K Number
K150321

Validate with FDA (Live)

Date Cleared
2015-07-13

(154 days)

Product Code
Regulation Number
888.3080
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EVOS Lumbar Interbody System (EVOS device) is an intervertebral body device intended for use in skeletally mature patients with Degenerative Disc Disease (DDD) of the lumbar spine with up to Grade 1 spondylolisthesis at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies.

The EVOS Lumbar Interbody System is intended to be used with autologous bone graft to facilitate fusion. It is to be used in patients who have had six months of non-operative treatment and is to be implanted via a direct posterior or transforaminal approach. The EVOS CURVED devices are implanted singly, while the EVOS ROTATE and EVOS STRAIGHT devices may be implanted singly or in pairs in the lumbosacral spine. The EVOS Lumbar Interbody System is intended to be used with supplemental fixation.

Device Description

The EVOS Lumbar Interbody System is an Intervertebral Body Fusion System for the lumbar spine. The EVOS Lumbar Interbody System includes surgical instruments to deliver the EVOS devices via a Posterior or Transforaminal approach. The generally rectangular design of the EVOS device incorporates an internal cavity for insertion of biological bone growth material through the top and bottom openings, as well as four small transverse holes (two on each side) that are designed to enhance bony ingrowth. Each device has a series of ridges or "teeth" on the top and bottom surfaces that are designed to ensure strong anchoring on the vertebral body endplates and resist backing out of the implant. The EVOS devices are provided sterile (gamma irradiated) for single use.

The EVOS device is made of radiolucent PEEK-OPTIMA® LT1 or PEEK-OPTIMA® HA Enhanced material. In addition, tantalum beads are embedded in the spacers to allow for radiographic visualization. The geometric shape of the EVOS devices are categorized as either STRAIGHT, CURVED, or ROTATE. All are available in a range of sizes, as well as flat and biconvex endplates, and with various degrees of lordosis to accommodate variations in patients' anatomy. Most of the EVOS STRAIGHT devices are offered with an option of a Standard nose, Wide nose, or Narrow nose. The devices are available in various heights, widths, and lengths to accommodate various surgical technique options.

AI/ML Overview

The provided text describes the EVOS Lumbar Interbody System, an intervertebral body fusion device, and its substantial equivalence to predicate devices, but does not contain information about studies related to acceptance criteria for a device that relies on artificial intelligence (AI) or machine learning.

The document is a 510(k) summary for a medical device approval. It focuses on:

  • Device Description: The physical characteristics, material, and types of the EVOS Lumbar Interbody System.
  • Indications for Use: The medical conditions and patient population for which the device is intended.
  • Comparison to Predicate Devices: How the EVOS system is similar to other legally marketed devices, primarily regarding technological characteristics and indications for use.
  • Performance Data: Lists mechanical testing performed on the device to demonstrate its structural integrity and safety. These are standard tests for spinal implants.

Therefore, I cannot provide the requested information regarding AI acceptance criteria or studies proving performance based on this document. The questions you posed (sample size for test/training sets, data provenance, expert ground truth, adjudication, MRMC studies, standalone performance, type of ground truth, etc.) are highly relevant to AI/ML medical devices, which this document does not cover.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around it. Inside the circle is a stylized image of three human profiles facing right, layered on top of each other to create a sense of depth.

July 13, 2015

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Cutting Edge Spine, LLC Mr. John Souza, Sr. Director of Engineering 101 Waxhaw Professional Park Drive, Suite A Waxhaw, North Carolina 28173

Re: K150321

Trade/Device Name: EVOS Lumbar Interbody System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: July 2, 2015 Received: July 7, 2015

Dear Mr. Souza:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

Page 2 - Mr. John Souza, Sr.

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K150321

Device Name EVOS Lumbar Interbody System

Indications for Use (Describe)

The EVOS Lumbar Interbody System (EVOS device) is an intervertebral body device intended for use in skeletally mature patients with Degenerative Disc Disease (DDD) of the lumbar spine with up to Grade 1 spondylolisthesis at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies.

The EVOS Lumbar Interbody System is intended to be used with autologous bone graft to facilitate fusion. It is to be used in patients who have had six months of non-operative treatment and is to be implanted via a direct posterior or transforaminal approach. The EVOS CURVED devices are implanted singly, while the EVOS ROTATE and EVOS STRAIGHT devices may be implanted singly or in pairs in the lumbosacral spine. The EVOS Lumbar Interbody System is intended to be used with supplemental fixation.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

I. SUBMITTER

Cutting Edge Spine, LLC 101 Waxhaw Professional Park Dr., Suite A Waxhaw, NC 28173

Tel: (704) 243-0892 ext 12 Fax: (704) 731-2559

Contact Person: John Souza Sr., Director of Engineering Date Prepared: February 06, 2015

II. DEVICE

Name of Device: EVOS Lumbar Interbody System Common or Usual Name: Intervertebral fusion device with bone graft, lumbar Classification Name: Intervertebral body fusion device (21 CFR 888.3080) Regulatory Class: II Product Code: MAX

III. PREDICATE DEVICE

510(k) NumberDeviceManufacturer
PrimaryPredicateK102957EVOL Intervertebral BodyFusion DeviceCutting Edge Spine, LLC
AdditionalPredicateK071724Lucent Intervertebral BodyFusion DeviceSpinal Elements
AdditionalPredicateK073291Capstone Spinal SystemMedtronic
AdditionalPredicateK120368Capstone Control SpinalSystemMedtronic

No reference devices were used in this submission.

IV. DEVICE DESCRIPTION

The EVOS Lumbar Interbody System is an Intervertebral Body Fusion System for the lumbar spine. The EVOS Lumbar Interbody System includes surgical instruments to deliver the EVOS devices via a Posterior or Transforaminal approach. The generally rectangular design of the EVOS device

{4}------------------------------------------------

incorporates an internal cavity for insertion of biological bone growth material through the top and bottom openings, as well as four small transverse holes (two on each side) that are designed to enhance bony ingrowth. Each device has a series of ridges or "teeth" on the top and bottom surfaces that are designed to ensure strong anchoring on the vertebral body endplates and resist backing out of the implant. The EVOS devices are provided sterile (gamma irradiated) for single use.

The EVOS device is made of radiolucent PEEK-OPTIMA® LT1 or PEEK-OPTIMA® HA Enhanced material. In addition, tantalum beads are embedded in the spacers to allow for radiographic visualization. The geometric shape of the EVOS devices are categorized as either STRAIGHT, CURVED, or ROTATE. All are available in a range of sizes, as well as flat and biconvex endplates, and with various degrees of lordosis to accommodate variations in patients' anatomy. Most of the EVOS STRAIGHT devices are offered with an option of a Standard nose, Wide nose, or Narrow nose. The devices are available in various heights, widths, and lengths to accommodate various surgical technique options.

V. INDICATIONS FOR USE

The EVOS Lumbar Interbody System (EVOS device) is an intervertebral body device intended for use in skeletally mature patients with Degenerative Disc Disease (DDD) of the lumbar spine with up to Grade 1 spondylolisthesis at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies.

The EVOS Lumbar Interbody System is intended to be used with autologous bone graft to facilitate fusion. It is to be used in patients who have had six months of non-operative treatment and is to be implanted via a direct posterior or transforaminal approach. The EVOS CURVED devices are implanted singly, while the EVOS ROTATE and EVOS STRAIGHT devices may be implanted singly or in pairs in the lumbosacral spine. The EVOS Lumbar Interbody System is intended to be used with supplemental fixation.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

Intervertebral Body Fusion through a Posterior or Transformanial approach is the technological principle for both the subject and predicate devices. It is based on the use of surgical instrumentation to insert PEEK implants into the intervertebral space in order to assist in fusion. At a high level, the subject and predicate devices are based on the following same technological elements:

  • Indications for Use
  • Structural support mechanism .

VII. PERFORMANCE DATA

The EVOS device was tested in compliance with FDA's guidance document titled "Class II Special Controls Guidance Document: Intervertebral Body Fusion Device" and demonstrated substantially equivalent performance characteristics to the identified predicate devices.

{5}------------------------------------------------

The EVOS device has been tested in the following test modes:

  • Mechanical Testing on the final device that has been gamma irradiated and aged. ●
    • o Expulsion Testing
    • o Static Axial Compression per ASTM F2077
    • Static Compressive Shear per ASTM F2077 O
    • Static Testing in Subsidence per ASTM F2267 O
    • Dynamic Axial Compression per ASTM F2077 O

VIII. CONCLUSIONS

The overall technology characteristics and mechanical performance data lead to the conclusion that the EVOS Lumbar Interbody System is substantially equivalent to the EVOL Intervertebral Body Fusion Device from Cutting Edge Spine (K102957), the Lucent Intervertebral Body Fusion Device from Spinal Elements (K071724), the Medtronic Capstone Spinal System (K073291), and the Medtronic Capstone Control Spinal System (K120368).

N/A