K Number
K151169

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-10-08

(160 days)

Product Code
Regulation Number
872.3630
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

OsteoReady ® ZircoSeal® Dental Abutments are intended to be used with an endosseous implant to aid prosthetic rehabilitation in order to restore the patient's chewing function. OsteoReady® ZircoSeal® Abutments are compatible with DENTIN® Dental Implant System - Prestige Implant System diameters 3.75, 4.2, 5 & 6;

Device Description

OsteoReady® ZircoSeal® Abutments are Straight Anti-Rotation Abutments made of titanium alloy. OsteoReady® ZircoSeal® Abutments are available in the following measurements:

Slim Anti-Rotation Titanium Abutment with ZircoSeal®: Length 5,7,9,11,13 mm with Diameter 3.75 mm.

Anti-Rotation Titanium Abutment with ZircoSeal®: Length 5,7,9,11,13 mm Diameter 4.5 mm.

Anti-Rotation Straight Titanium Abutment with ZircoSeal®: Shoulder Length 1,2,3,4 mm Diameter 4.5 mm.

OsteoReady® ZiroSeal® Abutments are straight only and are not intended to be modified chairside or by CAD/CAM to provide an angle correction.

The ZircoSeal surface treatment is the trade name of zirconium nitride.

The ZircoSeal® simply indicates the signature zirconium nitride surface treatment of the abutment. The trade name is unrelated to claims of "sealing" of implant/abutment connection platform.

AI/ML Overview

This document is a 510(k) premarket notification for the OsteoReady® ZircoSeal® Abutments, asserting substantial equivalence to previously marketed devices. It does not describe a study to prove a device meets specific acceptance criteria in the way one might expect for a diagnostic or AI-powered device.

Instead, the "acceptance criteria" here refer to meeting the requirements for demonstrating substantial equivalence to predicate devices, primarily through performance testing, biocompatibility testing, and comparison of device characteristics.

Here's an analysis based on the provided text, addressing your questions where applicable:


1. Table of Acceptance Criteria and Reported Device Performance

The document does not present a formal table of "acceptance criteria" paired with "reported device performance" in the context of clinical accuracy or performance metrics for an AI/diagnostic device. Instead, it lists the types of tests conducted and their successful outcomes as proof of equivalence and safety.

Test TypeAcceptance Criteria (Implied)Reported Device Performance
Sterilization ValidationAchieve a Safety Assurance Level (SAL) of 10⁻⁶Met all applicable criteria; SAL of 10⁻⁶ was achieved.
Biocompatibility TestingMeet all applicable criteria as per ISO 10993-5 (cytotoxicity)Met all applicable criteria.
Surface Treatment Eval.(Implied) Consistency with predicate device characteristicsChemical analysis and abrasion characteristics evaluated (details not provided beyond being "similar").
Material Composition(Implied) Same or equivalent to predicate devices (GR-5 Titanium)GR-5 Titanium Ti-6Al-4V ELI (consistent with most predicates).
Design Characteristics(Implied) Similar to predicate devicesStraight Anti-Rotation Abutments, various lengths/diameters, Internal Hex connection (similar to predicates).
Indications for Use(Implied) Substantially equivalent to predicate devicesMinor verbiage differences, but determined not to change intended use.
Risk AssessmentConducted in accordance with ISO 14971:2012Determined that differences raise no new safety or effectiveness issues.

2. Sample Size Used for the Test Set and Data Provenance

This document does not describe a test set or data provenance in the context of a clinical performance study (e.g., for AI/diagnostic algorithms). The "tests" mentioned are non-clinical, such as sterilization, biocompatibility, and material characterization. Therefore, information about "sample size for a test set" or "data provenance" (country of origin, retrospective/prospective) is not applicable or provided.

3. Number of Experts Used to Establish Ground Truth and Qualifications

Not applicable. This device is a dental abutment, not an AI or diagnostic device requiring expert interpretation for ground truth.

4. Adjudication Method

Not applicable. There is no clinical study involving human interpretation or ground truth adjudication in this submission.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. This document does not mention an MRMC study. The device is a physical dental abutment, not a diagnostic tool requiring human-AI comparative effectiveness evaluation.

6. Standalone Performance Study

Yes, in a non-clinical sense. The testing performed (sterilization, biocompatibility, surface treatment evaluation) assessed the standalone performance of the OsteoReady® ZircoSeal® Abutments against established standards (ISO, AAMI) and similarity to predicate devices. However, this is not a "standalone algorithm performance" study as would be relevant for an AI device.

7. Type of Ground Truth Used

The concept of "ground truth" as typically applied to diagnostic or AI studies is not directly relevant here. The "truth" in this context is established by:

  • Standards compliance: E.g., ISO 10993-5 for biocompatibility, ISO 17665-1/2 and ANSI/AAMI ST79:2010 for sterilization.
  • Material specifications: GR-5 Titanium Ti-6Al-4V ELI.
  • Manufacturing specifications: Device design features (diameters, lengths, connection type).
  • Chemical analysis: For surface treatment evaluation.

8. Sample Size for the Training Set

Not applicable. This is not an AI/machine learning device; therefore, there is no "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 8, 2015

OsteoReady LLC c/o Ms. Daniela Levy Regulatory Consultant Sterling Medical Registration 22817 Ventura Blvd Woodland Hills, California 91364

Re: K151169

Trade/Device Name: OsteoReady® ZircoSeal® Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: August 26, 2015 Received: September 9, 2015

Dear Ms. Levy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tina
Kiang -S

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image contains the logo for OsteoReady, a company that provides practical implant solutions. The logo consists of a stylized blue icon on the left, resembling concentric waves forming a shape. To the right of the icon is the company name, "OsteoReady," in a bold, dark blue font. Below the company name is the tagline "Practical Implant Solutions" in a smaller, lighter blue font.

Page 1 of

SECTION 4 -Indication for Use Statement
510(k) Number (if known):
K151169
Device Name:
OsteoReady® ZircoSeal® Abutments
Indications for Use ( Describe )
Indications for Use:
OsteoReady ® ZircoSeal® Dental Abutments are intended to be used with an endosseousimplant to aid prosthetic rehabilitation in order to restore the patient's chewing function.
OsteoReady® ZircoSeal® Abutments are compatible with DENTIN® Dental Implant System- Prestige Implant System diameters 3.75, 4.2, 5 & 6;

Type of Use (Select one or both, as applicable)

区Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use(21 CFR 801 Subpart

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image displays the logo for OsteoReady, a company specializing in practical implant solutions. Below the logo, the text "SECTION 5 - 510(k) Summary (21 CFR 807.92)" is present. This suggests the image is part of a document related to regulatory compliance, specifically a 510(k) summary as required by the Code of Federal Regulations for medical devices.

510(k) Number K151169

1SubmissionOsteoReady LLC
OwnerMichael D. Silverman, DMD, DICOI
106 Lenora Street
Seattle, WA 98121
USA
Phone : 206-812-7727
Fax : 800-721-4907
2Official CorrespondentSterling Medical Registration
Contact PersonDaniela Levy - Regulatory Consultant
22817 Ventura blvd.
Woodland Hills, CA 91364
Phone: 1-213-787-3027
Email: daniela@sterlingmedicalregistration.com
3Date PreparedOctober 05, 2015
4Device Trade NameOsteoReady® ZircoSeal® Abutments
5Regulation DescriptionEndosseous Dental Implant Abutment
6ClassificationDevice Name : Abutment,Implant,Dental,Endosseous
Product Code:NHA
Regulation No:872.3630
Class:II
Panel:Dental
7Reason for the Premarket Notification Submission:New Device
8Identification of Legally Marketed Predicate Devices :
OsteoReady® ZircoSeal® Abutments are substantially equivalent to the following:

Primary Predicate

  • PerioType X-Pert (K123386); .
    Reference Predicate

  • . AB Dental Devices Implants and Accessories (K112440)

  • A.B. Dental Devices (K051719) .

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for OsteoReady, a company that provides practical implant solutions. The logo consists of a stylized blue icon on the left, resembling a series of concentric arcs. To the right of the icon is the company name, "OsteoReady," in a bold, blue font. Below the company name is the tagline "Practical Implant Solutions" in a smaller, lighter blue font.

  • IQ Implant System(K130829)
    in terms of intended use, indication for use, raw material, technological characteristics and performance.

The predicate devices are a Class II medical device.

  • တ Device Description:
    OsteoReady® ZircoSeal® Abutments are Straight Anti-Rotation Abutments made of titanium alloy. OsteoReady® ZircoSeal® Abutments are available in the following measurements:

Slim Anti-Rotation Titanium Abutment with ZircoSeal®: Length 5,7,9,11,13 mm with Diameter 3.75 mm.

Anti-Rotation Titanium Abutment with ZircoSeal®: Length 5,7,9,11,13 mm Diameter 4.5 mm.

Anti-Rotation Straight Titanium Abutment with ZircoSeal®: Shoulder Length 1,2,3,4 mm Diameter 4.5 mm.

OsteoReady® ZiroSeal® Abutments are straight only and are not intended to be modified chairside or by CAD/CAM to provide an angle correction.

The ZircoSeal surface treatment is the trade name of zirconium nitride.

The ZircoSeal® simply indicates the signature zirconium nitride surface treatment of the abutment. The trade name is unrelated to claims of "sealing" of implant/abutment connection platform.

  • 10 Indication for Use:
    OsteoReady ® ZircoSeal® Dental Abutments are intended to be used with an endosseous implant to aid prosthetic rehabilitation in order to restore the patient's chewing function. OsteoReady® ZircoSeal® Abutments are compatible with DENTIN® Dental Implant System - Prestige Implant System diameters 3.75, 4.2, 5 & 6;

  • 11 Performance Testing and Special Controls
    Steam validation test was conducted in order to ensure safety assurance level of ( SAL) 10-6 related to OsteoReady® ZircoSeal® Abutments system.

Biocompatibility test was conducted according to ISO 10993-5, test results have met all applicable criteria. Additional biocompatibility justification was provided attesting that the raw material and surface treatments of the subject abutment is the same as the predicates.

Clinical Testing - No clinical data is included in this submission.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for OsteoReady, a company that provides practical implant solutions. The logo consists of a stylized blue icon on the left, resembling a tooth or bone structure. To the right of the icon is the company name "OsteoReady" in a bold, blue font. Below the company name, in a smaller font, is the tagline "Practical Implant Solutions."

  • ISO 7405 Second edition 2008-12-15 Dentistry Evaluation of biocompatibility . of medical devices used in dentistry.
  • ISO 5832-3:1996 Implants for surgery -- Metallic materials -- Part 3: . Wrought titanium 6-aluminium 4-vanadium alloy.
    • Sterilization validation according to to ISO 17665-1/2 and ANSI/AAMI . ST79:2010. Test results have demonstrated that the SAL of 10-6 was achieved and all testing requirements were met.
  • . Surface treatment evaluation per the FDA guidance document for Endosseous Dental Implants and Abutments including chemical analysis and abrasion characteristics of the surface.

FDA guidance document: Class II Special Controls Guidance Document: Rootform Endosseous Dental Implants and Endosseous Dental Abutments - Guidance for Industry and FDA Staff.

SubstantialEquivalent TableCHARACTERISTICCandidateSlim Anti RotationAbutment with ZircoSeal ®REFERENCE PREDICATEDEVICEP3 Anti rotation abutmentREFERENCEPREDICATE DEVICESlim Straight Abutment
510kK112440, K051719K130829
Company NameOsteoReady LLCA.B.Dental Devices Itd.IQ Implants Itd
Device DesignEndosseous dental implantabutmentEndosseous dental implantabutmentEndosseous dental implantabutment
ClassificationClass II 872.3630P.Code NHAClass II 872.3630P.Code NHAClass II 872.3630P.Code NHA
MaterialGR-5 Titanium Ti-6Al-4V ELIGR-5 Titanium Ti-6Al-4V ELIGR-5 Titanium Ti-6Al-4V ELI
Diameters mm3.753.753.75
Lengthsmm5, 7, 9, 11, 135, 7, 9, 12, 155, 7, 9, 11
Implant / AbutmentconnectionInternal HexInternal HexInternal Hex
SurfaceZircoSeal[Surface is similar to PerioTypeX-pert K123386]CNC MachinedCNC Machined
Anatomical SiteOral CavityOral CavityOral Cavity

12 Substantial Equivalence

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for OsteoReady, a company that provides practical implant solutions. The logo consists of a stylized blue symbol on the left, resembling a tooth or bone structure. To the right of the symbol is the company name, "OsteoReady," in a bold, blue font. Below the name is the tagline "Practical Implant Solutions" in a smaller, lighter font.

Indication for UseOsteoReady ® ZircoSeal® Dental Abutments are intended to be used with an endosseous implant to aid prosthetic rehabilitation in order to restore the patient's chewing function. OsteoReady® ZircoSeal® Abutments are compatible with DENTIN® Dental Implant System - Prestige Implant System diameters 3.75, 4.2, 5 & 6.The AB Dental Devices implants are intended for surgical placement in the maxillary mandibular and/or arch to support crowns, bridges, or overdentures in edlentulous or partially edentulous patients.IQ Implant System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. IQ Implant System is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
SterilizationProvided non sterile prior to their use sterilize in an autoclave.Provided non sterile prior to their use sterilize in an autoclave.Provided non sterile prior to their use sterilize in an autoclave.
UseSingle useSingle useSingle use
Substantial EquivalentTableCHARACTERISTICCandidateAnti Rotation Abutment withZircoSeal ®REFERENCE PREDICATE DEVICEP3* Abutment Anti rotation
510kK112440, K051719
Company NameOsteoReady LLCA.B.Dental Devices Itd.
Device DesignEndosseous dental implant abutmentEndosseous dental implant abutment
ClassificationClass II 872.3630P.Code NHAClass II 872.3630P.Code NHA
MaterialGR-5 Titanium Ti-6Al-4V ELIGR-5 Titanium Ti-6Al-4V ELI
Diameters mm4.54.5
Lengthsmm5, 7, 9, 11, 135, 7, 9, 12, 15
Implant / Abutment connectionInternal HexInternal Hex
SurfaceZircoSeal [Surface is similar to PerioType X-pert K123386]CNC Machined Surface
Anatomical SiteOral CavityOral Cavity
Indication for UseOsteoReady ® ZircoSeal® DentalAbutments are intended to be used withan endosseous implant to aid prostheticrehabilitation in order to restore thepatient'schewingfunction.OsteoReady® ZircoSeal® Abutments arecompatible with DENTIN® Dental ImplantSystem - Prestige Implant Systemdiameters 3.75, 4.2, 5 & 6.The AB Dental Devices implantsareintended for surgical placement in themaxillary mandibular and/or arch tosupport crowns, bridges, or overdlenturesin edlentulous or partially edentulouspatients.
SterilizationProvided non sterile prior to their usesterilize in an autoclave.Provided non sterile prior to their usesterilize in an autoclave.
UseSingle useSingle use

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for OsteoReady. The logo features a stylized blue icon on the left, resembling a series of concentric arcs. To the right of the icon is the company name, "OsteoReady," in a bold, blue font. Below the name is the tagline "Practical Implant Solutions" in a smaller, lighter font.

Substantial Equivalent TableCHARACTERISTICCandidateAnti Rotation Straight Abutment withZircoSeal ®PRIMARY PREDICATE DEVICEPerioType X-pert, ZircoSeal® AbutmentsNAS
510kK123386
Company NameOsteoReady LLCClinical House Europe Gmbh
Device DesignEndosseous dental implant abutmentEndosseous dental implant abutment
ClassificationClass II 872.3630P.Code NHAClass II 872.3630P.Code NHA
MaterialGR-5 Titanium Ti-6Al-4V ELITitanium Grade 4
Diameters mm4.54.2, 5.5
LengthsmmCuff height 1, 2, 3, 4Cuff height 1, 2, 3, 4
Implant / Abutment connectionInternal HexInternal Hex
SurfaceZircoSealZircoSeal
Anatomical SiteOral CavityOral Cavity
Indication for UseOsteoReady ® ZircoSeal® DentalAbutments are intended to be used withan endosseous implant to aid prostheticrehabilitation in order to restore thepatient's chewing function.OsteoReady® ZircoSeal® Abutmentsare compatible with DENTIN® DentalImplant System - Prestige ImplantSystem diameters 3.75, 4.2, 5 & 6.The PerioType X-Pert Abutments arestraight and angled dental implantabutments to be used in conjunction withthe PerioType X-Pert dental implant fixtureto aid in prosthetic rehabilitation.
SterilizationProvided non sterile prior to their usesterilize in an autoclave.Provided non sterile prior to their usesterilize in an autoclave.
UseSingle useSingle use

Summary of Equivalence:

OsteoReady® ZircoSeal® Abutment System shares similarity to A.B.Dental Devices Itd. K112440, K051719; IQ Implants K130829; PerioType X-Pert K123386 in terms of intended use, indication for use, technological characteristics and performance.

The differences are that A.B.Dental Devices Itd. K112440, K051719; IQ Implants K130829; use CNC machined surface to their abutments while OsteoReady uses the ZircoSeal® surface treatment which is identical surface to its other predicate device the PerioType X-Pert K123386. In addition, PerioType X-Pert K123386 uses titanium grade 4 while OsteoReady uses titanium alloy as in similar to its other predicate devices A.B.Dental Devices Itd. K112440, K051719; IQ Implants K130829; The verbiage of the Indications for Use of the subject device is slightly different than that of the declared predicates; however, these slight differences in wording does not

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains the logo for OsteoReady, a company that provides practical implant solutions. The logo consists of a blue circular graphic on the left, resembling a stylized wave or ripple effect. To the right of the graphic is the company name "OsteoReady" in a bold, blue font. Below the name, in a smaller font, is the tagline "Practical Implant Solutions."

change the intended use of the subject device has compared to the declared predicate. As demonstrated by the substantial equivalent table, the differences raise no new safety or/and effectiveness issues, thus OsteoReady® ZircoSeal® Abutments System considered to be substantial equivalent to its predicate devices.

Risk Assessment was conducted with accordance to ISO 14971:2012 Medical devices -Application of risk management to medical devices Class II Special Controls Guidance Document.

Therefore, OsteoReady® ZircoSeal® Abutments system is considered to be substantially equivalent to its predicate devices.

Conclusion:

As verified by substantial equivalence, risk assessment and bench testing OsteoReady® ZircoSeal® Abutment System shares similarity with its predicate devices in terms of intended use, indication for use, raw material, technological characteristics and performance.

Thus OsteoReady® ZircoSeal® Abutment System is considered to be substantially equivalent to its predicate devices.

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)