(282 days)
The Dimension Tacrolimus Flex® Reagent Cartridge (TAC) is an in vitro diagnostic test for the quantitative measurement of tacrolimus in human whole blood on the Dimension® clinical chemistry system. Measurements of tacrolimus are used as an aid in the management of tacrolimus therapy in renal and hepatic transplant patients.
The Dimension Tacrolimus Callbrator (TAC CAL) is an in vitro diagnostic product for the Tacrolimus (TAC) method on the Dimension® clinical chemistry system.
The automated Dimension® TAC method uses an immunoassay technique in which free and tacrolimus-bound antibody-enzyme conjugate is separated using magnetic particles. The assay is performed using a method specific Flex® reagent cartridge. The Flex® cartridge contains a pretreatment reagent, antibody-ß-galactosidase conjugate, tacrolimus immobilized on chromium dioxide particles, chlorophenol red ß-d-galactopyranoside (CPRG) substrate, and diluent to hydrate the tablets.
Dimension® Tacrolimus (TAC) Calibrator is five level frozen liquid, whole blood hemolysate containing purified tacrolimus. The provided in 4.0 mL vials, 2 vials per level. There are five calibrator levels per kit which span the assay range for the Dimension® Tacrolimus (TAC) assay. Calibrators are filled with 1.0 mL per vial except for Level 1 calibrator which contains 2.0 mL per vial.
Here's a summary of the acceptance criteria and study information for the Dimension Tacrolimus Flex® Reagent Cartridge (TAC) and Dimension Tacrolimus Calibrator (TAC CAL), based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state acceptance criteria in a quantitative format for most categories. Instead, it describes the type of testing performed and then indicates whether the device met certain expectations (e.g., "All samples met the acceptance criterion of ≤ 10% mean bias in recovery"). Where specific quantitative performance is given, it is included.
| Performance Category | Acceptance Criteria (Explicit or Implied) | Reported Device Performance |
|---|---|---|
| Precision | Not explicitly stated (implied to be acceptable for clinical use) | Typical precision at levels ranging from 1.8 to 27.4 ng/mL was ≤ 6.6% CV for Repeatability and ≤ 13.1% CV for Within Lab. |
| Linearity (Analytical Measuring Range) | Established by Limit of Quantitation (LoQ) and linearity study. | Assay range established as 1.0 - 30.0 ng/mL. Linear regression, 2nd, and 3rd order polynomial regressions performed. The specific "acceptance criterion" for linearity (e.g., R-squared value, deviation from linearity) is not explicitly stated, but the range was "established". |
| Specificity (Cross-reactivity to Metabolites) | Not explicitly stated (implied acceptable low cross-reactivity for relevant metabolites). | * M-I (13-O-desmethyl-tacrolimus): 1% cross-reactivity. * M-II (31-O-desmethyl-tacrolimus): 18% cross-reactivity. * M-III (15-O-desmethyl-tacrolimus): 15% cross-reactivity. * M-IV (12-O-hydroxyl-tacrolimus): 99% cross-reactivity. * M-V (15,31-O-didesmethyltacrolimus): 1% cross-reactivity. * M-VI (13,31-O-didesmethyl-tacrolimus): 1% cross-reactivity. * M-VII (13,15-O-didesmethyl-tacrolimus): 43% cross-reactivity. * M-VIII (unknown name): 0% cross-reactivity. |
| Specificity (Interference from Endogenous/Co-administered Drugs) | Bias less than 10% at tested levels. | Co-administered drugs and other compounds tested for interference, exhibited less than 10% bias at the levels tested. |
| Recovery | ≤ 10% mean bias in recovery. | All samples met the acceptance criterion of ≤ 10% mean bias in recovery. |
| Method Comparison (vs. Reference Method LC-MS/MS) | Not explicitly stated (implied acceptable correlation and agreement for clinical use based on regression). | Slope: 1.04, Intercept: -0.30, r: 0.966 (n=315, Range: 1.3-24.9 ng/mL). |
| Method Comparison (vs. Predicate ARCHITECT Tacrolimus Assay) | Not explicitly stated (implied acceptable correlation and agreement for clinical use based on regression). | Slope: 0.99, Intercept: -0.42, r: 0.979 (n=308, Range: 2.1-24.2 ng/mL). |
| Method Comparison (vs. Predicate Dimension Tacrolimus (TACR) method) | Not explicitly stated (implied acceptable correlation and agreement for clinical use based on regression). | Slope: 1.00, Intercept: -0.50, r: 0.957 (n=213, Range: 2.6 - 21.7 ng/mL). |
| Calibrator Stability (Unopened) | Stable until expiration date. | Stored frozen (-25 to -15 °C) until expiration date. |
| Calibrator Stability (Thawed) | Stable for 30 days when recapped and stored at 2-8°C. | Assigned values are stable for 30 days when recapped immediately after use and stored at 2-8°C. |
2. Sample Size Used for the Test Set and Data Provenance
-
Precision: No specific sample size for a single "test set" is provided, but reproducibility testing was conducted using different levels (1.8 to 27.4 ng/mL).
-
Linearity: No specific sample size mentioned, but the study was performed by mixing a sample with a known high tacrolimus concentration (40.3 ng/mL) with a normal patient pool with no tacrolimus in various ratios.
-
Specificity: No specific sample size mentioned for the interference or cross-reactivity studies, beyond stating "compounds tested" and "major tacrolimus metabolite... as well as minor metabolites."
-
Recovery: No specific sample size mentioned, but "samples spiked with USP tacrolimus" were used.
-
Method Comparison:
- Versus LC-MS/MS: n = 315 patient samples.
- Versus ARCHITECT Tacrolimus Assay: n = 308 patient samples.
- Versus Dimension Tacrolimus (TACR) method: n = 213 patient samples.
-
Data Provenance: The document states that method comparison studies, and some precision testing, were performed at two external evaluation sites in addition to internal testing. Patient samples for method comparison included a "nearly equal distribution of liver and kidney transplant patients." The country of origin is not specified but is implied to be within the scope of Siemens Healthcare Diagnostics operations, likely in the US given the FDA submission. The studies were retrospective in the sense that existing patient samples (whole blood) were used, though the testing itself was prospective for the device evaluation.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The concept of "experts" for ground truth (e.g., radiologists) is not applicable here as this is an in vitro diagnostic (IVD) assay. The ground truth for the method comparison studies was established using:
- LC-MS/MS reference method: This is an analytical "gold standard" for measuring tacrolimus concentrations.
- Predicate devices: ARCHITECT Tacrolimus Assay (K070820) and Dimension Tacrolimus (TACR) method (K060502), which are already legally marketed and accepted methods.
The experts involved would be the laboratory personnel performing these high-complexity reference and predicate assays, who are implicitly qualified to conduct such tests according to established laboratory practices. No specific number or qualifications (e.g., "10 years of experience") are provided for these individuals.
4. Adjudication Method for the Test Set
Not applicable. This is an IVD assay where quantitative measurements are obtained. Adjudication typically refers to resolving discrepancies between multiple human readers in image-based diagnostics. Here, the "truth" is established by a reference method or comparison to established predicate devices.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is an in vitro diagnostic (IVD) assay and does not involve human readers interpreting images, nor does it involve AI assistance in the context of imaging diagnostics. The comparison is between the new device's analytical performance and established reference/predicate methods.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Yes, the studies reported are essentially standalone performance evaluations of the Dimension Tacrolimus Flex® Reagent Cartridge (TAC) system. The described testing (precision, linearity, specificity, recovery, method comparison) evaluates the analytical performance of the automated assay system itself. Human intervention is limited to standard laboratory procedures like sample preparation, loading, and result interpretation, not in an "AI-assisted reader" capacity.
7. The Type of Ground Truth Used
- Analytical Measuring Range (Linearity) & Recovery: "USP tacrolimus" (United States Pharmacopeia tacrolimus), which refers to highly pure, standardized tacrolimus used to spike samples to known concentrations.
- Method Comparison:
- LC-MS/MS Assay for Tacrolimus: This is considered the analytical "reference method" and serves as a highly accurate ground truth.
- Predicate Devices: ARCHITECT Tacrolimus Assay and Dimension Tacrolimus (TACR) method. While not a "ground truth" in the strictest sense of an absolute standard, they serve as the established clinical "truth" for comparison, demonstrating substantial equivalence.
8. The Sample Size for the Training Set
No explicit "training set" or sample size for it is mentioned. This is a chemical assay, not a machine learning algorithm in the typical sense that requires a separate training set for model development. The development of the assay reagents and protocols would involve internal optimization, but this is not typically referred to as a "training set" in the context of regulatory submissions for IVD devices.
9. How the Ground Truth for the Training Set Was Established
Since there is no explicit "training set" or machine learning model, this question is not directly applicable. The development and optimization of the assay would rely on:
- Understanding of tacrolimus chemistry and pharmacology: To design reagents that specifically target tacrolimus.
- Standards and reference materials: Such as purified tacrolimus (used for calibrators and recovery studies) to ensure accurate concentration measurements.
- Internal validation and optimization studies: To fine-tune reagent concentrations, reaction conditions, and instrument parameters to achieve desired analytical performance characteristics.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. Inside the circle is a stylized image of three human profiles facing to the right, with three curved lines above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
SIEMENS HEALTHCARE DIAGNOSTICS KATHLEEN DRAY-LYONS REGULATORY AFFAIRS SPECIALIST 500 GBC DRIVE NEWARK DE 19714
November 4, 2015
Re: K150168
Trade/Device Name: Dimension Tacrolimus Flex® Reagent Cartridge (TAC) Dimension Tacrolimus Calibrator (TAC CAL) Regulation Number: 21 CFR 862.1678 Regulation Name: Tacrolimus test system Regulatory Class: II Product Code: MLM, JIT Dated: October 30, 2015 Received: November 2, 2015
Dear Ms. Dray-Lyons:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Katherine Serrano -S
For: Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K150168
Device Name
Dimension Tacrolimus Flex® Reagent Cartridge (TAC) Dimension Tacrolimus Calibrator (TAC CAL)
Indications for Use (Describe)
The Dimension Tacrolimus Flex® Reagent Cartridge (TAC) is an in vitro diagnostic test for the quantitative measurement of tacrolimus in human whole blood on the Dimension® clinical chemistry system. Measurements of tacrolimus are used as an aid in the management of tacrolimus therapy in renal and hepatic transplant patients.
The Dimension Tacrolimus Callbrator (TAC CAL) is an in vitro diagnostic product for the Tacrolimus (TAC) method on the Dimension® clinical chemistry system.
| Type of Use (Select one or both, as applicable) |
|---|
| ☒ Prescription Use (Part 21 CFR 801 Subpart D) |
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
K150168 The assigned 510(k) number is:
-
- Submitter
Siemens Healthcare Diagnostics Inc. P.O. Box 6101 Newark, DE 19714
- Submitter
Tel: 302-631-9352 FAX # 302-631-6299
Contact Person: Anna Marie (Kathy) Ennis Date of Preparation: September 23, 2015
2. Device
| Name of Device: | Dimension Tacrolimus Flex® reagent cartridge (TAC) |
|---|---|
| Common Name: | Tacrolimus Assay |
| Classification Name: | Tacrolimus Test System (21CFR§862.1678) |
| Regulatory Class: | II |
| Product Code: | MLM |
| Name of Device: | Dimension Tacrolimus Calibrator (TAC CAL) |
| Common Name: | Tacrolimus Calibrator (TAC CAL) |
| Classification Name: | Calibrator Secondary (21CFR§862.1150) |
| Regulatory Class: | II |
| Product Code: | JIT |
3. Predicate Device
Abbott® ARCHITECT Tacrolimus Test System - K070820 Dimension® Tacrolimus Calibrator – K060503 Reference Device: LC/MS -MS Assay for Tacrolimus
4. Guidance Document(s):
"Class II Special Controls Guidance Document: Cyclosporine and Tacrolimus Assays; Guidance for Industry and FDA" 11/16/2002
"Bundling Multiple Devices or Multiple Indications in a Single Submission" - 06/22/2007
"The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] Guidance for Industry and Food and Drug Administration Staff" 7/28/2014
{4}------------------------------------------------
5. Device Description(s):
Dimension Tacrolimus Flex® Reagent Cartridge (TAC)
The automated Dimension® TAC method uses an immunoassay technique in which free and tacrolimus-bound antibody-enzyme conjugate is separated using magnetic particles. The assay is performed using a method specific Flex® reagent cartridge. The Flex® cartridge contains a pretreatment reagent, antibody-ß-galactosidase conjugate, tacrolimus immobilized on chromium dioxide particles, chlorophenol red ß-d-galactopyranoside (CPRG) substrate, and diluent to hydrate the tablets.To perform the TAC assay, a sample cup with a whole blood sample to be analyzed and a TAC Flex® reagent cartridge are placed appropriately on the Dimension® system. The
Dimension® system mixes and lyses the whole blood sample in the presence of pretreatment reagent 1. This reagent contains a displacer which acts to displace tacrolimus in the sample from binding proteins. The lysed sample is then mixed with the antibody enzyme conjugate. The tacrolimus present in the sample is bound by the tacrolimus antibody. Magnetic particles coated with tacrolimus are added to bind free (unbound) antibody-enzyme conjugate. The reaction mixture is then separated magnetically. Following separation, the supernatant containing the tacrolimus-antibody-enzyme complex is transferred to a cuvette and mixed with the substrate; chlorophenol red ß-d-galactopyranoside (CPRG). ß-galactosidase catalyzes the hydrolysis of CPRG to produce chlorophenol red (CPR) that absorbs light maximally at 577 nm. The change in absorbance at 577 nm due to the formation of CPR is directly proportional to the amount of tacrolimus in the patient's sample and is measured using a bichromatic (577, 700 nm) rate technique.
| TACR+Ab-β-gal | → TACR-Ab-β-gal + Ab-β+gal |
|---|---|
| TACR-Ab-β-gal + Ab-β-gal + CrO2-TACR | Magnetic separation → TACR-Ab-β-gal (transferred to cuvette) |
| CPRG (non-absorbing at 577 nm) | → TACR-Ab-β-gal → CPR (absorbs at 577 nm) |
If the calculated tacrolimus concentration in the patient sample exceeds 20 ng/mL, the test is automatically repeated using pretreatment reagent 2 in place of pretreatment reagent 1. Pretreatment reagent 2 does not contain the displacer so the signal produced represents nonanalyte signal. If significant signal is detected during the second analysis, the potential exists that interference from non-specific signal has impacted the accuracy of the initial test. In this situation, the reported test result is suppressed and replaced by a non-reportable test result message.
The reagents for the Tacrolimus assay are packaged in an 8-well Flex® Reagent Cartridge represented (top view) in the drawing below:
| 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 |
|---|---|---|---|---|---|---|---|
| --- | --- | --- | --- | --- | --- | --- | --- |
{5}------------------------------------------------
The reagent placement is as follows:
| Wellsa | Form | Ingredient | Concentrationb | Source |
|---|---|---|---|---|
| 1 | Liquid | Ab-β-galactosidase | c | mouse,monoclonal |
| 2 | Liquid | Pretreatment reagent 2 | ||
| 3,4 | Tabletsd | Tacrolimus-CrO2 | 36 mg/tab | |
| 5,6 | Tabletsd | CPRG | 9.3 mg/tablet | |
| 7 | Liquid | Substrate diluent | ||
| 8 | Liquid | Pretreatment reagent 1 |
Reagent Placement Dimension® Tacrolimus Flex® Reagent Cartridge
Wells are numbered consecutively from the wide end of the cartridge. a.
b. Nominal value in hydrated cartridge.
Antibody titer and conjugate activity may vary from lot to lot. C.
Tablets contain excipients, buffers, and stabilizers. d.
Dimension Tacrolimus Calibrator (TAC CAL)
Dimension® Tacrolimus (TAC) Calibrator is five level frozen liquid, whole blood hemolysate containing purified tacrolimus. The provided in 4.0 mL vials, 2 vials per level. There are five calibrator levels per kit which span the assay range for the Dimension® Tacrolimus (TAC) assay. Calibrators are filled with 1.0 mL per vial except for Level 1 calibrator which contains 2.0 mL per vial.
6. Intended Use/ Indications for Use
Dimension Tacrolimus Flex® Reagent Cartridge (TAC)
The Dimension® Tacrolimus Flex Reagent Cartridge (TAC) method is an in vitro diagnostic test for the quantitative measurement of tacrolimus in human whole blood on the Dimension® clinical chemistry system.
Measurements of tacrolimus are used as an aid in the management of tacrolimus therapy in renal and hepatic transplant patients.
Dimension Tacrolimus Calibrator (TAC CAL)
The Dimension® Tacrolimus Calibrator (TAC CAL) is an in vitro diagnostic product for the calibration of the Tacrolimus (TAC) method on the Dimension® clinical chemistry system.
{6}------------------------------------------------
7. Comparison of Technological Characteristics
A comparison summary of the features of the products is included in the following table.
| Item | DeviceDimension Tacrolimus Flex®Reagent Cartridge (TAC) | PredicateARCHITECT Tacrolimus Assay(K070820) |
|---|---|---|
| Similarities | ||
| Intended Use | an in vitro diagnostic test for thequantitative measurement oftacrolimus in human whole blood | an in vitro diagnostic test for thequantitative measurement oftacrolimus in whole blood. |
| Indications for Use | used as an aid in the managementof tacrolimus therapy in renal andhepatic transplant patients | used as an aid in the managementof liver and kidney allograft patientsreceiving tacrolimus therapy |
| Sample Types | Whole blood in EDTA | Whole blood in EDTA |
| High Sample Dilution | Manual | Manual |
| Assay Type | Immunoassay | Immunoassay |
| Differences | ||
| Instrument | The Dimension® clinical chemistrysystem. | The Abbott ARCHITECT iSystem |
| Sample Pretreatment | No manual pretreatment | Manual pre-treatment |
| Cross reactivity Profile | ||
| M-I | 1% | 8% |
| M-II | 18% | 94% |
| M-III | 15% | 45% |
| M-IV | 99% | 9% |
| M-V | 1% | Not Available |
| M-VI | 1% | Not Available |
| M-VII | 43% | Not Available |
| M-VIII | 0% | Not Available |
| Measuring Range | 1.0 - 30 ng/mL | 2 - 30 ng/mL |
{7}------------------------------------------------
| Calibrator: | ||
|---|---|---|
| Item | DeviceDimension Tacrolimus Calibrator | PredicateTACR CAL (K060503) |
| Similarities | ||
| Intended Use | An in vitro diagnostic product forthe calibration of the Tacrolimus(TAC) method on the Dimension®clinical chemistry system. | An in vitro diagnostic productintended to be used tocalibrate the Tacrolimusmethod (TACR) for theDimension® clinical chemistrysystem. |
| Form | Frozen Liquid | Frozen Liquid |
| Matrix | Whole blood hemolysate | Whole blood hemolysate |
| Levels | Five | Five |
| Traceability | Purified tacrolimus | Purified tacrolimus |
| Differences | ||
| Assignment | Assigned for Dimension® TAC | Assigned for Dimension®TACR |
| Target ConcentrationRange | Level 1: - 0.5 to + 0.5 ng/mLLevel 2: 2.7 to 4.2 ng/mLLevel 3: 5.8 to 7.3 ng/mLLevel 4: 11.6 to 13.6 ng/mLLevel 5: ≥30.0 ng/mL | Level A: - 0.7 to + 0.7 ng/mLLevel 2: 2.5 to 4.0 ng/mLLevel 3: 5.5 to 7.0 ng/mLLevel 4: 11.0 to 13.0 ng/mLLevel 5: 31.0 to 34.0 ng/mL |
8. Performance Data
Precision
Reproducibility testing was conducted for the Dimension ® TAC method, at two external evaluation sites in addition to our internal testing. This testing was done in accordance with the CLSI/NCCLS Approved Guideline for User Evaluation of Precision Performance of Clinical Chemistry Devices EP5-A2. Typical precision at levels ranging from a mean of 1.8 to 27.4 ng/mL was ≤ 6.6% CV for Repeatability and ≤ 13.1% CV for Within Lab.
Linearity
The analytical measuring range of the TAC method was confirmed by the establishment of the Limit of Quantitation (LoQ), for the low end, and by a linearity study. The linearity study was performed by mixing of a sample with a known native high tacrolimus concentration (40.3 ng/mL) with a normal patient pool with no tacrolimus in various ratios. The testing procedure was based on the CLSI Guideline EP 06-A Evaluation of the Linearity of Quantitative Measurement. A linear regression, and a 200 - and 35 -order polynomial regressions of the mean observed analyte values vs. expected concentrations were performed. The assay range for the Dimension® TAC method was established as 1.0 - 30.0 ng/mL by these two studies.
Specificity
Interference and cross-reactivity testing was performed according to CLSI/NCCLS Approved Guideline Interference Testing in Clinical Chemistry: EP7-A2. The compounds tested included endogenous compounds, drugs that are commonly co-administered with tacrolimus, and the anticoagulant EDTA. The cross-reactivity testing was done with major tacrolimus metabolite, M I (13-O-desmethyl-tacrolimus), M II (31-O-desmethyl-tacrolimus); as well as minor metabolites M III (15-O-desmethyl-tacrolimus), M IV (12-O-hydroxyl-tacrolimus), M V (15,31-O-didesmethyltacrolimus); and second pass metabolites, M VI (13,31-O-didesmethyl-tacrolimus), M VII (13,15-O-didesmethyl-tacrolimus) and M VIII(unknown name). Testing was done as recommended in the FDA Guidance document "Class II Special Controls Guidance Document: cyclosporine and Tacrolimus Assays; Guidance for Industry and FDA" – 09/16/2002. These studies showed that the Dimension® TAC method has low cross-reactivity to the major (M-1) tacrolimus metabolite,
{8}------------------------------------------------
and partial cross reactivity to the minor and second pass metabolites of lower prevalence. Coadministered drugs and other compounds tested for interference, exhibited less than 10% bias at the levels tested.
Recovery
The recovery capability of the Dimension® TAC assay was challenged across the assay range using samples spiked with USP tacrolimus. All samples met the acceptance criterion of ≤ 10% mean bias in recovery.
Method Comparison
A method comparison study was performed at two external evaluation sites and internally per the "FDA Class II Special Controls Guidance document: Cyclosporine and Tacrolimus Assays: Guidance for Industry and FDA". Samples included a nearly equal distribution of liver and kidney transplant patients which covered >80% of the assay range (1.0 - 30 ng/mL). CLSI Guideline EP-9A2 "Method Comparison and Bias Estimation using Patient Samples" was used to develop the study design.
As recommended by the guidance, samples were evaluated, at two external sites as well as internally, versus the predicate, ARCHITECT Tacrolimus Assay (K070820), and an LCMS reference method. An additional comparative method. Dimension® TACR, was also run at two sites. Analysis was done by Passing - Bablok regression. Least Squares regression was used to compute the correlation coefficient.
| Dimension Tacrolimus (TAC) | Slope | Intercept | r | n | Range |
|---|---|---|---|---|---|
| method vs | (ng/mL) | ||||
| Reference Method LC-MS/MS | 1.04 | -0.30 | 0.966 | 315 | 1.3-24.9 |
| ARCHITECT Tacrolimus Assay(K070820) | 0.99 | -0.42 | 0.979 | 308 | 2.1-24.2 |
| Dimension Tacrolimus (TACR) method(K060502) | 1.00 | -0.50 | 0.957 | 213 | 2.6 - 21.7 |
The samples range for the proposed Dimension TAC assay, in these studies, was 1.2 - 25.6 ng/mL.
Calibrator Traceability
Siemens Tacrolimus (TAC) Calibrator is traceable to purified tacrolimus drug through standardization of the LC-MS/MS reference method with gravimetrically prepared standards.
Calibrator Stability
Unopened vials of Siemens Tacrolimus (TAC) Calibrator are stored frozen (-25 to -15 °C) until the expiration date printed on each carton. Once thawed, assigned values are stable for 30 days when recapped immediately after use and stored at 2-8°C.
{9}------------------------------------------------
9. Conclusion:
The Siemens Healthcare Diagnostics Dimension® TAC method and the ARCHITECT Tacrolimus Assay (K070820) are substantially equivalent based on their intended use, indications for use and performance characteristics. In performance studies, summarized above, the Dimension® Tacrolimus methods demonstrated performance similar to the predicate ARCHITECT Tacrolimus Assay (K070820).
The modified Siemens Healthcare Diagnostics Dimension ® TAC calibrator and the predicate Dimension® TACR calibrator (K060503) are also substantially equivalent in design, intended use, principle and performance.
Anna Marie Kathleen Ennis Regulatory Affairs Manager
{10}------------------------------------------------
§ 862.1678 Tacrolimus test system.
(a)
Identification. A tacrolimus test system is a device intended to quantitatively determine tacrolimus concentrations as an aid in the management of transplant patients receiving therapy with this drug. This generic type of device includes immunoassays and chromatographic assays for tacrolimus.(b)
Classification. Class II (special controls). The special control is “Class II Special Controls Guidance Document: Cyclosporine and Tacrolimus Assays; Guidance for Industry and FDA.” See § 862.1(d) for the availability of this guidance document.