K Number
K150214

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-07-31

(182 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The OneTouch Verio Flex™ Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip. The OneTouch Verio Flex™ Blood Glucose Monitoring System is intended to be used by a single patient and should not be be shared.

OneTouch Verio Flex™ Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The OneTouch Verio Flex™ Blood Glucose Monitoring System is not to be used for the diagnosis of or screening of diabetes, or for neonatal use.

The OneTouch® Verio Test Strips are for use with the OneTouch Verio Flex™ Blood Glucose Meter to quantitatively measure glucose drawn from the fingertips.

Device Description

The OneTouch Verio Flex Blood Glucose Monitoring System consists of the OneTouch Verio Flex Blood Glucose Meter, OneTouch® Verio® Test Strips, OneTouch® Verio® Level 3 and Level 4 Control Solutions, Lancing Device and Sterile Lancets. The OneTouch Verio Flex Blood Glucose Monitoring System measures the glucose content of a blood sample by means of an electrical current produced in the test strip and sent to the meter for measurement.

AI/ML Overview

OneTouch Verio Flex™ Blood Glucose Monitoring System Acceptance Criteria and Study Details

1. Acceptance Criteria and Reported Device Performance

The OneTouch Verio Flex™ Blood Glucose Monitoring System was designed and tested in accordance with ISO 15197:2013(E) guidelines for system accuracy. The key acceptance criteria and reported performance from the device's system accuracy study are summarized below.

Acceptance Criteria per ISO 15197:2013(E) Clause 6.3 - System Accuracy (Healthcare Professional Use):

Glucose Concentration RangeAcceptance Criterion (Percentage of samples within range)
< 100 mg/dL≥ 95% of results within ±15 mg/dL
≥ 100 mg/dL≥ 95% of results within ±15%
Overall (Combined)≥ 95% of results within ±15 mg/dL or ±15%

Reported Device Performance (Healthcare Professional Use):

Glucose Concentration RangeReported Performance (Percentage of samples within range)
< 100 mg/dL100% of results within ±15 mg/dL (Pooled Lots A-C)
≥ 100 mg/dL99.3% of results within ±15% (Pooled Lots A-C)
Overall (Combined)99.5% of results within ±15 mg/dL or ±15% (Pooled Lots A-C)

Acceptance Criteria per ISO 15197:2013(E) Clause 8 - User Performance Evaluation (Lay User):

Glucose Concentration RangeAcceptance Criterion (Percentage of samples within range)
< 100 mg/dL≥ 95% of results within ±15 mg/dL
≥ 100 mg/dL≥ 95% of results within ±15%
Overall (Combined)≥ 95% of results within ±15 mg/dL or ±15%

Reported Device Performance (Lay User Performance Evaluation):

Glucose Concentration RangeReported Performance (Percentage of samples within range)
< 100 mg/dL93.1% of results within ±15 mg/dL (Note: This is below the ISO 95% criterion)
≥ 100 mg/dL98.6% of results within ±15%
Overall (Combined)97.6% of results within ±15 mg/dL or ±15%

Note on Lay User Performance: While the overall and ≥100 mg/dL results met the ISO 15197:2013(E) Clause 8 criteria, the performance for glucose concentrations <100 mg/dL was 93.1%, which is below the 95% acceptance criterion mentioned in the ISO standard for this range. However, the summary statement on page 15 still concludes that the system meets "all design input specifications" and ISO 15197:2013 requirements, suggesting that possibly the specific internal acceptance criteria for this portion might have been different or considered acceptable within the context of the overall system accuracy. Without further information, this discrepancy for the <100 mg/dL range in lay user performance is noted.


2. Sample size used for the test set and the data provenance

Method Comparison Performance (Healthcare Professional Use):

  • Sample Size: 115 subjects
  • Data Provenance: Not explicitly stated, but implies prospective collection as it involved "fingersick samples from diabetic subjects" and testing against a reference method. The location is not specified.

System Accuracy in compliance with ISO 15197:2013(E) Clause 6.3 (Healthcare Professional Use):

  • Sample Size: 100 subjects
  • Data Provenance: Not explicitly stated, but implies prospective collection as it involved "fingertip capillary blood samples obtained by healthcare professionals". The location is not specified.

Lay User Performance Evaluation (Clause 8, first study summary):

  • Sample Size: 172 lay persons
  • Data Provenance: Not explicitly stated, but implies prospective collection as it involved "fingertip capillary blood samples obtained by lay persons". The location is not specified.

User Performance Evaluation in compliance with ISO 15197:2013(E) Clause 8 (second study summary):

  • Sample Size: 167 lay persons
  • Data Provenance: Not explicitly stated, but implies prospective collection as it involved "fingertip capillary blood samples obtained by lay persons". The location is not specified.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document does not specify the number or qualifications of experts used to establish the ground truth. Instead, the ground truth was established using a recognized laboratory reference method: the Yellow Springs Instrument (YSI) 2300 (or YSI 2300 STAT PLUS) glucose analyzer. This instrument itself is considered the "expert" for generating accurate glucose measurements.


4. Adjudication method for the test set

Not applicable. The ground truth was established by a laboratory reference instrument (YSI analyzer), not by human experts requiring adjudication.


5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a blood glucose monitoring system, not an AI-powered diagnostic imaging or interpretation tool that involves "human readers." Therefore, an MRMC comparative effectiveness study with AI assistance is not relevant to this product.


6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, a standalone performance evaluation was done. The "Method Comparison Performance" study and the "System Accuracy in compliance with ISO 15197:2013(E) Clause 6.3" study assessed the device's accuracy when operated by healthcare professionals, comparing its readings directly to the laboratory reference method (YSI). This represents the algorithm's performance with trained users, effectively functioning as a standalone assessment of the device's accuracy in a controlled environment. The "Lay User Performance Evaluation" also assessed the device's accuracy when operated by intended users, albeit with human interaction in using the device.


7. The type of ground truth used

The ground truth used was laboratory reference measurements obtained from a Yellow Springs Instrument (YSI) 2300 (or YSI 2300 STAT PLUS) glucose analyzer. This is considered a highly accurate and recognized method for determining glucose concentrations.


8. The sample size for the training set

The document does not provide a sample size specifically for a "training set." Blood glucose monitoring systems like the OneTouch Verio Flex™ typically do not use machine learning models that require a separate training set in the same way AI algorithms do. The development process would involve calibration and optimization based on chemical and electrical principles rather than iterative training on a dataset. The studies described are validation studies.


9. How the ground truth for the training set was established

As noted above, the concept of a "training set" in the context of machine learning does not directly apply to this device. For the development and calibration of the glucose monitoring system, the ground truth would have been established through highly accurate laboratory reference methods (like those provided by a YSI analyzer) during the chemical and electrical engineering phases, prior to the validation studies described in the document. The document doesn't detail these initial development and calibration phases.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of three human profiles facing right, arranged in a stacked formation. The profiles are depicted in black and white. The graphic is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 31, 2015

LIFESCAN EUROPE ALISON WILSON REGULATORY AFFAIRS PROJECT MANAGER BEECHWOOD PARK NORTH INVERNESS IV2 3ED, GREAT BRITAIN

Re: K150214

Trade/Device Name: One Touch Verio Flex™ Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: NBW, LFR Dated: June 25, 2015 Received: July 02, 2015

Dear Alison Wilson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Courtney

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K150214

Device Name

OneTouch Verio FlexTM Blood Glucose Monitoring System

Indications for Use (Describe)

The OneTouch Verio Flex™ Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip. The OneTouch Verio Flex™ Blood Glucose Monitoring System is intended to be used by a single patient and should not be be shared.

OneTouch Verio Flex™ Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The OneTouch Verio Flex™ Blood Glucose Monitoring System is not to be used for the diagnosis of or screening of diabetes, or for neonatal use.

The OneTouch® Verio Test Strips are for use with the OneTouch Verio Flex™ Blood Glucose Meter to quantitatively measure glucose drawn from the fingertips.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the LifeScan logo, which includes a blue globe on the left side and a blue swoosh above the text "LIFESCAN". Below the text, in a smaller, red font, it says "a Johnson & Johnson company". The logo is clean and professional, representing a well-established brand in the healthcare industry.

Division of Cilag GmbH International

Gubelstrasse 34 , CH-6300 Zug, Switzerland

510(k) Summary (as required by section 807.92(c))

SponsorLifeScan Europe, a Division of Cilag GmbH InternationalGubelstrasse 34Zug, Switzerland 6300
CorrespondentAlison Wilson, Regulatory Affairs Project ManagerLifeScan Scotland Ltd.Beechwood Park NorthInverness, IV2 3EDUnited KingdomPhone: +44(0) 1463 721256e-mail: awilson4@its.jnj.comAlternate 510(k) Contact:Oyinkan Donaldson, Senior Regulatory Affairs ManagerLifeScan Scotland LtdBeechwood Park NorthInverness, IV2 3EDUnited KingdomPhone: +44 01463 721259Mobile: +44 (0) 7909935151Fax: +44 (0)1463 722000e-mail: odonalds@its.jnj.com
Date PreparedMay 2015
Device Trade NameOneTouch Verio Flex Blood Glucose Monitoring System
Common NameGlucose Test System

LifeScan Europe, a Div. of Cilag GmbH International

Page 1 of 13

Part 2 510K Summary

Confidential and Proprietary Information

Traditional 510(k)

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the LifeScan logo. The logo features the word "LIFESCAN" in large, bold, blue letters. Below the company name is the text "a Johnson & Johnson company" in a smaller, red font. To the left of the company name is a blue globe and a curved blue line above the company name.

Division of Cilag GmbH International

Gubelstrasse 34 , CH-6300 Zug, Switzerland

ClassificationOneTouch Verio Flex™ Blood Glucose Meters and OneTouch Verio™Test Strips are Class II devices (21 CFR § 862.1345), Product CodeNBW, LFROneTouch Verio™ Control Solutions are Class I devices (21 CFR§862.1660), Product Code JJXLancing Device and Sterile Lancets are Class I (exempt) devices (21CFR § 878.4800), Product Code FMK
System DescriptionThe OneTouch Verio Flex Blood Glucose Monitoring System consistsof the OneTouch Verio Flex Blood Glucose Meter, OneTouch® Verio®Test Strips, OneTouch® Verio® Level 3 and Level 4 Control Solutions,Lancing Device and Sterile Lancets. The OneTouch Verio Flex BloodGlucose Monitoring System measures the glucose content of a bloodsample by means of an electrical current produced in the test strip andsent to the meter for measurement.
Predicate DeviceOneTouch® Verio™ Blood Glucose Monitoring System (K131363,Cleared 30th August 2013)
IntendedUse/Indications forUseThe OneTouch Verio Flex Blood Glucose Monitoring System isintended to be used for the quantitative measurement of glucose (sugar)in fresh capillary whole blood samples drawn from the fingertip. TheOneTouch Verio Flex Blood Glucose Monitoring System is intended tobe used by a single patient and should not be shared.OneTouch Verio Flex Blood Glucose Monitoring System is intended forself testing outside the body (in vitro diagnostic use) by people withdiabetes at home as an aid to monitor the effectiveness of diabetescontrol. The OneTouch Verio Flex Blood Glucose MonitoringSystem is not to be used for the diagnosis of or screening of diabetes, or
for neonatal use.
The OneTouch® Verio™ Test Strips are for use with the OneTouchVerio Flex Blood Glucose Meter to quantitatively measure glucosedrawn from the fingertips.
Comparison toPredicate DeviceThe Subject device is different from the predicate device in thefollowing aspects:• Meter:○ Ergonomic/physical design: Changes to size, shape and color○ Electronic/hardware: Strip port connector○ Software/Firmware changes: Modified Blood GlucoseAlgorithm and addition of a range indicator software feature.• Labelling:○ New branding and Instructions for UseThere are no changes to the OneTouch® Verio® Test Strips or theOneTouch® Verio® Level 3 (Mid) and Level 4 (High) Control Solutionscleared for use with the predicate OneTouch® Verio™ Blood GlucoseMonitoring System 510(k) (K131363).There have been no changes to the intended use, operating principle orscientific technology.
TechnologicalCharacteristicsThere has been no change to the fundamental scientific technology,which is amperometric detection. The operating principle remainselectrochemical reaction.
Summary ofPerformanceCharacteristicsThe OneTouch Verio Flex Blood Glucose Monitoring System (meter,strips, and control solutions) was designed and tested in accordance withISO 15197:2013(E). Analytical performance testing included system
Within ±5 mg/dLWithin ±10 mg/dLWithin ±15 mg/dL
77.8%100%100%
(42/54)(54/54)(54/54)

LifeScan Europe, a Div. of Cilag GmbH International

Page 2 of 13

Part 2 510K Summary

Confidential and Proprietary Information

Traditional 510(k)

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the LifeScan logo. The logo features a blue globe on the left, followed by the word "LIFESCAN" in large, bold, blue letters. Below the company name, in smaller red font, is the text "a Johnson & Johnson company".

Division of Cilag GmbH International

Gubelstrasse 34 , CH-6300 Zug, Switzerland

LifeScan Europe, a Div. of Cilag GmbH International

Page 3 of 13

Confidential and Proprietary Information

Part 2 510K Summary

Traditional 510(k)

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the LifeScan logo. The logo features a blue globe on the left, followed by the word "LIFESCAN" in large, bold, blue letters. Below the company name, in smaller red cursive font, is the text "a Johnson & Johnson company".

Division of Cilag GmbH International

Gubelstrasse 34 , CH-6300 Zug, Switzerland

accuracy, repeatability, intermediate precision and linearity testing. A user performance evaluation assessed accuracy of results and usability of the device in the hands of intended users. The OneTouch Verio Flex Blood Glucose Monitoring System performed similarly to both the predicate device as well as to a laboratory reference method, the Yellow Springs Instrument (YSI).

Method Comparison Performance

A study evaluating the glucose values from the OneTouch Verio Flex Blood Glucose monitoring System (obtained by healthcare professional fingersick samples from diabetic subjects) and compared to the glucose results obtained by the recognized glucose reference method (YSL 2300) from 115 subjects showed the following results:

NOTE: first test sample data is presented here.

Method Comparison Results for Glucose Concentrations <75 mg/dL

Percent (and number) of meter results that match the YSI reference

Method Comparison Results for Glucose Concentrations ≥75 mg/dL

Percent (and number) of meter results that match the YSI reference

Within ±5%Within ±10%Within ±15%Within ±20%
62.9%94.8%98.6%99.3%
(183/291)(276/291)(287/291)(289/291)

LifeScan Europe, a Div. of Cilag GmbH International

Page 4 of 13

Part 2 510K Summary

Confidential and Proprietary Information

Traditional 510(k)

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for LifeScan, a Johnson & Johnson company. The logo features a blue swoosh design above the company name, with a globe graphic to the left. The words "LIFESCAN" are written in a bold, sans-serif font, and below it, in a smaller, cursive font, is the phrase "a Johnson & Johnson company."

Division of Cilag GmbH International

Gubelstrasse 34 , CH-6300 Zug, Switzerland

System Accuracy Results for Glucose concentrations across the glucose range:

Percent (and number) of meter results that match the YSI reference

Within ±10 mg/dL (±0.83 mmol/L) or ±15%Within ±10 mg/dL (±0.83 mmol/L) or ±20%
98.8%(341/345)99.4%(343/345)

LifeScan Europe, a Div. of Cilag GmbH International

Page 5 of 13

Part 2 510K Summary

Confidential and Proprietary Information

Traditional 510(k)

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains the word "LifeScan" in a simple, sans-serif font. The text is black and appears against a white background. The word is presented in a straightforward, unadorned manner.

Image /page/8/Picture/1 description: The image shows the logo for LifeScan, a Johnson & Johnson company. The logo features the word "LIFESCAN" in large, bold, blue letters. Below the company name is the text "a Johnson & Johnson company" in a smaller, cursive font. To the left of the company name is a blue globe graphic.

A linear regression ananlysis of the method comparison study data demonstrated that the OneTouch®VerioFlex™ Blood Glucose Monitoring System (subject device) and the OneTouch® Verio" Blood Glucose Monitoring System (predicate device) are substantially equivalent; and that the subject device compared well to the laboratory reference instrument (YSI analyzer). NOTE: first test sample data is presented here and within the product labelling

BGMSLot ##Participants#TestsSlope[95% CI]Intercept[95% CI](mg/dL)Std.Error(Sy.x)(mg/dL)R2
OneTouch® Verio™D1151150.97[0.95 to 0.99]6.05[2.41 to 9.69]10.150.99
OneTouch®VerioFlex™A1151151.00[0.98 to 1.02]-0.31[-4.04 to3.42]10.410.99
B1151150.99[0.97 to 1.01]1.14[-2.91 to5.20]11.330.99
C1151151.01[0.99 to 1.03]0.01[-4.13 to4.16]11.570.99
3 lots3453451.00[0.99 to 1.01]0.28[-2.00 to2.56]11.11

Regression Statistics for the Subject and Predicate Device compared to YSI

Table above reflects regression limits of 15mg/dL <75mg/dl and % results within 15%>75mg/dL

LifeScan Europe, a Div. of Cilag GmbH International

Page 6 of 13

Part 2 510K Summary

Confidential and Proprietary Information

Traditional 510(k)

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the LifeScan logo. The logo features a blue globe on the left, followed by the word "LIFESCAN" in large, bold, blue letters. Below the word "LIFESCAN" is the phrase "a Johnson & Johnson company" in smaller, red letters. A blue swoosh appears above the globe and the word "LIFESCAN".

Lay User Performance Evaluation

A study to validate the accuracy performance of the OneTouch Verio Flex Blood Glucose Monitoring System in the hands of the user was undertaken. A comparison of the Lay User OneTouch Verio Flex Blood Glucose Monitoring System fingertip results compared to glucose results obtained on the recognized glucose reference method (YSI 2300 STAT PLUS glucose analyzer), are summarized below. Glucose values from fingertip capillary blood samples obtained by 172 lay persons showed the following results:

Subject Fingertip Results for Glucose Concentrations <75 mg/dL

Percent (and number) of meter results that match the YSI reference

Within ±5 mg/dLWithin ±10 mg/dLWithin ±15 mg/dL
76.2%95.2%95.2%
(16/21)(20/21)(20/21)

Subject Fingertip Results for Glucose Concentrations ≥75 mg/dL

Percent (and number) of meter results that match the YSI reference

Within ±5%Within ±10%Within ±15%Within ±20%
58.3%88.7%98.0%99.3%
(88/151)(134/151)(148/151)(150/151)

Precision (Repeatability)

Within Run Precision (300 Venous Blood Samples Tested per glucose level)

LifeScan Europe, a Div. of Cilag GmbH InternationalPage 7 of 13
Part 2 510K SummaryConfidential and Proprietary Information
Traditional 510(k)OneTouch Verio FlexTM Blood Glucose Monitoring System

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the LifeScan logo, which features the company name in bold, blue letters. Above the name is a blue swoosh, and to the left is a blue globe. Below the company name, in smaller, red cursive letters, is the text "a Johnson & Johnson company."

Division of Cilag GmbH International

Gubelstrasse 34 , CH-6300 Zug, Switzerland

Target GlucoseMean GlucoseStandard DeviationCoefficient of
(mg/dL)(mg/dL)(mg/dL)Variation (%)
2018.30.633.43
4036.50.832.26
9088.51.721.95
130127.92.501.95
200199.33.941.98
350344.66.181.79
600583.410.541.81

Results show that the greatest variability observed between test strips when tested with blood is 1.72mg/dl SD or less at glucose levels less than 100mg/dl, or 1.98% or less at glucose levels at 100mg/dl or above

Total Precision (Intermediate Precision)

(600 Control Solution Tests)

Glucose LevelRanges(mg/dL)Mean Glucose(mg/dL)Standard Deviation(mg/dL)Coefficient ofVariation (%)
Very Low (0-24)12.460.574.54
Low (25 - 49)37.140.922.48
Mid (102 - 138)117.682.412.05
High (298 – 403)348.998.312.38
Very High (446-604)515.1813.532.63

LifeScan Europe, a Div. of Cilag GmbH International

Page 8 of 13

Part 2 510K Summary

Confidential and Proprietary Information

Traditional 510(k)

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for LifeScan, a Johnson & Johnson company. The logo features a blue globe on the left, followed by the word "LIFESCAN" in large, bold, blue letters. Below the company name is the text "a Johnson & Johnson company" in a smaller, handwritten-style font.

System Accuracy performance in accordance with ISO 15197:2013(E)

Accuracy of the subject device was analysed and assessed in compliance with the product design requirements in accordance with the requirements of ISO 15197:2013(E) Clause 6.3 System Accuracy and Clause 8 User Performance Evaluation.

System Accuracy in compliance with ISO 15197:2013(E) Clause 6.3:

A study was conducted to evaluate glucose values from fingertip capillary blood samples obtained by healthcare professionals from 100 subjects in accordance with the glucose ranges required by ISO15197:2013 Clause 6.3.5. The analysis showed the following results: 100% for all 3 test strip lots within ±15mg/dl of the medical laboratory values at glucose concentrations below 100mg/dL and 99.3%, 100%, 98.7% within ±15% of the medical laboratory values at glucose concentrations at or above 100mg/dL.

System Accuracy Results for Glucose Concentrations <100 mg/dL

Percent (and number) of meter results that match the YSI reference

Test Strip Lot NumberWithin ±5mg/dLWithin ±10mg/dLWithin ±15mg/dL
A86%(43/50)98%(49/50)100%(50/50)
B80%(40/50)98%(49/50)100%(50/50)
C80%(40/50)98%(49/50)100%(50/50)
Pooled Results for Lots A-C82.0%(123/150)98.0%(147/150)100%(150/150)

System Accuracy Results for Glucose Concentrations ≥100 mg/dL

Percent (and number) of meter results that match the YSI reference

LifeScan Europe, a Div. of Cilag GmbH International Page 9 of 13 Confidential and Proprietary Information

Part 2 510K Summary

Traditional 510(k)

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the LifeScan logo. The logo features the company name in a bold, blue font, with a blue swoosh above the name and a globe to the left. Below the company name, it says "a Johnson & Johnson company" in a smaller, red font.

Division of Cilag GmbH International

Gubelstrasse 34 , CH-6300 Zug, Switzerland

Test Strip Lot NumberWithin ±5%Within ±10%Within ±15%
A65.3%(98/150)93.3%(140/150)99.3%(149/150)
B64.0%(96/150)96.0%(144/150)100%(150/150)
C59.3%(89/150)94.7%(142/150)98.7%(148/150)
Pooled Results for LotsA-C62.9%(283/450)94.7%(426/450)99.3%(447/450)

System accuracy results across the glucose range tested: concentrations between 38.2 mg/dL and 466.3 mg/dL

Percent (and number) of meter results that match the YSI reference

Test Strip Lot NumberWithin ±15mg/dL or ±15%
A99.5%(199/200)
B100%(200/200)
C99.0%(198/200)
Pooled Results for A-C99.5%(597/600)

NOTE: Where 38.2 mg/dL represents the lowest glucose reference value and 466.3 mg/dL represents the highest glucose reference value (YSI value).

User Performance Evaluation in compliance with ISO 15197:2013(E) Clause 8

LifeScan Europe, a Div. of Cilag GmbH InternationalPage 10 of 13
Part 2 510K SummaryConfidential and Proprietary Information
Traditional 510(k)OneTouch Verio Flex™ Blood Glucose Monitoring System

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the LifeScan logo. The logo features a blue globe on the left, followed by the word "LIFESCAN" in large, bold, blue letters. Below the word "LIFESCAN" is the text "a Johnson & Johnson company" in a smaller, red, cursive font. A blue swoosh design arches over the globe and the word "LIFESCAN".

Division of Cilag GmbH International Gubelstrasse 34 , CH-6300 Zug, Switzerland

A study to validate the accuracy performance of the OneTouch Verio Flex Blood Glucose Monitoring System in the hands of the Lay user in accordance with ISO 15197:2013(E) Clause 8 was undertaken. A comparison of the Lay User OneTouch Verio Flex Blood Glucose Monitoring System fingertip results compared to glucose results obtained on the recognized glucose reference method (YSI 2300 STAT PLUS glucose analyzer), are summarized below:

Glucose values from fingertip capillary blood samples obtained by 167 lay persons showed the following results:

93.1% within ±15mg/dl of the glucose reference (YSI) values at glucose concentrations below 100mg/dL and 98.6% within ±15% of the medical laboratory values at glucose concentrations at or above 100mg/dL.

97.6% of the total number of samples across the entire glucose range were within ±15 mg/dl or ±15% of the medical laboratory values.

Subject Fingertip Results for Glucose Concentrations <100 mg/dL

Percent (and number) of meter results that match the YSI reference

Within ±5 mg/dLWithin ±10 mg/dLWithin ±15 mg/dL
72.4%93.1%93.1%
(21/29)(27/29)(27/29)

Subject Fingertip Results for Glucose Concentrations ≥100 mg/dL

Percent (and number) of meter results that match the YSI reference

Within ±5%Within ±10%Within ±15%
58.7%(81/138)88.4%(122/138)98.6%(136/138)

Results for Glucose concentrations across the glucose range:

LifeScan Europe, a Div. of Cilag GmbH International

Page 11 of 13

Part 2 510K Summary

Traditional 510(k)

OneTouchVerio Flex™ Blood Glucose Monitoring System

Confidential and Proprietary Information

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the LifeScan logo, which features a blue globe on the left side and the word "LIFESCAN" in large, bold, blue letters. Above the word LIFESCAN is a blue curved line. Below the word LIFESCAN is the text "a Johnson & Johnson company" in a smaller, cursive font.

Division of Cilag GmbH International

Gubelstrasse 34 , CH-6300 Zug, Switzerland

Percent (and number) of meter results that match the YSI reference

Within ±15 mg/dL (±0.83 mmol/L) or ±15%

97.6% (163/167)

LifeScan Europe, a Div. of Cilag GmbH International

Page 12 of 13

Part 2 510K Summary

Confidential and Proprietary Information

Traditional 510(k)

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the logo for LifeScan, a Johnson & Johnson company. The logo features the company name in a bold, blue font, with a blue swoosh above it. To the left of the company name is a blue globe. Below the company name, in a smaller, red font, is the text "a Johnson & Johnson company."

Division of Cilag GmbH International Gubelstrasse 34 , CH-6300 Zug, Switzerland

Summary

Design verification and validation testing confirmed that the performance, safety, and effectiveness of the OneTouch® VerioFlex™ Blood Glucose Monitoring System were met against all design input specifications and the system can be considered substantially equivalent to that of the predicate device. The OneTouch Verio Flex Meter also meets the requirements of ISO15197:2013 and applicable recognized electrical and safety standards including FCC requirements.

Conclusions

The OneTouch® Verio Flex™ Blood Glucose Monitoring System is substantially equivalent in its intended use, performance, safety, effectiveness and underlying scientific and operating principles to the predicate, the OneTouch® Verio" Blood Glucose Monitoring System (K131363).

LifeScan Europe, a Div. of Cilag GmbH International

Page 13 of 13

Part 2 510K Summary

Confidential and Proprietary Information

Traditional 510(k)

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.