K Number
K150096
Date Cleared
2015-03-04

(43 days)

Product Code
Regulation Number
862.3610
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Healgen MDMA (Ecstasy) Test is an immunochromatographic assay for the qualitative determination of Methylenedioxymethamphetamine in human urine at a Cut-Off concentration of 500 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result: GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.

Healgen Phencyclidine Test is an immunochromatographic assay for the qualitative determination of Phencyclidine in human urine at a Cut-Off concentration of 25 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.

Device Description

Healgen MDMA (Ecstasy) Test and Healgen Phencyclidine Test are immunochromatographic assays for Methylenedioxymethamphetamine and Phencyclidine. Each assay test is a lateral flow system for the qualitative detection of Methylenedioxymethamphetamine and Phencyclidine (target analyte) in human urine. The products are in vitro diagnostic devices, which come in the form of: Strips, Cassettes, DipCards, or Cups. Each product contains a Test Device (in one of the four formats), and a package Each test device is sealed with a desiccant in an aluminum pouch. insert.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Healgen MDMA (Ecstasy) Test and Healgen Phencyclidine Test, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly stated in a single section as pass/fail thresholds. Instead, the performance characteristics studies demonstrate the device's accuracy and reliability against the established cut-off values as compared to GC/MS, which serves as the gold standard.

Healgen MDMA (Ecstasy) Test Performance Summary (Across all formats: Strip, Cassette, Cup, Dip Card)

Acceptance Criteria (Implied)Reported Device Performance (Summary from tables)
Precision: - Samples < -25% Cut-off: Negative result - Samples > +25% Cut-off: Positive result- Samples at Cut-off: Expected mix of positive/negativeMDMA (Ecstasy):- Samples at -100%, -75%, -50%, -25% Cut-off: 50-/0+ (All negative)- Samples at +25%, +50%, +75%, +100% Cut-off: 50+/0- (All positive)- Samples at Cut-off: Varied (e.g., Strip, Cassette: 22-/28+; Dip Card: 24-/26+; Cup: 30-/20+) - Indicates appropriate sensitivity around the cut-off.
Cut-off Verification: - All samples at and above +25% cut-off: Positive- All samples at and below -25% cut-off: NegativeMDMA (Ecstasy): Verified (all positive at/above +25% cut-off, all negative at/below -25% cut-off).Phencyclidine: Verified (all positive at/above +25% cut-off, all negative at/below -25% cut-off).
Interference: No clinically significant interference from common substances at 100 µg/mL.No differences observed for different formats. Numerous compounds listed showed no interference at 100 µg/mL.
Specificity: Cross-reactivity analysis as expected.MDMA (Ecstasy): 3,4-Methylenedioxyamphetamine HCl (MDA) 17% cross-reactivity, 3,4-Methylenedioxyethylamphetamine (MDEA) 167% cross-reactivity. d-methamphetamine 20% cross-reactivity. d-, l-amphetamine, l-methamphetamine not detected at >100,000 ng/mL.Phencyclidine: 4-Hydroxy Phencyclidine 28% cross-reactivity.
Effect of Urine Specific Gravity and pH: Accurate results across a range of specific gravities (1.000-1.035) and pH (4-9).Results all positive for samples at and above +25% cut-off and all negative for samples at and below -25% Cut-Off. No differences observed for different formats.
Method Comparison (Professional User): High agreement with GC/MS.Each format across MDMA and Phencyclidine tests showed high agreement with GC/MS, with discordant results primarily near the cut-off concentrations (as expected for qualitative tests). Example for MDMA Strip, Viewer A: 13/16 (81.25%) agreement at Near Cutoff Positive, 24/24 (100%) at High Positive. 10/10 (100%) at Negative, 15/15 (100%) at Low Negative.
Lay-User Performance: High percentage of correct results, easy-to-follow instructions.MDMA (Ecstasy): - Samples at -100%, -75%, -50% Cut-off: 100% correct negative.- Samples at -25% Cut-off: 95% correct negative.- Samples at +50%, +75% Cut-off: 100% correct positive.- Samples at +25% Cut-off: 95% correct positive.Similar results for Phencyclidine with some formats showing 90% at -25% and +25% Cut-off. All lay users indicated device instructions can be easily followed (Flesch-Kincaid Grade Level 7).

Healgen Phencyclidine Test Performance Summary (Across all formats: Strip, Cassette, Cup, Dip Card)

Acceptance Criteria (Implied)Reported Device Performance (Summary from tables)
Precision: - Samples < -25% Cut-off: Negative result - Samples > +25% Cut-off: Positive result- Samples at Cut-off: Expected mix of positive/negativePhencyclidine:- Samples at -100%, -75%, -50%, -25% Cut-off: 50-/0+ (All negative)- Samples at +25%, +50%, +75%, +100% Cut-off: 50+/0- (All positive)- Samples at Cut-off: Varied (e.g., Strip: 20-/30+; Cassette: 18-/32+; Dip Card: 22-/28+; Cup: 16-/34+) - Indicates appropriate sensitivity around the cut-off.
Cut-off Verification: - All samples at and above +25% cut-off: Positive- All samples at and below -25% cut-off: NegativeMDMA (Ecstasy): Verified (all positive at/above +25% cut-off, all negative at/below -25% cut-off).Phencyclidine: Verified (all positive at/above +25% cut-off, all negative at/below -25% cut-off).
Interference: No clinically significant interference from common substances at 100 µg/mL.No differences observed for different formats. Numerous compounds listed showed no interference at 100 µg/mL.
Specificity: Cross-reactivity analysis as expected.MDMA (Ecstasy): 3,4-Methylenedioxyamphetamine HCl (MDA) 17% cross-reactivity, 3,4-Methylenedioxyethylamphetamine (MDEA) 167% cross-reactivity. d-methamphetamine 20% cross-reactivity. d-, l-amphetamine, l-methamphetamine not detected at >100,000 ng/mL.Phencyclidine: 4-Hydroxy Phencyclidine 28% cross-reactivity.
Effect of Urine Specific Gravity and pH: Accurate results across a range of specific gravities (1.000-1.035) and pH (4-9).Results all positive for samples at and above +25% cut-off and all negative for samples at and below -25% Cut-Off. No differences observed for different formats.
Method Comparison (Professional User): High agreement with GC/MS.Each format across MDMA and Phencyclidine tests showed high agreement with GC/MS, with discordant results primarily near the cut-off concentrations (as expected for qualitative tests). Example for Phencyclidine Strip, Viewer A: 13/16 (81.25%) agreement at Near Cutoff Positive, 24/24 (100%) at High Positive. 10/10 (100%) at Negative, 15/15 (100%) at Low Negative.
Lay-User Performance: High percentage of correct results, easy-to-follow instructions.MDMA (Ecstasy): - Samples at -100%, -75%, -50% Cut-off: 100% correct negative.- Samples at -25% Cut-off: 95% correct negative.- Samples at +50%, +75% Cut-off: 100% correct positive.- Samples at +25% Cut-off: 95% correct positive.Similar results for Phencyclidine with some formats showing 90% at -25% and +25% Cut-off. All lay users indicated device instructions can be easily followed (Flesch-Kincaid Grade Level 7).

2. Sample size used for the test set and the data provenance

  • Precision Study:

    • Test set size: For each drug (MDMA/Phencyclidine) and each device format, 9 different concentration levels were tested. For each concentration level, 50 tests were performed (2 runs/day for 25 days). So, 9 concentrations * 50 tests/concentration = 450 tests per lot. Since there are 3 lots mentioned per format, this would be 3 lots * 450 tests/lot = 1350 tests per format for each drug.
    • Data Provenance: Samples were prepared by spiking drug in negative urine samples, suggesting these are prospectively prepared, controlled samples. The text does not specify the country of origin, but the submission is to the FDA, implying studies likely conducted in or for the US market. The samples were confirmed by GC/MS.
  • Cut-off Verification Study:

    • Test set size: 150 samples were tested for each drug, equally distributed at 5 concentrations (-50% cut-off, -25% cut-off, cut-off, +25% cut-off, +50% cut-off).
    • Data Provenance: Samples were prepared by spiking drug in negative samples, indicating prospectively prepared, controlled samples.
  • Interference Study:

    • Test set size: Not explicitly stated as a number of samples, but numerous potential interfering substances were added to drug-free urine and urine containing target drugs at 25% above cut-off levels. Each type of sample/interferent was tested using three batches of each device for all formats.
    • Data Provenance: Prepared samples, suggesting prospectively prepared, controlled samples.
  • Specificity Study:

    • Test set size: Not explicitly stated as a number of samples, but drug metabolites and other components were tested using three batches of each device for all formats.
    • Data Provenance: Prepared samples, suggesting prospectively prepared, controlled samples.
  • Effect of Urine Specific Gravity and pH Study:

    • Test set size: Urine samples across a range of specific gravities (1.000-1.035) and pH (4-9) were spiked with target drugs at 25% below and 25% above cut-off levels. These were tested using three batches of each device for all formats.
    • Data Provenance: Prepared samples, suggesting prospectively prepared, controlled samples.
  • Comparison Studies (Professional Users):

    • Test set size: For each drug (MDMA and Phencyclidine) and each format (Strip, Cassette, Cup, Dip Card), 80 unaltered clinical samples were used (40 negative and 40 positive). So, for MDMA, 4 formats * 80 samples/format = 320 samples. For Phencyclidine, 4 formats * 80 samples/format = 320 samples.
    • Data Provenance: Unaltered clinical samples, which were blind labeled. The text does not specify the country of origin, but it is retrospective in nature given they are "unaltered clinical samples." (The samples were analyzed by GC/MS, meaning the true drug concentration was known retrospectively).
  • Lay-User Study:

    • Test set size: 557 lay persons tested MDMA devices and 555 lay persons tested Phencyclidine devices. Total of 1112 individuals. For each drug, samples were prepared at 7 concentration levels (negative, +/-75%, +/-50%, +/-25% of cutoff). For each concentration, 19-20 samples were tested. Thus, for MDMA: (3 * 20 samples) + (1 * 19 samples for -50% cutoff) + (1 * 20 samples for -25% cutoff) + (1 * 20 samples for +25% cutoff) + (1 * 20 samples for +50% cutoff) + (1 * 19 samples for +75% cutoff) = ~139 unique samples tested across different concentrations for each format (though the table headers suggest 20 samples per concentration usually). The 557 and 555 refers to the number of users, implying each user got at least one sample.
    • Data Provenance: Samples were prepared by spiking drugs into drug-free pooled urine specimens, making this a prospectively prepared, controlled sample study. The concentrations were confirmed by GC/MS. The study was performed at "three intended user sites," but country of origin is not specified.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Professional User Comparison Study:

    • Ground Truth: Established by GC/MS (Gas Chromatography/Mass Spectrometry) results. GC/MS is a highly accurate analytical method, considered the gold standard for drug confirmation.
    • Experts for Ground Truth: Not applicable in the traditional sense of human readers. The ground truth is established by a diagnostic laboratory method (GC/MS).
    • Human Readers of the Device: Three "laboratory assistants" were used for each format. Their specific qualifications are not detailed beyond "laboratory assistants."
  • Precision, Cut-off, Interference, Specificity, Effect of Urine Specific Gravity and pH Studies:

    • Ground Truth: Established primarily by GC/MS for confirmation of spiked drug concentrations.
    • Experts for Ground Truth: Not applicable in the traditional sense.
  • Lay-User Study:

    • Ground Truth: Established by GC/MS confirmation of the spiked urine sample concentrations.
    • Experts for Ground Truth: Not applicable.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

  • Precision, Cut-off, Interference, Specificity, Effect of Urine Specific Gravity and pH Studies:

    • No adjudication method is described for interpreting the results of the device itself (e.g., if there were conflicting results). However, for the precision study, three different operators performed the tests, suggesting individual readings were recorded. The summary tables aggregate these results. The ground truth (spiked concentration) was confirmed by GC/MS prior to testing.
  • Comparison Studies (Professional Users):

    • For the discordant results in the comparison studies (Table: Discordant Results of MDMA (Ecstasy) Strip, etc.), it shows individual "Viewer" results (Viewer A, B, C) compared to the GC/MS result. This implies no adjudication method was applied to the device readings to establish a single device result. Instead, the individual results of the three viewers were compared directly to the GC/MS ground truth. They are presenting individual agreement/disagreement.
  • Lay-User Study:

    • Each lay person individually performed the test and interpreted the results. There was no adjudication method among lay users. Each lay user's result was compared to the GC/MS ground truth.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, an MRMC comparative effectiveness study involving AI assistance was NOT done.
    • This device is an in vitro diagnostic (IVD) test (immunochromatographic assay for drug detection in urine), not an AI-powered image analysis or diagnostic support system for human readers.
    • The "Viewers" mentioned in the comparison studies are human laboratory assistants interpreting the visual lines on the test strips/cassettes, not radiologists or clinicians integrating AI outputs. Their performance alone, both individually and in aggregate, is being evaluated against the GC/MS gold standard.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Yes, in essence, the "device" itself is a standalone test.
    • For an IVD like this, the "algorithm" is the biochemical reaction and visual detection mechanism. The testing (precision, cut-off, interference, specificity, specific gravity/pH) directly evaluates the device's performance in detecting the target analytes in urine samples. The "professional user" comparison and "lay user" study involve human interpretation, but the core performance characteristics are intrinsic to the device's design. There isn't a separate "algorithm" being evaluated beyond the physical test kit itself.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • The primary ground truth used for all performance studies (Precision, Cut-off, Interference, Specificity, Specific Gravity/pH Effect, Professional User Comparison, Lay-User Study) was Gas Chromatography/Mass Spectrometry (GC/MS).
    • GC/MS is a highly reliable and recognized analytical method for confirming the presence and concentration of drugs in biological samples and is considered the gold standard for such applications.

8. The sample size for the training set

  • There is no stated training set for this device as it is an immunochromatographic assay, not a machine learning or AI algorithm in the context of typical software device development. The device's performance is based on its biochemical design and manufacturing, not a learned model from a dataset.

9. How the ground truth for the training set was established

  • Not applicable, as there is no training set for this type of IVD device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, stacked on top of each other.

March 4, 2015

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

HEALGEN SCIENTIFIC LLC C/O JOE XIA BUSINESS DIRECTOR 504 EAST DIAMOND AVE. SUITE I GAITHERSBURG MD 20877

Re: K150096

Trade/Device Name: Healgen MDMA (Ecstasy) Test (Strip, Cassette, Cup, Dip Card), Healgen Phencyclidine Test (Strip, Cassette, Cup, Dip Card) Regulation Number: 21 CFR 862.3610 Regulation Name: Methamphetamine test system Regulatory Class: II Product Code: LAF, LCM Dated: January 12, 2015 Received: January 20, 2015

Dear Mr. Xia:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Katherine Serrano -A

For : Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K150096

Device Name Healgen MDMA (Ecstasy) Test (Strip, Cassette, Cup, Dip Card) Healgen Phencyclidine Test (Strip, Cassette, Cup, Dip Card)

Indications for Use (Describe)

Healgen MDMA (Ecstasy) Test is an immunochromatographic assay for the qualitative determination of Methylenedioxymethamphetamine in human urine at a Cut-Off concentration of 500 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result: GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.

Healgen Phencyclidine Test is an immunochromatographic assay for the qualitative determination of Phencyclidine in human urine at a Cut-Off concentration of 25 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

    1. Date: February 26, 2015
    1. Submitter: HEALGEN SCIENTIFIC LLC 5213 Maple St Bellaire, TX 77401
    1. Contact person: Jianqiu Fang HEALGEN SCIENTIFIC LLC 5213 Maple St Bellaire, TX 77401 Telephone: 713-733-8088 Fax: 713-733-8088 Email: bryan@healgen.com
    1. Device Name: Healgen MDMA (Ecstasy) Test (Strip, Cassette, Cup, Dip Card) Healgen Phencyclidine Test (Strip, Cassette, Cup, Dip Card)

Classification:

Product CodeCFR #Panel
LCMThis device has not been classified.
LAF21 CFR, 862.3610 Methamphetamine Test SystemToxicology
    1. Predicate Devices: K052115 First Check Multi Drug Cup 12
    1. Intended Use / Indications for Use

Healgen MDMA (Ecstasy) Test is an immunochromatographic assay for the qualitative determination of Methylenedioxymethamphetamine in human urine at a Cut-Off concentration of 500 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.

{4}------------------------------------------------

Healgen Phencyclidine Test is an immunochromatographic assay for the qualitative determination of Phencyclidine in human urine at a Cut-Off concentration of 25 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.

    1. Device Description
      Healgen MDMA (Ecstasy) Test and Healgen Phencyclidine Test are immunochromatographic assays for Methylenedioxymethamphetamine and Phencyclidine. Each assay test is a lateral flow system for the qualitative detection of Methylenedioxymethamphetamine and Phencyclidine (target analyte) in human urine. The products are in vitro diagnostic devices, which come in the form of: Strips, Cassettes, DipCards, or Cups. Each product contains a Test Device (in one of the four formats), and a package Each test device is sealed with a desiccant in an aluminum pouch. insert.

8. Substantial Equivalence Information

A summary comparison of features of the Healgen MDMA (Ecstasy) Test and Healgen Phencyclidine Test and the predicate device is provided in Table 1 & Table 2.

ItemDevicePredicate -K052115
Intended UseFor the qualitative determination of drugsof abuse in human urine.Same
Drug AnalyteMethylenedioxymethamphetamineSame
MethodologyCompetitive binding, lateral flowimmunochromatographic assays based onthe principle of antigen antibodyimmunochemistry.Same
Specimen TypeHuman UrineSame
Cut-Off Values500 ng/mLSame
IntendedPopulationFor over-the-counter and prescriptionuses.Forover-the-counteruse.
ConfigurationsStrip, Cassette, Cup, Dip CardCup

Table 1: Features Comparison of Healgen MDMA (Ecstasy) Test and the Predicate Device

Table 2: Features Comparison of Healgen Phencyclidine Test and the Predicate Device

{5}------------------------------------------------

ItemDevicePredicate - K052115
Intended UseFor the qualitative determination ofdrugs of abuse in human urine.Same
Drug AnalytePhencyclidineSame
MethodologyCompetitive binding, lateral flowimmunochromatographic assays basedon the principle of antigen antibodyimmunochemistry.Same
Specimen TypeHuman UrineSame
Cut-Off Values25 ng/mLSame
IntendedPopulationFor over-the-counter and prescriptionuses.For over-the-counteruse.
ConfigurationsStrip, Cassette, Cup, Dip CardCup

9. Test Principle

Healgen MDMA (Ecstasy) Test and Healgen Phencyclidine Test are rapid tests for the qualitative detection of Methylenedioxymethamphetamine and Phencyclidine in urine samples. Each assay test is a lateral flow chromatographic immunoassay. During testing, a urine specimen migrates upward by capillary action. If target drugs are present in the urine specimen below its cut-off concentration, it will not saturate the binding sites of its specific antibody (monoclonal mouse antibody) coated on the particles. The antibody-coated particles will then be captured by immobilized drug-conjugate and a visible colored line will show up in the test line region. The will not form in the test line region if the target drug level exceeds its cut-offconcentration because it will saturate all the binding sites of the antibody coated on the particles. A band should form in the control region of the devices regardless of the presence of drug or metabolite in the sample to indicate that the test has been performed properly.

10. Performance Characteristics

    1. Analytical Performance
    • a. Precision

Precision studies were carried out for samples with concentrations of -100% cut-off, -50% cut-off, -25% cut-off, at the cut-off, +25% cut-off, +75% cut-off, +75% cut-off and +100% cut-off. These samples were prepared by spiking drug in negative samples. Each drug concentration was confirmed by GC/MS. All sample aliquots were blind labeled and randomized by the person who prepared samples and did not take part in the sample testing. For each concentration, tests were performed two runs per day for 25 days by three different operators for

{6}------------------------------------------------

each format of devices. Different set of operators tested each format. The results obtained are summarized in the following tables:

MDMA (Ecstasy)

ResultCut-off
Drug-100% Cut-off-75% Cut-off-50% Cut-off-25% Cut-offCut-off+25% Cut-off+50% Cut-off+75% Cut-off+100% Cut-off
Lot:MDMA120100150-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-
Lot:MDMA120100250-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-
Lot:MDMA120100350-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-

Cassette Format

DrugResult-100% Cut-off-75% Cut-off-50% Cut-off-25% Cut-offCut-off+25% Cut-off+50% Cut-off+75% Cut-off+100% Cut-off
Lot:MDMA120100450-/0+50-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-
Lot:MDMA120100550-/0+50-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-
Lot:MDMA120100650-/0+50-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-

Dip Card Format

Result-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Drug
Lot:MDMA120100750-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot:MDMA120100850-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot:MDMA120100950-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-

CUP Format

DrugResult-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Lot:MDMA120101050-/0+50-/0+50-/0+50-/0+30-/20+50+/0-50+/0-50+/0-50+/0-
Lot:MDMA120101150-/0+50-/0+50-/0+50-/0+30-/20+50+/0-50+/0-50+/0-50+/0-
Lot:MDMA120101250-/0+50-/0+50-/0+50-/0+30-/20+50+/0-50+/0-50+/0-50+/0-

Phencyclidine

Strip Format

Result-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Drug
Lot:PCP111100150-/0+50-/0+50-/0+50-/0+20-/30+50+/0-50+/0-50+/0-50+/0-
Lot:PCP111100250-/0+50-/0+50-/0+50-/0+20-/30+50+/0-50+/0-50+/0-50+/0-
Lot:PCP111100350-/0+50-/0+50-/0+50-/0+20-/30+50+/0-50+/0-50+/0-50+/0-

{7}------------------------------------------------

Cassette Format

Result-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
DrugLot:PCP111100450-/0+50-/0+50-/0+50-/0+18-/32+50+/0-50+/0-50+/0-50+/0-
Lot:PCP111100550-/0+50-/0+50-/0+50-/0+18-/32+50+/0-50+/0-50+/0-50+/0-
Lot:PCP111100650-/0+50-/0+50-/0+50-/0+18-/32+50+/0-50+/0-50+/0-50+/0-

Dip Card Format

Result -100% -75%-50% ।-25%+25% '+50% +75% +100%
DrugCut-off Cut-off Cut-off Cut-offCut-offCut-off Cut-off Cut-off Cut-off
Lot:PCP111100750-/0+ 50-/0+ 50-/0+ 50-/0+ 22-/28+ 50+/0- 50+/0- 50+/0- 50+/0- 50+/0-
Lot:PCP111100850-/0+ 50-/0+ 50-/0+ 50-/0+ 22-/28+ 50+/0- 50+/0- 50+/0- 50+/0-
Lot:PCP111100950-/0+ 50-/0+ 50-/0+ 50-/0+ 22-/28+ 50+/0- 50+/0- 50+/0- 50+/0-

CUP Format

Result-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
DrugLot:PCP111101050-/0+50-/0+50-/0+50-/0+16-/34+50+/0-50+/0-50+/0-50+/0-
Lot:PCP111101150-/0+50-/0+50-/0+50-/0+16-/34+50+/0-50+/0-50+/0-50+/0-
Lot:PCP111101250-/0+50-/0+50-/0+50-/0+16-/34+50+/0-50+/0-50+/0-50+/0-
  • b. Linearity
    Not applicable.

  • c. Stability
    The devices are stable at 4-30℃ for 24 months based on the accelerated stability study at 45℃ and real time stability determination at both 4 ℃ and 30℃.

Control materials are not provided with the device. The labeling provides information on how to obtain control materials.

  • d. Cut-off
    A total of 150 samples equally distributed at concentrations of -50% cut-off; -25% cut-off; cut-off; +25% cut-off; +50% cut-off were tested using three different lots of each device by three different operators. Results were all positive at and above +25% cut-off and all negative at and below -25% cut-off for both Methylenedioxymethamphetamine and Phencyclidine. The following cut-off values for the test devices have been verified.
TestCalibratorCut-off(ng/mL)

{8}------------------------------------------------

MDMA (Ecstasy) TestMethylenedioxymethamphetamine500
Phencyclidine TestPhencyclidine25
  • e. Interference
    Potential interfering substances found in human urine of physiological or pathological conditions were added to drug-free urine and target drugs urine with concentration at 25% above cut-off levels. These urine samples were tested using three batches of each device for all formats.

Compounds that showed no interference at a concentration of 100µg/mL are summarized in the following tables. There were no differences observed for different formats.

AcetophenetidinEthyl MorphinePhenelzine
N-AcetylprocainamideEthyl-p-aminobenzoatePhenobarbital
Acetylsalicylic Acid(Aspirin)FenoprofenPhentermine
AminopyrineFurosemidePhenylephrine-L
AmitriptylineGentisic acidPhenylethylamine
AmoxicillinHemoglobinPhenylpropanolamine
AmobarbitalHydralazinePrednisolone Acetate
D-Amphetamine(+/-)-4-HydroxyamphetamineHCLPrednisone
L-AmphetamineHydrochlorothiazideProcaine(Novocaine)
Amphetamine SulfateHydrocodonePromazine
Ampicinine(Ampicillin)HydrocortisonePromethazine
Apomorphinea -Hydroxyhippuric acidPropoxyphene,d-
L-Ascorbic Acidp-HydroxymethamphetaminePropranolol
AspartameIbuprofenPseudoephedrine HCL
AtropineImipramineQuinidine
Benzilic acidIsoxsuprineQuinine
BenzphetamineIsoproterenol-(+/-)Ranitidine(Zantac)
Bezoic AcidKetamineSalicylic Acid
BilirubinLevorphanolSecobarbital
CaffeineLoperamideSerotonin
ChloramphenicolMaprotilineSulfamethazine
ChlordiazepoxideHCLMeprobamateSulindac
ChloroquineMethadoneTemazepam
ChlorothiazideMethoxyphenamine11-Nor-Δ9-Tetrahydrocannabinol

MDMA (Ecstasy)

{9}------------------------------------------------

ChlorpheniramineMethylphenidateTetracycline
ChlorpromazineNalbuphineTetrahydrozoline
CholesterolNalidixic acidThebaine
ClomipramineNaloxone hydrochlorideThiamine
Clonidine hydrochlorideNaltrexone hydrochlorideL-Thyroxine
CodeineNaproxenThioridazineHydrochloride
CortisoneNiacinamideTriamterene
Cotinine(-)NifedipineTriflupromazineHydrochloride
CreatinineNorethindroneTrimethoprim
DeoxyepinephrineNorpropoxypheneTrimipramine
DextromethorphanNoscapineTryptamine
DiazepamOxazepamDL-Tryptophan
DiflunisalOxycodoneTyramine
DigoxinOxymetazolineD/L-Tyrosine
DoxylaminePapaverineUric Acid
Ecgonine methylesterPenicillinVerapamil
R(-)-EpinephrinePentobarbitalZomepirac
ErythromycinPerphenazine
Estrone-3-sulfatePhencyclidine

Phencyclidine

AcetophenetidinEthyl-p-aminobenzoatePhenelzine
N-AcetylprocainamideFenoprofenPhenobarbital
Acetylsalicylic Acid(Aspirin)FurosemidePhentermine
AminopyrineGentisic acidPhenylephrine-L
AmitriptylineHemoglobinPhenylethylamine
AmoxicillinHydralazinePhenylpropanolamine
Amobarbital(+/-)-4-HydroxyamphetamineHCLPrednisolone Acetate
D-AmphetamineHydrochlorothiazidePrednisone
L-AmphetamineHydrocodoneProcaine(Novocaine)
Amphetamine SulfateHydrocortisonePromazine
Ampicinine(Ampicillin)a -Hydroxyhippuric acidPromethazine
Apomorphinep-HydroxymethamphetaminePropoxyphene,d-
L-Ascorbic AcidIbuprofenPropranolol
AspartameImipraminePseudoephedrine HCL
AtropineIsoxsuprineQuinidine
Benzilic acidIsoproterenol-(+/-)Quinine

{10}------------------------------------------------

BenzphetamineKetamineRanitidine(Zantac)
Bezoic AcidLabetalolSalicylic Acid
BilirubinLevorphanolSecobarbital
CaffeineLoperamideSerotonin
ChloramphenicolMaprotilineSulfamethazine
ChlordiazepoxideHCLMeprobamateSulindac
ChloroquineMethadoneTemazepam
ChlorothiazideMethoxyphenamine11-Nor-Δ9-Tetrahydrocannabinol
Chlorpheniramine(+/-)-Methylenedioxyamphetamine(MDA)Tetracycline
ChlorpromazineMethylphenidateTetrahydrozoline
CholesterolNalbuphineThebaine
ClomipramineNalidixic acidThiamine
Clonidine hydrochlorideNaloxone hydrochlorideL-Thyroxine
CodeineNaltrexone hydrochlorideThioridazineHydrochloride
CortisoneNaproxenTriamterene
Cotinine(-)NiacinamideTriflupromazineHydrochloride
CreatinineNifedipineTrimethoprim
DeoxyepinephrineNorethindroneTrimipramine
DextromethorphanNorpropoxypheneTryptamine
DiazepamNoscapineDL-Tryptophan
DiflunisalOxazepamTyramine
DigoxinOxycodoneD/L-Tyrosine
DoxylamineOxymetazolineUric Acid
Ecgonine methylesterPapaverineVerapamil
R(-)-EpinephrinePenicillinZomepirac
ErythromycinPentobarbital
Estrone-3-sulfatePerphenazine

f. Specificity

To test the specificity, drug metabolites and other components that are likely to interfere in urine samples were tested using three batches of each device for all formats. The obtained lowest detectable concentration was used to calculate the cross-reactivity. There were no differences observed for different formats.

MDMAResult%
-----------------

{11}------------------------------------------------

Methylenedioxymethamphetamine(Cut-off=500 ng/mL)Cross-Reactivity
MethylenedioxymethamphetaminePositive at 500 ng/mL100%
3,4-Methylenedioxyamphetamine HCl(MDA)Positive at 3000 ng/mL17%
3,4-Methylenedioxyethylamphetamine(MDEA)Positive at 300 ng/mL167%
d-methamphetaminePositive at 2500 ng/mL20%
d-amphetamine>100,000Not detected
l-amphetamine>100,000Not detected
l-methamphetamine>100,000Not detected
Phencyclidine(Cut-off=25 ng/mL)Result% Cross-Reactivity
PhencyclidinePositive at 25 ng/mL100%
4-Hydroxy PhencyclidinePositive at 90 ng/mL28%

g. Effect of Urine Specific Gravity and Urine pH

To investigate the effect of urine specific gravity and urine samples with of 1.000 to 1.035 specific gravity or urine samples with pH 4 to 9 were spiked with target drugs at 25% below and 25% above cut-off levels. These samples were tested using three batches of each device for all formats. Results were all positive for samples at and above +25% cut-off and all negative for samples at and below -25% Cut-Off. There were no differences observed for different formats.

    1. Comparison Studies
      The method comparison studies for the MDMA (Ecstasy) Test, and the Phencyclidine Test were performed in-house with three different laboratory assistants for each format of the device. Operators ran 80 (40 negative and 40 positive) unaltered clinical samples were blind labeled and compared to GC/MS results. The results are presented in the tables below:
IVALLA VIZA LEZULO V
StripformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)NearCutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan +50%)
Viewer APositive0001324
MDMA (Ecstasy)
------------------

{12}------------------------------------------------

Viewer BNegative10151530
Positive0001224
Viewer CNegative10151540
Positive0001324
Negative10151530

Discordant Results of MDMA (Ecstasy) Strip

ViewerSample NumberGC/MS ResultStrip FormatViewer Results
Viewer A1008ર૦તેNegative
Viewer A1033ર૦રNegative
Viewer A1076507Negative
Viewer B1008ર૦તેNegative
Viewer B1033ર૦રNegative
Viewer B1076507Negative
Viewer B1037211Negative
Viewer C1008ર૦તેNegative
Viewer C1033ર૦રNegative
Viewer C1076507Negative
Cassette formatNegativeLow Negative by GC/MS (less than -50%)Near Cutoff Negative by GC/MS (Between -50% and cut-off)Near Cutoff Positive by GC/MS (Between the cut-off and +50%)High Positive by GC/MS (greater than +50%)
Viewer APositive0001324
Negative10151530
Viewer BPositive0001324
Negative10151530
Viewer CPositive0001424
Negative10151520

Discordant Results of MDMA (Ecstasy) Cassette

Cassette Format
ViewerSample NumberGC/MS ResultViewer Results
Viewer A1008509Negative
Viewer A1033505Negative
Viewer A1076507Negative
Viewer B1033505Negative
Viewer B1037511Negative

{13}------------------------------------------------

ViewerSample NumberGC/MS ResultCassette FormatViewer Results
Viewer B1076507Negative
Viewer C1033505Negative
Viewer C1076507Negative
Cup formatNegativeLow Negative by GC/MS (less than -50%)Near Cutoff Negative by GC/MS (Between -50% and cut-off)Near Cutoff Positive by GC/MS (Between the cut-off and +50%)High Positive by GC/MS (greater than +50%)
Viewer APositive0001324
Negative10151530
Viewer BPositive0001224
Negative10151540
Viewer CPositive0001324
Negative10151530

Discordant Results of MDMA (Ecstasy) Cup

ViewerSample NumberGC/MS ResultCup Format
Viewer Results
Viewer A1008ર09Negative
Viewer A1033ર૦રNegative
Viewer A1076507Negative
Viewer B1008ર૦તેNegative
Viewer B1033૨૦૨Negative
Viewer B1076507Negative
Viewer B1037211Negative
Viewer C1008ર૦તેNegative
Viewer C1033ર૦રNegative
Viewer C1076507Negative
Dip CardformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Viewer APositive0001424

{14}------------------------------------------------

Negative10151520
Viewer BPositive0001524
Negative10151510
Viewer CPositive0001324
Negative10151530

Discordant Results of MDMA (Ecstasy) Dip Card

ViewerSample NumberGC/MS ResultDip Card FormatViewer Results
Viewer A1033505Negative
Viewer A1076507Negative
Viewer B1033505Negative
Viewer C1008509Negative
Viewer C1033505Negative
Viewer C1076507Negative
Phencyclidine
Strip formatNegativeLow Negative by GC/MS (less than -50%)Near Cutoff Negative by GC/MS (Between -50% and cut-off)Near Cutoff Positive by GC/MS (Between the cut-off and +50%)High Positive by GC/MS (greater than +50%)
Viewer APositive0001324
Negative10151530
Viewer BPositive0001324
Negative10151530
Viewer CPositive0001324
Negative10151530

Discordant Results of Phencyclidine Strip

ViewerSample NumberGC/MS ResultStrip FormatViewer Results
Viewer A77426Negative
Viewer A72125Negative
Viewer A77927Negative
Viewer B77426Negative
Viewer B72125Negative
Viewer B70227Negative
Viewer C72125Negative
Viewer C70427Negative

{15}------------------------------------------------

ViewerSample NumberGC/MS ResultStrip Format Viewer Results
Viewer C77927Negative
Cassette formatNegativeLow Negative by GC/MS (less than -50%)Near Cutoff Negative by GC/MS (Between -50% and cut-off)Near Cutoff Positive by GC/MS (Between the cut-off and +50%)High Positive by GC/MS (greater than +50%)
Viewer APositive0001424
Negative10151520
Viewer BPositive0001324
Negative10151530
Viewer CPositive0001224
Negative10151540

Discordant Results of Phencyclidine Cassette

ViewerSample NumberGC/MS ResultCassette Format Viewer Results
Viewer A77426Negative
Viewer A72125Negative
Viewer B77426Negative
Viewer B72125Negative
Viewer B70227Negative
Viewer C77426Negative
Viewer C72125Negative
Viewer C70427Negative
Viewer C77927Negative
Dip CardformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Viewer APositive0001324
Negative10151530

{16}------------------------------------------------

Viewer BPositive0001324
Negative10151530
Viewer CPositive0001224
Negative10151540
ViewerSample NumberGC/MS ResultDip Card Format Viewer Results
Viewer A77426Negative
Viewer A72125Negative
Viewer A77927Negative
Viewer B77426Negative
Viewer B72125Negative
Viewer B70227Negative
Viewer C77426Negative
Viewer C72125Negative
Viewer C77927Negative
Viewer C70427Negative

Discordant Results of Phencyclidine Dip Card

CupformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Viewer APositive0001324
Negative10151530
Viewer BPositive0001324
Negative10151530
Viewer CPositive0001324
Negative10151530

Discordant Results of Phencyclidine Cup

ViewerSample NumberGC/MS ResultCup Format Viewer Results
Viewer A77426Negative
Viewer A72125Negative
Viewer A77927Negative
Viewer B77426Negative
Viewer B72125Negative

{17}------------------------------------------------

ViewerSample NumberGC/MS ResultCup FormatViewer Results
Viewer B70227Negative
Viewer C72125Negative
Viewer C77927Negative
Viewer C70427Negative

Lay-user study

A lay user study was performed at three intended user sites with 557 lay persons testing the MDMA(Ecstasy) devices and another set of 555 persons testing the Phencyclidine devices. Total of 1112 individuals performed the study. A total of 230 females and 327 males tested the MDMA(Ecstasy) samples, and 226 females and 329 males tested the Phencyclidine samples. They had diverse educational and professional backgrounds and ranged in age from 21 to > 50 years. Urine samples were prepared at the following concentrations; negative, +/-75%, +/-50%, +/-25% of the cutoff by spiking drugs into drug free-pooled urine specimens. The concentrations of the samples were confirmed by GC/MS. Each sample was aliquoted into individual containers and blind-labeled. Each participant was provided with the package insert, 1 blind labeled sample and a device. The results are summarized below.

NumberofsamplesMDMA Concentration byGC/MS(ng/mL)Lay person resultsThe
% of CutoffNo. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100% Cutoff200020100%
-75% Cutoff20125020100%
-50% Cutoff19250019100%
-25% Cutoff2037511995%
+25% Cutoff2062519195%
+50% Cutoff20750200100%
+75% Cutoff20875200100%

Comparison between GC/MS and Lay Person Results (MDMA(Ecstasy) Strip)

Comparison between GC/MS and Lay Person Results (MDMA(Ecstasy) Cassette)

NumberMDMA Concentration byGC/MS(ng/mL)Lay person resultsThe
% of CutoffofsamplesNo. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100%Cutoff200020100%
-75%Cutoff19125019100%
-50% Cutoff20250020100%
-25% Cutoff20375119તે તે જે જ જીરી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં લોકોનો મુખ્ય
+25% Cutoff20ર્ભ (૧૯૮૮)191તે તે જેની જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં મુખ્યત્વે ખેત-ઉત

{18}------------------------------------------------

+50% Cutoff20750200100%
+75% Cutoff20875200100%

Comparison between GC/MS and Lay Person Results (MDMA(Ecstasy) DipCard)

NumberMDMA Concentration byLay person resultsThe
% of CutoffofsamplesGC/MS(ng/mL)No. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100%Cutoff200020100%
-75%Cutoff20125020100%
-50% Cutoff20250020100%
-25% Cutoff20375119તે તે જેની જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામમાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો
+25% Cutoff20ર્ભ (૧૯૮૮)191તે તે જેની જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામમાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો
+50% Cutoff20750200100%
+75% Cutoff20875200100%

Comparison between GC/MS and Lay Person Results (MDMA(Ecstasy) Cup)

% of CutoffNumberofsamplesMDMA Concentration byGC/MS(ng/mL)Lay person resultsThe
No. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100%Cutoff200020100%
-75%Cutoff20125020100%
-50% Cutoff20250020100%
-25% Cutoff20375119તે તે જે જ જીરી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં લોકોનો મુખ્ય
+25% Cutoff20ર્ભ (૧૯૮૮)191તે તે જે જ જીરી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં લોકોનો મુખ્ય
+50% Cutoff19750190100%
+75% Cutoff20875200100%

Comparison between GC/MS and Lay Person Results (Phencyclidine Strip)

NumberPhencyclidineConcentration by GC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
% of CutoffofsamplesNo. ofPositiveNo. ofNegative
-100%Cutoff200020100%
-75%Cutoff196019100%
-50% Cutoff2012.5020100%
-25% Cutoff201911995%
+25% Cutoff203119195%
+50% Cutoff2037.5200100%
+75% Cutoff2044200100%

{19}------------------------------------------------

NumberofsamplesPhencyclidineConcentration by GC/MS(ng/mL)Lay person resultsThe
% of CutoffNo. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100%Cutoff200020100%
-75%Cutoff206020100%
-50% Cutoff1912.5019100%
-25% Cutoff201911995%
+25% Cutoff203119195%
+50% Cutoff2037.5200100%
+75% Cutoff2044200100%

Comparison between GC/MS and Lay Person Results (Phencyclidine Cassette)

Comparison between GC/MS and Lay Person Results (Phencyclidine DipCard)

NumberofsamplesPhencyclidineConcentration by GC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
% of CutoffofsamplesPhencyclidineConcentration by GC/MS(ng/mL)No. ofPositiveNo. ofNegativeThepercentage ofcorrect results(%)
-100%Cutoff200020100%
-75%Cutoff206020100%
-50% Cutoff2012.5020100%
-25% Cutoff201921890%
+25% Cutoff203118290%
+50% Cutoff2037.5200100%
+75% Cutoff1944190100%

Comparison between GC/MS and Lay Person Results (Phencyclidine Cup)

% of CutoffNumberofsamplesPhencyclidineConcentration by GC/MS(ng/mL)Lay person resultsThe
No. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100%Cutoff200020100%
-75%Cutoff206020100%
-50% Cutoff1912.5019100%
-25% Cutoff201921890%
+25% Cutoff203119195%
+50% Cutoff2037.5200100%
+75% Cutoff1944190100%

{20}------------------------------------------------

Lay-users were also given surveys on the ease of understanding the package insert instructions. All lay users indicated that the device instructions can be easily followed. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.

    1. Clinical Studies Not applicable.
  • 11.Conclusion

Based on the test principle and acceptable performance characteristics including precision, cut-off, interference, specificity and method comparison of the devices, it's concluded that the Healgen MDMA(Ecstasy) Test, and Healgen Phencyclidine Test are substantially equivalent to the predicate.

§ 862.3610 Methamphetamine test system.

(a)
Identification. A methamphetamine test system is a device intended to measure methamphetamine, a central nervous system stimulating drug, in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of methamphetamine use or overdose.(b)
Classification. Class II (special controls). A methamphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).