(63 days)
LLX240AB01, LTX240AA01-A
Not Found
No
The summary describes a standard digital X-ray detector and its performance metrics (MTF, DQE, NPS, SNR). It mentions image processing by separate software but does not indicate the use of AI or ML within the detector itself or the described workflow. There is no mention of AI, DNN, or ML in the text.
No.
This device is an X-ray detector used for imaging and diagnosis, not for treating diseases or conditions.
Yes
The "Intended Use / Indications for Use" states that the device is "intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures." Additionally, the "Device Description" mentions that it "can be utilized to capture and digitalize X-ray images for radiographic diagnosis."
No
The device description explicitly states it is a "digital solid state X-ray detector" based on "flat-panel technology" and consists of a "scintillator coupled to an a-Si TFTsensor." This clearly describes hardware components, not a software-only device.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly states it's for "digital imaging solution designed for general radiographic system for human anatomy." This describes an in-vivo imaging device, not a device used to examine specimens taken from the body (which is the definition of in vitro).
- Device Description: The description details an X-ray detector that captures and digitalizes X-ray images for radiographic diagnosis. This is consistent with in-vivo imaging.
- Input Imaging Modality: The input modality is X-ray, which is used for imaging the inside of the body directly.
- Anatomical Site: The anatomical site is "human anatomy," indicating imaging of the body itself.
IVD devices are used to examine specimens such as blood, urine, or tissue outside of the body to provide information for diagnosis, monitoring, or screening. This device does not fit that description.
N/A
Intended Use / Indications for Use
1717SGN and 1717SCN Digital Flat Panel X-Ray Detector are indicated for digital imaging solution designed for general radiographic system for human anatomy. They are intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. Not to be used for mammography.
Product codes
MQB
Device Description
1717SGN and 1717SCN digital solid state X-ray detectors are based on flat-panel technology. Both radiographic image detector and processing units consist of a scintillator coupled to an a-Si TFTsensor. A digital flat panel X-ray detector needs to be integrated with a radiographic imaging system. It can be utilized to capture and digitalize X-ray images for radiographic diagnosis. The RAW files can be further processed as DICOM compatible image files by separate console SW (not part of this 510k submission) for a radiographic diagnosis and analysis.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
X-ray
Anatomical Site
human anatomy
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Clinical images are taken from both subject devices (1717SGN and 1717SCN) and reviewed by a licensed US radiologist. Both the test subject (1717SGN and 1717SCN) and control group (LLX240AB01 and LTX240AA01-A) are evaluated and compared by taking sample radiographs of similar age groups and anatomical structures in accordance with the test protocol of diagnostic radiography evaluation procedure.
Summary of Performance Studies
Non-clinical test report and clinical consideration report.
The non-clinical test report contains the MTF, DQE and NPS test results of 1717SGN and 1717SCN by using the identical test equipment and same analysis method described by IEC 6220-1. The comparative result of the MTF test for 1717SGN and 1717SCN detector with respective predicate demonstrated that the MTF of the both 1717SGN and 1717SCN performed better compared with the predicate devices. The DQE represents the ability to visualize object details of a certain size and contrast. 1717SGN demonstrated higher DQE performance than LLX240AB01 at all spatial frequencies as well as superior Signal-to Noise Ratio (SNR) transfer from the input to the output of the detector as a function of frequency. 1717SCN has higher DQE performance at high spartial frequencies, especially from 11p/mm to 3 lp/mm.
Clinical images are taken from both subject devices and reviewed by a licensed US radiologist to render an expert opinion. Both the test subject (1717SGN and 1717SCN) and control group (LLX240AB01 and LTX240AA01-A) are evaluated and compared by taking sample radiographs of similar age groups and anatomical structures in accordance with the test protocol of diagnostic radiography evaluation procedure. Images obtained with the 1717SGN / 1717SCN are comparable or superior to the same view obtained from a similar patient with the LLX240AB01 / LTX240AA01-A. In general, both the spatial resolution and soft tissue contrast are superior using the 1717SCN / 1717SCN digital flat panel detector. Specifically, the soft tissues on extremity films were seen with better clarity.
Key Metrics
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 892.1680 Stationary x-ray system.
(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
0
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of three human profiles facing to the right, stacked on top of each other. The profiles are rendered in a dark color, contrasting with the white background.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 27, 2015
Rayence Co., Ltd. % Mr. Dave Kim Medical Device Regulatory Affairs Mtech Group 8310 Buffalo Speedway HOUSTON TX 77025
Re: K150150
Trade/Device Name: 1717SGN / 1717SCN Digital Flat Panel X-ray Detector Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: January 20, 2015 Received: January 23, 2015
Dear Mr. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
1
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K150150
Device Name
1717SGN / 1717SCN Digital Flat Panel X-ray Detector
Indications for Use (Describe)
1717SGN and 1717SCN Digital Flat Panel X-Ray Detector are indicated for digital imaging solution designed for general radiographic system for human anatomy. They are intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. Not to be used for mammography.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CER 801 Subpart D) | |
---|---|
Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
1. 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date 510k summary prepared: March 19, 2015
Submitter's Name, address, telephone number, a contact person:
Submitter's Name : | Rayence Co., Ltd. |
---|---|
Submitter's Address: | 14, Samsung 1-ro 1-gil, Hwaseong-si, Gyeonggi-do, Korea |
Submitter's Telephone: | +82-31-8015-6459 |
Contact person: | Mr. Kee Dock Kim / RA Team Manager / +82-31-8015-6459 |
Official Correspondent: | Dave Kim (davekim@mtech-inc.net) |
(U.S. Designated agent) | |
Address: | 8310 Buffalo Speedway, Houston, TX 77025 |
Telephone: | +713-467-2607 |
Fax: | +713-583-8988 |
Name of the device, including the trade or proprietary name if applicable, the common or usual name and the classification name, if known:
Trade/proprietary name: | 1717SGN / 1717SCN |
---|---|
Common Name: | Digital Flat Panel X-ray Detector |
Classification Name : | 21CFR892.1680 / Stationary x-ray system |
Product Code: | MQB |
Predicate Device :
Manufacturer | : Samsung Mobile Display Co., Ltd. |
---|---|
Device | : LLX240AB01 & LTX240AA01-A |
510(k) Number | : K102587 & K090742 |
Decision Date | : DEC 1, 2010 & SEP 18, 2009 |
4
2. Device Description
1717SGN and 1717SCN digital solid state X-ray detectors are based on flat-panel technology. Both radiographic image detector and processing units consist of a scintillator coupled to an a-Si TFTsensor. A digital flat panel X-ray detector needs to be integrated with a radiographic imaging system. It can be utilized to capture and digitalize X-ray images for radiographic diagnosis. The RAW files can be further processed as DICOM compatible image files by separate console SW (not part of this 510k submission) for a radiographic diagnosis and analysis.
3. Indication for use
1717SGN and 1717SCN Digital Flat Panel X-Ray Detector is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. Not to be used for mammography.
4. Summarv of Design Control Risk management
1717SGN and 1717SCN digital X-ray detectors are the modification version of LLX240AB01(K102587) and LTX240AA01-A(K090742), respectively.
1717SGN and 1717SCN are slightly smaller in dimensions and substantially lighter in weight than LLX240AB01(K102587) and LTX240AA01-A(K090742).
The risks and the hazardous impact of the device modification were analyzed with FMEA method. The specific risk control and protective measures to mitigate the risks from the modification were reviewed and implemented in the new product design phase. The overall assessment concluded that all risks and hazardous conditions identified arising from the design change were successfully mitigated and accepted.
5. Summary of the technological characteristics of the device compared to the predicate device:
1717SGN and 1717SCN detector described in this 510(k) have the same indications for use and similar technical characteristics as each predicate device, LLX240AB01(K102587) and LTX240AA01-A(K090742) flat panel detector, manufacturerd by Samsung Mobile Display Co., Ltd.
5
5.1 Scintillator layer
1717SGN and 1717SCN detector have same Hardware, Software and components except the type of the scintillator layer.
( *scintillator : a phosphor that produces scintillations)
Scintillator Layer Type | Proposed | Predicate |
---|---|---|
Gd2O2S:Tb | ||
(Gadolinium Oxysulfide) | 1717SGN | LLX240AB01 |
CsI | ||
(Cesium Iodide) | 1717SCN | LTX240AA01-A |
Dimensions and Weight 5.2
| | | Proposed
1717SGN | Predicate
LLX240AB01 |
|--------------|------------|---------------------|-------------------------|
| | | 1717SCN | LTX240AA01-A |
| Detector | Model name | 1717SGN | LLX240AB01 |
| | W x L x H | 460 X 460 X 15.5 | 500 x 496.6 x 45mm |
| | Weight | 4 kg | 14.5kg |
| Power supply | Model name | RS1717 | SPS01 |
| | W x L x H | 188 X 92 X 41.5 | 290 x 245 x 68 mm |
| | Weight | 0.5 kg | 4 kg |
5.3 Power Requirements (Power supply)
Proposed | Predicate | ||
---|---|---|---|
1717SGN | |||
1717SCN | LLX240AB01 | ||
LTX240AA01-A | |||
Power | |||
supply | Model name | RS1717 | SPS01 |
W x L x H | AC 100 - 240 V~, | ||
50/60Hz | AC100-120 / 220-240 V~, | ||
50/60Hz |
5.4 Proposed Labeling
Proposed | Predicate | |
---|---|---|
1717SGN | LLX240AB01 | |
1717SCN | LTX240AA01-A |
6
Image /page/6/Picture/1 description: The image shows labels for Rayence and Samsung Mobile Display Co., Ltd products. The Rayence labels include models 1717SGN, 1717SCN, and RS1717, identifying medical image processing units and a power supply, with ratings around 24V and compliance certifications. The Samsung labels feature models LLX240AB01, LTX240AA01-A, and SPS01, detailing digital flat panel X-ray detectors and a power supply, with weight and dimension specifications, along with caution notices for power supply usage.
5.5 Comparison table
Proposed | Predicate | ||
---|---|---|---|
Characteristic | 1717SGN | LLX240AB01 | |
1717SCN | LTX240AA01-A |
7
510(k) number | LLX240AB01 | K102587 | ||
---|---|---|---|---|
LTX240AA01-A | K090742 | |||
Intended Use | 1717SGN and 1717SCN Digital | |||
Flat Panel X-Ray Detector is | ||||
indicated for digital imaging | ||||
solution designed for general | ||||
radiographic system for human | ||||
anatomy. It is intended to replace | ||||
film or screen based radiographic | ||||
systems in all general purpose | ||||
diagnostic procedures. Not to be | ||||
used for mammography. | Digital Flat Panel X-Ray Detector is | |||
indicated for digital imaging solution | ||||
designed for general radiographic | ||||
system for human anatomy. It is | ||||
intended to replace film or screen based | ||||
radiographic systems in all general | ||||
purpose diagnostic procedures. Not to | ||||
be used for mammography. | ||||
Manufacturer | Rayence Co., Ltd. | Samsung Mobile Display Co., Ltd. | ||
Detector Type | Amorphous Silicon, TFT | Amorphous Silicon, TFT | ||
Scintillator | 1717SGN | |||
1717SCN | Gd2O2S:Tb | |||
CsI:Tl | LLX240AB01 | |||
LTX240AA01-A | Gd2O2S:Tb | |||
CsI:Tl | ||||
Imaging Area | 17 x 17 inches | |||
Pixel matrix | 3072 x 3072 (9 million) | |||
Pixel pitch | 127 µm | 143 µm | ||
Resolution | 3.9 lp/mm | 3.5 lp/mm | ||
A/D conversion | 14 / 16 bit | 14 bit | ||
Preview time | ≤2 | 5 seconds per Image | ||
Data output | RAW | |||
*The RAW files are convertible into | ||||
DICOM 3.0 by console S/W | RAW | |||
*The RAW files are convertible into DICOM | ||||
3.0 by console S/W | ||||
Dimensions | 460 × 460 × 15.5 mm | 500 x 496.6 x 45 mm | ||
Weight | 4 kg | 14.5 kg | ||
Application | General Radiology system or Portable | |||
system | ||||
Available with upright stand, table, | ||||
universal stand. | General Radiology system or Portable system | |||
Available with upright stand, table, universal | ||||
stand. |
6.System requirements to operate with other radiographic system components
-
- Recommended Generator Specification:
Energy range: 40~150kVp
- Recommended Generator Specification:
-
mA range 10~1000mA(depending on the generator power)
[NOTE] To our best knowledge, the detector is compatible with the X-ray generators with the specifications described above. If you have questions regarding the compatibility issue for other generators, please contact your Rayence representative. -
- Software Development Kit (Library's Software Interface) for third party viewing software Peripheral hardware: DaVinci detector connected via Ethernet card.
8
CPU/Motherboard: 1 GHz (or faster) CPU, cache L2≥256KB RAM: ≥512MB, 2GB recommended Hard drive: no special requirements Display: At least 17-inch monitor is required. 21-inch monitor is suggested. OS: Windows 2K or XP professional Development environment: MS Visual C++ 6.0 or MS Visual Studio 2005
7. Summary of Performance Testing
1717SGN and 1717SCN Digital Flat Panel X-Ray Detector have the same indications for use, material, form factor, performance, and safety characteristics compared to the predicate devices, LLX240AB01 and LTX240AA01-A, respectively.
The non-clinical test report and clinical consideration report for each subject device were prepared and submitted to FDA separately to demonstrate the substantial equivalency of the subject devices compared to each respective predicate device. The non-clinical test report contains the MTF, DQE and NPS test results of 1717SGN and 1717SCN by using the identical test equipment and same analysis method described by IEC 6220-1. The comparative result of the MTF test for 1717SGN and 1717SCN detector with respective predicate demonstrated that the MTF of the both 1717SGN and 1717SCN performed better compared with the predicate devices. The DQE represents the ability to visualize object details of a certain size and contrast. 1717SGN demonstrated higher DQE performance than LLX240AB01 at all spatial frequencies as well as superior Signal-to Noise Ratio (SNR) transfer from the input to the output of the detector as a function of frequency. 1717SCN has higher DQE performance at high spartial frequencies, especially from 11p/mm to 3 lp/mm.
To further demonstrate the substantial equivalency of two devices, clinical images are taken from both subject devices and reviewed by a licensed US radiologist to render an expert opinion. Both the test subject (1717SGN and 1717SCN) and control group (LLX240AB01 and LTX240AA01-A) are evaluated and compared by taking sample radiographs of similar age groups and anatomical structures in accordance with the test protocol of diagnostic radiography evaluation procedure.
After comparing a broad review of plain radiographic images taken with 1717SGN / 1717SCN and LLX240AB01 / LTX240AA01-A, the images obtained with the 1717SGN / 1717SCN are comparable or superior to the same view obtained from a similar patient with the LLX240AB01 / LTX240AA01-A. In general, both the spatial resolution and soft tissue contrast are superior
9
using the 1717SCN / 1717SCN digital flat panel detector. Specifically, the soft tissues on extremity films were seen with better clarity. There is no difficulty in evaluating a wide range of anatomic structures necessary to provide a correct conclusion.
Clinical images were provided; these images were not necessary to establish substantial equivalence based on the modifications to the device but they provide further evidence in addition to the laboratory performance data to show that the subject devices operate as indicated.
Based on the non-clinical bench testing supplemented by the clinical consideration and the outcome of a comparative review by an expert for both devices, the sponsor demonstrated that the subject devices operate as indicated and are substantially equivalent to the predicate devices in terms of safety and effectiveness.
The manufacturing facility is in conformance with the design control procedure requirements and the relevant EPRC standards as specified in 21 CFR 802.30 and the records are available for review.
8. Summary for any testing in the submission:
Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1: 2005 + CORR.1(2006) + CORR(2007) (Medical electrical equipment Part 1:General requirements for basic safety and essential performance) was performed, and EMC testing were conducted in accordance with standard IEC 60601-1-2:2007.
Non-clinical & Clinical considerations according to FDA Guidance for the Submissions of 510(k)'s for Solid State X-ray Imaging Devices" was performed. All test results were satisfactory.
All test results were satisfactory
9. Conclusions:
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Rayence Co., Ltd. concludes that 1717SGN / 1717SCN are safe and effective and substantially equivalent in comparison with LLX240AB01 / LTX240AA01-A, the predicate devices as described herein.