(267 days)
Sonishield™ 100 Antimicrobial Ultrasound Gel is intended for general use as a non-sterile transmission media for acoustically coupling a transducer to a human body surface during external and therapeutic ultrasound imaging procedures. It is placed on the patient's skin prior to initiating an ultrasound examination. It is indicated for prescription use only.
Sonishield™ 100 Antimicrobial Ultrasound Gel consists of deionized water, hydroxyethylcellulose, propanediol and benzalkonium chloride. It is a type of conductive medium (i.e.) scanning gel used in ultrasound therapeutic and diagnostic imaging techniques. A scanning gel acts as a couplant that provides an acoustic pathway between the transducer and the skin. In addition, the gel eliminates air (a disruptive influence) from the interface and adapts the contours of the probe to the skin.
The major characteristics of Sonishield™ 100 Antimicrobial Ultrasound Gel include:
- Hypoallergenic, non-irritating
- Water soluble, non-staining and easily cleanable
- . Does not contain oil or fatty matter
- Free from formaldehyde and salt ●
- No toxic effects
- Produced as a completely harmless material
- Has no odor
- Vacuum treated production
- Does not damage the probe
- Does not contain air bubbles
- pH level ranges from 4.5 to 6.5 at 25°C ●
The document describes the Sonishield™ 100 Antimicrobial Ultrasound Gel and its substantial equivalence to predicate devices, rather than a study proving the device meets specific acceptance criteria based on performance in a diagnostic task.
However, from the provided text, we can infer some "acceptance criteria" through the non-clinical performance tests and the comparison to predicate devices. The study detailed is primarily a non-clinical performance evaluation to demonstrate safety and functional equivalence, not a clinical trial evaluating diagnostic accuracy with human interpretation.
Here's the breakdown of information based on your request, extracting what's available from the provided text:
1. Table of Acceptance Criteria (Inferred from Non-Clinical Performance) and Reported Device Performance
| Acceptance Criteria (Inferred) | Reported Device Performance (Sonishield™ 100 Antimicrobial Ultrasound Gel) |
|---|---|
| Acoustic Properties: | |
| Sound velocity (m/sec) at 30°C | 1497 |
| Density (kg/m³) at 30°C | 1.023 X 10⁻³ |
| Acoustic impedance (kg/m² sec) at 30°C | 1.53 |
| Attenuation coefficient as a function of frequency, a/f (dB/cm-MHz) | 0.04 ± 0.0042 |
| Comparison to predicate gels | Virtually identical to predicate Konix® Ultrasound Gel (Konix® values: 1516 m/sec, 0.98X 10⁻³ kg/m³, 1.49 kg/m² sec, <0.05 f dB/cm-MHz) |
| Antimicrobial Effectiveness: | |
| Log reduction > 4 from initial count at 14 days post-inoculation (for specified bacteria) | Achieved log reduction > 4 for S. Aureus, Pseudomonas aeruginosa, E.coli, S.Aureus (MRSA), and K. pneumoniae at 14 days and 28 days. |
| Biocompatibility - Skin Irritation (ISO 10993-10): | Non-irritating |
| Biocompatibility - Skin Sensitization (ISO 10993-10): | Non-sensitizing |
| Clinical Tolerance - Cutaneous Irritation: | No signs of cutaneous irritation |
| Clinical Hypoallergenic (Human Repeat Insult Patch Test - HRIPT): | Considered hypoallergenic; no pertinent reaction to test product during challenge phase. (Note: 9/55, 1/55, 2/55 volunteers showed 'barely noticeable' to Grade 2 erythema during induction phase, but this did not lead to a 'pertinent reaction' in the challenge phase for the overall conclusion of hypoallergenic.) |
2. Sample size used for the test set and the data provenance
- Acoustic Testing: Not explicitly stated, but the values are physical properties, likely measured in a lab setting. Data provenance is not specified beyond "Sonishield™ 100 Antimicrobial Ultrasound Gel was evaluated for its acoustic performance."
- Antimicrobial Effectiveness Testing: Not explicitly stated for specific sample counts of the gel, but performed using USP <51> Category 2 as a guide. Data provenance is not specified.
- Animal Biocompatibility (Skin Irritation and Sensitization): Standard animal testing for ISO 10993-10. Not explicitly stated how many animals were used. Data provenance is not specified.
- Clinical Tolerance Study: "two part study." No specific number of participants mentioned. Data provenance is not specified.
- HRIPT (Hypoallergenic study): 55 volunteers. Data provenance is not specified.
All studies mentioned appear to be prospective in nature, as they were conducted to evaluate the performance of the Sonishield™ 100 gel. The country of origin for the data is not specified.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This section is not applicable as the studies described are non-clinical performance evaluations and biocompatibility/tolerance tests for an ultrasound gel, not diagnostic accuracy studies involving medical image interpretation or human diagnoses. Therefore, "ground truth" in the diagnostic sense (e.g., confirmed disease status) and "experts" (e.g., radiologists) in that context are not relevant to the described tests.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
This is not applicable for the reasons stated above (non-clinical performance and biocompatibility studies).
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. The device is an ultrasound gel, not an AI-powered diagnostic tool. No MRMC study was conducted.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This is not applicable. The device is an ultrasound gel, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Acoustic Testing: The "ground truth" would be the physically measured acoustic properties of the gel.
- Antimicrobial Effectiveness Testing: The "ground truth" is the quantitative bacterial count reduction compared to a baseline, against established microbiological standards (USP <51>).
- Animal Biocompatibility: The "ground truth" is the observed biological reaction (or lack thereof) in living tissue, against the criteria of ISO 10993-10.
- Clinical Tolerance Study: The "ground truth" is the observed cutaneous reaction (or lack thereof) in human skin, assessed by study personnel.
- HRIPT (Hypoallergenic study): The "ground truth" is the observed skin reaction (or lack thereof) in human volunteers, assessed by study personnel, following a standardized protocol.
8. The sample size for the training set
This is not applicable. There is no "training set" as this is a physical product undergoing performance and safety testing, not a machine learning model.
9. How the ground truth for the training set was established
This is not applicable for the same reason as point 8.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines beneath them, resembling a stylized representation of water or wind.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 13, 2016
Quotient Medical % Ms. Marilyn Lockhart President Lockhart Regulatory Services, Inc. 1860 Appleby Line, Suite 348 Burlington, Ontario L7L 7H7 CANADA
Re: K151070
Trade/Device Name: Sonishield™ 100 Antimicrobial Ultrasound Gel Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: MUI Dated: January 4, 2016 Received: January 11, 2016
Dear Ms. Lockhart:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert Ochs
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K151070
Device Name
Sonishield™ 100 Antimicrobial Ultrasound Gel
Indications for Use (Describe)
Sonishield™ 100 Antimicrobial Ultrasound Gel is intended for general use as a non-sterile transmission media for acoustically coupling a transducer to a human body surface during external and therapeutic ultrasound imaging procedures. It is placed on the patient's skin prior to initiating an ultrasound examination. It is indicated for prescription use only.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
Sonishield 100 Antimicrobial Ultrasound Gel
Common name: Ultrasound gel
MUI Product code
Class II device,
21 CFR 892.1570
Traditional 510(k) Submission
April 15, 2015
General Information
-510(k) owner's name:
Quotient Medical 3365 Harvester Road, Suite 110, Burlington, Ontario, Canada L7N 3N2 Telephone: 416- 843-6177
-Contact person:
Contact person for Quotient Medical: David Okamoto Vice President. Telephone 416-843-6177
Preferred contact person for Quotient Medical: Marilyn Lockhart
President, Lockhart Regulatory Services Inc. Consultant to Quotient Medical Address: 1860 Appleby Line, Suite 348 Burlington, Ontario L7L 7H7 Canada. Telephone: 905-635-2855
-Prepared on March 20, 2015.
Device name
{4}------------------------------------------------
-Trade name Sonishield™ 100 Antimicrobial Ultrasound Gel
-Common name
Ultrasound gel
-Classified name
Diagnostic ultrasonic transducer/acoustic gel (21 CFR § 892.1570, Product code MUI)
Predicate devices
-Ecogel 100 Ultrasound Gel (K961757) -Konix® Ultrasound Gel (K101952)
Device Description
Sonishield™ 100 Antimicrobial Ultrasound Gel consists of deionized water,
hydroxyethylcellulose, propanediol and benzalkonium chloride. It is a type of conductive medium (i.e.) scanning gel used in ultrasound therapeutic and diagnostic imaging techniques. A scanning gel acts as a couplant that provides an acoustic pathway between the transducer and the skin. In addition, the gel eliminates air (a disruptive influence) from the interface and adapts the contours of the probe to the skin.
The major characteristics of Sonishield™ 100 Antimicrobial Ultrasound Gel include:
- Hypoallergenic, non-irritating
- Water soluble, non-staining and easily cleanable
- . Does not contain oil or fatty matter
- Free from formaldehyde and salt ●
- No toxic effects
- Produced as a completely harmless material
- Has no odor
- Vacuum treated production
- Does not damage the probe
- Does not contain air bubbles
- pH level ranges from 4.5 to 6.5 at 25°C ●
{5}------------------------------------------------
Intended use
Sonishield™ 100 Antimicrobial Ultrasound Gel is intended for general use as a non-sterile transmission media for acoustically coupling a transducer to a human body surface during external therapeutic and diagnostic ultrasound imaging procedures. It is placed on the patient's skin prior to initiating an ultrasound examination. It is indicated for prescription use only.
Technological characteristics
Sonishield™ 100 Antimicrobial Ultrasound Gel has substantially the same technological characteristics as the predicate devices. The three products are compared in the table below:
{6}------------------------------------------------
Substantial Equivalence Discussion
| Subject | Sonishield™ 100AntimicrobialUltrasound Gel | Ecogel 100 UltrasoundGel | Konix® UltrasoundGel |
|---|---|---|---|
| Intended Use | External | External | External |
| Ingredients | Salt freeDye freeAlcohol freeFormaldehyde freePerfume free | Salt freeDye freeAlcohol freeFormaldehyde freePerfume free | Salt freeDye freeAlcohol freeFormaldehyde freePerfume free |
| PhysicalProperties | Twist cap for accuratedispensingFlip-top can for quickrefillingVery high clarityHypoallergenic,bacteriostatic, non-sensitizing | Twist cap for accuratedispensingFlip-top can for quickrefillingGood clarityHypoallergenic,bacteriostatic, non-sensitizing | Twist cap for accuratedispensingFlip-top can for quickrefillingHypoallergenic, non-irritating |
| ChemicalProperties | pH 4.5 - 6.5Density (g/mL) = 1.009Very clear screen imagewith high viscosity andvacuum process. Norapid melting from high-viscosity gel.Viscosity 80,000 –120,000 CPSBoiling point >200°CWater soluble high MWpolymerNo irritation | pH 6.5 ± 0.75Density (g/mL) = 0.99It has low viscosity. Itmelts immediately fromlow viscosityViscosity 35,000 –40,000 CPSBoiling point 100°CWater soluble high MWpolymerNo irritation | pH 6.5 ± 0.75Density (g/mL) = 0.983Very clear screenimage with highviscosity and vacuumprocess. No rapidmelting from high-viscosity gel.Viscosity 100,000 –120,000 CPSBoiling point >200°CWater soluble highMW polymerNo irritation |
| Process | It has a rapidmanufacturing processSonishield™ employs asoft bottle for ease ofuse.Sonishield™ productionemploys a closed-loopsystem so there is nopollution transmission.Product is manufacturedvery cleanly.Standard productionarea | Normal processStandard bottleNormal processStandard productionarea | a rapid manufacturingprocessemploys a soft bottlefor ease of useproduction employs aclosed-loop system sothere is no pollutiontransmission. Productis manufactured verycleanly.Standard productionarea |
| Subject | Sonishield™ 100AntimicrobialUltrasound Gel | Ecogel 100 UltrasoundGel | Konix® UltrasoundGel |
| Standard processmanufactured to releasespecifications. | Standard process | Manufactured in aclean room (1/100000class) | |
| Label | Standard information onpolyethylene label toprevent loss of lettering | Standard information | Standard informationon polyethylene labelto prevent loss oflettering |
| Design | Bottle diameterdesigned to becompatible withultrasound device.Bottle cap is designedfor ease of opening andclosing with one hand. | Conical cap | Bottle diameterdesigned to becompatible withultrasound device.Bottle cap is designedfor ease of opening andclosing with one hand. |
| Safety | Sonishield™ labelcontains appropriatewarnings andcharacteristics (Latex-free, PVC-free) | Standard information | Label containsappropriate warningsand characteristics(Latex-free, PVC-free) |
| Environment ofuse | Hospital | Hospital | Hospital |
| Target population | Pediatric and adult | Pediatric and adult | Pediatric and adult |
| Anatomical site | Body (abdomen) | Body (abdomen) | Body (abdomen) |
| Use | Multiple uses | Multiple uses | Multiple uses |
| Material(package) | Polyethylene | Polyethylene | Polyethylene |
| Patient contactmaterials | Probe | Probe | Probe |
| Energy type | Electricity only for theultrasound device | Electricity only for theultrasound device | Electricity only for theultrasound device |
{7}------------------------------------------------
{8}------------------------------------------------
Non-clinical performance
Acoustic:
Sonishield™ 100 Antimicrobial Ultrasound Gel was evaluated for its acoustic performance. Results indicate that the acoustic properties of the gel are:
-
- Virtually identical to that of human skin.
-
- Similar to other coupling gels commonly used in the United States.
The acoustic properties of Sonishield™ 100 Antimicrobial Ultrasound Gel are as follows:
| • Sound velocity (m/sec) at 30°C | 1497 |
|---|---|
| • Density (kg/m³) at 30°C | 1.023 X 10-3 |
| • Acoustic impedance (kg/m² sec) at 30°C | 1.53 |
| • Attenuation coefficient as a function of frequency, a/f (dB/cm-MHz) | 0.04 ± 0.0042 |
The acoustic properties of the predicate gels are virtually identical. Those properties for the Konix® Ultrasound Gel are provided below:
| • Sound velocity (m/sec) at 30°C | 1516 |
|---|---|
| • Density (kg/m³) at 30°C | 0.98X 10⁻³ |
| • Acoustic impedance (kg/m² sec) at 30°C | 1.49 |
| • Attenuation coefficient as a function of frequency, a/f (dB/cm-MHz) | <0.05 f |
Antimicrobial Effectiveness Testing:
Using USP <51> Category 2 as a guide, antimicrobial effectiveness testing performed on Sonishield™ 100 Antimicrobial Ultrasound Gel demonstrated the bacteria (S. Aureus, Pseudomonas aeruginosa. E.coli. S.Aureus (MRSA) and K. pneumoniae) showed a log reduction greater than 4 from the initial count at 14 days and at 28 days.
Animal tests:
Biocompatibility (ISO 10993-10) testing was conducted for Skin Irritation and Skin Sensitization. Conclusions from these studies: Sonishield™ 100 Antibacterial Ultrasound Gel was found to be non-sensitizing and non-irritating.
Clinical tests:
In clinical studies using Sonishield™ 100 Antimicrobial Ultrasound Gel, the following were concluded:
-In a two part study of the Tolerance of Sonishield™ 100 Antimicrobial Ultrasound Gel, the product produced no signs of cutaneous irritation.
{9}------------------------------------------------
-In another study, (HRIPT with 55 volunteers), no pertinent reaction to the test product was observed during the challenge phase of the study and the product can be considered as hypoallergenic. In the induction phase of the study, 9 of 55 volunteers showed "barely noticeable" reactions, 1 of 55 volunteers showed Grade 1 and 2 volunteers showed Grade 2 erythema, while all other volunteers had no reaction to the gel during the induction phase of the study.
CONCLUSIONS
The above-referenced comparisons of the technological and non-clinical performance characteristics indicate that the Sonishield™ Antimicrobial 100 Ultrasound Gel is nearly identical to the predicate gels and certainly substantially equivalent to both and other coupling gels commonly used in the United States.
§ 892.1570 Diagnostic ultrasonic transducer.
(a)
Identification. A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.(b)
Classification. Class II.