(375 days)
The OVA1 Next Generation test is a qualitative serum test that combines the results of five immunoassays into a single numeric result. It is indicated for women who meet the following criteria: over age 18, ovarian adnexal mass present for which surgery is planned, and not yet referred to an oncologist.
The OVA1 Next Generation test is an aid to further assess the likelihood that malignancy is present when the physician's independent clinical and radiological evaluation does not indicate malignancy. The test is not intended as a screening or stand-alone diagnostic assay.
The OVA1 Next Generation (NG) test consists of software, instruments, assays and reagents. The software incorporates the results of serum biomarker concentrations from five immunoassays to calculate a single, unitless numeric result indicating a low or high risk of ovarian malignancy.
The assays used to generate the numeric result (OVA1 NG test result) are APO, CA 125 II, FSH, HE4 and TRF.
Biomarker values are determined using assays on the Roche cobas® 6000 system, which is a fully automated, software-controlled system for clinical chemistry and immunoassay analysis. The biomarker assays are run according to the manufacturer's instructions as detailed in the package insert for each reagent.
The OVA1 NG software (OvaCalc v4.0.0) contains a proprietary algorithm that utilizes the results (values) from the five biomarker assays, (APO, CA 125 II, FSH, HE4 and TRF). The assay values from the cobas 6000 system are either imported into OvaCalc through a .csv file or manually entered into the OvaCalc user interface to generate an OVA1 NG test result between 0.0 and 10.0. A low- or high-risk result is then determined by comparing the software-generated risk score to a single cutoff (low-risk result <5. high-risk result ≥5).
Here's an analysis of the acceptance criteria and study findings for the OVA1 Next Generation device, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state pre-defined acceptance criteria for the OVA1 Next Generation device in terms of specific performance thresholds (e.g., "Sensitivity must be >X%"). Instead, the study focuses on demonstrating substantial equivalence to a predicate device (the original OVA1) and showing improvements in certain metrics.
However, based on the clinical performance evaluation and comparisons to the predicate, we can infer the primary goal was to at least maintain sensitivity while significantly improving specificity and positive predictive value. The "clinically small" and "clinically significant" definitions also act as implicit criteria for comparison with the predicate.
Here's a table summarizing the reported comparative performance:
| Metric (vs. Predicate OVA1) | OVA1 Next Generation Performance (with PA) | OVA1 Next Generation Performance (Standalone) | Goal/Inferred Acceptance Criterion |
|---|---|---|---|
| Specificity (Overall) | Improved by ~14% (64.8% vs 50.9%) | Improved by ~16% (69.1% vs 53.6%) | Significantly improved specificity while maintaining sensitivity |
| Specificity (Postmenopausal) | Improved ~23% (60.9% vs 37.8%) | Improved ~24% (65.4% vs 41.0%) | Significantly improved specificity |
| Specificity (Premenopausal) | Improved ~8% (67.3% vs 59.2%) | Improved ~10% (71.4% vs 61.6%) | Significantly improved specificity |
| Sensitivity (Overall) | Differences "clinically small" (~ -2.17%) (93.5% vs 95.7%) | Differences "clinically small" (~ -1.09%) (91.3% vs 92.4%) | Maintain similar sensitivity |
| Sensitivity (Postmenopausal) | Differences "clinically small" (~ -1.64%) (95.1% vs 96.7%) | Identical (91.8% vs 91.8%) | Maintain similar sensitivity |
| Sensitivity (Premenopausal) | Differences "clinically small" (~ -3.23%) (90.3% vs 93.5%) | Differences "clinically small" (~ -3.23%) (90.3% vs 93.5%) | Maintain similar sensitivity |
| Positive Predictive Value (PPV) (Overall) | N/A (Only Standalone data provided for PPV comparison) | Improved by 9% (40.4% vs 31.4%) | Significantly improved PPV |
| Negative Predictive Value (NPV) (Overall) | Differences "substantially equivalent" (~0.35%) (97.7% vs 98.1%) | Differences "substantially equivalent" (~0.35%) (97.2% vs 96.8%) | Maintain similar NPV |
| Precision (%CV) | 1.54% (Overall) | 1.54% (Overall) | Better than or equivalent to predicate (Predicate was 4.09%) |
| Reproducibility (%CV) | 1.63% (Overall) | 1.63% (Overall) | Better than or equivalent to predicate (Predicate was 2.80%) |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size:
- Clinical Performance Evaluation: 493 evaluable subjects (from an initial 519 enrolled). Split into 276 premenopausal and 217 postmenopausal.
- Clinical Specificity - Healthy Women Study: 152 healthy women (68 premenopausal, 84 postmenopausal).
- Clinical Specificity - Other Cancers and Disease States: 401 samples from women with various non-ovarian cancers and benign conditions.
- Method Comparison (Archived Samples): 133 samples (28 primary ovarian malignancies, 105 benign ovarian conditions) for a direct comparison with the predicate.
- Data Provenance: The primary clinical study used a banked sample set from a prospective, multi-site pivotal study of OVA1 – the OVA500 Study. The archived samples were used to conduct a side-by-side clinical validation for Substantial Equivalence purposes. The method comparison study (Table 10) also used archived samples collected from selected larger prospective studies, tested within one year of collection. The document does not specify the country of origin for the OVA500 study, but it is implied to be a US study given the FDA submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The ground truth for the clinical performance evaluation was established by postoperative pathology diagnosis, which was recorded at each enrolling site and independently reviewed.
The document does not specify the number of experts or their explicit qualifications (e.g., "radiologist with 10 years of experience") for this pathology review. However, "postoperative pathology diagnosis" generally implies review by trained pathologists.
4. Adjudication Method for the Test Set
The document explicitly states that postoperative pathology diagnosis was "independently reviewed." However, the exact adjudication method (e.g., 2+1, 3+1, none) for discrepancies in pathology results is not detailed in the provided text.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done in the traditional sense of evaluating multiple human readers' performance with and without AI assistance.
The study compared the performance of:
- The OVA1 Next Generation device itself (standalone).
- The original OVA1 device (predicate) itself (standalone).
- A "dual assessment" combining Physician Assessment (PA) OR the device result (OVA1 Next Generation or original OVA1).
While Physician Assessment (PA) involves human readers (clinicians), the study design evaluates the addition of the device's score to the physician's assessment, rather than directly measuring improvement of human readers performing a task with AI assistance vs without.
- Effect Size of Human Readers Improvement with AI vs. Without AI Assistance: Not directly measurable from the provided data in the context of an MRMC study. The "with PA" results show the combined performance, where the device acts as an "aid" to PA.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Yes, a standalone (algorithm only) performance evaluation was done.
Table 6 and Table 7 explicitly present "Standalone Specificity" and "Standalone Sensitivity" for the OVA1 Next Generation device without Physician Assessment (PA) in the risk calculation.
7. The Type of Ground Truth Used
The primary ground truth used for the clinical performance evaluation was pathology (postoperative pathology diagnosis), which was independently reviewed.
8. The Sample Size for the Training Set
The document does not provide the sample size for the training set for the OVA1 Next Generation algorithm. It describes the device's algorithm and its inputs but focuses on the validation study using a banked sample set without detailing the original training cohort.
9. How the Ground Truth for the Training Set Was Established
Since the sample size for the training set is not provided, the method for establishing its ground truth is also not detailed. It is implied that the algorithm was developed (and likely trained/tuned) using similar pathology-confirmed data, but specifics are absent in this document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three faces in profile, stacked on top of each other, with flowing lines representing hair or fabric.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 26, 2016
VERMILLION, INC. BENJAMIN A. KIMBALL SENIOR DIRECTOR OF REGULATORY AND QUALITY AFFAIRS, COMPLIANCE OFFICER 12117 BEE CAVES ROAD BUILDING 3 SUITE 100 AUSTIN, TEXAS 78738
Re: K150588
Trade/Device Name: OVA1 Next Generation Regulation Number: 21 CFR §866.6050 Regulation Name: Ovarian adnexal mass assessment score test system Regulatory Class: Class II Product Code: ONX Dated: February 22, 2016 Received: March 1, 2016
Dear Mr. Kimball:
This letter corrects our substantially equivalent letter of March 18, 2016.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kelly Olir er -S
FOR Leonthena Carrington, MS, MBA, MT(ASCP) Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health (OIR) Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K150588
Device Name OVA1 Next Generation
Indications for Use (Describe)
The OVA1 Next Generation test is a qualitative serum test that combines the results of five immunoassays into a single numeric result. It is indicated for women who meet the following criteria: over age 18, ovarian adnexal mass present for which surgery is planned, and not yet referred to an oncologist.
The OVA1 Next Generation test is an aid to further assess the likelihood that malignancy is present when the physician's independent clinical and radiological evaluation does not indicate malignancy. The test is not intended as a screening or stand-alone diagnostic assay.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary OVA1 Next Generation
510(k) Number: K150588
| Manufacturer Identification | |
|---|---|
| Submitted by: | Vermillion, Inc. |
| 12117 Bee Caves Rd | |
| Building 3, Suite 100 | |
| Austin, Texas 78738 | |
| 512.519.0435 | |
| Contact Information: | |
| Benjamin A. Kimball | |
| Senior Director of Regulatory and Quality Affairs | |
| Vermillion, Inc. | |
| 12117 Bee Caves Rd | |
| Building 3, Suite 100 | |
| Austin, Texas 78738 | |
| 512.519.0435 | |
| bkimball@vermillion.com |
| Date Prepared: | 03.06.2015 |
|---|---|
| Proprietary Name | OVA1 Next Generation |
| Common Name | Ovarian adnexal mass assessment score testsystem |
| Device Classification | 21 CFR 866.6050 |
| Proposed Regulatory Class | Class II |
| Device Product Code | ONX |
Purpose of this Special 510(k)
This Traditional 510(k) seeks clearance for a new test.
Device Description
The OVA1 Next Generation (NG) test consists of software, instruments, assays and reagents. The software incorporates the results of serum biomarker concentrations from five immunoassays to calculate a single, unitless numeric result indicating a low or high risk of ovarian malignancy.
The assays used to generate the numeric result (OVA1 NG test result) are APO, CA 125 II, FSH, HE4 and TRF.
{4}------------------------------------------------
Biomarker values are determined using assays on the Roche cobas® 6000 system, which is a fully automated, software-controlled system for clinical chemistry and immunoassay analysis. The biomarker assays are run according to the manufacturer's instructions as detailed in the package insert for each reagent.
The OVA1 NG software (OvaCalc v4.0.0) contains a proprietary algorithm that utilizes the results (values) from the five biomarker assays, (APO, CA 125 II, FSH, HE4 and TRF). The assay values from the cobas 6000 system are either imported into OvaCalc through a .csv file or manually entered into the OvaCalc user interface to generate an OVA1 NG test result between 0.0 and 10.0. A low- or high-risk result is then determined by comparing the software-generated risk score to a single cutoff (low-risk result <5. high-risk result ≥5).
| Analyte | Reagent and Calibrator | Instrument |
|---|---|---|
| Apolipoprotein A-1 | cobas APO A1, C.F.A.S. Lipids | Roche cobas 6000:Roche cobas® c501 |
| CA 125 | cobas CA 125 Gen 2, CA 125 II Cal Set | Roche cobas 6000:Roche cobas® e601 |
| Follicle Stimulating Hormone(FSH) | cobas FSH, FSH Cal Set II | Roche cobas 6000:Roche cobas® e601 |
| Human epididymis protein 4(HE4) | cobas HE4, HE4 Cal Set | Roche cobas 6000:Roche cobas® e601 |
| Transferrin | cobas Transferrin, C.F.A.S. Proteins | Roche cobas 6000:Roche cobas® c501 |
The analytes and corresponding analytes and calibrators are as follows:
Substantial Equivalence Information:
Predicate Device and K number: OVA1: K081754
See Tables 1 and 2 below for a comparison with predicate technological characteristics.
| Table 1 - Similarities |
|---|
| Item | Subject Device | Predicate OVA1 |
|---|---|---|
| Intended use | The OVA1 Next Generation testis a qualitative serum test thatcombines the results of fiveimmunoassays into a singlenumeric result. It is indicated forwomen who meet the followingcriteria: over age 18, ovarianadnexal mass present for whichsurgery is planned, and not yetreferred to an oncologist.The OVA1 Next Generation testis an aid to further assess the | The OVA1™ Test is a qualitativeserum test that combines theresults of five immunoassays intoa single numerical score. It isindicated for women who meet thefollowing criteria: over age 18;ovarian adnexal mass present forwhich surgery is planned, and notyet referred to an oncologist. TheOVA1 Test is an aid to furtherassess the likelihood thatmalignancy is present when thephysician's independent clinical |
| Item | Subject Device | Predicate OVA1 |
| likelihood that malignancy ispresent when the physician'sindependent clinical andradiological evaluation does notindicate malignancy. The test isnot intended as a screening orstand-alone diagnostic assay. | and radiological evaluation doesnot indicate malignancy. The testis not intended as a screening orstand-alone diagnostic assay. | |
| Indications for Use | The OVA1 Next Generation testis a qualitative serum test thatcombines the results of fiveimmunoassays, ApolipoproteinA-1 (APO), cancer antigen 125(CA 125 II), Follicle StimulatingHormone (FSH), Humanepididymis protein 4 (HE4), andTransferrin (TRF) into a singlenumeric result. It is indicated forwomen who meet the followingcriteria: over age 18, ovarianadnexal mass present for whichsurgery is planned, and not yetreferred to an oncologist. TheOVA1 Next Generation test is anaid to further assess thelikelihood that malignancy ispresent when the physician'sindependent clinical andradiological evaluation does notindicate malignancy. The test isnot intended as a screening orstand-alone diagnostic assay. | Substantially the same as Subjectdevice |
| Boxed Warning | Should not be used without anindependent clinical and imagingevaluation and is not intended tobe a screening test or todetermine whether a patientshould proceed to surgery.Incorrect use carries the risk ofunnecessary testing, surgery, and/ or delayed diagnosis. | Same as Subject device |
| Sample Matrix | Serum | Same as Subject device |
| Type of Test | Algorithm | Same as Subject device |
| Analytes | APO, CA 125, FSH, HE4, TRF | APO, CA 125, TRF |
| Item | Subject Next Generation | Predicate OVA1 |
| Analytes | FSH | Beta-2 microglobulin, Prealbumin |
| Measurement | Score based on 5 analytes | Score based on 5 analytes |
| Clinical Cutoff | 5.0 | 5.0 Premenopausal,4.4 Postmenopausal |
| Platform | Roche cobas e601, c501 | BNII, Roche Elecsys 2010 |
{5}------------------------------------------------
Premarket Notification – OVA1 Next Generation
{6}------------------------------------------------
Table 2 - Differences
Intended Use of the Device
The OVA1 Next Generation test is a qualitative serum test that combines the results of five immunoassays into a single numeric result. It is indicated for women who meet the following criteria: over age 18, ovarian adnexal mass present for which surgery is planned, and not yet referred to an oncologist.
The OVA1 Next Generation test is intended to be part of the preoperative evaluation to aid in assessing whether a woman who presents with an ovarian adnexal mass is at high or low likelihood of finding malignancy at surgery. The OVA1 Next Generation test must be interpreted in conjunction with an independent clinical and imaging evaluation. The test is not intended as a screening or stand-alone diagnostic assay.
PRECAUTION: The OVA1 Next Generation test should not be used without an independent clinical and imaging evaluation and is not intended to be a screening test or to determine whether a patient should proceed to surgery. Incorrect use of the OVA1 Next Generation test carries the risk of unnecessary testing, surgery, and/or delayed diagnosis.
Indications for Use
The OVA1 Next Generation test is a qualitative serum test that combines the results of five immunoassays, Apolipoprotein A-1 (APO), cancer antigen 125 (CA 125 II), Follicle Stimulating Hormone (FSH), Human epididymis protein 4 (HE4), and Transferrin (TRF) into a single numeric result. The OVA1 Next Generation test is intended to aid in assessing whether a woman who presents with an ovarian adnexal mass is at high or low likelihood of finding malignancy as part of the preoperative evaluation. It is indicated for women who meet the following criteria: over age 18, ovarian adnexal mass present for which surgery is planned, and not yet referred to an oncologist. The OVA1 Next Generation test must be interpreted in conjunction with an independent clinical and imaging evaluation. The test is not intended as a screening or stand-alone diagnostic assav.
Clinical Performance Evaluation:
Clinical and Analytical testing included:
- Clinical Performance on OVA500 cohort ●
- Clinical Specificity Healthy Pre- and Postmenopausal women ●
{7}------------------------------------------------
- Clinical Specificity Other Cancers and Diseases ●
- Precision
- Reproducibility ●
- Sample Stability ●
- . Interference
Clinical Performance Evaluation
Description of Subjects:
The clinical study used a banked sample set from a prospective, multi-site pivotal study of OVA1 - the OVA500 Study - which was published in a leading peer-reviewed specialty journal in Feb 2013. Four hundred ninety-three of the banked samples were used to conduct a side-by-side clinical validation for Substantial Equivalence purposes.
The external sites received the samples for blinded validation testing and did not have access to patient or physician information. The validation sites sent the OvaCalc reports to an independent 3rd party statistician for analysis.
Clinicians were required to document the results of physical examination, family history, imaging, laboratory tests (including CA 125, when available, but not OVA1), and to make a formal pre-surgical prediction of malignancy. In cases where the formal prediction was done by a clinician other than the enrolling physician, the referral history and the specialty of the clinician who made the prediction were recorded, as was the specialty of the surgeon who ultimately operated on each patient. In order to reflect their routine clinical judgment and referral behavior, physicians were not asked to either follow any specific prediction algorithm or justify their prediction. Postoperative pathology diagnosis was recorded at each enrolling site and independently reviewed.
A total of 519 subjects were reported on for this study, of which 493 were evaluable for OVA1 and Physician Assessment (PA). The demographics are detailed in Table 3 below:
| All EnrolledSubjects(N= 519) | All EvaluableSubjects(N= 493) | Pre-menopausalWomen(N= 276) | Post-menopausalWomen(N= 217) | |
|---|---|---|---|---|
| Age, years | ||||
| N | 519 | 493 | 276 | 217 |
| Mean (SD) | 48.4 (14.32) | 48.6 (14.16) | 39.5 (8.96) | 60.2 (10.74) |
| Median | 47 | 48 | 41 | 60 |
| Range (min, max) | 18, 87 | 18, 87 | 18, 60 | 33, 87 |
| Ethnicity/race, n(%) | ||||
| Asian | 13 (2.5) | 13 (2.6) | 8 (2.9) | 5 (2.3) |
| Black or African American | 86 (16.6) | 81 (16.4) | 54 (19.6) | 27 (12.4) |
Table 3 - Specimen and subject disposition - Demography
{8}------------------------------------------------
| Vermillion Inc. | ||
|---|---|---|
| Premarket Notification - OVA1 Next Generation |
| All EnrolledSubjects(N= 519) | All EvaluableSubjects(N= 493) | Evaluable Subjects | ||
|---|---|---|---|---|
| Pre-menopausalWomen(N= 276) | Post-menopausalWomen(N= 217) | |||
| Native Hawaiian/Pacific islander | 1 (0.2) | 1 (0.2) | 1 (0.4) | 0 (0.0) |
| White | 365 (70.3) | 347 (70.4) | 173 (62.7) | 174 (80.2) |
| Other | 5 (1.0) | 5 (1.0) | 4 (1.4) | 1 (0.5) |
| Hispanic or Latino | 49 (9.4) | 46 (9.3) | 36 (13.0) | 10 (4.6) |
| No. of pregnancies, n(%) | ||||
| None | 87 (16.8) | 80 (16.2) | 56 (20.3) | 24 (11.1) |
| 1 | 87 (16.8) | 86 (17.4) | 52 (18.8) | 34 (15.7) |
| 2 | 141 (27.2) | 131 (26.6) | 70 (25.4) | 61 (28.1) |
| 3 | 97 (18.7) | 94 (19.1) | 50 (18.1) | 44 (20.3) |
| 4 or more | 107 (20.6) | 102 (20.7) | 48 (17.4) | 54 (24.9) |
| No. of live births, n(%) | ||||
| None | 123 (23.7) | 116 (23.5) | 84 (30.4) | 32 (14.7) |
| 1 | 94 (18.1) | 91 (18.5) | 54 (19.6) | 37 (17.1) |
| 2 | 163 (31.4) | 152 (30.8) | 82 (29.7) | 70 (32.3) |
| 3 | 87 (16.8) | 84 (17.0) | 39 (14.1) | 45 (20.7) |
| 4 or more | 52 (10.0) | 50 (10.1) | 17 (6.2) | 33 (15.2) |
Sensitivity and Specificity:
Specificity and sensitivity are primary clinical measures of the performance of diagnostic tests. When comparing specificities and sensitivities across multiple diagnostic tests (Subject compared to the Predicate), the different tests are applied to groups of subjects with the same disease status for a disease or medical condition under consideration, in this case ovarian cancer. Therefore, direct comparisons to clinical performance can be drawn from one device to another, in light of the "gold standard," pathology, as discussed above. This is especially informative when the different methods are compared in matched subjects; i.e. head to head, such that direct inferences can be made about concordance or differences on a per-subject basis, rather than a similar population.
In the case of ovarian cancer triage, specificity measures the percent of benign masses (a mass is diagnosed as benign after surgery and pathology review) that are correctly predicted as low risk.
Sensitivity measures the percent of all subjects with a malignant mass (adnexal malignancy is diagnosed after surgery and pathology review) that are correctly identified as such (i.e., the percentage of subjects who are correctly predicted as high risk).
Specificity - with PA:
Overall specificity for the Subject device improved by ~14% as compared to the Predicate OVA1 device. Postmenopausal specificity improved ~23% and premenopausal
{9}------------------------------------------------
specificity improved ~8% as compared to the Predicate OVA1 device. Please see Table 4 below for the Specificity with PA.
| All EvaluableSubjects(N= 493) | Pre-menopausalWomen(N= 276) | Post-menopausalWomen(N= 217) | |
|---|---|---|---|
| All benign | |||
| PA OR Subject OVA1 Next Generation | |||
| Specificity, % | 64.8 | 67.3 | 60.9 |
| n/N | 260/ 401 | 165/ 245 | 95/ 156 |
| 95% CI | 60.0 to 69.4 | 61.2 to 72.9 | 53.1 to 68.2 |
| PA OR Predicate OVA1 | |||
| Specificity, % | 50.9 | 59.2 | 37.8 |
| n/N | 204/ 401 | 145/ 245 | 59/ 156 |
| 95% CI | 46.0 to 55.7 | 52.9 to 65.2 | 30.6 to 45.6 |
| PA OR Subject OVA1 Next Generation vs PA OR Predicate OVA1 | |||
| Difference in specificity, % | 13.97 | 8.16 | 23.08 |
| 95% CI for difference | 8.52 to 19.41 | 1.75 to 14.58 | 13.53 to 32.62 |
| Ratio of specificity | 1.275 | 1.138 | 1.610 |
| 95% CI for ratio | 1.158 to 1.410 | 1.027 to 1.267 | 1.309 to 2.010 |
Table 4 – Specificity with PA
Note: In this table, a positive combined test result is where the women has a high risk index score OR the pre-surgical PA was malignant. When the women has both a low risk index score and the pre-surgical PA was benign the combined test result is negative.
Sensitivity – with PA
Differences between the sensitivity of the Subject device and Predicate OVA1 device were clinically small. Clinically small is defined as the difference in any characteristic as a difference where; the 95% CI of the difference between the subject OVA1 Next Generation and the predicate OVA1 values bounds or contains the value 0, or; where the 95% CI of the ratio between the Subject OVA1 Next Generation and the predicate OVA1 values bounds or contains the value 1. Values where differences do not bound or contain 0. or where ratios do not bound or contain 1 will be described as "clinically significant."
For all subjects, premenopausal women, and postmenopausal women, the differences were ~2%, ~3%, and ~2%, respectively. Please see Table 5 below for the Sensitivity with PA.
{10}------------------------------------------------
| All EvaluableSubjects(N= 493) | Pre-menopausalWomen(N= 276) | Post-menopausalWomen(N= 217) | |
|---|---|---|---|
| All malignancies | |||
| PA OR Subject OVA1 Next Generation | |||
| Sensitivity, % | 93.5 | 90.3 | 95.1 |
| n/N | 86/ 92 | 28/ 31 | 58/ 61 |
| 95% CI | 86.5 to 97.0 | 75.1 to 96.7 | 86.5 to 98.3 |
| PA OR Predicate OVA1 | |||
| Sensitivity, % | 95.7 | 93.5 | 96.7 |
| n/N | 88/ 92 | 29/ 31 | 59/ 61 |
| 95% CI | 89.3 to 98.3 | 79.3 to 98.2 | 88.8 to 99.1 |
| PA OR Subject OVA1 Next Generation vs PA OR Predicate OVA1 | |||
| Difference in sensitivity, % | -2.17 | -3.23 | -1.64 |
| 95% CI for difference | -7.37 to 3.03 | -14.12 to 7.67 | -7.19 to 3.91 |
| Ratio of sensitivity | 0.977 | 0.966 | 0.983 |
| 95% CI for ratio | 0.908 to 1.043 | 0.808 to 1.136 | 0.899 to 1.066 |
Table 5 – Sensitivity with PA
Note: In this table, a positive combined test result is where the women has a high risk index score OR the pre-surgical PA was malignant. When the women has both a low risk index score and the pre-surgical PA was benign the combined test result is negative.
Specificity – Standalone:
The overall standalone specificity for the Subject device improved ~16% as compared to the Predicate OVA1 device. For premenopausal women the improvement was ~10% and for postmenopausal women the improvement was ~24%. Please see Table 6 below for a summary of the specificity data.
| Table 6 - Standalone Specificity | ||
|---|---|---|
| All EvaluableSubjects(N= 493) | Pre-menopausalWomen(N= 276) | Post-menopausalWomen(N= 217) | |
|---|---|---|---|
| Specificity | |||
| Subject OVA1 Next Generation | |||
| Specificity, % | 69.1 | 71.4 | 65.4 |
| n/N | 277/401 | 175/245 | 102/156 |
| 95% CI | 64.4 to 73.4 | 65.5 to 76.7 | 57.6 to 72.4 |
| Predicate OVA1 | |||
| Specificity, % | 53.6 | 61.6 | 41.0 |
| n/N | 215/401 | 151/245 | 64/156 |
| 95% CI | 48.7 to 58.4 | 55.4 to 67.5 | 33.6 to 48.9 |
| Subject OVA1 Next Generation vs Predicate OVA1 | |||
| Difference in specificity, % | 15.46 | 9.80 | 24.36 |
| 95% CI for difference | 9.79 to 21.13 | 3.12 to 16.47 | 14.41 to 34.31 |
{11}------------------------------------------------
| All EvaluableSubjects(N= 493) | Pre-menopausalWomen(N= 276) | Post-menopausalWomen(N= 217) | |
|---|---|---|---|
| Ratio of specificity | 1.288 | 1.159 | 1.594 |
| 95% CI for ratio | 1.172 to 1.423 | 1.047 to 1.290 | 1.304 to 1.974 |
Sensitivity - Standalone:
Standalone sensitivity was substantially equivalent for the Subject and Predicate OVA1 device. For all subjects the difference was clinically small at ~1%. For post-menopausal women both the Subject and Predicate had 91.8% performance and for pre-menopausal women there was a ~3% difference. Please see Table 7 below for a summary of the sensitivity data.
Table 7 - Standalone Sensitivity
| All EvaluableSubjects(N= 493) | Pre-menopausalWomen(N= 276) | Post-menopausalWomen(N= 217) | |
|---|---|---|---|
| All malignancies | |||
| Subject OVA1 Next Generation | |||
| Sensitivity, % | 91.3 | 90.3 | 91.8 |
| n/N | 84/92 | 28/31 | 56/61 |
| 95% CI | 83.8 to 95.5 | 75.1 to 96.7 | 82.2 to 96.4 |
| Predicate OVA1 | |||
| Sensitivity, % | 92.4 | 93.5 | 91.8 |
| n/N | 85/92 | 29/31 | 56/61 |
| 95% CI | 85.1 to 96.3 | 79.3 to 98.2 | 82.2 to 96.4 |
| Subject OVA1 Next Generation vs Predicate OVA1 | |||
| Difference in sensitivity, % | -1.09 | -3.23 | 0.00 |
| 95% CI for difference | -7.47 to 5.30 | -14.12 to 7.67 | -7.87 to 7.87 |
| Ratio of sensitivity | 0.988 | 0.966 | 1.000 |
| 95% CI for ratio | 0.909 to 1.071 | 0.808 to 1.136 | 0.898 to 1.113 |
Sensitivity - Subtype of Ovarian Malignancy
All malignancies were evaluated and stratified by five subtypes;
-
- Epithelial ovarian cancer (EOC)
-
- Non-EOC malignancies
-
- Low malignant potential (LMP)
-
- Malignancies metastatic to the ovaries, and
-
- Other, non-ovarian malignancies
The sensitivity of detection across these five ovarian cancer subtypes was retained in the Subject OVA1 Next Generation device as compared to the Predicate OVA1 device. The Subject OVA1 Next Generation device identified one less non-ovarian malignancy than the Predicate device. However, guidelines for GO referral focus on primary ovarian
{12}------------------------------------------------
cancer, since the specialized GO procedures are effective for primary ovarian cancer but clinically irrelevant for non-ovarian pelvic malignancies. In any case, when one considers the balance between specificity and sensitivity; and the improvement in the specificity of the Subject device, performance differences between the Subject and Predicate OVA1 device are de minims and clinically small. Please see Table 8 below for the Sensitivity based on Subtype of Ovarian Malignancy.
| Epithelialovariancancer | Non-EOCmalignancies | Lowmalignantpotential | Malignanciesmetastatic tothe ovaries | Othernon-ovarianmalignancies | |
|---|---|---|---|---|---|
| All evaluable subjects | |||||
| Subject OVA1 Next Generation | |||||
| Sensitivity, % | 95.0 | 80.0 | 82.4 | 100.0 | 75.0 |
| n/N | 57/60 | 4/5 | 14/17 | 6/6 | 3/4 |
| 95% CI | 86.3 to 98.3 | 37.6 to 96.4 | 59.0 to 93.8 | 61.0 to 100.0 | 30.1 to 95.4 |
| Predicate OVA1 | |||||
| Sensitivity, % | 95.0 | 80.0 | 82.4 | 100.0 | 100.0 |
| n/N | 57/60 | 4/5 | 14/17 | 6/6 | 4/4 |
| 95% CI | 86.3 to 98.3 | 37.6 to 96.4 | 59.0 to 93.8 | 61.0 to 100.0 | 51.0 to 100.0 |
| Premenopausal women | |||||
| Subject OVA1 Next Generation | |||||
| Sensitivity, % | 100.0 | 80.0 | 80.0 | 100.0 | 0.0 |
| n/N | 18/18 | 4/5 | 4/5 | 2/2 | 0/1 |
| 95% CI | 82.4 to 100.0 | 37.6 to 96.4 | 37.6 to 96.4 | 34.2 to 100.0 | 0.0 to 79.3 |
| Predicate OVA1 | |||||
| Sensitivity, % | 100.0 | 80.0 | 80.0 | 100.0 | 100.0 |
| n/N | 18/18 | 4/5 | 4/5 | 2/2 | 1/1 |
| 95% CI | 82.4 to 100.0 | 37.6 to 96.4 | 37.6 to 96.4 | 34.2 to 100.0 | 20.7 to 100.0 |
| Postmenopausal women | |||||
| Subject OVA1 Next Generation | |||||
| Sensitivity, % | 92.9 | -- | 83.3 | 100.0 | 100.0 |
| n/N | 39/42 | 0/0 | 10/12 | 4/4 | 3/3 |
| 95% CI | 81.0 to 97.5 | -- | 55.2 to 95.3 | 51.0 to 100.0 | 43.9 to 100.0 |
| Predicate OVA1 | |||||
| Sensitivity, % | 92.9 | -- | 83.3 | 100.0 | 100.0 |
| n/N | 39/42 | 0/0 | 10/12 | 4/4 | 3/3 |
| 95% CI | 81.0 to 97.5 | -- | 55.2 to 95.3 | 51.0 to 100.0 | 43.9 to 100.0 |
Table 8 - Sensitivity - Subtype of Ovarian Malignancy
Sensitivity - Stage of Primary Ovarian Malignancy:
The Subject OVA1 Next Generation device performed substantially equivalent to the Predicate OVA1 device for sensitivity across all early stage malignancies. For all evaluable subjects with early stage malignancies (stage I or II) the Subject OVA1 Next Generation device had a sensitivity of ~89% as compared to ~91% for the Predicate OVA1 device. For all evaluable subjects with late stage malignancies (stage III or IV) the sensitivity of the Subject device was 100% as compared to the Predicate OVA1 device at ~97%. Please see Table 9 below for a summary of results.
{13}------------------------------------------------
| Stage I | Stage II | AllEarly Stage(I or II) | Stage III | Stage IV | AllLate Stage(III or IV) | |
|---|---|---|---|---|---|---|
| All evaluable subjects | ||||||
| Subject OVA1 Next Generation | ||||||
| Sensitivity, % | 85.7 | 100.0 | 88.6 | 100.0 | 100.0 | 100.0 |
| n/N | 24/28 | 7/7 | 31/35 | 25/25 | 5/5 | 30/30 |
| 95% CI | 68.5 to94.3 | 64.6 to100.0 | 74.0 to95.5 | 86.7 to100.0 | 56.6 to100.0 | 88.6 to100.0 |
| Predicate OVA1 | ||||||
| Sensitivity, % | 89.3 | 100.0 | 91.4 | 96.0 | 100.0 | 96.7 |
| n/N | 25/28 | 7/7 | 32/35 | 24/25 | 5/5 | 29/30 |
| 95% CI | 72.8 to96.3 | 64.6 to100.0 | 77.6 to97.0 | 80.5 to99.3 | 56.6 to100.0 | 83.3 to99.4 |
| Premenopausal women | ||||||
| Subject OVA1 Next Generation | ||||||
| Sensitivity, % | 88.9 | 100.0 | 90.9 | 100.0 | 100.0 | 100.0 |
| n/N | 8/9 | 2/2 | 10/11 | 10/10 | 2/2 | 12/12 |
| 95% CI | 56.5 to98.0 | 34.2 to100.0 | 62.3 to98.4 | 72.2 to100.0 | 34.2 to100.0 | 75.8 to100.0 |
| Predicate OVA1 | ||||||
| Sensitivity, % | 88.9 | 100.0 | 90.9 | 100.0 | 100.0 | 100.0 |
| n/N | 8/9 | 2/2 | 10/11 | 10/10 | 2/2 | 12/12 |
| 95% CI | 56.5 to98.0 | 34.2 to100.0 | 62.3 to98.4 | 72.2 to100.0 | 34.2 to100.0 | 75.8 to100.0 |
| Postmenopausal women | ||||||
| Subject OVA1 Next Generation | ||||||
| Sensitivity, % | 84.2 | 100.0 | 87.5 | 100.0 | 100.0 | 100.0 |
| n/N | 16/19 | 5/5 | 21/24 | 15/15 | 3/3 | 18/18 |
| 95% CI | 62.4 to94.5 | 56.6 to100.0 | 69.0 to95.7 | 79.6 to100.0 | 43.9 to100.0 | 82.4 to100.0 |
| Predicate OVA1 | ||||||
| Sensitivity, % | 89.5 | 100.0 | 91.7 | 93.3 | 100.0 | 94.4 |
| n/N | 17/19 | 5/5 | 22/24 | 14/15 | 3/3 | 17/18 |
| 95% CI | 68.6 to97.1 | 56.6 to100.0 | 74.2 to97.7 | 70.2 to98.8 | 43.9 to100.0 | 74.2 to99.0 |
Table 9 – Summary of Results
ª Characterization evaluated stand-alone risk stratification versus cutoff, without regard to results of physician assessment. OVA1 Next Generation is not intended as a standalone diagnostic test.
Table 10 shows a comparison of clinical performance of OVA1 Next Generation and OVA1®, with samples collected from selected larger prospective studies. These larger prospective studies recruited premenopausal and postmenopausal women presenting with an adnexal mass requiring surgical intervention. The purpose of the comparison was to demonstrate that for samples archived less than one year prior to testing, on the OVA1 Next Generation showed equivalent clinical performance when compared to OVA1. This blinded study included twenty eight patients confirmed by pathology to have primary ovarian malignancy, along with 105 block-randomized patients with benign conditions,
{14}------------------------------------------------
selected to balance the malignancy rate within each menopausal subgroup as well as to approximate the prevalence of primary ovarian malignancies found in OVA1 pivotal clinical trials. All serum samples had been archived at -65 ℃ to -85 ℃ and tested for OVA1 Next Generation and OVA1 tests no more than one year after collection. Tables 13 and 14 show comparisons of OVA1 Next Generation and OVA1 in this set of samples.
Table 10. Comparison of OVA1 Next Generation and OVA1 performance for a selected set of serum samples from patients confirmed by pathology to have primary ovarian malignancies (N=28) or benign ovarian conditions (N=105). Samples were tested within one year of collection
| OVA1 Next | OVA1 | Difference | ||||
|---|---|---|---|---|---|---|
| Generation | (OVA1 Next Generation -OVA1) | |||||
| All subjects | ||||||
| Sensitivity % | 78.6 | 82.1 | -3.6 | |||
| n/N | 22/28 | 23/28 | 1/28 | |||
| 95% CI | 60.5 to 89.9 | 64.4 to 92.1 | -19.2 to 12.0 | |||
| Specificity % | 74.3 | 57.1 | 17.2 | |||
| n/N | 78/105 | 60/105 | 18/105 | |||
| 95% CI | 65.2 to 81.7 | 47.6 to 66.2 | 7.1 to 27.2* |
-
- performance was considered statistically different if the 95% CI of the difference did not bound or contain zero.
| Stage | N | OVA1 Next Generation% Sensitivity(n/N) | OVA1% Sensitivity(n/N) |
|---|---|---|---|
| I | 10 | 90 (9/10) | 90 (9/10) |
| II | 1 | 100.0 (1/1) | 100.0 (1/1) |
| III | 9 | 88.9 (8/9) | 88.9 (8/9) |
| IV | 3 | 66.7 (2/3) | 100.0 (3/3) |
| Not Staged | 5 | 40.0 (2/5) | 40.0 (2/5) |
Table 11. OVA1 Next Generation and OVA1 test sensitivity by stage of primary ovarian malignancy for a selected set of samples tested within one year of collection
Positive and Negative Predictive Value:
Positive and negative predictive values are proportions of positive and negative results representing true positive and true negative diagnostic (pathology) findings. Positive predictive value (PPV) is the percent of subjects with a positive test result who truly have the disease. Negative predictive value (NPV) is the percent of subjects with a negative test result who truly do not have a malignancy.
It should be noted that the PPV is not intrinsic to the test as it critically depends also on the prevalence of the disease state. However, PPV of the Subject OVA1 Next Generation and the Predicate OVA1 device can be compared here because both results were generated on the same set of patient serum samples.
{15}------------------------------------------------
Positive Predictive Value (PPV):
As with the improved standalone specificity discussed above, the standalone PPV for the Subject device improved significantly as well, with a 9% increase overall in PPV as compared to the Predicate OVA1 device. Please see Table 12 for a summary of PPV results.
| All EvaluableSubjects(N= 493) | PremenopausalWomen(N= 276) | PostmenopausalWomen(N= 217) | |
|---|---|---|---|
| Positive Predictive Value, % | |||
| Subject OVA1 Next Generation | 40.4 | 28.6 | 50.9 |
| n/N | 84/ 208 | 28/ 98 | 56/ 110 |
| 95% CI | 33.9 to 47.2 | 20.6 to 38.2 | 41.7 to 60.1 |
| Predicate OVA1 | 31.4 | 23.6 | 37.8 |
| n/N | 85/ 271 | 29/ 123 | 56/ 148 |
| 95% CI | 26.1 to 37.1 | 16.9 to 31.8 | 30.4 to 45.9 |
| Subject OVA1 Next Generation vs Predicate OVA1 | |||
| Difference in PPV, % | 9.02 | 4.99 | 13.07 |
| 95% CI for difference | 5.02 to 13.02 | 0.24 to 9.75 | 6.88 to 19.27 |
| Ratio of PPV | 1.288 | 1.212 | 1.345 |
| 95% CI for ratio | 1.155 to 1.435 | 1.018 to 1.443 | 1.171 to 1.546 |
| 95% CI | 93.6 to 98.5 | 95.4 to 99.6 | 84.1 to 96.9 |
Table 12 – Positive Predictive Values
Negative Predictive Value (NPV):
The NPV for the Subject device was substantially equivalent to the Predicate OVA1 device at ~97% for both. Please see Table 13 for a summary of NPV results.
Table 13 – Negative Predictive Values
| All EvaluableSubjects(N= 493) | Pre-menopausalWomen(N= 276) | Post-menopausalWomen(N= 217) | |
|---|---|---|---|
| Negative Predictive Value, % | |||
| Subject OVA1 Next Generation | 97.2 | 98.3 | 95.3 |
| n/N | 277/ 285 | 175/ 178 | 102/ 107 |
| 95% CI | 94.6 to 98.6 | 95.2 to 99.4 | 89.5 to 98.0 |
| Predicate OVA1 | 96.8 | 98.7 | 92.8 |
| n/N | 215/ 222 | 151/ 153 | 64/ 69 |
| 95% CI | 93.6 to 98.5 | 95.4 to 99.6 | 84.1 to 96.9 |
| Subject OVA1 Next Generation vs Predicate OVA1 | |||
| Difference in NPV, % | 0.35 | -0.38 | 2.57 |
| 95% CI for difference | -1.97 to 2.66 | -2.38 to 1.62 | -3.20 to 8.35 |
| Ratio of NPV | 1.004 | 0.996 | 1.028 |
| 95% CI for ratio | 0.980 to 1.028 | 0.976 to 1.017 | 0.966 to 1.093 |
{16}------------------------------------------------
Please see Table 14 below for the Specificity, Sensitivity, PPV, NPV - standalone values. Please see Table 15 below for values with PA.
| Comparison of the standalone performance ofthe subject OVA1 Next Generation to the Predicate OVA1 | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Sensitivity | Specificity | Positive Predictive Value | Negative Predictive Value | |||||||||
| All | Pre | Post | All | Pre | Post | All | Pre | Post | All | Pre | Post | |
| OVA1 NextGeneration(95% CI) | 91.3%83.8 to95.5 | 90.3%75.1 to96.7 | 91.8%82.2 to96.4 | 69.1%64.4 to73.4 | 71.4%65.5 to76.7 | 65.4%57.6 to72.4 | 40.4%39.9 to47.2 | 28.6%20.6 to38.2 | 50.9%41.7 to60.1 | 97.2%94.6 to98.6 | 98.3%95.2 to99.4 | 95.3%89.5 to98.0 |
| OVA1(95% CI) | 92.4%85.1 to96.3 | 93.5%79.3 to98.2 | 91.8%82.2 to96.4 | 53.6%47.8 to58.4 | 61.6%55.4 to67.5 | 41.0%33.6 to48.9 | 31.4%26.1 to37.1 | 23.6%16.9 to31.8 | 37.8%30.4 to45.9 | 96.8%93.6 to98.5 | 98.7%95.4 to99.6 | 92.8%84.1 to96.9 |
| Table 14 - Sensitivity, Specificity, PPV, NPV - Standalone | ||
|---|---|---|
| Table 15 - Sensitivity, Specificity, PPV, NPV, with PA |
|---|
| Sensitivity | Specificity | Positive PredictiveValue | Negative PredictiveValue | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| All | Pre | Post | All | Pre | Post | All | Pre | Post | All | Pre | Post | |
| OVA1 NextGeneration(95% CI) | 93.5%86.5 to97.0 | 90.3%75.1 to96.7 | 95.1%86.5 to98.3 | 64.8%60.0 to69.4 | 67.3%61.2 to72.9 | 60.9%53.1 to68.2 | 37.9%31.8 to44.3 | 25.9%18.6 to34.9 | 48.7%39.9 to57.6 | 97.7%95.2 to99.0 | 98.2%94.9 to99.4 | 96.9%91.4 to99.0 |
| OVA1(95% CI) | 95.7%89.3 to98.3 | 93.5%79.3 to98.2 | 96.7%88.8 to99.1 | 50.9%46.0 to55.7 | 59.2%52.9 to65.2 | 37.8%30.6 to45.6 | 30.9%25.8 to36.5 | 22.5%16.1 to30.4 | 37.8%30.6 to45.6 | 98.1%95.2 to99.2 | 98.6%95.2 to99.6 | 96.7%88.8 to99.1 |
Likelihood ratios (LR):
There are two likelihood ratios; The LR+ represents the probability of a person who has a malignancy testing positive divided by the probability of a person who does not have the disease testing positive.
Conversely, the LR- represents the probability of a person who has the disease testing negative divided by the probability of a person who does not have the disease testing negative.
A likelihood ratio of greater than 1 indicates the test result is associated with the disease. A likelihood ratio less than 1 indicates that the result is associated with absence of the disease.
{17}------------------------------------------------
The Subject device demonstrated an improved LR+ as compared to the Predicate device. as well as an improved LR-. Please see Table 16 for a summary of the Likelihood ratios.
| All EvaluableSubjects(N= 493) | Pre-menopausalWomen(N= 276) | Post-menopausalWomen(N= 217) | |
|---|---|---|---|
| Positive Likelihood Ratio | |||
| Subject OVA1 Next Generation | 2.953 | 3.161 | 2.652 |
| 95% CI | 2.518 to 3.463 | 2.514 to 3.975 | 2.111 to 3.332 |
| Predicate OVA1 | 1.992 | 2.438 | 1.557 |
| 95% CI | 1.766 to 2.247 | 2.029 to 2.930 | 1.339 to 1.810 |
| Subject OVA1 Next Generation vs Predicate OVA1 | |||
| Ratio of positive LR | 1.482 | 1.297 | 1.704 |
| 95% CI for ratio | 1.242 to 1.769 | 1.012 to 1.661 | 1.311 to 2.214 |
| Negative Likelihood Ratio | |||
| Subject OVA1 Next Generation | 0.126 | 0.135 | 0.125 |
| 95% CI | 0.065 to 0.245 | 0.046 to 0.398 | 0.054 to 0.293 |
| Predicate OVA1 | 0.142 | 0.105 | 0.200 |
| 95% CI | 0.069 to 0.291 | 0.027 to 0.401 | 0.084 to 0.472 |
| Subject OVA1 Next Generation vs Predicate OVA1 | |||
| Ratio of negative LR | 0.887 | 1.294 | 0.627 |
| 95% CI for ratio | 0.374 to 2.106 | 0.382 to 4.383 | 0.188 to 2.097 |
Table 16 - Positive and Negative Likelihood Ratios
Clinical Specificity -Healthy Women Study:
A Clinical Specificity Study on healthy women was conducted to determine test result reference intervals of the Subject device on healthy women. Healthy women are not within the definition of the intended use population and would therefore not be considered for testing according to labeling. Nevertheless, the characterization study data will help physicians understand the representative range and distribution of OVA 1 Next Generation test results in the intended use population, and to answer questions on how they might differ or compare with healthy women.
Study Design:
Study subjects were healthy women ages 18 to 92 years old. "Healthy" was defined as: no viral or bacterial infection, no substance abuse, no chronic disease state (for example, diabetes, lupus, or hepatitis), and no diagnosis of malignancy in the last 10 years, with the exception of non-melanoma skin cancer.
{18}------------------------------------------------
A total of 152 results were obtained from 68 premenopausal and 84 postmenopausal subjects (there were no unevaluable results). Single samples were run using the same calibrator reference curve, the same kit reagents lots, and the same control lots for the entire study duration. Serum concentrations of each protein biomarker were determined using the cobas 6000 instrument. Each operator generated the test result for the test he or she ran using the Subject OvaCalc software.
Study Results:
The Subject OVA1 Next Generation test demonstrated approximately 50% reduction in the test-positive subjects, as compared to the Predicate device. For example, ~ 13% of premenopausal subjects were test-positive with the Subject OVA1 Next Generation compared with 29% for the Predicate OVA1; and ~17% of postmenopausal subjects were test-positive for Subject OVA1 Next Generation compared with ~33% for the Predicate OVA 1. These improvements are consistent with the design intent of lowering the percent of test-positive non-malignant subjects. However, no claim is made since this study was designed to inform physicians on the representative range and distribution of OVA 1 Next Generation test results in healthy women, and to answer questions on how they might differ or compare with the intended use population. Therefore, we conclude the results to be substantially equivalent.
Please see Table 17 below for a summary of the clinical specificity of the Subject and Predicate OVA1 device. The Study demonstrated that the performance of the Subject device is substantially equivalent to the Predicate device.
| All HealthySubjects | Pre-menopausalWomen | Post-menopausalWomen | |
|---|---|---|---|
| N | 152 | 68 | 84 |
| Mean (SD) | 3.94 (0.984) | 3.72 (0.938) | 4.12 (0.989) |
| Median | 3.90 | 3.60 | 4.05 |
| Range (min, max) | 2.2, 7.1 | 2.2, 6.1 | 2.5, 7.1 |
| Percentile (5% to 95%) | 2.5, 5.9 | 2.4, 5.3 | 2.9, 5.9 |
| OVA1 Next Generation Result, n (%)* | |||
| Positive | 23 (15.1%) | 9 (13.2%) | 14 (16.7%) |
| Negative | 129 (84.9%) | 59 (86.8%) | 70 (83.3%) |
| Predicate OVA1 | |||
| N | 147 | 69 | 78 |
| Positive | 46 (31.3%) | 20 (29.0%) | 26 (33.3%) |
Table 17 - Clinical Specificity -Healthy Women
Clinical Specificity - Other Cancers and Disease States:
A Clinical Specificity Study was conducted to determine the representative range of the Subject OVA1 Next Generation test results from women with other (non-ovarian) cancers and benign conditions including the following:
{19}------------------------------------------------
- Bladder cancer ●
- Breast cancer ●
- Cervical cancer
- Colon cancer
- Endometrial cancer
- Leukemia
- . Lung cancer
- Lymphoma
- Autoimmune disease
- Cardiac disease
- Diabetes
- . Endometriosis
- Hepatitis
- Kidney disease ●
- . Pregnancy
Similar to the Clinical Specificity Study - Healthy Women, this characterization study was conducted to help physicians better understand and compare the representative range and distribution of OVA1 Next Generation test results in the intended use population to various benign and malignant diseases they may encounter when managing patients.
Study Design:
Four hundred and one single samples were run using the Subject OVA1 Next Generation device, using the same calibrator reference curve, the same kit reagent lots, and the same control lots for the entire study duration. Serum concentrations of each protein biomarker were determined using the cobas 6000 instrument. Each operator generated the test result for the test he or she ran using the Subject OvaCalc software.
Study Results:
As previously discussed; the nature of the test is to assign a high risk or low risk of malignancy. It is therefore not unexpected that the Subject device might detect certain other (non-ovarian cancer) cancers, disease states and conditions - as previously shown for the Predicate OVA1 device. Indeed the results of the Clinical Specificity for Other Cancers and Diseases indicate that other cancers and disease conditions can yield positive OVA1 Next Generation test result in some cases. However, the clinical study was not designed and Sponsor is not making any claims regarding other disease states. This study was designed simply to characterize and report the expected range and distribution of Subject OVA1 Next Generation test results for various benign and malignant diseases they may encounter when managing patients, relative to the Predicate OVA1 device.
{20}------------------------------------------------
Nevertheless the risk of detecting other non-ovarian cancers is mitigated by including physician assessment as part of the intended use for the Subject device. Lastly, both the Predicate OVA1 device has been on the market for several years without any reported events related to misdiagnosis versus other cancers, evidencing that the PA, as required for proper use of the Subject device, is adequate to mitigate risk related to other cancers.
The characterization study results and comparison to the Predicate OVA1 characterization results support a conclusion that the performance of the Subject OVA1 Next Generation device is substantially equivalent to the Predicate device. Please see Table 18 for a summary of results.
{21}------------------------------------------------
Premarket Notification – OVA1 Next Generation
| AllEvaluableSubjects | BladderCancer | BreastCancer | CervicalCancer | ColonCancer | EndometrialCancer | Leukemia | LungCancer | Lymphoma | |
|---|---|---|---|---|---|---|---|---|---|
| N | 221 | 20 | 40 | 20 | 40 | 40 | 11 | 40 | 10 |
| OVA1 Next Generation Result, n (%) | |||||||||
| Positive | 103(46.6) | 10(50.0) | 6(15.0) | 13(65.0) | 18(45.0) | 20(50.0) | 10(90.9) | 18(45.0) | 8(80.0) |
| Negative | 118(53.4) | 10(50.0) | 34(85.0) | 7(35.0) | 22(55.0) | 20(50.0) | 1(9.1) | 22(55.0) | 2(20.0) |
| Specificity, % | 53.4 | 50.0 | 85.0 | 35.0 | 55.0 | 50.0 | 9.1 | 55.0 | 20.0 |
| AllEvaluableSubjects | BladderCancer | BreastCancer | CervicalCancer | ColonCancer | EndometrialCancer | Leukemia | LungCancer | Lymphoma | |
| N | NA | 16 | 45 | 12 | 40 | 44 | 10 | 13 | 13 |
| Predicate OVA1 | |||||||||
| Positive | NA | 6 | 11 | 8 | 18 | 15 | 9 | 3 | 6 |
| Negative | NA | 10 | 34 | 4 | 22 | 29 | 1 | 10 | 7 |
| Specificity % | NA | 62.5 | 75.6 | 33.3 | 55.0 | 65.9 | 10 | 76.9 | 53.8 |
| AllEvaluableSubjects | AutoimmuneDisease | CardiacDisease | Diabetes | Endo-metriosis | Hepatitis | KidneyDisease | PregnantWomen | ||
| N | 180 | 20 | 20 | 40 | 40 | 20 | 20 | 20 | |
| OVA1 Next Generation Result, n (%) | |||||||||
| Positive | 99(55.0) | 11(55.0) | 15(75.0) | 14(35.0) | 11(27.5) | 11(55.0) | 18(90.0) | 19(95.0) | |
| Negative | 81(45.0) | 9(45.0) | 5(25.0) | 26(65.0) | 29(72.5) | 9(45.0) | 2(10.0) | 1(5.0) | |
| Specificity, % | 45.0 | 45.0 | 25.0 | 65.0 | 72.5 | 45.0 | 10.0 | 5.0 | |
| AllEvaluableSubjects | AutoimmuneDisease | CardiacDisease | Diabetes | Endo-metriosis | Hepatitis | KidneyDisease | PregnantWomen | ||
| N | NA | 10 | 12 | 40 | 40 | 10 | 12 | 10 | |
| Predicate OVA1 | |||||||||
| Positive | NA | 5 | 7 | 10 | 17 | 3 | 12 | 3 | |
| Negative | NA | 5 | 5 | 30 | 23 | 10 | 0 | 7 |
Table 18 - Clinical Specificity - Other Cancers and Diseases States
{22}------------------------------------------------
Analytical Performance Validation:
Precision Study:
The Precision Study established total precision for the risk score algorithm and individual analyte measurements in the subject device. End users were instructed to follow the package insert from the manufacturer of each respective immunoassay.
Study Design:
The sample set consisted of five pooled serum samples were numbered 6 – 10 for Subject device and were 1 – 5 for Predicate OVA1 submission) spanning the Subject device test result range (low test result, high test result, and close to the cutoff at 5.0), as well as two control levels for each assay per run.
The five samples were tested using two separate aliquots on two runs on each day. Multiple operators were used to run the five samples were run over 20 separate days and used the same kit reagents lots and the same control lots for the entire study.
Serum concentrations of each protein biomarker were determined using the cobas 6000 instrument. Each operator generated the test result for the test they ran using the Subject OvaCalc software.
A total of 400 results were obtained - on each of the 20 days, the same five samples were analyzed in duplicate for two runs per day. There were no unevaluable results.
Study Results:
The overall coefficient of variation (%CV) was 1.54% across all days and pools, which demonstrates that the errors of measurement were well within the acceptable limits (< 10%) established in the Product Design Specification.
In addition, the precision for the Subject OVA1 Next Generation test was notably improved from the Predicate OVA1, as seen in comparison in Table 19 below. The %CVs for the Subject device are equivalent to or less than the %CVs for the Predicate OVA1 test. Pool 6 exhibited little variance. This was likely due to that fact that biomarker concentrations are at limits such that their individual variation does not affect the calculated OVA1 Next Generation test result.
Analysis of Results:
The overall %CV for the Subject device was 1.54% across all days and pools. The overall %CV for the Predicate OVA1 device in contrast was 4.09% (please refer to OVA1 precision data in K081754).
The Study demonstrated that the performance of the Subject device is substantially equivalent to the Predicate device.
{23}------------------------------------------------
| Serum Pool | ||||||
|---|---|---|---|---|---|---|
| OVA1 Next GenerationTest Value | 1 | 2 | 3 | 4 | 5 | All Pools |
| N | 80 | 80 | 80 | 80 | 80 | 400 |
| Mean | 8.50 | 8.16 | 5.08 | 4.11 | 3.30 | 5.83 |
| SD of Error | 0.000 | 0.055 | 0.161 | 0.085 | 0.065 | 0.090 |
| %CV of overall error | 0.00 | 0.67 | 3.16 | 2.06 | 1.95 | 1.54 |
| Table 19 - Precision Study, subject device compared to predicate OVA1 | ||||
|---|---|---|---|---|
| Serum Pool | ||||||
|---|---|---|---|---|---|---|
| OVA1 Test Value | 1 | 2 | 3 | 4 | 5 | All Pools |
| N | 80 | 80 | 80 | 80 | 80 | 400 |
| Mean | 2.74 | 3.39 | 3.74 | 4.69 | 9.94 | 4.90 |
| SD of Error | 0.091 | 0.159 | 0.192 | 0.349 | 0.098 | 0.200 |
| %CV of overall error | 3.31 | 4.69 | 5.11 | 7.43 | 0.98 | 4.09 |
Reproducibility Study:
Study Design:
The study was run using the same five pooled serum samples used in the Precision Study. Pooled samples (low test result, high test result, and close to the cutoff at 5.0) were run over six, non-consecutive days at three sites and two operators as per site.
A total of 360 test results were obtained (one result was rejected due to operator error: operator used APO B, rather than APO A1) over the six days.
Study Results:
The Subject device showed very little variability over different sites, days, operators and runs. The overall %CV including all sites was 1.63% as compared to the Predicate OVA1 device with an overall %CV of 2.80% across all sites (please reference reproducibility study in K081754). Please see Table 20 below for a summary of reproducibility study results.
The Study demonstrated that the performance of the Subject device is substantially equivalent to the Predicate device.
{24}------------------------------------------------
| OVA1 Next Generation Test Value | Total - All Pools |
|---|---|
| SD | 0.10 |
| %CV | 1.6 |
| Predicate OVA1 | |
| SD | 0.27 |
| %CV | 2.8 |
Table 20 - Reproducibility Study, subject device compared to predicate device (OVA1)
Stability Study:
A sample stability study was conducted to verify specimen sample stability for use with the Subject device. The study duration and temperatures represented instructions for use that include shipping and laboratory storage prior to running the assays and provide estimates of specimen stability for the Subject device test result and individual biomarker assays.
Study Design:
The sample set consisted of the seven pooled serum samples (Pools A-G) run for the other analytical studies, with two control levels for each assay per run.
Four independent aliquots of each pool's samples were removed from the freezer, thawed on the bench top, and then placed in the refrigerator (2-8°C) for time points from zero to eight days. Samples were run using the same calibrator reference curve, the same kit reagents lots, and the same control lots for the entire study duration.
Serum concentrations of each protein biomarker were determined using the cobas 6000 instrument. Each operator generated the test result for the ran using the Subject OvaCalc software.
Study Results:
Results from the Sample Stability study confirmed that the specimen sample provides stable OVA1 Next Generation test results over eight days of storage. For all pools, results at eight days of storage between 2℃ and 8℃ were within 10% of the initial (Day 0) value, which meets the acceptance criteria established in the Product Design Specification.
The Study demonstrated that the performance of the Subject device is substantially equivalent to the Predicate device. Please see Table 21 below for a summary of stability study results.
{25}------------------------------------------------
Premarket Notification – OVA1 Next Generation
| Table 21 - Sample Stability | ||||||
|---|---|---|---|---|---|---|
| Analysis Metric | Day 0 | Day 2 | Day 6 | Day 8 | Day 9 | |
| Pool A | Mean | 7.90 | 7.90 | 7.90 | 7.90 | 7.90 |
| SD | 0.000 | 0.000 | 0.000 | 0.000 | 0.000 | |
| Mean Change | - | 0.00 | 0.00 | 0.00 | 0.00 | |
| %Mean Change from Day 0 | - | 0.0% | 0.0% | 0.0% | 0.0% | |
| 95% CI of change | - | 0.00 to 0.00 | 0.00 to 0.00 | 0.00 to 0.00 | 0.00 to 0.00 | |
| Pool B | Mean | 8.30 | 8.35 | 8.35 | 8.30 | 8.30 |
| SD | 0.000 | 0.071 | 0.071 | 0.000 | 0.000 | |
| Mean Change | - | 0.05 | 0.05 | 0.00 | 0.00 | |
| %Mean Change from Day 0 | - | 0.6% | 0.6% | 0.0% | 0.0% | |
| 95% CI of change | - | -0.02 to 0.12 | -0.02 to 0.12 | -0.07 to 0.07 | -0.07 to0.07 | |
| Pool C | Mean | 6.00 | 6.00 | 6.05 | 6.00 | 6.10 |
| SD | 0.100 | 0.141 | 0.071 | 0.141 | 0.141 | |
| Mean Change | - | 0.00 | 0.05 | 0.00 | 0.10 | |
| %Mean Change from Day 0 | - | 0% | 0.8% | 0% | 1.7% | |
| 95% CI of change | - | -0.22 to 0.22 | -0.17 to 0.27 | -0.22 to 0.22 | -0.12 to0.32 | |
| Pool D | Mean | 7.96 | 7.85 | 7.90 | 7.95 | 8.00 |
| SD | 0.055 | 0.071 | 0.000 | 0.071 | 0.000 | |
| Mean Change | - | -0.11 | -0.06 | -0.01 | 0.04 | |
| %Mean Change from Day 0 | - | -1.4% | -0.8% | -0.1% | 0.5% | |
| 95% CI of change | - | -0.21 to -0.01 | -0.16 to 0.04 | -0.11 to 0.09 | -0.06 to0.14 | |
| Pool E | Mean | 8.20 | 8.20 | 8.20 | 8.20 | 8.20 |
| SD | 0.000 | 0.000 | 0.000 | 0.000 | 0.000 | |
| Mean Change | - | 0.0 | 0.0 | 0.0 | 0.0 | |
| %Mean Change from Day 0 | - | 0.0% | 0.0% | 0.0% | 0.0% | |
| 95% CI of change | - | 0.00 to 0.00 | 0.00 to 0.00 | 0.00 to 0.00 | 0.00 to 0.00 | |
| Pool F | Mean | 4.00 | 4.00 | 4.05 | 4.05 | 4.05 |
| SD | 0.071 | 0.000 | 0.071 | 0.354 | 0.071 | |
| Mean Change | - | 0.00 | 0.05 | 0.05 | 0.05 | |
| %Mean Change from Day 0 | - | 0.0% | 1.3% | 1.3% | 1.3% | |
| 95% CI of change | - | -0.27 to 0.27 | -0.22 to 0.32 | -0.22 to 0.32 | -0.22 to0.32 | |
| Pool G | Mean | 3.10 | 2.80 | 3.05 | 2.90 | 2.95 |
| SD | 0.100 | 0.000 | 0.354 | 0.000 | 0.212 | |
| Mean Change | - | -0.30 | -0.05 | -0.20 | -0.15 | |
| %Mean Change from Day 0 | - | -9.7% | -1.6% | -6.5% | -4.8% | |
| 95% CI of change | - | -0.61 to 0.01 | -0.36 to 0.26 | -0.51 to 0.11 | -0.46 to0.16 |
Table 21 - Sample Stability
{26}------------------------------------------------
Interference Study:
An interference study was conducted to screen common interfering substances for potential effects to the Subject OVA1 Next Generation test. The study was adapted from and is consistent with the Clinical and Laboratory Standards Institute (CLSI) Interference Testing in Clinical Chemistry; Approved Guideline-Second Edition (EP07-A2). Additionally, the study was designed to be consistent with the Predicate OVA1 submission interference study in order to evaluate possible interfering substance bias for establishing substantial equivalence to the Predicate OVA1.
Study Design:
The study tested three serum sample pools (Pools 6, 8 and 10) spanning the OVA1 Next Generation test result range (low, close to the cutoff, and high), each spiked with various levels of five potential interfering substances. These were three of the five pools tested in the other analytical studies above. Vehicle control samples without potential interfering substances but with the same amount of solvent were also tested. Potential interfering substances and concentrations tested were as listed in Table 22 below.
| Substance | Concentrations Tested |
|---|---|
| Hemoglobin | 5.0 g/L |
| Hemoglobin | 9.0 g/L |
| Bilirubin, conjugated | 0.3 g/L |
| Bilirubin, conjugated | 0.9 g/L |
| Bilirubin, unconjugated | 0.3 g/L |
| Bilirubin, unconjugated | 0.9 g/L |
| Triglycerides | 2.0 g/L |
| Triglycerides | 4.6 g/L |
| Triglycerides | 10.0 g/L |
| Rheumatoid factor | 250 IU/mL |
| Rheumatoid factor | 1000 IU/mL |
Four replicates of each experimental group were prepared, each from an independent sample aliquot. Two control levels for each assay were run prior to each sample run. Samples were run using the same calibrator reference curve, the same kit reagents lots, and the same control lots for the entire study duration. Serum concentrations of each protein biomarker were determined using the cobas 6000 instrument. Each operator generated the test result for the test he or she ran using the Subject OvaCalc software.
Study Results:
All potential interfering substances tested were within acceptable limits established in the Product Design Specification and consistent with the assay manufacturer's instructions for use for interfering substances. For the purpose of this study, an interfering substance showed no effect if the 95% CI of the treated pooled sample was within the 10% margin of the untreated control.
{27}------------------------------------------------
Method Comparison:
The comparison of performance for risk stratification between dual assessment of PA with OVA1 Next Generation (PA + OVA1 Next Generation) and dual assessment of PA with OVA1 Test (PA + OVA1 Test) for all evaluable subjects, and malignant and benign cases as determined by pathology is summarized in Table 23. Results showed that PA+OVA1 Next Generation and PA+OVA1 Test agreed on 187 high risk cases and 168 low risk cases for a total percentage agreement of 355 of 493 cases, or 72%. For risk stratification agreement of malignant cases, PA+OVA1 Next Generation and PA+OVA1 Test agreed on 88 high risk cases and 2 low risk cases (misclassified) for a total percentage agreement of 86 of 92 cases, or 93.5%. For benign cases, PA+OVA1 Next Generation and PA+OVA1 Test agreed on the classification of 166 of 401 benign cases (41% of all benign cases) but incorrectly classified 103 benign cases as high risk (26%). PA+OVA1 Next Generation correctly classified 94 benign cases as low risk which PA+OVA1 Test classified as high risk (23% of benign cases correctly classified by PA+OVA1 Next Generation but not PA+OVA1 Test). PA+OVA1 Next Generation incorrectly classified 38 benign cases as high risk which PA+OVA1 Test classified as low risk (9% of benign cases correctly classified by PA+OVA1 Test but not PA+OVA1 Next Generation). Overall, PA+OVA1 Next Generation showed a net improvement of 14% in the classification of benign subjects.
| All Evaluable Subjects | |||
|---|---|---|---|
| Dual assessment of PA withOVA1 Next Generation | Dual assessment of PA withOVA1 Test | Total | |
| High risk | Low risk | ||
| High risk | 187 | 40 | 227 |
| Low risk | 98 | 168 | 266 |
| Total | 285 | 208 | 493 |
| Positive Percent Agreement: 65.6% (187/285) 95% CI: 59.9% to 70.9%Negative Percent Agreement: 80.8% (168/208) 95% CI: 74.9% to 85.5%Total Percent Agreement: 72.0% (355/493) 95% CI: 67.9% to 75.8% | |||
| Malignant Cases | |||
| Dual assessment of PA withOVA1 Next Generation | Dual assessment of PA withOVA1 Test | Total | |
| High risk | Low risk | ||
| High risk | 84 | 2 | 86 |
| Low risk | 4 | 2 | 6 |
| Total | 88 | 4 | 92 |
| Total Percent Agreement: 93.5% (86/92) 95% CI: 86.5% to 97.0% |
Table 23 - Dual assessment of PA with OVA1 Next Generation versus dual assessment of PA with OVA1 Test
{28}------------------------------------------------
| Benign Cases | ||||
|---|---|---|---|---|
| Dual assessment of PA withOVA1 Test | ||||
| High risk | Low risk | Total | ||
| Dual assessment of PA withOVA1 Next Generation | High risk | 103 | 38 | 141 |
| Low risk | 94 | 166 | 260 | |
| Total | 197 | 204 | 401 | |
| Total Percent Agreement: 67.1% (269/401) 95% CI: 62.3% to 71.5% |
Conclusions:
The subject OVA1 Next Generation Test device is substantially equivalent in indications for use, intended use, and functionality to the predicate device cleared in K081754.
OVA1 510(k) K081754
All data indicates that the device will perform as intended.
§ 866.6050 Ovarian adnexal mass assessment score test system.
(a)
Identification. An ovarian/adnexal mass assessment test system is a device that measures one or more proteins in serum or plasma. It yields a single result for the likelihood that an adnexal pelvic mass in a woman, for whom surgery is planned, is malignant. The test is for adjunctive use, in the context of a negative primary clinical and radiological evaluation, to augment the identification of patients whose gynecologic surgery requires oncology expertise and resources.(b)
Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Ovarian Adnexal Mass Assessment Score Test System.” For the availability of this guidance document,see § 866.1(e).(c)
Black box warning. Under section 520(e) of the Federal Food, Drug, and Cosmetic Act these devices are subject to the following restriction: A warning statement must be placed in a black box and must appear in all advertising, labeling, and promotional material for these devices. That warning statement must read: