K Number
K150869
Date Cleared
2015-12-10

(253 days)

Product Code
Regulation Number
870.2300
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Checkme Pro Health Monitor is intended to be used for measuring, reviewing and storing of ECG (adults only), oxygen saturation and pulse rate (adults only for continuous data collection and recording, adults and pediatrics for spot checking) and temperature in the home or in healthcare facilities.

This device is not intended to subspital diagnostic ECG device and also not to be used on patients with implanted cardiac devices, such as pacemakers and/or implanted cardio-defibrillators (ICDs).

Device Description

The Checkme Pro Health Monitor is a lightweight, portable health monitor for use in the home or health care facility. The Checkme Pro Health Monitor is used for measuring, displaying, reviewing and storing of multiple physiological parameters including ECG, pulse oxygen saturation (SpO2), pulse rate and temperature. This device is not intended to substitute for a hospital diagnostic ECG device. The device is capable of measuring and storing data for up to twelve users.

AI/ML Overview

The provided text describes the regulatory clearance for the Viatom Checkme Pro Health Monitor, which includes functionalities for ECG, oxygen saturation (SpO2), pulse rate, and temperature measurement. The document details performance testing for each function against recognized standards.

Here's a breakdown of the requested information based on the provided text, focusing on the ECG, SpO2, and Thermometer functionalities as these are the core medical measurements. The Pedometer function is also listed but does not have detailed acceptance criteria or performance study information similar to the medical functions within this document.

1. Table of Acceptance Criteria and Reported Device Performance

FunctionParameterAcceptance Criteria (Standard Reference)Reported Device Performance
ECGECG algorithmIEC 60601-2-47: Medical electrical equipment Part 2: Particular requirements for the basic safety and essential performance of ambulatory electrocardiographic systems.The ECG algorithm used in the Checkme Pro Health Monitor has been tested according to the requirements of IEC 60601-2-47. (Implies compliance)
Heart Rate RangePredicate K062894 (InstantCheck RTD-ECG Monitor): 45 to 180 bpmSubject Device (Checkme Pro): 30~250 bpm, ±2 bpm or ±2% whichever is larger. This indicates the Checkme Pro has a wider range with a specified accuracy.
SpO2Accuracy (Integrated Sensor)ISO 80601-2-61: Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment.The Checkme Pro Health Monitor has been tested and shown to comply with the requirements of ISO 80601-2-61. (Implies compliance). Specific accuracy stated as: 70%-100%: ± 3% (Arms:2.11) with further breakdown: 70-80%: ± 3%, 80-90%: ± 3%, 90-100%: ± 2%. 0% - 69% not defined. This is compared to the predicate's (K131762's external sensor) 70%-100%: ± 2% (Arms:1.88) with 70-80%: ± 3%, 80-90%: ± 2%, 90-100%: ± 2%. While slightly different, the subject device's performance is listed and implicitly accepted through the substantial equivalence claim.
Accuracy (External Sensor)ISO 80601-2-61: Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment. Additional predicate (K131762) has an accuracy of 70% - 100%: ± 2% (0% - 69% not defined).Same compliance as above. Specific accuracy stated as: 70%-100%: ± 2% (Arms:1.88) with further breakdown: 70-80%: ± 3%, 80-90%: ± 2%, 90-100%: ± 2%. 0% - 69% not defined. This meets the predicate's overall accuracy for 70-100%.
PR AccuracyPredicate K131762: 25-250 BPM, ± 1 bpm or ± 2% whichever is larger.Subject Device (Checkme Pro): 30-250 BPM, ± 2 bpm or ± 2% whichever is larger. While the predicate has slightly better absolute bpm accuracy, the overall accuracy within 2% is similar.
ThermometerBody Mode AccuracyISO 80601-2-56:2009: Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement.The Checkme Pro has been tested and shown to comply with the requirements of ISO 80601-2-56:2009. (Implies compliance). Specific accuracy stated as: ±0.4°F. This is compared to the predicate's (K100953) 36°C to 39°C (96.8°F to 102.2°F): +/- 0.2°C / 0.4°F, and 34.0°C to 35.9°C (93.2°F to 96.6°F), 39.1°C to 42.2°C (102.4°F to 108.0°F): +/- 0.3°C/0.5°F. The subject device's +/- 0.4°F falls within the predicate's accuracy range for its primary operating range.

2. Sample size used for the test set and the data provenance

The document does not provide specific sample sizes (number of patients/cases) for the clinical performance tests (ECG, SpO2, Temperature). It only states that the device was tested according to relevant IEC and ISO standards. These standards typically involve specific test methodologies, which may include human subjects or simulations, but the details of the sample size or whether the data was retrospective or prospective, or country of origin are not disclosed in this summary.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided in the document. The performance claims are based on compliance with international standards, which would outline how ground truth is established within their testing protocols, but the specifics (e.g., number of expert clinicians, their qualifications) are not detailed here.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

The adjudication method for establishing ground truth within the test sets is not specified in this document.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There is no mention of a multi-reader multi-case (MRMC) comparative effectiveness study, nor any evaluation of human reader improvement with AI assistance. The device is a monitor that provides measurements and waveforms, not an AI-assisted diagnostic tool that interprets complex medical images or data requiring human reader assessment of improvement.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

The performance testing described (ECG algorithm testing, SpO2 compliance, Thermometer compliance) assesses the device's inherent function, which is a standalone performance of the algorithms and sensors. For instance, the ECG algorithm's compliance with IEC 60601-2-47 means its performance in calculating heart rate and waveform display meets the standard without human intervention for results generation. Similarly for SpO2 and temperature, the listed accuracies are for the device's direct measurements.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The document does not explicitly state the type of ground truth used for each test. However, for devices like these, ground truth would typically be established by:

  • ECG: Reference highly accurate diagnostic ECG devices or expert interpretation of standard 12-lead ECGs for heart rate and morphology analysis (as per IEC 60601-2-47).
  • SpO2: Co-oximetry for blood oxygen saturation levels for absolute accuracy, or a calibrated reference pulse oximeter for comparative accuracy (as per ISO 80601-2-61).
  • Temperature: Calibrated reference thermometers (e.g., rectal thermometers for core body temperature, as per ISO 80601-2-56).

8. The sample size for the training set

The document does not mention any training set size. This device appears to be primarily a physiological monitor with direct measurement capabilities rather than a complex machine learning model that requires explicit training data in the context of this regulatory submission. The "ECG algorithm" mentioned would likely refer to signal processing algorithms for heart rate detection and waveform filtering, which are typically designed and validated through engineering methods rather than machine learning training sets in the same sense.

9. How the ground truth for the training set was established

Since no training set is mentioned in the context of machine learning, this information is not applicable and therefore not provided in the document.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized graphic of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a silhouette effect. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the graphic.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 10, 2015

Viatom Technology Co., Ltd. % Donna-Bea Tillman Senior Consultant Biologics Consulting Group, Inc. 400 N. Washington St. Suite 100 Alexandria, Virginia 22314

Re: K150869

Trade/Device Name: CheckMe Pro Health Monitor Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer and Rate Alarm) Regulatory Class: Class II Product Code: MWI Dated: November 6, 2015 Received: November 10, 2015

Dear Donna-Bea Tillman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Mitchell Stein

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K150869

Device Name Checkme Pro Health Monitor

Indications for Use (Describe)

The Checkme Pro Health Monitor is intended to be used for measuring, reviewing and storing of ECG (adults only), oxygen saturation and pulse rate (adults only for continuous data collection and recording, adults and pediatrics for spot checking) and temperature in the home or in healthcare facilities.

This device is not intended to subspital diagnostic ECG device and also not to be used on patients with implanted cardiac devices, such as pacemakers and/or implanted cardio-defibrillators (ICDs).

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Device Common Name:Multiparameter Monitor
Device Trade Name:Checkme Pro Health Monitor
Applicant:Viatom Technology Co., Ltd.C607,Languang Park, No. 7 Xinxi Road, Hi-TechPark(north) Nanshan, Shenzhen, 518057, P.R.Chinawww.viatomtech.com
Contact:Zhou SaixinGeneral ManagerViatom Technology Co., Ltd.C607,Languang Park, No. 7 Xinxi Road, Hi-TechPark(north) Nanshan, Shenzhen, 518057, P.R.ChinaPhone: +86 755 86638929Fax: +86 755 22649904Email: zhousaixin@viatomtech.com
Prepared by:Donna-Bea Tillman, Ph.D.Senior Consultant, DevicesBiologics Consulting Group, Inc.Phone: 410-531-6542Email: dtillman@bcg-usa.com
Date Prepared:November 6, 2015
Classification Regulation:21 CFR 870.2300 – Cardiac Monitor
Panel:Cardiovascular
Product Code:MWI, Physiological Patient Monitor
Secondary Product Codes:DQA (21 CFR 870.2700)
DPS (21 CFR 870.2340)
FLL (21 CFR 880.2910)
DRT (21 CFR 870.2300)
Predicate Device:K131762, M800 Handheld Monitor
Secondary Predicates:K062894, InstantCheck RTD-ECG Monitor
K100953, Microlife Non-Contact Infrared Forehead

{4}------------------------------------------------

Indications for Use:

The Checkme Pro Health Monitor is intended to be used for measuring, displaying, reviewing and storing of ECG (adults only), oxygen saturation and pulse rate (adults only for continuous data collection and recording, adults and pediatrics for spot checking) and temperature in the home or in healthcare facilities.

This device is not intended to substitute for a hospital diagnostic ECG device and also not to be used on patients with implanted cardiac devices, such as pacemakers and/or implanted cardiodefibrillators (ICDs).

Device Description:

The Checkme Pro Health Monitor is a lightweight, portable health monitor for use in the home or health care facility. The Checkme Pro Health Monitor is used for measuring, displaying, reviewing and storing of multiple physiological parameters including ECG, pulse oxygen saturation (SpO2), pulse rate and temperature. This device is not intended to substitute for a hospital diagnostic ECG device. The device is capable of measuring and storing data for up to twelve users.

FunctionsMeasured parametersFeatures
ECGECG waveformHeart Rate (HR)Cable or cable-free, four modes, 30s waveformreplay, voice memo, 100 measurements storage
Pulse Oximeter:Spot-checkPleth waveformSpO2, Pulse Rate (PR)Cable or cable-free, 20 second measurement, 100measurements storage
Pulse Oximeter:ContinuousRecording(adult only)Total duration, <90% STAT,Average saturation, LowestsaturationRecording for up to 10 hrs
ThermometerTemperatureIntegrated infrared sensor, 3 second measurement,or °F configurable
PedometerSteps, distance, calories, fatburned, speedBuilt-in high accuracy 3D smart sensor, algorithmto calculate distance

Measurements Provided by Checkme Pro

Performance Testing:

The performance testing provided for the Checkme Pro Health Monitor consists of the following:

ECG Testing

The ECG algorithm used in the Checkme Pro Health Monitor has been tested according to the requirements of IEC 60601-2-47. Medical electrical equipment Part 2: Particular requirements for the basic safety and essential performance of ambulatory electrocardiographic systems.

{5}------------------------------------------------

SpO2 Testing

The Checkme Pro Health Monitor has been tested and shown to comply with the requirements of ISO 80601-2-61. Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment

Thermometer Testing

The Checkme Pro has been tested and shown to comply with the requirements of the FDArecognized standard ISO 80601-2-56:2009: Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement.

Home Healthcare Testing

The Checkme Pro has been tested and shown to comply with the following FDA-recognized standard: IEC 60601-1-11 MEDICAL ELECTRICAL EQUIPMENT - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.

System Level Testing

Viatom conducted system level performance testing to demonstrate that the device meets its specifications.

The device passed all tests and performed according to specifications.

Biocompatibility Testing:

It is a surface contacting device with less than 24 hours contact duration. The patient contacting materials were testing in accordance with ISO 10993-1: 2009/AC 2010.

Software Verification and Validation:

Similar to the predicate device, the software level of concern for the Checkme Pro Health Monitor is MODERATE. Per FDA Guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices - Guidance for Industry and FDA Staff", the required software documentation for a MODERATE Level of Concern device has been provided.

EMC and Electrical Safety:

The following EMC and Electrical Safety tests were performed on the device:

  • . IEC 60601-1: 2005 + Corrigendum 1 + Corrigendum 2 + AM1 (2012) IEC 60601-1 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
  • EN 60601-1-2: 2007 + AC: 2010 medical electrical equipment part 1-2: general . requirements for basic safety and essential performance - collateral standard: electromagnetic compatibility - requirements and tests. (General II (ES/EMC))

The results of these tests demonstrate the electrical safety and electromagnetic compatibility of the Checkme Pro device.

{6}------------------------------------------------

The Checkme Pro Health Monitor is provided with a rechargeable lithium ion polymer battery. Therefore, Viatom has also conducted battery safety testing in accordance with the FDArecognized standard IEC 62133:2012 Secondary cells and batteries containing alkaline or other non-acid electrolytes – Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications. The battery passed all required tests, and therefore has been shown to be safe for the proposed intended use.

The device comparison table below compares the primary features of the proposed Checkme device and the predicate M800 device.

Subject DevicePredicate Device
510(k) NumberK150869K131762
ApplicantViatom Technology Co. Ltd.GUANGDONG BIOLIGHTMEDITECH CO., LTD
Device NameCheckme Pro Health MonitorM800 Handheld Monitor
ClassificationRegulation870.2300 - Physiological PatientMonitor870.2300 – Physiological PatientMonitor
Product Code(Primary)MWIMWI
Indications for UseThe Checkme Pro HealthMonitor is intended to be used formeasuring, displaying, reviewingand storing of ECG (adults only),oxygen saturation and pulse rate(adults only for continuous datacollection and recording, adultsand pediatrics for spot checking)and temperature in the home or inhealthcare facilities.This device is not intended tosubstitute for a hospitaldiagnostic ECG device and alsonot to be used on patients withimplanted cardiac devices, suchas pacemakers and/or implantedcardio-defibrillators (ICDs).M800 handheld monitor is intended forcontinuously monitoring or spotchecking of SpO2, PR, ECG and HR ofadult, pediatric and neonatal patients inhospital, hospital type facilities as wellas in the home care environment.
PhysiologicalParameters MonitoredECG, HR, SpO2, TemperatureECG, HR, SpO2
LocationHome and HospitalHome and Hospital
Rx or OTCRxRx
Power Supply560mAh rechargeable lithium-ionpolymer batteryRechargeable Lithium Battery or AAalkaline or lithium batteries

Device Comparison Table

{7}------------------------------------------------

The comparison of ECG Measurements table below provides a more detailed comparison of the Checkme device with the additional predicate for ECG monitoring, the DailyCare InstantCheck RTD-ECG Monitor (K062894)

Comparison of ECG Measurements

Subject DevicePredicate Device (Additional)
510(k) NumberK150869K062894
ApplicantViatom Technology Co. Ltd.Daily Care BioMedical, Inc.
Device NameCheckme Pro Health MonitorInstantCheck RTD-ECG Monitor
ClassificationRegulation870.2300 – Physiological PatientMonitor870.2340 - Electrocardiograph
Measured ECGParameterECG waveform, Heart Rate (HR)ECG waveform, Heart Rate (HR), STsegment and QRS interval
ECG RhythmClassificationNot includedRegular Heart RateFast Heart RateSlow Heart RateHigh ST ValueLow ST ValueHigh QRS ValueIrregular Heart RateImpossible to Analyze
ECG Lead TypeExternal ECGCable andElectrodesIntegratedECGElectrodesExternal and Integrated ECG Electrodes
Input impedance> 10 M Ohm> 20 M Ohm
Input dynamic range+/- 3 mV+/- 3 mV
Bandwidth0.05 – 40 Hz0.1 - 40 Hz
A/D conversion16 bit12 bit
Sampling Rate500 samples/sec250 samples/sec
Measurement Time30 seconds30 seconds
Display400*240 Dot-matrix LCD display240 X 128 Dot-matrix LCD display
InputDry conduction electrodes and/orexternal auxiliary electrodesDry conduction electrodes and/or externalauxiliary electrodes
Heart rate Range30~250 bpm, ±2 bpm or ±2%whichever is larger45 to 180 bpm
Subject DevicePredicate Device
510(k) NumberK150869K131762
Device NameCheckme Pro Health MonitorM800 Handheld Monitor
ClassificationRegulation870.2300 - Physiological PatientMonitor870.2300 - Physiological PatientMonitor
Display DataSpO2, Pulse Rate (PR)SpO2, Pulse Rate (PR)
ModeContinuous Recording and Spot-CheckContinuous Recording and Spot-Check
Sensor TypesIntegrated and externalExternal
PopulationAdult and Pediatric (continuousrecording only for adult)Adult, Pediatric and Neonate
SpO2 Display Range0 - 100%0 - 100%
AccuracyIntegratedSensor70%-100%:$\pm$ 3%(Arms:2.11)70-80%: $\pm$ 3%80-90%: $\pm$ 3%90-100%: $\pm$ 2%0% - 69% notdefinedExternal Sensor70%-100%: $\pm$ 2%(Arms:1.88)70-80%: $\pm$ 3%80-90%: $\pm$ 2%90-100%: $\pm$ 2%0% - 69% notdefined70% - 100%: $\pm$ 2%0% - 69% not defined
Resolution1%1%
PR range30-250 BPM25-250 BPM
PR Accuracy30-250 BPM, $\pm$ 2 bpm or $\pm$ 2%whichever is larger25-250 BPM, $\pm$ 1 bpm or $\pm$ 2%whichever is larger
Resolution1bpm1bpm
Subject DevicePredicate Device (Additional)
510(k) NumberK150869K100953
ApplicantViatom Technology Co. Ltd.Microlife Intellectual Property, GMBH
Device NameCheckme Pro Health MonitorMicrolife Non-Contact InfraredForehead Thermometer
ClassificationRegulation870.2300 - Physiological Patient Monitor880.2910 - Thermometer, Electronic,Clinical
Measurementtechniqueinfrared sensor (thermopile)infrared sensor (thermopile)
MeasurementsiteForeheadForehead
Unit°F°C/°F
Body modemeasuringrange93.2°F to 108.0°F34°C to 42.2°C (93.2°F to 108.0°F)
Body modeAccuracy±0.4°F36°C to 39°C (96.8°F to 102.2°F): +/-0.2°C /0.4°F34.0°C to 35.9°C (93.2°F to 96.6°F),39.1 °C to 42.2°C (102.4°F to 108.0°F):

{8}------------------------------------------------

The Comparison of Pulse Oximeter Measurement and Analysis table below provides a more detailed comparison of the pulse oximetry feature of the Checkme device with the primary predicate, theM800 Handheld Monitor.

Comparison of Pulse Oximeter Measurement and Analysis

The Comparison of Temperature Measurement and Analysis table below provides a more detailed comparison of the Checkme and the additional predicate for temperature measurement, the Microlife Non-Contact Infrared Forehead Thermometer (K100953).

{9}------------------------------------------------

Comparison of Temperature Measurement and Analysis

Substantial Equivalence Conclusion:

The primary predicate device for the Checkme Pro Health Monitor is M800 Handheld Monitor cleared in K131762. The M800 is also a handheld device for monitoring physiological parameters in the home or health care environment.

The Checkme Pro Health Monitor has three primary medical device functions, and each of these has been compared to a predicate device. While there are some minor differences in technological characteristics, this do not raise different questions of safety and effectiveness. Each function has been tested in accordance with the appropriate standards, and the results of testing demonstrate that the Checkme Pro Health Monitor complies with the standards and has equivalent performance to the reference predicate devices.

Therefore, the Checkme Pro Health Monitor has been shown to be substantially equivalent to the M800 Handheld Monitor.

§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).

(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).