolol | Triamterene |
D-Pseudoephedrine | Ketoprofen | Trifluoperazine |
Dextromethorphan | Norethindrone | Trimethoprim |
ß-Dglucuronide | D-Norpropoxyphene | Tyramine |
Diclofenac | Noscapine | D,L-Tryptophan |
Diflunisal | D,L-Octopamine | Urine acid |
Digoxin | Oxalic acid | Verapamil |
Diphenhydramine | Oxolinic acid | Zomepirac |
f. Specificity
To test specificity, drug metabolites and other components that are likely to interfere in urine samples were tested using three batches of each device. The lowest concentration that caused a positive result for each compound are listed below. There were no differences observed for different devices.
Secobarbital | 0/0 | |
---|---|---|
(Cut-off=300 ng/mL) | Result | Cross-Reactivity |
Secobarbital | Positive at 300 ng/mL | 100% |
Amobarbital | Positive at 1000 ng/mL | 30% |
Alphenol | Positive at 62.5 ng/mL | 480% |
Aprobarbital | Positive at 250 ng/mL | 120% |
Butabarbital | Positive at 100 ng/mL | 300% |
Butathal | Positive at 500 ng/mL | 60% |
Butalbital | Positive at 5000ng/mL | 6% |
Cyclopentobarbital | Positive at 500 ng/mL | 60% |
Pentobarbital | Positive at 200 ng/mL | 150% |
Phenobarbital | Positive at 300 ng/mL | 100% |
| Oxycodone
(Cut-off=100 ng/mL) | Result | %
Cross-Reactivity |
|----------------------------------|--------------------------|-----------------------|
| Oxycodone | Positive at 100 ng/mL | 100% |
| Codeine | Positive at 100000 ng/mL | 0.1% |
| Acetylmorphine | Positive at 100000 ng/mL | 0.1% |
| Oxymorphone | Positive at 250 ng/mL | 40% |
13
Dihydrocodeine | Positive at 100000 ng/mL | 0.1% |
---|---|---|
Hydromorphone | Positive at 100000 ng/mL | 0.1% |
Hydrocodone | Positive at 3125 ng/mL | 3.2% |
Morphine | Positive at 100000 ng/mL | 0.1% |
Buprenorphine | Positive at 100000 ng/mL | 0.1% |
Ethylmorphine | Positive at 100000 ng/mL | 0.1% |
g. Effect of Urine Specific Gravity and Urine pH
To investigate the effect of urine specific gravity and urine pH, urine samples, with 1.000 to 1.035 specific gravity or urine samples with pH 4 to 9 were spiked with target drugs at 25% below and 25% above Cut-Off levels. These samples were tested using three batches of each device . Results were all positive for samples at and above +25% Cut-Off and all negative for samples at and below -25% Cut-Off. There were no differences observed for different devices.
2. Comparison Studies
Method comparison studies for the candidate devices were performed in-house with three laboratory assistants for each device. Operators ran 80 (40 negative and 40 positive) unaltered clinical samples. The samples were blind labeled and compared to GC/MS results. The results are presented in the tables below:
| Strip | | Negative | Low
Negative by
GC/MS
(less than
-50%) | Near Cutoff
Negative by
GC/MS
(Between
-50% and
cutoff) | Near Cutoff
Positive by
GC/MS
(Between the
cutoff and
+50%) | High Positive
by GC/MS
(greater than
+50%) |
|---------------------|----------|----------|----------------------------------------------------|------------------------------------------------------------------------|----------------------------------------------------------------------------|-----------------------------------------------------|
| Secobarbital | | | | | | |
| Viewer | Positive | 0 | 0 | 1 | 15 | 25 |
| A | Negative | 10 | 20 | 9 | 0 | 0 |
| Viewer | Positive | 0 | 0 | 1 | 15 | 25 |
| B | Negative | 10 | 20 | 9 | 0 | 0 |
| Viewer | Positive | 0 | 0 | 1 | 15 | 25 |
| C | Negative | 10 | 20 | 9 | 0 | 0 |
Discordant Results of Secobarbital Strip
Viewer | Sample Number | GC/MS Result | Strip Viewer Results |
---|---|---|---|
Viewer A | 817130 | 286 | Positive |
Viewer B | 817130 | 286 | Positive |
Viewer C | 817130 | 286 | Positive |
14
| Panel
Dip | | Negative | Low Negative by
GC/MS
(less than
-50%) | Near Cutoff Negative by
GC/MS
(Between
-50% and
cutoff) | Near Cutoff Positive by
GC/MS
(Between the
cutoff and
+50%) | High Positive
by GC/MS
(greater than
+50%) |
|--------------|----------|----------|-------------------------------------------------|---------------------------------------------------------------------|-------------------------------------------------------------------------|-----------------------------------------------------|
| Viewer
A | Positive | 0 | 0 | 1 | 14 | 25 |
| | Negative | 10 | 20 | 9 | 1 | 0 |
| Viewer
B | Positive | 0 | 0 | 1 | 15 | 25 |
| | Negative | 10 | 20 | 9 | 0 | 0 |
| Viewer
C | Positive | 0 | 0 | 1 | 15 | 25 |
| | Negative | 10 | 20 | 9 | 0 | 0 |
Discordant Results of Secobarbital Panel Dip
| Viewer | Sample Number | GC/MS Result | Panel Dip
Viewer Results |
|----------|---------------|--------------|-----------------------------|
| Viewer A | 817130 | 286 | Positive |
| Viewer B | 817130 | 286 | Positive |
| Viewer C | 817130 | 286 | Positive |
| Viewer A | 933321 | 306 | Negative |
| Turn-
Key
Split
Cup | | Negative | Low Negative by
GC/MS
(less than
-50%) | Near Cutoff
Negative by
GC/MS
(Between
-50% and
cutoff) | Near Cutoff
Positive by
GC/MS
(Between the
cutoff and
+50%) | High Positive
by GC/MS
(greater than
+50%) |
|------------------------------|----------|----------|-------------------------------------------------|------------------------------------------------------------------------|----------------------------------------------------------------------------|-----------------------------------------------------|
| Viewer
A | Positive | 0 | 0 | 1 | 14 | 25 |
| | Negative | 1 | 20 | 9 | 1 | 0 |
| Viewer
B | Positive | 0 | 0 | 0 | 14 | 25 |
| | Negative | 1 | 20 | 10 | 1 | 0 |
| Viewer
C | Positive | 0 | 0 | 1 | 14 | 25 |
| | Negative | 1 | 20 | 9 | 1 | 0 |
Discordant Results of Secobarbital Turn-Key Split Cup
| Viewer | Sample Number | GC/MS Result | Turn-Key Split
Cup |
|----------|---------------|--------------|-----------------------|
| Viewer A | 817130 | 286 | Positive |
| Viewer A | 862382 | 310 | Negative |
| Viewer C | 817130 | 286 | Positive |
| Viewer B | 933321 | 306 | Negative |
| Viewer C | 722825 | 336 | Negative |
15
| Quick
Cup | | Negative | Low Negative by
GC/MS
(less than
-50%) | Near Cutoff Negative by
GC/MS
(Between
-50% and
cutoff) | Near Cutoff Positive by
GC/MS
(Between the
cutoff and
+50%) | High Positive
by GC/MS
(greater than
+50%) |
|--------------|----------|----------|-------------------------------------------------|---------------------------------------------------------------------|-------------------------------------------------------------------------|-----------------------------------------------------|
| Viewer
A | Positive | 0 | 0 | 1 | 14 | 25 |
| | Negative | 10 | 20 | 9 | 1 | 0 |
| Viewer
B | Positive | 0 | 0 | 1 | 15 | 25 |
| | Negative | 10 | 20 | 9 | 0 | 0 |
| Viewer
C | Positive | 0 | 0 | 1 | 15 | 25 |
| | Negative | 10 | 20 | 9 | 0 | 0 |
Discordant Results of Secobarbital Quick Cup
Viewer | Sample Number | GC/MS Result | Quick Cup Viewer Results |
---|---|---|---|
Viewer A | 216041 | 241 | Positive |
Viewer B | 126635 | 252 | Positive |
Viewer C | 126635 | 252 | Positive |
Viewer A | 358232 | 315 | Negative |
Oxycodone (OXY)
| Strip | | Negative | Low
Negative by
GC/MS
(less than
-50%) | Near Cutoff
Negative by
GC/MS
(Between
-50% and
cutoff) | Near Cutoff
Positive by
GC/MS
(Between the
cutoff and
+50%) | High Positive
by GC/MS
(greater than
+50%) |
|-------------|----------|----------|----------------------------------------------------|------------------------------------------------------------------------|----------------------------------------------------------------------------|-----------------------------------------------------|
| Viewer
A | Positive | 0 | 0 | 1 | 14 | 25 |
| Viewer
A | Negative | 10 | 20 | 9 | 1 | 0 |
| Viewer
B | Positive | 0 | 0 | 1 | 15 | 25 |
| Viewer
B | Negative | 10 | 20 | 9 | 0 | 0 |
| Viewer
C | Positive | 0 | 0 | 1 | 15 | 25 |
| Viewer
C | Negative | 10 | 20 | 9 | 0 | 0 |
Discordant Results of OXY Strip
Viewer | Sample Number | GC/MS Result | Strip Viewer Results |
---|---|---|---|
Viewer A | 638805 | 96 | Positive |
Viewer A | 757887 | 102 | Negative |
Viewer B | 638805 | 96 | Positive |
Viewer C | 638805 | 96 | Positive |
16
| Panel
Dip | | Negative | Low Negative by
GC/MS
(less than -50%) | Near Cutoff Negative by
GC/MS
(Between -50% and
cutoff) | Near Cutoff Positive by
GC/MS
(Between the
cutoff and +50%) | High Positive
by GC/MS
(greater than
+50%) |
|--------------|----------|----------|----------------------------------------------|------------------------------------------------------------------|----------------------------------------------------------------------|-----------------------------------------------------|
| Viewer
A | Positive | 0 | 0 | 1 | 15 | 25 |
| | Negative | 10 | 20 | 9 | 0 | 0 |
| Viewer
B | Positive | 0 | 0 | 0 | 14 | 25 |
| | Negative | 10 | 20 | 10 | 1 | 0 |
| Viewer
C | Positive | 0 | 0 | 1 | 15 | 25 |
| | Negative | 10 | 20 | 9 | 0 | 0 |
Discordant Results of OXY Panel Dip
| Viewer | Sample Number | GC/MS Result | Panel Dip
Viewer Results |
|----------|---------------|--------------|-----------------------------|
| Viewer A | 603216 | 90 | Positive |
| Viewer C | 638805 | 96 | Positive |
| Viewer B | 757887 | 102 | Negative |
| Turn-Key
Split Cup | | Negative | Low Negative by
GC/MS
(less than -50%) | Near Cutoff Negative by
GC/MS
(Between -50% and
cutoff) | Near Cutoff Positive by
GC/MS
(Between the cutoff and
+50%) | High Positive
by GC/MS
(greater than +50%) |
|-----------------------|----------|----------|----------------------------------------------|------------------------------------------------------------------|----------------------------------------------------------------------|--------------------------------------------------|
| Viewer
A | Positive | 0 | 0 | 1 | 14 | 25 |
| | Negative | 1 | 20 | 9 | 1 | 0 |
| Viewer
B | Positive | 0 | 0 | 1 | 15 | 25 |
| | Negative | 1 | 20 | 9 | 0 | 0 |
| Viewer
C | Positive | 0 | 0 | 0 | 15 | 25 |
| | Negative | 1 | 20 | 10 | 0 | 0 |
Discordant Results of OXY Turn-Key Split Cup
Viewer | Sample Number | GC/MS Result | Split Cup Viewer Results |
---|---|---|---|
Viewer A | 638805 | 96 | Positive |
Viewer A | 285248 | 110 | Negative |
Viewer B | 638805 | 96 | Positive |
17
| Quick
Cup | | Negative | Low Negative by
GC/MS
(less than
-50%) | Near Cutoff
Negative by
GC/MS
(Between
-50% and
cutoff) | Near Cutoff
Positive by
GC/MS
(Between the
cutoff and
+50%) | High Positive
by GC/MS
(greater than
+50%) |
|--------------|----------|----------|-------------------------------------------------|------------------------------------------------------------------------|----------------------------------------------------------------------------|-----------------------------------------------------|
| Viewer
A | Positive | 0 | 0 | 1 | 14 | 25 |
| | Negative | 10 | 20 | 9 | 1 | 0 |
| Viewer
B | Positive | 0 | 0 | 0 | 14 | 25 |
| | Negative | 10 | 20 | 10 | 1 | 0 |
| Viewer
C | Positive | 0 | 0 | 1 | 14 | 25 |
| | Negative | 10 | 20 | 9 | 1 | 0 |
Discordant Results of OXY Quick Cup
| Viewer | Sample Number | GC/MS Result | Quick Cup
Viewer Results |
|----------|---------------|--------------|-----------------------------|
| Viewer A | 638805 | 96 | Positive |
| Viewer C | 603216 | 90 | Positive |
| Viewer A | 757887 | 102 | Negative |
| Viewer B | 508044 | 116 | Negative |
| Viewer C | 285248 | 110 | Negative |
Lay-user study
A lay user study was performed at three intended user sites with 1060 lay persons. The lay users had diverse educational and professional backgrounds and ranged in age from 18 to > 50 years. Urine samples were prepared at the following concentrations; negative, +/-75%, +/-50%, +/-25% of the cutoff by spiking drug(s) into drug free-pooled urine specimens. The concentrations of the samples were confirmed by GC/MS. Each sample was aliquoted into individual containers and blind-labeled. Each participant was provided with the package insert, 1 blind labeled sample and a device. Each device was tested.
Comparison between GC/MS and Lay Person Results for Secobarbital Strip
| % of Cutoff | Number of samples | Secobarbital
Concentration by GC/MS
(ng/mL) | Lay person results | | The
percentage of
correct results
(%) |
|--------------|-------------------|---------------------------------------------------|--------------------|--------------------|------------------------------------------------|
| -100% Cutoff | 20 | 0 | No. of
Positive | No. of
Negative | 100 |
| -75% Cutoff | 20 | 73 | 0 | 20 | 100 |
| -50% Cutoff | 20 | 149 | 0 | 20 | 100 |
| -25% Cutoff | 20 | 228 | 2 | 18 | 90 |
| +25% Cutoff | 20 | 369 | 20 | 0 | 100 |
| +50% Cutoff | 20 | 455 | 20 | 0 | 100 |
| +75% Cutoff | 20 | 530 | 20 | 0 | 100 |
18
Number | Oxycodone Concentration | Lay person results | The | ||
---|---|---|---|---|---|
% of Cutoff | of | ||||
samples | by GC/MS (ng/mL) | No. of | |||
Positive | No. of | ||||
Negative | percentage of | ||||
correct results | |||||
(%) | |||||
-100% Cutoff | 20 | 0 | 0 | 20 | 100 |
-75% Cutoff | 20 | 22 | 0 | 20 | 100 |
-50% Cutoff | 20 | 48 | 0 | 20 | 100 |
-25% Cutoff | 20 | 80 | 1 | 19 | 95 |
+25% Cutoff | 20 | 127 | 19 | 1 | 95 |
+50% Cutoff | 20 | 158 | 20 | 0 | 100 |
+75% Cutoff | 20 | 176 | 20 | 0 | 100 |
Comparison between GC/MS and Lay Person Results for OXY Strip
Comparison between GC/MS and Lay Person Results for BAR/OXY Panel DipCard
| % Cutoff | No of
samples | Concentration by
GC/MS(ng/mL) | | Lay person
results | | Correct Results
(%) | |
|----------|------------------|----------------------------------|-----------|-----------------------|--------|------------------------|-----|
| | | Secobarbital | Oxycodone | BAR | OXY | BAR | OXY |
| -100% | 20 | 0 | 0 | 0+/20- | 0+/20- | 100 | 100 |
| -75% | 20 | 73 | 22 | 0+/20- | 0+/20- | 100 | 100 |
| -50% | 20 | 149 | 48 | 0+/20- | 0+/20- | 100 | 100 |
| -25% | 20 | 228 | 80 | 1+/19- | 1+/19- | 95 | 95 |
| +25% | 20 | 369 | 127 | 19+/1- | 18+/2- | 95 | 90 |
| +50% | 20 | 455 | 158 | 20+/0- | 20+/0- | 100 | 100 |
| +75% | 20 | 530 | 176 | 20+/0- | 20+/0- | 100 | 100 |
Comparison between GC/MS and Lay Person Results for BAR/OXY Turn-Key Split Cup
| % Cutoff | No of
samples | Concentration by
GC/MS(ng/mL) | | Lay person
results | | Correct Results
(%) | |
|----------|------------------|----------------------------------|-----------|-----------------------|--------|------------------------|-----|
| | | Secobarbital | Oxycodone | BAR | OXY | BAR | OXY |
| -100% | 20 | 0 | 0 | 0+/20- | 0+/20- | 100 | 100 |
| -75% | 20 | 73 | 22 | 0+/20- | 0+/20- | 100 | 100 |
| -50% | 20 | 149 | 48 | 0+/20- | 0+/20- | 100 | 100 |
| -25% | 20 | 228 | 80 | 2+/18- | 1+/19- | 90 | 95 |
| +25% | 20 | 369 | 127 | 20+/0- | 19+/1- | 100 | 95 |
| +50% | 20 | 455 | 158 | 20+/0- | 20+/0- | 100 | 100 |
| +75% | 20 | 530 | 176 | 20+/0- | 20+/0- | 100 | 100 |
Comparison between GC/MS and Lay Person Results for BAR/OXY Quick Cup
| % Cutoff | No of
samples | Concentration by
GC/MS(ng/mL) | | Lay person results | | Correct Results
(%) | |
|----------|------------------|----------------------------------|-----------|--------------------|--------|------------------------|-----|
| | | Secobarbital | Oxycodone | BAR | OXY | BAR | OXY |
| -100% | 20 | 0 | 0 | 0+/20- | 0+/20- | 100 | 100 |
| -75% | 20 | 73 | 22 | 0+/20- | 0+/20- | 100 | 100 |
| -50% | 20 | 149 | 48 | 0+/20- | 0+/20- | 100 | 100 |
| -25% | 20 | 228 | 80 | 1+/19- | 1+/19- | 95 | 95 |
19
+25% | 20 | ૩૯તે રાજ્યના ઉત્તર પ્રદેશના પાકની ખેતી કરવવામાં આવે છે. આ ગામમાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસ | 127 | 20+/0- | 20+/0- | 100 | 100 |
---|---|---|---|---|---|---|---|
+50% | 20 | 455 | 158 | 20+/0- | 20+/0- | 100 | 100 |
+75% | 20 | 530 | 176 | 20+/0- | 20+/0- | 100 | 100 |
Lay-users were also given surveys on the ease of understanding the package insert instructions. All lay users indicated that the device instructions can be easily followed. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.
3. Clinical Studies
Not applicable.
11. Conclusion
Based on the test principle and acceptable performance characteristics including precision, cut-off, interference, specificity, method comparison, and lay-user studies of the devices, it's concluded that Secobarbital Strip, AssureTech Oxycodone the AssureTech Strip, AssureTech Secobarbital/Oxycodone Panel Dip, AssureTech Secobarbital/Oxycodone Quick Cup and AssureTech Secobarbial/Oxycodone Turn Key-Split Cup are substantially equivalent to the predicate.