K Number
K150404
Date Cleared
2016-06-23

(491 days)

Product Code
Regulation Number
880.5725
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The QiF Blood and Fluid Warmer device is intended for warming blood, blood products, and intravenous fluids prior to administration. It is intended to be used by healthcare professionals in hospital, clinics and field environments, to help prevent hypothermia.

Device Description

The Quality in Flow Ltd. QiF Blood and Fluid Warmer device is a portable, sterile Fluid Path, in-line Blood and Fluid Warmer. The device is comprised of a Base Unit and a sterile disposable cartridge (Disposable Unit). The device is located between the fluid container (intravenous fluid, blood or blood product) and treated patient, outside of the patient's body. The Disposable Unit is composed of a plastic oval box encasing a spiral stainless steel heat exchanger tube. The Base Unit contains Firmware (software) and electronics (HW). The base Unit controls the performance of the system and the fluid outflow temperature. The power source is a rechargeable detachable battery located within the Base Unit. The disposable unit cartridge (blood/fluid) is provided with inlet and outlet luers that connect to standard IV tubing.

AI/ML Overview

Here's a summary of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Device: QiF Blood and Fluid Warmer

1. Table of Acceptance Criteria and Reported Device Performance

Test/CharacteristicAcceptance Criteria (where explicitly stated)Reported Device Performance
Hemolysis StudyHemolysis of up to 1% (in compliance with ASTM 2172:[(2002) Reapproved 2011)])All samples met the acceptance criteria (hemolysis of up to 1%). Tested under normal and worse-case conditions (stop flow, high flow rate, air bubble), and with old RBC blood product at high and low flow rates. No protocol deviations.
Temperature Indication AccuracyDetermined according to product design input requirement specifications (not explicitly stated in document)All tests met the acceptance criteria.
Outflow Temperature ProfileDetermined according to product design input requirement specifications (not explicitly stated in document)All tests met the acceptance criteria. Tested under normal and abnormal (stop flow, high flow rate, air bubble) flow scenarios.
Fluid Path Integrity under PressureDetermined according to product design input requirement specifications (in compliance with ASTM 2172:[(2002) Reapproved 2011)])All tests met the acceptance criteria.
Battery CapacityDetermined according to product design input requirement specifications (in terms of minimal heated volume)All tests met the acceptance criteria.
Leachability TestingNo significant change in the concentration of new materials not present in the fluid prior to heating; no significant change in the concentration of any of the detected materials between control sample and test article.All test samples met the acceptance criteria. Analysis performed by external certified lab. No protocol deviations.
Software V&VCompliance with "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (May 11, 2005) and EN IEC 62304:2006/AC:2008.Performed in compliance. Software level of concern is moderate.
BiocompatibilityCompliance with ISO 10993-5:2009, 10993-10:2010, 10993-11:2006, ASTM F756:2008Testing performed. (Implied compliance by "performed the following testing... to evaluate the QiF Blood and Fluid Warmer device and demonstrate substantial equivalence")
Sterilization & Shelf LifeCompliance with ISO 11135-1:2007, ISO 11737-1:2006, ISO 10993-7:2008, and FDA CDRH, Shelf Life of Medical Devices guidance, April 1991Validation conducted in compliance. (Implied device meets criteria by "conducted in compliance")
Electrical Safety & EMCCompliance with IEC 60601-1:2012, 60601-1-2:2007 ed. 3.0, 60601-1-6:2010, 60601-1-11:2010Testing performed. (Implied compliance by "performed the following testing...")
Battery TestingCompliance with IEC 62133:2012Testing performed. (Implied compliance by "performed the following testing...")

Study That Proves the Device Meets Acceptance Criteria:

The document describes a series of bench testing and design control activities performed by Quality In Flow Ltd. to evaluate the QiF Blood and Fluid Warmer device. These tests were conducted according to established international standards and internal product design input requirement specifications.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: The document does not explicitly state the sample sizes used for each specific test (e.g., number of units tested for hemolysis, temperature accuracy). It refers to "all samples" meeting criteria for hemolysis and leachability, suggesting multiple samples were indeed tested.
  • Data Provenance: The studies were performed by Quality in Flow Ltd., the device manufacturer. The data is prospective, as it was generated specifically for the 510(k) submission to demonstrate device performance. The country of origin of the data is Israel, where Quality in Flow Ltd. is based.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

This information is not provided in the document. The "ground truth" for these performance tests is typically the measured physical properties against objective standards or specifications, rather than expert consensus on subjective interpretations. For example, for hemolysis, the ground truth is the chemical measurement of free hemoglobin in the fluid.

4. Adjudication Method for the Test Set:

Not applicable. The tests performed are objective performance measurements against defined criteria, not subjective assessments requiring adjudication.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This device is a medical warming device, not an imaging or diagnostic AI device that would involve human readers.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not explicitly stated in these terms. However, the performance data presented (hemolysis, temperature accuracy, etc.) represents the intrinsic performance of the device's hardware and software (algorithm only) without direct human intervention during the measurement process, aside from setting up and operating the device. The software verification and validation is a key component of this "algorithm only" assessment.

7. The Type of Ground Truth Used:

The ground truth used for these performance tests is primarily objective physical and chemical measurements against:

  • Established international standards (e.g., ASTM 2172 for hemolysis and fluid path integrity, ISO 10993 for biocompatibility, IEC 60601 for electrical safety).
  • The manufacturer's own product design input requirement specifications (e.g., for temperature indication accuracy, outflow temperature profile, battery capacity).

8. The Sample Size for the Training Set:

Not applicable. This is a medical device (fluid warmer), not an AI/ML model that requires a training set in the conventional sense. The "training" for such a device would be its iterative design, engineering, and prototyping phases.

9. How the Ground Truth for the Training Set was Established:

Not applicable, as there is no "training set" in the context of an AI/ML model for this device. The design and performance specifications were established through engineering principles, regulatory requirements, and user needs for a medical fluid warmer.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are black and have a flowing, abstract design. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 23, 2016

Ouality in Flow, Ltd. % John Smith, M.D., J.D. Partner Hogan Lovells US LLP 555 Thirteenth Street NW Washington, DC 20004

Re: K150404

Trade/Device Name: QiF Blood and Fluid Warmer Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: LGZ, BSB Dated: April 22, 2016 Received: April 22, 2016

Dear Dr. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - John Smith, M.D., J.D.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Image /page/1/Picture/8 description: The image shows a signature and the name Tina Kiang. The signature is a stylized, looping design to the left of the name. Below the name, there is a small "-s".

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known) K150404

Device Name

QiF Blood and Fluid Warmer Indications for Use (Describe)

The QiF Blood and Fluid Warmer device is intended for warming blood, blood products, and intravenous fluids prior to administration. It is intended to be used by healthcare professionals in hospital, clinics and field environments, to help prevent hypothermia.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

Quality in Flow Ltd. QiF Blood and Fluid Warmer

K150404

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Name:Address:Quality in Flow Ltd.Kibutz EinatPOB 29Israel 4880500
Phone:+972-54-6600-146
Contact Person:Neta ShermanDirector Product Realization
Additional ContactPerson:Tsvia ErlichSenior Regulatory Affairs Consultant
Date Prepared:June 21, 2016

Name of Device and Name/Address of Sponsor

Trade Name:QiF Blood and Fluid Warmer
-----------------------------------------
Sponsor Address:Quality in Flow Ltd.
Kibutz EinatPOB 29Israel 4880500

Common or Usual Name

Sterile Fluid Path, in-line Blood Fluid Warmer

Classification Code and Name

LGZ - Warmer, Thermal, Infusion Fluid, Class 2 (21 C.F.R 880.5725) BSB - Warmer, Blood, Non-electromagnetic radiation, Class 2

Predicate Devices

VITAL SIGNS, Inc. (a GE Healthcare Company), enFlow IV Fluid Warmer (K112902)

Intended Use / Indications for Use

The QiF Blood and Fluid Warmer device is intended for warming blood, blood products, and intravenous fluids prior to administration. It is intended to be used by healthcare professionals in hospital, clinics and field environments, to help prevent hypothermia.

{4}------------------------------------------------

Technological Characteristics

The Quality in Flow Ltd. QiF Blood and Fluid Warmer device is a portable, sterile Fluid Path, in-line Blood and Fluid Warmer. The device is comprised of a Base Unit and a sterile disposable cartridge (Disposable Unit). The device is located between the fluid container (intravenous fluid, blood or blood product) and treated patient, outside of the patient's body. The Disposable Unit is composed of a plastic oval box encasing a spiral stainless steel heat exchanger tube. The Base Unit contains Firmware (software) and electronics (HW). The base Unit controls the performance of the system and the fluid outflow temperature. The power source is a rechargeable detachable battery located within the Base Unit. The disposable unit cartridge (blood/fluid) is provided with inlet and outlet luers that connect to standard IV tubing.

Performance Data

Quality In Flow Ltd. performed the following testing and Design Control activities, as described within the 510(k) submission, to evaluate the QiF Blood and Fluid Warmer device and demonstrate substantial equivalence to the identified predicate:

  • Software Verification and Validation was performed in compliance with ● "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", issued on May 11, 2005 (the Software Guidance)", and EN IEC 62304:2006/AC:2008. Software level of concern is moderate.
  • Bench testing to evaluate the performance and functionality of the device . includina:
    • Hemolysis study evaluated by flowing and warming blood O products under normal and worse-case conditions (such as stop flow, high flow rate and air bubble), in compliance with ASTM 2172:[(2002) Reapproved 2011)] and in addition flowing and heating old RBC blood product in high and low flow rates. No protocol deviations that affected the results took place. All samples met the acceptance criteria which is hemolysis of up to 1%.
    • Bench testing to demonstrate temperature indication accuracy, O outflow temperature profile under normal and abnormal (such as stop flow, high flow rate and air bubble) flow scenarios, fluid path integrity under pressure in compliance with ASTM 2172: [(2002) Reapproved 2011)], and battery capacity in terms of minimal heated volume. No protocol deviations that affected the results took place. All the tests met the acceptance criteria which were determined according to the product design input requirement specifications.
  • Biocompatibility testing (ISO 10993-5:2009, 10993-10:2010, 10993-● 11:2006, ASTM F756:2008)
  • . Leachability testing was performed on fluid that was flown and heated by the QiF device, the analysis of the fluid was done by external certified lab. There were no protocol deviations. Pass criteria - no significant change in the concentration of new materials not present in the fluid prior to heating

{5}------------------------------------------------

and no significant change in the concentration of any of the detected materials between results obtained with control sample and test article. All test samples met the acceptance criteria.

  • Sterilization and Shelf Life validation were conducted in compliance with ● ISO 11135-1:2007, ISO 11737-1:2006, ISO 10993-7:2008, and FDA CDRH, Shelf Life of Medical Devices guidance, April 1991.
  • . Electrical Safety and EMC testing (IEC 60601-1:2012, 60601-1-2:2007 ed. 3.0, 60601-1-6:2010, 60601-1-11:2010).
  • Battery testing per IEC 62133:2012. ●

In all instances, the QiF Blood and Fluid Warmer device performed as intended and met the acceptance criteria as set by the company or by the applicable standards.

Substantial Equivalence

The QiF Blood and Fluid Warmer device is substantially equivalent to the cleared Enginivity LLC enFlow IV Fluid Warmer. The QiF Blood and Fluid Warmer has the same intended uses and indications. The technological characteristics and principles of operation are substantially similar, both the QiF Blood and Fluid Warmer and the enFlow IV Fluid Warmers are composed of a warming reusable unit and a sterile, disposable, single patient use cartridge. The reusable warming unit of the subject and predicate devices controls the operation of the system and the outflow temperature of the infusate and contains software, hardware and the power source. The subject and predicate device use the same heating method, i.e., resistive heating. The disposable cartridge of both the subject and predicate devices contains a fluid path that serves as a heat exchanger. The heat exchanger of the subject and predicate devices contains metal that heat the blood, blood product or intravenous fluid by resistive heating method. The only main technological differences between the QIF device and its predicate are: the material composition of the heat exchanger fluid path, maximum output temperature, and presence of electrical and thermal insulation by shielding the QIF device Disposable Unit. These differences do not raise any new questions of safety and effectiveness and equivalent performance has been demonstrated by the hemolysis tests. Thus, the QiF Blood and Fluid Warmer is substantially equivalent. Detailed comparison between the predicate and the subject device is provided in the table below:

{6}------------------------------------------------

Substantial Equivalence Chart:

QiF Blood and Fluid Warmer(Subject Device)enFlow IV Fluid Warmer(K112902)
Indication for UseIntended for warming blood, blood products and intravenous solutions prior to administration. It is intended to be used by healthcare professionals in hospital, clinical and field environments to help prevent hypothermia.Intended for warming blood, blood products and intravenous solutions prior to administration. It is intended to be used by healthcare professionals in hospital, clinical and field environments to help prevent hypothermia.
Intended UseMedical emergencies or surgeries where warm fluid administration is required to treat the patient.Whenever parenteral introduction of normothermic fluid are desired or indicatedMedical emergencies or surgeries where warm fluid administration is required to treat the patient.Whenever parenteral introduction of normothermic fluid are desired or indicated
User PopulationHealthcare professional (i.e. paramedic, nurse doctor etc.)Healthcare professional (i.e. paramedic, nurse doctor etc.)
User Interface / NotificationsVisual (LCD display) and audioVisual (LED) and audio
Notification TypesOverheatUnder heatLow batteryStop FlowOverheatUnder heatLow battery
Usage environmentClinic and FieldClinic and Field
System componentsWarmer with display and a sterile disposable heat exchangerWarmer with display and a sterile disposable heat exchanger
Infusion temp.$38±2°C$Up to $40°C$(± 2 °C)
Heating MethodResistive heatingResistive heating
Fluid PathLocated within the sterile disposable cartridges (DU).Spiral stainless steel tube and a short segment of a PVC tubeThe sterile disposable cartridges consist of a plastic housing and biocompatible coated aluminum extrusion
The tube serves as theconductor of electrical current.form an enclosed fluidpath.
Flow RateBased on Gravity,Based on Gravity,
~ 160 ml/minUp to 200 ml/min
Warmer TypeInline warmerInline warmer
Power SourceRechargeable batteryRechargeable battery
110-120 or 220-240 VAC
BiocompatibilityThe fluid path is made ofbiocompatible Stainless Steeland PVCThe fluid path is made ofbiocompatible coatedaluminum extrusion
SoftwareThe software control theheating process and theoperation of the deviceThe software control theheating process and theoperation of the device
SterilizationThe disposable unit isprovided sterile for singlepatient useThe disposable unit isprovided sterile for singlepatient use
Product specificStandard with whichthe Device CompliesASTM 2172:2002,Standardspecification forblood/Intravenous FluidIrrigation Fluid WarmersASTM2172:2002,Standardspecification forblood/Intravenous FluidIrrigation Fluid Warmers

{7}------------------------------------------------

Conclusions

Based on comparison to the predicate device and completed performance testing, including biocompatibility, hemolysis, and leachability testing, as well as bench testing intended to demonstrate that the device meets it specifications and performs as intended, the QiF Blood and Fluid Warmer is substantially equivalent to the predicate device.

§ 880.5725 Infusion pump.

(a)
Identification. An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.(b)
Classification. Class II (performance standards).